DOSIMETRY IN DIAGNOSTIC RADIOLOGY Chien-chuan, Andersen Chen Medical Physicist

DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

Page 1: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

DOSIMETRY IN DIAGNOSTIC RADIOLOGY

Chien-chuan, Andersen Chen

Medical Physicist

Page 2: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

In the beginning you should know

• Radiation units

• Absorbed dose (Gy), Equivalent dose (Sv), Effective

dose (Sv), and measurement

• Physics of Radiology

• X-ray generation and properties (kV, mAs, beam quality),

and interactions with matters

• Fundamental Principles of Medical Imaging

• Digital radiography, Computed Tomography, and their

imaging techniques in clinical examinations

Page 3: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Exercise

• An exposure is measured and the readout is 1 R (roentgen),

please answer the following questions:

1. What is the absorbed dose in SI unit ?

2. What is the equivalent dose when the source of

radiation is X-ray?

3. What is the equivalent dose when the source of

radiation is alpha particles?

Page 4: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Modern imaging modalities in radiology department

• Plain X-ray imaging (including Mammography)

• Computed Radiography

• Digital Radiography

• Fluoroscopy

• Computed Tomography

• Ultrasonography

• Magnetic Resonance Imaging

Page 5: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Radiation exposure and protection • Global averaged natural background radiation: 2.4 mSv/year

• India, Brazil, Iran have higher: 5-15 mSv/year

• Opacification of the lens of the eye may occur following exposure to significantly lower doses of ionizing radiation (perhaps without a threshold, previous threshold: 2-5 Gy)

• ICRP 2011, Occupational dose limits to lens: 150 mGy/year • Whole-body effective dose limits for workers:

• 20 mSv/year (averaged over 5-year period; 100 mSv in 5 years, no single year exceeding 50 mSv)

• Use protective tools less than 1 mSv/year

Page 6: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

NCRP Report 168, 2011

• Potential risks at varying dose level :

• <0.1 mSv = negligible

• 0.1-1 mSv = minimal

• 1-10 mSv = minor

• 10-100 mSv = low

• >100 mSv = acceptable in context of the expected benefit

Page 7: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 8: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 9: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

• Deterministic:

• visible, documented, confirmed within a relative short time

• Skin erythema, hair loss, cataract, infertility, circulatory disease

• Stochastic:

• estimated, years or decades to manifest

• Cancer, genetic effects

Health effects of ionizing radiation

Page 10: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 11: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

• Have thresholds that are typically quite high

• Skin erythema

• Hair loss

• Cataracts (even in low doses of radiation)

• 5 Sv for protracted exposures

• 2 Sv for acute exposures

• Epidemiological evidence suggesting thresholds

(equivalent dose):

• Lens of eye: 0.5 Sv

• Circulatory system: 0.5 Sv

Tissue reactions (Deterministic effects)

Page 12: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 13: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

• Detriment-adjusted nominal risk coefficient at low dose rate:

• Cancer – 5.5 % per Sv

• Genetic effects – 0.2 % per Sv (non-human species)

• Cancer risks are estimated on the basis of probability

• Organ dose > 100 mGy carcinogenic effects

• Stochastic risks have no threshold

Stochastic effects

Tissue weighting factor of gonads: 0.2 0.08 (ICRP, 2007)

1 chest CT scan ~ 8 mSv 20 mGy to breast

5 ~ 15 CT scans carcinogenic effects

Page 14: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

http://www.rerf.jp/radefx/late_e/cancrisk.html

For the average radiation exposure of survivors within 2,500 meters (about 0.2 Gy), the increase is

about 10% above normal age-specific rates. For a dose of 1.0 Gy, the corresponding cancer excess

is about 50% (relative risk = 1.5)

The excess number of solid cancers is estimated as 848 (10.7%)

Stochastic effects Hiroshima and Nagasaki

Page 15: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

The dose-response relationship appears to be linear, without any apparent threshold below which

effects may not occur

Page 16: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

The probability that an A-bomb survivor will have a cancer caused by A-bomb radiation (excess

lifetime risk) depends on the (1)dose received, (2)age at exposure, and (3)sex.

Other analyses (not shown) indicate that females have somewhat higher risks of cancer from

radiation exposure than males do.

Page 17: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

DOSIMETRY IN DIAGNOSTIC RADIOLOGY

Exposure Indicator (CR, DR)

Dose-Area Product (RF)

Dose Index (CR, DR, RF)

Average Glandular Dose (MG)

Computed Tomography Dose Index (CT)

Page 18: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Exposure to pixel value

Page 19: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Exposure indicator

• Modality: plain x-ray exams (CR or DR)

• Terms:

Pixel value = 2000 + 1000 × 𝑙𝑜𝑔10(𝑒𝑥𝑝𝑜𝑠𝑢𝑟𝑒)

i.e. Carestream (for plain radiography)

exposure is measured in mR Coded value (CV)

Page 20: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

System calibration for dose indicator • Carestream CR/DR system:

• a heavily filtered (an additional 0.5 mm copper and 1 mm aluminum) 80 kVp x-ray beam

• Choose proper mAs approximate to have 1 mR exposure

• 15 minutes delay after exposure

• Read the imaging plate

• Acceptable range: +/- 100 (diff. of calculated and displayed EIs)

EI =1000 log(E) + 2000 (or +1000 for MG)

Page 21: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Exercise

• During annual QA test for a Carestream CR system, the

exposure dose measured at 80 kVp, 25 mAs is 1.25 mR,

and the displayed EI is 1991. Does this CR system need a

new calibration for EI ?

Page 22: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

System calibration for dose indicator • Fujijfilm CR/DR system:

• a filtered (~ 3 mm Al HVL) 80 kVp x-ray beam

• Choose proper mAs approximate to have 1 mR exposure

• 10 minutes delay after exposure

• Read the imaging plate

• Acceptable range: +/- 20%

𝑆 = 4 × 10(4−𝑆𝑘)

L = 1023 x (S1-S2) / (Q1-Q2)

Raw data is recorded in 12-bit pixel depth

Image data is recorded in 10-bit pixel depth

𝑷𝒊𝒙𝒆𝒍 𝒗𝒂𝒍𝒖𝒆 = 𝟏𝟎𝟐𝟒

𝑳× 𝒍𝒐𝒈𝟏𝟎(𝒌 × 𝑺 × 𝑬) + 𝟓𝟏𝟏

𝑒𝑥𝑝𝑜𝑠𝑢𝑟𝑒 ≅ 200/𝑆

In automatic processing mode:

S is inverse proportional to exposure

Page 23: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Exercise

• A Sk value of 2.30 corresponds to an incident exposure of

2.0 mR in automatic processing mode.

1. What is the S number for this image ?

2. What is the pixel value of 2.0 mR ?

3. What is the k value in this system ?

4. What is the exposure (mR) when pixel value = 1023

while L = 1 ?

Page 24: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Exercise

• For a given S number, describe the effects when the L

number is doubled manually.

Page 25: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

𝐼𝑔𝑀 = 1.9607 + log𝐸 𝜇𝐺𝑦

2.5+ log (

𝑆𝑝𝑒𝑒𝑑𝐶𝑙𝑎𝑠𝑠

400)

Page 26: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Dose area product (kerma area product)

DAP is measured in Gy∙cm2

Page 27: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 28: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 29: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 30: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 31: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Deviation index

𝐷𝑒𝑣𝑖𝑎𝑡𝑖𝑜𝑛 𝐼𝑛𝑑𝑒𝑥 = 10 × 𝑙𝑜𝑔10

𝐾𝐼𝑁𝐷

𝐾𝑇𝐺𝑇

Dev Index mGy cm2

K3x 4.8 30

K2x 3.0 20

KTGT 1.0 10

K1/2 -3.0 5

K1/4 -6.0 2.5

Page 32: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

AVERAGE (MEAN) GLANDULAR DOSE

Procedures:

1. Determine the beam quality (HVL)

2. Make at least 4 exposures for ACR

phantom using clinical mode

3. Record the readouts from dose

meter

4. Calculate the mean value of the

exposures (readouts)

Page 33: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 34: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Conversion factor (K) table for Mo/Mo

𝐴𝐺𝐷 = 𝐸𝑥𝑝𝑜𝑠𝑢𝑟𝑒 × 𝐾

AGD < 300 mrad or 3.0 mGy

HVL of 26 kVp = 0.34 mmAl

Find K = 172

AGD using 26 kVp

= 172 x 686.6 / 1000 = 118.1 mrad

Page 35: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Exercise

• Using previous measurements in the AGD test, answer

following questions:

1. What is the AGD when using 28 kVp ?

2. Why the AGDs in different kVp is different ?

3. Describe the pros-and-cons in different kVp ?

Page 36: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

CTDI (CT dose index) 32 cm 16 cm

15

cm

Procedures:

1. Find the routine CT scan protocols (i.e adult abdomen, adult head, …) and the scan parameters

2. Setup the phantom and align it with the bore center of the gantry

3. Make at least 5 exposures with axial scan mode (change the ion chamber positions – center

and 4 peripherals) and using the parameters found in step 1

4. Record the readouts

Page 37: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Computed Tomography Dose Index • Weighted CTDI : CTDIw

CTDI100,center

CTDI100, P1

CTDI100, P2

CTDI100, P3

CTDI100, P4

CTDI100,edge = 4

CTDI100, P1+P2+P3+P4

Page 38: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Calculation for each readout

- f=8.7 mGy/R (if readout in R)

C: calibration factor for electrometer (1.0-2.0)

E: average measured value

L: active length of pencil ion chamber (100 mm or 160 mm)

N: actual number of data channels (axial)

T: nominal slice width (axial)

Example:

• Multi-slice scanner

• 120 kVp, 400 mA, 0.8-s scan, 4×2.5 mode

• Reading: 540 mR

• Calculation:

100CTDIf C E L

N T

100

8.7 1.0 0.54 100CTDI

4 2.5

=47 mGy

Page 39: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

DOSE CALCULATION

pcw CTDI3

2CTDI

3

1CTDI

wvol

CTDICTDI

Pitch

volDLP=CTDI total scan length

DLPE k

C P1

P2

P4

P3

(mGy)

(mGy)

(mGy cm)

(mSv)

Page 40: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

COMPUTED TOMOGRAPHY DOSE INDEX

Volume CTDI: CTDIvol

In axial scan : 4 mm

1 mm

Pitch = 4 ÷ (4 + 1) = 0.8 Example:

Slice thickness = 4 mm

Inter-slice gap = 1 mm

Page 41: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Exercise

• Please calculate the CTDIw, and the CTDIvol using the data

shown below.

(the scan pitch are: brain=0.875, abd=1.375)

Page 42: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

DLP – dose length product

DLPE k (mSv)

Page 43: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Exercise

• A CT examination on an adult brain is performed, the

CTDIvol and scan length of this examination are 55 mGy

and 15 cm in the dose report respectively. What is the

relative effective dose in this CT scan ?

Page 44: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

DRL – diagnostic reference level • DRLs should be set for representative examinations or

procedures performed in the local area, country or region where they are applied.

• NDRL (National DRL): set on the basis of wide scale surveys of the median doses representing typical practice for a patient group (e.g. adults or children of different sizes) at a range of representative healthcare facilities for a specific type of examination or procedure.

• LDRL (Local DRL): represent the typical local practice at a single large centre or group of healthcare facilities, set as the third quartile of the median doses determined from samples of patients in the different healthcare facilities of the group.

Page 45: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Set DRLs • Define DRL scale (national

or local)

• Define patient group (adult/child, body size/weight)

• Collect dose report data (Dose index, DAP, CTDI, AGD, …)

• Find the 3rd quarter of the dataset DRL

• Repeat setup periodically

Dose length product (DLP) distribution for chest

CT and local diagnostic reference level(LDRL).

References: Department of Radiology,La Rabta

Hospital,Tunis,Tunisia 2015.

Page 46: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 47: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

National DRLs

• CT:

• adult and pediatric

• General radiology and Fluoroscopy:

• individual radiographs on adult patients

• diagnostic examinations on adult patients

• interventional procedures on adult patients

• diagnostic examinations on paediatric patients

• Dental radiography

Page 48: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 49: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 50: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 51: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

U.S. Diagnostic Reference Levels and Achievable Doses for 10 Adult CT

Examinations

Page 52: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 53: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Exercise

• Discuss the pros-and-cons of NDRL and LDRL.

Dose length product (DLP) distribution for abdomen and pelvis CT (AP) and local diagnostic

reference level(LDRL). References: Department of Radiology, Rabta Hospital,Tunis,Tunisia 2015.

Page 54: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 55: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 56: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Introduction-1

• Terms

• Shall indicates a recommendation that is necessary to meet the currently accepted standards of radiation protection.

• Should indicates an advisory recommendation that is to be applied when practicable or practical.

• DT×WR=HT

• HT×WT=E

Page 57: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Introduction-2

• Controlled area and uncontrolled area

• Recommendation for controlled areas—

Shielding design goal (P) (in air kerma):

0.1 mGy/week (5 mGy/y or 2.5 μGy/h)

• Recommendation for uncontrolled areas—

Shielding design goal (P) (in air kerma):

0.02 mGy/week (1 mGy/y or 0.5 μGy/h)

Shielding criteria in Taiwan (from Atomic Energy Council website)

Page 58: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Concepts and terminology

• Shielding design goals (P)

• Distance to the occupied area

• Occupancy factors

• Workload and workload distribution

• Use factor

• Primary barriers

• Secondary barriers

Page 59: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Distance to the occupied area (d)

• The source to the nearest likely approach of the

sensitive organs of a person to the barrier

• For a wall this may be assumed to be not <0.3 m.

• For a source located above potentially occupied

spaces, the sensitive organs of the person below

can be assumed to be not >1.7 m above the lower

floor, while for ceiling transmission the distance of

at least 0.5 m above the floor of the room above is

generally reasonable.

Page 60: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Elevation (left) and plan (right) views of a

representative radiographic (or radiographic and

fluoroscopic) room.

Points A, B, C, D and E represent a distance of 0.3 m

from the respective walls. Point F is 1.7 m above the

floor below. Point G is taken at 0.5 m above the floor

of the room above.

1.7 m

0.5 m

0.3 m

Page 61: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Occupancy factors (T)

• The average fraction of time that the maximally

exposed individual is present while the x-ray

beam is on.

• the fraction of the working hours in the week

that a given person would occupy the area

Page 62: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 63: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Workload (W) and workload distribution

• The time integral of the x-ray tube current over a specified period

• Milliampere-minutes (mA-min)

• The total workload per week (Wtot)

Wtot = N Wnorm

• The normalized workload (Wnorm), the average workload per patient

• The average number of patients per week (N)

Page 64: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 65: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Use factor (U)

• The fraction of the primary beam workload that

is directed toward a given primary barrier.

• Type of radiation installation

Page 66: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 67: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Primary barriers • Unshielded Primary Air Kerma

• The weekly unshielded primary air kerma

[KP(0)] in the occupied area due to N

patients examined per week in the room

• the unshielded primary air kerma per

patient at 1 m

• Preshielding

2

1

)0(p

p

pd

UNKK

1

pK

Page 68: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Pre-shielding

Page 69: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Secondary barriers

• Leakage radiation

• Scatter radiation

• Total contribution from secondary radiation

• The air kerma from unshielded secondary

radiation [Ksec(0)] at a distance dsec for N

patients

2

sec

1

secsec )0(

d

NKK

Page 70: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Shielding for primary barriers

• The barrier transmission factor (Bp)

• The structural barrier thickness (xbarrier)

UNK

d

T

PxxB

p

p

prebarrierp 1

2

)(

pre

p

p

barrier xPd

NTUK

x

1

ln1

2

1

Page 71: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 72: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Example: A dedicated chest unit that is used to image 300

patients per week. Assume that the x-ray beam in

this room is always directed horizontally toward a

wall-mounted chest-bucky image receptor of area

1,535 cm2 (at 1.83 m SID).

Let the room behind the image-receptor wall be a

fully-occupied, uncontrolled office, so that P/T = 0.02

mGy week–1. Assume a primary distance dP = 3 m.

The wall on which the image receptor is mounted

will therefore serve as a primary barrier to the x-ray

beam with a use factor U = 1.

UNK

d

T

PxxB

p

p

prebarrierp 1

2

)(

Page 73: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

The transmission required for the primary barriers is therefore:

Page 74: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

B = 5 x 10-4

Thickness of lead required

Page 75: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

pre

p

p

barrier xPd

NTUK

x

1

ln1

2

1

Page 76: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Shielding for secondary barriers

• The barrier transmission factor Bsec(xbarrier)

• The thickness of secondary barrier

NK

d

T

PxB barrier 1

sec

2

secsec )(

1

ln1

1

sec

1

sec

Pd

NTUK

xbarrier

Page 77: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 78: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 79: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 80: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Beam quality (HVL) measurement

collimator

4.5 cm

4 cm

Al sheet

paddle

radiation detector

Page 81: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Exercise

• According to the results shown below, what is the HVL of

24 kVp.

Page 82: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),

Exercise

• Please prove the formula of HVL.

xn is the thickness

Yn is the recorded exposure

Page 83: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 84: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 85: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 86: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 87: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),
Page 88: DOSIMETRY IN DIAGNOSTIC RADIOLOGY › TMPweb › images › news › 20171220110425.pdf · 2017-12-20 · In the beginning you should know • Radiation units • Absorbed dose (Gy),