Upload
al-james
View
212
Download
0
Embed Size (px)
Citation preview
Contralateral suppression of DPOAE measured in real time
A.L. JAMES,� R.J. MOUNT� & R.V. HARRISON�y�Auditory Science Laboratory, Department of Otolaryngology and Brain and Behaviour Division, Hospital for Sick
Children and yDepartment of Physiology, University of Toronto, Toronto, Ontario, Canada
Accepted for publication 15 October 2001
JA M E S A.L . , M O U N T R.J . & H A R R I S O N R.V.
(2002) Clin. Otolaryngol. 27, 106–112
Contralateral suppression of DPOAE measured in real time
The aim of this study was to measure contralateral suppression of distortion product otoacoustic emissions (DPOAE)
in real time. A total of 10 human subjects were studied with a novel device to record DPOAE without signal
time averaging, using digital narrow band pass filtering. Real time DPOAE levels were recorded at 2f1-f2 using
primary tone settings of f2/f1¼ 1.22 and L1¼ 70 dB SPL, L2¼ 65 dB SPL, at five values of f2 between 2.2
and 7.7 kHz. An acoustic stimulus was applied intermittently to the contralateral ear to cause DPOAE suppression.
Characteristic features of contralateral suppression were identified and distinguished from small spontaneous
variations in the real time DPOAE signal. Magnitude of suppression increased with contralateral stimulus intensity.
Onset latency of suppression was around 43 ms (31–95 ms). Potential clinical applications are discussed in the
light of these findings, including a role in improving the specificity of neonatal hearing screening.
Keywords distortion product otoacoustic emission contralateral suppression latency
Distortion product otoacoustic emissions (DPOAEs) are
widely used in a variety of clinical settings, most commonly
in neonatal hearing testing.1 Their discovery was first reported
in 19792 and, since then, much work has been done on
defining optimum characteristics of the primary frequencies
that evoke these emissions, and on the interpretation of
emission data from normal and abnormal ears. Among their
many research applications, DPOAEs have been used to study
the function of the cochlear efferent neurones. These fibres
mediate suppression of otoacoustic emissions (OAEs) via the
medial olivocochlear pathway after acoustic stimulation of the
contralateral ear.3
To date, all commercially available devices that measure
DPOAEs use signal time-averaging to distinguish the small
DPOAE signal from physiological and other background
noise. We report a preliminary study of a novel DPOAE
device (Vivo 600DPR, Vivosonic, Toronto, Canada; http://
www.vivosonic.com), which discriminates DPOAE signals
using a digitally implemented narrow band pass filter,
thus allowing measurement of emissions in real time.4 In
addition to considerable research applications, we consider
that this may have significant advantages in certain clinical
settings.
Our previous studies with a prototype device have assessed
the accuracy and reliability of real time DPOAE measurement
in comparison with conventional signal averaging devices.5,6
We have used real time DPOAE measurement to assess the
latency and frequency specificity of contralateral suppres-
sion in an animal model.7,8 In this report, we demonstrate
the characteristics of real time contralateral DPOAE sup-
pression in human subjects and discuss potential clinical
applications.
Patients and methods
Real time DPOAEs were measured in 10 human subjects (ages
from 9 days to 60 years) using a Vivo 600DPR. Recordings
were made in the presence and absence of acoustic stimulation
of the contralateral ear. All measurements were made in a
sound-attenuating booth. Primary frequencies were set at f2/
f1¼ 1.22 for 5-values of f2 between 2.2 and 7.7 kHz, with
intensities of L1¼ 70 dB SPL and L2¼ 65 dB SPL. The dis-
tortion product emission at 2f1-f2 was recorded for periods
ranging from 15 s to 3 min.
Clin. Otolaryngol. 2002, 27, 106–112
106 # 2002 Blackwell Science Ltd
Correspondence: Professor R.V. Harrison, Auditory Science La-boratory, Department of Otolaryngology, Hospital for Sick Children,555 University Avenue, Toronto, Ontario M5G 1X8, Canada(e-mail: [email protected]).
pr i nc i ple s of dpoa e e st im at ion i n r eal t im e
The Vivo 600DPR uses a Kalman filtering paradigm to
estimate DPOAE signal levels. The process consists of three
steps: (i) modelling of the measured signal and background
noise; (ii) processing the measured signal using the properties
of the model to reduce the number of computational opera-
tions and processing time; and (iii) post processing that
includes signal presentation and recording.4
The Kalman filter represents the optimal solution to the
discrete-data linear filtering problem9,10 namely the instanta-
neous pressure state at the probe microphone due to the two
primary tones and the DPOAE. It recursively conditions the
current state estimate on all past measurements, and converges
very rapidly to the least square error. The optimal estimate is
equal to the best prediction of its value before the measure-
ment is taken, plus an optimally weighed correction factor
times the difference between the actual measurement and the
best prediction of its value before it is taken (called the
residual). If the a priori estimate error covariance is small,
the adaptive gain factor is small, and the gain weighs the
residual less heavily. Thus, any uncorrelated process or mea-
surement noise is filtered out. This process out-performs
traditional frequency and synchronous time averaging algo-
rithms in its capacity to track DPOAE level in real time
(Harrison et al., submitted). Empirical projections predict a
processing time for analogue-to-digital conversion and for
detection of DPOAE signal in noise in the order of 1 ms.
contr alat e r al dpoa e suppr e s s i on
To elicit contralateral suppression, real time DPOAE record-
ings were made with simultaneous presentation of a broad
band noise stimulus to the opposite ear. In three subjects,
narrow band noise signals (1/3 octave, centred at 1, 2, 3, 4 and
6 kHz) and pure tones (range 2–8 kHz) were also used as
contralateral stimuli. Stimuli were applied intermittently via
an Etymotic Research ER-2 transducer (Etymotic Research,
Elk Grove Village, IL). The signal duration was 300–1000 ms;
the ‘off’ period between stimuli was approximately 50%
greater than the stimulus duration. Rise/fall time was 0 for
noise signals and 4 ms for tones. Intensity of the stimulus was
set between 0 and 80 dB SPL. Stimulus onset was recorded in
synchrony with DPOAE recording.
cal i br at ion
Calibration of the DPOAE probe was performed in a 2-ml
coupler using a spectrum analyser (SR760 FFT, Stanford
Research Systems, Sunnyvale, CA). The contralateral broad
band noise stimulus was calibrated with a sound level meter.
Processing time of the Vivo 600DPR was calibrated using a
modulated acoustic signal in the 2 ml coupler, and found to be
in the order of 1–2 ms. This matched the empirical estimation
of the model as described above.
analy s i s o f r e sult s
DPOAEs recorded without contralateral stimulation were
plotted as DP-grams (i.e. DPOAE level versus f2 frequency).
Real time DPOAE recordings before and during contralateral
stimulation were analysed to reveal the magnitude of DPOAE
suppression at different intensities of contralateral stimulus.
The onset latency period between the start of contralateral
stimulation and the onset of DPOAE suppression was mea-
sured on averaged recordings. Tests for statistical significance
were made with the t-test using SIGMA STAT (SPSS,
Chicago, IL).
Results
dpoa e m ea su r em ent
Real time DPOAEs were recorded in all 20 ears. Examples of
DP-grams (i.e. DPOAE versus f2) from six ears in four
subjects are shown in Fig. 1. The emission level ranges from
0 to 27 dB SPL (solid lines) in these examples, which was a
characteristic finding. Typically OAE level was >10 dB above
the noise floor (dashed lines). Emissions were highest in the
youngest subject (age 9 days) and tended to be lowest in the
oldest subjects.
Figure 2 illustrates some typical forms of spontaneous
variation that occurred in OAE levels over different time-
courses. Amplitude changes of around 0.25 dB occurred every
1–2 ms (Fig. 2a) and those of around 1–2 dB occurred over
periods of 1–2 s (Fig. 2b). Regularly repeating larger spikes of
2–5 dB occasionally occurred at the same rate as the heart rate
Figure 1. DP-Grams from six ears in four subjects. DPOAE level(2f1-f2) (solid lines) and noise floor (dashed lines) are plotted againstf2 frequency.
Real time DPOAE suppression 107
# 2002 Blackwell Science Ltd, Clinical Otolaryngology, 27, 106–112
and are considered consistent with changes in OAE transmis-
sion from pulses of blood flow through the middle ear, altering
middle ear compliance (Fig. 2c). Occasionally, OAE levels
fluctuated in phase with respiratory rate. Although DPOAE
levels were typically stable during recordings (up to 3 min
duration), a downward drift of around 1–2 dB/h was usually
found over a whole recording session. This occurred inde-
pendently of contralateral stimulation.
det ect ion of contr alat e r al suppr e s s i on
The effects of intermittent contralateral stimulation on real
time DPOAE levels are shown in Fig. 3a. These recordings are
from three subjects at three different values of f2 (2.2, 4.4 and
6.6 kHz) with a 1-s duration contralateral broad band noise
signal of 50 dB SPL (indicated by the uppermost bar symbols).
In each case, a suppression of DPOAE can be seen to occur
following a short latency after onset of the contralateral
stimulus. At contralateral stimulus offset, DPOAE values
return to prestimulus levels with a similar latency. When
the magnitude of suppression was small compared with spon-
taneous variation in the DPOAE signal, averaging of multiple
responses from repeated stimulation was used. Figure 3b
shows examples of such averages corresponding to the adja-
cent real time traces. When no response was visible in an
averaged trace from a 3 min recording (up to 70 stimulations),
contralateral suppression was considered not to be detectable.
In the majority of cases, between one and 10 stimulations were
required to provide adequate data for quantification of the
suppression response. Using these criteria, suppression of real
time DPOAEs by contralateral broad band noise was found in
nine out of 10 subjects. In two subjects (ages 35 and 48 years),
suppression was only detected in one ear. Maximal suppres-
sion usually occurred when measuring the distortion product
at f2¼ 4.4 kHz. Suppression was also seen at f2¼ 2.2, 3.1, 5.4
and 6.6 kHz in some ears. Suppression was not detectable
when the DPOAE level was low, for example below 12 dB
SPL for f2¼ 4.4 kHz.
magn itude of contr alat e r al suppr e s s i on
The magnitude of suppression was found to vary between
subjects. For example, the averages in Fig. 3b show suppres-
sion of 4.0, 1.3 and 0.35 dB. However, there was little intra-
subject variation in suppression magnitude both between and
within recording sessions. This is shown clearly in the two
examples illustrated in Fig. 4, in which DPOAE magnitude
with and without contralateral stimulation is plotted in
successive recordings. In Fig. 4a, mean suppression¼1.15 dB� 0.13 SD in six successive 3-min recordings, and
in Fig. 4b, from the same subject 1 week later, mean suppres-
sion¼ 1.01 dB� 0.09 SD in eight successive recordings.
Although DPOAE level varies during each session (Fig. 5a,
mean¼ 18.7 dB SPL� 0.32 SD; Fig. 5b, 16.9 dB SPL�1.2 SD), the DPOAE level during contralateral stimulation
varies coincidentally, and there is no significant difference in
suppression magnitude between the two sessions (P> 0.05,
t-test).
Figure 5 shows the increase in magnitude of suppression
with increasing intensity of contralateral stimulus from one
recording session, a finding common to all subjects. In this
example, a small suppression response was seen with 10 dB
SPL stimulation, though more commonly �30dB SPL was
required to elicit suppression. There was no evidence of
significant acoustic cross-talk at intensities below 80 dB SPL.
To date we have not found a consistent suppression
response to contralateral pure tones in humans (though in a
parallel animal study, pure tone suppression has been found in
chinchilla7). However, contralateral stimulation with narrow
band noise at 50–60 dB SPL elicited DPOAE suppression in
the three subjects tested.
Despite the variation in magnitude of suppression between
subjects, the shape of the suppressed emission recording
remained consistent. As shown in Fig. 6, a single exponential
function fits both the suppression time-course and the recov-
ery phase.
Figure 2. Real time DPOAE traces at different time scales to showtypes of spontaneous variation in DPOAE signal levels (recordingsfrom one subject at f2¼ 4.4 kHz). Interval of 1.15 s between spikesin lowest trace (c) correlates with observed heart rate of 70 bpm.
# 2002 Blackwell Science Ltd, Clinical Otolaryngology, 27, 106–112
108 A.L. James et al.
Figure 3. (a) Real time DPOAE recordings from three subjects during intermittent contralateral broad band noise (indicated by upperhorizontal bar). Recordings are at f2¼ (i) 6.6 kHz; (ii) 2.2 kHz; and (iii) 4.4 kHz. (b) Synchronized averages of 3-min recordings, whichincluded responses from up to 70 contralateral stimulus presentations. The averages are taken from the same recordings as the adjacent realtime examples.
Figure 4. Graphs from two separate recording sessions in onesubject (DM) showing stability of DPOAE level in successiverecordings (filled symbols). The suppressed DPOAE level duringcontralateral stimulation (open symbols) mirrors the unsuppressedlevel, showing stability of the magnitude of suppression. (Record-ings at f2¼ 4.4 kHz; longer periods between data-points were spentrecording DPOAEs at other frequencies).
Figure 5. Graph of magnitude of DPOAE suppression versusintensity of contralateral stimulus showing growth of suppressionwith larger stimuli.
# 2002 Blackwell Science Ltd, Clinical Otolaryngology, 27, 106–112
Real time DPOAE suppression 109
lat ency of contr alat e r al suppr e s s i on
As was seen in Fig. 3, suppression of the DPOAE consistently
occurred after a short latency. To measure this latency, aver-
aged data from 3-min recordings were obtained to minimize
interference from spontaneous variation in the DPOAE signal.
An example is shown in Fig. 7 (upper), with the period of onset
of contralateral DPOAE suppression shown below on an
expanded timescale (synchronized average of 70 responses;
f2¼ 4.4 kHz; contralateral broad band noise 1 s duration at
50 dB SPL). Onset latencies were measured at the intercept
of linear fits, as illustrated (lower). In this example, onset
latency is 39 ms. The median onset latency from 20 three-
min recordings in four ears (f2¼ 4.4 kHz) was 43 ms (range
31–95 ms).
Discussion
r eal t im e dpoa e m ea su r em ent
It is possible to measure DPOAEs in real time, without the
time averaging that is the standard method of signal detection
in conventional DPOAE recording devices. Otoacoustic emis-
sions were detectable in all our subjects and, as with conven-
tional measurement, DPOAE levels tended to be higher in the
younger subjects.1 Emissions were highest in the youngest
subject; a neonate age 9 days. Previous studies have shown
real time DPOAE levels are similar to measurements made
with conventional devices and have similar intertest varia-
bility.6
DPOAE measurement was stable for the duration of
recording sessions (up to 45 min) but small fluctuations in
the real time signal were observed. There are obviously
multiple causes for such ‘spontaneous’ amplitude changes,
only some may relate directly to outer hair cell activity. Some
signal variability is attributable to cardiovascular and respira-
tory events. It is also likely that some variation can be
attributed to the limits of device accuracy in the presence
of fluctuating background and other physiological noise.
These variations are usually too small to prevent the detection
of DPOAE suppression in response to contralateral acoustic
stimulation.
contr alat e r al suppr e s s i on
Contralateral suppression of DPOAE was identified in all
subjects except one who had very low DPOAE levels. In
our neonatal subject, contralateral suppression was robustly
present and this finding is consistent with reports showing full
functional maturity of the olivocochlear contralateral reflex in
full-term neonates.11
We propose that the following features should be identified
to distinguish a contralateral suppression response from
Figure 6. Demonstration of exponential curve (continuous line)fitted to the falling and rising components of the DPOAEsuppression response. (Mean square error of fit<0.0001; MatlabSoftware).
Figure 7. Determination of onset latency for contralateral suppres-sion of DPOAE. Averaged suppression response from 70 contral-ateral stimulations (DPOAE at f2¼ 4.4 kHz; contralateral broadband noise stimulus at 50 dB SPL). Boxed section in upper panelshown with enlarged scale (lower). Linear regression used to fitstraight lines to baseline (first 30 ms after stimulus onset) and firstpart of suppression response (55–155 ms) (SIGMA PLOT Software).Onset latency (arrow) defined by intercept of straight lines¼ 39 ms.
# 2002 Blackwell Science Ltd, Clinical Otolaryngology, 27, 106–112
110 A.L. James et al.
normal baseline variation in real time DPOAE. First, suppres-
sion of DPOAE is synchronized with the contralateral stimu-
lus and occurs after a short latency following onset of the
contralateral stimulus. In humans, this latency is around
45 ms. Second, suppression appears to develop and decay
exponentially. Third, the suppression response is maintained
for at least the duration of the contralateral stimulus. To be
confirmed as a suppression response, maximum suppression
must be greater than spontaneous variations in DPOAE level.
Fourth, following contralateral stimulus offset, DPOAE
values return to prestimulus levels after a short delay.
Contralateral suppression of OAEs was first demonstrated
in humans in 198912 but, to date, assessment is not part of
mainstream clinical practice. Real time measurement may
allow assessment to be made more readily and should be
particularly feasible in younger subjects with larger emissions.
In many cases, suppression can be seen ‘live’ on the real
time display, and recording of three responses is often suffi-
cient to quantify magnitude and latency of suppression. Thus
contralateral suppression can be detected with only a few
seconds of recording. We consider this has potential as a
viable means of objective hearing assessment, based on the
concept that detection of contralateral suppression of
DPOAEs is indicative of intact brainstem connections and
normal function in the contralateral ear. We have shown that
the magnitude of contralateral suppression, measured with
this technique, increases with contralateral stimulus intensity,
and that suppression can be elicited with narrow band
noise. Thus frequency range-specific hearing is assessed in
the contralateral ear by this test. Limitations of the test are that
suppression is usually only detectable with a contralateral
stimulus �30 dB SPL and that OAEs are not reliably detect-
able with �40 dB HL hearing loss. However, these limitations
would not preclude its use in, for example, assessment of
those unable to co-operate with subjective tests (e.g. because
of behavioural abnormalities or functional hearing loss), or
indeed as an adjunct in neonatal hearing screening. As
presently used in neonatal hearing screening and diagnosis,
OAE measurements only indicate the functional status of
outer hair cells and middle ear.1 Thus they fail to detect
conditions affecting the inner hair cells and retrocochlear
causes of hearing loss. Assessment of contralateral suppres-
sion would improve specificity by detecting rare causes of
false negative results, such as auditory neuropathy (which
arises from inner hair cell/cochlear nerve disorders13) deaf-
ness from neonatal hyperbilirubinaemia (which probably
affects the cochlear nucleus14) and congenital auditory nerve
absence.
lat ency of contr alat e r al suppr e s s i on
Previous studies have measured the onset latency of the
olivocochlear reflex in humans with OAEs but accuracy has
been limited by the temporal resolution achievable with time
or spectral averaging. In a study by Moulin, using a conven-
tional DPOAE device, the suppression latency was found to be
<2.6 s, which was the temporal resolution of the technique.15
Using spontaneous OAEs, a value of 40–200 ms was obtain-
ed14 and with transient evoked OAEs, <40–140 ms.16 These
two studies had resolutions of 40 and 20 ms respectively. Our
measurement of mean onset latency of 45 ms is in line with
these findings, but should be more reliable given our temporal
resolution of 1–2 ms. Most of the suppression latency is
composed of the transmission time of the intercochlear neural
pathway from inner hair cells, via afferent cochlear neurones
and the cochlear nucleus, to the medial superior olive, olivo-
cochlear efferents and outer hair cells.3 Sound conduction
between the probes and both cochleas is unlikely to account
for >10 ms in total.17 As conduction along the afferent path-
way is fast (namely ABR latencies of wave II around 2 ms for
cochlea to the cochlear nucleus), and animal studies show
conduction along the myelinated efferent pathway is also
fast18, it is likely that much of the delay occurs in the
brainstem. Our findings in chinchilla suggest brainstem
transmission accounts for around 40% of the onset latency
(Harrison et al., submitted). It is likely that central synapses
are under the control of descending neurones19–21, thus levels
of attention22 might influence latency through neuromodula-
tion at the cochlear nucleus or medial superior olive.23 This
may account for the wide range of suppression latencies
observed (31–95 ms). Reduction of activity in descending
pathways is thought to occur during anaesthesia24 and anaes-
thesia may influence latency (unpublished observations on
chinchilla). We are presently assessing the utility of measuring
contralateral suppression for monitoring depth of anaesthesia
clinically.
Brainstem and cerebello-pontine angle disease affects the
magnitude of contralateral suppression25 and could be
expected to influence the suppression latency. Thus latency
measurement might have value in the monitoring of neuro-
otological conditions. The instantaneous monitoring of
cochlea function provided by real time DPOAE measurement
might assist hearing preservation during acoustic neuroma
surgery by giving the surgeon more rapid feedback of vascular
compromise to the cochlea (as has been suggested with
DPOAE phase changes26).
Conclusions
We have demonstrated the ability of a novel device to measure
DPOAE in real time, without time averaging. It provides an
instant record of changes in the emission, which has allowed
us to measure the onset and offset latencies of suppression of
DPOAEs using a contralateral acoustic stimulus. This tech-
nique is simple to perform and many potential clinical appli-
cations are envisaged.
# 2002 Blackwell Science Ltd, Clinical Otolaryngology, 27, 106–112
Real time DPOAE suppression 111
Acknowledgements
This research was supported by MRC (Canada) and the
Masonic Foundation of Ontario.
References
1 LONSBURY-MARTIN B.L., MARTIN G.K. & WHITEHEAD M.L.(1997) Distortion product otoacoustic emissions. In OtoacousticEmissions: Clinical Applications, LONSBURY-MARTIN, B.L.,MARTIN, G.K. & WHITEHEAD, M.L. (eds), pp. 83–109. Thieme,New York
2 KEMP D.T. (1979) Evidence of mechanical nonlinearity andfrequency selective wave amplification in the cochlea. Arch.Otorhinolaryngol. 224, 37–45
3 WARR W.B. (1980) Efferent components of the auditory system.Ann. Otol. Rhinol. Laryngol. Supplement 89, 114–120
4 LI X., SOKOLOV Y. & KUNOV H. (2000) Method for real-timerecording of distortion product otoacoustic emissions(DPOAEs). XXV. International Congress of Audiology; TheHague, The Netherlands, p. 158
5 YOUNG V., HARRISON R., DOLAN A. et al. (2000) Comparison ofthree distortion product otoacoustic emission devices. Can.Acoustic 28, 136–137
6 YOUNG V.W. (2000) Pre-clinical testing of real-time distortionproduct otoacoustic emission devices. Master of Health Science,University of Toronto, Toronto
7 JAMES A.L., MOUNT R.J. & HARRISON R.V. (2001) Frequencyspecificity of the contralateral cochlear efferent pathway: realtime recordings with distortion product otoacoustic emissions.American Society of Pediatric Otolaryngology 16th AnnualMeeting; Scottsdale, Arizona, p. 77
8 JAMES A.L., SZAINWALD L., MOUNT R.J. et al. (2001) Dynamicsof contralateral suppression of distortion product otoacousticemissions measured in real time. Triological Society, EasternSection Annual Meeting; Toronto, Ontario, p. 131
9 KALMAN R.E. (1960) A new approach to linear filtering andprediction problems. Transaction of the ASME – Journal ofBasic Engineering March, 35–45
10 MAYBECK P. (1979) Stochastic Models, Estimation, and Control,Academic Press, New York
11 ABDALA C., MA E. & SININGER Y.S. (1999) Maturation ofmedial efferent system function in humans. J. Acoust. Soc. Am.105, 2392–2402
12 MOTT J.B., NORTON S.J., NEELY S.T. et al. (1989) Changes inspontaneous otoacoustic emissions produced by acoustic stimu-lation of the contralateral ear. Hear. Res. 38, 229–242
13 HARRISON R.V. (1998) An animal model of auditory neuropathy.Ear Hear. 19, 355–361
14 RHEE C.K., PARK H.M. & JANG Y.J. (1999) Audiologicevaluation of neonates with severe hyperbilirubinemia usingtransiently evoked otoacoustic emissions and auditory brainstemresponses. Laryngoscope 109, 2005–2008
15 MOULIN A. & CARRIER S. (1998) Time course of the medialolivocochlear efferent effect on otoacoustic emissions inhumans. Neuroreport 9, 3741–3744
16 LIND O. (1994) Contralateral suppression of TEOAE. Attemptsto find a latency. Br. J. Audiol. 28, 219–225
17 SILVA A. & YSUNZA A. (1998) Effect of contralateral maskingon the latency of otoacoustic emissions elicited by acousticdistortion products. Int. J. Pediatr. Otorhinolaryngol. 44,125–132
18 FEX J. (1973) Neuropharmacology and potentials of the innerear. In Basic Mechanisms in Hearing, MOLLER, A.R. (ed.), pp.377–420. Academic Press, New York
19 MULDERS W.H. & ROBERTSON D. (2000) Evidence for directcortical innervation of medial olivocochlear neurones in rats.Hear. Res. 144, 65–72
20 MULDERS W.H. & ROBERTSON D. (2000) Effects on cochlearresponses of activation of descending pathways from the inferiorcolliculus. Hear. Res. 149, 11–23
21 SCATES K.W., WOODS C.I. & AZEREDO W.J. (1999) Inferiorcolliculus stimulation and changes in 2f1-f2 distortion productotoacoustic emissions in the rat. Hear. Res. 128, 51–60
22 FERBER-VIART C., DUCLAUX R., COLLET L. et al. (1995)Influence of auditory stimulation and visual attention onotoacoustic emissions. Physiol. Behav. 57, 1075–1079
23 ROBERTSON D. & MULDERS W.H. (2000) Distribution andpossible functional roles of some neuroactive peptides in themammalian superior olivary complex. Microsc. Res. Techn. 51,307–317
24 HAREL N., KAKIGI A., HIRAKAWA H. et al. (1997) The effects ofanesthesia on otoacoustic emissions. Hear. Res. 110, 25–33
25 PRASHER D., RYAN S. & LUXON L. (1994) Contralateralsuppression of transiently evoked otoacoustic emissions andneuro-otology. Br. J. Audiol. 28, 247–254
26 TELISCHI F.F., STAGNER B., WIDICK M.P. et al. (1998)Distortion-product otoacoustic emission monitoring of cochlearblood flow. Laryngoscope 108, 837–842
# 2002 Blackwell Science Ltd, Clinical Otolaryngology, 27, 106–112
112 A.L. James et al.