Conclusions Our quality assurance system and databases allowed us to quickly evaluate our test results in response to reports of increased numbers of false positive tests in San Francisco, New York and Los Angeles. False positive OraQuick results in New Jersey were not an issue. Specificity remained within the claims of the manufacturer. The CDC presented NJHIV’s data at national meetings (Figure 6) and included our findings in their decision not to change current rapid test recommendations. Results Data comparison among databases showed that data was complete in the NJHIV database for 22 of 23 sites tested. After remediation of recordkeeping practices, NJHIV data at all sites is complete through June 2006. One cluster of 3 simultaneous discordants was discovered at one site. Further analysis did not yield an explanation for these discordants. Consistent with the manufacturer’s product performance claims, there was a slightly higher false positive rate when using oral fluid than when using fingerstick whole blood specimens (Figure 5). occasionaly drops below manufacturer’s claims for individual months. Figure 5: Specificity by month Analysis of discordant results did not show any other pattern when examined by month, site, reagent lot or sample type. 0.991 0.992 0.993 0.994 0.995 0.996 0.997 0.998 0.999 1 O raQ uick S pecificity June July Aug Sept O ct Nov Dec Jan Feb M ar Apr M ay June M onth (2005-2006) Blood Oral Figure 1 OMT: A New Form of Rapid Testing Introduced NJHIV participated in a CDC sponsored evaluation of OMT testing through May 2005. OMT testing was introduced at New Jersey testing sites in June 2005. OraQuick’s specificity claim (lower bound of confidence interval) for fingerstick testing is 99.8% and for OMT testing is 99.6%. There was a gradual transition from blood to OMT testing (Figure 2). Figure 2: Rapid HIV sample types NJHIV's Rapid Testing Program Keeps False NJHIV's Rapid Testing Program Keeps False Positives in Check Positives in Check Evan M. Cadoff, MD 1 ; Eugene G. Martin, PhD 1 ; Sindy M. Paul, MD, MPH 2 ; Sharon A. Holswade, MBA 1 ; Jake Nelson1; Vivian H. Shih, BA 1 ; Dolores Van Pelt, MSN 1 ; Nisha J. Intwala, BS (MT) 1 ; Franchesca N. Jackson, BS 1 ; Gratian Salaru, MD 1 1 UMDNJ-RWJMS, New Brunswick, NJ and 2 New Jersey Department of Health and Senior Services, Trenton, NJ. Figure 4: Test records entered into database which tracks frequency of testing by month, primary site, satellite, reagent lot numbers, sample type and test result. Figure 4a: Records selected for review based on a single parameter. All oral fluid tests were selected in the diagram to the left and are shown here. Figure 4b: Records selected for review based on a second parameter. Here, all discordant results were selected from among the oral fluid tests in the diagram to the left. Abstract Background:FDA approval of rapid HIV testing kits is partly based on sensitivity and specificity claims. Recently, in San Francisco, New York and Los Angeles, health care workers confronted a spike of false positive results (positive rapid HIV test with negative confirmatory testing) on oral fluid (OMT). By reviewing NJHIV quality assurance (QA) monitoring data, we were able to assess the prevalence and significance of false positive results in NJ. Methods:The NJHIV program began rapid testing under a medical school QA program in late 2003. The program has expanded to include 117 sites by 2005. A critical factor in the organization of testing was the use of a centralized core staff of clinical laboratorians who were responsible for development of common, uniform policies, procedures, training and recertification, inventory control and validation, proficiency testing, and a core communication hub (www.njhiv.org) all part of a rigorous quality assurance initiative. Data from all sites is analyzed monthly. Results:We reviewed NJHIV data since March 2005, including over 24,000 tests, of which more than 5000 were OMT. Specificity of fingerstick testing exceeded 99.9% . A slight increase in the false positive rate (99.5% specificity) was noted when sites moved from blood to OMT testing, but specificity remained within manufacturer’s claims. Reviewing data from each of our sites, each lot of reagent, each month of testing and each specimen type, we did not record an out of label FP spike as was seen in some other states. Conclusion:Data routinely collected in our QA program allowed rapid assessment of HIV testing performance during the introduction of oral mucosal testing. No increase in false positives occurred beyond what would be expected with the transition to oral fluid testing. This is largely attributable to the organization of the program. Background The NJHIV Rapid HIV Testing Program • Built upon existing UMDNJ-Robert Wood Johnson Medical School, multi-facility, point-of-care-testing program. • Centralized quality assurance process using pathologists, technologists and informaticians. • OraQuick fingerstick whole blood testing began November 2003 at the first NJ licensed site – a primary site in close proximity to the medical school. Currently there are 140 sites including fixed and mobile venues. (Figure 1) • Testing activities are reviewed monthly by a medical technologist. • Preliminary positive rapid tests are confirmed by Western Blot at the NJDHSS – PHEL laboratory in Trenton. Quality Assurance Plan • Management by a board certified Pathologist • Supervisory control through site coordinators • Central lab oversees: Regulatory issues and proficiency testing Acquisition and validation of supplies Inventory control • Common procedures and core policies • Uniform administration at all locations • Common training, certification of personnel, forms Core communication hub www.njhiv.org • Quality Control Rules • Standardized monthly site visits – ‘The Report Card’ • Intra and inter site comparisons to insure that Methods • NJHIV testing sites maintain logs of each test performed (Figure 3). Figure 3: NJHIV testing log • Data from the logs is entered into an Excel spreadsheet (Figure 4). • Data from the logs is compared to the NJ Counseling and Testing database and the database of confirmatory tests run at the NJDHSS - Public Health and Environmental Laboratories. • The NJHIV database was queried for discordant results and reviewed by month, by testing site, by OraQuick reagent lot number, and by sample matrix--ie, blood or oral fluid. (Figures 4a and 4b). S.F. Clinics Getting High False- Positive Rate on Oral HIV Test Sabin Russell Friday, December 9, 2005 “A promising new oral HIV test being considered for home use has produced at least 47 false positives at San Francisco public health clinics, throwing a scare into those who received the results and raising questions about the test's suitability for widespread use in the United States and abroad. The OraQuick ADVANCE HIV test approved for professional use by the Food and Drug Administration in March 2004 detects, in just 20 minutes, antibodies to the virus that causes AIDS, using fluid swabbed from the mouth. Its speed and ease of use make the test a particularly suitable candidate for HIV prevention and treatment efforts that stress frequent testing and speedy access to care for those who test positive. San Francisco Department of Public Health officials were alarmed last week when their analyses of more than 6,000 oral tests since Spring turned up 47 instances in which the results were positive, but follow-up tests showed those patients were not infected.” PRESS RELEASE AVALANCHE •“SF Clinics Getting High False Positive Rates on Oral HIV Test” San Francisco Chronicle 12/10/05 •“False Positives from HIV Test” New York Times 12/11/05 •“Facility Halts Use of Oral HIV Test” LA Times 12/16/05 •“More Sites Drop Oral HIV Test” LA Times 12/20/05 November 2003 1 st site Rollout of primary sites April 2004 28 primary sites QA plan adapted to fixed satellite sites June 2005 102 fixed satellites Temperature, lighting, transportation issues resolved, to support large mobile vans and converted minivans November 2005 10 mobile vans March 2006 140 licensed sites Sites include: CTS sites, Emergency Departments, HIV prevention programs, hospital inpatients,hospital outpatient clinics, STD clinics TB clinics, Faith-based initiatives, mobile vans CDC's PMS-2 study begins July 2004 December 2005 PMS-2 ends October 2004 CDC's OMT study begins May 2005 OMT ends Expansion of rapid HIV testing sites, and participation in CDC-sponsored studies 0 500 1000 1500 2000 2500 N um beroftests June A ug O ct D ec Feb Apr June M onth (2005-2006) R apid HIV tests Blood Oral

Conclusions Our quality assurance system and databases allowed us to quickly evaluate our test results in response to reports of increased numbers of false

Embed Size (px)

Citation preview

Page 1: Conclusions Our quality assurance system and databases allowed us to quickly evaluate our test results in response to reports of increased numbers of false

Conclusions• Our quality assurance system and databases allowed us to quickly evaluate

our test results in response to reports of increased numbers of false positive tests in San Francisco, New York and Los Angeles.

• False positive OraQuick results in New Jersey were not an issue. Specificity remained within the claims of the manufacturer.

• The CDC presented NJHIV’s data at national meetings (Figure 6) and included our findings in their decision not to change current rapid test recommendations.

Figure 6: Slide from CDC presentation at CROI meeting, February 2006

Results• Data comparison among databases showed that data was complete in the

NJHIV database for 22 of 23 sites tested.• After remediation of recordkeeping practices, NJHIV data at all sites is

complete through June 2006.• One cluster of 3 simultaneous discordants was discovered at one site.

Further analysis did not yield an explanation for these discordants.• Consistent with the manufacturer’s product performance claims, there was

a slightly higher false positive rate when using oral fluid than when using fingerstick whole blood specimens (Figure 5). Specificity occasionaly drops below manufacturer’s claims for individual months.

Figure 5:Specificity by month

• Analysis of discordant results did not show any other pattern when examined by month, site, reagent lot or sample type.

0.991

0.992

0.993

0.994

0.995

0.996

0.997

0.998

0.999

1

Ora

Qu

ick

Sp

ecif

icit

y

June July Aug Sept Oct Nov Dec Jan Feb Mar Apr May June

Month (2005-2006)

Blood

Oral

Figure 1

OMT: A New Form of Rapid Testing Introduced• NJHIV participated in a CDC sponsored evaluation of OMT testing through

May 2005.• OMT testing was introduced at New Jersey testing sites in June 2005.• OraQuick’s specificity claim (lower bound of confidence interval) for

fingerstick testing is 99.8% and for OMT testing is 99.6%.• There was a gradual transition from blood to OMT testing (Figure 2).

Figure 2: Rapid HIV sample types QC Usage - Year OneCtrl 1 First Day of the WeekCtrl 2 Temperature Out of RangeCtrl 3 Annonymous TestingCtrl 4 New ShipmentCtrl 5 Invalid ResultCtrl 6 QC FailureCtrl 7 Coordinator AssignedCtrl 8 - Other

NJHIV's Rapid Testing Program Keeps False NJHIV's Rapid Testing Program Keeps False Positives in CheckPositives in Check

Evan M. Cadoff, MD1; Eugene G. Martin, PhD1; Sindy M. Paul, MD, MPH2; Sharon A. Holswade, MBA1; Jake Nelson1; Vivian H. Shih, BA1; Dolores Van Pelt, MSN1; Nisha J. Intwala, BS (MT)1; Franchesca N. Jackson, BS1; Gratian Salaru, MD1

1 UMDNJ-RWJMS, New Brunswick, NJ and 2 New Jersey Department of Health and Senior Services, Trenton, NJ.

Figure 4:Test records entered into database which tracks frequency of testing by month, primary site, satellite, reagent lot numbers, sample type and test result.

Figure 4a:Records selected for review based on a single parameter. All oral fluid tests were selected in the diagram to the left and are shown here.

Figure 4b: Records selected for review based on a second parameter. Here, all discordant results were selected from among the oral fluid tests in the diagram to the left.

Abstract

Background:FDA approval of rapid HIV testing kits is partly based on sensitivity and specificity claims. Recently, in San Francisco, New York and Los Angeles, health care workers confronted a spike of false positive results (positive rapid HIV test with negative confirmatory testing) on oral fluid (OMT). By reviewing NJHIV quality assurance (QA) monitoring data, we were able to assess the prevalence and significance of false positive results in NJ.Methods:The NJHIV program began rapid testing under a medical school QA program in late 2003. The program has expanded to include 117 sites by 2005. A critical factor in the organization of testing was the use of a centralized core staff of clinical laboratorians who were responsible for development of common, uniform policies, procedures, training and recertification, inventory control and validation, proficiency testing, and a core communication hub (www.njhiv.org) all part of a rigorous quality assurance initiative. Data from all sites is analyzed monthly.Results:We reviewed NJHIV data since March 2005, including over 24,000 tests, of which more than 5000 were OMT. Specificity of fingerstick testing exceeded 99.9% . A slight increase in the false positive rate (99.5% specificity) was noted when sites moved from blood to OMT testing, but specificity remained within manufacturer’s claims. Reviewingdata from each of our sites, each lot of reagent, each month of testing and each specimen type, we did not record an out of label FP spike as was seen in some other states.Conclusion:Data routinely collected in our QA program allowed rapid assessment of HIV testing performance during the introduction of oral mucosal testing. No increase in false positives occurred beyond what would be expected with the transition to oral fluid testing. This is largely attributable to the organization of the program.

Background

The NJHIV Rapid HIV Testing Program• Built upon existing UMDNJ-Robert Wood Johnson Medical School, multi-facility, point-of-

care-testing program.• Centralized quality assurance process using pathologists, technologists and

informaticians.• OraQuick fingerstick whole blood testing began November 2003 at the first NJ licensed

site – a primary site in close proximity to the medical school. Currently there are 140 sites including fixed and mobile venues. (Figure 1)

• Testing activities are reviewed monthly by a medical technologist.• Preliminary positive rapid tests are confirmed by Western Blot at the NJDHSS – PHEL

laboratory in Trenton.

Quality Assurance Plan• Management by a board certified Pathologist• Supervisory control through site coordinators• Central lab oversees:

Regulatory issues and proficiency testingAcquisition and validation of supplies Inventory control

• Common procedures and core policies• Uniform administration at all locations • Common training, certification of personnel, forms • Core communication hub www.njhiv.org• Quality Control Rules • Standardized monthly site visits – ‘The Report Card’• Intra and inter site comparisons to insure that requirements for quality and process control

are maintained

Blood OralJune 1575 202July 1096 660Aug 1054 982Sept 752 1164Oct 845 1371Nov 704 1316Dec 727 1207Jan 894 1584Feb 807 1676Mar 1052 2323Apr 859 2067May 918 1982June 996 2286

0

500

1000

1500

2000

2500

Nu

mb

er

of te

sts

June Aug Oct Dec Feb Apr June

Month (2005-2006)

Rapid HIV tests

Blood

Oral

Methods• NJHIV testing sites maintain logs of each test performed (Figure 3).

Figure 3:NJHIV testing log

• Data from the logs is entered into an Excel spreadsheet (Figure 4).• Data from the logs is compared to the NJ Counseling and Testing database

and the database of confirmatory tests run at the NJDHSS - Public Health and Environmental Laboratories.

• The NJHIV database was queried for discordant results and reviewed by month, by testing site, by OraQuick reagent lot number, and by sample matrix--ie, blood or oral fluid. (Figures 4a and 4b).

S.F. Clinics Getting High False-Positive Rate on Oral HIV Test

Sabin RussellFriday, December 9, 2005

“A promising new oral HIV test being considered for home use has produced at least 47 false positives at San Francisco public health clinics, throwing a scare into those who received the results and raising questions about the test's suitability for widespread use in the United States and abroad.

The OraQuick ADVANCE HIV test approved for professional use by the Food and Drug Administration in March 2004 detects, in just 20 minutes, antibodies to the virus that causes AIDS, using fluid swabbed from the mouth. Its speed and ease of use make the test a particularly suitable candidate for HIV prevention and treatment efforts that stress frequent testing and speedy access to care for those who test positive.

San Francisco Department of Public Health officials were alarmed last week when their analyses of more than 6,000 oral tests since Spring turned up 47 instances in which the results were positive, but follow-up tests showed those patients were not infected.”

PRESS RELEASE AVALANCHE• “SF Clinics Getting High False Positive Rates on

Oral HIV Test” San Francisco Chronicle 12/10/05• “False Positives from HIV Test”

New York Times 12/11/05• “Facility Halts Use of Oral HIV Test”

LA Times 12/16/05• “More Sites Drop Oral HIV Test”

LA Times 12/20/05

November 2003 1st site

Rolloutof primary

sitesApril 2004 28 primary sites

QA plan adapted to

fixed satellite sites

June 2005 102 fixed satellites

Temperature, lighting, transportationissues resolved, to support

large mobile vans and converted minivansNovember 2005 10 mobile vans

March 2006140 licensed

sites

Sites include:CTS sites, Emergency Departments, HIV prevention programs,

hospital inpatients,hospital outpatient clinics, STD clinicsTB clinics, Faith-based initiatives, mobile vans

CDC's PMS-2 study beginsJuly 2004

December 2005 PMS-2 ends

October 2004 CDC's OMT study begins

May 2005 OMT ends

Expansion of rapid HIV testing sites, and participation in CDC-sponsored studies

0

500

1000

1500

2000

2500

Nu

mb

er

of te

sts

June Aug Oct Dec Feb Apr June

Month (2005-2006)

Rapid HIV tests

Blood

Oral