27
Foto: Ann-Sofi Ingman Continuous Positive Airway Pressure (CPAP) and Work of Breathing Baldvin Jónsson MD, PhD Karolinska Institutet and University Hospital, Stockholm, Sweden

Continuous Positive Airway Pressure (CPAP) and Work of Breathing

  • Upload
    others

  • View
    10

  • Download
    1

Embed Size (px)

Citation preview

Page 1: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

Foto Ann-Sofi Ingman

Continuous Positive Airway Pressure

(CPAP) and Work of Breathing

Baldvin Joacutensson MD PhD

Karolinska Institutet and University Hospital Stockholm Sweden

Static pressure-volume curves of lung (PL) Chest wall (PW)

and total respiratory system Arrows indicate static forces

Chest wall recoil is outward and lung recoil is inward

Pulmonary physiology

Lung volume is related to transpulmonary pressure

Transpulmonary pressure = alveolar pressure - pleural pressure

Compliance is mL)cm H2O

Work of breathing = integral of pressure applied and resulting volume change (W = P dVminkg)

Preterm breathing strategy is directed

toward control of absolute lung volume

High respiratory rate to achieve auto end-expiratory-pressure

Active expiration against laryngeal narrowing

Tonic activity of diaphragm and intercostal muscles

Respiratory fatique in newborns

Diaphragmatic muscle fatique demonstarted in newborns on

CPAP and mechanical ventilation (Miller et al J Appl Physiol 1979)

Controversy exists about the relative importance of central or

peripheral mechanisms that contribute to fatique (Roussos and

Zakyntos Intensive Care Med 1996)

Hypoxia accentuates fatique during loaded breathing in an

animal model (Radell P et al J Appl Physiol 2000)

9 juni 2012 Baldvin Joacutensson 5

Work of Breathing

CONTINUOUS POSITIVE

AIRWAY PRESSURE (CPAP)

How can we help the infant

9 juni 2012 Baldvin Joacutensson 8

The Perfect CPAP

Should keep airway pressure as constant as possible to reduce

WOB

any deviations in pressure add to WOB done by the patient

FRC should rise proportionately to the increase in level of CPAP

This requirement means that the device must have highly variable

flow characteristics to be able to

accelerate flow on inspiration without time lag

on expiration pressures should not overshoot thus maintaining a low

WOB

Differences in failure rates between studies can possibly be

attributable to the performance of the devices used (dePaoli 2003)

9 juni 2012 Baldvin Joacutensson 9

5 week old male 840g spontaneous breaths with

airway pressure measured in nasopharynx

Pressure variations during spontaneous

breathingno CPAP applied

Argyle nasal prong without CPAP attached

Argyle nasal prong with 4cm CPAP

Jet flow CPAP 4cm

Total work of breathing (Campbell diagram)

Increase by

9 juni 2012 Baldvin Jonsson 13

CPAP applied via ET-tube or head box (Gregory et al N EJM 19712841333-40)

New device

Old device

Physiologic tests for WOB

CLINCAL studies comparing different devices

9 juni 2012 Baldvin Joacutensson 16

Pandit et al 2001 Ventilator vs Variable Flow

Courtney et al 2003 Variable Flow vs Variable Flow

Liptsen et al 2005 Bubble vs Variable Flow

Courtney et al 2011 Bubble vs Ventilator

VF gt Ventilator

Bubble ge ventilator when intraprong P is same

VF 1 gt VF 2

VF gt Bubble

Clinical studies using CPAP system

specification

Early CPAP DR feasability (Pediatrics 2004)

rdquoNeoPuff T-piece for CPAP

Coin Study (NEJM 2009)

System for CPAP use not specified

SUPPORT (NEJM 2010)

rdquoCPAP administered via T-piece systemrdquo

9 juni 2012 Baldvin Joacutensson 17

Compare existing systems used for

NCPAP care using simulated neonatal

breathing

Devices 2011

9 juni 2012 Baldvin Joacutensson 19

Mechanical lung model (ASL 5000)

NCPAP system

Mechanical lung model (ASL 5000)

Imposed WOB = Expiratory Imposed WOB + Inspiratory Imposed WOB

NeoPuff 4 cm H2O Breath profile from 35 kg infant

Inspiratory

WOB

(Area)

Expiratory

WOB

(Area)

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 2: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

Static pressure-volume curves of lung (PL) Chest wall (PW)

and total respiratory system Arrows indicate static forces

Chest wall recoil is outward and lung recoil is inward

Pulmonary physiology

Lung volume is related to transpulmonary pressure

Transpulmonary pressure = alveolar pressure - pleural pressure

Compliance is mL)cm H2O

Work of breathing = integral of pressure applied and resulting volume change (W = P dVminkg)

Preterm breathing strategy is directed

toward control of absolute lung volume

High respiratory rate to achieve auto end-expiratory-pressure

Active expiration against laryngeal narrowing

Tonic activity of diaphragm and intercostal muscles

Respiratory fatique in newborns

Diaphragmatic muscle fatique demonstarted in newborns on

CPAP and mechanical ventilation (Miller et al J Appl Physiol 1979)

Controversy exists about the relative importance of central or

peripheral mechanisms that contribute to fatique (Roussos and

Zakyntos Intensive Care Med 1996)

Hypoxia accentuates fatique during loaded breathing in an

animal model (Radell P et al J Appl Physiol 2000)

9 juni 2012 Baldvin Joacutensson 5

Work of Breathing

CONTINUOUS POSITIVE

AIRWAY PRESSURE (CPAP)

How can we help the infant

9 juni 2012 Baldvin Joacutensson 8

The Perfect CPAP

Should keep airway pressure as constant as possible to reduce

WOB

any deviations in pressure add to WOB done by the patient

FRC should rise proportionately to the increase in level of CPAP

This requirement means that the device must have highly variable

flow characteristics to be able to

accelerate flow on inspiration without time lag

on expiration pressures should not overshoot thus maintaining a low

WOB

Differences in failure rates between studies can possibly be

attributable to the performance of the devices used (dePaoli 2003)

9 juni 2012 Baldvin Joacutensson 9

5 week old male 840g spontaneous breaths with

airway pressure measured in nasopharynx

Pressure variations during spontaneous

breathingno CPAP applied

Argyle nasal prong without CPAP attached

Argyle nasal prong with 4cm CPAP

Jet flow CPAP 4cm

Total work of breathing (Campbell diagram)

Increase by

9 juni 2012 Baldvin Jonsson 13

CPAP applied via ET-tube or head box (Gregory et al N EJM 19712841333-40)

New device

Old device

Physiologic tests for WOB

CLINCAL studies comparing different devices

9 juni 2012 Baldvin Joacutensson 16

Pandit et al 2001 Ventilator vs Variable Flow

Courtney et al 2003 Variable Flow vs Variable Flow

Liptsen et al 2005 Bubble vs Variable Flow

Courtney et al 2011 Bubble vs Ventilator

VF gt Ventilator

Bubble ge ventilator when intraprong P is same

VF 1 gt VF 2

VF gt Bubble

Clinical studies using CPAP system

specification

Early CPAP DR feasability (Pediatrics 2004)

rdquoNeoPuff T-piece for CPAP

Coin Study (NEJM 2009)

System for CPAP use not specified

SUPPORT (NEJM 2010)

rdquoCPAP administered via T-piece systemrdquo

9 juni 2012 Baldvin Joacutensson 17

Compare existing systems used for

NCPAP care using simulated neonatal

breathing

Devices 2011

9 juni 2012 Baldvin Joacutensson 19

Mechanical lung model (ASL 5000)

NCPAP system

Mechanical lung model (ASL 5000)

Imposed WOB = Expiratory Imposed WOB + Inspiratory Imposed WOB

NeoPuff 4 cm H2O Breath profile from 35 kg infant

Inspiratory

WOB

(Area)

Expiratory

WOB

(Area)

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 3: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

Pulmonary physiology

Lung volume is related to transpulmonary pressure

Transpulmonary pressure = alveolar pressure - pleural pressure

Compliance is mL)cm H2O

Work of breathing = integral of pressure applied and resulting volume change (W = P dVminkg)

Preterm breathing strategy is directed

toward control of absolute lung volume

High respiratory rate to achieve auto end-expiratory-pressure

Active expiration against laryngeal narrowing

Tonic activity of diaphragm and intercostal muscles

Respiratory fatique in newborns

Diaphragmatic muscle fatique demonstarted in newborns on

CPAP and mechanical ventilation (Miller et al J Appl Physiol 1979)

Controversy exists about the relative importance of central or

peripheral mechanisms that contribute to fatique (Roussos and

Zakyntos Intensive Care Med 1996)

Hypoxia accentuates fatique during loaded breathing in an

animal model (Radell P et al J Appl Physiol 2000)

9 juni 2012 Baldvin Joacutensson 5

Work of Breathing

CONTINUOUS POSITIVE

AIRWAY PRESSURE (CPAP)

How can we help the infant

9 juni 2012 Baldvin Joacutensson 8

The Perfect CPAP

Should keep airway pressure as constant as possible to reduce

WOB

any deviations in pressure add to WOB done by the patient

FRC should rise proportionately to the increase in level of CPAP

This requirement means that the device must have highly variable

flow characteristics to be able to

accelerate flow on inspiration without time lag

on expiration pressures should not overshoot thus maintaining a low

WOB

Differences in failure rates between studies can possibly be

attributable to the performance of the devices used (dePaoli 2003)

9 juni 2012 Baldvin Joacutensson 9

5 week old male 840g spontaneous breaths with

airway pressure measured in nasopharynx

Pressure variations during spontaneous

breathingno CPAP applied

Argyle nasal prong without CPAP attached

Argyle nasal prong with 4cm CPAP

Jet flow CPAP 4cm

Total work of breathing (Campbell diagram)

Increase by

9 juni 2012 Baldvin Jonsson 13

CPAP applied via ET-tube or head box (Gregory et al N EJM 19712841333-40)

New device

Old device

Physiologic tests for WOB

CLINCAL studies comparing different devices

9 juni 2012 Baldvin Joacutensson 16

Pandit et al 2001 Ventilator vs Variable Flow

Courtney et al 2003 Variable Flow vs Variable Flow

Liptsen et al 2005 Bubble vs Variable Flow

Courtney et al 2011 Bubble vs Ventilator

VF gt Ventilator

Bubble ge ventilator when intraprong P is same

VF 1 gt VF 2

VF gt Bubble

Clinical studies using CPAP system

specification

Early CPAP DR feasability (Pediatrics 2004)

rdquoNeoPuff T-piece for CPAP

Coin Study (NEJM 2009)

System for CPAP use not specified

SUPPORT (NEJM 2010)

rdquoCPAP administered via T-piece systemrdquo

9 juni 2012 Baldvin Joacutensson 17

Compare existing systems used for

NCPAP care using simulated neonatal

breathing

Devices 2011

9 juni 2012 Baldvin Joacutensson 19

Mechanical lung model (ASL 5000)

NCPAP system

Mechanical lung model (ASL 5000)

Imposed WOB = Expiratory Imposed WOB + Inspiratory Imposed WOB

NeoPuff 4 cm H2O Breath profile from 35 kg infant

Inspiratory

WOB

(Area)

Expiratory

WOB

(Area)

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 4: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

Preterm breathing strategy is directed

toward control of absolute lung volume

High respiratory rate to achieve auto end-expiratory-pressure

Active expiration against laryngeal narrowing

Tonic activity of diaphragm and intercostal muscles

Respiratory fatique in newborns

Diaphragmatic muscle fatique demonstarted in newborns on

CPAP and mechanical ventilation (Miller et al J Appl Physiol 1979)

Controversy exists about the relative importance of central or

peripheral mechanisms that contribute to fatique (Roussos and

Zakyntos Intensive Care Med 1996)

Hypoxia accentuates fatique during loaded breathing in an

animal model (Radell P et al J Appl Physiol 2000)

9 juni 2012 Baldvin Joacutensson 5

Work of Breathing

CONTINUOUS POSITIVE

AIRWAY PRESSURE (CPAP)

How can we help the infant

9 juni 2012 Baldvin Joacutensson 8

The Perfect CPAP

Should keep airway pressure as constant as possible to reduce

WOB

any deviations in pressure add to WOB done by the patient

FRC should rise proportionately to the increase in level of CPAP

This requirement means that the device must have highly variable

flow characteristics to be able to

accelerate flow on inspiration without time lag

on expiration pressures should not overshoot thus maintaining a low

WOB

Differences in failure rates between studies can possibly be

attributable to the performance of the devices used (dePaoli 2003)

9 juni 2012 Baldvin Joacutensson 9

5 week old male 840g spontaneous breaths with

airway pressure measured in nasopharynx

Pressure variations during spontaneous

breathingno CPAP applied

Argyle nasal prong without CPAP attached

Argyle nasal prong with 4cm CPAP

Jet flow CPAP 4cm

Total work of breathing (Campbell diagram)

Increase by

9 juni 2012 Baldvin Jonsson 13

CPAP applied via ET-tube or head box (Gregory et al N EJM 19712841333-40)

New device

Old device

Physiologic tests for WOB

CLINCAL studies comparing different devices

9 juni 2012 Baldvin Joacutensson 16

Pandit et al 2001 Ventilator vs Variable Flow

Courtney et al 2003 Variable Flow vs Variable Flow

Liptsen et al 2005 Bubble vs Variable Flow

Courtney et al 2011 Bubble vs Ventilator

VF gt Ventilator

Bubble ge ventilator when intraprong P is same

VF 1 gt VF 2

VF gt Bubble

Clinical studies using CPAP system

specification

Early CPAP DR feasability (Pediatrics 2004)

rdquoNeoPuff T-piece for CPAP

Coin Study (NEJM 2009)

System for CPAP use not specified

SUPPORT (NEJM 2010)

rdquoCPAP administered via T-piece systemrdquo

9 juni 2012 Baldvin Joacutensson 17

Compare existing systems used for

NCPAP care using simulated neonatal

breathing

Devices 2011

9 juni 2012 Baldvin Joacutensson 19

Mechanical lung model (ASL 5000)

NCPAP system

Mechanical lung model (ASL 5000)

Imposed WOB = Expiratory Imposed WOB + Inspiratory Imposed WOB

NeoPuff 4 cm H2O Breath profile from 35 kg infant

Inspiratory

WOB

(Area)

Expiratory

WOB

(Area)

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 5: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

Respiratory fatique in newborns

Diaphragmatic muscle fatique demonstarted in newborns on

CPAP and mechanical ventilation (Miller et al J Appl Physiol 1979)

Controversy exists about the relative importance of central or

peripheral mechanisms that contribute to fatique (Roussos and

Zakyntos Intensive Care Med 1996)

Hypoxia accentuates fatique during loaded breathing in an

animal model (Radell P et al J Appl Physiol 2000)

9 juni 2012 Baldvin Joacutensson 5

Work of Breathing

CONTINUOUS POSITIVE

AIRWAY PRESSURE (CPAP)

How can we help the infant

9 juni 2012 Baldvin Joacutensson 8

The Perfect CPAP

Should keep airway pressure as constant as possible to reduce

WOB

any deviations in pressure add to WOB done by the patient

FRC should rise proportionately to the increase in level of CPAP

This requirement means that the device must have highly variable

flow characteristics to be able to

accelerate flow on inspiration without time lag

on expiration pressures should not overshoot thus maintaining a low

WOB

Differences in failure rates between studies can possibly be

attributable to the performance of the devices used (dePaoli 2003)

9 juni 2012 Baldvin Joacutensson 9

5 week old male 840g spontaneous breaths with

airway pressure measured in nasopharynx

Pressure variations during spontaneous

breathingno CPAP applied

Argyle nasal prong without CPAP attached

Argyle nasal prong with 4cm CPAP

Jet flow CPAP 4cm

Total work of breathing (Campbell diagram)

Increase by

9 juni 2012 Baldvin Jonsson 13

CPAP applied via ET-tube or head box (Gregory et al N EJM 19712841333-40)

New device

Old device

Physiologic tests for WOB

CLINCAL studies comparing different devices

9 juni 2012 Baldvin Joacutensson 16

Pandit et al 2001 Ventilator vs Variable Flow

Courtney et al 2003 Variable Flow vs Variable Flow

Liptsen et al 2005 Bubble vs Variable Flow

Courtney et al 2011 Bubble vs Ventilator

VF gt Ventilator

Bubble ge ventilator when intraprong P is same

VF 1 gt VF 2

VF gt Bubble

Clinical studies using CPAP system

specification

Early CPAP DR feasability (Pediatrics 2004)

rdquoNeoPuff T-piece for CPAP

Coin Study (NEJM 2009)

System for CPAP use not specified

SUPPORT (NEJM 2010)

rdquoCPAP administered via T-piece systemrdquo

9 juni 2012 Baldvin Joacutensson 17

Compare existing systems used for

NCPAP care using simulated neonatal

breathing

Devices 2011

9 juni 2012 Baldvin Joacutensson 19

Mechanical lung model (ASL 5000)

NCPAP system

Mechanical lung model (ASL 5000)

Imposed WOB = Expiratory Imposed WOB + Inspiratory Imposed WOB

NeoPuff 4 cm H2O Breath profile from 35 kg infant

Inspiratory

WOB

(Area)

Expiratory

WOB

(Area)

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 6: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

Work of Breathing

CONTINUOUS POSITIVE

AIRWAY PRESSURE (CPAP)

How can we help the infant

9 juni 2012 Baldvin Joacutensson 8

The Perfect CPAP

Should keep airway pressure as constant as possible to reduce

WOB

any deviations in pressure add to WOB done by the patient

FRC should rise proportionately to the increase in level of CPAP

This requirement means that the device must have highly variable

flow characteristics to be able to

accelerate flow on inspiration without time lag

on expiration pressures should not overshoot thus maintaining a low

WOB

Differences in failure rates between studies can possibly be

attributable to the performance of the devices used (dePaoli 2003)

9 juni 2012 Baldvin Joacutensson 9

5 week old male 840g spontaneous breaths with

airway pressure measured in nasopharynx

Pressure variations during spontaneous

breathingno CPAP applied

Argyle nasal prong without CPAP attached

Argyle nasal prong with 4cm CPAP

Jet flow CPAP 4cm

Total work of breathing (Campbell diagram)

Increase by

9 juni 2012 Baldvin Jonsson 13

CPAP applied via ET-tube or head box (Gregory et al N EJM 19712841333-40)

New device

Old device

Physiologic tests for WOB

CLINCAL studies comparing different devices

9 juni 2012 Baldvin Joacutensson 16

Pandit et al 2001 Ventilator vs Variable Flow

Courtney et al 2003 Variable Flow vs Variable Flow

Liptsen et al 2005 Bubble vs Variable Flow

Courtney et al 2011 Bubble vs Ventilator

VF gt Ventilator

Bubble ge ventilator when intraprong P is same

VF 1 gt VF 2

VF gt Bubble

Clinical studies using CPAP system

specification

Early CPAP DR feasability (Pediatrics 2004)

rdquoNeoPuff T-piece for CPAP

Coin Study (NEJM 2009)

System for CPAP use not specified

SUPPORT (NEJM 2010)

rdquoCPAP administered via T-piece systemrdquo

9 juni 2012 Baldvin Joacutensson 17

Compare existing systems used for

NCPAP care using simulated neonatal

breathing

Devices 2011

9 juni 2012 Baldvin Joacutensson 19

Mechanical lung model (ASL 5000)

NCPAP system

Mechanical lung model (ASL 5000)

Imposed WOB = Expiratory Imposed WOB + Inspiratory Imposed WOB

NeoPuff 4 cm H2O Breath profile from 35 kg infant

Inspiratory

WOB

(Area)

Expiratory

WOB

(Area)

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 7: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

CONTINUOUS POSITIVE

AIRWAY PRESSURE (CPAP)

How can we help the infant

9 juni 2012 Baldvin Joacutensson 8

The Perfect CPAP

Should keep airway pressure as constant as possible to reduce

WOB

any deviations in pressure add to WOB done by the patient

FRC should rise proportionately to the increase in level of CPAP

This requirement means that the device must have highly variable

flow characteristics to be able to

accelerate flow on inspiration without time lag

on expiration pressures should not overshoot thus maintaining a low

WOB

Differences in failure rates between studies can possibly be

attributable to the performance of the devices used (dePaoli 2003)

9 juni 2012 Baldvin Joacutensson 9

5 week old male 840g spontaneous breaths with

airway pressure measured in nasopharynx

Pressure variations during spontaneous

breathingno CPAP applied

Argyle nasal prong without CPAP attached

Argyle nasal prong with 4cm CPAP

Jet flow CPAP 4cm

Total work of breathing (Campbell diagram)

Increase by

9 juni 2012 Baldvin Jonsson 13

CPAP applied via ET-tube or head box (Gregory et al N EJM 19712841333-40)

New device

Old device

Physiologic tests for WOB

CLINCAL studies comparing different devices

9 juni 2012 Baldvin Joacutensson 16

Pandit et al 2001 Ventilator vs Variable Flow

Courtney et al 2003 Variable Flow vs Variable Flow

Liptsen et al 2005 Bubble vs Variable Flow

Courtney et al 2011 Bubble vs Ventilator

VF gt Ventilator

Bubble ge ventilator when intraprong P is same

VF 1 gt VF 2

VF gt Bubble

Clinical studies using CPAP system

specification

Early CPAP DR feasability (Pediatrics 2004)

rdquoNeoPuff T-piece for CPAP

Coin Study (NEJM 2009)

System for CPAP use not specified

SUPPORT (NEJM 2010)

rdquoCPAP administered via T-piece systemrdquo

9 juni 2012 Baldvin Joacutensson 17

Compare existing systems used for

NCPAP care using simulated neonatal

breathing

Devices 2011

9 juni 2012 Baldvin Joacutensson 19

Mechanical lung model (ASL 5000)

NCPAP system

Mechanical lung model (ASL 5000)

Imposed WOB = Expiratory Imposed WOB + Inspiratory Imposed WOB

NeoPuff 4 cm H2O Breath profile from 35 kg infant

Inspiratory

WOB

(Area)

Expiratory

WOB

(Area)

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 8: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

The Perfect CPAP

Should keep airway pressure as constant as possible to reduce

WOB

any deviations in pressure add to WOB done by the patient

FRC should rise proportionately to the increase in level of CPAP

This requirement means that the device must have highly variable

flow characteristics to be able to

accelerate flow on inspiration without time lag

on expiration pressures should not overshoot thus maintaining a low

WOB

Differences in failure rates between studies can possibly be

attributable to the performance of the devices used (dePaoli 2003)

9 juni 2012 Baldvin Joacutensson 9

5 week old male 840g spontaneous breaths with

airway pressure measured in nasopharynx

Pressure variations during spontaneous

breathingno CPAP applied

Argyle nasal prong without CPAP attached

Argyle nasal prong with 4cm CPAP

Jet flow CPAP 4cm

Total work of breathing (Campbell diagram)

Increase by

9 juni 2012 Baldvin Jonsson 13

CPAP applied via ET-tube or head box (Gregory et al N EJM 19712841333-40)

New device

Old device

Physiologic tests for WOB

CLINCAL studies comparing different devices

9 juni 2012 Baldvin Joacutensson 16

Pandit et al 2001 Ventilator vs Variable Flow

Courtney et al 2003 Variable Flow vs Variable Flow

Liptsen et al 2005 Bubble vs Variable Flow

Courtney et al 2011 Bubble vs Ventilator

VF gt Ventilator

Bubble ge ventilator when intraprong P is same

VF 1 gt VF 2

VF gt Bubble

Clinical studies using CPAP system

specification

Early CPAP DR feasability (Pediatrics 2004)

rdquoNeoPuff T-piece for CPAP

Coin Study (NEJM 2009)

System for CPAP use not specified

SUPPORT (NEJM 2010)

rdquoCPAP administered via T-piece systemrdquo

9 juni 2012 Baldvin Joacutensson 17

Compare existing systems used for

NCPAP care using simulated neonatal

breathing

Devices 2011

9 juni 2012 Baldvin Joacutensson 19

Mechanical lung model (ASL 5000)

NCPAP system

Mechanical lung model (ASL 5000)

Imposed WOB = Expiratory Imposed WOB + Inspiratory Imposed WOB

NeoPuff 4 cm H2O Breath profile from 35 kg infant

Inspiratory

WOB

(Area)

Expiratory

WOB

(Area)

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 9: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

5 week old male 840g spontaneous breaths with

airway pressure measured in nasopharynx

Pressure variations during spontaneous

breathingno CPAP applied

Argyle nasal prong without CPAP attached

Argyle nasal prong with 4cm CPAP

Jet flow CPAP 4cm

Total work of breathing (Campbell diagram)

Increase by

9 juni 2012 Baldvin Jonsson 13

CPAP applied via ET-tube or head box (Gregory et al N EJM 19712841333-40)

New device

Old device

Physiologic tests for WOB

CLINCAL studies comparing different devices

9 juni 2012 Baldvin Joacutensson 16

Pandit et al 2001 Ventilator vs Variable Flow

Courtney et al 2003 Variable Flow vs Variable Flow

Liptsen et al 2005 Bubble vs Variable Flow

Courtney et al 2011 Bubble vs Ventilator

VF gt Ventilator

Bubble ge ventilator when intraprong P is same

VF 1 gt VF 2

VF gt Bubble

Clinical studies using CPAP system

specification

Early CPAP DR feasability (Pediatrics 2004)

rdquoNeoPuff T-piece for CPAP

Coin Study (NEJM 2009)

System for CPAP use not specified

SUPPORT (NEJM 2010)

rdquoCPAP administered via T-piece systemrdquo

9 juni 2012 Baldvin Joacutensson 17

Compare existing systems used for

NCPAP care using simulated neonatal

breathing

Devices 2011

9 juni 2012 Baldvin Joacutensson 19

Mechanical lung model (ASL 5000)

NCPAP system

Mechanical lung model (ASL 5000)

Imposed WOB = Expiratory Imposed WOB + Inspiratory Imposed WOB

NeoPuff 4 cm H2O Breath profile from 35 kg infant

Inspiratory

WOB

(Area)

Expiratory

WOB

(Area)

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 10: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

Total work of breathing (Campbell diagram)

Increase by

9 juni 2012 Baldvin Jonsson 13

CPAP applied via ET-tube or head box (Gregory et al N EJM 19712841333-40)

New device

Old device

Physiologic tests for WOB

CLINCAL studies comparing different devices

9 juni 2012 Baldvin Joacutensson 16

Pandit et al 2001 Ventilator vs Variable Flow

Courtney et al 2003 Variable Flow vs Variable Flow

Liptsen et al 2005 Bubble vs Variable Flow

Courtney et al 2011 Bubble vs Ventilator

VF gt Ventilator

Bubble ge ventilator when intraprong P is same

VF 1 gt VF 2

VF gt Bubble

Clinical studies using CPAP system

specification

Early CPAP DR feasability (Pediatrics 2004)

rdquoNeoPuff T-piece for CPAP

Coin Study (NEJM 2009)

System for CPAP use not specified

SUPPORT (NEJM 2010)

rdquoCPAP administered via T-piece systemrdquo

9 juni 2012 Baldvin Joacutensson 17

Compare existing systems used for

NCPAP care using simulated neonatal

breathing

Devices 2011

9 juni 2012 Baldvin Joacutensson 19

Mechanical lung model (ASL 5000)

NCPAP system

Mechanical lung model (ASL 5000)

Imposed WOB = Expiratory Imposed WOB + Inspiratory Imposed WOB

NeoPuff 4 cm H2O Breath profile from 35 kg infant

Inspiratory

WOB

(Area)

Expiratory

WOB

(Area)

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 11: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

9 juni 2012 Baldvin Jonsson 13

CPAP applied via ET-tube or head box (Gregory et al N EJM 19712841333-40)

New device

Old device

Physiologic tests for WOB

CLINCAL studies comparing different devices

9 juni 2012 Baldvin Joacutensson 16

Pandit et al 2001 Ventilator vs Variable Flow

Courtney et al 2003 Variable Flow vs Variable Flow

Liptsen et al 2005 Bubble vs Variable Flow

Courtney et al 2011 Bubble vs Ventilator

VF gt Ventilator

Bubble ge ventilator when intraprong P is same

VF 1 gt VF 2

VF gt Bubble

Clinical studies using CPAP system

specification

Early CPAP DR feasability (Pediatrics 2004)

rdquoNeoPuff T-piece for CPAP

Coin Study (NEJM 2009)

System for CPAP use not specified

SUPPORT (NEJM 2010)

rdquoCPAP administered via T-piece systemrdquo

9 juni 2012 Baldvin Joacutensson 17

Compare existing systems used for

NCPAP care using simulated neonatal

breathing

Devices 2011

9 juni 2012 Baldvin Joacutensson 19

Mechanical lung model (ASL 5000)

NCPAP system

Mechanical lung model (ASL 5000)

Imposed WOB = Expiratory Imposed WOB + Inspiratory Imposed WOB

NeoPuff 4 cm H2O Breath profile from 35 kg infant

Inspiratory

WOB

(Area)

Expiratory

WOB

(Area)

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 12: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

New device

Old device

Physiologic tests for WOB

CLINCAL studies comparing different devices

9 juni 2012 Baldvin Joacutensson 16

Pandit et al 2001 Ventilator vs Variable Flow

Courtney et al 2003 Variable Flow vs Variable Flow

Liptsen et al 2005 Bubble vs Variable Flow

Courtney et al 2011 Bubble vs Ventilator

VF gt Ventilator

Bubble ge ventilator when intraprong P is same

VF 1 gt VF 2

VF gt Bubble

Clinical studies using CPAP system

specification

Early CPAP DR feasability (Pediatrics 2004)

rdquoNeoPuff T-piece for CPAP

Coin Study (NEJM 2009)

System for CPAP use not specified

SUPPORT (NEJM 2010)

rdquoCPAP administered via T-piece systemrdquo

9 juni 2012 Baldvin Joacutensson 17

Compare existing systems used for

NCPAP care using simulated neonatal

breathing

Devices 2011

9 juni 2012 Baldvin Joacutensson 19

Mechanical lung model (ASL 5000)

NCPAP system

Mechanical lung model (ASL 5000)

Imposed WOB = Expiratory Imposed WOB + Inspiratory Imposed WOB

NeoPuff 4 cm H2O Breath profile from 35 kg infant

Inspiratory

WOB

(Area)

Expiratory

WOB

(Area)

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 13: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

Physiologic tests for WOB

CLINCAL studies comparing different devices

9 juni 2012 Baldvin Joacutensson 16

Pandit et al 2001 Ventilator vs Variable Flow

Courtney et al 2003 Variable Flow vs Variable Flow

Liptsen et al 2005 Bubble vs Variable Flow

Courtney et al 2011 Bubble vs Ventilator

VF gt Ventilator

Bubble ge ventilator when intraprong P is same

VF 1 gt VF 2

VF gt Bubble

Clinical studies using CPAP system

specification

Early CPAP DR feasability (Pediatrics 2004)

rdquoNeoPuff T-piece for CPAP

Coin Study (NEJM 2009)

System for CPAP use not specified

SUPPORT (NEJM 2010)

rdquoCPAP administered via T-piece systemrdquo

9 juni 2012 Baldvin Joacutensson 17

Compare existing systems used for

NCPAP care using simulated neonatal

breathing

Devices 2011

9 juni 2012 Baldvin Joacutensson 19

Mechanical lung model (ASL 5000)

NCPAP system

Mechanical lung model (ASL 5000)

Imposed WOB = Expiratory Imposed WOB + Inspiratory Imposed WOB

NeoPuff 4 cm H2O Breath profile from 35 kg infant

Inspiratory

WOB

(Area)

Expiratory

WOB

(Area)

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 14: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

Clinical studies using CPAP system

specification

Early CPAP DR feasability (Pediatrics 2004)

rdquoNeoPuff T-piece for CPAP

Coin Study (NEJM 2009)

System for CPAP use not specified

SUPPORT (NEJM 2010)

rdquoCPAP administered via T-piece systemrdquo

9 juni 2012 Baldvin Joacutensson 17

Compare existing systems used for

NCPAP care using simulated neonatal

breathing

Devices 2011

9 juni 2012 Baldvin Joacutensson 19

Mechanical lung model (ASL 5000)

NCPAP system

Mechanical lung model (ASL 5000)

Imposed WOB = Expiratory Imposed WOB + Inspiratory Imposed WOB

NeoPuff 4 cm H2O Breath profile from 35 kg infant

Inspiratory

WOB

(Area)

Expiratory

WOB

(Area)

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 15: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

Compare existing systems used for

NCPAP care using simulated neonatal

breathing

Devices 2011

9 juni 2012 Baldvin Joacutensson 19

Mechanical lung model (ASL 5000)

NCPAP system

Mechanical lung model (ASL 5000)

Imposed WOB = Expiratory Imposed WOB + Inspiratory Imposed WOB

NeoPuff 4 cm H2O Breath profile from 35 kg infant

Inspiratory

WOB

(Area)

Expiratory

WOB

(Area)

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 16: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

9 juni 2012 Baldvin Joacutensson 19

Mechanical lung model (ASL 5000)

NCPAP system

Mechanical lung model (ASL 5000)

Imposed WOB = Expiratory Imposed WOB + Inspiratory Imposed WOB

NeoPuff 4 cm H2O Breath profile from 35 kg infant

Inspiratory

WOB

(Area)

Expiratory

WOB

(Area)

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 17: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

Mechanical lung model (ASL 5000)

NCPAP system

Mechanical lung model (ASL 5000)

Imposed WOB = Expiratory Imposed WOB + Inspiratory Imposed WOB

NeoPuff 4 cm H2O Breath profile from 35 kg infant

Inspiratory

WOB

(Area)

Expiratory

WOB

(Area)

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 18: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

Mechanical lung model (ASL 5000)

Imposed WOB = Expiratory Imposed WOB + Inspiratory Imposed WOB

NeoPuff 4 cm H2O Breath profile from 35 kg infant

Inspiratory

WOB

(Area)

Expiratory

WOB

(Area)

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 19: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

Breath profiles used

9 juni 2012 Baldvin Joacutensson 22

Recorded flow from a healthy male

newborn with a body weight of 34 kg

Sample from Moa et al

Recorded flow from a healthy

preterm (26 weeks) female at 28

weeks corrected age with a weight of

13 kg on recording

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 20: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

9 juni 2012 Baldvin Joacutensson 23

Drevhammar T Nilsson K Zetterstroumlm H Jonsson BComparison of seven infant continuous

positive airway pressure systems using simulated neonatal breathing Pediatr Crit Care Med

2012 13e113ndash e119

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 21: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

Effect of leak

9 juni 2012 Baldvin Joacutensson 24

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 22: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

System summary

9 juni 2012 Baldvin Joacutensson 25

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 23: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

Conclusions

Large variations in pressure stability and iWOB exist between

the different devices

Clinical significance if any needs to be assessed

9 juni 2012 Baldvin Joacutensson 26

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011

Page 24: Continuous Positive Airway Pressure (CPAP) and Work of Breathing

9 juni 2012 Baldvin Joacutensson 27

Nikischin et al Pediatr Crit Care Med 2011