Determination Determination ofof HH22SSin in WasteWaste WaterWater andand
PetrochemicalPetrochemical ProductsProducts
1
Dr. Michael Hahn (Chemistry) Dr. Dorit Wilke (Chemistry)
ECH Elektrochemie Halle GmbH, Germany
Bildquelle: ECHwww.detmolder-abwasser.de
Outline
1. ECH – Area of Business
2. H2S Analyzer by ECH- Basics
2
- Basics
- Applications for petrochemical products,
e.g. liquids, viscous samples, solids and gases
- Applications for water, typical measurements and
methods for waste water treatment
3. Summary
1. ECH – Area of Business� Development, manufacturing and marketing of analyti cal
instruments and measurement methods
� Main fields of application are:(A) laboratory analysis(B) factory - and process monitoring for technical control
3
(B) factory - and process monitoring for technical control(C) on-site, online and mobile analysis
(A) Products for Laboratory Analysis
Titrator AQUA 40.00Gas Module
ECH Laboratory GC
Titrator AQUA 40.00
H2S-Analyzer ElectrochemicalAnalyzer SnLayer
4
(B) Typical On -Site ProductsB.2 Analyzer including sample preparationB.1 Analyzers
ECH MobilGC
H2S-Analyzer: automatic gas analyzer for 2 sample points (e.g.: biofilters or waste water tanks)
H2S-Analyzer (manual injection)5
(C) Typical Online ProductsOnline H 2S-Analyzer Online gas chromatograph
online monitoring systembased on the measuring principle of
gas chromatography
liquid samples
gaseous samples6
2. H2S-Analyzers by ECH
7
Applications for the determination of H 2S:
� water and waste water
� solids, oil and gas
� in the laboratory, on-site or online
http://www.deutschebp.de
Problems caused by H2S
1. Odour trouble- sensitization of population
2. Health risk- toxicity is comparable to hydrogen cyanide (HCN)hydrogen cyanide (HCN)
3. Sulphuric acid corrosion - concrete reducing up to 10 mm per year- high investment costs to repair
4. Blocking of biological activity of microorganism s- increasing of bulking sludge inside aeration tanks of waste water treatment plants
8
Effects of H2Sc(H2S)
[ppm ]
Effect
0.002 – 0.2 Odour threshold
3 – 5 Odour nuisance
10 MAC (Maximum Allowable Concentration)10 MAC (Maximum Allowable Concentration)
> 10 Irritation to eyes and respiratory tract
> 50 Serious eye defect (damage of cornea)
150 – 250 Loss of sense of smell
300 – 500 Damage of breathing
>1000 Mortal danger in minutes or seconds
9
Different Types of H 2S-containing Samples in Refineries
Sample Example Method
#1 liquid gasoline, kerosene, diesel fuel, crude oil, mineral oil, hydraulic oil, engine oil, transformer oil
indirect by oven and gas extraction
chamber
10
oil, engine oil, transformer oil chamber
#2 highly viscous and solid
tar, oil sludgebitumen, sulphur
indirect by oven and headspace-technique
#3 gaseous LPG, LNG, CNG gas injection module
#4 water extraction and washing solutions,waste water
indirect by gas extraction chamber
and acidification
Principle of H 2S-Analyzer
Advantages:
� no sample preparation
� fast gas extraction procedure of H 2S from
11
procedure of H 2S from sample
� low cross sensitivity
� H2S is purged completely from sample and transferred to the sensor
H2S-Analyzer Cubiform
12
Features:
� short measuring time 5 – 10 min
� free adjustable sample volume 0.01 – 20 mL
� wide measuring range and low detection limit of 0.0 1 ppm
� automatic peak evaluation at the end of the measure ment
H2S – Analyzer with Autosampler
Advantages:� fully automated analytical
procedure
� sample volume 0.01 – 20 mL
13
0.01 – 20 mL
� low demand of chemicals
� robust and fast analysis(up to 40 measurementsper hour)
Headspace -Module for the Determination of H 2S
14
Without sample
preparation
Headspace vial is inserted into the oven and the
measurement is started
Principle of the Headspace -Module
15
Temperature: up to 250°C
Application: Determination of H 2S in Oil
16Determination of H 2S in a mineral oil sample
Application : Determination of H 2S in heavy Oil
17
Determination of H 2S in heavy oil
Determination of H 2S in Oil
18
Result listshows the high repeatability
of the measurement
Calibration function
Features:� automatic sampling
� direct analysis of sulphide/H 2S-content of waste water and light oils
Online H 2S- Analyzer
� low demand of chemicals (H3PO4)
� data transfer via 4-20 mA or modem/internet
� alarm settings
19
Online application directly in the waste water
Typical Applications� surface water
� ground water
� mineral water
� leachate
� seweage water
20
� seweage water
� pressure pipes
� pump sumps
� grease separators
� refineries, chemical industries
� mineral oil, crude oil
Special Features� detector is separated from the
sample by the gas extraction chamber
� high sensitivity (>0,01 ppm )
21
� high accuracy of the sample injection (1%)
� sample inlet can be rinsed backwards to prevent biofouling and sedimentation
Self Cleaning Sampling
Problem:
Biofouling on surface of electrodes or measuring systems
22
Answer:
� back flushing of sample tubes with fresh water (4 b ar)
� immediate cleaning after injection of waste water s ample
� low maintenance and high robustness
Determination of Sulphide in the Waste Water of Refineries
23Online-analysis of waste water
Comparison of Results from the Laboratory and Online H 2S-Analyzer
200
250
300
350
Online-Sulfid
Laborwerte
Back flush cleaning of the waste water pipe
Online-sulphide
Lab results
0
50
100
150
200
2.6
3.6
4.6
5.6
6.6
7.6
8.6
9.6
10
.6
11
.6
12
.6
Su
lph
ide
[m
g/L
]
24
Typical H 2S- ProfileThe odour nuisance was caused by peaks of H2S from the waste water
60
70
80
90
100
S [p
pm]
--- inlet of pressurized sewer pipe --- outlet of pressurized sewer pipe 25
0
10
20
30
40
50
28.8 29.8 30.8 31.8 1.9 2.9
H2S
[ppm
]
Self-Learning Treatment System
Formation of H 2S depends on:
- flow rate of waste water - residence time of waste water in pressure pipes
26
- ageing characteristics of waste water- inflowing H 2S-content - weather conditions, temperature and season
Optimized Waste Water Treatment
30
40
50
60
70
80
90
100
6
8
10
12
14
16
18
20
H2S
[ppm
]
H2S
[ppm
]
waste water chemical: Fe(II)-salt-solutionconsumption: 1.800 L / dwaste water: 10.000 m3 / dlength of pressure pipe: 3 km
Mean H2S value at outlet: 1,1 ppm
27
--- inlet of pressurized sewer pipe --- outlet of pressurized sewer pipe
0
10
20
30
0
2
4
6
25.7 26.7 27.7 28.7 29.7 30.7 31.7 1.8 2.8 3.8 4.8
SummaryHigh sensitivity and accuracy
Wide linear measuring range
Low demand of chemicals
Low detection limit
28
Low demand of chemicals
Low maintenance effort
No sample preparation is needed
For laboratory, on-site and online applications
ECH Elektrochemie Halle GmbH
Weinbergweg 23D-06120 Halle/S.
Tel: 0345/5583711Internet: www.ech.de
Email: [email protected]
Different kinds of petrochemical samples
Experts in Analysis
ECH Elektrochemie Halle GmbHWeinbergweg 23D-06120 Halle/S.Tel: +49 (0)345 5583 711Fax: +49 (0)345 5583 710web: www.ech.de
… and we will take care ofyour analytical problem …
29
web: www.ech.deEmail: [email protected]