HIRL Birmingham 1

CORRECT ENDOSCOPE REPROCESSING

Christina BradleyLaboratory Manager

Hospital Infection Research LaboratoryQueen Elizabeth Hospital

Birmingham

Page 2: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham

�“Inadequate decontamination procedures and equipment malfunction were two leading causes of post endoscopic infection and contamination. More than 91% of the infections could be prevented if quality control systems were improved”

Kovaleva J et al Clin Microbiol Rev 2013;26(2):231-54

Page 3: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham

Transmission of infection by flexible gastrointesti nal endoscopy and bronchoscopy

TYPE OF FLEXIBLE ENDOSCOPIES

No. of REPORTED

INFECTIONS/OUTBREAKS

REASONS FOR FAILURE

BRONCHOSCOPY 51 • Inapprpopriate cleaning and disinfection – povidone iodine• Contaminated AER• Incorrect connectors• Rinsing with tap water• Hole in the endoscope sheath (no leak testing)

ERCPEndoscopic Retrograde CholangioPancreaticography

23 • Inapprpopriate cleaning and disinfection – povidone iodine, cetrimide• Contaminated AER• Incorrect connectors• Failure to irrigate all channels• Rinsing with tap water• Contaminated water bottle

UPPER GASTROINTESTINAL ENDOSCOPY

19 • Inapprpopriate cleaning and disinfection – povidone iodine• Contaminated AER• Incorrect connectors• Rinsing with tap water• Hole in the endoscope sheath (no leak testing)

SIGMOIDOSCOPY &COLONOSCOPY

5 • Inapprpopriate cleaning and disinfection – povidone iodine, BKC, cetrimide• Contaminated AER• Incorrect connectors• Contaminated water bottle• Biopsy forceps not sterilized

Kovaleva J et al Clin Microbiol Rev 2013;26(2):231-54

Page 4: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham 4

Page 5: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham 5

DECONTAMINATION OF FLEXIBLE ENDOSCOPES

�Meticulous cleaning�Proper immersion in disinfectant�Thorough rinsing after disinfection

�Endoscope Washer Disinfectors are the preferred option

Page 6: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham

ENDOSCOPE DECONTAMINATION PROBLEMS

� Instruments and accessories are expensive� Damaged by heat and pressure� Complex, difficult to clean and dry� Penetration of channels uncertain� Rapid throughput� Risk of recontamination during final rinse

from water supply/EWD

Page 7: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham

ENVIRONMENT DESIGN AND LAYOUT

� Designated and dedicated decontamination area

� Separate dirty, clean and storage areas� One-way flow for equipment� Suitable facilities for manual cleaning� Dedicated hand wash basin

Page 8: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham

ENDOSCOPE CHANNELS

Page 9: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham

DECONTAMINATION STAGES

�Pre clean at patient bed side�Leak test�Manual clean�Endoscope Washer Disinfector�Storage

Page 10: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham

CLEANING

Page 11: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham

PRE CLEANING

� At point of use– Air/water channel– External surfaces– Suction channel

� Manufacturers instructions areessential

Page 12: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham

MANUAL CLEANING� At the sink in dedicated decontamination room

– Brushing Flushing

� Single use brushes must be used for cleaning at all times (UK recommendation)

� Staff must ensure meticulous adherence to manufacturers instructions for cleaning

Important to ensure� Access to all channels whether they have been used or not e.g. forceps

raiser channel, auxiliary water channel� Irrigation of all channels that cannot be brushed

Page 13: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham

DISINFECTANTS

Page 14: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham 14

ENDOSCOPE DISINFECTANTS

�Glutaraldehyde�Peracetic acid�Chlorine dioxide�Superoxidised water�OPA�Etc, etc ……….

Page 15: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham 15

DISINFECTANTS AND ENDOSCOPE DECONTAMINATION

Ensure disinfectant�has broad spectrum of antimicrobial

activity� is compatible with instruments and

processing equipment� is used at effective concentration� is in contact with all surfaces

Page 16: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham 16

CHANNEL IRRIGATION DEVICES

Page 17: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham 17

Page 18: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham 18

ENDOSCOPE DECONTAMINATION : POTENTIAL PROBLEMS

� Disinfectant– Spectrum of activity– Over use

� Water quality– Bacterial contamination– Chemical contamination

� Training� Accessories e.g. water bottle, biopsy forceps

Page 19: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham 19

Page 20: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham 20

Page 21: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham 21

BACTERIA FREE WATER

�Cystoscopes�Bronchoscopes�ERCP

�Potential for infection and/or misdiagnosis of infection

Page 22: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham 22

Page 23: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham 23

ENDOSCOPE WASHER DISINFECTORS

The WD440 machines:The WD440 machines:

�Single chamber �Twin chamber (independent operated)

Page 24: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham

ENDOSCOPE WASHER DISINFECTORS : PROBLEMS

�Disinfectant concentration�Machine contamination�Water quality�Maintenance�Validation

Page 25: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham 25

ENDOSCOPE STORAGE

Endoscopes which have been reprocessed by high level disinfection and have been left for periods >3 hours should be reprocessed.

MDA DB2002(05)

Page 26: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham

TRANSPORTATION

�Protection of endoscopes and the environment from contamination and/or possible damage

�Separation of dirty and clean endoscopes

Page 27: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham

DOCUMENTATION

Types:-� Policies: Statement of intent e.g. ‘We will “Provide

decontaminated endoscopes”� Procedures: Sets of instructions – How policy

is implemented e.g. ‘How to decontaminate endoscopes’

� Forms: Evidence that procedures are carried out. e.g. ‘List of items processed’

� Others: Info, Guidelines, standards etc

Page 28: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham

STANDARD OPERATING PROCEDURES

� Purpose of SOPs– SOP’s are required for all tasks which could have an

effect on the quality of the decontamination process

� SOP’s– Must clearly describe specific tasks– Must be easy to understand– Must change if the task changes– Are an essential tool in training

Page 29: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham 29

STAFF TRAINING

Page 30: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham 30

STAFF TRAINING

� At induction and regular updates� Anatomy/construction of endoscopes� Use of washer disinfectors (if applicable)� Traceability systems� Importance of decontamination record

keeping

� Records kept of individuals training� Sufficient time allowed for training

Page 31: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham

SUMMARY

It is important to ensure that� Staff receive comprehensive training � Cleaning takes place prior to sterilization or

disinfection� All surfaces of the endoscope are exposed during the

decontamination procedure� An effective disinfectant is used at the correct

concentration� The final rinse does not re-contaminate the

endoscope� Testing, in accordance with national guidance, takes

place and is audited

Page 32: CORRECT ENDOSCOPE REPROCESSING · CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory ... Transmission of infection by flexible

HIRL Birmingham 32

THANK YOU FOR LISTENING