2
378 CT Evaluation of Plastic Intraocular Foreign Bodies Glenn C. Henrikson', Mahmood F. Mafee,' Adam E. Flanders,' Robert J. Kriz,' and Gholam A. Peyman 2 The high sensitivity of CT in detecting metallic orbital foreign bodies has been previously described [1-6] . While there has been some success in detecting glass fragments, detecting intraorbital foreign bodies made of wood has been very un- satisfactory [1, 2, 6- 9]. Plastic orbital implants are easily recognized on CT as high-attenuation-value images [6, 7]. This report presents two cases of plastic foreign bodies within the globe that appeared as low-attenuation-value images. The studies were performed on a GE 8800 CT fT scanner, and the globe was evaluated by contiguous 1.5-mm axial sections obtained parallel to the canthomeatalline. The gantry was not angulated to facilitate multi planar reconstruction imaging. A phantom containing eight different types of plas- tics, having a wide range of appearance by CT, was used to further evaluate the appearance of plastic foreign bodies. Case Reports Case 1 An 8-year-old boy was struck in the left eye by an unknown object after a firecracker exploded near him. The next day his eye became swollen and inflamed. On admission, the conjunctiva was injected and a full-thickness corneal laceration was noted. The lens showed a dense, white cortical cataract that precluded a vi ew of the posterior chamber. The right eye was normal on examination. A 8-scan sonogram of the left eye showed trace vitreous debris but no retinal detachment or foreign body, although the patient was minimally cooperative for the examination. A CT scan showed perior- bital edema and an irregular and deformed lens secondary to trauma. A foreign body was not seen on the initial viewing. A low-attenuation- value image was seen on the left in an axial section obtained tangen- tial to the superior pole of the globes (Fig . 1). This was thought to be an artifact arising from the naso-orbital angle. The patient was taken to surgery for repair of the corneal lacera- tion, a pars plana lensectomy, and vitrectomy. During the vitrectomy, a plastic foreign body was found embedded in the superior retina and was removed. The foreign body measured 7.0 mm x 2.5 mm x 0.6 mm (10 .5 mm 3 ). Fulminant endopthalmitis was noted in the posterior chamber and intensive local and intravenous antibiotic therapy was started. After surgery, the patient had only light perception in the left eye. A repeat 8-scan sonogram showed a total retinal detachment. Reevaluation of the CT scan showed the plastic fragment as a lucent band in the superior pole of the left globe. The lucent foreign body measured a -20 CT number and was seen better at a narrow Recei ved January 18, 1985; accepted after revision May 9, 1985. window width of 150 (Fig . 1). An oblique coronal reconstruction image, along the long axis of the lucent band, confirmed the intraocu- lar location of the foreign body. Case 2 A 9-year-old boy had been playing with a bottle rocket when it exploded close to his face. He noted blood from his left eye. Ocular examination on admission showed only light perception in the left eye and a full-thickness corneal laceration. The right eye was normal. A preoperative CT scan showed vitreous and choroidal hemor- rhage and a ruptured lens in the left eye. A lucent object was seen in the mid-aspect of the left globe measuring a -70 CT number (Fig . 2A) . This was thought to represent an intraocular foreign body and was confirmed in the parasagittal reconstruction image (Fig. 28). The patient was operated on for repair of the corneal laceration. A limited 8-scan sonogram at the time of surgery showed echoes within the globe that were thought to be secondary to vitreous hemorrhage. Attenuation of the sonographic beam did not locate a foreign body. A follow-up sonogram 3 days later showed a retinal detachment that was confirmed on the repeat CT study. The foreign body was redemonstrated on CT, but again not definitely located with sonog- raphy because of considerable vitreous debris. At the second surgery, a large retinal tear was observed but a foreign body was not found and the surgery was stopped because of uncontrolled bleeding. A Fig. 1.-Case 1: LOW-density structure (arrow) represent- ing plastic intraocular foreign body is seen in superior pole of left globe. CT scan was obtained at a window width of 150, which shows the foreign body better than at the usual width of 300. , Department of Radiology, Eye and Ear Infirmary, University of Illinois, 1855 W. Taylor Street, Chicago, IL 60612. Address reprint requests to M. F. Mafee. 2 Department of Ophthalmology, Eye and Ear Infirmary, University of Illinoi s, 1855 W. Taylor Street, Chicago, IL 60612. AJNR 8:378-379, March/ April 1987 0195-6108/87/0802-0378 © American Society of Neuroradiology

CT Evaluation of Plastic Intraocular Foreign BodiesThe patient was taken to surgery for repair of the corneal lacera tion, a pars plana lensectomy, and vitrectomy. During the vitrectomy,

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: CT Evaluation of Plastic Intraocular Foreign BodiesThe patient was taken to surgery for repair of the corneal lacera tion, a pars plana lensectomy, and vitrectomy. During the vitrectomy,

378

CT Evaluation of Plastic Intraocular Foreign Bodies Glenn C. Henrikson', Mahmood F. Mafee,' Adam E. Flanders,' Robert J . Kriz,' and Gholam A. Peyman2

The high sensitivity of CT in detecting metallic orbital foreign bodies has been previously described [1-6] . While there has been some success in detecting glass fragments , detecting intraorbital foreign bodies made of wood has been very un­satisfactory [1, 2, 6- 9]. Plastic orbital implants are easily recognized on CT as high-attenuation-value images [6, 7]. This report presents two cases of plastic foreign bodies within the globe that appeared as low-attenuation-value images.

The studies were performed on a GE 8800 CT fT scanner, and the globe was evaluated by contiguous 1.5-mm axial sections obtained parallel to the canthomeatalline. The gantry was not angulated to facilitate multi planar reconstruction imaging. A phantom containing eight different types of plas­tics, having a wide range of appearance by CT, was used to further evaluate the appearance of plastic foreign bodies.

Case Reports

Case 1

An 8-year-old boy was struck in the left eye by an unknown object after a firecracker exploded near him. The next day his eye became swollen and inflamed. On admission, the conjunctiva was injected and a full-thickness corneal laceration was noted. The lens showed a dense, white cortical cataract that precluded a view of the posterior chamber. The right eye was normal on examination.

A 8-scan sonogram of the left eye showed trace vitreous debris but no retinal detachment or foreign body, although the patient was minimally cooperative for the examination. A CT scan showed perior­bital edema and an irregular and deformed lens secondary to trauma. A foreign body was not seen on the initial viewing. A low-attenuation­value image was seen on the left in an axial section obtained tangen­tial to the superior pole of the globes (Fig . 1). This was thought to be an artifact arising from the naso-orbital angle.

The patient was taken to surgery for repair of the corneal lacera­tion , a pars plana lensectomy, and vitrectomy. During the vitrectomy, a plastic foreign body was found embedded in the superior retina and was removed. The foreign body measured 7.0 mm x 2.5 mm x 0.6 mm (10.5 mm3

) . Fulminant endopthalmitis was noted in the posterior chamber and intensive local and intravenous antibiotic therapy was started. After surgery, the patient had only light perception in the left eye. A repeat 8-scan sonogram showed a total retinal detachment.

Reevaluation of the CT scan showed the plastic fragment as a lucent band in the superior pole of the left globe. The lucent foreign body measured a -20 CT number and was seen better at a narrow

Received January 18, 1985; accepted after revision May 9, 1985.

window width of 150 (Fig. 1). An oblique coronal reconstruction image, along the long axis of the lucent band, confirmed the intraocu­lar location of the foreign body.

Case 2

A 9-year-old boy had been playing with a bottle rocket when it exploded close to his face. He noted blood from his left eye. Ocular examination on admission showed only light perception in the left eye and a full-thickness corneal laceration. The right eye was normal.

A preoperative CT scan showed vitreous and choroidal hemor­rhage and a ruptured lens in the left eye. A lucent object was seen in the mid-aspect of the left globe measuring a -70 CT number (Fig. 2A). This was thought to represent an intraocular foreign body and was confirmed in the parasagittal reconstruction image (Fig. 28).

The patient was operated on for repair of the corneal laceration. A limited 8-scan sonogram at the time of surgery showed echoes within the globe that were thought to be secondary to vitreous hemorrhage. Attenuation of the sonographic beam did not locate a foreign body. A follow-up sonogram 3 days later showed a retinal detachment that was confirmed on the repeat CT study. The foreign body was redemonstrated on CT, but again not definitely located with sonog­raphy because of considerable vitreous debris. At the second surgery, a large retinal tear was observed but a foreign body was not found and the surgery was stopped because of uncontrolled bleeding. A

Fig. 1.-Case 1: LOW-density structure (arrow) represent­ing plastic intraocular foreign body is seen in superior pole of left globe. CT scan was obtained at a window width of 150, which shows the foreign body better than at the usual width of 300.

, Department of Radiology, Eye and Ear Infirmary, University of Illinois , 1855 W. Taylor Street, Chicago, IL 60612. Address reprint requests to M. F. Mafee. 2 Department of Ophthalmology, Eye and Ear Infirmary, University of Illinois, 1855 W. Taylor Street, Chicago, IL 60612.

AJNR 8:378-379, March/ April 1987 0195-6108/87/0802-0378 © American Society of Neuroradiology

Page 2: CT Evaluation of Plastic Intraocular Foreign BodiesThe patient was taken to surgery for repair of the corneal lacera tion, a pars plana lensectomy, and vitrectomy. During the vitrectomy,

AJNR:8, March/April 1987 INTRAOCULAR FOREIGN BODIES 379

Fig. 2.-A, Case 2: CT scan shows lucent foreign body (arrow) in mid-as­pect of left globe.

B, Arrow points to lucent intraocular foreign body.

Fig. 3.-Eight diHerent plastics in a water phantom evaluated by CT. Starting from lowest attenuation value to highest density in clockwise rotation: PP = polypropylene (- 125), UHMWPE = ultra-high molecular weight polyethylene (-97), HOPE = high-density polyethylene (-80), PSTY = polystyrene (-35), Nylon = nylon (+94), PMMA = polymethyl methacry­la te (+120), PS = polysulfone (+ 125), Acetal = acetal (+364).

third operation was performed 6 days later. After vitrectomy, a plastic foreign body emerged from the vitreous cavity that measured 11 mm x 5 mm x 1.1 mm (60.5 mm3). At the time of discharge, the boy's vision was limited to light perception only.

Discussion

Optimal evaluation of intraocular pathology, in our experi­ence, is with contiguous 1 .5-mm axial sections parallel to the canthomeatal line. Thin sections are particularly valuable when evaluating small foreign bodies. Orbital reconstructions are helpful in confirming the intraocular location of a foreign body, especially when it is peripherally located. Direct coronal scanning is the best means of evaluating suspected foreign bodies at the 6 or 12 o'clock positions in the globe. The

B

plastic intraocular foreign bodies were seen better at a narrow window width, as used for intraorbital wood fragments [6 , 7).

Since a description of the varied CT appearance of plastics is not found in the literature, we determined the CT charac­teristics of eight different types of plastics by using a water­filled phantom (Fig . 3). The plastics seen in the image of the phantom are (from the lowest CT number to the highest) polypropylene, ultrahigh molecular weight polyethylene, high­density polyethylene, polystyrene, nylon, poly methyl methac­rylate (plexiglass), polysulfone, and acetal. It is apparent that the plastics have a wide range of CT attenuation values (-125 to + 364). The true range of plastics is greater, as Teflon has a CT density of approximately +1000. The plastic foreign bodies in the two cases presented are in the -20 to - 70 range. These plast iCS are most likely polyethylene and poly­styrene mixtures, which are inexpensive plastics and can be found in cheap, disposable products such as fireworks.

ACKNOWLEDGMENT

The authors thank Wayne Swanberg for photographic preparation, Karen Henrikson for editorial advice, Kirk Packo for expert advice, and Leah Freeman for secretarial assistance.

REFERENCES

1. Tate E, Cupples H. Detection of orbital foreign bodies with computed tomography. Current limits. AJR 1981;137 :493- 495

2. Lobes LA. Computed tomography in the detection of intraocular foreign bodies. In! Ophthalmol Clin 1981 ;22:219-234

3. Lobes LA, Grand MG, Reece J, Penkrot RJ. Computerized axial tomog­raphy in the detection of intraocular foreign bodies. Ophthalmology 1981;88:26- 29

4. Wilheim JL, lakov IN, Weinstein MA, Berlin LA, legarra H, Gulman FA. Localization of suspected intraocular foreign bodies with a modified delta 2020 scanner. Ophthalmic Surg 1981;12:633- 641

5. Gaster RN, Duda EE. Localization of intraocular foreign bodies by c0m­

puted tomography. Ophthalmic Surg 1980;11 :25- 29 6. Grove AS. Orbital trauma evaluation by computed tomography. In! Oph­

thalmology Clin 1982;133- 153 7. Grove AS. Orbital trauma and computed tomography. Ophthalmology

1980;403--4 11 8. MaCrae JA. Diagnosis and management of a wooden orbital foreign body:

case report. Br J Ophthalmo/ 1979;63 :848-851 9. Weisman RA, Savino PJ, SChut L, SChatz NJ. Computed tomography in

penetrating wounds oJ the orbit with retained foreign bodies. Arch Otolar­yngo/1983;109:265- 268