Upload
kowsilax
View
32
Download
2
Embed Size (px)
DESCRIPTION
densitometry
Citation preview
Densitometry / Chung 1
Densitometry
Bob Chung, R • I • T
A densitometer measures light absorption due to the presenceof colorant, such as ink and toner. It mainly measures colorantamount, and not color.
(Rev., 062)
Topics How a densitometer works and what it
measures—Reflection and transmission densitometers—Wide-band and narrow-band densitometers—Hand-held and scanning densitometers
Density-derived values—Dot gain / Print contrast—Ink trapping—Hue error / Grayness
Densitometry / Chung 2
Densitometer Fundamentals A densitometer is an electronic instrument
consisting of—Light source—Photocell—Associated logic and circuitry—To measure the degree of light absorption of a
sample Two types of densitometer
—Reflection and transmission
Reflection Densitometer For press sheet analysis
—CMYK inks or colorants are printed.
1.36
light source
optics
sample
color filter
detector electronics display
Densitometry / Chung 3
Transmission Densitometer For film density and dot area measurement
Reflection Density Density is defined as the logarithm (base 10)
of the reciprocal of the reflectance of a sample.— Absolute density is relative to the calibration white point.— Relative density is the absolute density minus the paper
density.
1.28
Densitometry / Chung 4
Reflection Density Correlate with the ability of a sample to
absorb light—Density ranges from 0.0 to 2.0 (typical)
1.28
Transmission Density Transmission density is log of the
reciprocal of the transmittance (T) of thesample—Transmission densities range from 0.0 to 5.0
Zero density is calibrated with no sample (or air).
3.00
Densitometry / Chung 5
% Film Dot Area Halftones are measured in % dot area.
% dot area = (1 - T) x 100 = (1 - 10-D) x 100
0.00
0.50
1.00
1.50
2.00
2.50
3.00
3.50
0 20 40 60 80 100
% Film dot area
Tra
ns.
D
Density
D Trans.
= log [ 1
100 ! % FDA
100
]
3.00
% Film Dot Area Halftone positive vs. halftone negative
% dot area (pos) = 100 - % dot area (neg)
3.00
Densitometry / Chung 6
Densitometer Calibration Adjust the instrument to known values
—Zero (white point) adjust—Upper scale (black point) adjust
1.28
Transmittance of Filters Applicable to reflection densitometers
—Red Wratten #25
—Green Wratten #58
—Blue Wratten #47
—Amber Wratten #106
For color separation
To convert photocell to the luminosityresponse of the eye
1.28
Densitometry / Chung 7
Transmittance of Filters Spectral transmittance curves of RGB filters
—Wide band filters transmit about 100 nm orone-third of the visible spectrum.
1.28
Reflectance of Process InksCyan ink
Magenta ink
Yellow ink
Densitometry / Chung 8
Filter Densities Cyan ink absorbs long wavelength light
—Red filter density implies cyan ink amount
1.28
Density and Visual Sensation Equal density-incremented scale is not
visually uniform.—Density difference in light region (0.4 - 0.3)
has greater visual difference than the samedensity difference in the shadow region (1.4 -1.3). Density differentiates dark shades.
— Effective process control in pressroom
Densitometry / Chung 9
Wide-band Densitometer Reflection densitometry practices in the
U.S. are wide band ANSI Status T.—CGATS.4
0/45 geometry Black opaque backing The source-filter-detector combination define its spectral responses.
Narrow-band Densitometer The densitomertry practices in Europe is
narrow-band. Advantages
—Less variations due to lamp & photocell—Higher sensitivity to ink film variation—Better inter-instrument agreement
Disadvantages—Not Status T
Densitometry / Chung 10
Polarizing Filter Use of polarizing filters
—Better agreement between wet ink and dry inkdensity readings Minimizing the difference in light scattering
between gloss and matte surfaces—Measurements are different from non-
polarizing densitometers Practices in the United State are non-polarizing
Densitometer Limitations Only measure partial print quality
—Other attributes that affect print quality include Color fidelity Tone reproduction Sharpness Gloss, texture, visual demerits, etc.
May not agree with visual judgment—Visual judgment of the okay press sheet should
precede the density measurement.
Densitometry / Chung 11
Densitometer Limitations A densitometer measures light absorption.
—It does not measure color. It may be termed as “anomalous trichromat”
600400 500 700
wavelength ( )
0
50
100 ! "#
$
ρ is the red (L) cone sensitivityγ is the green (M) cone sensitivityβ is the blue (S) cone sensitivity
Densitometer Usefulness Consistent over time
—No "eye fatigue" problems Not influenced by surrounds
—Eyes are often fooled by surrounds.
Densitometry / Chung 12
Densitometer Usefulness Characterize the input-output relationship
from film, to platemaking, and to printing—% digital dot vs. density of print.—Plate/press curve or amplitude response curve
Provide quantitative data for process control
020 40 60 80 1000
0.5
1.0
1.5
2.0
Drepro
% Dot Area (film pos)
Ds
50%tints
Solid
D50
Density-derived Values Ink-paper-press characteristics
—Dot gain (use relative density values)—Print contrast (use absolute density values)—Trapping (use relative density values)
Process ink characteristics—Hue error / Grayness / Efficiency
Use relative density values
Densitometry / Chung 13
Dot Gain The increase in halftone dot size from the
film stage to the reproduction stagefilm
plate
paper
Dot Gain% Dot Gain = % Paper Dot - % Film Dot
% film dot is measured by a transmission densitometeror defined by digital values.
% paper dot is calculated by the Murray-Davies formula.
Densitometry / Chung 14
Murray-Davies Formula Complementary filter densities of the solid
(Ds) and its tint (Dt) are used—The default dot size on film is 50%
% Paper Dot = 1- 10-Dt
1- 10-Ds ! 100
020 40 60 80 1000
0.5
1.0
1.5
2.0
Drepro
% Dot Area (film pos)
Ds
50%tints
Solid
D50
Print Contrast A measure of the shadow contrast rendered
by the printing process.—Using absolute density values
% Print Contrast = Ds - D75
Ds
! 100
020 40 60 80 1000
0.5
1.0
1.5
2.0
Drepro
% Dot Area (film pos)
Ds
75%tints
Solid
D75
Densitometry / Chung 15
Ink Trap A measure of how well the second ink is
transferred (or trapped) on top of the first ink.—Influence the hue and strength of two-color
overlaps, i.e., red, green, and blue
Ink Trap Formula The most commonly used ink trap formula
was developed by Frank Preucil (GATF).
—All three densities are measured with thecomplementary filter of the second ink layer.
Densitometry / Chung 16
Standard-based Printing GRACoL 6 has been a density-based
printing standard.—Density and % dot gain depend on paper grade.
Higher paper grade can produce higher SID withlower dot gain.
Paper Paper Paper
Name Description (Db) K C M Y K C M Y
Premium gloss/Grade 1 & 2 dull coated
matte coated 0.09 1.60 1.30 1.40 1.00 24 22 22 20#3/#5 coated SWOP 0.14 1.60 1.30 1.40 1.00 +/- 0.10 26 22 22 20 +/- 3%Supercal SC 0.17 1.40 1.10 1.15 0.95 28 26 26 24Uncoated offset 0.17 1.25 1.00 1.12 0.95 28 26 26 24Newsprint SNAP 0.28 1.05 0.90 0.90 0.85 32 30 30 28
* Status T, Absolute, Dry density
Solid Ink Density (aim)*
1.70 1.40 1.50 1.05
Solid Ink
Density
0.07
% Dot Gain (aim) % Dot
Gain tol.
22 20 20 18
Process Ink Evaluation Process inks are evaluated against the ideal
inks.—An ideal process ink absorbs 1/3 of the visible
spectrum completely (high wanted density)and reflects the other 2/3 (zero unwanteddensity).
Solid ink patch
Filter Yellow Magenta Cyan
Red 0 0 High
Green 0 High 0
Blue High 0 0
Densitometry / Chung 17
Typical Process Inks Wanted densities are not high enough.
—Indicated by density of solid tone Unwanted densities are not zero.
—Indicated by medium and low densities
Solid ink patch
Filter Yellow Magenta Cyan
Red low low High
Green medium High medium
Blue High medium low
Process Ink Evaluation Based on the high, medium, and low values
of the three filter densities.
% Hue Error = Dmed - DlowDhigh - Dlow
x 100
% Grayness = DlowDhigh
x 100
% Efficiency = 1 - Dmed + Dlow
2 x Dhigh x 100
Densitometry / Chung 18
Features and Benefits Hand-held densitometers are for lab use.
—Specify printing requirements for legacy jobs, suchas reprint
—Ensure that quality of color printing conforms tocustomer’s requirements.
Features and Benefits Scanning densitometers are for production use.
—One-dimensional scanning densitometer scanscolor control bar.
—Two-dimensional scanning densitometer scanscolor control bar and printed image spots.