21
Uppsala University Bachelor project 15hp Vt15 Determination of trace and major elements in oil by ICP-AES after microwave digestion Amalia Lundholm Supervisor: Jean Pettersson

Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

Embed Size (px)

Citation preview

Page 1: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

Uppsala University

Bachelor project 15hp

Vt15

Determination of trace and major elements in oil by

ICP-AES after microwave digestion

Amalia Lundholm

Supervisor: Jean Pettersson

Page 2: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

2

Table of content Abbreviations .......................................................................................................................................... 2

Introduction ............................................................................................................................................. 3

Material and method .............................................................................................................................. 4

Chemicals and glassware ..................................................................................................................... 4

Apparatus ............................................................................................................................................ 4

Calibration curves ................................................................................................................................ 4

Samples and sample preparation ........................................................................................................ 4

ICP-AES measurements ....................................................................................................................... 6

Method validation ............................................................................................................................... 6

Results and discussion ............................................................................................................................. 7

Standard oils ........................................................................................................................................ 7

Lignin oils ............................................................................................................................................. 8

Conclusion and further testing .......................................................................................................... 14

References ............................................................................................................................................. 15

Appendix ................................................................................................................................................ 16

Standard oils ...................................................................................................................................... 16

Lignin oils ........................................................................................................................................... 18

Abbreviations ICP-AES Inductively coupled plasma atomic emission spectroscopy

JL JL166DV, liquid lignin oil

BO 161 botton, liquid lignin oil

RF RAFAP 2e91, solid lignin oil

STDAV Standard deviation

RSD Relative standard deviation

LOQ Limit of quantification

Page 3: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

3

Introduction A company called RenFuel got its laboratories at BMC, Uppsala University. Their business plan is to

convert lignin, which is a by-product in the forest industry, into a biochemical oil that later can be used

as fuel to reduce our dependence on fossil fuel. (RenFuel, 24-05-2015) RenFuel sends their oils for

analysis to the analytical chemistry department at Uppsala University, to measure trace and major

elements in the lignin oils.

Lignin can be found in plant cell walls, as the essential glue between the cellulose microfibrilis. Lignin

is a large aromatic polymer that makes up 15-30% of the plants weight. The paper industry produces

large amount of lignin, but most of it are not being used for anything, but is burned as low value fuel.

(Zakzeski, Bruijnincx, Jongerius, & Weckhuysen, 2010)

Lignin oil or LIGNOL® is produced at atmospheric pressure, below the boiling point with low use of

energy, 1 ton lignin becomes 1 m3 oil. This is made by a catalytic process on the lignin that takes a

couple of hours. (Zakzeski, Bruijnincx, Jongerius, & Weckhuysen, 2010)

The method of analysing Lignin oil currently being used at Uppsala University are open vessel digestion

followed by inductively coupled plasma atomic emission spectroscopy (ICP-AES). Quantitative

measurements of various different elements can be made with sufficient accuracy, but the method is

considered to be too slow. This is because of the open vessel digestion, which takes approximately

four days.

The reason why oil need to be digested prior to analysing it with ICP-AES are so that the large organic

compounds wont clog up the system, also, the sample need to be water soluble for the ICP-

measurements. (Harris, 2010)

Open vessel digestion uses acid and heat to break down the oil, which can take time since the oil isn’t

soluble in the water soluble reagents, and therefore have a small reaction surface. Furthermore, the

temperature can’t be too high since this would lead to losses of the volatile elements, also, the

temperature need to be below the boiling point of the reagents. Microwave digestion can be used as

a closed system, and can therefore use both temperature and pressure to break down the oil. Since

the system is closed there are less risk of contamination and losses of volatile product, which means

that the temperatures can be higher than in an open system, and that the pressure can be varied also

speeds up the digestion process. (Harris, 2010)

Microwave digestion were suggested as a replacement for the open vessel digestion, and several

articles describing the method of microwave digestion to make oil samples ready for ICP-AES analysis

have been published, and the ones that are closest to this project studied different kinds of edible oils.

In one study the authors had determined trace elements in pumpkin oil, and were successful in

detecting Ca, Cd, Co, Cr, Cu, Fe, K, Mg, Mn, Mo, Na, Ni, Pb, Ti, V and Zn, but could not measure Al, Cr,

P, since limit of detection was too high and the concentrations too low. The recoveries for all elements

were >95% with exception for S, that only had a recovery of <50%. They used nitric acid and hydrogen

peroxide for the microwave digestion. (Juranovic, Breinhoelder, & Steffan, 2003)

Other articles that have given good results with the microwave digestion are a study to determine

trace elements in olive oil (Zeinera, Steffana, & Cindric, 2005), a quantitative study on metals in virgin

olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of nickel in

cottonseed oil. (Menga & Zhang, 2011) All of these mentioned articles use nitric acid and hydrogen

peroxide for the digestion.

Page 4: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

4

The theory where that microwave digestion could be used as a replacement for the open vessel

digestion, and after searching after articles on the subject of oil analysis this seemed to be a valid guess.

The purpose of this project where to optimize and validate a method to determine trace and major

elements in Lignin oil using microwave digestion to break down the oil and make it water soluble, and

then make quantitative measurements of the elements using ICP-AES.

Material and method

Chemicals and glassware The chemicals being used during this project where nitric acid (HNO3, 65%) and hydrogen peroxide

(H2O2, 30%). The nitric acid where sub-boiled before being used, to remove traces of metals.

Two different oil analysis standards were used for method validation, CONSTAN Sulphur (0,5% Wt.)

and CONSTAN S-21+K (300 ppm Wt.), a multi-element oil containing Au, Al, B, Ba, Ca, Cd, Cr, Cu, Fe, K,

Mg, Mn, Mo, Na, Ni, P, Pb, Si, Sn, Ti, V and Zn.

To calibrate the ICP-AES calibration solutions were prepared from a multi-element standard solution

(5% HNO3) containing 10 ppm of As, B, Ba, Cd, Co, K, Li, Mg, Mn, Mo, Na, P, Rb, Sr, Ti, V and Y, and 9.7

ppm of Ag, Al, Bi, Ca, Cr, Cu, Fe, Ni, Pb, S, Se, Zn.

Glassware used during the open vessel digestion were volumetric flasks made of quartz glass.

All standards and samples were diluted with Milli-Q water in sterile Polypropylene Conical Falcon

tubes.

Apparatus The ICP-AES used was a SPECTRO CIROSCCD whit a Modified Lichte nebulizer spraychamber and

SPECTRO Smart Analyzer Vision Software.

The microwave digestion was done using a Titan MPSTM 16 position microwave sample preparation

system, with 75mL (40 Bar) TFM digestion vessels.

During the open vessel digestion the samples were heated with a programmable EUROTHERM heater

in an aluminium block.

Calibration curves The ICP regression lines for the different elements were calculated by the smart analyser program from

measurements on standard solutions. Three solutions were prepared from a multi element standard

and diluted with nitric acid and Milli-Q water to 0, 1 and 5 ppm. The amount of nitric acid varied

depending on the acid concentration in the digested and diluted samples. The ICP were recalibrated

with new standard solutions before each set of measurements.

Samples and sample preparation During this project three different Lignin oils have been studied, and in table 1 follows a short

description of the samples.

Table 1: Description of lignin oil samples.

Sample name Abbreviation State Comment

JL166DV JL Liquid Viscous, difficult to work whit

161 Botton BO Liquid More fluid than JL, not as thick and viscous

RFAP 2e91 RF Solid Solid, but soft, pieces could be removed whit spatula

Page 5: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

5

Open vessel digestion

The Glassware used during the open vessel digestion where thoroughly rinsed with Milli-Q water and

nitric acid. Before adding the samples and reagents the glassware where dried in an oven set to 80°C

and then cooled at room temperature.

Approximately 0.2000g of the oils where weighed directly into volumetric flasks. 4.00mL HNO3 and

2.00ml H2O2 where added to each sample, and at the same time two blanks were prepared with the

same amount of HNO3 and H2O2.

The samples got two hours to react in room temperature, then the temperature was raised to 80°C for

a couple of hours. The samples were left over night at 105°C, and then raised again to 150°C. The

solutions were monitored during the days, and if they started to boil to much the flasks were removed

from the aluminium block to cool, and then slowly heated to 150°C again.

The oils were completely digested after approximately four days, but more often than not the samples

were left at 150°C over the weekend, and then found to be digested on Monday, so it is difficult to say

how long time it actually took. The solutions were transferred to 14ml Falcon tubes quantitative, which

means that the flasks were rinsed with an amount Milli-Q water three times, so that more of the oil

could be transferred to the tubes. Finally the oils were diluted with Milli-Q water to 14ml.

After measurements with the ICP, some of the oils were diluted further, because that the sulphur

concentrations were too high, and therefore not in the linear range. 500µl of the sample were

transferred to a 14ml Falcon tube, 5ml HNO3 was added and the solution was diluted with Milli-Q

water.

Microwave digestion

Before using the Tritan MPSTM the digestion vessels had to be cleaned properly, this was done by filling

the vessels with 5ml HNO3 and 3ml H2O2, which is the amount of reagents later being used for digesting

the samples, and then running the temperature program described in table 2. After the program was

done, the vessels were left in the microwave for 20 minutes, so they wouldn’t be too hot. Finally, the

vessels were rinsed with Milli-Q water before adding the samples.

The digestion vessels for the microwave digestion were quite large, 1kg each, and were therefore too

heavy for the analytical scale, so the oils could not be weighed directly into the vessels. Several

different ways of deciding the exact amount of oil were tried. Two methods were most efficient for

the liquid oils, the first were some of the oil was transferred to a plastic, one use only, container, the

oil was weighted and then poured into the vessel, the container was then weighted again to see how

much of the oil that had stuck to the walls, this was subtracted from the result. The other method was

to drop the oil on a filter paper, the filter paper could be weighted at the same time as the oil was

added, which gave more control over how much sample that was put into the digestion vessels.

Approximately 0.2400g of oil was used for each sample. An oil-free paper was added to the blanks.

Then 5ml HNO3 and 3ml H2O2 were added to the oils, and three blanks were prepared using 5ml HNO3

and 3ml H2O2. The vessels were left open for 10-15min before being closed and put into the microwave.

The microwave was programed with the temperature program found in table 2.

After the digestion all samples were transferred to 14ml Falcon tubes and diluted with Milli-Q water

the same way as described for the open vessel digestion. Some of these samples also had a too high

sulphur concentration, and was diluted further for more measurements.

Page 6: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

6

Table 2: Temperature program, microwave digestion

Step Temperature [°C] Pressure [bar] Ramp time [min] Hold time [min] Power [%]

1 170 40 2 5 60

2 190 40 2 10 80

3 210 40 2 15 100

4 50 40 1 10 0

The temperature program was found in the reference notebook of microwave applications, belonging

to the microwave. The pressures was however altered, they had to be reduced since the temperature

program was originally made for the 100ml digestion vessels, not the 75ml that was used during this

project.

ICP-AES measurements Table 3 shows the ICP-AES settings.

Table 3: ICP-AES settings

Coolant Flow 14L/min

Auxiliary Flow 0.9L/min

Nebulizer Flow 0.9l/min

Pump Flow Rate 2.5ml/min

It took approximately three minutes to measure one sample, and that includes preflush. The ICP

measures three times on each sample, before calculating a mean value. The RSD-values for the ICP are,

however, not discussed in this report.

In the SPECTRO Smart Analyzer Vision Software recommended wavelengths for measurements on the

different elements can be found, most elements have been measured with two different wavelengths.

In the results, however, are only one presented. The one that has been selected is less interfered by

other elements, or is proven better at detecting low concentrations.

Method validation In order to analyse the methods ability to give correct values measurements were made on standard

oils with a known concentration of different elements. The open vessel digestion as well as the

microwave digestion was done as described in the sample preparation, but the oils wasn’t weighted

on filter papers for the microwave digestion, instead, the first method was used, where a tube

containing the oil were weighted before and after.

From the data obtained the standard deviation, STDAV, as well as the relative standard deviation, RSD,

could be calculated for each element and method respectively, so that the methods could be

compared. LOQ, limit of quantification, was calculated from the measurements on the blanks as seen

in eq1, and made it easier to see if the measured element concentrations could be distinguished from

the blanks.

𝐿𝑂𝑄 = 3 × 𝑆𝑇𝐷𝐴𝑉(𝑏𝑙𝑎𝑛𝑘𝑠) × 3.3 (𝑒𝑞1)

Page 7: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

7

Results and discussion

Standard oils In table 4 below are the results from the ICP-AES measurements made on the multi-element standard

oil. Two replicate samples was made with the open vessel digestion and three replicate samples was

made with the microwave digestion, table 3 shows the mean value of the calculated concentrations.

In appendix A1 the original measurements for all the replicates are presented, before they were

recalculated with regard to the dilution and weight of the oil sample. The recalculated values can be

found in appendix A2.

Table 4: Mean values of ICP-AES measurements on multi-element oils (300ppm), using the two different digestion methods, microwave digestion (Micro) and open vessel digestion (Open).

Element Wavelength [nm]

Al 396.152

Ag 338.289

Ba 455.404

Ca 393.366

K 766.491

Mg 279.553

Mo 202.030

MV Micro [ppm] 276 302 298 325 189 333 304

MV Open [ppm] 289 326 326 341 210 346 288

Element Wavelength [nm]

Na 589.592

Ni 231.604

Pb 220.353

Zn 213.856

Cu 324.754

Fe 259.941

Cr 267.716

MV Micro [ppm] 228 323 313 290 322 318 309

MV Open [ppm] 245 344 328 322 330 273 314

Element Wavelength [nm]

B 208.959

Cd 214.438

Mn 257.611

P 177.495

Ti 334.941

V 292.464

MV Micro [ppm] 252 296 334 320 318 307

MV Open [ppm] 282 319 342 207 149 310

The result are also presented in the bar chart below, figure 1. The black line shows 300ppm.

Figure 1: Mean values of ICP-AES measurements on multi-element oils (300ppm), using the two different digestion methods, microwave digestion (Micro) and open vessel digestion (Open).

It seems that the microwave digestion gives good recovery for all elements except potassium and

sodium. Boron also seems to be a tricky element. The concentrations for the other elements only

deviates from 300ppm with a few %. The open vessel digestion have similar results compared to the

microwave digestion, except for phosphorus and titanium that gives low recoveries.

0

50

100

150

200

250

300

350

400

[pp

m]

Micro Open

Page 8: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

8

The reason why sodium and potassium can be difficult to measure in oil often depends on that they

are easily ionized, which means that the ICP can’t decide the correct value. (Harris, 2010)

The measurements after microwave digestion was made on three samples, but the open vessel

digestion only had two. The results therefore have a degree of uncertainty, especially for the open

vessel digestion. By looking in appendix A2 one can see that the deviation between some of the

replicates are quite large, as for example Ba that in the first sample, M1, had a concentration of

326ppm, and a concentration of 384ppm for the second sample, M2.

But the overall trend appears to be that the open vessel digestion give slightly larger values than the

microwave digestion. The reason for this is difficult to say, it might be a coincidence, since some of the

values for the open vessel digestion differs so much for some elements. More tests are needed.

More measurements needs to be done to be sure, but it seems like the microwave digestion gives

better results, both in recovery, but also better standard deviation for the replicates.

The results from measurements on the standard oil with 5000ppm sulphur can be found in table 5

below. For the full data, se appendix A3 and A4.

Table 5: Mean values of ICP-AES measurements on standard oil with 5000ppm sulphur, using the two different digestion methods, microwave digestion (Micro) and open vessel digestion (Open).

Element Wavelength [nm]

S 182.034

Micro S [ppm] 5858

Open S [ppm] 6542

The results from this measurements are high, both for the microwave digestion and the open vessel

digestion. This probably depends on that it was difficult to get good regression lines for sulphur during

the ICP-calibration. But also here can a higher value for the open vessel digestion be seen.

Lignin oils Even for the samples, only elements present in the multi-element oil can be found in the results, i.e.

Au, Al, B, Ba, Ca, Cd, Cr, Cu, Fe, K, Mg, Mn, Mo, Na, Ni, P, Pb, Si, Sn, Ti, V and Zn.

In the following three tables 6, 7 and 8, ICP-results after the microwave digestion on the three lignin

oils can be found. Only elements that gave signals over the limit of quantification, LOQ, and therefore

was distinguishable from the blanks, are presented. The concentrations have been recalculated to take

account of the dilution and weight of the oil sample. For all the measurements, see appendix A5, A6

and A7.

Page 9: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

9

Table 6: Oil JL166DV, microwave digestion on three replicates. Mean value (MV), standard deviation (STDEV) and relative standard deviation (RSD).

Element Wavelength [nm]

Mg 279.553

Zn 213.856

Cr 267.716

B 208.959

Cd 214.438

Micro JL1 [ppm] 0.131 0.113 0.050 0.925 0.029

Micro JL2 [ppm] 0.057 0.076 0.084 0.847 0.021

Micro JL3 [ppm] 0.040 0.167 0.060 0.681 0.028

MV JL [ppm] 0.076 0.119 0.065 0.818 0.026

STDEV [ppm] 0.05 0.05 0.02 0.12 0.004

RSD [%] 64 39 27 15 17

Element Wavelength [nm]

P 177.495

Ti 334.941

V 292.464

S 182.034

Micro JL1 [ppm] 8.13 0.139 0.014 834

Micro JL2 [ppm] 4.74 0.101 0.026 832

Micro JL3 [ppm] 5.71 0.133 -0.010 805

MV JL [ppm] 6.19 0.124 0.010 824

STDEV [ppm] 1.7 0.02 0.02 16

RSD [%] 28 16 183 1.9

For the JL166DV oil, only nine out of twenty elements gave a high enough concentration-signal to be

distinguished from the blanks, but the relative standard deviation is quite high for all the elements.

This is probably because the concentrations are quite low, which gives high uncertainty while

measuring with the ICP-AES.

Magnesium and vanadium have really high RSD-values, this is due to that the magnesium

concentration measured in JL1 is too high and the vanadium concentration measured in JL3 is too low,

relative the other replicates. They can probably be seen as outliners, but to be sure more

measurements needs to be made.

Table 7: Oil 161 Botton, microwave digestion on three replicates. Mean value (MV), standard deviation (STDEV) and relative standard deviation (RSD).

Element Wavelength [nm]

Al 396.152

Mg 279.553

Mo 202.030

Na 589.592

Ni 231.604

Zn 213.856

Fe 259.941

Micro BO1 [ppm] 55.9 0.435 0.202 8.01 2.33 0.656 0.921

Micro BO2 [ppm] 0.543 0.312 0.068 7.60 2.35 0.457 0.679

Micro BO3 [ppm] 0.515 0.472 0.092 9.03 2.71 0.806 1.03

MV BO [ppm] 19.0 0.406 0.121 8.21 2.46 0.639 0.878

STDEV [ppm] 32 0.08 0.07 0.74 0.21 0.17 0.18

RSD [%] 168 21 59 9.0 8.6 27 21

Element Wavelength [nm]

Cr 267.716

B 208.959

Cd 214.438

P 177.495

Ti 334.941

V 292.464

S 182.034

Micro BO1 [ppm] 0.082 1.18 0.038 20.9 0.333 0.553 4814

Micro BO2 [ppm] 0.085 0.831 0.041 21.3 0.172 0.504 4522

Micro BO3 [ppm] 0.114 1.32 0.042 17.2 0.237 0.502 4758

MV BO [ppm] 0.093 1.11 0.040 19.8 0.247 0.519 4698

STDEV [ppm] 0.02 0.25 0.002 2.3 0.08 0.03 155

RSD [%] 19 23 4.9 11 33 5.6 3.3

For the 161 botton oil more elements could be quantified, the concentrations are higher as well, which

probably is the main reason for the better RSD-results. The high RSD-values for aluminium and

Page 10: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

10

molybdenum are due to the high values measured in the first replicate, and are marked in blue in the

table, they can probably be seen as outliners if more measurements were to be made.

Table 8: Oil RAFAP 2e91, microwave digestion on three replicates. Mean value (MV), standard deviation (STDEV) and relative standard deviation (RSD).

Element Wavelength [nm]

Al 396.152

Ba 455.404

Ca 393.366

K 766.491

Mg 279.553

Mo 202.030

Micro RF1 [ppm] 3.15 5.69 191 37.0 9.09 1.14

Micro RF2 [ppm] 2.34 5.74 191 43.3 8.92 1.14

Micro RF3 [ppm] 3.07 5.44 223 39.3 9.17 1.09

MV RF [ppm] 2.85 5.62 202 39.9 9.06 1.12

STDEV [ppm] 0.45 0.16 18 3.2 0.13 0.03

RSD [%] 16 2.9 9.1 8.0 1.4 2.5

Element Wavelength [nm]

Na 589.592

Zn 213.856

Cu 324.754

Fe 259.941

Cr 267.716

B 249.773

Micro RF1 [ppm] 479 11.3 0.09 3.26 0.655 17.7

Micro RF2 [ppm] 507 10.9 0.55 3.49 0.592 17.3

Micro RF3 [ppm] 486 10.3 0.73 3.57 0.868 17.1

MV RF [ppm] 491 10.8 0.46 3.44 0.705 17.4

STDEV [ppm] 14 0.52 0.33 0.16 0.14 0.31

RSD [%] 2.9 4.8 72 4.6 21 1.8

Element Wavelength [nm]

Cd 214.438

Mn 257.611

P 177.495

Ti 334.941

V 292.464

S 182.034

Micro RF1 [ppm] 0.276 8.65 53.3 0.171 13.7 7116

Micro RF2 [ppm] 0.258 8.54 52.8 0.079 13.5 6348

Micro RF3 [ppm] 0.303 8.73 59.4 0.176 13.3 7107

MV RF [ppm] 0.279 8.64 55.2 0.142 13.5 6857

STDEV [ppm] 0.02 0.10 3.6 0.05 0.18 441

RSD [%] 8.1 1.1 6.6 38 1.3 6.4

For the solid oil RAFAP 2e91 it was only nickel and lead that gave signals below LOQ, and the RSD-

values looks better.

In table 9 and 10 below, are the results from the ICP-AES measurements made with the open vessel

digestion. The concentrations have been recalculated to take into account the dilution and weight of

the oil sample. For all the measurements, see appendix A8, A9 and A10.

Table 9: Oil JL166DV, open vessel digestion on three replicates. Mean value (MV), standard deviation (STDEV) and relative standard deviation (RSD).

Element Wavelength [nm]

Na 589.592

Fe 259.941

P 177.495

Ti 334.941

S 182.034

Open JL1 [ppm] 5.57 0.023 18.4 0.130 1326

Open JL2 [ppm] 5.38 7.18 54.2 50.2 1374

Open JL3 [ppm] 4.49 0.707 28.4 33.7 1793

MV JL [ppm] 5.14 2.64 33.7 28.0 1498

STDEV [ppm] 0.6 4.0 18 26 257

RSD [%] 11 150 55 91 17

A different set of elements could be quantified in JL166DV when using the open vessel digestion

compared with the microwave digestion. Only phosphorus, titanium and sulphur could be quantified

in both. The values are lower for the microwave, but the RSD, however, are better. This becomes

Page 11: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

11

strange when the results are compared to the measurements on the multi-element oil, since they gave

too low values on both phosphor and titanium for the open vessel digestion, but good recovery with

the microwave digestion.

The 161 botton oil sample weren’t analysed using the open vessel digestion, but the results from the

solid oil RAFAP 2e91 can be found in table 10 below.

Table 10: Oil RAFAP 2e91, open vessel digestion on three replicates. Mean value (MV), standard deviation (STDEV) and relative standard deviation (RSD).

Element Wavelength [nm]

Al 396.152

Ba 455.404

Ca 393.366

K 766.491

Mg 279.553

Mo 202.030

Na 589.592

Open RF1 [ppm] 5.93 11.9 281 62.5 13.0 0.609 795

Open RF2 [ppm] 6.91 9.66 234 82.1 15.2 8.11 946

Open RF3 [ppm] 5.72 13.7 309 71.9 14.4 0.326 891

MV RF [ppm] 6.19 11.8 275 72.2 14.2 3.01 877

STDEV [ppm] 0.63 2.0 38 9.8 1.1 4.4 77

RSD [%] 10 17 14 14 7.7 146 8.7

Element Wavelength [nm]

Ni 231.604

Zn 213.856

Cu 324.754

Fe 259.941

Cr 267.716

B 208.959

Cd 214.438

Open RF1 [ppm] 0.315 17.8 0.510 6.17 0.396 26.6 0.182

Open RF2 [ppm] 0.371 21.6 0.649 10.9 1.90 32.5 0.185

Open RF3 [ppm] 0.273 20.1 0.544 6.80 0.467 29.4 0.242

MV RF [ppm] 0.320 19.8 0.568 7.94 0.920 29.5 0.203

STDEV [ppm] 0.05 1.9 0.07 2.5 0.85 2.9 0.03

RSD [%] 15 9.6 13 32 92 10 17

Element Wavelength [nm]

Mn 257.611

P 177.495

Ti 334.941

V 292.464

S 182.034

Open RF1 [ppm] 12.6 72.1 1.30 20.3 17351

Open RF2 [ppm] 14.4 116 15.5 25.1 15945

Open RF3 [ppm] 14.3 79.3 -0.044 23.1 16781

MV RF [ppm] 13.8 89.0 5.57 22.9 16692

STDEV [ppm] 1.0 23 8.6 2.4 707

RSD [%] 7.4 26 154 10 4.2

All elements except for lead could be found using the open vessel digestion, but compared to the

microwave digestion, the open vessel digestion gives higher values on the standard deviation as well

as the RSD. For some of the elements, Mo, Cr and Ti, the RSD-values can be blamed on outliners, which

are marked in blue in the table. But the RSD are also higher for almost all the other elements as well.

The measured concentrations are higher for the open vessel compared to the microwave digestion,

but this will be discussed more below.

In figure 2, 3 and 4 below are bar charts comparing the mean values of the elements for the microwave

digestion and the open vessel digestion. The standard deviation-values are presented as well. The

reason that the results are divided in three different bar charts is because the concentrations of the

elements vary. Figure 3 shows the highest concentration, figure 4 the lowest and figure 2 the rest.

Page 12: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

12

Figure 2: Oil RAFAP 2e91, ICP-AES measurements made after microwave digestion and open vessel digestion, with standard deviation for the methods on respective element.

Figure 3: Oil RAFAP 2e91, ICP-AES measurements made after microwave digestion and open vessel digestion, with standard deviation for the methods on respective element. Elements with relative high concentrations.

-5

0

5

10

15

20

25

30

35

Al 396,152 Ba 455,404 Mg 279,553 Zn 213,856 Fe 259,941 B 249,773 Mn 257,611 Ti 334,941 V 292,464

MV Micro RF MV Open RF

0

200

400

600

800

1000

1200

Ca 393,366 K 766,491 Na 589,592 P 177,495

MV Micro RF MV Open RF

Page 13: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

13

Figure 4: Oil RAFAP 2e91, ICP-AES measurements made after microwave digestion and open vessel digestion, with standard deviation for the methods on respective element. Elements with relative low concentrations.

In the three figures it is easy to see the trend that concentrations measured after open vessel digestion

are higher than those measured after microwave digestion. But what also can be seen is that the

microwave digestion gives lower standard deviation for the elements, which means that the

microwave digestion gives better repeatability. For the low-concentration elements the uncertainty of

the values are considerably larger, especially for the open vessel digestion. Which probably is due to

that the low concentrations, almost as low as the blank, give less accuracy for the results.

It is difficult to say why the concentrations are systematically lower after microwave digestion

compared with the open vessel digestion. Since it is for all elements, one can rule out contamination

of the open vessel. Losses due to vaporization in the microwave bombs are also quite unlikely, since it

is a closed system, but there could still be losses in the sense that the oil have not been completely

digested. If the sample did not have enough time to digest, this would mean that some elements are

still bound in molecules and complexes, and could not be measured by the ICP.

The multi-element standard oils did, on the other hand, not seem to suffer from this problem, since

the recovery for the most elements were good, see results from standard oils. However, one credible

guess is that the lignin oils are more difficult to digest than the standard oil, since the lignin polymer

probably is bigger than the oil-molecule present in the multi-standard, which would make it more

difficult to break down.

-2

-1

0

1

2

3

4

5

6

7

8

Mo 202,030 Cu 324,754 Cr 267,716 Cd 214,438

MV Micro RF MV Open RF

Page 14: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

14

Table 11 below shows the final compilation of the results.

Table 11: LOQ-values for the microwave digestion and open vessel digestion. Mean values from measurements on three different kinds of lignin oil after microwave digestion, and two oils after open vessel digestion.

Element Wavelength [nm]

Al 396.152

Ba 455.404

Ca 393.366

K 766.491

Mg 279.553

Mo 202.030

Na 589.592

Micro LOQ [ppm] 0.187 0.103 0.052 0.204 0.004 0.004 0.324

Micro JL [ppm] <0.187 <0.103 <0.052 <0.204 0.076 <0.004 <0.324

Micro BO [ppm] 0.529 <0.103 <0.052 <0.204 0.406 0.121 8.21

Micro RF [ppm] 2.85 5.62 202 39.9 9.06 1.12 491

Open LOD [ppm] 0.054 0.017 0.025 0.556 0.010 0.064 0.307

Open JL [ppm] <0.054 <0.017 <0.025 <0.556 <0.010 <0.064 5.14

Open RF [ppm] 6.19 11.8 275 72.2 14.2 3.01 877

Element Wavelength [nm]

Ni 231.604

Pb 220.353

Zn 213.856

Cu 324.754

Fe 259.941

Cr 267.716

B 208.959

Micro LOQ [ppm] 0.020 0.089 0.063 0.028 0.101 0.002 0.044

Micro JL [ppm] <0.020 <0.089 0.119 <0.028 <0.101 0.065 0.818

Micro BO [ppm] 2.46 <0.089 0.639 <0.028 0.878 0.093 1.11

Micro RF [ppm] <0.020 <0.089 10.81 0.458 3.44 0.705 17.4

Open LOD [ppm] 0.013 0.067 0.245 0.543 0.017 0.014 0.119

Open JL [ppm] <0.013 <0.067 <0.245 <0.543 2.64 <0.014 <0.119

Open RF [ppm] 0.320 0.121 19.8 0.568 7.94 0.920 29.5

Element Wavelength [nm]

Cd 214.438

Mn 257.611

P 177.495

Ti 334.941

V 292.464

S 182.034

Micro LOQ [ppm] 0.002 0.002 0.279 0.009 0.004 3.33

Micro JL [ppm] 0.026 <0.002 6.19 0.124 0.020 824

Micro BO [ppm] 0.040 <0.002 19.8 0.247 0.519 4698

Micro RF [ppm] 0.279 8.64 55.2 0.142 13.5 6857

Open LOD [ppm] 0.014 0.013 0.108 0.014 0.023 1.523

Open JL [ppm] <0.014 <0.013 33.7 28.0 <0.023 1498

Open RF [ppm] 0.203 13.8 89.0 5.57 22.9 16692

As can be seen in table 11, the LOQ-values are for the most part lower for the microwave digestion

compared with the open vessel, this is due to the lower standard deviation for the microwave

Conclusion and further testing Using microwave digestion as sample preparation shows good promise, since it gives good

repeatability, and low RSD- and LOQ-values, but more tests are needed, especially to decide the reason

for the elevated concentrations measured with the open vessel digestion compared with the

microwave digestion.

A recovery-test on the lignin oils using the method of spiking the sample with a known amount of an

element could give useful information about the microwave digestion.

If the multi-element standard oil were to be tested again, an ionization suppressor could be added, to

see if recovery for sodium and potassium can get better. CsCl could work, since Cs is more easily ionized

than K. (Harris, 2010)

Page 15: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

15

References Cindric, I. J., Zeiner, M., & Steffan, I. (2007). Trace elemental characterization of edible oils by ICP–AES

and GFAAS. Microchemical Journal, 85(1):136-139.

Harris, D. C. (2010). Quantitative Chemical Analysis, Eighth Edition. USA: W.H. Freeman and Company.

Juranovic, I., Breinhoelder, P., & Steffan, I. (2003). Determination of trace elements in pumpkin seed

oils and pumpkin seeds by ICP-AES. Journal of analytical atomic spectrometry, 18(1):54-58.

Llorent-Martı´nez, E. J., Co´rdova, M. L.-d., Ortega-Barrales, P., & Ruiz-Medina, A. (2014). Quantitation

of Metals During the Extraction of Virgin Olive Oil from Olives Using ICP-MS after Microwave-

assisted Acid Digestion. Journal of the American Oil Chemists' Society, 91(10):1823-1830.

Mello, P. d., Pereira, J. S., Moraes, D. P., Dressler, V. L., Flores, E. M., & Knap, G. (2009). Nickel,

vanadium and sulfur determination by inductively coupled plasma optical emission

spectrometry in crude oil distillation residues after microwave-induced combustion. Journal of

analytical atomic spectrometry, 24(7):911-916.

Menga, Z., & Zhang, N. (2011). Rapid Analysis of Trace Nickel in Hydrogenated Cottonseed Oil by

Microwave Digestion Prior to Its Inductively Coupled Plasma Mass Spectrometry

Determination. Spectroscopy letters, 45(4):296-300.

RenFuel. (24-05-2015). RenFuel. Retrieved from http://renfuel.se/

Zakzeski, J., Bruijnincx, P. C., Jongerius, A. L., & Weckhuysen, B. M. (2010). The Catalytic Valorization of

Lignin for the Production of Renewable Chemicals. Chemical reviews, 110(6):3552–3599.

Zeinera, M., Steffana, I., & Cindric, I. J. (2005). Determination of trace elements in olive oil by ICP-AES

and ETA-AAS:. Microchemical Journal, 82(2):171-176.

Page 16: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

16

Appendix

Standard oils A 1: ICP-AES measurements made on multi-element standard oil after open vessel digestion and microwave digestion.

Al

167.078 Al

396.152 Ag

328.068 Ag

338.289 Ba

455.404 Ba

233.527 Ca

396.847 Ca

393.366 K

766.491 Mg

279.553

Open Blank 1 -0.05055 0.023031 0.000489 -0.00263 -0.01576 -0.01146 -0.08765 -0.12001 0.12255 -0.03394

Open Blank 2 -0.05072 0.021304 -0.00089 -0.00421 -0.01581 -0.0117 -0.08786 -0.1202 0.08442 -0.03444

Open M1 1.354681 1.193474 1.332566 1.333714 1.31388 1.26493 1.318221 1.334951 0.99131 1.39177

Open M2 1.369916 1.224087 1.337959 1.333038 1.57232 1.5262 1.310461 1.337879 0.94043 1.38559

Micro Blank 1 -0.04559 0.02619 -0.00064 -0.0046 -0.01578 -0.01183 -0.09021 -0.12279 -0.03906 -0.03516

Micro Blank 2 0.03856 0.0869 -0.00158 -0.0056 -0.01587 -0.01199 -0.09113 -0.12375 -0.06741 -0.03531

Micro M1 2.397804 2.042126 1.551448 1.556547 1.52032 1.48296 1.583912 1.554438 0.93438 1.6873

Micro M2 1.520471 1.32971 1.417117 1.388987 1.36853 1.32618 1.413593 1.387732 0.82646 1.51586

Micro M3 1.667136 1.444374 1.517623 1.510682 1.47979 1.44752 1.542816 1.561219 0.88257 1.61934

Mg

280.270 Mo

202.030 Mo

281.615 Na

589.592 Na

588.995 Ni

231.604 Pb

220.353 Pb

168.215 Zn

213.856 Zn

206.191

Open Blank 1 -0.02233 -0.00853 0.00723 0.09836 0.08151 -0.02251 -0.01279 -0.01895 -0.01326 -0.03359

Open Blank 2 -0.02277 -0.01221 0.00294 0.09426 0.07649 -0.0225 -0.01312 -0.01974 -0.01339 -0.03391

Open M1 1.31143 1.11661 1.10095 1.05961 1.09167 1.390384 1.322199 1.274455 1.335975 1.343646

Open M2 1.32127 1.23203 1.2057 1.05475 1.07741 1.392609 1.497671 1.457599 1.278513 1.306465

Micro Blank 1 -0.02344 -0.00859 0.00889 0.06467 0.04536 -0.02269 -0.01124 -0.02164 -0.01185 -0.03245

Micro Blank 2 -0.02357 -0.01255 0.00514 0.06501 0.0455 -0.02235 -0.01047 -0.0193 -0.01338 -0.03422

Micro M1 1.60309 1.55943 1.53398 1.18621 1.21984 1.642931 1.612772 1.566861 1.490223 1.559947

Micro M2 1.43132 1.39078 1.35889 1.07439 1.10152 1.468171 1.445603 1.384644 1.323943 1.378984

Micro M3 1.53926 1.51372 1.48615 1.16164 1.18571 1.602333 1.53996 1.480193 1.435956 1.495491

Cu

324.754 Cu

327.396 Fe

259.941 Fe

238.204 Cr

267.716 Cr

205.552 B

249.773 B

208.959 Cd

214.438 Cd

226.502

Open Blank 1 -0.01322 -0.01326 -0.01079 -0.01649 -0.00462 -0.01711 -0.02593 -0.02785 -0.00826 -0.00859

Open Blank 2 -0.01322 -0.01351 -0.0115 -0.01724 -0.00484 -0.01693 -0.02819 -0.03047 -0.0083 -0.00873

Open M1 1.32664 1.296365 1.102427 1.095059 1.281592 1.265841 1.13585 1.13601 1.31511 1.31444

Open M2 1.359869 1.337559 1.420569 1.427778 1.29188 1.282856 1.14683 1.12487 1.28441 1.30333

Micro Blank 1 -0.01302 -0.01265 -0.00957 -0.01557 -0.00418 -0.01701 -0.02788 -0.03291 -0.00825 -0.00899

Micro Blank 2 -0.01304 -0.01351 -0.01095 -0.01661 -0.00468 -0.01696 -0.03122 -0.03619 -0.00836 -0.00895

Micro M1 1.636292 1.611535 1.639505 1.639104 1.59791 1.576926 1.31125 1.27003 1.53544 1.52412

Micro M2 1.48733 1.470547 1.458764 1.457403 1.416031 1.395358 1.17948 1.13531 1.35695 1.35597

Micro M3 1.605359 1.579356 1.582551 1.587997 1.539399 1.507816 1.27879 1.2261 1.46072 1.47499

Mn

257.611 Mn

259.373 P

177.495 P

178.287 Ti

334.941 Ti

336.121 V

292.464 V

311.071

Open Blank 1 -0.01938 -0.01579 0.00607 -0.04315 -0.02109 -0.01928 -0.00692 -0.00175

Open Blank 2 -0.01948 -0.01588 0.00576 -0.04222 0.03646 0.03839 -0.00695 -0.00115

Open M1 1.38297 1.37093 0.64194 0.49628 0.304 0.3033 1.24085 1.27415

Open M2 1.3902 1.37536 0.86392 0.78089 0.62387 0.62511 1.2881 1.32598

Micro Blank 1 -0.0194 -0.01575 0.00117 -0.03563 -0.02083 -0.01866 -0.00737 -0.00125

Micro Blank 2 -0.0195 -0.01593 0.00104 -0.03667 -0.02104 -0.01894 -0.00747 -0.0011

Micro M1 1.69817 1.68153 1.63863 1.70861 1.61297 1.62395 1.58114 1.64054

Micro M2 1.53892 1.52226 1.50326 1.57182 1.45942 1.46007 1.40152 1.49336

Micro M3 1.65254 1.64156 1.59184 1.68437 1.56534 1.56946 1.53479 1.58511

Page 17: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

17

A 2: ICP-AES measurements made on multi-element standard oil after open vessel digestion and microwave digestion, recalculated by subtracting the blank and taking dilution and weight of the sample into account.

Al

167.078 Al

396.152 Ag

328.068 Ag

338.289 Ba

455.404 Ba

233.527 Ca

396.847 Ca

393.366 K

766.491 Mg

279.553

Open M1 344.95 287.51 327.14 328.21 326.38 313.33 345.11 357.16 217.92 350.01

Open M2 343.13 290.32 323.23 322.82 383.60 371.44 337.73 352.17 202.16 342.94

Micro M1 466.64 385.85 301.70 303.47 298.51 290.49 325.41 326.02 191.92 334.73

Micro M2 325.92 272.28 303.30 298.14 296.06 286.16 321.70 323.14 188.13 331.71

Micro M3 335.88 279.02 305.33 304.74 300.69 293.41 328.40 338.66 188.14 332.64

MV Open 344.04 288.91 325.18 325.51 326.38 313.33 341.42 354.66 210.04 346.48

MV Micro 330.90 275.65 303.45 302.12 298.42 290.02 325.17 329.27 189.40 333.03

Mg

280.270 Mo

202.030 Mo

281.615 Na

589.592 Na

588.995 Ni

231.604 Pb

220.353 Pb

168.215 Zn

213.856 Zn

206.191

Open M1 327.44 276.63 268.99 236.45 248.57 346.81 327.72 317.57 331.20 338.09

Open M2 324.59 300.10 290.00 231.51 241.16 341.82 364.88 356.75 312.04 323.72

Micro M1 316.09 305.09 296.73 217.91 228.22 323.64 315.51 308.46 292.04 309.61

Micro M2 311.13 299.69 289.11 215.90 225.85 318.80 311.48 300.50 285.83 302.03

Micro M3 314.19 306.45 297.37 220.51 229.25 326.67 311.78 301.70 291.23 307.36

MV Open 326.02 288.36 279.50 233.98 244.87 344.31 327.72 317.57 321.62 330.91

MV Micro 313.80 303.74 294.40 218.11 227.77 323.03 312.92 303.55 289.70 306.34

Cu

324.754 Cu

327.396 Fe

259.941 Fe

238.204 Cr

267.716 Cr

205.552 B

249.773 B

208.959 Cd

214.438 Cd

226.502

Open M1 328.88 321.49 273.34 272.93 315.74 314.89 285.45 286.00 324.84 324.77

Open M2 331.66 326.31 345.82 348.95 313.19 313.98 283.55 278.75 312.24 316.91

Micro M1 320.51 315.70 320.59 321.65 311.38 309.74 260.55 253.51 299.99 297.92

Micro M2 320.86 317.29 314.16 315.12 303.78 302.04 258.56 250.18 291.97 291.90

Micro M3 325.37 320.15 320.23 322.49 310.38 306.55 263.04 253.45 295.34 298.34

MV Open 330.27 323.90 273.34 272.93 314.47 314.44 284.50 282.38 318.54 320.84

MV Micro 322.25 317.71 318.33 319.75 308.51 306.11 260.72 252.38 295.77 296.06

Mn

257.611 Mn

259.373 P

177.495 P

178.287 Ti

334.941 Ti

336.121 V

292.464 V

311.071

Open M1 344.23 340.39 156.12 132.29 72.73 72.10 306.28 313.11

Open M2 340.49 336.04 207.25 198.93 148.84 148.68 312.81 320.64

Micro M1 333.78 329.84 318.21 339.05 317.51 319.23 308.70 319.03

Micro M2 333.27 328.93 321.25 343.88 316.59 316.27 301.31 319.62

Micro M3 336.15 333.21 319.81 345.90 318.91 319.31 310.05 318.92

MV Open 342.36 338.22 207.25 198.93 148.84 148.68 309.55 316.87

MV Micro 334.40 330.66 319.76 342.94 317.67 318.27 306.69 319.19

A 3: ICP-AES measurements made on sulphur standard oil after open vessel digestion and microwave digestion.

S

180.731 S

182.034

Open S1 2.9667 2.7664

Open S2 3.0489 2.9147

Micro S1 2.6935 2.5602

Micro S2 2.6585 2.4907

Micro S3 2.6744 2.4709

Page 18: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

18

A 4: ICP-AES measurements made on sulphur standard oil after open vessel digestion and microwave digestion, recalculated by subtracting the blank and taking dilution and weight of the sample into account.

S 180.731 S 182.034

Open S1 6965.8 6584.0

Open S2 6716.8 6500.8

Micro S1 6147.7 5958.9

Micro S2 6049.6 5783.9

Micro S3 6182.5 5831.2

Lignin oils A 5: ICP-AES measurements made on JL166DV, 161 botton and RAFAP 2e91 after microwave digestion.

Al

167.078 Al

396.152 Ba

455.404 Ba

233.527 Ca

396.847 Ca

393.366 K

766.491 Mg

279.553 Mg

280.270 Mo

202.030

Micro Blank 1 0.04551 0.11631 -0.02243 -0.02199 0.09291 0.09068 -0.06535 -0.0371 -0.00672 -0.00362

Micro Blank 2 -0.01079 0.07852 -0.03462 -0.03412 0.09393 0.09227 -0.10661 -0.03766 -0.00723 -0.01167

Micro Blank 3 0.0143 0.09616 -0.04314 -0.04268 0.10254 0.10102 -0.08757 -0.03795 -0.00751 -0.01315

Micro JL1 0.0123 0.0972 -0.05105 -0.04807 0.10604 0.10504 -0.14206 -0.03531 -0.00496 -0.01317

Micro JL2 -0.00985 0.07746 -0.05756 -0.05547 0.07919 0.07475 -0.17752 -0.0366 -0.00615 -0.01299

Micro JL3 0.05474 0.12333 -0.06268 -0.06062 0.08267 0.07966 -0.19077 -0.03689 -0.00651 -0.01352

Micro BO1 1.43296 1.05342 -0.07119 -0.06996 0.08404 0.08 -0.1656 -0.03013 -0.00002 -0.00603

Micro BO2 0.00798 0.10613 -0.07449 -0.07403 0.07065 0.06739 -0.17426 -0.03233 -0.00227 -0.00833

Micro BO3 0.01228 0.10578 -0.07705 -0.07465 0.26512 0.26911 -0.1655 -0.02953 0.00059 -0.00791

Micro RF1 0.06643 0.14893 0.06032 0.06709 3.24369 3.20804 0.52361 0.1123 0.13356 0.00935

Micro RF2 0.03458 0.13637 0.0632 0.06464 3.312 3.33006 0.64295 0.11248 0.13267 0.00969

Micro RF3 0.05837 0.14789 0.0569 0.06446 3.79593 3.80191 0.56588 0.11457 0.13365 0.00864

Mo

281.615 Na

589.592 Na

588.995 Ni

231.604 Pb

220.353 Pb

168.215 Zn

213.856 Zn

206.191 Cu

324.754 Cu

327.396

Micro Blank 1 0.00327 0.20564 0.19244 -0.01288 -0.02833 -0.22374 0.02459 0.00775 0.00273 0.00251

Micro Blank 2 -0.00246 0.23229 0.22058 -0.01594 -0.03901 -0.31529 0.01304 -0.00398 0.00298 0.00289

Micro Blank 3 -0.00417 0.27076 0.26332 -0.01677 -0.04627 -0.34014 0.01433 -0.00268 0.00768 0.00723

Micro JL1 -0.00387 0.20229 0.18856 -0.02015 -0.04392 -0.46199 0.01927 0.00213 0.00023 -0.00072

Micro JL2 -0.00511 0.19657 0.18154 -0.01374 -0.0478 -0.38476 0.01862 0.00095 -0.00074 -0.00126

Micro JL3 -0.00409 0.19464 0.17986 -0.01542 -0.05335 -0.42953 0.02015 0.00289 -0.0016 -0.00237

Micro BO1 0.00509 0.37331 0.37301 0.02463 -0.05961 -0.76794 0.02854 0.01371 0.00092 0.00015

Micro BO2 0.00387 0.36392 0.36116 0.02435 -0.06306 -0.77291 0.025 0.00986 -0.00139 -0.00148

Micro BO3 0.00217 0.39014 0.39471 0.03092 -0.06455 -0.67403 0.03105 0.01623 -0.00103 -0.0018

Micro RF1 0.02604 8.13018 7.85625 -0.02306 -0.05445 -0.60707 0.20374 0.26489 0.006 0.0055

Micro RF2 0.02522 8.7659 8.3354 -0.0279 -0.05848 -0.67349 0.20004 0.25605 0.01374 0.01265

Micro RF3 0.02573 8.30986 7.97432 -0.02956 -0.05523 -0.67874 0.18782 0.26192 0.01656 0.01583

Fe

259.941 Fe

238.204 Cr

267.716 Cr

205.552 B

249.773 B

208.959 Cd

214.438 Cd

226.502 Mn

257.611 Mn

259.373

Micro Blank 1 0.0111 0.00427 0.00503 -0.01235 0.00348 0.00831 -0.01172 -0.00094 -0.01281 -0.00971

Micro Blank 2 0.0303 0.02356 0.00546 -0.01167 0.00199 0.00366 -0.01174 -0.00103 -0.0127 -0.0097

Micro Blank 3 0.01486 0.00727 0.00521 -0.01223 -0.00263 -0.00058 -0.012 -0.00096 -0.01234 -0.00938

Micro JL1 0.01478 0.00769 0.0061 -0.01191 0.02271 0.0197 -0.01132 -0.00165 -0.01291 -0.01004

Micro JL2 0.02376 0.01619 0.00668 -0.01141 0.01858 0.01833 -0.01146 -0.0017 -0.01296 -0.01017

Micro JL3 0.01411 0.00669 0.00624 -0.01173 0.0158 0.01531 -0.01134 -0.00164 -0.01276 -0.00993

Micro BO1 0.03452 0.02763 0.00663 -0.01267 0.03329 0.02398 -0.01117 -0.00156 -0.01269 -0.00988

Micro BO2 0.03016 0.02384 0.00666 -0.01246 0.03102 0.01777 -0.01113 -0.00158 -0.01286 -0.01

Micro BO3 0.03639 0.02885 0.00717 -0.01155 0.03288 0.02634 -0.01111 -0.00146 -0.01268 -0.00989

Micro RF1 0.07251 0.07239 0.01603 -0.01555 0.31561 0.29538 -0.00728 -0.00585 0.12998 0.12535

Micro RF2 0.07747 0.0767 0.0152 -0.01577 0.31846 0.29463 -0.00747 -0.00554 0.13104 0.12797

Micro RF3 0.07798 0.07878 0.01965 -0.01622 0.3148 0.28739 -0.00679 -0.00803 0.13223 0.12709

Page 19: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

19

P

177.495 P

178.287 Ti

334.941 Ti

336.121 V

292.464 V

311.071 S

180.731 S

182.034

Micro Blank 1 0.13738 0.12534 0.00256 0.00485 -0.0046 0.01187 -1.10561 -1.32242

Micro Blank 2 0.18028 0.16312 0.00183 0.00422 -0.0053 0.01128 -1.45734 -1.77339

Micro Blank 3 0.19062 0.16509 0.0007 0.00265 -0.00529 0.01056 -1.67594 -1.98134

Micro JL1 0.30917 0.26215 0.00408 0.00655 -0.00483 0.0108 13.0052 12.6443

Micro JL2 0.2507 0.20305 0.00343 0.00604 -0.00461 0.01075 12.9093 12.5814

Micro JL3 0.2659 0.22136 0.00395 0.00653 -0.00523 0.01025 12.3708 11.9233

Micro BO1 0.52733 0.45658 0.00739 0.00966 0.0044 0.02012 85.6114 86.2401

Micro BO2 0.528 0.46373 0.00459 0.00706 0.0034 0.0197 83.8525 84.0457

Micro BO3 0.46309 0.40771 0.00574 0.0084 0.00349 0.01975 86.5369 87.1792

Micro RF1 1.048 1.09058 0.00452 0.01206 0.21993 0.24877 119.445 121.804

Micro RF2 1.05867 1.10276 0.00303 0.00988 0.2215 0.25111 118.849 120.711

Micro RF3 1.15496 1.20064 0.00461 0.01341 0.21571 0.24989 117.827 119.806

A 6: ICP-AES measurements made on further diluted sampels of 161 botton and RAFAP 2e91 after microwave digestion.

S

180.731 S

182.034

Blank 1d -1.57374 -1.5991

Blank 2d -1.3493 -1.35288

Blank 3d -0.8492 -0.87703

BO1d 1.76321 1.66622

BO2d 1.52024 1.43785

BO3d 1.72428 1.62

RF1d 2.99316 2.91166

RF2d 2.57286 2.53881

RF3d 2.971 2.93729

A 7: ICP-AES measurements made on JL166DV, 161 botton and RAFAP 2e91 after microwave digestion, recalculated by subtracting the blank and taking dilution and weight of the sample into account.

Al

396.152 Ba

455.404 Ca

393.366 K

766.491 Mg

279.553 Mo

202.030 Na

589.592 Ni

231.604

Micro LOQ 0.18720 0.10305 0.05237 0.20444 0.00428 0.00428 0.32411 0.02028

Micro Blank 0.09700 -0.03340 0.09646 -0.08651 -0.03757 -0.00948 0.23623 -0.01520

Micro JL1 0.01183 -1.02678 0.55721 -3.23099 0.13145 -0.21462 -1.97408 -0.28810

Micro JL2 -1.13822 -1.40777 -1.00616 -5.30229 0.05651 -0.20449 -2.31061 0.08487

Micro JL3 1.55753 -1.73201 -0.81563 -6.16662 0.04022 -0.23895 -2.45991 -0.01321

Micro BO1 55.88450 -2.20829 -0.72571 -4.62129 0.43472 0.20159 8.00968 2.32710

Micro BO2 0.54342 -2.44499 -1.53566 -5.22099 0.31177 0.06842 7.59737 2.35297

Micro BO3 0.51537 -2.56139 9.89623 -4.63479 0.47175 0.09212 9.03076 2.70592

Micro RF1 3.15157 5.68718 190.98925 37.02505 9.09484 1.14270 479.04335 -0.47719

Micro RF2 2.33967 5.74004 191.07623 43.34652 8.91638 1.13913 506.85645 -0.75487

Micro RF3 3.06586 5.43956 222.85964 39.30060 9.16506 1.09157 486.36325 -0.86526

Pb

220.353 Zn

213.856 Cu

324.754 Fe

259.941 Cr

267.716 B

208.959 Cd

214.438 Mn

257.611

Micro LOQ 0.08934 0.06266 0.02761 0.10073 0.00214 0.04402 0.00155 0.00243

Micro Blank -0.03787 0.01732 0.00446 0.01875 0.00523 0.00380 -0.01182 -0.01262

Micro JL1 -0.35189 0.11342 -0.24623 -0.23110 0.05041 0.92500 0.02908 -0.01706

Micro JL2 -0.57853 0.07574 -0.30315 0.29169 0.08428 0.84672 0.02097 -0.02000

Micro JL3 -0.91559 0.16738 -0.35863 -0.27464 0.05954 0.68097 0.02839 -0.00848

Micro BO1 -1.27028 0.65559 -0.20704 0.92126 0.08161 1.17933 0.03798 -0.00428

Micro BO2 -1.49877 0.45695 -0.34826 0.67868 0.08488 0.83139 0.04105 -0.01448

Micro BO3 -1.56547 0.80562 -0.32232 1.03484 0.11364 1.32274 0.04166 -0.00372

Page 20: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

20

Micro RF1 -1.00616 11.31287 0.09325 3.26222 0.65519 17.69470 0.27551 8.65346

Micro RF2 -1.22470 10.85772 0.55125 3.48911 0.59225 17.28212 0.25849 8.53647

Micro RF3 -1.04578 10.27108 0.72871 3.56787 0.86847 17.08394 0.30301 8.72570

P

177.495 Ti

334.941 V

292.464 S

182.034

Micro LOQ 0.27948 0.00928 0.00397 3.3348

Micro Blank 0.16943 0.00170 -0.00506 -1.6924

Micro JL1 8.12799 0.13862 0.01357 833.8744

Micro JL2 4.73503 0.10098 0.02641 831.5979

Micro JL3 5.70607 0.13328 -0.00986 805.3214

Micro BO1 20.91255 0.33267 0.55295 4814.1996

Micro BO2 21.33458 0.17215 0.50356 4521.7219

Micro BO3 17.23087 0.23725 0.50187 4758.4408

Micro RF1 53.31611 0.17133 13.6537 7116.1452

Micro RF2 52.84128 0.07923 13.4630 6347.7822

Micro RF3 59.36948 0.17550 13.2996 7107.3221

A 8: ICP-AES measurements made on JL166DV and RAFAP 2e91 after open vessel digestion.

Al

167.078 Al

396.152 Ba

455.404 Ba

233.527 Ca

396.847 Ca

393.366 K

766.491 Mg

279.553 Mg

280.270 Mo

202.030

Open Blank1 -0.03929 0.04548 -0.01034 -0.01173 -0.01911 -0.04506 -0.05402 -0.03882 -0.01775 -0.00361

Open Blank2 -0.04684 0.03777 -0.01277 -0.01319 -0.01561 -0.04146 -0.13346 -0.03746 -0.01648 -0.0127

Open JL1 -0.06303 0.04517 -0.01357 -0.01316 -0.019 -0.04504 -0.139 -0.03893 -0.01791 -0.01396

Open JL2 -0.02914 0.06965 -0.00519 -0.00658 -0.02472 -0.05103 -0.15371 -0.03596 -0.01516 0.06996

Open JL3 -0.07526 0.04455 -0.01437 -0.01407 -0.02417 -0.05022 -0.21038 -0.03959 -0.01857 -0.01416

Open RF1 0.04021 0.12881 0.15606 0.15526 3.89832 3.90428 0.78405 0.14478 0.15438 0.00039

Open RF2 0.05326 0.14119 0.12311 0.12265 3.15119 3.2146 1.05091 0.17372 0.183 0.1049

Open RF3 0.03918 0.13209 0.18586 0.18492 4.4121 4.39767 0.93997 0.16891 0.17688 -0.00346

Mo

281.615 Na

589.592 Na

588.995 Ni

231.604 Pb

220.353 Pb

168.215 Zn

213.856 Zn

206.191 Cu

324.754 Cu

327.396

Open Blank1 0.00431 0.01709 -0.00343 -0.0102 -0.00498 -0.02075 0.05123 0.02555 0.07519 0.07584

Open Blank2 -0.00204 0.06101 0.04369 -0.01209 -0.01461 -0.02187 0.01624 -0.00828 -0.00239 -0.00417

Open JL1 -0.00052 0.12015 0.10517 -0.0206 -0.0201 -0.45756 0.01606 -0.00715 -0.00521 -0.00883

Open JL2 0.08213 0.11427 0.09799 -0.01886 -0.01956 -0.45584 0.01806 -0.00425 -0.00448 -0.00826

Open JL3 0.00177 0.10063 0.08389 -0.02786 -0.01997 -0.6577 0.01561 -0.00735 -0.00582 -0.00942

Open RF1 0.01763 11.1928 10.4313 -0.00672 -0.01001 -0.40894 0.28363 0.29464 0.00477 0.00111

Open RF2 0.11964 13.2252 12.233 -0.00597 -0.00962 -0.43727 0.33452 0.34724 0.00666 0.00328

Open RF3 0.01456 12.8609 11.809 -0.00722 -0.00452 -0.47446 0.32335 0.35041 0.00544 0.00283

Fe

259.941 Fe

238.204 Cr

267.716 Cr

205.552 B

249.773 B

208.959 Cd

214.438 Cd

226.502 Mn

257.611 Mn

259.373

Open Blank1 -0.00244 -0.00777 -0.00504 -0.0138 0.00191 0.01998 0.00573 0.00543 -0.03094 -0.0211

Open Blank2 -0.00488 -0.00962 -0.00703 -0.01568 -0.01246 0.003 0.00366 0.00412 -0.03283 -0.02282

Open JL1 -0.00332 -0.00885 -0.00735 -0.01728 -0.01802 -0.0132 0.00346 0.00446 -0.03287 -0.02269

Open JL2 0.09685 0.091 0.02782 0.01976 -0.01975 -0.0121 0.0034 0.00456 -0.02979 -0.01913

Open JL3 0.00604 0.00084 -0.00686 -0.0173 -0.00741 -0.00934 0.00349 0.00419 -0.03289 -0.02273

Open RF1 0.08296 0.08056 -0.00048 -0.01324 0.4159 0.3853 0.00725 0.00579 0.14476 0.15234

Open RF2 0.14763 0.14579 0.02043 0.00729 0.49503 0.46478 0.00727 0.00596 0.16882 0.17712

Open RF3 0.0941 0.09219 0.00069 -0.01343 0.47896 0.43475 0.00818 0.00567 0.1735 0.18124

P

177.495 P

178.287 Ti

334.941 Ti

336.121 V

292.464 V

311.071 S

180.731 S

182.034

Open Blank1 -0.02002 0.01861 -0.00938 -0.01176 0.00054 0.00439 -4.01893 -4.07807

Open Blank2 -0.00456 -0.00049 -0.01134 -0.01422 -0.00271 0.00178 -3.67654 -3.86047

Open JL1 0.25514 0.23081 -0.00847 -0.01256 -0.00323 0.0033 15.4605 15.3447

Open JL2 0.74597 0.76292 0.69159 0.68592 0.02006 0.0335 15.5089 15.262

Open JL3 0.37741 0.35292 0.45206 0.44963 -0.00288 0.00856 21.5686 20.6183

Page 21: Determination of trace and major elements in oil by ICP ...825116/FULLTEXT01.pdf · olive oil (Llorent-Martı´nez, Co´rdova, Ortega-Barrales, & Ruiz-Medina, 2014) and analysis of

21

Open RF1 1.00008 1.10833 0.00782 0.00575 0.28451 0.29996 220.819 225.657

Open RF2 1.59862 1.79457 0.20527 0.20321 0.34885 0.37141 263.381 272.12

Open RF3 1.12855 1.25937 -0.01099 -0.01175 0.33194 0.34923 240.078 244.907

A 9: ICP-AES measurements made on further diluted sampels of RAFAP 2e91 after open vessel digestion.

S

180.731 S

182.034

Open R1d 1.3859 0.90152

Open R2d 1.0046 0.47708

Open R3d 1.4048 0.85882

A 10: ICP-AES measurements made on JL166DV and RAFAP 2e91 after open vessel digestion, recalculated by subtracting the blank and taking dilution and weight of the sample into account.

Al

396.152 Ba

455.404 Ca

393.366 K

766.491 Mg

279.553 Mo

202.030 Na

589.592 Ni

231.604 Pb

220.353 Zn

213.856

Open LOQ 0.00386 0.00122 0.00180 0.03972 0.00068 0.00455 0.02196 0.00095 0.00482 0.01750

Open JL1 -1.37082 -0.13835 -0.12222 -3.10760 -0.05424 -0.39858 5.56842 -0.64919 -0.70755 -1.21359

Open JL2 0.99515 0.45487 -0.55528 -4.28576 0.15579 5.58249 5.37560 -0.55135 -0.69786 -1.12021

Open JL3 -2.34755 -0.20526 -0.50750 -8.50500 -0.10573 -0.43786 4.49021 -1.21880 -0.74193 -1.32161

Open RF1 5.9331 11.9420 281.2497 62.5397 13.0325 0.60880 794.669 0.31527 -0.01532 17.80422

Open RF2 6.9086 9.6584 233.6580 82.0958 15.1949 8.10845 945.728 0.37116 0.01255 21.57269

Open RF3 5.7171 13.7230 308.7042 71.8567 14.3928 0.32637 891.290 0.27284 0.36668 20.13213

Cu

324.754 Fe

259.941 Cr

267.716 B

208.959 Cd

214.438 Mn

257.611 P

177.495 Ti

334.941 V

292.464 S

182.034

Open LOQ 0.03879 0.00122 0.00100 0.00849 0.00104 0.00095 0.00773 0.00098 0.00163 0.10880

Open JL1 -0.19362 0.02334 -0.09029 -1.69524 -0.08480 -0.06763 18.36204 0.12977 -0.14728 1326.119

Open JL2 -0.14936 7.18295 2.41945 -1.68586 -0.09255 0.14972 54.18908 50.16488 1.51113 1374.363

Open JL3 -0.25010 0.70729 -0.06016 -1.51885 -0.08786 -0.07328 28.41563 33.71813 -0.13089 1792.844

Open RF1 0.51013 6.17140 0.39578 26.63277 0.18204 12.58539 72.12814 1.29527 20.34774 17351.415

Open RF2 0.64908 10.85072 1.89810 32.51055 0.18468 14.39483 115.53658 15.46527 25.09780 15944.903

Open RF3 0.54429 6.79563 0.46748 29.42224 0.24225 14.27701 79.30367 -0.04379 23.14970 16780.849