1701 - 11/04/2017 10:00 - Medical Equipment AuctionSale
11-04-2017 10:00
Condition Report
Sale Date
These reports are compiled by our Medical Engineering Team and are for guidance only - all lots are sold asseen, as per our Conditions of Sale. We recommend that wherever possible you inspect lots prior to purchase.
Malmesbury, Wiltshire, UKLocation:
Viewing:
Removal:
Friday 7th April 12pm-5pm, Monday 10th April 10am-6pm and morning of the sale from8.30amBy 4pm on Tuesday 18th April 2017 - Storage charges will be incurred after this date
Spacelabs Ultracare Anaesthetic Machine with Ultracare SLP Monitor
The anaesthetic machine powers up and completes the system self test successfully.Ventilator display screen and control panel are in good working order. Flowmeterassembly, breathing system, bellows and absorber are all intact and the medical gashoses are included. The battery is also working (at the time of test). Ventilation modesinclude Volume Control, Pressure Control, SIMV and PSVPro. The patient monitor(YOM 2003) also appears to be fully functional with display screen and control panel ingood working order. Parameters include ECG, SpO2, NIBP, IBP, Temp, Spirometry andGas monitoring (CO2, O2, N2O, AA, id, p&v). Overall the unit is in good physicalcondition with minor scuff marks. The system has not been fully tested. No patientmonitoring accessories included with the unit.SN: AMRH01323. Software version: 4.5 PSVPro.
4.5 PSVPro
Lot 1AMRH01323
Software:
Serial No:
Condition Report:
Acuson Aspen Advanced Ultrasound to include Acuson L10 Transducer Serial
This ultrasound had the hard drive formatted due to patient data found. Thereforeunable to perform any further tests.Transducer 4C1, serial 01516202 is in good physical and cosmetic condition.Transducer C7, serial 80102309 has some light crystal membrane descoloration andplastic cracks on its locking handle mechanism.Transducer L10, serial number 74700298 has initial signs of cracks on its plastic lockinghandle mechanism.Transducer L5, serial 80104906 is in good physical and cosmetic condition.Castors, breaks and overall physical condition is good. A few scuffs noticed.
Lot 2
Condition Report:
Toshiba Xario Ultrasound with No Probes - Serial No. CIC M8T H8LLot 399E0726503Serial No:
Page 1 Updated 10 April 2017
Ultrasound powers up and the system initiation completed successfully.CRT display, trackball, keyboard, LCD touch screen, control console function keys,sliders and knobs are in good working order.Probe identification and selection is working fine. The unit was tested with phantom andsatisfactory image quality was displayed along with average grey scale and resolution.Accurate caliper measurement was obtained.All of the available modes (M, PW, Power, CDI, Dynamic Flow and 2D) were tested andare working fine. Some sample images have been stored.Printer is in good working order and two of the stored images been printed.The DVD drive seems working ok but not fully tested.Cart, castors and brakes are in good condition.This unit has no transducers and is in satisfactory overall condition.
Software DVD's:SSDM30-(00374, 00358, 00350, 00348)
Foot pedal
Accessories:
Condition Report:
Acuson Sequoia C512 Ultrasound With 4 VIC Transducer, 2.0mhz Doppler and
This ultrasound hard drive had to be formatted due to patiend data found.The ultrasoud seemed to work fine but had not been tested in full.Console function keys has a crack on the right corner of its plastic enclosure.The probe 4V1C, serial number 41111502 is in good physical and cosmetic condition.Castors and breaks are fully functional.Overal the ultrasound is in good physical condition with few scuffs.
Lot 4
Condition Report:
Philips Heartstart MRX Defibrillator - 10055/Serial No. U500214850
The defibrillator powers up and successfully completes its initial self-test. Display,control panel function keys and therapy knob are all in good working order. AED andvoice commands are functional. Printer is functional. Backup battery not working at thetime of testing. Overall the unit is in fairly good physical condition.
Lot 8
Condition Report:
Draeger Oxlog 3000 Ventilator - Serial No. W031769
The ventillator powers and completes its initial self-test successfully. All buttons anddisplay are functional. Modes include: IPPV, SIMV ASB, CPAP ASB & BIPAP ASBwhich appear to be in working order but were not tested. Backup battery was functionalat the time of testing. Power adapter is not included. Overall the ventillator is insatisfactory physical condition.
1.22
Lot 9
Software:Condition Report:
Page 2 Updated 10 April 2017
Osiris Ventilator
The unit shows signs of wear and tear. There seems to be a loose part inside the unit.Rotary knobs and gauge are intact, but overall the unit is in poor cosmetic condition.Not tested.
Lot 14
Condition Report:
Sonosite M-Turbo Ultrasound With 8.5 MHz Probe
This unit has no batteries, no external power supply and asks for software license keywhen powered up (using spare power supply).Its only one probe has deep dent on plastic enclosure, border with crystal membrane.Unable to perform further tests.
Transducer C11X / 8-5, serial 03F32D, DoM 12/2008.
Lot 1903D47M08 2008Manufactured:
Accessories:
Serial No:
Condition Report:
Critikon Dinamp Pro 1000 Patient Monitor
Unit powers up and successfully completes its initial self-test. Display, function keys androtory knob are all in good working order. Alarms functional. Printer was not tested.NIBP pump failed to work after initiating a measurement. Backup battery was functionalat the time of testing. Overall the monitor is in satisfactory physical condition.
Lot 25
Condition Report:
Philips iU22 Ultrasound and Sony Printer (H Drive Removed)
This ultrasound had the hard drive been removed by the trust.Printer and DVD drives seems working fine but not been tested fully.LCD monitor is working ok. LCD monitor arm is with some fatige and cannot holdmonitor up. Sideways positioning is working ok.Front panel console keys , knobs, sliders and positioning mechanism, all are workingfine.This unit has no probes.Front console function keys have signs of cracks on top corner of its plastic enclosure.Castors, breaks and overall physical condition is good.
Lot 30
Condition Report:
Philips CX50 POC Ultrasound on Trolley With CS-1, L12-3, S5-1 Probes - HDD
This units hard drive have been removed by the trust.LCD screen is working fine but unit is unable to boot up.Keyboard, function keys, sliders and knobs seems to be working fine but not beentested.Also printer seems to be working ok but had not been fully tested.Transducer S5-1, serial number 453561265582, good physical and cosmetic condition.Transducer L12-3, serial number 453561616351, good physical and cosmeticcondition.transducer C5-1, serial number 453561301481, good physical condition but has smallmarks / scuffs on crystal membrane.
Lot 31
Condition Report:
Page 3 Updated 10 April 2017
This unit contains ECG leads and ethernet cable.Castors, breaks and overall condition is good.
Olympus PSD-60/Olympus Endoplasma Diathermy on Trolley With Footswitch
Unit powers up and completes its initial self-test. All the keys and display are fulllyfunctional. Footswitch was not tested. Overall it is in satisfactory physical condition.
Lot 32
Condition Report:
Datex Ohmeda S/5 Patient Monitor, Datascope Passport XG Patient Monitor,
One unit comes up with service due and battery failure, did not pass start up tests andwas therefore not tested.Another unit requires a special power adaptor that is not included, no compatible poweradaptor available therefore not tested. It is in poor cosmetic condition, scratches onunit.Third unit was able to power on, display has dark patches. Buttons work and unit is intact but was not fully tested.
Lot 33
Condition Report:
Draeger Oxlog 3000 Ventilator - Serial No. W031056
The ventillator powers and completes its initial self-test successfully. All buttons anddisplay are functional. Modes include: IPPV, SIMV ASB, CPAP ASB & BIPAP ASBwhich appear to be in working order but were not tested. Backup battery was functionalat the time of testing. Power adapter is not included. Overall the ventillator is insatisfactory physical condition.
Lot 36
Condition Report:
Stryker Arthroscopy System
Unit powers up and sucessfully completes its initial self-test. Display and keys are ingood working condition. No further tests carried out. Overall the unit is in satisfactoryphysical condition.
Lot 43
Condition Report:
Sonosite 180 Plus Portable Ultrasound with 2 Probes - C15/4-2MHz and L38/10-5
The ultrasound powers up and completes the initial self test successfully.All function buttons, rotary knobs, track ball and display are working ok.Ultrasound modes D, CPD, M and 2D all appear to be functioning well.Caliper measurements were accurate to set measurements.Unit also showed clear images with poor to reasonable resolution and poor grey scale.The probes and their membranes are in good physical condition.Battery was not working at the time of tests (not charged).Castors, trolley, breaks and overall physical condition are all good.
Transducer L38/ 10-5, serial 0325FT, DoM 12/2003;Transducer C15/ 4-2, serial number 03319G, DoM 09/2004;External power supply unit.
Lot 45031Zh312 2003Manufactured:
Accessories:
Serial No:
Condition Report:
Page 4 Updated 10 April 2017
Haag-Streit Microruptor Opthalmic Laser (No Eyepiece/Head)
This unit power on ok.PSU fan is noisy.Table hight adjustments is working fine.Slit lamp body with operating switches is in good condition.Laser control panel – with movable LCD display on the table (preset laser energy)seems working fine but has not been fully tested. Two red light dots are present at theend of optics.There is glasses and foot pedal (not tested) with the unit.
Lot 57
Condition Report:
Storz Electronic Endoflator
Unit powers up and completes its self-test. Display indicators and keys are functional.Overall it is in satisfactory physical condition.
Lot 63
Condition Report:
Storz Aida Control II Endoskope - 20046020
This unit is not booting up (hard drive been formatted or removed by the trust).Unable to perform further tests on this unit.Physical condition is good. There is a few scuffs on eclosure.
Lot 70
Condition Report:
Sonosite i Look Ultrasound on Rollstand with Probe
The ultrasound powers up and completes the initial self test successfully.All function buttons, rotary knobs, track ball and display are working ok.Unit showed poor to reasonable images with poor resolution and grey scale.The probe and its membrane is in good physical condition.Battery was working at the time of tests but needs a good charge.Castors, trolley, breaks and overall physical condition are all good.
Lot 82039D4W02 2008Manufactured:
Serial No:
Condition Report:
Valleylab Ligasure x 2
Both units power on fine and complete their initialisation process. All indicators andsoftbuttons are functional. Cosmetically they are in average condition and show fewsigns of wear and tear. One of the units has a few dents on the front panel. Overall theyappear to be intact, but have not been fully tested.
Lot 83
Condition Report:
Storz Arthropump PlusLot 90
Page 5 Updated 10 April 2017
All indicators come on, unit is in tact but was not fully tested. Cosmetically in goodcondition.
Condition Report:
Aloka SSD 500 Ultrasound with Probe - ID 0026
This ultrasound powers up and passes it initial self tests.CRT screen and function console keys amongst its keyboard seems to be working fine.Probe model UST-944B-3.5 serial number 2A13917 has strong descoloration on itscrystal membrane.The ultrasound itself seems working although unable to perform further tests due toprobe not producing any image on screen and also not having any other compatibleprobe available.The ultrasound has a carring bag and probe carring hard case.The overall condition is average. The ultrasound seems aged.
Lot 91
Condition Report:
Datascope 98XT Intra-Aortic Balloon Pump System
The unit powers up and completes the system self test successfully. Display screen,control panel and printer are in good working order. The doppler is in place but thebattery cover is damaged. The doppler was not tested due to flat battery. Thecondensate removal / safety disk is in place but the helium gas cylinder is not included.Overall the unit is in satisfactory physical condition. No other accessories included.Operating hours: 14420.1.
Lot 106
Condition Report:
14420.1Operating Hours:
Zeiss IOL Master
Powers up and completes the system initialisation successfully. Display screen, joystickcontrol, chin rest adjustment and keyboard/touchpad are in good working order. Theavailable measurement modes are ALM, KER and ACD but the system does notproduce any image or readings from the test eye (showing noise and vertical lines onthe image section of the display screen). The table height adjustment (manual) isfunctional and overall the unit is in satisfactory physical condition. YOM: 2001. Softwareversion: 3.01.0294.
3.01.0294.
Lot 1072001Manufactured:
Software:Condition Report:
Storz Stack System Including Hub 1 Processor, Camera Head, Xenon 3000 Light
All units on the stack system power on. The display has scratches but is functional. Theprocessor, lightsource and pump, and camera hub are all intact. Printer powers on butwas not tested. The lightsource does not include a bulb ( or frame to hold the bulb). TheH3 camera head has a few scratches on the handle. Images produced are fine, remotebuttons appear to be functional ( One did not carry out any function).
Lot 108
Condition Report:
Page 6 Updated 10 April 2017
Diomed 630 PDT Laser with Accessories x 1
Unit power up with the numerical displays, LCD and selectors soft keys working. Laserfibre optic attachment is not included therefore output was not fully tested. System is infair physical condition but not fully tested. Laser On/Off key not included.
Version 18.03
Fibre optic calibration wellHand switch
Lot 1126300567R2001Manufactured:
Software:
Accessories:
Serial No:
Condition Report:
Pentax F1-10P2 Fibrescope
- Insertion Tube - Damaged and in poor condition- Bending mechanism & Angulation Control - ok movement- Light Guide Cable & Light Transmission - Good light transmission- Objective Lens & Focussing - //- Remote Control Switches - //- Distal End - Average condition- Image Quality - Average- Working Channel (s) -Clear- Leak Test - Failed- Overall Condition - Poor condition Dimension: 60x4.06mm
Lot 150
Condition Report:
Pentax FB10X Fibrescope
- Insertion Tube - Blemished but intact- Bending mechanism & Angulation Control - Good movement andbreaks- Light Guide Cable & Light Transmission - Good light transmission- Objective Lens & Focussing - Lens in ok condition- Remote Control Switches - //- Distal End - Average condition- Image Quality - Average- Working Channel (s) -Clear- Leak Test - Passed- Overall Condition - Average condition Dimension: 60x3.92mm
Lot 151
Condition Report:
Olympus BF Type 1T20D Videoscope x1
- Insertion Tube - Average condition- Bending mechanism & Angulation Control - ok movement, limited D
Lot 152
Condition Report:
Page 7 Updated 10 April 2017
movement , good breaks- Light Guide Cable & Light Transmission - Good light transmission- Objective Lens & Focussing - Lens in ok condition, few pixel missing- Remote Control Switches - //- Distal End - Average condition- Image Quality - Average- Working Channel (s) -Clear- Leak Test - Passed- Overall Condition - Average condition Dimension:
Karl Storz Cystoscope x 1
- Insertion Tube - Average condition- Bending mechanism & Angulation Control - No movement- Light Guide Cable & Light Transmission - N/A- Objective Lens & Focussing - Poor- Remote Control Switches - //- Distal End - Average condition- Image Quality - poor- Working Channel (s) -Clear- Leak Test - Fail- Overall Condition - Poor condition Dimension:
Lot 153
Condition Report:
Olympus CYF-230L Videocolonoscope
- Insertion Tube - Average condition- Bending mechanism & Angulation Control - Good movement + locks- Light Guide Cable & Light Transmission - Good light transmission- Objective Lens & Focussing - ok- Remote Control Switches - //- Distal End - slightly scratched- Image Quality - poor image, vertical stripes- Working Channel (s) -clear- Leak Test - Passed- Overall Condition - Average condition Dimension: 14 x 1600
Lot 154
Condition Report:
Olympus CYF-4 Endoscope
- Insertion Tube - Average condition- Bending mechanism & Angulation Control - ok movement, limited Dmovement
Lot 155
Condition Report:
Page 8 Updated 10 April 2017
- Light Guide Cable & Light Transmission - Good light transmission- Objective Lens & Focussing - Lens in ok condition, few pixel missing- Remote Control Switches - //- Distal End - Average condition- Image Quality - Poor- Working Channel (s) -Clear- Leak Test - Passed- Overall Condition - Average condition Dimension: 350 x 5.6
Olympus Type Q230 Videoscope
- Insertion Tube - Distal end cover damaged- Bending mechanism & Angulation Control - ok movement, limited Dmovement- Light Guide Cable & Light Transmission - Good light transmission- Objective Lens & Focussing - ok- Remote Control Switches - //- Distal End - Average ok- Image Quality - No image, disturbed image- Working Channel (s) -Clear- Leak Test - Failed- Overall Condition - Average condition Dimension: 1000 x 12
Lot 156
Condition Report:
Olympus OSF-2 Fibrescope - 2704910
- Insertion Tube - worn but intact- Bending mechanism & Angulation Control - ok movement, limited Dmovement- Light Guide Cable & Light Transmission - Good light transmission- Objective Lens & Focussing - Lens ok- Remote Control Switches - //- Distal End - Average condition- Image Quality - Blurry- Working Channel (s) -Clear- Leak Test - Passed- Overall Condition - Average condition Dimension: 600 x 13.2
Lot 157
Condition Report:
Olympus ENF-6P Laryngoscope - AN 12611
- Insertion Tube - Worn but intact- Bending mechanism & Angulation Control - good movement
Lot 158
Condition Report:
Page 9 Updated 10 April 2017
- Light Guide Cable & Light Transmission - N/A- Objective Lens & Focussing - Lens in ok condition- Remote Control Switches - //- Distal End - Average condition, needs clean- Image Quality - Good- Working Channel (s) -//- Leak Test - Fail- Overall Condition - Average condition Dimension: 250 x 4.3
Olympus CYF-4 Flexible Endoscope - SN W304636
- Insertion Tube - Average condition- Bending mechanism & Angulation Control - ok movement, limited Dmovement , ok breaks- Light Guide Cable & Light Transmission - Good light transmission- Objective Lens & Focussing - Lens in ok condition, few pixel missing- Remote Control Switches - //- Distal End - Average condition- Image Quality - A few black pixels visible- Working Channel (s) -Clear- Leak Test - Passed- Overall Condition - Average condition Dimension: 350 x 5.4
Lot 159
Condition Report:
Olympus CYF-4 Flexible Endoscope - SN W304206
- Insertion Tube - Average condition- Bending mechanism & Angulation Control - ok movement- Light Guide Cable & Light Transmission - Good light transmission- Objective Lens & Focussing - Lens in ok condition, few pixel missing- Remote Control Switches - //- Distal End - Average condition- Image Quality - Few black pixels visible- Working Channel (s) -Clear- Leak Test - Failed- Overall Condition - Average condition Dimension: 350 x 5.3
Lot 160
Condition Report:
Olympus Video Scope GF-VE 260 AL5
- Insertion Tube - Average condition- Bending mechanism & Angulation Control - slightly limited, good locks- Light Guide Cable & Light Transmission - Good light transmission
Lot 161
Condition Report:
Page 10 Updated 10 April 2017
- Objective Lens & Focussing - Lens in ok condition- Remote Control Switches - One of the buttons did not work (enhance)- Distal End - Average condition- Image Quality - Average- Working Channel (s) -Clear- Leak Test - Passed- Overall Condition - Average condition Dimension: 1200 x 14.9
Olympus Video Scope GF-VE 260 AL5
- Insertion Tube - Good condition- Bending mechanism & Angulation Control - slightly limited, good locks- Light Guide Cable & Light Transmission - Good light transmission- Objective Lens & Focussing - Lens in ok condition- Remote Control Switches - Functional- Distal End - Average condition- Image Quality - Average- Working Channel (s) -Clear- Leak Test - Passed- Overall Condition - Average condition Dimension: 1200 x 14.9
Lot 162
Condition Report:
Pentax FI-1OBS Intubation Fibrescope
- Insertion Tube - Slightly worn- Bending mechanism & Angulation Control - good, good locks- Light Guide Cable & Light Transmission - Not tested- Objective Lens & Focussing - Lens in ok condition- Remote Control Switches - //- Distal End - Average condition- Image Quality - Poor- Working Channel (s) -//- Leak Test - Passed- Overall Condition - Poor Dimension: 600 x 3.8
Lot 163
Condition Report:
DP Medical TF6 Flexible Scope
- Insertion Tube - Good condition- Bending mechanism & Angulation Control - Good- Light Guide Cable & Light Transmission - //- Objective Lens & Focussing - ok- Remote Control Switches - //
Lot 164
Condition Report:
Page 11 Updated 10 April 2017
- Distal End - Average condition- Image Quality - Average but with patches ( possibly liquid )- Working Channel (s) -//- Leak Test - //- Overall Condition - Average condition Dimension: 300 x 3.83
Pentax FB-15v Bronchoscope
- Insertion Tube - Dents- Bending mechanism & Angulation Control - good, good locks- Light Guide Cable & Light Transmission - Good light transmission- Objective Lens & Focussing - Lens in ok condition- Remote Control Switches - Functional- Distal End - Average condition- Image Quality - Average- Working Channel (s) -Clear- Leak Test - Passed- Overall Condition - Average condition Dimension:
Lot 165
Condition Report:
Olympus JF Type 240 Videoscope
- Insertion Tube - Good condition- Bending mechanism & Angulation Control - good, good locks- Light Guide Cable & Light Transmission - Good light transmission- Objective Lens & Focussing - Lens in ok condition- Remote Control Switches - Functional- Distal End - Average condition- Image Quality - Average- Working Channel (s) -Clear- Leak Test - Passed- Overall Condition - Average condition Dimension: 110 x 11.7
Lot 166
Condition Report:
Olympus CF240AL Videoscope
- Insertion Tube - Dents- Bending mechanism & Angulation Control - good, good locks- Light Guide Cable & Light Transmission - Good light transmission- Objective Lens & Focussing - Lens in ok condition- Remote Control Switches - Functional- Distal End - Good condition- Image Quality - Good
Lot 167
Condition Report:
Page 12 Updated 10 April 2017
- Working Channel (s) -Clear- Leak Test - Passed- Overall Condition - Good condition Dimension: 1000 x 13.3
Olympus GIFXQ240 Videoscope
- Insertion Tube - Good condition- Bending mechanism & Angulation Control - limited D movement, goodlocks- Light Guide Cable & Light Transmission - Good light transmission- Objective Lens & Focussing - Lens in ok condition- Remote Control Switches - Functional- Distal End - Good condition- Image Quality - Good- Working Channel (s) -Clear- Leak Test - Failed- Overall Condition - Average condition, leaks Dimension: 1000 x 10.2
Lot 168
Condition Report:
Olympus CF240AL Videoscope
- Insertion Tube - Average condition, blemishes on handle- Bending mechanism & Angulation Control - good, good locks- Light Guide Cable & Light Transmission - Good light transmission- Objective Lens & Focussing - see comment on image below- Remote Control Switches - Functional- Distal End - Average condition- Image Quality - Poor, vertical stripes- Working Channel (s) -Clear- Leak Test - Passed- Overall Condition - Poor condition as image is not good Dimension: 1600 x 12.9
Lot 169
Condition Report:
Olympus GIFXQ240 Videoscope
- Insertion Tube - Ok- Bending mechanism & Angulation Control - limited D movement, goodlocks- Light Guide Cable & Light Transmission - Good light transmission- Objective Lens & Focussing - Lens in ok condition- Remote Control Switches - Functional- Distal End - Good condition- Image Quality - Good
Lot 170
Condition Report:
Page 13 Updated 10 April 2017
- Working Channel (s) -Clear- Leak Test - Passed- Overall Condition - Good condition Dimension: 1000 x 10.2
Olympus CF 1T200L Videoscope
- Insertion Tube - Blemishes on handle- Bending mechanism & Angulation Control - good, good locks- Light Guide Cable & Light Transmission - Good light transmission- Objective Lens & Focussing - Lens in ok condition- Remote Control Switches - Functional- Distal End - Good condition- Image Quality - Good- Working Channel (s) -Clear- Leak Test - Passed- Overall Condition - Good condition Dimension: 1600 x 15.3
Lot 171
Condition Report:
Olympus CF200S Videoscope
- Insertion Tube - Blemishes on tube- Bending mechanism & Angulation Control - good, good locks- Light Guide Cable & Light Transmission - Good light transmission- Objective Lens & Focussing - Lens in ok condition- Remote Control Switches - Functional- Distal End - Good condition- Image Quality - Good- Working Channel (s) -Clear- Leak Test - Very small leak- Overall Condition - Good performance, cosmetically in averagecondition, leaks Dimension: 600 x 14.9
Lot 172
Condition Report:
Olympus JF 1T20 Flexiscope
- Insertion Tube - Average with blemishes- Bending mechanism & Angulation Control - average, loose locks on L+R- Light Guide Cable & Light Transmission - Good light transmission, tubeblemished- Objective Lens & Focussing - Lens in ok condition- Remote Control Switches - //- Distal End - Average condition
Lot 173
Condition Report:
Page 14 Updated 10 April 2017
- Image Quality - ok but with black pixels- Working Channel (s) -Clear- Leak Test - Passed- Overall Condition - Average condition Dimension: 1100 x 12.2
Olympus TJF 1T20 Flexiscope
- Insertion Tube - Good- Bending mechanism & Angulation Control - Limited D movement, looselocks- Light Guide Cable & Light Transmission - Good light transmission, tubeblemished- Objective Lens & Focussing - Lens in ok condition- Remote Control Switches - //- Distal End - Good condition- Image Quality - Good, very few black pixels- Working Channel (s) -Clear- Leak Test - Passed- Overall Condition - OK condition Dimension: 1200 x 13.9
Lot 174
Condition Report:
Olympus JF 1T30 Flexiscope
- Insertion Tube - Good- Bending mechanism & Angulation Control - Limited movement, goodlocks- Light Guide Cable & Light Transmission - Good light transmission, tubeblemished- Objective Lens & Focussing - Lens in ok condition- Remote Control Switches - Functional- Distal End - Needs clean- Image Quality - Good, very few black pixels- Working Channel (s) -Clear- Leak Test - Passed- Overall Condition - OK condition Dimension: 1100 x 11.7
Lot 175
Condition Report:
Richard Wolfe 730S Flexiscope
- Insertion Tube - Dents- Bending mechanism & Angulation Control - Limited but working- Light Guide Cable & Light Transmission - //- Objective Lens & Focussing - Lens in ok condition, debris
Lot 176
Condition Report:
Page 15 Updated 10 April 2017
- Remote Control Switches - //- Distal End - Good condition- Image Quality - Poor, dirt and black pixels- Working Channel (s) -Clear- Leak Test - //- Overall Condition - Average - poor condition Dimension: 330 x 5.9
Olympus URF Type P2 Flexible Scope
- Insertion Tube - Dents- Bending mechanism & Angulation Control - Poor- Light Guide Cable & Light Transmission - Light was yellow- Objective Lens & Focussing - ok- Remote Control Switches - //- Distal End - ok condition- Image Quality - Poor- Working Channel (s) -Clear- Leak Test - Fail- Overall Condition - Poor and leaking Dimension: 650 x 4
Lot 177
Condition Report:
Olympus ENF-P4 Endscope
- Insertion Tube - Marks and dents- Bending mechanism & Angulation Control - Good- Light Guide Cable & Light Transmission - Good- Objective Lens & Focussing - OK- Remote Control Switches - //- Distal End - ok condition- Image Quality - Poor- Working Channel (s) -//- Leak Test - Fail- Overall Condition - Poor and leaking Dimension: 300 x 4
Lot 178
Condition Report:
Sonosite 180plus Ultrasound With 2MHz Probe
Unit powers on fine. The display is blemished but does not obstruct readings. Buttons,trackball, and gain controls are all functional. The airscan was of average quality, nomajor issues detected, no further tests conducted. The probe cable has a kink,membrane of the probe was in poor condition, damaged. No power adaptor included.
Lot 215
Condition Report:
Page 16 Updated 10 April 2017
Storz Image 1 Autoclavable A3 Camera
The camera head was tested with a compatible processor (lot 399) and the systemproduced clear images. Zoom, focus and white balance settings are all in good workingorder. Overall in good physical condition.
Lot 216
Condition Report:
Olympus 30° Autoclaveable Rigid Scope
In good physical condition with clear optics and good light transmission. Size: 50 x2.7mm
Lot 217
Condition Report:
Sonosite C60/5-2 MHz Transducer
This probe is in good physical and cosmetic condition.When connected to two different sonosite 180 it keeps resetting / rebooting theultrasound.Unable to perform further tests.
Lot 228Erased12 2009Manufactured:
Serial No:
Condition Report:
Sonosite P21/5-2 MHz Transducer
This probe is in good physical and cosmetic condition.When connected to two different sonosite 180 it keeps resetting / rebooting theultrasound.Unable to perform further tests.
Lot 229Erased08 2011Manufactured:
Serial No:
Condition Report:
Pentax EG-2970TK Videoscope - No Case
- Insertion Tube - Dented- Bending mechanism & Angulation Control - Limited D movement, goodlocks- Light Guide Cable & Light Transmission - //- Objective Lens & Focussing - //- Remote Control Switches - //- Distal End - average condition- Image Quality - Not tested- Working Channel (s) -Clear- Leak Test - Fail
Lot 230
Condition Report:
Page 17 Updated 10 April 2017
- Overall Condition - Poor, leaking and not tested Dimension: 1000 x 11.2
Pentax ED-3480TK Videoscope - No Case
- Insertion Tube - Blemished tube- Bending mechanism & Angulation Control - Good, good locks- Light Guide Cable & Light Transmission - //- Objective Lens & Focussing - //- Remote Control Switches - //- Distal End - average condition- Image Quality - Not tested- Working Channel (s) -Clear- Leak Test - Fail- Overall Condition - Poor, leaking and not tested Dimension: 1200 x 13
Lot 231
Condition Report:
Olympus GIF XQ-20 Fibrescope (No Case)
- Insertion Tube - Damaged distal end- Bending mechanism & Angulation Control - Limited- Light Guide Cable & Light Transmission - OK- Objective Lens & Focussing - OK- Remote Control Switches - //- Distal End - OK condition- Image Quality - Poor,blurry+black pixels- Working Channel (s) -Clear- Leak Test - Fail- Overall Condition - Poor and leaks Dimension: 900 x 11
Lot 244
Condition Report:
Olympus GIF Type Q230 Videoscope
- Insertion Tube - Good- Bending mechanism & Angulation Control - OK and good locks- Light Guide Cable & Light Transmission - Good- Objective Lens & Focussing - Good- Remote Control Switches - Functional- Distal End - OK condition- Image Quality - Good- Working Channel (s) -Clear- Leak Test - Fail- Overall Condition - Good but leaks Dimension: 1000 x 12.3
Lot 255
Condition Report:
Page 18 Updated 10 April 2017
Olympus CF Type 1 T200L Videoscope
- Insertion Tube - Blemished tube- Bending mechanism & Angulation Control - Average, slightly stiff- Light Guide Cable & Light Transmission - Good- Objective Lens & Focussing - OK- Remote Control Switches - Functional- Distal End - OK condition- Image Quality - Poor, verticle lines- Working Channel (s) -Clear- Leak Test - Fail- Overall Condition - Poor and leaks Dimension: 1600 x 15.5
Lot 256
Condition Report:
Olympus JF Type 200 Videoscope
- Insertion Tube - Good- Bending mechanism & Angulation Control - Poor, locks work ok- Light Guide Cable & Light Transmission - Good- Objective Lens & Focussing - OK- Remote Control Switches - Functional- Distal End - OK condition- Image Quality - Good- Working Channel (s) -Clear- Leak Test - Very small leak- Overall Condition - Average with slight leak Dimension: 1100 x 12.1
Lot 257
Condition Report:
Olympus URF Type PS Flexible Scope
- Insertion Tube - Distal end poor- Bending mechanism & Angulation Control - Poor- Light Guide Cable & Light Transmission - //- Objective Lens & Focussing - OK- Remote Control Switches - //- Distal End - Poor, damaged- Image Quality - Poor, black spots- Working Channel (s) -Clear- Leak Test - Fail- Overall Condition - Poor Dimension: 700 x 3
Lot 258
Condition Report:
Page 19 Updated 10 April 2017
Toshiba Convex Array Transducer PVT-674BT Probe
This probe is in good physical and cosmetic condition. Crystal membrane is also ingood condition.When tested this probe produced good image quality and average resolution andgreyscale However the probe has some crystal dropout and some signal noise at oneof the corners.
Lot 26299A152513102 2015Manufactured:
Serial No:
Condition Report:
Ultrasonic CS-2/60 Probe x 2
Both transducers are in good physical condition and also their crystal membranes.Transducer serial number ending in 2755 has a repair in the midle of the cable and it iscovered by white tape.Unable to perform further tests due to no compatible ultrasound machine at the time oftests.
Lot 2631) TRA1.2-C52.27552) TRA1.0-C52RP.1843
Serial No:
Condition Report:
Micromaxx L2SE/13-6 MHz Probe
The transducer L25e/13-6 is in poor physical condition. The crystal membrane hascracks / is broken on the corners and has dents / marks all accross.When connected with two different sonosite 180 ultrasounds it keeps reseting /rebooting the ultrasopunds.No further tests been carried out.
Lot 26403895703 2007Manufactured:
Serial No:
Condition Report:
Toshiba Convex Array Transducer PVT-375BT
The transducer physical condition is good and cable has an initial light discoloration.Crystal membrane is in good condition.When tested with a compatibe Toshiba ultrasound, the probe showed initial signs ofcrystal dropout on one side but the obtained image is clear with average greyscalelevels and average resolution.
Lot 265FDA128326308 2012Manufactured:
Serial No:
Condition Report:
Page 20 Updated 10 April 2017
Toshiba Convex Array Transducer PVT-674BT
Small dent on corner of plastic enclosure adjacent to crystal membrane. The membraneitself is in good condition and is showing signs of normal ageing.The overall physical condition is good.When tested with compatible toshiba ultrasound, the transducer presented initial signsof crystal dropout and some noise accros but image quality still good with averageresolution and average greyscale levels.
Lot 26699A129203209 2012Manufactured:
Serial No:
Condition Report:
Storz 0 Degree Rigid Scope
In good physical condition with clear optics and good light transmission. Size: 350 x3mm
Lot 267
Condition Report:
Dyonics Camera Mount Coupler
The camera head coupler is intact with lens and focus control. Tested with a Smith &Nephew 460H camera head and the system produced clear images. The coupler is in aDyonics box but the actual details are IMAGES 4533.
Lot 278
Condition Report:
Siemens CH4-1 Probe
The convex probe was tested on a compatible ultrasound system, fair scan with thecrystal intact. Probe, cable and connector in good physical condition.
Lot 286218600762012Manufactured:
Serial No:
Condition Report:
Siemens CH4-1 Probe
The convex probe was tested on a compatible ultrasound system, Air scan shows twosmall lines of drop out. Probe, cable and connector in good physical condition.
Lot 294804233342008Manufactured:
Serial No:
Condition Report:
Stack System To Include Sony LCD Monitor, Olymlpus CLV-S20 Light Source,
The CLV-S20 light source power up with the xenon lamp and brightness control in goodworking order. LCD is working fine as well. The digital image capture unit power up butnot booting up. It's hard drive had been removed. System is in fair physical condition.
Lot 328
Condition Report:
Sedia Urine Analyser (HDD Removed)
This unit fails to initialize. HDD removed.
Lot 334
Condition Report:
Page 21 Updated 10 April 2017
Unable to perform further tests.The physical condition is good.Cart, castors and the breaks are funtional.
GE Vivid i Ultrasound With 2.0MHz Doppler (No HDD)
The Ultrasound fails to boot up (no opetation system found).LCD screen seems working fine and console function keys, keyboard, sliders andknobs seems in good condition.The overall physical and cosmetic condition is good.Cart, castors and breaks are funtional.
Lot 337
Condition Report:
Pentax Stack Trolley With Radiance Monitor and Pentax Keyboard
LCD monitor is working fine and the trolley is in good physical condition.
Lot 342
Condition Report:
Huntleigh Birthing Bed - Serial no. 874133
Plastic bottom case has cracks. Unit is stacked and therefore no tests conducted.
Lot 365
Condition Report:
Huntleigh Birthing Bed - Serial no. 891257
Plastic bottom case has cracks. Unit is stacked and therefore no tests conducted.
Lot 366
Condition Report:
Huntleigh Birthing Bed - Serial no. 891260
Plastic bottom case has cracks. Unit is stacked and therefore no tests conducted.
Lot 367
Condition Report:
Huntleigh Birthing Bed - Serial no. 891255
Plastic bottom case has cracks. Unit is stacked and therefore no tests conducted.
Lot 368
Condition Report:
Huntleigh Birthing Bed (No Siderails) - Serial no. 020790
Plastic bottom case has cracks. Unit is stacked and therefore no tests conducted.
Lot 369
Condition Report:
Huntleigh Birthing Bed (No Siderails) - Serial no. 891256
Plastic bottom case has cracks. Unit is stacked and therefore no tests conducted.
Lot 370
Condition Report:
Huntleigh Birthing Bed - Serial no. 891259 (No Side Rails)
Plastic bottom case has cracks. Unit is stacked and therefore no tests conducted.
Lot 371
Condition Report:
Huntleigh Birthing Bed (No Siderails) - Serial no. 891258
Plastic bottom case has cracks. Unit is stacked and therefore no tests conducted.
Lot 372
Condition Report:
Huntleigh Birthing Bed - Serial no. 891263
Plastic bottom case has cracks. Unit is stacked and therefore no tests conducted.
Lot 373
Condition Report:
Page 22 Updated 10 April 2017
Huntleigh Birthing Bed
Plastic bottom case has cracks. Unit is stacked and therefore no tests conducted.
Lot 374
Condition Report:
Acuson CV70 Ultrasound With P4-2 Probe, Mitsubishi P9-1 Printer and Sony
Unit fails to power up. Appears as if the computer unit for ultrasound has beendisconnected. Unit possibly does not have a HDD. The probe and ultrasound are insatisfactory physical condition.
Lot 377
Condition Report:
Olympus Stack Trolley with Sony Monitor, Olympus Visera CLV-S40 Light
The light source power up with lamp and brightness control working fine. The camerahead and controller power up ok, displayed image is slightly green. White balance fail tocomplete. Image not of the best of quality. Lot is in fair physical condition.
Lot 378
Condition Report:
Clement Clarke Synoptophore x 1
Unit is in poor condition, power cable has been cut, only one bulb present. There areseveral cracks and damages on the unit.
Lot 392
Condition Report:
Clement Clarke Synoptophore x 1
Unit powers on fine. Both lamps were working at the time of testing. All mechanicalpositiong adjustments were in good working order, although slightly stiff. Optics are inneed of a clean.
Lot 393
Condition Report:
Clement Clarke Synoptophore x 1
Unit powered up fine, both bulbs working at the time of testing. 1 cap for the hyper leftand right adjustment is missing, but is still able to operate. Optics are in need of aclean.
Lot 394
Condition Report:
Olympus Evis Stack Trolley, Olympus CV-240 Light Source, Olympus CLV-U40
The system was tested with a compatible vodeoscope. Processor and light sourcepower up ok, with xenon lamp, brightness control and air pump working fine. Theprocessor produces good image quality when tested with the video scope and videocable. Good image displayed on the test monitor. System is in fair physical condition.The diathermy unit and suction is working as well. All in fair physical condition.
Lot 396
Condition Report:
Erbe ICC 200 Diathermy - D-1177, 10128-058
Unit powers up and completes its self test successfully. All indicators come on and softbuttons are functional. Unit is in tact and appears to be working but was not fully tested.
Lot 415
Condition Report:
Erbe ICC 200 Diathermy - D-1485
Unit powers up and completes its self test successfully. All indicators come on and soft
Lot 416
Condition Report:
Page 23 Updated 10 April 2017
buttons are functional. Unit is in tact and appears to be working but was not fully tested.
Laser Aperature with Footswitch in Box
Laser aperture with slit lamp adapter BMBQ. Plus footswitch all in fair condition. Mainlaser system is not included
Carry Case
Lot 4291040-11-a-3
Accessories:
Serial No:
Condition Report:
Natus Neo Blue LED Phototherapy x 3
All three LED phototherapy lamps powers up and both double and single light modesare working fine. Loot is in good physical condition.
Lot 452
Condition Report:
Philips iv22 Ultrasound (Incomplete, No HDD, Spares and Repairs)
This unit has no monitor, no HDD, plastic knobs and plastic sliders are missing.The body is in average condition. LCD touch screen is in good physical aspect.Castors and breaks are fully operational.
Lot 461
Condition Report:
Toshiba Convex Array Transducer PVT-674BT
The physical condition of probe and the crystal membrane is good but there is crystaldamage on one corner with high levels of noise and crystal dropout.
Lot 47999A14X472210 2014Manufactured:
Serial No:
Condition Report:
Siemens VFX 13.5 Probe
The linear array probe was tested on a compatible ultrasound system, fair scan with thecrystal intact. There are a few small scratches on the probe's membrane and sign ofbroken wires when the cable is bent. Fair physical condition.
Lot 482X13C1102472011Manufactured:
Serial No:
Condition Report:
Siemens VFX9-4 Probe
The linear array probe was tested on a compatible ultrasound system, fair scan with thecrystal intact. There is one small tear in the corner of the probe's membrane. Fairphysical condition.
Lot 483X94C0706032007Manufactured:
Serial No:
Condition Report:
Page 24 Updated 10 April 2017
Sonosite 180 Plus Ultrasound With 2 x Probes - 28761
Unit powers on fine. Display is blemished, but trackball, buttons, and gain controls areall functional. Some scratches on the unit but overall in satisfactory condition. ProbeC60/5 caused the system to reboot repeatedly. Probes L38/10 produced a satisfactoryairscan, membrane of probe has some scratches. Probe C11/7 is functional and alsoproduced an image of similar quality as previous probe. The membrane is in okcondition, the casing has some scratches. Adaptor is included and trolley is in goodworking condition but shows signs of wear and tear.
Lot 490
Condition Report:
GE Logiq 9 Ultrasound Machine With 4C Probe - 33763, 82604USO (No HDD)
This unit is not powering on and HDD has been removed.Function console keys (rotary knobs, sliders, track ball) are in good physical conditionbut they have not been tested. The CRT monitor has black tape holding the front plasticframe in place.The ultrasound has foot pedal, Printer and VCR but also not been tested.The probe is in good physical condition (membrane also) but cable adjacent to probe'sbody looks iffy (protective rubber detached).Overall condition is reasonable. Castors are iffy (intermittently holding / no freemovement) and breaks are ok.
Transducer 4C, serial 48214PDI, DoM: 11/2005.
Lot 52182604USD?Manufactured:
Accessories:
Serial No:
Condition Report:
Zeiss IOL Master
Powers up and completes the system initialisation successfully. Display screen, joystickcontrol, chin rest adjustment and keyboard/touchpad are in good working order. Theavailable and functional measurement modes are ALM, KER and ACD (with minorinterference on the image). The table height adjustment (motorised) is functional andoverall the unit is in good physical condition. Test eye included. YOM: 2002. Softwareversion: 3.01.0294.
3.01.0294.
Lot 5232002Manufactured:
Software:Condition Report:
Olympus Stack System to include OTV-S6 Camera, CLV-S20 Lightsource, OTV-S6
All units within the stack system power up ok. The camera controller and the camerahead are functional and produces satisfactory images (slightly dark) on the CRTmonitor. The light source is in good working order. The printer also appears to beworking – not fully tested. Overall in satisfactory physical condition.
Lot 574
Condition Report:
Page 25 Updated 10 April 2017
Dinamap Patient Monitor x 3 - 04520, 04791 and 04189
Pro 1000 monitor fails to power up. Continuously alarming when plugged into mains.
Pro 300 & 400V2 power up and sucessfully complete their initial self-test. Display andkeys are functional. NIBP pump is in good working condition. Alarms are functional.Backup battery not tested.
All the monitors are in satisfactory physical condition.
Lot 583
Condition Report:
Fresenius 4008 Dialysis Machine
Unit was not tested. Overall in poor cosmetic condition.
Lot 605
Condition Report:
Fresenius 4008 Dialysis Machine
The unit does not boot up, continuous bleep and multicolored stripes on display. Nofurther tests conducted.
Lot 606
Condition Report:
GE M-CAiOV Module x 4, GE M-CAio x 1, Philips Temp M1029AT Module x 1
2 Gas modules include a water trap. Cosmetically all 5 modules are in satisfactorycondition. They were tested on a compatible monitor, all passed the initial calibrationand were able to sense airflow. The temperature module is in tact but was not tested.
Lot 622
Condition Report:
Siemens Acuson Sequoia C512 Ultrasound With 3V2c x 2 Probes and TE-VSMs x
This unit has no HDD.LCD monitor is working fine. Diskette and cd drives at front panel and also the computerwith two video boards (4 DVI ports), diskette and DVD drive seems ok but unable totest.
Transducer 3V2c, serial 34805702, DoM: 2003
Transducer 3V2c, serial 62650601, DoM: 2006
Transducer Aux CW Tee, serial 64301013, DoM: 2006
Transducer TE-V5Ms, serial 84102753, DoM: 2008
Transducer TE-V5Ms, serial 82302474, DoM: 2008
Lot 633666282006Manufactured:
Accessories:
Serial No:
Condition Report:
Page 26 Updated 10 April 2017
The console panel function keys, knobs, sliders and track ball are in good physicalcondition.All transducers are in good physical overall condition (including the crystal membrane).The TE-V5Ms's transducers movement and locking mechanism are working ok.Castors and breaks are working fine and the overall physical condition of the ultrasoundis good.
GE Logiq 3 Pro Ultrasound with E8C, 3.5C Probes AN: 015647 SN: 35477WX9,
This unit does not power up.CRT monitor, front console function keys, keyboard, rotary knobs and sliders are ingood physical condition. Also printer and DVD drive seems to be working ok but notfully tested.Both transducers are in good physical condition (this also includes the crystalmembrane).Castors and breaks are functioning well.The overall physical condition is reasonable.
Transducer 3.5c, serial 47569WP9, DoM 01/2006Transducer E8C, serial 35477WX9, DoM: 12/2005
Lot 634
Accessories:
Condition Report:
Gambro Primaflex Dialysis Machine with Barkey Auto Control - 1250668, PA1632
The dialysis machine fails to power up. It is in fairly good physical condition. The barkeyauto control powers up OK. The display and keypad is functional. It is in satisfactoryphysical condition.
Lot 641PA1632Serial No:
Condition Report:
17049Operating Hours:
Gambro Primaflex Dialysis Machine with Barkey Auto Control - 1250666, PA1649
Unit powers up and displays "Error detected (CD Unit Touchscreen)". After a fewseconds the unit sucessfully loads the application software. Touchscreen was notworking. It is in fairly good physical condition. Barkey Autocontrol powers up OK andsucessfully completes its self-test. Display and keys are functional. Overall it is insatisfacory physical condition.
6.10
Lot 642PA1649
Software:
Serial No:
Condition Report:
2726Operating Hours:
Gambro Primaflex Dialysis Machine with Barkey Auto Control - 1250271, PA1627
6.10
Lot 643PA1627
Software:
Serial No:
17865Operating Hours:
Page 27 Updated 10 April 2017
Dialysis machine powers up and successfully completes its initial self-test. Touchscreen display and all the blood pumps appeared to be in good working order. No faultsappeared when testing this unit. Overall it is in fairly good condition.The Auto Control Barkey 3XPT is also working fine. Display and keys are functional. Itis in satisfactory physical condition.
Condition Report:
Stack System With Sony Monitor, Smith & Nephew Dyonics Power Unit, Camera
The LCD display, power control unit, light source, and camera unit all power on fine.The display is in satisfactory condition. All indications on the power control unit light up,intact and appears to be working but was not fully tested. The light source is also insatisfactory condition, xenon lamp was working at the time of testing.The camera head has a few scratches on the handle but is otherwise in satisfactorycondition and was able to produce good images. Printer is in good working order.Recorder and PC appear to be in tact but were not tested.
Lot 657
Condition Report:
Olympus WM-60 Workstation with CV-240 Processor and Connector, CLU-U40
The system was tested with a compatible vodeoscope. Processor and light sourcepower up ok, with xenon lamp, brightness control and air pump working fine. Theprocessor produces good image quality when tested with the video scope and videocable. Good image displayed on the test monitor. System is in fair physical condition.The pigtail connection is slightly stiff to fit on the videscope socket.
Lot 659
Condition Report:
Olympus PSD-20 Diathermy and Footswitch
Unit power up and completed the self-test. The numerical displays and power outputselectors are working fine. The patient return plate monitoring circuit is functional. Thefootswitches are working. RF output was not fully tested. Unit is in fair physicalcondition. Accessories not included.
Lot 689
Condition Report:
HGM Elite Laser with Foot Switch
Unit power up with the numerical displays and selectors soft keys working. Laseraperture is not included therefore output was not fully tested. System is in fair physicalcondition but not fully tested. Laser On/Off key not included.
Footswitch
Lot 690001538411 2000Manufactured:
Accessories:
Serial No:
Condition Report:
UW Ultrasonic Bath
The ultrasonic bath is working ok.It has analog timer, lid and instruments basket.This unit has no drainage system.The overall physical condition is good and cosmetic is average.
Lot 702
Condition Report:
Page 28 Updated 10 April 2017
Clement Clarke Synoptophore
Unit powers up fine and both bulbs were working fine at the time of testing. Switchesand position adjustments are functional. Chin rest height adjustment is in good workingorder. Optics are in need of a clean.
Lot 703
Condition Report:
Triad 2000 Light Cure Unit - 0005691336, Ultrasonic Bath and Heat Plate, Basic
The ultrasonic bath is working ok. This is a basic unit (no timer, no drainage system, noinstruments basket) with lid and on/off switch, only.The cure light is also in good working condition. Plate rotation and high adjustments,door open sensor, analog timer and light are all working fine.The other units has not been tested (basic ventilator and heat plate).The overall physical condition is good and cosmetic condition is average.
Lot 708
Condition Report:
Stryker SDC HD High Definition Digital Capture Monitor (No HDD)
Powers up but the hard drive has been removed and therefore the unit is not functional.In satisfactory physical condition.
Lot 744
Condition Report:
Stack Trolley with Olympus OTV -SX2 - 1101885, Processor and Camera Head,
Both the camera controller and the light source power up ok but the image is not clearfrom the camera head (interference and blur). The lamp needs replacing in the lightsource. Overall in satisfactory physical condition. Monitor not included.
Lot 747
Condition Report:
Stealth Treon Surgical Navigation System
Unit powers up and completes its initial self-test. The touch screen display, keyboardand mouse are fully functional. All the modes except one were accessible andappeared to be working fine. Unit is in satisfactory physical condition.
Lot 750
Condition Report:
Draeger Titus Anaesthetic Machine with PM8050 MRI Ventilator
The anaesthetic machine is intact with breathing system, bellows and medical gashoses in place. The ventilator powers up and the display screen and control panel arein good working order. No further checks carried out. Overall the unit is in good physicalcondition. The AGSS receiver is missing.
Lot 753
Condition Report:
Zeiss Visulas Yag II Plus (Case Damaged)Lot 755
Page 29 Updated 10 April 2017
This unit does not power on. The control console is broken.The ophthalmic laser unit seems in good physical and cosmetic condition.Optics, knobs and other adjustments seems working OK however it has not been fullytested.Table electric hight adjustments, castors and breaks are in good working order.The overall condition is average.
Control console: 774673Ophthalmic laser: 7765851998Manufactured:
Serial No:
Condition Report:
Oculus Pentacam Slit Lamp
The pentacam slit lamp is in good physical condition but it was not been tested. Thisunit has no computer. LCD monitor seems working ok but has not been fully tested.Table electric high adjustments is working ok. Castors and breaks are working ok.The overall condition is average.
Lot 761
Condition Report:
Draeger Primus Anaesthetic Machine
The anaesthetic machine powers up successfully. Display screen and control panel arein good working order. The breathing system is intact and the medical gas hoses areincluded. The battery is also working (at the time of test). Overall the unit is in goodphysical condition and appears to functional (not fully tested). Ventilation modes: ManSpont, Vol, Press and Press Supp. Software version: 4.30. Operating hours: Mixer25985, Vent 3544, Power supply 49129, Battery 279. SN: ARUH-0099.
4.30
Lot 766ARUH-0099
Software:
Serial No:
Condition Report:
Philips iU22 Ultrasound With L17-5 and L12-5 Probes
After been data wiped this ultrasound developed a fault. During boot up it displays: "Reboot and select proper boot device or insert boot media in selected boot device andpress a key".Therefore unable to perform complete tests.LCD monitor is working fine.Console panel has a crack on both lower corners.Function console keys, track ball, sliders, rotary knobs and keyboard are all in goodcondition and seems working fine. Also DVD drive and printer seems in working order
Probe L17-5, serial 032326Probe L12-5, serial 032B9Z
Lot 783
Accessories:
Condition Report:
Page 30 Updated 10 April 2017
but not fully tested.The transducers are both in good physical and cosmetic condition.Castors, breaks and console panel hight adjustment are all in working order.Overall condition of the ultrasound is average.
Philips HD11XE Ultrasound With 2 x Probes C5-2/L12-3, Sony Printer Up-D895MD
After been data wiped this ultrasound developed a fault. During boot up it displays: "Novideo input".Therefore unable to perform complete tests.LCD monitor is working fine.Left hand side probe holder is broken but still in place.Function console keys, track ball, sliders, rotary knobs and keyboard are all in goodcondition and seems working fine.The transducers are both in good physical and cosmetic condition but transducer C5-2crystal membrane is showing signs of age..Castors, breaks and console panel hight adjustment are all in working order.Overall condition of the ultrasound is average.
Probe L12-3, serial B09Z5Kprobe C5-2, serial 03FMRJ
Lot 784
Accessories:
Condition Report:
Stryker Patient Trolley - AN 109889
Unit is stacked. Not able to test.
Lot 797
Condition Report:
Merivaara Emergo Patient Trolley - AN 119962
Unit is stacked. Not able to test.
Lot 798
Condition Report:
Stryker Vision Elect HDTV Surgical Viewing Monitor
Display is in good cosmetic condition ( still protected ). No power adaptor availabletherefore not tested.
Lot 808
Condition Report:
Maquet Operating Table
All motorised positioning adjustments and the base locking system are in good workingorder. The operating table does not come with a hand control but a compatible handcontrol was used for testing. The emergency control panel is functional and the batteryis also working (at the time of test). Overall the operating table is in satisfactory physicalcondition with scuff marks to the base of the unit. The back and head rest cushions aremissing. Includes two spare back rest boards. YOM: 2003.
Lot 8612003Manufactured:
Condition Report:
Page 31 Updated 10 April 2017
Maquet Operating Table
All motorised positioning adjustments and the base locking system are in good workingorder. The hand control is also included and it's operational. The emergency controlpanel is functional and the battery is also working (at the time of test). Overall theoperating table is in satisfactory physical condition with scuff marks to the base of theunit. The corners of the footboard has been damaged. Only one cushion is included(centre section). YOM: 2002.
Lot 8702002Manufactured:
Condition Report:
Olympus Stack Trolley with Olympus CV240 Camera Processor, CLV U40 Light
The system was tested with a compatible videoscope. Processor and light sourcepower up ok, with xenon lamp, brightness control and air pump working fine. The lightoutput mode is stuck in the Evis 200 mode. The processor produces good image qualitywhen tested with the video scope and video cable. Good image displayed on the testmonitor. System is in fair physical condition.
Lot 878
Condition Report:
Fresenius 4008H Dialysis MC x 2 - IXCA7012, IXCA7011 (Cracked Screens)
Both units power up and complete the system self tests successfully.Unit serial number 1XCA7012 has 21114 working hours.Unit serial number 1XCA7011 has 23381 working hours and on screen "F11configuration" message.The software is V4.3 on both units.Both LCD screen glass are cracked but LCD screen is working. LED bar graph andfunction keys are in good working order and system menus are accessible.Both units contains the following modules:
* Two Blood Pump* Air Bubble Detector* Heparin* Diasafe filter* Bibag
System electronics appears to be working ok but unit was not fully tested.Diasafe filter needs replacing.The rollers and doors of the blood pump are intact.Online Clearance Monitoring (oCM) software installed.Heparin was not powering on at the time of tests.Missing whater and drain hoses.Overall, apart from the screens broken glass, both units are in reasonable physicalcondition.
V 4.3
Lot 879
Software:Condition Report:
Page 32 Updated 10 April 2017
Fresenius 4008S Dialysis MC - 2VCAE929 (Incomplete)
The unit powers up and completes the system self tests successfully.LCD screen, LED bar graph and function keys are in good working order and systemmenus are accessible.The unit contains the following modules:
* Heparin* Diasafe filter* Bibag
System electronics appears to be working ok but unit was not fully tested.Diasafe filter needs replacing.The blood pump and air buble detector are missing.The wired hand control apears to be working ok.Online Clearance Monitoring (oCM) software installed.Heparin was not powering on at the time of tests.Missing mains plug, whater and drain hoses.Overall the unit is in poor to reasonable physical condition.
Lot 880
Condition Report:
Diamed Dia Cent - 12 Centrifuge
The centrifuge does not power up. In satisfactory physical condition.
Lot 892
Condition Report:
Maquet Operating Table (One Cushion Only)
All motorised positioning adjustments and the base locking system are in good workingorder. The operating table does not come with a hand control but a compatible handcontrol was used for testing. The emergency control panel is functional and the batteryis also working (at the time of test). Overall the operating table is in satisfactory physicalcondition with scuff marks to the base of the unit. One corner of the footboard has beendamaged. Only one cushion is included (centre section). YOM: 2003.
Lot 9162003Manufactured:
Condition Report:
Olympus PSD-20 Diathermy With Footswitch
Unit power up and completed the self-test. The numerical displays and power outputselectors are working fine. The patient return plate monitoring circuit is functional. The
Monopolar footswitch
Lot 10297701370
Accessories:
Serial No:
Condition Report:
Page 33 Updated 10 April 2017
footswitches are working. RF output was not fully tested. Unit is in fair physicalcondition. Accessories not included.
Nidek Green Laser Photocoagulator System
This is only a diode pump solid-state laser generator unit (model GYC-2000).This unit is powering on ok but unable to fully test as this is not a complete system.Overall physical and cosmetic condition is good.
Lot 10392106410 2001Manufactured:
Serial No:
Condition Report:
Ultrawave Ultrasonic Bath x 2 (One Lid Missing)
Both ultrasonic bath's are working ok (timer and ultrasound).One has an analog timer, lid, instruments basket but is missing two rubber foot at back.The other has a digital timer, instruments basket but is missing the lid.Both unit has no drainage system.The overal physical condition is good and cosmetic condition is average to good.
Lot 1044
Condition Report:
Maquet Operating Table (Missing End Section, one cushion only)
All motorised positioning adjustments and the base locking system are in good workingorder. The hand control is included and it's operational. The emergency control panel isfunctional and the battery is also working (at the time of test). Overall the operatingtable is in good physical condition with minor scuff marks to the base of the unit. Onlyone cushion is included (head rest section). The back rest board is not included and thelocking adaptors for the back rest board is missing. YOM: 2012
Lot 10692012Manufactured:
Condition Report:
Xario Ultrasound Scanner With No Probes - ID 0553
The unit powers up and the system initiation completed successfully.Display, trackball, keyboard, touch screen, control console function keys, rotary knobsand sliders are in good working order.Probe identification and selection is working fine.With spare probes this unit was able to display satisfactory image quality along with
Software DVD's:USDF-770A/XA, USDI-770/XA, USQI-770A/XA and USDI-772B/XA with Sony preinterdrivers and operating manuals.
Lot 109299E072651702 2007Manufactured:
Accessories:
Serial No:
Condition Report:
Page 34 Updated 10 April 2017
average grey scale and resolution. Accurate caliper measurements obtained.All of the available modes (M, PW, Power, CDI, Dynamic Flow and 2D) were tested andseems working fine.The system is in satisfactory overall condition.Cart, castors and brakes are in good condition.
Page 35 Updated 10 April 2017