Upload
hirokazu
View
216
Download
1
Embed Size (px)
Citation preview
Development of a human-tissue-like phantom for 3.0-T MRI
Yusuke Ikemoto and Wataru TakaoDepartment of Radiology, Okayama Kyokuto Hospital, Okayama, Okayama 703-8265, Japan
Keisuke Yoshitomi and Seiichiro OhnoCentral Division of Radiology, Okayama University Hospital, Okayama, Okayama 700-8558, Japan
Takashi HarimotoDepartment of Radiology, Kitagawa Hospital, Wake-gun, Okayama 709-0497, Japan
Susumu KanazawaDepartment of Radiology, Graduate School of Medicine, Dentistry and Pharmaceutical Sciences,Okayama University, Okayama, Okayama 700-8558, Japan
Koichi Shibuya, Masahiro Kuroda, and Hirokazu Katoa)
Department of Radiological Technology, Graduate School of Health Sciences, Okayama University, Okayama,Okayama 700-8558, Japan
(Received 18 April 2011; revised 29 September 2011; accepted for publication 6 October 2011;
published 31 October 2011)
Purpose: A 3.0-T MRI phantom having human-tissue-equivalent relaxation times was developed.
Methods: The ingredients of the phantom are carrageenan (for gelatinization), GdCl3 (as a
T1-relaxation modifier), agarose (as a T2-relaxation modifier), and NaN3 (as an antiseptic agent).
Numerous samples with varying concentrations of GdCl3 and agarose were prepared, and T1 and T2
were measured using 3.0-T MRI.
Results: Relaxation times of the phantom samples ranged from 395 to 2601 ms for T1 values and
29 to 334 ms for T2 values. Based on the measured results, empirical formulae were devised to
express the relationships between the concentrations of relaxation modifiers and relaxation times.
Conclusions: Adjustment of GdCl3 and agarose concentrations allows arbitrary setting of relaxa-
tion times, and the creation of a phantom that can mimic relaxation times of human-tissue. Carra-
geenan is considered the most suitable as a gelling agent for an MRI phantom, as it permits the
relatively easy and inexpensive production of a large phantom such as for the human torso, and
which can be easily shaped with a knife. VC 2011 American Association of Physicists in Medicine.
[DOI: 10.1118/1.3656077]
Key words: 3.0-T MRI, phantom, carrageenan, GdCl3, agarose
I. INTRODUCTION
The signal to noise ratio (SNR) and contrast to noise ratio
(CNR) of 3.0-T MRI are much greater than those of 1.5-T
MRI,1,2and thus 3.0-T MRI systems have gradually become
widely used in clinical settings. Benefits of 3.0-T MRI
include a shorter examination time, reduction of contrast
agent in MR angiography examinations, improved spatial re-
solution for biliary and pancreatic duct images, and
improved frequency resolution for MR spectroscopy exami-
nation.3 However, artifacts increase in 3.0-T MRI because of
increased magnetic susceptibility effects and chemical shift
effects, leading to a reduction of image contrast and prolon-
gation of scanning time. In addition, increased frequency of
RF magnetic fields due to the strengthening of the static
magnetic field from 1.5 T to 3.0 T induces adverse effects.
The specific absorption rate (SAR) in the human body from
RF magnetic fields at 3.0 T is approximately four times
greater than that at 1.5 T, making it necessary to set appro-
priate pulse sequences and flip angles to prevent burning
from the high SAR.4 Nonuniformity of MR images occurs in
3.0-T equipment due to attenuation and shortened wave-
length of the RF magnetic field passing through the human
body, induced by the high conductivity and permittivity of
the human body. Research to investigate and overcome these
problems raises the need of an MRI phantom having human-
tissue-equivalent relaxation times and dielectric properties.
A phantom’s T1 and T2 are modified by a static magnetic
field, while its conductivity and permittivity are dependent
on the frequency of applied RF magnetic fields. We previ-
ously successfully developed a phantom using carrageenan
gel (CAGN phantom), for use with 1.5-T MRI.5–9 In the
present study, we have now developed a new phantom hav-
ing human-tissue-equivalent relaxation times for 3.0-T MRI.
Numerous sample phantoms with varying amounts of T1 and
T2 relaxation modifiers were prepared, and the relaxation
times of each sample were measured. Based on these meas-
ured results, empirical formulae were obtained by the nonlin-
ear least-squares method, which showed the relationships
between the concentrations of the two modifiers and both T1
and T2 values. The creation of a human-tissue-like 3.0-T
phantom with two arbitrarily-set relaxation times is possible
with the appropriate blend of T1 and T2 modifiers according
to the empirical formulae.
6336 Med. Phys. 38 (11), November 2011 0094-2405/2011/38(11)/6336/7/$30.00 VC 2011 Am. Assoc. Phys. Med. 6336
Carrageenan, normally used as a low-cost food additive,
consists of saccharides extracted from red algae (seaweed).
These saccharides have molecular weights of 100 000–500
000, comprised mainly of galactose and 3,6-anhydrogalac-
tose. Carrageenan is a gelling material similar to agar though
relatively cheaper. While agarose10–12 and agar13,14 have
been widely used as gelling agents for MRI phantoms, the
benefits of carrageenan over other gels such as polysaccha-
ride or gelatin,15 include greater elasticity and strength,
being formable into a large and stable phantom, and having
an easily modified shape.5 In contrast to agarose, carra-
geenan has a little influence on T2, making possible a phan-
tom with a long T2 value.
II. MATERIALS AND METHODS
II.A. Sample preparation
Materials used for the phantoms were carrageenan
(KC-200S: Chuo Kasei Co., Ltd., Osaka, Japan) for gelati-
nization; GdCl3 (Sigma Chemical Corp., St. Louis, MO,
USA) as a T1 modifier; agarose (Type 1, #A-6013: Sigma
Chemical Corp., St. Louis, MO, USA) as a T2 modifier;
NaN3 (Katayama Chemical, Osaka, Japan) as an antisep-
tic; and, distilled water.5 Agarose was chosen as a T2
modifier as it affects that value in MRI. Conversely, carra-
geenan was selected as the gelling agent because it has
almost no impact on T2 values. As a control phantom, a
solution of NiCl2(Sigma Chemical Corp., St. Louis, MO)
was used.
The total weight of each sample was 100 g, including
distilled water. Sample size was limited by the total num-
ber of samples needed to evaluate more than 180 combina-
tions of T1 and T2 modifier concentrations necessary for
measuring the various relaxation times. Previous studies
have shown that phantom size is irrelevant due to the phan-
tom’s homogeneity.9 The concentrations of carrageenan
and NaN3 were fixed at 3 and 0.03 w/w%, respectively.
Based on previous reports, we used the lowest possible
concentration of carrageenan that would yield a phantom
physically strong enough to maintain its shape, and the
lowest possible concentration of NaN3 necessary to prevent
the occurrence of mold.5 The concentration of GdCl3ranged from 0 to 180 lmol/kg, while that of agarose ranged
from 0 to 2.0 w/w%.
All ingredients for each phantom were mixed together in
a Pyrex tube (diameter: 40 mm, height: 130 mm). The mix-
ture was then heated while being stirred in a hot water bath
at 90 �C to dissolve the carrageenan and agarose. This mix-
ture was then boiled in a microwave oven at 90–100 �C to
complete the dissolving of the agarose.9 Heating in the
microwave oven serves to prevent the occurrence of many
defects when imaging the phantom. Next, the mixture was
cooled to 25 6 1 for gelling purposes. Finally, each sample
phantom tube was sealed with a rubber stopper to prevent
water evaporation. For verification of MRI equipment set-
tings and the validity of our data, a 100 g control phantom
was prepared, consisting of NiCl2 solution set at 14 mmol/
kg.
II.B. MRI measurement
Measurements of T1 and T2 were made using a 3.0-T
Magnetic Resonance Imager (Signa Excite HDx 3.0 T, GE
Healthcare ). The Pyrex tube containing the control phantom
was placed at the center of a custom-made acrylic case, and
18 Pyrex tubes, each containing a phantom sample, were
placed inside the case around the control. Figure 1(a) shows
an array of samples for measurement in which 18 sample
phantoms and 1 control at the center of the array are set in
the head coil. Figure 1(b) is an SE image of such an array,
obtained at TR¼ 15 000 ms, TE¼ 15 ms. An region-of-in-
terest (ROI) was identified on each sample, and T1 and T2
values were then calculated.
The samples and control were then scanned with an axial
projection using a head coil. The scan parameters were: slice
thickness, 10 mm; matrix, 256� 256; FOV, 220 mm; band
width, 6 15.63 kHz; and, number of acquisitions, 1. To com-
pare the relaxation times at 3.0 T with those at 1.5 T, previ-
ously reported pulse sequences5 were used here for
measuring T1 and T2 values . T1 was measured using the sat-
uration recovery method with a constant TE value of 15 ms
and TR values of 133, 167, 217, 300, 400, 533, 717, 950,
1683, 3000, 5283, 9317, and 15 000 ms. T2 was measured
using the spin echo method with a constant TR value of 10
000 ms and TE values of 15, 22, 29, 39, 52, 69, 93, 125, 167,
224, and 300 ms. Calculation of T1 and T2 values was per-
formed using “IMAGE J” by setting a ROI 20 mm in diameter
(448 pixels) in the center of each sample.
II.C. Examination of the relationship betweenconcentrations of modifiers and relaxation times
Based on the obtained data, and using methods previously
reported,5 the relationships between the concentrations of
the two modifiers and the two relaxation times were both for-
mulated using the nonlinear least-squares method (Gauss–-
Newton Method) installed in MATLABVR
(The MathWorks,
Inc., Natick).
III. RESULTS
III.A. Relaxation times of samples
The relaxation times of the varied mixtures of GdCl3 and
agarose are plotted in Fig. 2. Figure 2(a) shows the effect of
FIG. 1. Sample phantoms within an acrylic case inserted in a head coil (a)
and MR image of sample phantoms (b).
6337 Ikemoto et al.: Development of a human-tissue-like phantom for 3.0-T MRI 6337
Medical Physics, Vol. 38, No. 11, November 2011
the GdCl3 and agarose concentrations on T1 and T2. Each
solid line corresponds to a different iso-agarose concentra-
tion, while each broken line indicates an iso-GdCl3 concen-
tration. T1 decreases with an increased GdCl3 concentration,
irrespective of the agarose concentration. T2 decreases with
an increased agarose concentration and further decreases
with the increase of the GdCl3 concentration. In Fig. 2(b),
the relaxation times for a range of human-tissues2,16–19have
been superimposed on our data. The range of relaxation
times obtained for our various samples fall within those of
most types of human-tissues.
III.B. Relationship between concentration of modifierand 1/T1
Based on the measured T1 values, Fig. 3 shows the T1
relaxation rate (ms� 1) relative to GdCl3 concentrations
(lmol/kg) of the samples having identical agarose concen-
trations. The open circles represent the data for 0% agarose
concentration. The solid line, plotted as a quadratic function,
is the least-squares fit to the data indicated by the open
circles. Likewise, each broken line, plotted as a quadratic
function, is the least-squares fit to the data relative to the dif-
ferent agarose concentrations within the range of
0.2%–2.0%. As the concentration of agarose increases, the
broken lines tend to move toward the x-axis. The relation-
ship between the GdCl3 concentration and T1 relaxation rate
could be approximated by the quadratic functions, and the
coefficients of determination (R2s) of each regression curve
were over 0.9 for all agarose concentrations (0%–2.0%).
Each quadratic curve can be expressed as follows:
1
T1
¼ a1ðAÞ þ b1ðAÞGþ c1ðAÞG2; (1)
where G is the variable of the GdCl3 concentration in lmol/
kg. a1(A), b1(A), and c1(A) are the coefficients of each term
of the Eq. (1), and also the function of the agarose concentra-
tion as a percentage, A (%).
Figure 4 shows the relationship between the agarose con-
centration and the coefficients of Eq. (1). The coefficients,
a1(A), b1(A), can be expressed by a linear equation, while
c1(A) can be expressed by a quadratic equation, that is,
a1(A)¼ a1þ a2A, b1(A)¼ a3þ a4A, c1(A)¼ a5þ a6Aþ a7A2,
where a1–a7 are the coefficients of the equations. Accordingly,
the relationship between the concentrations of GdCl3 and aga-
rose and T1 can be expressed by the following equation:
FIG. 2. T1 and T2 for all samples (*) and human tissues (�). (a) T1 and T2 for all sample mixtures of GdCl3 and agarose. Each solid line corresponds to a
different iso-agarose concentration, while each broken line shows a different iso-GdCl3 concentration. The GdCl3 concentrations range from 0 to 180
lmol/kg, while the agarose concentrations range from 0 to 2%. (b) The relaxation times of human tissues are superimposed on all the sample data. Note that
the range of relaxation times obtained for the samples covers the relaxation times of most human tissues.
FIG. 3. T1 relaxation rate relative to GdCl3 concentration. The open circles
represent the data for 0% agarose. The solid line, plotted as a quadratic func-
tion, is the least-squares fit to the data indicated by the open circles. Each
broken line, plotted as a quadratic function, is the least-squares fit to
the data relative to different agarose concentrations within a range of
0.2%–2.0%. The value of R2 for each regression curve is greater than 0.9.
As the concentration of agarose increases, the broken lines tend to move
toward the x-axis.
6338 Ikemoto et al.: Development of a human-tissue-like phantom for 3.0-T MRI 6338
Medical Physics, Vol. 38, No. 11, November 2011
1
T1
¼ a1 þ a2 Aþ ða3 þ a4 AÞG
þ ða5 þ a6 Aþ a7 A2ÞG2 (2)
III.C. Relationship between concentration of modifierand 1/T2
Based on the measured T2 values, Fig. 5 shows the T2
relaxation rate (ms� 1) relative to the agarose concentrations
(%) of the samples having the same GdCl3 concentration.
The open circles represent the data for 0 lmol/kg of GdCl3concentration. The solid line, plotted as a linear function, is
the least-squares fit to the data indicated by the open circles.
Likewise, each broken line, plotted as a linear function, is
the least-squares fit to the data relative to the different GdCl3concentrations within a range of 4–180 lmol/kg. The value
of R2 for each regression line is over 0.9. As the concentra-
tion of GdCl3 increases, the broken lines tend to move away
from the x-axis. Each linear equation can be expressed by
the following equation:
1
T2
¼ a2ðGÞ þ b2ðGÞA; (3)
where A (%) is the variable of agarose concentration in per-
centage, a2(G) is the value of the intercept, and b2(G) is the
inclination of the line, which are functions of the GdCl3 con-
centration expressed as G (lmol/kg).
Figure 6 shows the relationship between the GdCl3 con-
centration and the coefficients of Eq. (3) for each line of Fig.
5. The coefficient, a2 (G), can be expressed by a quadratic
equation, while b2(G) can be expressed by a linear equation,
that is, a2 (G)¼ b1þ b2 Gþ b3 G2, b2¼ b4þ b5 G, where
b1–b5 are the coefficients of the equations. Accordingly, the
relationship between the concentrations of GdCl3 and aga-
rose and 1/T2 can be indicated by the following equation:
1
T2
¼ b1 þ b2 Gþ b3 G2 þ ðb4 þ b5 GÞA: (4)
III.D. Formulation of relationship betweenconcentrations of modifiers and relaxation times
Based on all measured data, the coefficients of Eqs. (2)
and (4) were calculated using the nonlinear least-squares
method (Gauss-Newton method). The obtained coefficients
(95% confidence intervals) for Eq. (2) were as follows:
FIG. 4. Relationship between agarose concentration and coefficients of Eq. (1). (a) a1 vs agarose concentration. (b) b1 vs agarose concentration. (c) c1 vs
agarose concentration.
6339 Ikemoto et al.: Development of a human-tissue-like phantom for 3.0-T MRI 6339
Medical Physics, Vol. 38, No. 11, November 2011
a1 ¼ 3:98� 10�4 ½3:92� 10�4; 4:04� 10�4�;
a2 ¼ 9:58� 10�6 ½3:93� 10�6; 1:52� 10�5�;
a3 ¼ 9:94� 10�6 ½9:35� 10�6; 1:05� 10�5�;
a4 ¼ �2:00� 10�8 ½�5:65� 10�7; 5:25� 10�7�;
a5 ¼ 8:46� 10�9 ½�3:39� 10�10; 1:73� 10�8�;
a6 ¼ �9:13� 10�9 ½�2:51� 10�8; 6:84� 10�9�;
a7 ¼ 1:94� 10�9 ½�5:03� 10�9; 8:90� 10�9�: (5)
The obtained coefficients (95% confidence interval) for Eq.
(4) were as follows:
b1 ¼ 3:13� 10�3 ½3:08� 10�3; 3:18� 10�3�;
b2 ¼ 4:67� 10�5 ½4:33� 10�5; 5:01� 10�5�;
b3 ¼ 4:81� 10�8 ½1:61� 10�8; 8:02� 10�8�;
b4 ¼ 1:33� 10�2 ½1:31� 10�2; 1:36� 10�2�;
b5 ¼ �9:20� 10�6 ½�1:56� 10�5;�2:81� 10�6�: (6)
Thus, the empirical formulae for calculating T1 and T2 as a
function of the concentrations of GdCl3 and agarose were
obtained.
III.E. Graphical expression of the empirical formulae
Figure 7 is a graphical expression of the empirical for-
mulae, Eqs. (2) and (4). The solid lines indicate the relaxa-
tion times for iso-agarose concentrations of 0, 0.2, 0.3, 0.4,
0.5, 0.6, 0.7, 0.8, 1.0, 1.2, 1.4, 1.6, 1.8, and 2.0%, while
the broken lines show those for iso-GdCl3 concentrations
of 0, 4, 7, 15, 20, 25, 33, 40, 60, 80, 110, 140, and
180 lmol/kg. The values of X-axis and Y-axis, at the
intersection point of the solid line and broken line, show
the calculated results of the relaxation time. The open
circles indicate the experimental data corresponding to the
concentration.
IV. DISCUSSION
This study resulted in the development of a human-tissue-
like phantom for use with 3.0-T MRI, comprised of carra-
geenan for gelatinization, GdCl3 as a T1-relaxation modifier,
and agarose as a T2-relaxation modifier. Based on the meas-
ured results, empirical formulae were devised to express the
relationships between the concentrations of the relaxation
modifiers and relaxation times.
Figure 7 shows the relationship between the measured val-
ues and the calculated values of the empirical formulae
obtained with the nonlinear least-squares method. There is
much greater error between the representative values among
the larger T1 and T2 values, with smaller error among the
smaller T1 and T2 values. This is due to the relation of the
inverse between relaxation rate and relaxation time. The non-
linear least-squares method was applied to the empirical for-
mulae Eqs. (2) and (4), expressing the relaxation rates; Fig. 7
expresses the relaxation times, the inverse of the relaxation
rates. Mean error, that is, fR[(measured value� calculated
value)/calculated value]/number of samplesg * 100%,
was� 0.2% for T1, and 0.7% for T2. The standard deviation
was 3.8% for T1, and 3.9% for T2.
The developed phantom was found to have T1 values
ranging from 395 to 2601 ms and T2 values ranging from 29
to 334 ms when the GdCl3 concentration varied from 0 to
180 lmol/kg and the agarose concentration varied from 0 to
2.0%, and with all samples having a fixed carrageenan con-
centration of 3%. Based on the concentrations of T1 and T2
relaxation modifiers, we determined formulae by which to
obtain T1 and T2 values for purposes of developing the 3.0-T
MRI phantom, using previously established methods.5–9 It is
of interest to illustrate the relationship between the relaxa-
tion times at 1.5 T and 3.0 T, when the same concentrations
of the same relaxation modifiers are used. In our previous
phantom study, with concentrations of GdCl3 between 0 and
140 lmol/kg and of agarose between 0 and 1.6%, the T1 and
T2 values of the phantom ranged from 202 to 1904 ms and
38 to 423 ms, respectively.5
FIG. 5. T2 relaxation rate relative to agarose concentrations. The open
circles represent the data at 0 lmol/kg of GdCl3. The solid line, plotted as a
linear function, is the least-squares fit to the data indicated by the open
circles. Each broken line, plotted as a linear function, is the least-squares fit
to the data relative to the different GdCl3 concentrations within a range of
4–180 lmol/kg. The value of R2 for each regression line is greater than 0.9.
As the concentration of GdCl3 increases, the broken lines tend to move
away from the x-axis.
6340 Ikemoto et al.: Development of a human-tissue-like phantom for 3.0-T MRI 6340
Medical Physics, Vol. 38, No. 11, November 2011
Figure 8 is a comparison of the relaxation times at 1.5 T
and at 3.0 T, with identical concentrations of relaxation
modifiers. The isoconcentration curves are depicted with
varying concentrations of GdCl3 (0, 4, 7, 15, 20, 25, 33, 40,
60, 80, 110, and 140 lmol/kg) and agarose (0, 0.2, 0.3, 0.4,
0.5, 0.6, 0.7, 0.8, 1.0, 1.2, 1.4, and 1.6%). The solid lines are
graphical expressions of the formulae for 3.0-T MRI, and the
broken lines represent the same for 1.5-T MRI.
The Bloembergen-Purcell-Pound theory (BPP theory)
states that the T1 and T2 of a pure substance depend on a mu-
tual relation between the Larmor frequency of a magnetic
nucleus and a correlation time in which the nucleus is per-
turbed by the magnetic fields associated with neighboring
nuclei with Brownian motion.20 According to the BPP
theory, in a region of the correlation time where T1 is much
greater than T2, T1 becomes much longer when the static
magnetic field increases from 1.5 T to 3.0 T, while T2
becomes only slightly longer. In human-tissues, T1 at 3.0 T
is more prolonged than that at 1.5 T, whereas T2 at 3.0 T is
slightly shorter or slightly longer than that at 1.5 T.2,16,18,19
In our experiment, T1 became much longer, while T2 became
slightly shorter, when the static magnetic field increased
from 1.5 T to 3.0 T.
The maximum concentrations of the relaxation modifiers
for the 3.0-T MRI phantom samples were set higher than
were used in prior reports on the 1.5-T MRI phantom. The
range of GdCl3 concentration was extended from 0–140
lmol/kg to 0–180 lmol/kg due to the prolongation of T1 val-
ues at 3.0 T. The range of agarose concentration was
expanded from 0%–1.6% to 0%–2.0% for 3.0-T to
FIG. 7. Graphical expression of the empirical formulae. The solid lines indi-
cate the relaxation times for iso-agarose concentrations of 0, 0.2, 0.3, 0.4,
0.5, 0.6, 0.7, 0.8, 1.0, 1.2, 1.4, 1.6, 1.8, and 2.0%, while the broken lines are
for iso-GdCl3 concentrations of 0, 4, 7, 15, 20, 25, 33, 40, 60, 80, 110, 140,
and 180 lmol/kg. The open circles indicate the experimental data of T1 and
T2, corresponding to the concentrations.
FIG. 8. Comparison of relaxation times between 1.5-T phantom and 3.0-T
phantom. Solid lines: 3.0-T MRI phantom. Broken lines: 1.5-T MRI phan-
tom. Both the solid and broken lines are depicted within the same ranges of
GdCl3 and agarose concentrations. Iso-concentration curves of GdCl3 are
for 0, 4, 7, 15, 20, 25, 33, 40, 60, 80, 110, and 140 lmol/kg, and those of
agarose are for 0, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8, 1.0, 1.2, 1.4, and 1.6%.
FIG. 6. Relationship between GdCl3 concentrations and the coefficients of Eq. (3). (a) a2 vs GdCl3 concentrations. The curved line, plotted as a quadratic
function, is the least-squares fit to the data. (b) b2 vs GdCl3 concentrations. The straight line is the least-squares fit to the data.
6341 Ikemoto et al.: Development of a human-tissue-like phantom for 3.0-T MRI 6341
Medical Physics, Vol. 38, No. 11, November 2011
compensate for human-tissues having shorter T2 values com-
pared with those at 1.5 T.
The SAR and nonuniformity of B1 in a human body
exposed to 3.0-T MRI are greater than occur with 1.5-T MRI
and are dependent on the frequency of B1 and the dielectric
properties of the body. Conductivity and permittivity vary
with tissue-type, and change relative to the frequency of B1.
The loading property of a phantom to the RF coil depends
on the phantom’s conductivity and permittivity. While the
permittivity of a phantom is difficult to modify, the permit-
tivity of high water-content tissue such as muscle is nearly
the same as that of water at the resonance frequency of 3.0
T. In contrast, a phantom’s conductivity is relatively easy to
modify by changing its concentration of sodium chloride;
phantoms with and without sodium chloride have been pre-
viously developed for 1.5-T MRI.5–7 In future study, we
intend to develop a phantom containing sodium chloride for
3.0-T MRI use. We anticipate that the T1, T2, and conductiv-
ity aspects of such a phantom will each be independently
modifiable.
V. CONCLUSIONS
The human-tissue-like 3.0-T MRI phantom we developed
with carrageenan has high elasticity and great strength, is
easy to produce, and its shape can be easily modified. The T1
and T2 values of this phantom can be arbitrarily adjusted by
changing the concentrations of GdCl3 and agarose, accord-
ing to the developed calculation formulae. The phantom is
believed to be a valuable contribution to research into 3.0-T
high field MRI for optimum use of such systems in clinical
practice.
ACKNOWLEDGMENTS
This work was supported in part by Grants in Aid for Sci-
entific Research from the Ministry of Education, Culture,
Sports, Science and Technology, Japan (#19591418,
#22591334)
a)Author to whom correspondence should be addressed. Electronic mail:
[email protected]; Telephone:þ 81-86-235-6845.1H. Fukatsu, “3T MR for clinical use: Update,” Magn. Reson. Med. Sci. 2,
37–45 (2003).2G. E. Gold, E. Han, J. Stainsby, G. Wright, J. Brittain, and C. Beaulieu,
“Musculoskeletal MRI at 3.0 T: Relaxation times and image contrast,”
Am. J. Roentgenol. 183, 343–351 (2004).3K. J. Chang, I. R. Kamel, K. J. Macura, and D. S. Bluemke, “3.0-T MR
imaging of the abdomen: Comparison with 1.5 T,” RadioGraphics 28,
1983–1998 (2008).
4A. Boss, H. Graf, A. Berger, U. A. Lauer, H. Wojtczyk, C. D. Claussen,
and F. Schick, “Tissue warming and regulatory responses induced by radio
frequency energy deposition on a whole body 3-Tesla magnetic resonance
imager,” J. Magn. Reson. Imaging 26, 1334–1339 (2007).5K. Yoshimura, H. Kato, M. Kuroda, A. Yoshida, A. Hanamoto, A. Tanaka,
M. Tsunoda, S. Kanazawa, K. Shibuya, S. Kawasaki, and Y. Hiraki,
“Development of a tissue-equivalent MRI phantom using carrageenan
gel,” Magn. Reson. Med. 50, 1011–1017 (2003).6H. Kato, K. Yoshimura, M. Kuroda, A. Yoshida, K. Hanamoto, S. Kawa-
saki, K. Shibuya, Y. Yamamoto, M. Tsunoda, M. Takemoto, and Y. Hir-
aki, “Development of a phantom compatible for MRI and hyperthermia
using carrageenan gel-Relationship between dielectric properties and
NaCl concentration,” Int. J. Hyperthermia 20, 529–538 (2004).7A. Yoshida, H. Kato, M. Kuroda, K. Hanamoto, K. Yoshimura, K. Shi-
buya, S. Kawasaki, M. Tsunoda, S. Kanazawa, and Y. Hiraki,
“Development of a phantom compatible for MRI and hyperthermia using
carrageenan gel-Relationship between T1 and T2 values and NaCl concen-
tration,” Int. J. Hyperthermia 20, 803–814 (2004).8H. Kato, M. Kuroda, K. Yoshimura, A. Yoshida, K. Hanamoto, S. Kawa-
saki, K. Shibuya, and S. Kanazawa, “Composition of MRI phantom equiv-
alent to human tissues,” Med. Phys. 32, 3199–3208 (2005).9S. Ohno, H. Kato, T. Harimoto, Y. Ikemoto, K. Yoshitomi, S. Kadohisa,
M. Kuroda, and S. Kanazawa, “Production of a human-tissue-equivalent
MRI phantom: Optimization of material heating,” Magn. Reson. Med. Sci.
7, 131–140 (2008).10G. D. Waiter and M. A. Foster, “Lanthanide-EDTA doped agarose gels for
use in NMR imaging phantoms,” Magn. Reson. Imaging 15, 929–938
(1997).11W. D. D’Souza, E. L. Madsen, O. Unal, K. K. Vigen, G. R. Frank, and
B. R. Thomadsen, “Tissue mimicking materials for a multi-imaging mo-
dality prostate phantom,” Med. Phys. 28, 688–700 (2001).12T. A. J. Hopper, B Vasilic, J. M. Pope, C. E. Jones, C. L. Epstein, H. K.
Song, and F. W. Wehrli, “Experimental and computational analyses of the
effects of slice distortion from a metallic sphere in an MRI phantom,”
Magn. Reson. Imaging 24, 1077–1085 (2006).13A. Kumar and P. A. Bottomley, “Optimized quadrature surface coil
designs,” Magn. Reson. Mater. Phys., Biol., Med. 21, 41–52 (2008).14G. Niculescu, J. L. Nosher, M. D. B. Schneider, and D. J. Foran, “A de-
formable model for tracking tumors across consecutive imaging studies,”
Int. J. Comput. Assist. Radiol. Surg. 4, 337–347 (2009).15E. L. Madsen, M. A. Hobson, G. R. Frank, H Shi, J Jiang, T. J. Hall,
T. Varghese, M. M. Doyley, and J. B. Weaver, “Anthropomorphic breast
phantoms for testing elastography systems,” Ultrasound Med. Biol. 32,
857–874 (2006).16C. M. de Bazelaire, G. D. Duhamel, N. M. Rofsky, and D. C. Alsop, “MR
imaging relaxation times of abdominal and pelvic tissues measured
in vivo at 3.0 T: Preliminary results,” Radiology 230, 652–659 (2004).17J. P. Wansapura, S. K. Holland, R. S. Dunn, and W. S. Ball, Jr.. “NMR
relaxation times in the human brain at 3.0 tesla,” J. Magn. Reson. Imaging
9, 531–538 (1999).18R. Rakow-Penner, B. Daniel, H. Yu, A. Sawyer-Glover, and G. H. Glover,
“Relaxation times of breast tissue at 1.5T and 3T measured using IDEAL,”
J. Magn. Reson. Imaging 23, 87–91 (2006).19G. J. Stanisz, E. E. Odrobina, J. Pun, M. Escaravage, S. J. Graham, M. J.
Bronskill, and R. M. Henkelman, “T1, T2 relaxation and magnetization
transfer in tissue at 3T,” Magn. Reson. Med. 54, 507–512 (2005).20N. Bloembergen, E. M. Purcell, and R. V. Pound, “Relaxation effects in
nuclear magnetic resonance absorption,” Phys. Rev. 73, 679–712 (1948).
6342 Ikemoto et al.: Development of a human-tissue-like phantom for 3.0-T MRI 6342
Medical Physics, Vol. 38, No. 11, November 2011