Khaldoon Alaswad, MD, FACC, FSCAI

Assistant Professor of Medicine Wayne State University

Director, Cardiac Catheterization Laboratory

Henry Ford Hospital and Health System

Device Troubleshooting:

IABP, Impella, TandemHeart, and

ECMO

SCAI Shock 201810 m

Page 2: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

IABP Inflation & Deflation

Page 3: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

Wrong inflation or deflation time

▪ Early deflation: systolic augmentation can be lost

▪ Late deflation increase afterload

▪ Early inflation increase afterload

▪ Late inflation less consequential but decreases systolic

augmentation

Created on Wed, 06/17/2015 - 19:17Last updated Mon, 05/14/2018 - 19:49TopicCardiothoracic Intensive Care

Page 4: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

Impella potential Troubles and Troubleshooting

▪ Abnormal Position

– Aortic Position

– Ventricular

▪ Suction

▪ Hypotension

▪ Hemolysis

Page 5: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

Impella Correct Position

Page 6: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

Impella 5.0 & Impella LD

Page 7: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on
Page 8: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

Impella CP is too deep

▪ Placement Signal:

▪ ?

▪ Motor Current:

▪ ?

▪ Actions

▪ 1. Reduce Flow Rate to P-2.

▪ 2. Echo guidance to reposition the pump until the inlet area measures 3.5 cm below the aortic valve annulus.

▪ 4. Lock Touhy Valve.

Page 9: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

Impella CP is not deep enough

▪ Placement Signal:

▪ ?

▪ Motor Current:

▪ ?

▪ Actions

▪ 1. Reduce Flow Rate to P-2.

▪ 2. Echo guidance to reposition the pump until the inlet area measures 3.5 cm below the aortic valve annulus.

▪ 4.Lock Touhy Valve.

Page 10: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

If the Impella motor current is not restored by

pushing the catheter

▪ Cancel Echo and move to the cath lab

Page 11: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

Impella 5.0 or LD

▪ Placement Signal:

▪ ?

▪ Motor Current:

▪ ?

▪ Actions

▪ 1. Reduce Flow Rate to P-2.

▪ 2. Using echo guidance, reposition the pump until the inlet area measures 3.5 cm below the aortic valve annulus.

▪ 3. Resume previous P-level setting.

▪ 4. Lock Touhy Valve.

Page 12: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

Impella Suction Alarm

Risks

▪ Less support

▪ Hemolysis

Causes

▪ Wrong Impella inlet

position

▪ Low LV filling pressure:

–Low volume

–RV Failure

Page 13: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

Troubleshooting of suction alarm

▪ Decrease the flow rate until the alarm stop

▪ Assess volume status

▪ Assess RV function

▪ Rule out tamponade

Page 14: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

Interference with Device Operation results in hemolysis

• Ventricular structures obstruct inlet area

• Blood subjected to higher shear in cannula

• Higher shear = hemolysis

Inflow obstruction

• Aortic valve or wall obstruct outlet area

• Force with which blood contacts obstruction = hemolysis

Outflow obstruction

• Clot or fibers narrow cannula

• Blood subject to higher shear in cannula

• Higher shear = hemolysis

Obstruction within cannula

HCS-PP00080-024 rF

Page 15: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

Impella position in the LV determine its function

Page 16: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

Hemolysis

▪ Consider adding volume if hemolysis is accompanied

by CVP or PCWP (PAD) < 10mmHg

▪ May reduce the P-level if hemolysis is present

▪ Goal is to run at the lowest possible level to achieve

the highest amount of flow

HCS-PP00080-023 rH

Page 17: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on
Page 18: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

IMPELLA

PCI

Right Heart CathCPO < 0.6 CPO ≥ 0.6 andPAPI > 0.9

Continue to Titrate ↓ Pressors/InotropesPAPI < 0.9

Possible RV Failure

Consider RV Support

RV Normal

Consider ↑ LV Support

PAPI > 0.9

CARDIAC POWER OUTPUT(CPO)

CPO = MAP x CO / 451

PULMONARY ARTERY PULSATILITY INDEX

(PAPI)PAPI = sPA – dPA / RA

Calculate PAPI

NCSI Algorithm

Page 19: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

ECPella Troubleshooting

▪ 57 y/o EcPella post LMCA

acute stent thrombosis

▪ Previous procedure was

complicated by possible

perforation

Page 20: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

Suction all night long, CVP 8, MBP 60 mmHg

Page 21: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

In the Cath Lab

Page 22: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

After US and Fluoroscopic Guided

Pericardiocentesis

▪ CVP down to 4

▪ MBP 75

Page 23: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

TROUBLESHOOTING: Purge Pressure Highnormal 300-1100

Page 24: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

TROUBLESHOOTING: Purge Pressure Low

Page 25: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

TandemHeart Troubles and Troubleshooting

▪ Catheters misposition

–Left atrial

–Arterial

▪ Low flow

Page 26: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

TandemHeart Left Atrial Catheter too Deep

Page 27: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

Troubleshooting Left Atrial Catheter is too Deep

▪ The left atrial catheter can

migrate into the pulmonary

vein ▪ Pull the catheter slowly under

TEE or fluoroscopic guidance

▪ If the blood in the catheter

became darker push forward

Page 28: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

TandemHeart Left Atrial Catheter Pulled back to

the right atria

▪ Decrease rpm to 3500

▪ Clamp the outflow

▪ Reposition under fluoroscopic

guidance

▪ Or add an oxygenator and

convert to ECMO

Page 29: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

TandemHeart Low Flow Alarm

Inflow catheter (LA) issues

▪ Catheter position

▪ Check volume status

▪ Check RV status

Outflow catheter (arterial) issues

▪ Kinks

▪ Arterial stenosis

Page 30: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

ECMO Troubleshooting

Low Flow

Inflow catheter (LA) issues

▪ Kinks

▪ Catheter position

▪ Check volume status

Outflow catheter (arterial) issues

▪ Kinks

▪ Arterial stenosis

Page 31: Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO · 2018-12-01 · Device Troubleshooting: IABP, Impella, TandemHeart, and ECMO SCAI Shock ... augmentation Created on

North South Syndrome

When the head receives deoxygenated blood