DIATHERMY. PT 156: Physical Agents 1. Aila Nica J. Bandong , PTRP Instructor Department of Physical Therapy UP-College of Allied Medical Professions. Objectives. At the end of the lecture, the students should be able to: Define diathermy Identify the parts of the SWD/MWD machine - PowerPoint PPT Presentation

Citation preview

Page 1: DIATHERMY
Page 2: DIATHERMY

At the end of the lecture, the students should be able to:

Define diathermy Identify the parts of the SWD/MWD machine Describe how heat is produced in SWD/MWD Enumerate the therapeutic effects of diathermy Differentiate the types of SWD Enumerate the indications, contraindications,

and precautions/guidelines for the use of the modality

Use clinical decision making skill in choosing appropriate modality

Apply evidence regarding use of SWD/MWD

Page 3: DIATHERMY

The use of non-ionizing electromagnetic energy from the radio-frequency spectrum as therapeutic agent

Page 4: DIATHERMY

Long waveLong wave- longest wavelength 300 – 30 m- most penetrating- no longer utilized due to high potential of causing burns and interference with radio transmissions

ShortwaveShortwave MicrowaveMicrowave

Page 5: DIATHERMY
Page 6: DIATHERMY

Superficial and Superficial and deep heating deep heating modalitymodality

FrequencyFrequency- 27.12 MHz

WavelengthWavelength- 11 m

Method of Method of HeatHeat TransferTransfer- Conversion

Manner of Manner of DeliveryDelivery- continuous- pulsed

Page 7: DIATHERMY
Page 8: DIATHERMY

Pulse Repetition Rate (PRR)Pulse Repetition Rate (PRR)- 15 to 800 Hz

Pulse Duration (PD)Pulse Duration (PD)- 25 to 400 microseconds

Peak Pulse Power (PPP)Peak Pulse Power (PPP)- 100 to 1000 watts

DurationDuration- 20 minutes (5-15 acute; 10-20 chronic)

Page 9: DIATHERMY

1.1. Cycle Duration = 1000 / PRRCycle Duration = 1000 / PRR

2.2. % cycle SWD delivered =% cycle SWD delivered =

(PD x 100) / Cycle Duration(PD x 100) / Cycle Duration

3.3. Mean Power delivered =Mean Power delivered =

PPP x % cycle SWDPPP x % cycle SWD

Page 10: DIATHERMY
Page 11: DIATHERMY

Dependent on:SPECIFIC ABSORPTION RATE

Tissue conductivity

charged molecules

dipolar molecules

non-polar molecules

Electrical field magnitude

Page 12: DIATHERMY

Ions and certain proteins

Molecules are accelerated along lines of electric force

Most efficient way of heat production

+

+

+

+

-

-

-

-

Page 13: DIATHERMY

Water and some proteins

Positive pole of the molecule aligns itself to the negative pole of the electric field (vice versa)

Moderately efficient heat production

Page 14: DIATHERMY

Fat cells Electron cloud is distorted but

negligible heat is produced Least efficient heat production

Page 15: DIATHERMY

Blood, having high ionic content, is a good conductor vascular tissues as well

Metal and sweat are good conductors if metal implants and sweat are present within the electric field, may cause burn

Page 16: DIATHERMY
Page 17: DIATHERMY

Patient’s tissues are used as DIELECTRIC between the conducting electrodes

Oscillation and rotation of the molecules of the tissues produces heat

Either flexible metal plates (malleable) or rigid metal discs can be used as electrodes

Can be applied in 3 ways: contraplanar, coplanar, or longitudinal

Page 18: DIATHERMY

ContraplanarContraplanar- aka Transverse positioning- plates are on either side of the limb

Page 19: DIATHERMY

CoplanarCoplanar- plates parallel with the longitudinal section of the body part; same side

Page 20: DIATHERMY

LongitudinalLongitudinal- plates are placed at each end of the limb

Page 21: DIATHERMY

No conclusive evidence as to the technique of application that will produce the most effect on the heated tissue (Kitchen and Bazin, 1996)

Page 22: DIATHERMY

Electrodes should be:Electrodes should be: Equal in size Slightly larger than the area treated Equidistant and at right angles to

the skin surface

Page 23: DIATHERMY

Patient is in the electromagnetic field or the electric circuit produce strong magnetic field induce electrical currents within the body (EDDY currents)

Utilizes either an insulated cable or an inductive coil applicator

Page 24: DIATHERMY

Monode:Monode: coil arranged in one plane

Hinged Diplode:Hinged Diplode: permits electrode to be positioned at various angles around the three sides of the body part, or in one plane

Page 25: DIATHERMY

Some studies argue that inductive diathermy produces greater increase in temperature of deeper tissues compared to condenser/capacitive technique

Any deep effects following capacitive technique requires considerable increase in superficial tissue temperature

Page 26: DIATHERMY

Increase blood flow Assist in resolution of inflammation Increase extensibility of deep

collagen tissue Decrease joint stiffness Relieve deep muscle pain and

spasm

Page 27: DIATHERMY

Soft tissue healing- conflicting evidence as regards effectiveness of SWD- controlled animal studies revealed insignificant results as well as trials involving human subjects (Kitchen and Bazin, 1996); to date, no studies in the treatment setting was conducted

Page 28: DIATHERMY

Recent ankle injuries- inconclusive results following three double-blind protocols (Kitchen and Bazin)

Page 29: DIATHERMY

Pain Syndromes- Pulsed SWD may provide better pain relief in some musculoskeletal conditions (neck and back) than SWD

A. Nerve Regeneration- studies were done on cats and rats- PSWD induced regeneration of axons, acceleration and recovery of nerve conduction

B. Osteoarthritis- no established effect

C. Post-operative- insignificant (abdominal surgery

Page 30: DIATHERMY
Page 31: DIATHERMY

Superficial and deep heatingSuperficial and deep heating Frequency:Frequency: 300 MHz to 300 GHz Wavelength:Wavelength: 1m to 1mm Therapeutic Parameters:Therapeutic Parameters:

A. 122.5 mm – 2456 MHzB. 327 mm – 915 MHzC. 690 mm – 433.9 MHz

Dosage:Dosage: acute 5 to 15 minutes chronic 10 to 20 minutes

Page 32: DIATHERMY
Page 33: DIATHERMY

Direct current (DC) is shunted to the cathode in the magnetron valve

Release of electrons from the cathode to the multi-cavity anode valve

Electrons oscillate at predetermined frequency

High frequency alternating current is transmitted along a coaxial cable

Coaxial cable transmits energy to a director

Page 34: DIATHERMY

AbsorbedAbsorbed- energy is taken up by the material

TransmittedTransmitted- pass through the material without being absorbed

RefractedRefracted- direction of propagation is altered

ReflectedReflected- turned back from the surface

Page 35: DIATHERMY

Increased blood flow or circulation to the area

Increased tissue temperature Increased metabolism Facilitate relaxation Increased pain threshold Decreased blood viscosity

Page 36: DIATHERMY

Soft tissue injury Mobilization Pain relief

Page 37: DIATHERMY

Pacemakers Metal implants Impaired

sensation Pregnancy Hemorrhage Ischemic

Tissue Testicles and

eyes

Malignant CA Active TB Fever Thrombosis X-ray exposure Uncooperative

patient Areas of poor

circulation

Page 38: DIATHERMY

Operator should observe caution when handling the machine: same contraindications apply

Page 39: DIATHERMY

Gorgon, E. J. (2004). Lecture notes on high frequency currents: Shortwave and microwave diathermy. University of the Philippines- College of Allied Medical Professions.

Hayes, K. W. (1993). Manual for physical agents (4th Ed). Connecticut: Appleton and Lange.

Hecox, B., Mehreteab, T. A., and Weisberg, J. (1994). Physical agents: A comprehensive text for physical therapists. Connecticut: Appleton and Lange.

Kitchen, S. and Bazin, S. (1996). Clayton’s electrotherapy (10th ed). Philadelphia: W.B. Saunders Company.

Low, R. Reed, A. (1995). Electrotherapy explained: Principles and practice (2nd Ed). Oxford: Butterworth-Heinemann Ltd.

Michlovitz, S. L. (1996). Thermal agents in rehabilitation (3rd Ed). Philadelphia: F. A. Davis Company.