9
Datalogger DL.OCS/N/RS485 CUSTOMER BENEFITS Multi-parameter probe: pressure, temperature and conductivity High-precision digital data logger with user-friendly software Thanks to the huge memory (1.5 Mio. values per channel) and the long battery lifetime (up to 10 years) the cost of ownership can be reduced to a minimum For varying conditions of use, 16 measuring series can be set up parallel Stainless steel and titanium version for use in acidic or otherwise aggressive media www.stssensors.com Version: 28.03.2017 1

DL.OCS/N/RS485 - Automated Control › ... › 2017 › 07 › Datasheet-DL.OCS-N … · DL.OCS/N/RS485 CUSTOMER BENEFITS • Multi-parameter probe: pressure, temperature and conductivity

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

  • Datalogger

    DL.OCS/N/RS485

    CUSTOMER BENEFITS

    • Multi-parameter probe: pressure, temperature and conductivity

    • High-precision digital data logger with user-friendly software

    • Thanks to the huge memory (1.5 Mio. values per channel) and the long battery lifetime (up to 10 years) the cost of ownership can be reduced to a minimum

    • For varying conditions of use, 16 measuring series can be set up parallel

    • Stainless steel and titanium version for use in acidic or otherwise aggressive media

    www.stssensors.comVersion: 28.03.2017

    1

  • Technical Specifications

    PRESSURE MEASURING RANGE (MH2O)

    2 ... 5 > 5 ... 20 > 20 ... 250

    Overpressure ≥ 3 bar ≥ 3 x FS (≥ 3 bar) ≥ 3 x FS

    Burst pressure, (1) > 200 bar > 200 bar > 200 bar

    Accuracy (± % FS) ≤ 0.15 ≤ 0.05 ≤ 0.03

    Total Error, (2), (3)(± % FS)

    -5 ... 50°C, (typ. / max.) ≤ ± 0.20 / 0.40 ≤ ± 0.10 / 0.20 ≤ ± 0.5 / 1.0 cmH2O

    ≤ ± 0.05 / 0.10

    -5 ... 80°C, (typ. / max.) ≤ ± 0.50 / 1.00 ≤ ± 0.10 / 0.20 ≤ ± 1.0 / 2.0 cmH2O

    ≤ ± 0.10 / 0.20

    Long term stability, (4) < 0.5 % FS / < 4 mbar < 0.2 % FS / < 4 mbar < 0.1 % FS / < 0.2 % FS

    (1) Transducer(2) Total error including accuracy, hysteresis, repeatability and temperature influences(3) The error values are valid within the corresponding temperature range(4) 1 year (typ. / max.)

    TEMPERATURE MEASURING RANGE, (1) (°C)

    -5 ... 50 -5 ... 80

    Accuracy (2) ≤ ± 0.5 °C ≤ ± 1.5 °C

    Response time, (3), (4)

    T 0.50 9 s 9 s

    T 0.63 15 s 15 s

    T 0.90 27 s 27 s

    (1) Temperature measurement included(2) Accuracy of the equipment ± 2 °C(3) Time in seconds that the sensor needs to carry out eg 63% of a temperature change(4) Time of measurement for liquid medium

    CONDUCTIVITY MEASURING RANGE

    Standard 0 ... 200 mS / cm

    Accuracy

    0 ... 200 µS / cm ≤ ± 2.5 % FS

    0 ... 2 mS / cm ≤ ± 1.5 % FS

    0 ... 20 mS / cm ≤ ± 1.5 % FS

    0 ... 200 mS / cm ≤ ± 1.5 % FS

    www.stssensors.com 2

  • TEMPERATURE RANGE

    Operating temperature, (1) -5 ... 50°C / -5 ... 80°C

    Process temperature, (1) -5 ... 50°C / -5 ... 80°C

    Storage temperature -40 ... 85°C

    (1) Depending on cable type, pressure range, seal, measuring medium

    ELECTRICAL SPECIFICATIONS

    Resolution

    Pressure 14 Bit

    Temperature 14 Bit

    Conductivity 14 Bit

    Output

    Interface RS485

    Protocol STS Layer 8

    Baudrate 38'400 bps

    Power supply (1) Lithium Batterie SAFT LS 14500, 3.6V AA 2250 mAh

    Standby current 6 uA / 15 uA (typ./max.)

    Operating current < 25 mA

    Battery life, (2) > 10 years

    Max. cable length 300 m

    (1) External power supply (9...30VDC) on request(2) Lifetime at 1 measurement / hour, 0...40°C

    www.stssensors.com 3

  • FUNCTIONS

    Data format Data are stored in ASCII format

    Data memory Up to 1.5 Mio measurement values per channel, data remains in

    memory even without battery, each measurement

    value is correlated with time and date

    Data transfer Read out data per measurement series, Read out all stored data, Read

    out data for a defined time-period

    Real-time clock Quartz-precision clock with date, Start-time of

    datalogging configurable

    Identification Each datalogger has a unique serial number, as well as a user-definable

    description

    Battery indicator Battery level indicator, calculated by considering the ambient temperature and the battery properties

    Configuration Sample rate, threshold values, Identification (f.e. measuring series), Taring of measurement value,

    Density of the measuring medium, Measurement

    units

    SYSTEM REQUIREMENTS

    PC / Notebook Min. 1.6 GHz Dual Core x86, Memory: Min. 10 GB,

    RAM: Min. 2 GB

    Tablet PC Please contact STS

    Operating System (1) Windows XP SP3, 7, 8, 8.1, 8 Pro (32-/ 64-Bit), 10

    (1) Not compatible with Windows 8 RT

    www.stssensors.com 4

  • QUALIFICATIONS

    Description Level Typical interferences

    EN 61000-4-2EN 61326-1

    Electrostatic discharge 8 kV contact / 15 kV air

    EN 61000-4-4EN 61326-1

    Transients (burst) 4 kV Motors, valves

    EN 61000-4-5EN 61326-1

    Surge 2 kV Overvoltage

    EN 61000-4-6EN 61326-1

    Conducted RF 10 V (0.15 ... 80 MHz) Frequency converters

    IEC 60068-2-6 Vibration 2 G (5...2000Hz)

    PHYSICAL SPECIFICATIONS

    Materials

    Transducer Stainless steel (316L / 1.4435), titanium (Gr. 2)

    Housing level transmitter Stainless steel (316L / 1.4435), titanium (Gr. 2)

    Housing suspension Stainless steel (316L / 1.4435), titanium (Gr.2)

    Seals Viton (standard), EPDM, Kalrez, NBR

    Cable PUR, FEP, PE

    Weight (1) 150 g

    (1) Specification for a DL.OCS/N/RS485, closed, without cable

    CABLE SPECIFICATIONS

    Pressure Temperature

    PUR ≤ 25 bar -5 … 50°C

    FEP ≤ 25 bar -5 … 80°C

    PE ≤ 25 bar -5 … 80°C

    www.stssensors.com 5

  • Accessories

    OVERVIEW

    10.00.0091 Accessories overview

    SOFTWARE / FIRMWARE

    114209 DL.OCS PC Application Software

    Additional documents

    OPERATING AND SAFETY INSTRUCTIONS

    Article number

    10.88.0397 DMM033

    www.stssensors.com 6

  • Ordering informationX. XX XX. XX XX. XX. XXX

    Type

    DL.OCS/N/RS485

    Pressure type

    Gauge 1

    Absolute (vacuum) 2

    Pressure measuring range (1)

    0 ... 2 mH2O and 0 ... 250 mH2O XX

    Offset, special adjustment 99

    Model

    With connection housing (Fig. 1) 1

    With connection housing (for ext. power supply) (Fig. 1) 3

    Without connection housing (Fig. 3) 0

    Cable

    PUR cable, IP 68, black (2) 0

    PE cable, IP 68, black (2) 1

    FEP cable, IP 68, black (2) 2

    PUR cablel, black, IP 68, plug-in (Fig. 7), (2) 4

    PVC cable, blue (ACS certified), (2) 5

    Without cable (Fig.3) 3

    Process connection

    Closed (Fig. 4) 57

    Open (Fig. 5) 58

    Closed, 1.4435 (Fig. 4) (4) 59

    G 1/4 A 11

    G 1/2 A 13

    Transmitter housing material

    316L Stainless steel 1.4435 0

    Titanium CP Grade 2 1

    Titanium CP Grade 2 (absolute type) 1

    Connector housing material

    316L Stainless steel 1.4435 0

    Titanium CP Grade 2 1

    without connection housing (Fig.3) 2

    Seal material

    Viton (Standard) 0

    EPDM 1

    Kalrez 2

    NBR (ACS) (4) (5) 3

    Temperature range

    -5 ... 50°C compensated (allowed process temperature: -5 ... 50°C)

    4

    -5 ... 80°C compensated (allowed process temperature: -5 ... 80°C), (3)

    5

    Option

    Conductivity (316L Stainless steel 1.4435) (Fig. 6) D

    Conductivity (Titanium CP Grade 2) (Fig. 6) P

    Ballast weight 316L Stainless steel 1.4435) B

    www.stssensors.com 7

  • (1) Other pressure ranges on request

    (2) Please specify the required cable length

    (3) Cable type PE & FEP, pressure range < 10 bar

    (4) For ACS Certification

    (5) For ACS Certification only without conductivity

    Flooding protection 316L Stainless steel 1.4435 I

    www.stssensors.com 8

  • Technical drawings

    www.stssensors.com 9