SPECIFICATIONS Table 1. Measurable components and range limits: No. of components Component Lower limit *1 Upper limit *1 Gas temperature Purge gas 4th digit 1 HCl 10 ppm *2 5000 ppm ≤ 400°C Air C NH3 15 ppm *2 5000 ppm ≤ 450°C W CO 2 vol% 100 vol% ≤ 300°C A CO (high temperature) 10 vol% 100 vol% ≤ 1200°C *3 B CO (low level) 200 ppm 2 vol% ≤ 400°C M CO2 2 vol% 100 vol% ≤ 300°C G CO2 (high temperature) 10 vol% 100 vol% ≤ 1200°C *3 H CH4 100 ppm 100 vol% ≤ 300°C R O2 4 vol% *5 100 vol% ≤ 300°C N2 P O2 (high temperature) 4 vol% 100 vol% ≤ 1200°C Q O2 (air purge) *4 25 vol% 100 vol% 400–1200°C Air T 2 HCl + H2O (50 vol%) *4 50 ppm (HCl) 1000 ppm (HCl) 130–400°C F NH3 + H2O (50 vol%) *4 50 ppm (NH3) 1000 ppm (NH3) ≤ 450°C X CO + CO2 2.5 vol% 100 vol% ≤ 300°C K CO + CO2 (high temperature) 10 vol% 100 vol% ≤ 1200°C *3 L Notes: 1. The lower and upper limits of the measuring range shown in the above table are for a measuring path length of 1 meter. See the “calculation method” below for other path lengths. 2. Consult us for the ranges below 50 ppm. ● Calculation method of measuring range for optical path lengths other than 1m Measuring range = [Lower/upper limit ÷ path length] Example. 1) HCl analyzer, path length 5m Upper limit: 5,000ppm ÷ 5m = 1,000ppm Lower limit: 10ppm ÷ 5m = 2ppm Therefore, measuring range is between 0 to 2 ···1,000ppm Example. 2) HCl analyzer, path length 0.5m Upper limit: 5,000ppm ÷ 0.5m = 10,000ppm Lower limit: 10ppm ÷ 0.5m = 20ppm Therefore, measuring range is between 0 to 20 ···10,000ppm DATA SHEET ZSS-S EDS3-135m CROSS STACK LASER GAS ANALYZER (Single beam version) Date Control unit Transmitter unit Receiver unit Apr. 4, 2017 Cross stack laser gas analyzer (ZSS) provides continuous measurement of HCl contained in flue gas in waste incin- eration plants and industrial waste disposal facilities, NH3 in power plants and denitration equipment, CO, CO2, and O2 in iron and steel plants and chemical plants with high speed response. Because ZSS is installed directly on the stack or the pipe through which the target gas passes, sample gas conditioning is not required. Being highly tolerant to dust, ZSS can be installed on the upstream of a bug filter where the gas sampling is usually difficult. ZSS is the first laser gas analyzer in Japan that is designed for environmental moni- toring and process monitoring. FEATURES 1. Low maintenance 2. Low operating-cost: with no gas-sampling device and few parts to be replaced 3. Superior long-term stability 4. Hardly affected by cross interference from other gases 5. Fast response within 2 seconds 6. Tolerant to high temperature and high dust 7. Environment friendly: 75 VA, low power consumption 3. Consult us for the ranges above 400°C. 4. Consult us if you select the O2 (air purge) version or the H2O measuring version. 5. Consult us for the ranges below 10 vol%.

CROSS STACK LASER GAS ANALYZER - Istec International · the gas sampling is usually diffi cult. ZSS is the fi rst laser gas analyzer in Japan that is designed for environmental moni-toring

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

Page 1: CROSS STACK LASER GAS ANALYZER - Istec International · the gas sampling is usually diffi cult. ZSS is the fi rst laser gas analyzer in Japan that is designed for environmental moni-toring

SPECIFICATIONS Table 1. Measurable components and range limits:

No. of components Component Lower limit *1 Upper limit *1 Gas

temperature Purge gas 4th digit

1

HCl 10 ppm *2 5000 ppm ≤ 400°C

Air

CNH3 15 ppm *2 5000 ppm ≤ 450°C WCO 2 vol% 100 vol% ≤ 300°C ACO (high temperature) 10 vol% 100 vol% ≤ 1200°C *3 BCO (low level) 200 ppm 2 vol% ≤ 400°C MCO2 2 vol% 100 vol% ≤ 300°C GCO2 (high temperature) 10 vol% 100 vol% ≤ 1200°C *3 HCH4 100 ppm 100 vol% ≤ 300°C RO2 4 vol% *5 100 vol% ≤ 300°C

N2P

O2 (high temperature) 4 vol% 100 vol% ≤ 1200°C QO2 (air purge) *4 25 vol% 100 vol% 400–1200°C

Air

T

2

HCl + H2O (50 vol%) *4 50 ppm (HCl) 1000 ppm (HCl) 130–400°C FNH3 + H2O (50 vol%) *4 50 ppm (NH3) 1000 ppm (NH3) ≤ 450°C XCO + CO2 2.5 vol% 100 vol% ≤ 300°C KCO + CO2 (high temperature) 10 vol% 100 vol% ≤ 1200°C *3 L

Notes:1. The lower and upper limits of the measuring range shown in the above table are for a measuring path length of 1 meter. See the “calculation method” below for other path lengths.2. Consult us for the ranges below 50 ppm.

Calculation method of measuring range for optical path lengths other than 1mMeasuring range = [Lower/upper limit ÷ path length]

Example. 1) HCl analyzer, path length 5m Upper limit: 5,000ppm ÷ 5m = 1,000ppm Lower limit: 10ppm ÷ 5m = 2ppm Therefore, measuring range is between 0 to 2 ···1,000ppm

Example. 2) HCl analyzer, path length 0.5m Upper limit: 5,000ppm ÷ 0.5m = 10,000ppm Lower limit: 10ppm ÷ 0.5m = 20ppm Therefore, measuring range is between 0 to 20 ···10,000ppm

DATA SHEET ZSS-S

EDS3-135m

CROSS STACK LASER GAS ANALYZER(Single beam version)

Date

Control unit

Transmitter unit

Receiver unit

Apr. 4, 2017

Cross stack laser gas analyzer (ZSS) provides continuous measurement of HCl contained in fl ue gas in waste incin-eration plants and industrial waste disposal facilities, NH3 in power plants and denitration equipment, CO, CO2, and O2 in iron and steel plants and chemical plants with high speed response. Because ZSS is installed directly on the stack or the pipe through which the target gas passes, sample gas conditioning is not required. Being highly tolerant to dust, ZSS can be installed on the upstream of a bug fi lter where the gas sampling is usually diffi cult. ZSS is the fi rst laser gas analyzer in Japan that is designed for environmental moni-toring and process monitoring.

FEATURES 1. Low maintenance2. Low operating-cost: with no gas-sampling device and few

parts to be replaced3. Superior long-term stability 4. Hardly affected by cross interference from other gases

5. Fast response within 2 seconds6. Tolerant to high temperature and high dust7. Environment friendly: 75 VA, low power consumption

3. Consult us for the ranges above 400°C.4. Consult us if you select the O2 (air purge) version or the

H2O measuring version.5. Consult us for the ranges below 10 vol%.

Page 2: CROSS STACK LASER GAS ANALYZER - Istec International · the gas sampling is usually diffi cult. ZSS is the fi rst laser gas analyzer in Japan that is designed for environmental moni-toring

ZSS-S

2

Measurement principle: Non-dispersive infrared (NDIR), cross-stack system

Display contents:Gas components, concentration (instantaneous value, average value, O2 corrected instantaneous value, O2 corrected average value), alarm

Light source: Semiconductor laser, Class 1 (O2 analyzer for high-temperature gas and O2 analyzer with instrument air purge use Class 3B laser)

Analog output:4–20 mA DC, 0–20 mA DC, 0–1 V DC, 0–5 V DC, or 1–5 V DCNo-insulation, 2 or 4 points Allowable load: 4–20 mA DC, 0–20 mA DC: ≤ 550Ω

0–1 V DC: ≥ 100 kΩ0–5 V DC, 1–5 V DC: ≥ 500 kΩ

Analog input (for concentration compensation, O2 correction, and alarm output)

4–20 mA DC, 2 points or 6 points (as specified by model code)Input type (user can set the content for each channel among the followings): gas pressure, gas temperature, gas flow velocity, O2 concentration, H2O concentration, purge gas pressure

Contact output:1 from A (normally open) relay contact, 4 points (device failure, low light transmission, out of range, during hold/during calibration)1 from B (normally closed) relay contact, 1 point (power interruption)Contact capacity: 24 V DC, 1 A (resistive load)

Contact input (option):Voltage input (12–24 V DC, ≤ 20 mA), 3 points (output hold, switchover between instantaneous value/average value, average reset)Between contacts and between each contact and the internal circuit are insulated by photo coupler.

Power supply:Rated voltage: 100–240 V ACOperating voltage: 90–264 V ACRated frequency: 50/60 Hz

Power rating: ≤ 75 VA

Purge:To remove dust and/or moisture from the transmitter unit and the receiver unit, and to decrease temperature inside the units*Purging equipment is available. See page 6.

Purge gas:Instrument air that contains neither dust nor mistUse N2 for explosive gas or combustible gas, or for O2 measurement that cannot use instrument air purgePurge gas pressure: ≥ 0.3 Mpa (depends on flow rate, pipe diameter, and length of purge line)

Purge gas flow rate:Purge line is required for each of the transmitter side and the receiver sideMinimum flow rate (L/min) for 10k50A flange:

Flue gas flow velocity (m/s) × 10 L/min, or 20 L/min Note that required flow rate varies with operating envi-ronment.

Storage conditions:Ambient temperature: -20°C to +60°CAmbient humidity: ≤ 95% RH

Dimensions:Receiver unit: approx. 155 × 180 × 400 mmTransmitter unit: approx. 195 × 240 × 400 mmControl unit: approx. 440 × 255 × 137 mm

Weight:Control unit unit and receiver unit: approx. 10 kg eachTransmitter: approx. 8 kg

Cable length:Between transmitter unit and receiver unit: standard 2 m (Maximum 25 m)

Between transmitter unit and receiver unit: standard 5 m (Maximum 100 m)

Finish color:Transmitter box and receiver box: greyControl unit case: silverControl unit front cover: blue

Enclosure:For outdoor installation, dust and rain proof (IP65)Transmitter unit and receiver unit: aluminum + stainless steelControl unit: aluminum

Major materials of gas contacting parts:Stainless steel 316, BK7, PTFE, glass-cloth, silicone

Purge gas port:Joint for 10/8 PTFE tube (Rc1/4 when the joint is removed)

FUNCTIONS CONTACT OUTPUTLow light transmission:

Contact is closed when the amount of the light received by the receiver unit is insufficient.

Out of range:Contact is closed when the measured value is outside the range.

Device failure:Contact is closed when the laser is failed, the laser tem-perature control is failed, the received light is excessive, and/or a communication error occurred.

During hold / during calibration:Contact is closed during calibration and during the analog output hold. The analog output is held at the preceding value or at the user-specified value.

Power interruption:Contact is closed when the power supply is interrupted.

CONTACT INPUT (OPTION)Average reset:

By applying rectangular-wave voltage (with a minimum pulse width of two seconds) to the average value resetting input terminal, a user can reset the O2-corrected average value to the initial value. By opening the contact, a user can re-start the output.

Switchover between the instantaneous value and the moving average value:

By applying rectangular-wave voltage (with a minimum pulse width of two seconds) to the instantaneous and moving average switching input terminal, a user can switch the analog output between the instantaneous value and the moving average value.

Output hold:By applying rectangular-wave voltage (with a minimum pulse width of two seconds) to the AO holding signal input terminal, a user can remotely hold the analog input. Apply-ing the voltage again allows him/her to release the hold.

OTHER FUNCTIONSO2 correction:

Conversion of measured CO gas concentrations into values at reference O2 concentrationCorrection formula:

C = 21 – On 21 – Os

× Cs

C: Converted concentrationCs: Measured concentration of sample gasOs: Measured O2 concentration (Upper limit settable

between 1 and 20%)On: Reference O2 concentration (changeable between

0 and 19%)The result of calculation is indicated and transmitted as an analog output signal.

Page 3: CROSS STACK LASER GAS ANALYZER - Istec International · the gas sampling is usually diffi cult. ZSS is the fi rst laser gas analyzer in Japan that is designed for environmental moni-toring

3

Dry/wet value output:A user can switch the indication and the output between wet values (concentrations which include moisture) and dry values (concentration from which the moisture content is eliminated). The moisture concentration can be set by key or by the analog input.

Communication interface:USB (for adjustment data loading)

PERFORMANCE at 1 atm and 25°C, except for H2O measurementRepeatability:

±1.0% FS (depends on gas, range, optical path length)Linearity:

±1.0% FS (depends on gas, range, optical path length)Zero point drift:

±2.0% FS per 6 months (depends on gas, range, optical path length)

Span drift:±3.0% FS per 6 months (depends on gas, range, optical path length)

Response time (T90):Approx. 1 to 2 seconds

Cross interference:≤ ±2% FSNotes:

1. The amount of some gases (e.g. HCl, NH3) adsorbed by water or other substances are not counted.

2. H2O measurement is a function to detect the pres-ence of moisture and its change. Note that its measurement accuracy is not guaranteed.

3. There may be some difference in measured concen-tration between the laser gas analyzer and manual analysis or other kinds of gas analyzers, because some degree of compensation error is generated depending on measuring environment (moisture concentration, pressure broadening, temperature, pressure, and/or purge gas flow rate).

4. Indication adjustment with manual analysis or other gas analyzers is provided with charge, only when requested.

Warm-up time:≤ 90 min

Minimum detectable limit:1% of range

GAS CONDITIONSTemperature:

See Table1 on page 1, “Measurable components and range limits”.*For measurement of high temperature gas of 500°C or above, install a pressure sensor on the purge line to moni-tor the purging status. If the purge is suspended under the high temperature gas environment, the product may be seriously damaged.

Pressure:±10 kPa (Consult us if the gas pressure exceeds this limit.)

Dust (when the optical path length is 1 m):≤ 5 g/m3 (N) Dust resistant version (22th code H): ≤ 15 g/m3 (N)O2 measurement: ≤ 10 g/m3 (N)Low level CO: 20 g/m3 (N)Consult us if your application has high dust. Dust resis-tance varies with the target gases, specification, optical path length, particle size of dust, and other conditions.

Flow velocity:Depends on the amount of the purge gasFor 50A pipe:Flow velocity (m/s) ≈ purge flow rate (L/min) of each purge line / 10

Moisture:≤ 50 vol%, or not saturated (no condensation)*Measurement may be unavailable when the gas contains

25 vol% or more moisture, the concentration of a target gas component is low, and the component has a tendency to be adsorbed by water.

CALIBRATIONCycle:

Every 6 months*the cycle may change according to the operating en-vironment.

Method:Remove the transmitter and the receiver from the stack, attach them to the calibration cells, and flow the standard gas with a flow rate of 1.5 to 2.0 L/min. (calibration cells and standard gas are to be prepared separately)

Zero gas:Dry N2

Span gas:80% to 100% concentration of each range, balance N2

(when the calibration cell is 1 m)

INSTALLATION ENVIRONMENTLocation:

Where the transmitter unit and the receiver unit can be directly mounted across the stackWhere the scaffolding for maintenance can be providedWhere the transmitter unit and the receiver unit can be monitored without difficultyWhere the transmitter unit, the receiver unit, and the control unit are not exposed to direct sunlight

Vibration:≤ 0.5 G (≤ 0.2 G for a frequency band from 20 Hz to 40 Hz)

Ambient temperature (including the case temperature affected by radiant and/or conductive heat):

Transmitter unit and receiver unit: -20°C to +55°CControl unit: -5°C to +45°C

Ambient humidity:≤ 90% RH

Mounting angle:≤ ±5° from the horizontal(No dew shall accumulate on the lenses of the transmitter unit and the receiver unit)

Optical path length (stack inner diameter):0.5 m to 10 m

Mounting flange:JIS 10K 50A (standard), others (as specified)

PurgeInstrument air (if not available, a compressor is required. The air shall contain neither oil nor mist.)Use N2 for O2 analyzers other than the instrument air purge version.Purge gas pressure: ≥ 0.3 MPa (depends on the flow velocity of target gas)

Calibration cycle:Every 6 months (the cycle may vary depending on the installation environment)

EC DIRECTIVE COMPLIANCE Low voltage (LVD):

EN 61010-2-201: 2013EN 62311: 2008Installation category: II

Electromagnetic Compatibility (EMC):EN 55011: 2009, A1: 2010EN 61000-3-2: 2006, +A1: 2009, +A2: 2009EN 61000-3-3: 2008EN 61326-1: 2013EN 61326-2-3: 2013

Measurement category:CAT II

Page 4: CROSS STACK LASER GAS ANALYZER - Istec International · the gas sampling is usually diffi cult. ZSS is the fi rst laser gas analyzer in Japan that is designed for environmental moni-toring

ZSS-S

4

CODE SYMBOLSWhen ordering, be sure to submit the order sheet on the last page of this Data Sheet.

Note 1) Consult us when selecting H2O measurement.Note 2) Specify the same range for CO and CO2. If different

range is desired for CO and CO2, select “X” at 6 digit and give a description of each range.

Note 3) Specify the measuring range within the limit calculated based on the optical path length (See Page 1).

Note 4) Cable length between receiver and control unit: when you select the code “X”, available length is 10 m or longer.

Note 5) Cable length between receiver and transmitter: when you select the code “X”, available length is 5 m or longer.

Note 6) When the optical path length is 10 m, select “9” in 18th, 19th, and 20th codes.

4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22

ZSS Y 7 - - 0 - NDigit Specification Note Code4 Components CO A

CO (High temperature) BCO (low-level) MHCl CHCl + H2O (50 vol%) Note 1 FCO2 GCO2 (High temperature) HCO + CO2 KCO + CO2 (High temperature) LO2 PO2 (High temperature) QO2 (Air purge) TCH4 RNH3 WNH3 + H2O (50 vol%) Note 1 X

5 Unit ppm 1mg/m3 3vol% 5ppm (1st comp), vol% (2nd comp) 7vol% (1st comp), vol% (2nd comp) 9

6 Measuring 0 to 2 Note 2, 3 Krange 0 to 2.5 Q

0 to 4 S0 to 5 L0 to 10 V0 to 15 00 to 20 10 to 25 T0 to 50 A0 to 100 B0 to 200 C0 to 250 D0 to 400 J0 to 500 E0 to 1000 F0 to 2000 G0 to 5000 H0 to 6000 MOthers X

7 − None Y9 Flange rating 10K 50A (JIS B 2212) A

10K 100A BDN50/PN10 CANSI #150 2B D

10 Number of analog 2 points 0

output points 4 points 1

11 Number of analog 2 points A

input points 6 points B12 Analog output 4 to 20mA DC 1

0 to 20mA DC 20 to 1V DC 30 to 5V DC 41 to 5V DC 5

13 Contact input/ 5 output points, No input 0output 5 output points, 3 input points 1

Digit Specification Note Code14 Cable length 5m Note 4 A

between 10m Breceiver and 20m Ccontrol unit 30m D

40m E50m F80m G100m HOthers X

15 Cable length 2m Note 5 Abetween 5m Breceiver and 10m Ctransmitter 15m D

20m E25m FOthers X

16 Language Japanese JEnglish EChinese C

17 − − 018 Optical path 0m Note 6 0

length 1m 1(ones place) 2m 2

3m 34m 45m 56m 67m 78m 89m 9

19 Optical path 0.0m Note 6 0length 0.1m 1(tenths place) 0.2m 2

0.3m 30.4m 40.5m 50.6m 60.7m 70.8m 80.9m 9

20 Optical path 0.00m Note 6 0length 0.05m 5(hundredths place) (Used only when 10m is specified) 9

21 − − N22 High-dust

versionNo NYes H

Page 5: CROSS STACK LASER GAS ANALYZER - Istec International · the gas sampling is usually diffi cult. ZSS is the fi rst laser gas analyzer in Japan that is designed for environmental moni-toring

5

BASIC PRINCIPLEZSS uses the near-infrared semiconductor laser as the light source, and photo-diode as photodetector. Each gas component has its own wavelength range where it absorbs the light, and a wavelength range consists of a set of many absorption lines, as shown in the figures below. As ZSS can aim at only one absorption-line among them, the measurement principally receives no interference from other gases. ZSS also focus on the amplitude of modulated signal, instead of the amount of change in light intensity, to detect gas concentration.

750nm 1000nm 1250nm 1500nm 1750nm 2000nm 2250nm

1720nm 1740nm 1760nm 1780nm 1800nm 1820nm 1840nm

Absorption line used for measurement

Absorbing spectrum map of measuring gases

Wavelength

Wavelength

Abs

orpt

ion

inte

nsity

Absorption lines of HCl

NH3 NH3NH3

HCl

CH4

CO2 CO

HF HF

H2S H2S

NO

N2O N2O

O2

LIST FOR COMBINATIONS OF MEASURABLE COMPONENTS, UNITS AND MEASUREMENT RANGESComponent Measuring range

CO 0 to 2, 2.5, 4, 5, 10, 15, 20, 25, 50, 100 vol%

CO (high temp.) 0 to 10, 15, 20, 25, 50, 100 vol%

CO (low concentration) 0 to 200, 250, 400, 500, 1000, 2000, 5000 ppm or mg/m3

0 to 2 vol%

HCl 0 to 10, 15, 20, 25, 50, 100, 200, 250, 400, 500, 1000, 2000, 5000 ppm or mg/m3

CO2 0 to 2, 2.5, 4,5, 10, 15, 20, 25, 50, 100 vol%

CO2 (high temp.) 0 to 10, 15, 20, 25, 50 vol%

O2 0 to 4, 5, 10, 15, 20, 25, 50, 100 vol%

O2 (high temp.) 0 to 4, 5, 10, 15, 20, 25, 50, 100 vol%

O2 (air purge) 0 to 25, 50, 100 vol%

CH40 to 100, 200, 250, 400, 500, 1000, 2000, 5000 ppm or mg/m3

0 to 2, 2.5, 4, 5, 10, 15, 20, 25, 50, 100 vol%

NH3 0 to 15, 20, 25, 50, 100, 200, 250, 400, 500, 1000, 2000, 5000 ppm or mg/m3

CO + CO2

1st comp.: CO 0 to 2.5, 4, 5, 10, 15, 20, 25, 50, 100 vol%

2nd comp.: CO2 0 to 2.5, 4, 5, 10, 15, 20, 25, 50, 100 vol%

CO + CO2

(high temp.)1st comp.: CO 0 to 10, 15, 20, 25, 50, 100 vol%

2nd comp.: CO2 0 to 10, 15, 20, 25, 50, 100 vol%

Page 6: CROSS STACK LASER GAS ANALYZER - Istec International · the gas sampling is usually diffi cult. ZSS is the fi rst laser gas analyzer in Japan that is designed for environmental moni-toring

ZSS-S

6

NECESSARY ITEMS TO BE ORDERED SEPARATELY1. Purging equipment Purging equipment is indispensable for the laser gas analyzer to remove dust and mist from the transmitter unit and the re-

ceiver unit. We offer the following three types: A set of equipment in a box, a set of equipment without box, or each single equipment.

1.1 A set of purging equipment in a box

Item ModelPurging equipment in a box: flowmeter scale 4–50 L/min ZZP*ZSSTK7N6685C2Purging equipment in a box: flowmeter scale 20–100 L/min ZZP*ZSSTK7N6685C1Purging equipment in a box: flowmeter scale 30–300 L/min ZZP*ZSSTK7N6685C3

Inlet for instrument air or N2

(Coupling joint)

to transmitter purge inlet

Piping: ø10/8 PTFE tube

2-Flow meterMist separator

Filter regulator

Piping system diagram

(IN)(OUT)

(IN)

(OUT)

(IN)

(OUT)

Total 20m1m

0.5m 0.5m

for ø10/8 tube (metallic tube recommended)

to receiver purge inlet

Receiver purge outlet

Rc 1/4

Rc 1/4 Transmitter purge outlet

Inlet for instrument air or N2

Rc 1/4

Drain outletPipe retainer for ø10 pipe

4 - ø11

Mist separator

Filter regulator

Flowmeter

(Mounting hole)

340

300

370

420

500

150 110 6543

58

120

200

223

Inlet for instrument air or N2

Filter regulator

Drain outlet

Receiver purge outlet

Transmitter purge outlet

10-mm OD pipe (not supplied)

Mist separator

2 - Flowmeter

Piping diagram

(IN)

(IN) (IN)

(IN)

(OUT)

(OUT) (OUT)

(OUT)

1.2 A set of purging equipment without box

Item Model RemarksPurging equipment without box: flowmeter scale 4–50 L/min ZZP*ZSSTK7P1433C2

With PTFE tubePurging equipment without box: flowmeter scale 20–100 L/min ZZP*ZSSTK7P1433C1Purging equipment without box: flowmeter scale 30–300 L/min ZZP*ZSSTK7P1433C3

Page 7: CROSS STACK LASER GAS ANALYZER - Istec International · the gas sampling is usually diffi cult. ZSS is the fi rst laser gas analyzer in Japan that is designed for environmental moni-toring

7

1.3 Purging equipment

Item Q’ty ModelFlowmeter with 4–50 L/min scale 2 ZZP*ZSSTK7N4624P2Flowmeter with 20–100 L/min scale 2 ZZP*ZSSTK7N4624P1Flowmeter with 30–300 L/min scale 2 ZZP*ZSSTK7N8849P1Filter regulator 1 ZZP*ZSSTK7N7466P1Mist separator 1 ZZP*ZSSTK7H8049P1R 1/4 cap nut for mist separator 1 ZZP*ZSSTK738114P7

2. Zero/span calibration equipment To carry out calibration, remove the angle adjustment unit, and install the following equipment, and then flow the zero gas or

span gas.Item Q’ty Model

(1) Calibration gas cell (for HCl, NH3, CO, CO2, CO + CO2, CH4)*1 1 ZZP*ZSSTK4J3676C1Calibration gas cell (for HCl + H2O, NH3 + H2O, O2)*1 1 ZZP*ZSSTK4J5026C1

(2) Cable btwn the receiver unit and the control unit (5 m) 1 ZZP*ZSSTK4J1271C2(3) Cable btwn the receiver unit and the transmitter unit (1.6 m) 1 ZZP*ZSSTK4J0641C3(4) Pressure regulator 1 ZBD6(5) Flowmeter (1.5–2.0 L/min) 1 ZBD4

*1: The standard length of the calibration gas cell is 1 m. Consult us if a low concentration gas cylinder is difficult to obtain.

3. Optical axis adjusting tool You can adjust the optical axis by aiming the laser pointer to the center of the target.3.1 Optical axis adjusting tool

Item Q’ty Model(1) Optical axis adjusting tool (laser pointer)

Optical axis adjusting tool (target)1 set ZZP*ZSSTK4J1274C1

(3) Cable btwn the receiver unit and the transmitter unit (1.6 m)

(2) Cable btwn the receiver unit and the control unit (5 m)

(1) Calibration gas cell (1m)

IN

(4) Pressure regulator

Pipe (PTFE tube, etc)

(1) Calibration gas cell

OUT

(5) Flowmeter (1.5 to 2.0 L/min)

Cyl

inde

r

Stack

(1) Optical axis adjusting tool (laser pointer)(1) Optical axis adjusting tool (target)

Angle adjustment unit

3.2 BNC cable for optical adjustment

After installing the transmitter unit, the receiver unit, and the control unit, connect a digital voltmeter through the BNC cable to the transmitter unit or the receiver unit. Adjust the optical axis so that the voltage indicated on the voltmeter becomes the relevant value.

Item Q’ty ModelBNC cable for optical adjustment 1 ZZP*ZSSTK7P2524C1 Receiver unit

Digital voltmeter BNC cable for optical axis adjustment

Page 8: CROSS STACK LASER GAS ANALYZER - Istec International · the gas sampling is usually diffi cult. ZSS is the fi rst laser gas analyzer in Japan that is designed for environmental moni-toring

ZSS-S

8

DIMENSIONS (in mm)

100100

Approx. 400

Approx. 400

150 to 200

0.5 to 10 m

150 to 200

Receiver box (120×180×100 mm)

Transmitter box (160×240×100 mm)

Angle adjustment unit

Purge gas inlet (with bell-and-spigot joint for 10/8 tube)

Purge gas inlet

Filter regulator

Mistseparator

Note 1)

Note 1: Be sure to put a R 1/4 cap nut (ZZP*ZSSTK738114P7) on the drain outlet of the mist separator.

120

160

ø155

Receiver unit Transmitter unit

Cable between receiver unit and transmitter unit (standard: 2m)

Notes:• For O2 analyzers other than the instrument air purge version, use N2. For all the other cases, use instrument air. If you use air that contains oil and/or mist, the purge gas flow decreases due to the contaminant, which adversely affect the measurement. In such a case, frequent maintenance is required, and in some cases you may have to install a filter additionally. • Once you install the analyzer, you have to purge the equipment with instrument air or N2 regardless if the analyzer and the furnace are in operation or not. If you operate the furnace without purging, it may cause the fatal damage to the optical part.

Angle adjustment unit

Inlet for instrument air or N2

16

Bolt and nut (M16 × 55)Flange not supplied

JIS10K50A FF(with packing)

Stack

420 32

0

240255

440

1302-ø12

ø10/8 PTFE tube (not supplied)

2 - Flowmeter

20 to 300 L/min

Purging equipment (to be ordered separately)

74

Cable between receiver unit and control unit (standard: 5 m)

Control unit

DI/DOAI

AOPow

er

supp

ly

134 3

ø34(Connector)

ø34(Connector)

ø30

Connector

Cable length(200 mm)

7- Waterproof grand (inner diameter 12 mm)

Mounting hole

1274

ZSSLASER GAS ANALYZER

2-ø1274

420

ø14

ø10.5

4. IR card (IR visualizer) for NH3

Item Q’ty ModelIR card (for NH3) 1 ZZP*ZSSTK7N4505P1

IR CARD

Transmitter unitIR card

laser light (visualized)

Page 9: CROSS STACK LASER GAS ANALYZER - Istec International · the gas sampling is usually diffi cult. ZSS is the fi rst laser gas analyzer in Japan that is designed for environmental moni-toring

9

CONNECTION DIAGRAMThere are two types of output/input terminal boards.

• Two Analog Outputs Version (10th code "0")

AO terminal

Al terminal

DI/DO terminal

Notes: 1. Do not use an unassigned terminal as a relay terminal because it may be connected to the internal circuit. 2. AI terminal, AO terminal, and DI/DO terminal are all on the same board.

21 22 23 24 25 26

1 2 3 4 5 6

27 28 29 30 31 32

7 8 9

33 34 35 36 37 38

13 14 15 16 17 18

39 40

19 2010 11 12

1 AO1+2 AO1-

Screw diameter : M2 or the equivalent

21 22 23 24 25 26

1 2 3 4 5 6

27 28 29 30 31 32

7 8 9

33 34 35 36 37 38

13 14 15 16 17 18

39 40

19 2010 11 12

3 AI1+4 AI1-5 AI2+6 AI2-7 AI3+8 AI3-

Screw diameter : M2 or the equivalent

21 22 23 24 25 26

1 2 3 4 5 6

27 28 29 30 31 32

7 8 9

33 34 35 36 37 38

13 14 15 16 17 18

39 40

19 2010 11 12

9 DO1 10 DO1 11 DO2 12 DO2 13 DO3 14 DO3 15 DO4 16 DO4 17 DO5 18 DO5 19 DO6 20 DO6

Screw diameter : M2 or the equivalent

Analog output 1

Low light transmission

Device failure

During hold / during calibration

Outside the range

29 DO7 30 DO7 31 DO8 32 DO8 33 DI1 34 DI1 35 DI2 36 DI2 37 DI3 38 DI3 39 DI4 40 DI4

Power interruption

Average value reset signal (option)

Switchover between instantaneous value/moving average value (option)

Output hold (option)

Analog input 3 (AI extension board is required)

Analog input 2

Analog input 123 AI4+24 AI4-25 AI5+26 AI5-27 AI6+28 AI6-

Analog input 6 (AI extension board is required)

Analog input 5 (AI extension board is required)

Analog input 4 (AI extension board is required)

21 AO2+22 AO2-

Analog output 2

1 2 3

Power supply terminal

100 to 240 V AC(50/60Hz) FG Screw diameter : M4

ANGLE ADJUSTMENT UNIT

ø120

Purge gas inlet (with bell-and-spigot joint for 10/8 tube)

164 to 194

3 - M16

6 - M8

HCl + H2ONH3 + H2OO2

Other than above

68 (receiver unit)74 (transmitter unit)

d (mm)Component

68 (receiver unit)68 (transmitter unit)

3 - M12

ød

ø120

ø155

Companion flange Transmitter box / receiver box

ø155

4-ø19

Page 10: CROSS STACK LASER GAS ANALYZER - Istec International · the gas sampling is usually diffi cult. ZSS is the fi rst laser gas analyzer in Japan that is designed for environmental moni-toring

ZSS-S

10

• Four Analog Outputs Version (10th code "1")

1 2 3

7 8 9 10 11 12

1 2 3 4 5 6

Power supply terminal

AO terminal

AC100 to 240V50/60Hz

AO1

AO4

AO2 AO3

Unassigned Unassigned

PE

+ -

+ -

+ - + -

7 8 9 10 11 12

1

1 2 3 4 5 6 7 8 9 10 11 12

13 14 15 16 17 18 19 20 21 22 23 24

25 26 27 28 29 30 31 32 33 34 35 36

2 3 4 5 6

AI terminal

DI/DO terminal

AI1 AI2 AI3

AI5 AI6AI4

+ - + - + -

+ - + - + -

Note) Do not use an unassigned terminal as a relay terminal because it may be connected to the internal circuit.

Screw diameter: M4

1 AO1+ 2 AO1-

Analog output 1Analog output 3(AO extension board is required)

5 AO3+ 6 AO3-

3 AO2+ 4 AO2-

Analog output 2Analog output 4(AO extension board is required)

7 AO4+ 8 AO4-

1 AI1+ 2 AI1-

Analog input 1Analog input 4(AI extension board is required)

7 AI4+ 8 AI4-

3 AI2+ 4 AI2-

Analog input 2Analog input 5(AI extension board is required)

9 AI5+10 AI5-

5 AI3+ 6 AI3-

Analog input 3(AI extension board is required)

Analog input 6(AI extension board is required)

11 AI6+12 AI6-

1 DI1 2 DI1 3 DI2 4 DI2 5 DI3 6 DI3 7 DI4 8 DI4 9 DI5 10 DI5 13 DO1 14 DO1 15 DO2 16 DO2 17 DO3 18 DO3

19 DO4 20 DO4 21 DO5 22 DO5 23 DO6 24 DO6 25 DO7 26 DO7 27 DO8 28 DO8 29 DO9 30 DO9 31 DO10 32 DO10 33 DO11 34 DO11

Average value reset (option)

Switchover between instantaneous value/moving average value (option)

Outside the range(ch.1)

Outside the range(ch.2)

Output hold (option)

Power interruption

Low light transmission

Device failure

During hold/duringcallibration

STANDARD ACCESSORIESItem Q’ty Specification

Bolt 8 or 16 *1 M16 × 55 (70) *2, stainless steel

Nut 8 or 16 *1 M16, stainless steel

Spring washer 8 or 16 *1 M16, stainless steel

Flat washer 8 or 16 *1 M16, stainless steel

Companion flange packing or flange packing for use in high temperature

2 According to flange specification

Bolt for angle adjustment 6 Hex socket bolt M8 × 70

Power supply fuse 2

Bolt for connecting receiving unit and transmitter unit

12 Hex socket bolt M5 × 12

Notes:1. When the 9th code is “B”, 16 pieces are provided. For

other cases, 8 pieces are provided.2. When the 9th code is “A”, the bolt length is 55 mm.

When the 9th code is “B, “C”, or “D”, the bolt length is 70 mm. Inch-sized bolts are not supplied.

SPARE PARTS FOR ONE YEAR (ZBN1SS12)Name Quantity Remarks

Silicon packing A 2 pieces For bellows (ZZP*ZSSTK7N3508P1)

O-ring 2 pieces (ZZP*ZSSTK7P2530P5)

SCOPE OF DELIVERY• Control unit• Transmitter unit• Receiver unit• Angle adjustment units (two units, one for transmitter unit

and the other for receiver unit)• Cable between the transmitter unit and the receiver unit

(specified length)• Cable between the receiver unit and the control unit (speci-

fied length)• Standard accessories• Instruction manual

NECESSARY ITEMS TO BE ORDERED SEPA-RATELY• Purging equipment …………………………… See Page 6• Zero/span calibration equipment *1 ………… See Page 7• Optical axis adjustment tool *1 ……………… See Page 7• Spare parts for one year

*1: The calibration equipment and the optical axis adjustment tool are not required for every gas analyzer, but required at least one set for one site.

Page 11: CROSS STACK LASER GAS ANALYZER - Istec International · the gas sampling is usually diffi cult. ZSS is the fi rst laser gas analyzer in Japan that is designed for environmental moni-toring

11

MOUNTING DIMENSIONS (in mm)

c

b

Stack

Companion flange (*5, *6)

(*6)

Angle adjustment unit (*6)

Scaffold (*4)

a: inside diameter ≥ 50 mmb: 20 to 70 mm (*1)c: ≥ approx. 200 mm (*2)d: ≥ 400 mm (*3)e: 500 to 1500 mm(*3)

a

d

e

*1: When the flue gas contains a large amount of dust, ensure the minimum length of 50 mm.*2: When the flue gas temperature is high (500°C or more), ensure the minimum distance of 400 mm. Note that in that case you have to be even more careful in adjusting the angle of the companion flange because the angle range within which the companion flange can be adjusted is narrow.*3: Make sure to install the analyzer in the place where it is easy to operate, and where there is enough clearance.*4: Scaffold is required for both the receiver unit side and the transmitter unit side.*5: Use a flanged valve where there is a risk of gas blowout or where the operator may be exposed to a dangerous situation.*6: Tightening torque for the companion flange shall be 118±14 N·m. When tightening the bolts of the flange or the angle adjustment unit, apply grease to the bolts. (Recommended grease is the one contains molybdenum.)

Keep purging the analyzer once you install it on a stack. Otherwise, the contamination of optical surface may result in the analyzer failure.

ROUGH GUIDE TO MOUNT THE COMPANION FLANGEWhen θ is larger than 5°, attach the companion flange with the γ angle ≤ 5°inside the circumference of radius A.

θ = tan-1

A ≈ 0.087 × (Z + Y + 125)

tan5° ≈ 0.087

X: Inner diameter of flange of receiver unit and transmitter unitY: Distance between the outer wall of the stack and each flange of the transmitter unit and the receiver unitZ: Inner diameter of stack 2 (Y + 125)

X

Y

X

Z

γ

Page 12: CROSS STACK LASER GAS ANALYZER - Istec International · the gas sampling is usually diffi cult. ZSS is the fi rst laser gas analyzer in Japan that is designed for environmental moni-toring

Printed in Japan

Caution on Safety

*Before using this product, be sure to read its instruction manual.

Information in this catalog is subject to change without notice.

Global Sales SectionInstrumentation & Sensors Planning Dept.1, Fuji-machi, Hino-city, Tokyo 191-8502, Japanhttp://www.fujielectric.comPhone: +81-42-514-8930 Fax: +81-42-583-8275http://www.fujielectric.com/products/instruments/

ZSS-S

Please specify the following items when you contact us for inquiry.

1. Code symbol ZSS

2. Parameter of target gas

Item Min. value Typical value Max. value

Concentration range of target gas

Temperature (°C)

Pressure (kPa)

Flow velocity (m/s)

Moisture (vol%)

Dust (mg/m3 (N))

Other gas component (vol%/ppm)

Other gas component (vol%/ppm)

Other gas component (vol%/ppm)

3. Stack inner diameter (optical path length) A m

4. Distance from stack (outside wall) to flange B m

5.Distance from stack (outside wall) to flange C m

6. Cable between receiver unit and control unit D m

7. Cable between receiver unit and transmitter unit E m

8. Spares for 1-year measurement no need / necessary

9. Separate order items no need / necessary

10.Output of O2 correction value (only for HCl, NH3 and CO meter) no need / necessary

11.Reference O2 concentration value (vol%) (When "necessary" is selected in 10) vol%

12.Presence of vibration No / Yes ( G)

AB C

E

Receiver unit

Control unit

Transmitter unit

Stack

D

Instrument air or N2 Instrument air or N2