Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
Nobuo Suga
Precision Length
Measurement to
Improve Quality and Productivity
in Manufacturing
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
A B
Comparison
Measurement
“Known” “Unknown”
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
Artifact Standard
1: End Standard
2: Line Standard Gage Blocks
Glass Scales
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
Carl E. Johansson (1864-1943)
“The Master of Measurement,” also known as “the most precise man”reached 1µm at the turn of the last century
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
Johansson’s Gage Block Set patented in 1911
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
µ
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
633 633 633 633
Gage Block
L End Standard
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
633 633 633 633
L
End Standard
Zero Cera Block
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
633 633 633 633
L
End Standard
Zero Cera Block
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
633 633 633 633
L
End Standard
Zero Cera Block
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
633 nm
Wavelength of Helium-Neon Laser (and Fringe Lines)
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
Test Model No. 1 632 991 382 fm
No. 2 632 991 412 fm
No. 3 632 991 379 fm
No. 4 632 991 331 fm
No. 5 632 991 388 fm
No. 6 632 991 381 fm
633 633 633 633 Wavelength of randomly selected
Laser Interferometers tested by Mitutoyo (was accurate within one picometer)
Nano
Pico Femto
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
Is traceability to NIST required?
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
Measurement Uncertainty (U) @ NIST, Washington, DC
50 mm End Standard = 31 nm
50 mm Line Standard = 8 nm
Temperature controlled = 0.005 C
°
In comparison, JIS standard “0.5” Grade = 0.5 C
°
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
End Standard
G a g e B l o c k
0.577L
(Airy Point)
µ
Narrow Range accuracy calibrated at NIST for 50 mm was 0.1 m
50 mm
µ
Sealed Chamber: Temperature kept at: 0.005 C
°
Calibrated at NIST
0.05 C °
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
m International System of Units (SI)
299 792 4xx m/s
299 910 xxx m/s
C = 299 792 458 m/s
1879 Michelson
1953 Michelson
1983 CIPM
(C = Speed of light in vacuum)
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
s
m Unit of Length
Unit of Time
Accuracy of 1 Second: NIST–F2 Model (under development now)
0.000 000 000 000 000 001 1 Second error =Ten to the power of negative 18
SI
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
Standard Temperature
23 25 20
°
C °
°
C
C
68 °
F At the turn of the last century, Johansson
calibrated his gage blocks at 20 degree C,
which was then the average machine shops
temperature in northern Europe.
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
Standard Temperature
20
0.01
(in degree C)
29.76 Triple Point of
water Standard
Temperature
Melting point
of Gallium
International Temperature Scale (IT-90)
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
Room Temperature
20
5
(in degree C)
35
The Definition of so-called “Room Temperature”
in accordance with JIS Standard Z 8703
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
G a g e B l o c k
0.577L
(Airy Point)
Sir George Airy (1801-1892) Astronomer at Greenwich Observation Laboratory
End Standard
ZERO CERA BLOCK
“Step Gage” (Check Master) to calibrate CMMs
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
Temperature is raised and lowered to study the effect of expansion
“The most accurate man is the man with the most accurate thermometer.” Ted Doiron, Metrologist, NIST
Mitutoyo Institute of Metrology / Mitutoyo America Corporation / Chicago USA
Artifact Standard NMC Seminar in Singapore
Thank you,