Development of TOF-MS from intellectual curiosity to practical technique Marvin Vestal SimulTOF Systems The Jim Waters Symposium at PittCon March 2, 2014

Development of TOF-MS from intellectual curiosity to practical technique

  • Upload
    lucita

  • View
    41

  • Download
    0

Embed Size (px)

DESCRIPTION

Development of TOF-MS from intellectual curiosity to practical technique . Marvin Vestal SimulTOF Systems The Jim Waters Symposium at PittCon March 2, 2014. A Brief History of Time (of flight) with apologies to Stephen Hawking. 1946 W.E Stephens, Phys. Rev. 69 ,641 - PowerPoint PPT Presentation

Citation preview

Page 1: Development of TOF-MS from intellectual curiosity to practical technique

Development of TOF-MS from intellectual curiosity to practical technique

Marvin VestalSimulTOF Systems

The Jim Waters Symposium at PittConMarch 2, 2014

Page 2: Development of TOF-MS from intellectual curiosity to practical technique

A Brief History of Time (of flight)with apologies to Stephen Hawking

• 1946 W.E Stephens, Phys. Rev. 69,641– “Advances in electronics seem to make practical a type of mass

spectrometer in which microsecond pulses of ions are selected every millisecond from an ordinary low-voltage ion source. In travelling down the vacuum tube, ions of different M/e have different velocities and consequently separate into groups spread out in space. … This type of mass spectrometer should offer many advantages over present types. The response time should be limited only by the repetition rate (milliseconds)… Magnets and stabilization equipment would be eliminated. Resolution would not be limited by smallness of slits or alignment. Such a mass spectrometer should be well suited for composition control, rapid analysis, and portable use.”

Page 3: Development of TOF-MS from intellectual curiosity to practical technique

Brief History of Time (of flight)

• 1948 Cameron & Eggers, Rev. Sci. Instr. 19, 605.– First working TOF

• 1953 Wiley & McLaren, Rev. Sci. Instr. 26, 1150– First practical TOF. Energy & Time Lag Focusing.

• 1959 Gohlke, Anal. Chem. 31, 535– GC-TOF

• 1963 Vestal & Wharhaftig, ASMS, 358.– Coincidence TOF, first ion counting TDC

• 1973 Mamyrin et al, Sov. Phys. JETP 37, 45.– Reflectron, higher resolution

• 1974 Macfarlane et al, Biochem. Biophys. Res. Comm. 60, 616– 252Cf Plasma desorption. Proteins Fly!!!!

Page 4: Development of TOF-MS from intellectual curiosity to practical technique

Brief History of Time (of flight)

• 1988 Karas & Hillenkamp, Anal. Chem. 60, 2299.– MALDI - Really big proteins fly!!!

• 1991 Dodenov et al, 12th Int. MS Conf.– O-TOF - Electrospray works with TOF

• 1993 Kaufman, Spengler & Lutzenkirchen, RCM 7, 902.– Post-source decay MALDI

• 1994 Brown & Lennon, Sunriver, p63.– Delayed extraction MALDI - Makes MALDI-TOF routine

• 1996 Morris et al, RCM 10, 889.– Q-TOF

• 2001– TOF-TOF

Page 5: Development of TOF-MS from intellectual curiosity to practical technique

Karras &Hillenkamp 1988

Early History of MALDI-TOF

Page 6: Development of TOF-MS from intellectual curiosity to practical technique

Beavis &Chait 1989

Page 7: Development of TOF-MS from intellectual curiosity to practical technique

Beavis &Chait 1989

Page 8: Development of TOF-MS from intellectual curiosity to practical technique

Va

Vg

0

d0 d1

Ds

Dv

Sample Plate

Extraction Electrode

Space focus

Detector

Pulsed AccelerationBrown &Lennon 1992, Wiley &Mclaren 1953

Initial velocity of ions from MALDI is high and independent of mass.This is the major contribution to time dispersion in static MALDI.

Page 9: Development of TOF-MS from intellectual curiosity to practical technique

Va

d0

Sample plate

Extractionelectrode

0

V1

V2

d3d4

0

d4

Drift tube

velocityfocus

Dm

2-stage reflector

D

Dodonov 1986

Delayed extraction source plus 2-stage reflectorMakes high resolution MALDI practical

Page 10: Development of TOF-MS from intellectual curiosity to practical technique
Page 11: Development of TOF-MS from intellectual curiosity to practical technique

Contributions to relative peak width, Dm/m

Initial position, dx: Rs1 = [(Dv-Ds)/De](dx/d0y)

Initial velocity, dv0: Rv1 =(4d0y/De)(dv0/vn)[(1-(m/m*)1/2] Rv2 =2[2d0y/(Dv-Ds)]2 (dv0/vn)2

Rv3 =2[2d0y/(Dv-Ds)]3 (dv0/vn)3

Time error, dt: Rt = 2dt/t = 2dtvn/De

Trajectory error, dL: RL = 2dL/De

Voltage error, dV: RV = dV/V

Resolving power:R-1 = [Rs1

2 + Rv2 + Rt

2 + RL2 + RV

2 ]-1/2

Not small!!!

Page 12: Development of TOF-MS from intellectual curiosity to practical technique

Contributions to relative peak width, Dm/m, with 1st and2nd order velocity focusing (size matters)Initial position, dx: Rs1 = [2/K](dx/De)

Initial velocity, dv0: Rv1 =(4d0y/De)(dv0/vn)[(1-(m/m*)1/2] Rv2 =2[K]2 (dv0/vn)2=0 Rv3 =2[K]3 (dv0/vn)3

Time error, dt: Rt = 2dt/t = 2dtvn/De

Trajectory error, dL: RL = 2dL/De

Voltage error, dV: RV = dV/V

Resolving power:R-1 = [Rs1

2 + Rv2 + Rt

2 + RL2 + RV

2 ]-1/2

Page 13: Development of TOF-MS from intellectual curiosity to practical technique

40000

50000

60000

70000

80000

90000

100000

110000

120000

130000

1 1.5 2 2.5 3 3.5 4

focus mass (kDa)

Reso

lvin

g Po

wer

e

14 m,RP/2

5 m

70 m, RP/10

3.2 m

Maximum resolving power at focus mass under optimized conditions.

Does not include contributions from trajectory error, HV noise, and collisions

dv0=400 m/s, dx0=0.01 mm, dt=0.75 ns

60,000

90,000

230,000

1,000,000

Page 14: Development of TOF-MS from intellectual curiosity to practical technique

BSA digest 3.2m reflector500 shots

1.3 ns

Resolution of commercial MALDI

Page 15: Development of TOF-MS from intellectual curiosity to practical technique

6 m single mirror14 m effective

Source is on firstfloor and mirroron second floor

Page 16: Development of TOF-MS from intellectual curiosity to practical technique

3 ns

14m single reflector BSA digest

Page 17: Development of TOF-MS from intellectual curiosity to practical technique

1.1 ns

14mSingleReflector

HighestResolvingPower Recorded For MALDI-TOF

MATRIX TRIMER

Page 18: Development of TOF-MS from intellectual curiosity to practical technique

Chromatogram

N/A

0 200 400 600

0

1

2

3

Spot Position

RMS

(PPM

)

RMS error for 10 peaks in spectrum of tryptic digest of BSA for all 675spots on a 102x108 mm sample plate with automatic 2-point internal Calibration for 3.2 m analyzer

Calibration Equation: m1/2=D0 +D1t[1+D2t+D3t2]Determine coefficients by least square fit to multiplePeaks covering broad mass range

-2

0

2

4

800 1000 1200 1400 1600 1800 2000

mass (Da)

mas

s er

ror (

mD

a)

Average=-0.1 mDa

RMS =0.99 ppm

Page 19: Development of TOF-MS from intellectual curiosity to practical technique

JEOL Spiraltron 17m

Page 20: Development of TOF-MS from intellectual curiosity to practical technique

F

Page 21: Development of TOF-MS from intellectual curiosity to practical technique
Page 22: Development of TOF-MS from intellectual curiosity to practical technique

High Resolution MALDI• Resolving power >20,000 for peptides and small molecules is

routine• Resolving Power approaching 100,000 has been demonstrated

with long flight distances• RMS Mass error 1-5 ppm across the plate over the full mass

range with single peak automatic calibration is routine• MALDI ions often metastable thus longer flight time imply lower

sensitivity particularly for larger, more fragile ions• Sensitivity generally limited by chemical noise, thus higher

resolving power may yield higher sensitivity

Page 23: Development of TOF-MS from intellectual curiosity to practical technique

For Linear Analyzer (size matters more)Initial position, dx: Rs1 = 2[(Dv-Ds)/2d0y](dx/De)

Initial velocity, dv0: Rv1 =(4d0y/De)(dv0/vn)[(1-(m/m*)1/2] Rv2 =2[2d0y/(Dv-Ds)]2 (dv0/vn)2=0 Rv3 =2[2d0y/(Dv-Ds)]3 (dv0/vn)3

Time error, dt: Rt = 2dt/t = 2dtvn/De

Trajectory error, dL: RL = 2dL/De

Voltage error, dV: RV = dV/V

Resolving power:R-1 = [Rs1

2 + Rv2 + Rt

2 + RL2 + RV

2 ]-1/2

Page 24: Development of TOF-MS from intellectual curiosity to practical technique

0

1000

2000

3000

4000

5000

6000

7000

8000

9000

10000

0 10 20 30 40 50 60 70 80 90 100

m/z (kDa)

Res

olvi

ng P

ower

3

1030 60

Maximum Resolving Power for Linear Analyzer.

Isotopelimit

Sensitivity, Dynamic Range, and Reproducibility are Key Metrics

Page 25: Development of TOF-MS from intellectual curiosity to practical technique

Linear @10 kV -- Higher Resolving Power than the old XL Reflector

Page 26: Development of TOF-MS from intellectual curiosity to practical technique

1 ns

Typical single ion pulsewith fast scintillator

+/-30 kV +/-30 kV

-1 kV 0 V

Photonis detector

Microchannelplate

scintillator

photomultiplier

Signalout

lighte-

Ionbeam

+20 kV

0 V

Potential diagram for linear detector

Page 27: Development of TOF-MS from intellectual curiosity to practical technique

Fig. 9 BSA (1 picomole/ml) in sinnipinic acid matrix 10,000 laser shots in 2 s.

M+

2M+

M2+

M3+

3M+3M2+

More exotic detectorsAre not required for MALDI

Page 28: Development of TOF-MS from intellectual curiosity to practical technique

Conventional Wisdom

•MALDI• is Not •Quantitive

Page 29: Development of TOF-MS from intellectual curiosity to practical technique

4 spectra 50 shots ea. On one sample spotIntensity varies by factor of 20@10 Hz requires 20 s

Your Old MALDI

Page 30: Development of TOF-MS from intellectual curiosity to practical technique

Data is acquired over mass range 1000-20,000 Da using 2 ns bins2 kHz laser rate50 shots averaged per spectrum1 mm/s snake raster for 5 passes at 500 mm over 3 mm dia spot Total travel 12 mm in 12 s generating 480 spectra with 50 mm resolutionSpectra with no significant peaks are not savedOnly 10% of sample used, 240,000 total laser shots possible

Can MALDI be Quantitative????Data Acquisition, One Example

Data Processing

Spot AverageBaseline CorrectionSmoothCalibrate and Detect PeaksNormalizeBin by mass rather than timeQuantifyReport

Page 31: Development of TOF-MS from intellectual curiosity to practical technique

347 spectra 50 shots ea.12s @2000 Hz

Page 32: Development of TOF-MS from intellectual curiosity to practical technique

0.5 mm

3 mm

17,400 shots recorded out of 24,000 shots applied10% of sample scanned in 12 sSpot 176

Page 33: Development of TOF-MS from intellectual curiosity to practical technique

Spectra from 4 different sample spots super imposed

Page 34: Development of TOF-MS from intellectual curiosity to practical technique

Linear MALDI• Resolving power is mass dependent

– 5000-8000 over limited range at low mass– 500-1000 over wide range limited by width of isotope distribution

• Each spot will yield up to 200,000 shots without degrading resolving power or accuracy and giving dynamic range limited only by chemical noise

• Mass error <30 ppm across the plate over the full mass range with single peak automatic calibration

• Dynamic range up to100,000• High laser rate and data processing makes MALDI quantitative

Page 35: Development of TOF-MS from intellectual curiosity to practical technique

Dual mirror TOF-TOF

laser mirror#1TIS#1

fragmentation chamber

pulsed acceleratorrefocuses at MS-2 source

pulsed accelerator for MS-2

DC acc. detector

lineardetector

drift tube

TIS#2

mirror#2

low mass

highmass

High resolution reflector for MS-1

X

X

X

firstfocus

thirdfocus

secondfocus

Page 36: Development of TOF-MS from intellectual curiosity to practical technique

Multiplexed selection of 10 peaks

No selection

Page 37: Development of TOF-MS from intellectual curiosity to practical technique

1/21/2011angiotensin

Isotopic envelope selected

Mono-Isotopic precursor selected

Page 38: Development of TOF-MS from intellectual curiosity to practical technique

y2

b2--17

b3-17

a3

d3 y3

y4

y5

b4

a4

a5

d5

DRVYIHPFHL

b5

b2b3

angiotensin 1296.685

Page 39: Development of TOF-MS from intellectual curiosity to practical technique

angiotensin DRVYIHPFHL 1296.685

Dual mirror TOF-TOF529,000 laser shots

Tandem Linear/reflector TOF-TOF1,000 laser shots

>200x sensitivity

Page 40: Development of TOF-MS from intellectual curiosity to practical technique

Status of TOF-TOF

• Trade-off between high resolution precursor selection and high sensitivity for fragments

• 5 kHz laser makes LC-MALDI-MS-MS practical• Multiplexing up to 10 precursors/laser shot

practical for MRM measurements• Both low energy fragmentation (psd) and high

energy(cid) are practical• Still not widely accepted relative to ESI with traps

Page 41: Development of TOF-MS from intellectual curiosity to practical technique

Applications• Pathogen Identification• Cancer typing directly from serum, tissue extracts, and other

bodily fluids• Tissue imaging

– Proteins for cancer typing– Small molecules for drug disposition

• Biomarker Identification and Validation– Mass Spec Immunoassay– Peptide quantitation (SISCAPA and others)

• Clinical assays of biomarkers for diagnosis and treatment • Protein array reader