34
Computer-Integrated Femoral Head Fracture Reduction System M.Mitsuishi 1 , N.Sugita 1 , S.Warisawa 1 , T.Ishizuka 1 , T.Nakazawa 2 , N.Sugano 3 , K.Yonenobu 4 and I.Sakuma 5 1 The University of Tokyo, 2 THK Co. Ltd. 3 Osaka University, 4 Osaka Minami National Hospital, 5 The University of Tokyo

Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Computer-Integrated Femoral Head Fracture Reduction System

M.Mitsuishi1, N.Sugita1, S.Warisawa1, T.Ishizuka1, T.Nakazawa2, N.Sugano3, K.Yonenobu4 and I.Sakuma5

1 The University of Tokyo, 2 THK Co. Ltd.3 Osaka University, 4 Osaka Minami National Hospital,

5 The University of Tokyo

Page 2: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Background

Page 3: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Conventional method and its problems

The force that must be exerted by the surgeons is large while pulling the inferior branch.

Surgeons, medical staffs and patients are exposed to radiation during the operation.

Fluoroscopy

Reduction device

Reduction accuracy depends on the ability of the surgeon.

(1) Separation of a piece of bone by pulling the leg (inferior branch)

(2) Reduction motion by pulling and rotation of an inferior branch

(3) Intra-operative confirmation by X-ray

Page 4: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Fracture reduction robot system construction

Page 5: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Schematic diagram of the systemThe optimum reduction path is calculated preThe optimum reduction path is calculated pre--operatively operatively using the CT and Xusing the CT and X--ray images at the navigation system.ray images at the navigation system.The calculated path is sent to the reduction robot.The calculated path is sent to the reduction robot.The reduction motion is performed by the robot.The reduction motion is performed by the robot.

Page 6: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Solutions(1) Preoperative full 3D reconstruction(2) Preoperative path planning(3) 6 d.o.f. reduction robot(4) Multi operation modes(5) Fail-safe mechanism

Automatic mode Power assisted mode JOG mode

Page 7: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Layout of the system3D fluoroscopy imaging system

Carbon operation table Reduction robot

Patient Medical doctor

Page 8: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

System overviewFluoroscopy

imaging system

Carbon operation table

Reduction robot

Optical 3D tracking system

Navigation system

Patient

Page 9: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Fracture reduction system

Page 10: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Fracture reduction robot

boot Pitch

Roll

Telescopic

caster

ForceSenser

Handle

Air Lift(built-in)

‘Holizontal is perpendicular to the paper)

Swing-Arm for connection to the surgical table

l=2267

h=10

17

880

(w=663)

Axis Stroke

Horizontal 202.43mm Telescopic 182.44mm

Vertical 303.12mm Pitch 679.84mrad Roll 4882.67mrad Yaw 672.80mrad

Weight 250kg

Vertical

•The axes of all the rotational d.o.f. intersect at the same point using a hollow motor.•Attaching and detaching operations should be easy while the robot is connected with the existing operation table.

•Fail-safe mechanism is necessary.•The same position can be maintained by a lift after the determination of a fracture reduction robot.•The motion of the robot must be synchronized with that of an operation table while moving up and down.

Yaw

Page 11: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

SpecificationSpecificationStrokeX:±150 mm,Y:-10~+300 mm,Z:-10~+300 mmA:±20 deg.,B:±25 deg.,C:±135 deg.

Force sensorFx, Fy:400 N,Fz:800 NMx, My, Mz:40 Nm

Fail-safe mechanismY:300 N,+100 mm, -30 mmB:28 Nm±135 deg.

A-axis

Z-axis

Y-axis

X-axis

C-axis

B-axis

Page 12: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Overview of the fracture reduction robot

Teaching pendant

Force sensor

FixtureEmergency stop bottom

Foot switch for the power assisted mode Connecting arm

Page 13: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Robot system controllerRobot system controllerReal Time Linux

Robot Control Task

Socket communi-cation

Teaching pendantcontrol task

Page 14: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

FailFail--safe mechanismsafe mechanism1. High rigidity should be maintained in the normal state.2. Large displacement occurs when an excessive force is

applied to the system.3. Emergency signal is generated and the system is stopped.4. The system returns to its normal state when the excessive

force is removed.

Concept

Configuration

Page 15: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Operation modes

JOG modePower assisted mode

Automatic (NAVI) mode

Page 16: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

JOG modeJOG mode

Interactive motion control using a teaching Interactive motion control using a teaching pendantpendantStep motionStep motionUseful for preparation of the surgeryUseful for preparation of the surgery

Page 17: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Power assisted modeTo reduce the load on a surgeon when pulling and rotating the patient’s inferior branch– Approaching the fixture device to a foot

To provide the same adjustment environment as the conventional one by holding a fixture device– Separating fracture segments

To move the inferior branch to an arbitrary position and posture

Page 18: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Power assisted modePower assisted mode3 different viscosities were prepared.

1. 1. UltraUltra--soft modesoft mode–– Appropriate for Appropriate for

the wide range the wide range motion (coarse motion (coarse positioning)positioning)

2. 2. Soft modeSoft mode–– Not necessaryNot necessary

3. 3. Hard modeHard mode–– Suitable for the Suitable for the

precise precise positioningpositioning

Page 19: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Control in power assisted modeControl in power assisted mode• The force which is applied to the patient’s branch by a surgeon is detected by the multi-axis force sensor.• The assist force is considered at the origin when the foot switch is pressed.• The variance of the force while pressing the foot switch is considered as an assist force.• The motion of the fracture reduction robot is controlled to make the variance force zero.

Page 20: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Automatic (NAVI) modeAutomatic (NAVI) modeUsed in the fracture reduction processUsed in the fracture reduction processControl points scheduled by NAVIControl points scheduled by NAVISmooth motion requiredSmooth motion requiredThe third order The third order splinespline curve interpolationcurve interpolationVelocity control: position not guaranteedVelocity control: position not guaranteed

t [sec]v

Page 21: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Video of automatic modeVideo of automatic mode

Page 22: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Experiments

Page 23: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Synchronized motion of the robot with an operation table

The fracture reduction robot moves up and The fracture reduction robot moves up and down down synchroniticallysynchronitically with an operation table by with an operation table by the operation to the operation table.the operation to the operation table.

Page 24: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Fail-safe mechanism

Page 25: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Evaluation of the fail-safe mechanism

A-axisB-axis

C-axisX-axis

Y-axis

Z-axis

Active point Active point

Loading LoadingRefuge motion,Excessive force absorption

Refuge motion,Excessive force absorption

Rated valueY-axis: 300 NB-axis: 28 Nm

(1) Y-axis, minus direction (2) B-axis, minus direction

Page 26: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Power assisted mode

Page 27: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Evaluation of the power assisted function (B-axis, ultra-soft mode)

ON OFFON OFFFoot switch Foot switch

(1) Position (2) Force

Page 28: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Evaluation of the power assisted function (B-axis, hard mode)

ON OFFON OFFFoot switch Foot switch

(1) Position (2) Force

Page 29: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Biomechanical data acquisition to apply the developed system

for the clinical use

Page 30: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Pain

Pain

Limit

Limit

Motion range of the inferior branch (Subject A: Female, Normal volunteer)(a) Right leg external rotation

(b) Right leg internal rotation

Page 31: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Motion range of the inferior branch (Subject A: Female, Normal volunteer)

LimitPain in pubis

Pain in knee joint

(c) Right leg traction

Page 32: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

• Limit value for internal rotation is smaller than that for external rotation.• Limit value is different for the right and the left legs.

Motion range of the inferior branch (Subject A: Female, Normal volunteer)(d) Left leg external rotation

(e) Left leg internal rotation

Limit

Limit

Pain in knee joint

Pain in hip joint

Page 33: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

Motion range of the inferior branch (Subject A: Female, Normal volunteer)

LimitPain in knee joint

Pain in hip joint

(f) Left leg traction

Page 34: Computer-Integrated Femoral Head Fracture Reduction System · robotic system, an X-ray system, and an operation table. The robot has 6 d.o.f., such as three translation d.o.f. and

ConclusionsThe authors have developed a robot system for femoral head fracture reduction.The system consists of a navigation system, a robotic system, an X-ray system, and an operation table.The robot has 6 d.o.f., such as three translation d.o.f. and three rotational d.o.f.Three operating modes were prepared: JOG mode, power assisted mode, and an automatic mode.The biomechanical data to apply the developed system for clinical use were obtained.