DPK TECHOLYSER, GAT NO. 188, BEHIND S.P.AUTO PARTS, JYOTIBA NAGAR, TALWADE, PUNE -412 114
DPK TECHOLYSER
Gas Sampling Probe SP-01
FEATURES
� Rugged & Modular Construction
� Heated and Non-heated versions available
� Easy filter replacement
� Low maintenance
� Stainless Steel enclosure
� Microprocessor based temperature controller
� Settable Low or High temperature contact
� Status & Fault Alarms on Front Display
DPK TECHOLYSER, GAT NO. 188, BEHIND S.P.AUTO PARTS, JYOTIBA NAGAR, TALWADE, PUNE -412 114
GENERAL The Model SP-01 of heated gas sample probe is designed for continuous use with extractive type Sampling Systems even
where the sample contains dust and Aerosols. Water vapor and high dew-point corrosive gases must be kept above the
dew-point to prevent corrosion and sample degradation prior to hot analysis or sample conditioning. The probe is
available in several versions which may be custom configured to meet user's specific requirements. The probe
incorporates a non-corrosive heated, replaceable sintered SS filter element with a filter porosity of 5 microns. The filter
element is mounted in an electrically heated and thermally isolated Stainless steel housing, which is installed in a
Weather protection housing. The temperature regulation is done by a maintenance free, fully electronic Temperature
Controller with under / over temperature alarm. The heated line can be easily connected to the probe by a suitable SS
Bulkhead fitting.
TECHNOLOGY
The complete unit, consisting of the heated filter Head, temperature controller, mounting flange and Probe assembly can
be mounted anywhere horizontally or vertically. The probe can be connected with a Standard 3 inch, 150# flange
directly. The Flange material is stainless steel. When mounted horizontally, the Probe should be inclined at an Angle of
5° to 15° to allow any condensate formed to drain away from the filter head. A broad range of different materials for In-
Situ probe tubes, and large In-Situ – pre Filters make the SP-01 versatile in application.
FEATURES
A temperature alarm contact signalizes fault and over / under Temperature. An over temperature protection avoids
overheating. The filters can be replaced easily without any tools and without disconnecting the heated line.
TECHANICAL DATA
DPK TECHOLYSER, GAT NO. 188, BEHIND S.P.AUTO PARTS, JYOTIBA NAGAR, TALWADE, PUNE -412 114
SPECIFICATIONS
Filter element SS316; pore size 5 µm;
50 x 40 x 135 mm
Operating pressure 50 ... 150 kPa abs.
Mounting flange
Option
3” ANSI flange, SS316, 150 lb
Variations possible.
Mounting angle Recommended 5°-15° inclined to
allow condensate drainage
Sample gas inlet 3/4“ NPT female thread
Sample gas outlet 1/4“ NPT female thread
Test gas connection Optional
Gas wetted parts SS316, Silicon
Temperature adjustment
Option
5 ... 200°C
5 ... 315°C
Factory setting 150° …. 180°C
Ambient temperature -20°C ... +60°C
Power consumption Approx. 300 Watts
Heater element Approx. 250 Watts
Temperature sensor PT100
Power supply
Option
230VAC / 50 Hz or
110VAC / 50 Hz
Junction box Optional
Electrical connection 3C x 2.5 sq. mm
Alarm contact 230VAC, 5 A
Protection class IP65 with junction box
Weight Approx. 11 kg
Dimensions As per variations