DOPPLER ECHOCARDIOGRAPHY-1
BASIC PHYSICS,PULSE WAVE AND CONTINUOS WAVE DOPPLER
DR PRADEEP SREEKUMAR
• Sound is a mechanical vibration transmitted through an elastic medium
• Ultrasound-portion of sound spectrum having frequency greater than 20,000 cycles /sec
• Use of ultrasound to study the structure and function of heart and great vessels-echocardiography
• Advantages of ultrasoundCan be directed as a beam and focussedObeys laws of reflection and refractionProduce longitudnal waves
Generation Of An Ultrasound Image
Machines
There are 5 basic components of an ultrasound scanner that are required for generation, display and storage of an ultrasound image.
1. Pulse generator - applies high amplitude voltage to energize the crystals
2. Transducer - converts electrical energy to mechanical (ultrasound) energy and vice versa
3. Receiver - detects and amplifies weak signals
4. Display - displays ultrasound signals in a variety of modes
5. Memory - stores video display
• Depicted as sine wave-peaks and troughs• One cylce=one compression + one rarefaction• Distance between 2 similar points represent
wavelength• 0.15 to 1.5 mm in soft tissue• Frequency- number of wavelengths per unit time• V=f X λ(v=velocity,f =frequency, λ is wavelength)
• Velocity of sound=1540 m/sec in soft tissue
• Wavelength=1.54/f
• AmplitudeMeasure of strength of the sound waveIndicated by height of sine wave above and below baseline
• Higher the frequency greater the resolution
• Higher frequency,lesser the penetration
• Loss of ultrasound as it propogates through a medium is called attenuation
PRICIPLES OF PEIZO ELECTRIC CRYSTALS
The charges in a piezoelectric crystal are exactly balanced, even if they're not symmetrically arranged.
The effects of the charges exactly cancel out, leaving no net charge on the crystal faces
the electric dipole moments—vector lines separating opposite charges—exactly cancel one another out.
If you squeeze the crystal , you force the charges out of balance.
• Now the effects of the charges (their dipole moments) no longer cancel one another out and net positive and negative charges appear on opposite crystal faces.
• By squeezing the crystal, voltage is produced across its opposite faces- piezoelectricity
• The piezoelectric effect was discovered in 1880 by two French physicists, brothers Pierre and Paul-Jacques Curie, in crystals of quartz, tourmaline, and Rochelle salt (potassium sodium tartrate). They took the name from the Greek work piezein, which means "to press."
• The phenomenon of generation of a voltage under mechanical stress is referred to as the direct piezoelectric effect
• mechanical strain produced in the crystal under electric stress is called the converse piezoelectric effect.
• Ferro electrics,barium tianate,lead zirconate titanate are used as peizo electric crystals.
• Dampening material-shortens the ringing responseAlso absorbs backward and laterally transmitted acoustic energy
• Frequency emitted by transducer is directly proportional to propagation speed within crystal and inversely related to thickness
• Important feature of ultrasound is ability to direct or focus the beam as it leaves the transducer
• Proximal cylindrical and distally divergent• Proximal zone –Fresnel zone• Divergent field is called Fraunhofer zone• Imaging is optimal in near field• Decreasing wavelength or increasing transducer size
increase near field
Haemo”dynamics”
• Blood flow is a complex phenomenon• Not a uniform liquid• Flow pulsatile• Vessel walls are elastic
Properties of Blood
• Density-mass of blood per unit volume• Measure of resistance to accelaration• Greater the density,greater the resistance to
flow• Viscosity:resistance to flow offered by fluid in
motion• 0.035 poise at 37 degree.
Factors determining flow
• Flow rate is determined by– Pressure gradient– Resistance • Viscosity of blood• Radius of lumen• Length of vessel
Types of flow
Laminar flowShape of parabolaConcentric layers,each parallel to vessel wallVelocity of each layer differsMaximal velocity is at centre of vesselDecreasing profile towards peripheries
Turbulent flow
• Obstruction produce increased velocities, flow vortices• Whirlpools shed off in different directions producing
variable velocities- chaos• Predicted by Reynolds number• Reynolds number depends on
Re=( ρ x c x D)/v ρ-Density of bloodD-Vessel diameterc-Velocity of flowV-viscosity
The Reynolds number is dimensionless
If Re is less than 1200 the flow will be -laminar
1200-2000 flow is described as -transitional
Greater than 2000 -turbulent
Doppler Principle
First described by Johann Christian Doppler, an Austrian mathematician and scientist who lived in the first half of the19th century.Doppler’s initial descriptions referred to changes in the wavelength of light as applied to astronomical events.
In 1842, he presented a paper entitled "On the Coloured Light of Double Stars and Some Other Heavenly Bodies" where he postulated that certain properties of light emitted from stars depend upon the relative motion of the observer and the wave source.
• Doppler effect describes the frequency shift of the signal in relation to the relative motion of a source and an observer.
• The wave generated by a source that moves away from an observer/receiver appears to him to be of lower frequency than the wave generated by a stationary source, or generated by a source moving toward the observer.
• The frequency of the signal detected by the receiver moving toward the still source is higher, compared to the frequency detected by the still receiver, or a receiver moving away from the source.
• Applied in echocardiography to determine flow direction,flow velocities,flow characteristics
• Stationary rbc-zero doppler shift(received frequency= transmitted frequency)
• Positive doppler shift-RBCs moving towards transducer ,received frequency >transmitted frequency
• Negative doppler shift:RBC’s moving away from transducer- transmitted frequency more than receiving frequency
• Doppler shift represents difference between received and transmitted frequencies ,which occur due to motion of RBC’s relative to the ultrasound beam
• Fd = (2f V cos Ø)/C
Why the factor 2?
• Double doppler shift
1st shift-transducer stationary source,RBC the moving receiver
2nd shift is when,RBCs are moving source and transducer is the stationary receiver.
Doppler equation- rearranged
Factors affecting doppler equation
• Estimation of blood flow velocity is dependent on incident angle between ultrasound beam and blood flow
• When RBCs parallel-maximum velocity• When RBCs perpendicular-no doppler shift• When angle between ultrasound beam and blood
flow is less than or equal 20 degree,cosine close to 1 and percent error is less than or equal to 7%
The Effect of Angle
Angle Cosine Percentage error0 1 010 0.98 220 0.94 730 0.87 1360 0.5 5090 0 100
Angle correction
• It is possible to correct for angle
• Not recommended as in most cases its possible to align ultrasound beam parallel by utilising multiple views, serial assessment difficult unless same angle correction used
• It is assumed that angle between ultrasound beam and direction of blood flow is parallel
• By adjusting according to the direction of assumed flow, it changes the angle calculations in the Doppler equation resulting in different estimates of flow velocity.
• The use of this control does not actually change the direction of the Doppler beam and its use does not alter the quality of either the audio output or the spectral recording
Effect of frequency
• Lower the frequency,higher the velocity detected
• A 2 MHz transducer detects higher velocity compared to a 5 MHz transducer
SPECTRAL DOPPLER DISPLAY
Flow velocity
• Displayed on y axis
• Velocity of RBCs within sampled volume is calculated
• Absence of velocity-zero baseline
Spectral velocity recordings
Direction of flow
• Flow direction also displayed on Y axis
• Positive doppler shift-flow towards transducerTraditionally displayed above baseline
• Negative doppler shift-flow away from transducerDisplayed below zero baseline
Intensity or amplitude
• Blood cells do not move at equal velocities• Produce different frequency shifts• Amplitude or intensity of doppler signal reflects
number of blood cells moving within a range of velocities at a particular point of time
• Bright signal-strong doppler shift frequency at a particular point of time .
• Darker regions-weak doppler shift
Timing
• Time is displayed along x axis
• Displayed along with ECG.
• Change in blood velocity,flow direction can be accurately timed in relation to cardiac cycle.
Doppler Audio signals
• Doppler shift frequencies are in audible range
• Guide for localising blood flow and for proper aligning ultrasound beam parallel to flow
• Laminar flow-smooth tone
• Turbulent flow-harsh sound.
Pulsed and Continuous Wave Doppler
Continuous Wave Doppler
• older and electronically more simple• continuous generation of ultrasound waves• continuous ultrasound reception• two crystal transducer• Blood flow along entire beam is observed
• ADVANTAGEability to measure high blood velocities accurately
• DISADVANTAGE1)lack of selectivity or depth discrimination2)no provision for range gating to allow selective placing of a given Doppler sample volume in space
Pulsed Wave Doppler
• Ultrasound impulses are sent out in short bursts or pulses
• transducer that alternates transmission and reception of ultrasound
• ability to provide Doppler shift data selectively from a small segment along the ultrasound beam- sample volume can be selected.
• The transducer functions as receiver for a limited time period
• Time corresponds to the interval required for sound to return from specified area.
• Another burst of sound waves are not transmitted until previous impulses are received.
• Pulse repetition frequency (PRF)–frequency at which transducer transmits pulses.
• PRF determines sampling rate.
Inability to accurately measure high blood flow velocities- aliasing
“Alias” means false
Fig.1.24
Aliasing
• The aliasing phenomenon occurs when the velocity exceeds the rate at which the pulsed wave system can record it properly
Aliasing is represented on the spectral trace as a cut-off of a given velocity with placement of the cut section in the opposite
channel or reverse flow direction
Nyquist Limit
• The Nyquist limit defines when aliasing will occur using PW Doppler.
• The Nyquist limit specifies that measurements of frequency shifts (and, thus, velocity) will be appropriately displayed only if the pulse repetition frequency (PRF) is at least twice the maximum velocity (or Doppler shift frequency) encountered in the sample volume.
The Nyquist Limit The simplest sound wave is an oscillation between two amplitudes. A sampled waveform thus needs at least two sample points per cycle. Thus the wave's frequency must not be above half the sampling frequency. This limit is called the Nyquist limit of a given sampling frequency
• Shannon's sampling theorem(Claude E. Shannon, born 1916, American mathematician) Also known as the Nyquist criterion, a general "rule" for measurement of frequencies, stating that the measurement (sampling) frequency must be at least twice the maximum frequency to be measured. Whenever Shannon's sampling theorem is not fulfilled, aliasing occurs
• Nyquist limit specifies the maximum velocity that can be recorded without aliasing.
Avoiding aliasing
Increase the Nyquist limit-1)altering variables in Doppler equation 2)high PRF mode3 )Change from PW to CW
• V = C × PRF4 f COS Ø
(PRF= Δf× 2= 2f V cos Ø × 2) C
Max velocity can be increased by1)Increasing PRF2)Decreasing transmitted frequency3)Increasing speed of sound in tissue4)Decreasing cosØ
High PRF Doppler
• It is desirable to use as high a PRF as possible for recording abnormally elevated velocity jets.
• Maximum PRF is limited by the distance the sample volume is placed into the heart.
Range ambiguity
• some of the range selectivity used in precisely locating the sample volume is lost.
• pulsing sequence is carried on over and over, some data is returned to the transducer
• data from all these volumes are added together
Baseline shift ("zero shift" or "zero off-set" )
Electronic cut and paste
Moves the aliased doppler signal upward or downward(unwrapping)
Repositioning baseline effectively increases the maximum velocity at the expense of other direction.
"baseline shift"
Pulse Wave VS Continuous Wave Doppler
CW PWDepth resolution no yesSample volume large smallHigh velocity detection yes noAliasing no yes Sensitivity more
lessControl of sample volume placement poor
good
Continuity equation
• The continuity equation states that the amount of blood flow through one cardiac chamber (or valve orifice) is the same as the blood flow through the other chambers and orifices
• It is based on the principle of conservation of mass.
• “Whatever mass flows in must flow out.”
The Continuity Equation
VELOCITY TIME INTEGRAL• Calculation of volumetric flow is complex as
flow velocity is not constant• Blood flow is pulsatile• Hence integrated velocity over time is taken• VTI is equal to area under curve cm/sec X sec.• Measure of distance that blood moves with
each heart beat.• VTI is also referred to as “stroke distance”
• Volumetric flow or stroke volume can be thus calculated as
SV=CSA X VTIContinuity equation can be rewritten as-CSA1 X VTI 1=CSA2 X VTI 2
CSA2 =(CSA1 x VTI 1) /VTI 2
• VTI is obtained by tracing the leading edge of velocity spectrum.
Clinical Applications
• Calculation of valve areas
• Calculation of regurgitant volumes and fractions
• Calculation of regurgitant orifice areas
• Calculation of intra cardiac shunt ratios
LIMITATIONS OF CONTINUITY PRINCIPLE
Erroneous determination of CSA• Measurement of diameter during wrong phase of
cardiac cycle.• Inconsistent annulus measurements
Erroneous determination of VTI• Incorrect placement of pulse wave sample volume• Significant angle between doppler beam and blood
flow• Incorrect filter/gain settings
• Bernoulli's principle is named after Swiss mathematecian Daniel Bernoulli who published his principle in his book Hydrodynamica in 1738
Bernoulli equation• Bernoulli described the conversion of energy in a fluid from one
form to another, as occurs when fluid flow in a tube that suddenly either increases or decreases its diameter.
• Bernoulli's law states that total energy at all points along a tube is the same—conservation of energy.
• Energy (the ability to do work) is composed of pressure energy (pressure x the volume of fluid),kinetic energy (fluid in motion has kinetic energy ,which is proportional to the mass and the velocity squared; mass is density x volume), and gravitational energy (the product of density, volume, height above a surface, and the gravitational constant).
Limitations of Bernoulli Equation
• Significant flow acceleration-prosthetic valves
• Significant viscous forces-muscular VSD,tunnel subaortic stenosis
• Increased proximal velocity-AS+AR,AS + HOCM
• Altered blood viscosity-polycythemia
Discrepancies between Catheter derived and Doppler derived Pressure Gradients
• Pressure gradients derived at cardiac cathetrisation and measured by Doppler may differ.
• Catheter derived gradient is peak to peak gradient between LV and Aorta
• Non simultaneous measurement• Doppler gives peak instantaneous gradient
and is greater than peak to peak gradient.
PRESSURE RECOVERY PHENOMENON
• Complex hemodymanic concept- pressure of fluid decreases as velocity increases.
• Once flow passes through a narrowing pressure drops and increases towards original value
• Rate and magnitude of pressure recovery is variable• In prosthetic valves,3 effective orifices-2 large orifices
by sides and a central small orifice.• Maximum velocity and lowest pressure is at
narrowest orifice.
• Immediately distal to orifice, pressure increases (recovers)and velocity decreases.
• Doppler gradients are measured at narrowest orifice, while catheter gradients are recorded downstream to prosthetic valve where the pressure has already recovered.
• Hence, Doppler derived pressure gradients are more compared to catheter derived pressure gradients.
Determination of Pressure Gradients
Pressure gradients are useful for assessment of severity of valvular stenosis and estimation of intracardiac pressures .
Pressure gradients commonly determined by Doppler include
1)Maximum instantaneous pressure gradient
2)Mean pressure gradient
• Maximum Instantaneous pressure gradient =4v2
(v=peak velocity)
• Mean pressure gradient cannot be obtained from the mean (or average) velocity, but must be calculated by making multiple instantaneous gradient (and hence pressure calculations) measurements, and then averaging those instantaneous pressures.
Estimation of RVSP by TR
• TR doppler signal represents the pressure difference between RV and RA during Systole
• RVSP-RAP=4(V TR 2 )
• In absence of RVOT obstructionRVSP=Pulmonary artery systolic pressure.
RVSP= BP SYSTLOLIC -4(V vsd )2
PASP=BP SYSTOLIC -4(V PDA )2
Mean Pulmonary Artery Pressure=4 (V PR-PEAK)2
PAEDP=4 (V PR-ED)2 + RVEDP
LAP= BP SYSTOLIC -4 (V MR)2
LVEDP =BP DIASTOLIC - 4 (V AR-ED)2
Pressure half-time anddeceleration time
The pressure half-time (PHT) is defined as the time (in milliseconds) required for the peak initial pressure to drop by one half .
Doppler signal measures velocity
The modified Bernoulli equation is applied to convert velocity to pressure.
PHT is the time (in milliseconds) for the velocity to drop to 0.707 of the maximum velocity . V2 = (0.707) x V1
The DT is the time required for the velocity,beginning with the peak value, when extrapolated, to cross the zero baseline
The PHT is clinically used most frequently in evaluating mitral stenosis and aortic regurgitation
MVA = 220/PHT
PHT = 0.29 xDT
MVA =220/0.29 x DT=759/DT
Limitations of PHT in calculation of MVA
• Non linear early diastolic slope
• Post balloon mitral valvuloplasty
• Significant Aortic regurgitation
• Cardiac rhythm disturbances
Non linear early diastolic slope
• Non linear or curvilinear decay –lead to erroneous calculation of PHT.
• Part of slope considered most representative should be chosen.
Post balloon mitral valvuloplasty
• Accuracy of calculated MVA by PHT declines immediate post BMV.
• PHT is directly related to chamber compliance and peak transmitral gradient
• Following BMV,abrupt changes in left atrial pressure and compliance occur altering the relationship between PHT and Mitral valve area.
• This effect on PHT lasts 24 to 48 hours.
Significant Aortic regurgitation
• Severe AR shortens PHT.• This is due to markedly elevated LVEDP which
reduces diastolic pressure gradient between LA and LV.
• Mitral valve area is thus overestimated.
Cardiac rhythm disturbances
• Tachycardia-deceleration slope is shortens,PHT shortens and thus valve area is overestimated.
Proximal Isovelocity Surface Area (PISA)
• Proximal Isovelocity Surface Area (PISA) method is based on the continuity equation.
• When a flow passes through a narrow orifice, as it approaches the narrowest region, there is a flow convergence and flow acceleration.
• PISA is the surface area of the hemisphere at the aliasing region of the flow convergence.
• PISA increases with lower aliasing velocity.
• Radius is measured from the orifice to point of colour change.
• If the flow convergence is not a true hemisphere, the angle subtended by the flow convergence at the orifice has to be measured and divided by 180 to get a correction .
• Good correlation between angiographic estimates of regurgitant flow and PISA based estimates have been reported.
Valve area calculation
A0=(2 Л r2x VN)/V0
A0=area of narrowed orifice
V0=velocity at narrowed orifice
R=radius of shellVN=aliased velocity identified as Niquist limit
Angle correction
PISA principle is for flow approaching narrow planar surfaceIn Mitral Stenosis,mitral leaflets may be funnel shaped To account for altered shape,angle corection factor ά /180 is applied.MVA= ( ά /180 ) x (2 Л r2x VN)/VMS
Limitations of PISA method
Errors in measurement of radiusValve area is proportional to square of radius-even small errors are magnified.
Errors in measurement of angle Measurement is done offline using a protractor
Angle measured in one dimension may not be true representation of valve leaflet geometry.
THANK YOU
MCQs
Maximum detectable velocity for a 5 MHz probe ,(assuming Doppler shift frequency as 5 KHz ,speed of sound in tissue as 1500m/sec and doppler beam being parallel to blood flow) ,would be
a)0.62 m/secb)0.75 m/secc)0.92 m/secd)1.0 m/sec
Percentage error when ultrasound beam makes an angle of 20 degrees to blood column
a)3b)7c)15d)30
When frequency of ultrasound probe is reduced,
a)Resolution increasesb)Maximum detectable velocity increasesc) Attenuation increasesd)All of the above
Following statements are true about pulse wave doppler compared to continuous wave doppler except
a)Better depth resolutionb)Larger sample volumec)More problem of aliasingd) Less sensitivity
Resistance to flow through a tube is directly proportional to
a)Fourth power of radiusb)Second power of radiusc)Length of tubed)None of the above
Turbulent flow typically occurs when reynolds number just exceeds
a)800b)1000c)2000d)5000
Pulse wave is preferred to continuous wave doppler to assess flow in all except
a)Pulmonary veinb)Superior venacavac)Pulmonary valved)Aortic valve
Doppler shift is directly proportional to all excepta)Speed of sound in tissueb)Velocity of blood flowc)Frequency of transducerd)Cosine of angle between interrogating beam and
blood flow
True about Mitral valve area assessment by PHT in severe AR is
a)PHT increasesb)Valve area is underestimatedc)MS severity is underestimatedd)Increases diastolic gradient between LA and
LV.
Mean pulmonary artery pressure of a patient with TR jet velocity of 3 m/sec, Peak PR velocity of 2m/sec is (in mm Hg)
a)36b)16c) 9d)20
KEY
Maximum detectable velocity for a 5 MHz probe ,assuming Doppler shift frequency as 5 KHz ,speed of sound in tissue as 1500m/sec and doppler beam being parallel to blood flow, would be
a)0.62 m/secb)0.75 m/secc)0.92 m/secd)1.0 m/sec
Percentage error when ultrasound beam makes an angle of 20 degrees to blood column
a)3b)7c)15d)30
When frequency of ultrasound probe is reduced,
a)Resolution increasesb)Maximum detectable velocity increasesc) Attenuation increasesd)All of the above
Following statements are true about pulse wave doppler compared to continuous wave doppler except
a)Better depth resolutionb)Larger sample volumec)More problem of aliasingd) Less sensitivity
Resistance to flow through a tube is directly proportional to
a)Fourth power of radiusb)Second power of radiusc)Length of tubed)None of the above
Turbulent flow typically occurs when reynolds number just exceeds
a)800b)1000c)2000d)5000
Pulse wave is preferred to continuous wave doppler to assess flow in all except
a)Pulmonary veinb)Superior venacavac)Pulmonary valved)Aortic valve
Doppler shift is directly proportional to all excepta)Speed of sound in tissueb)Velocity of blood flowc)Frequency of transducerd)Cosine of angle between interrogating beam and
blood flow
True about Mitral valve area assessment by PHT in severe AR is
a)PHT increasesb)Valve area is underestimatedc)MS severity is underestimatedd)Increases diastolic gradient between LA and
LV.
Mean pulmonary artery pressure of a patient with TR jet velocity of 3 m/sec, Peak PR velocity of 2m/sec is (in mm Hg)
a)36b)16c) 9d)20