50
The Pennsylvania State University The Graduate School College of Engineering COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING A Thesis in Mechanical Engineering by Justin A. Jones 2017 Justin A. Jones Submitted in Partial Fulfillment of the Requirements for the Degree of Master of Science May 2017

COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

The Pennsylvania State University

The Graduate School

College of Engineering

COMPLIANT MEDICAL TOOL DESIGN

FOR SOFT TISSUE CUTTING

A Thesis in

Mechanical Engineering

by

Justin A. Jones

2017 Justin A. Jones

Submitted in Partial Fulfillment

of the Requirements

for the Degree of

Master of Science

May 2017

Page 2: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

ii

The thesis of Justin A. Jones was reviewed and approved* by the following:

Jason Z. Moore

Assistant Professor of Mechanical Engineering

Thesis Advisor

Reuben Kraft

Shuman Early Career Professor

Karen Thole

Distinguished Professor of Mechanical Engineering

Head of the Department of Mechanical Engineering

*Signatures are on file in the Graduate School

Page 3: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

iii

ABSTRACT

Needles are some of the most commonly used medical devices in minimally invasive

procedures. Examples of these procedures include drug delivery, biopsies and brachytherapy

cancer treatment. The efficacy of these procedures is highly dependent on the ability for precise

placement of the needle. However, high insertion forces can lead to complications resulting in

poor needle placement and possibly a poor outcome for the patient. Ongoing concerns for

physician include the needle diverging from the target path and deformation or rotation of the soft

tissue, both the result of high insertion forces. Several studies have been carried out to reduce the

insertion force in the needle insertion procedure. These include geometry studies, the speed of

insertion, the use of vibrational insertion, and more recently the use of a novel compliant hinge to

act as a cutting mechanism, however the work with the novel compliant hinge was limited to the

exploration of the concept. This research advances knowledge in tissue cutting using a novel

compliant needle design. The properties of the compliant needle design are empirically derived

and used to determine the cutting motion of a dynamically inserted needle. By understanding how

the compliant mechanism aids in transverse cutting motion an improved needle can be achieved.

Finite element analysis is used to determine empirical compliant equations for varying

hinge geometries. Polynomial regression equations are generated to create the empirical equations

for each directional stiffness. The limitation to the empirical equations are determined to be a

function of the distance between the hinges and the thickness of the hinge. It was found that if the

distance between the hinges is less than 1.68 times the thickness than the cross hinge is a primary

flexible member. After this, its contribution to the compliance quickly diminishes. To verify the

compliant equations the models are compared to FEA results. The percent error for a distance

ratio of 1.04, 1.57, 3.15, and 4.72 was 4.7, 23, 164, and 438 percent respectively. This research

allows for the design of the compliant mechanisms to be simplified for these ranges. Outside of

Page 4: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

iv

these ranges it is possible to use compliant equations developed by Lobontiu et al.[1] found in

Appendix A.

Page 5: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

v

TABLE OF CONTENTS

List of Figures .......................................................................................................................... v

List of Tables ........................................................................................................................... vi

Acknowledgements .................................................................................................................. vii

Chapter 1 Introduction ............................................................................................................. 1

Motivation ........................................................................................................................ 1 Needle Cutting Background ............................................................................................. 2 Role of Compliant Mechanism ........................................................................................ 5 Research Outline .............................................................................................................. 6

Chapter 2 Flexure Hinge Design Theory ................................................................................. 7

Compliant Mechanisms .................................................................................................... 7 Pseudo-Rigid Body Method ............................................................................................. 11 Finite Element Method ..................................................................................................... 14

Chapter 3 Empirical Compliant Equations for Asymmetric Transition ................................... 16

Variable Geometric Study ................................................................................................ 17 Finite Element Analysis Setup ......................................................................................... 17 Results .............................................................................................................................. 19

FEA Empirical results for d/t study .......................................................................... 20 FEA Empirical Results for h/t study ........................................................................ 22 FEA Empirical Results for t/w Study ....................................................................... 24

Validation of FEA Based Empirical Equations................................................................ 25 Limitation of Lumped Hinge Model ................................................................................ 29

Chapter 4 Conclusions ............................................................................................................. 33

Limitations and Future Work ........................................................................................... 33

References ................................................................................................................................ 35

Appendix Selected Compliant Equations ....................................................................... 40

Page 6: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

vi

LIST OF FIGURES

Figure 1-1. Common Complications from (a) needle deflection and (b) soft tissue

deflection. ......................................................................................................................... 2

Figure 1-2. Fracture Component Forces for Needle Insertion. ................................................ 3

Figure 1-3. (a) Novel compliant hinge design and (b) the transverse cutting motion

perpendicular to insertion direction. ................................................................................ 5

Figure 2-1. Rigid-link vise grip with its compliant mechanism counterpart [51]. ................... 7

Figure 2-2. Common Flexural Hinge Geometries for Planer Motion. ..................................... 8

Figure 2-3. (a) Flexible hinge connecting two rigid links and (b) its equivalent discrete

spring model counterpart [1] ............................................................................................ 9

Figure 2-4. (a) Circular Flexure Hinge and (b) transversely asymmetric hinge [1]. ................ 10

Figure 2-5. Transition of the (a) circular symmetric hinge to a (b) transversely

asymmetric hinge. ............................................................................................................ 11

Figure 2-6. (a) A flexible cantilever beam with end load, and (b) along with its PRBM

counterpart [51] ................................................................................................................ 12

Figure 3-1. Geometric parameters for the asymmetric hinge design. ...................................... 16

Figure 3-2. Free Body Diagram of Simulation Setup. ............................................................. 18

Figure 3-3. Example of meshing for FEA simulation of flexure hinge with rigid body A

and B. ............................................................................................................................... 19

Figure 3-4. Rotational Stiffness Empirical Model for d/t study. .............................................. 21

Figure 3-5. Rotation of the symmetric hinge into two aysmmetric hinges with short center

element. ............................................................................................................................ 22

Figure 3-6. Rotational Stiffness Empirical Model for h/t study. .............................................. 23

Figure 3-7. Rotational Stiffness Model for t/w study ............................................................... 25

Figure 3-8. Verification Simulation Set Parameters. ............................................................... 26

Figure 3-9. Rotation Angle at End of Beam for d/t of 1.04. .................................................... 27

Figure 3-10. Rotation Angle at End of Beam for d/t of 1.57. .................................................. 27

Figure 3-11. Rotation Angle at End of Beam for d/t of 3.15. .................................................. 28

Figure 3-13. AStrain Energy of compliant mechanism from FEA. ......................................... 30

Page 7: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

vii

Figure 3-14. Analytical Strain Energy calculation to determine transition period. ................. 31

Figure 4-1. (a) Flexible hinge connecting two rigid links and (b) its equivalent discrete

spring model counterpart.................................................................................................. 34

Page 8: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

viii

LIST OF TABLES

Table 3-1. Design Parameters for Geometric Study. .............................................................. 17

Table 3-2. Coefficients for Empirical Stiffness Equations for d/t ratio. ................................. 21

Table 3-3. Coefficients for Empirical Stiffness Equations for h/t ratio. ................................. 23

Table 3-4. Coefficients for Empirical Stiffness Equations for t/w ratio.................................. 24

Table 3-5. Needle Geometric Parameters for Verification. .................................................... 26

Table 3-6. Verification Results. Rotation from M = -50 Nmm load ........................................ 29

Page 9: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

ix

ACKNOWLEDGEMENTS

To my Wife.

Page 10: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

1

Chapter 1

Introduction

Motivation

Needles are among the mostly widely used medical devices. Some minimally invasive

procedures include tissue biopsy, blood sampling, drug delivery, and the cancer treatment of

interstitial brachytherapy [2]. The ability to accurately place the needle with respect to the

location of the target is paramount for the procedures to be successful. For example, a poor needle

placement during tissue biopsy can lead to a false negative diagnosis or unwanted tissue damage

and result in a poor prognosis. Another example is that of internal radiation therapy, also known

as brachytherapy treatment, a common treatment plan for many types of cancer. During

interstitial brachytherapy radioactive seeds are placed inside of the body in a precise pattern near

or inside of the tumor. It can be seen then that if the placement is poor, the outcome may not be

optimal.

One such type of cancer commonly treated with interstitial brachytherapy is that of

prostate cancer. Prostate cancer is the most frequently diagnosed cancer in men with an estimated

238,590 new cases of prostate cancer to arise in the United States in 2013 [3]. In permanent

brachytherapy 50-70 radiation seeds, roughly the size of a grain of rice, are placed by hand using

thin gauged needles into and around the tumor using ultrasound guidance [4, 5]. The accuracy of

the needle placement is typically between 2 to 3 mm inside the tissue [6]. The level of accuracy is

constrained due to the chance of edema from multiple needle repositioning. An edema may

further lead to complications such as shifting of anatomy or the seed position after the procedure

[7]. With this knowledge, physicians will commonly implant more seeds than necessary to

Page 11: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

2

compensate [8]. This type of over seeding will lead to a larger dose of radiation and can lead to

complications and a lower quality of life for the patient [4, 9]. A needle that can improve the

accuracy during these minimally invasive procedures is needed to improve patient outcomes.

Needle Cutting Background

In many of these minimally invasive procedures, small diameter needles are chosen in

order to lessen pain and reduce recovery time [10]. However, with a thinner needle, there is an

increased chance of deflection from insertion forces [11, 12]. This type of needle deflection can

lead to inaccurate needle placement as seen in Figure1-1(a). Additionally, the soft tissue target is

free to rotate and deflect as seen in Figure1-1(b). This soft tissue deflection is hard to predict due

to the heterogeneous nature of soft tissue. Prostate tissue is especially prone to tissue deflection.

One study showed that the prostate is capable of rotations up to a maximum of 13.8° [13]. This

high degree of rotation would greatly affect the dosage of radiation in brachytherapy [14].

Figure 1-1. Common Complications from (a) needle deflection and (b) soft tissue deflection.

(a)

(b)

Page 12: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

3

Previous studies have investigated ways to improve the needle insertion process. Some of

these methods included a predictive and correction method that predicts the deformation and

needle trajectories by modeling the needle-tissue interaction [15, 16], measuring needle insertion

forces [17-20], and the use of robotic-assisted tools [21-24]. However, these types of predictive

methods are limited in use due to the large variability found in soft tissue.

The cause of poor needle placement can be attributed to the high tissue cutting force that

acts parallel to the direction of needle insertion as seen in Figure 1-2. The reduction of high tissue

cutting force that acts parallel to the needle insertion could allow for increased accuracy which

would greatly improve the outcomes of medical procedures.

Figure 1-2. Fracture Component Forces for Needle Insertion.

Previous work has investigated how to reduce these insertion forces. One method was to

look at the effects of needle tip geometry on insertion forces. Studies have shown that by

changing the needle tip geometry the needle can cut with less force [25, 26]. There has been work

on optimizing the standard hypodermic grind angles of the three plane bevel tip [27, 28].

Additionally, a five plane bevel-tipped needle has been shown to cut with less force [29-31]

Further investigation in reducing insertion force include reducing the diameter of the

needle. Reducing the gauge of a needle from 30 gauge to 20 gauge have shown to reduce patient

pain as well as a reduction in insertion force by 20% [10]. A large body of work has been

investigating microneedles. Microneedles have sub-micron diameters and have shown to

Page 13: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

4

significantly reduce insertion force as well as pain [32-35]. However, microneedles are limited in

that they cannot allow for larger particulates to be passed through the sub-micron diameters [36].

This would exclude them from many minimally invasive procedures including brachytherapy

seeding.

Insertion speed and vibratory insertion have been another area of research to reduce

insertion forces. A study by Heverly et al. [37] showed when the use of high insertion speed is

used there is a reduction in insertion force needed to puncture soft tissue. This has also been

demonstrated in porcine liver [38] and turkey breast tissue as well [39]. Conversely, Koelsman et

al. [40] showed an increase in insertion force of 12% when increasing the insertion speed into

skin like simulant, a tougher tissue type.

Vibrational cutting has been adopted in the medical field as well. For example harmonic

scalpels have been used to cut tissue and bone. These devices use high frequency vibration to

improve the safety, precision, and accuracy of the cutting operation [41]. Vibrational cutting with

needles has been shown to decrease insertion forces in excised animal skin [42], reduce tissue

deformation in chicken breast [43] as well as silicone gel [21]. One study by Huang et al. showed

that ultrasonic vibrations reduce needle insertion force by 28% [44]. Other work has shown a

reduction of insertion force of up to 70% with the use of microneedles and axial vibrations [45].

However, none of these studies investigates vibration using a needle designed to generate cutting

motions. It was shown by Barnet et al. [46] that cutting force is reduced when a compliant needle

is coupled with vibration. This type of design allows for transverse cutting, however, it was left to

future work to fully explore how compliance affects the cutting motion.

Page 14: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

5

Role of Compliant Mechanism

As mentioned in traditional needle insertion the cutting force acts parallel to the insertion

direction. A novel compliant hinge initially developed by Barnet et al. [46] can be seen in Figure

1-3(a). This compliant needle design will allow for the high cutting parallel force to be redirected

to a more perpendicular direction relative to the target tissue and the insertion direction as seen in

Figure 1-3(b). This cutting motion will reduce the forces acting on the needle, thus reducing the

chance of the needle to deflect, as well as the forces pushing on the soft body target and in turn

reducing inaccurate placement of the needle with respect to the target. In all other ways the

needle geometry, grind angles and gauge, can remain the same as a traditional needle allowing for

this design to be available for a large number of applications.

Figure 1-3. (a) Novel compliant hinge design and (b) the transverse cutting motion perpendicular

to insertion direction.

Cutting force direction

d D

t

h

(a)

(b)

r

Page 15: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

6

The use of compliance in needle design is not solely novel. There have been several

studies that have investigated the use of built in compliance to allow for steerable needle

insertions and path prediction. Chen and Chen [47] used a multi-hinged design to generate an

algorithm to predict the path direction of a beveled needle during a traditional needle insertion

procedure. Swaney et. al [48] used a flexural hinge coupled with an internal actuator to reduce

tissue damage but maintain steerability of a needle compared to the more traditional kinked

bevel-tip needle. Again, there is little research on how a compliant mechanism could be used to

generate transverse cutting motion and improve the cutting mechanics. The work presented here

investigates the compliance for several asymmetric designs and will allow for the cutting motion

to be optimized using standard mechanism design practices.

Research Outline

The purpose of this research is to study the utility of a compliant hinge coupled with

dynamic insertion to reduce cutting forces. In order to design the needle, it is necessary to know

the compliance of the hinge. In Chapter 2 background on compliant mechanisms is provided. In

Chapter 3 a geometric design study is carried out using finite element analysis to generate

empirical compliant equations for a lumped asymmetric flexure hinge. The empirical equations

are then compared to FEA results to determine the error. Further, the range of when it is

appropriate to use the empirical model vs other models is explored. Chapter 4 provides a

summary of the findings as well as limitations and potential future works.

Page 16: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

7

Chapter 2

Flexure Hinge Design Theory

Compliant Mechanisms

The general definition of a mechanism is a mechanical device that is used to transfer

energy, force or motion [49, 50]. This can further be broken down into the more traditional rigid-

body mechanism, and compliant mechanisms. Where the rigid-body mechanism is one that is

made of rigid links and kinematic pairs, such as revolute joints, a compliant mechanism uses

flexible members to transfer input forces, energy or displacements using stored strain energy in

the flexible members of the mechanism [51].

Compliant mechanisms have gained favor recently over their rigid-body mechanisms

counterpart due to their cost reduction, increased performance, and reduction in parts, wear,

weight and maintenance [51]. For these reasons, compliant mechanisms are widely used in

microscale mechanisms, such as Micro-Electro-Mechanical Systems (MEMS). An example of

how a compliant mechanism can replace a traditional joint pair is shown in Figure 2-1.

Figure 2-1. Rigid-link vise grip with its compliant mechanism counterpart [51].

Flexible

members

Rigid body vice

grip

Page 17: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

8

One commonly used compliant mechanism is the flexure hinge, typically used to replace

a revolute joint. Like its rigid-body counterpart it acts as a kinematic pair that restricts movement

in the translation directions, however, unlike a the revolute joint, a flexural hinge can only create

a limited angular motion about its rotational axis due to the lack of freedom to fully rotate without

failure [52]. For a flexural hinge to be effective it needs to provide increased compliance about its

favored axis, but remain relatively rigid about the remaining axes to minimize any unwanted

motion and stay within the materials yield stress [52].

Generally, flexural hinges are classified within two categories; primitive flexural hinges

and complex flexural hinges. Complex flexure hinges provide multiple degrees of freedom and

are typical for three-dimensional applications. Primitive flexural hinges, the more common of the

two, predominantly focus on two-dimensional applications [53]. A few examples of the primitive

flexural hinge, which includes the circular flexural hinge, corner filleted, and elliptical, can be

seen in Figure 2-2. The monolithic design of these flexural hinges allows for high reproducible

motion, reduced friction and reduced wear [51].

Figure 2-2. Common Flexural Hinge Geometries for Planer Motion.

Circular Hinge Elliptical Hinge

Right Fillet

Hinge

Axially Symmetric

Transversely

Symmetric

Hyperbolic

Hinge

Page 18: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

9

For any flexural hinge designs, the defining characteristic is the compliance of the hinge,

which allows for the relationship between displacement and force to be determined. In Figure 2-3

a flexure hinge with a discrete spring model can be seen. If only two dimensional planer motion is

assumed, common in primitive flexure hinges, then this relationship is defined as:

M

F

F

C

u

u

u

y

x

y

x

][ (2.1)

where ux, uy, uϑ are the displacements in the axial, transverse and rotational direction, Fx,, Fy, Fϑ

are the applied forces in the axial, transverse or rotational direction. The compliance, C is defined

as:

zy

zy

x

MF

MyFy

Fx

CC

CC

C

C

0

0

00

(2.2)

Here, the axial, transverse, and rotational compliances are denoted by the Cx, Cy, and Cθ

respectively. It is common to report either the compliance or the stiffness of the hinge. Here the

stiffness will be used to describe the hinge where stiffness K is defined as:

1 CK (2.3)

Figure 2-3. (a) Flexible hinge connecting two rigid links and (b) its equivalent discrete spring model

counterpart [1]

Flexible Hinge

Member Rigid

Link

Rigid

Link Kx

Ky

Kϑ Rigid

Link Rigid

Link

(a)

(b)

Page 19: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

10

There has been considerable research on analyzing and designing compliant mechanisms.

In 1965 Paros and Weisbord [54] published the first work on a compliance-based approach to

flexure hinges. They provided the compliance equations for a symmetric circular and right

circular flexure hinge. Smith et al. [53] used similar methods to provide the closed form

compliance equations for the elliptical flexure hinge. Compliant equations were further provided

by Lobontiu et al. who, using the reciprocity principle and Castigliano’s displacement theorem,

formulated closed-form equations for the symmetric corner-filleted [55], parabolic and hyperbolic

[56], and the symmetric circular [57] hinge types. Lobontiu’s compliance equations for a

symmetric circular hinge as well as the transversely asymmetric hinge, Figure 2-4, are based on

the capacity of rotation about its sensitive axis and is a function of the thickness of the hinge. For

the compliant needle design either of these equations could be used if d is zero, fully symmetric

hinge, or far enough apart, transversely asymmetric, that there would be no coupling effects

between the two hinges. As previously stated, the purpose of this study is to determine a

compliance model for this transition phase as seen in Figure 2-5.

Figure 2-4. (a) Circular Flexure Hinge and (b) transversely asymmetric hinge [1].

t

t(x) t(x)

t

x

(a) (b)

Page 20: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

11

Figure 2-5. Transition of the (a) circular symmetric hinge to a (b) transversely asymmetric hinge.

Pseudo-Rigid Body Method

The Pseudo-Rigid-Body Method (PRBM) was developed to simplify the design and

analysis of compliant mechanisms. Previously a common method to analyze the deflection for

flexible mechanisms was to use elliptic integrals. However, this can be exceedingly difficult for

anything but simple geometries and not provide intuitive results [51]. The PRBM creates a

parametric approximation model of the deflection of flexible members using rigid-body

components that have equivalent force-deflection characteristics [51]. Additionally, the PRBM

can be used to compute the forces required for a given deflection and works for linear as well as

non-linear applications. A straight cantilever beam PRBM with a vertical end load developed by

Howell et al. [58] can be seen in Figure 2-6 and has been provided to give context to the theory.

(a)

(b)

Transition as d increases

Page 21: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

12

Figure 2-6. (a) A flexible cantilever beam with end load, and (b) along with its PRBM counterpart

[51]

Figure 2-6 (b) shows the PRBM of a simple cantilever beam with and end load. The

PRBM consists of two rigid links, connected by a characteristic pivot point, γ. The characteristic

pivot is the fraction of the beam length where the path of the rigid beam matches closely to the

path of the compliant beam. Once the location of the characteristic pivot is determined, the

deflection path may be described in terms of the pseudo-rigid-body angle, Θ, as described by

equations (2.6) – (2.8). The distance from the end of the beam to the pivot point is known as the

EI

F

Undeflected Position

Torsional

Spring, K

y

x

y

x0

F

y0

x

γl

Θ

x0

y0

(a)

(b)

L

Characteristic

pivot, γ

ϑ0

F Fn

Ft

Page 22: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

13

characteristic radius, γl. The angle of rotation, ϑ0, of the compliant beam can also be determined

using equation (2.9), where Cϑ is the angle coefficient equal to 1.2385 [51, 59]. The PRBM has

been shown to be able to predict the deflection of the end of the beam within 0.5% of the closed-

form elliptic integral solutions for quite large deflections by using a γ of 0.8517 allowing a

maximum Θ of 64.3 degrees[59]. A limitation of this method is that the accuracy decreases as the

relative length of the hinge increases [51].

)cos1(10 l

x (2.6)

sin0 l

y (2.7)

)1(tan

0

01

lx

y (2.8)

C0 (2.9)

The PRBM also incorporates a torsional spring located at the characteristic pivot location

that is used to model the stiffness, K, in the beam. From Figure 2.6(b) the initial force causing the

beam to bend tangential component can be described by equation 2.10.

l

KFFt

)sin( (2.10)

This transverse force creates the torque, T, at the characteristic pivot point.

lFT t (2.11)

Substituting equation (2.11) into (2.12) it is then possible to write the equation for torque as a

function of the spring stiffness and pseudo rigid body angle.

KT (2.12)

Howell presented a simplified way for calculating the consent spring stiffness, K. Using the beam

geometry, I/l, and Young’s Modulus, E, as well as the PRBM constants γ and Kϑ the spring

Page 23: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

14

stiffness can be calculated using equation (2.13). Howell et al. [59] presented an average Kϑ of

2.61 for load angles of 11.3 ͦ < φ < 174.3 ͦ.

l

EIKK (2.13)

Finally the equation (2.13) can be used to accurately predict the forces.

2l

EIKFt

(2.14)

Finite Element Method

The use of numerical methods such as the finite element method (FEM) have been

utilized more due to the inherent difficulties of using the previously mentioned analytical

methods. Where analytical methods can typically only be used for simple geometries and require

complicated derivations, FEM allows for more general compliant mechanisms to be modeled

more efficiently. There have been several research groups that have used FEM to derive empirical

compliant equations and compared the results to known analytical solutions. Yong et al. used

FEM to create empirical stiffness equations for a circular flexural hinge where the ratio of the

thickness of the hinge with respect to the radius of the hinge was iterated through [60]. They then

compared analytically derived equations of the rotational and translational compliance to the

FEM predictions. They showed the percentage error of the empirical equations was less than 3%.

Meng et al. [61] used FEM to explore the existing stiffness equations for corner filleted flexural

hinges. They created three empirical stiffness equations for corner-filleted flexural hinges based

on finite element analysis by iterating over the thickness with respect to the length of the hinge. In

this study, they found that the empirical models were a closer approximation, 6% error, than the

Page 24: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

15

analytical methods developed by other research groups. A similar approach is used to formulate

the empirical compliant equations for the lumped asymmetric flexure hinge design.

Page 25: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

16

Chapter 3

Empirical Compliant Equations for Asymmetric Transition

It is necessary to design an asymmetric flexure hinge so when an axially vibration is

applied a transverse cutting motion will be induced. This type of hinge can be seen in Figure 3-1.

As previously mentioned in order to design and analyze a compliant mechanism the compliance

of the flexure hinge is required. For a simple longitudinal asymmetric hinge we could potentially

use an equation similar to the one described by Lobontiu et al. [1], however, it is necessary to

fully understand the interactions of the two hinges when they are in close proximity.

A geometric study is performed using the finite analysis method. The resulting empirical

stiffness equations are then presented. It is shown that there is a transition phase between when a

lumped model is appropriate and the more traditional compliant equations can be used. To better

determine this transition period the elastic strain energy of the mechanisms is investigated.

Figure 3-1. Geometric parameters for the asymmetric hinge design.

D

h

L

Rigid

Body B

Rigid Body A

t

d

* *

x

y

RP1 RP2

Page 26: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

17

Variable Geometric Study

Three design parameters were chosen to explore due to their physical relationship to

beam bending and can be seen in Figure 3-1 and described in Table 3-1. The first design explored

was how the distance between the two hinges affected the compliance of the lumped hinge

mechanism. This was done by iterating the distance between the hinges such that the ratio was

0.05 ≤ d/t ≤ 8, where d is the distance between the hinges and, t is the thickness of the hinges. The

second design explored was how the thickness, t, with respect to the length, h, of the hinge

affected the compliance of the hinge mechanism. This was done by iterating the thickness and the

length of the hinge between 0.1 ≤ h/t ≤ 0.5 with a fixed d of 1/6th the total length, L. For the third

study, the effect of the thickness of the hinge, t with respect to the thickness of the rod, w was

explored. This was done by iterating the thickness of the hinge from 0.1 ≤ t/w ≤ 0.9.

Table 3-1. Design Parameters for Geometric Study.

Study d(mm) t(mm) L(mm) h(µm) w(mm) Number of iterations

d/t 0.032 – 5.08 0.635 5-10.08 90 1.27 25

h/t 0.8 0.635 5 63.5- 318 1.27 25

t/w 0.8 0.127 - 1.14 5 90 1.27 25

Finite Element Analysis Setup

ABAQUS software was used to model the hinge designs. To calculate the stiffness for

the parametric studies, the hinges were modeled as three-dimensional cantilever beams fully fixed

on the left end and free on the right end as seen in Figure 3-2. To exclude any deformation caused

by either end of the beam, both the areas left and right of the hinge were modeled as rigid bodies.

Page 27: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

18

For the loading conditions either an end axial load, Fx, of 1N, or end moment load, Mϑ, of 1N*D

was applied at the reference point, RP1 seen in Figure 3-2. For transverse loading, Fy, a load of

1N was applied perpendicular to the beam at reference point RP2 in the negative direction. To

isolate the stiffness in the transverse direction a negative moment was applied to the non-loaded

hinge equal to:

dFM yfy (3.1)

The displacement in the axial, transverse and rotational directions were recorded from

RP2 and used to calculate the stiffness in the respective direction using equation 2.1. The material

properties chosen for the study were that of AISI 304 stainless steel with a Modulus of Elasticity

of 200 GPa, density of 8000 kg/m3 and a Poisson’s ratio of 0.29.

Figure 3-2. Free Body Diagram of Simulation Setup.

The mesh chosen for this study were quadratic hexahedral elements (C3D8R) and an

example of the mesh can be seen in Figure 3-3. Since the right and the left end of the geometry

are modeled as rigid bodies, it is not critical to have a fine mesh. However, it is critical that the

mesh for the hinge is adequate to capture an accurate approximation of the material response. A

mesh convergence study was performed to determine the optimal seeding for the geometry. It was

determined that after a mesh density of approximately 700 elements/mm3 there is no appreciable

improvement, so this seeding was chosen for all hinges design models.

Fy Fx

Mz

Mfy RP2 RP1

* *

Page 28: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

19

Figure 3-3. Example of meshing for FEA simulation of flexure hinge with rigid body A and B.

Results

For all design studies, the recorded displacements were used to calculate the stiffness for

each design iteration. This stiffness was then plotted versus the design ratio. A polynomial

regression line was then fit to determine the function of stiffness, R2 > 0.9. It was found that the

rotational stiffness was the dominating factor of the three principle directions, as expected. The

rotational stiffness plots from each study can be seen in Figures 3-4, 3-6 and 3-7. Similar graphs

were produced for the stiffness’s in each of the remaining direction to determine the polynomial

function. The form of the empirical stiffness equations are presented in Equations 3.2(a)-(d):

n

i

in

inx a

Ew

K

0

)( (3.2a)

n

i

in

inbEw

K

0

)( (3.2b)

n

i

in

in

yc

Ew

K

0

)( (3.2c)

n

i

in

in

fyd

Ew

K

0

)(

(3.2d)

Page 29: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

20

Where the Young’s modulus (E), and the diameter (w) of the rod are divided out in order to create

a non-dimensional function. The design ratio, α, is multiplied by its corresponding polynomial

coefficients that can be found in Tables 3-2 to - 3-4.

FEA Empirical results for d/t study

Table 3-2 and Figure 3-4 show the empirical stiffness equations for the study that

explores how the hinge stiffness changes with respect to the length between the hinges. It can be

seen that Figure 3-4 is broken into two distinct phases. In Phase A, the distance between the two

hinges is very small with respect to the hinge thickness and the loading conditions on both hinges

are nearly equivalent. The stiffness of these hinge configurations are a product of the hinge

thickness, t and the cross hinge thickness, Tc, between the two hinges as seen in Figure 3-5. As d

is increased further, the cross hinge thickness increases until the mechanism becomes two distinct

asymmetric hinges with a short center section that can be considered a short rod element. For

some distance this rod elements contribution to stiffness can be ignored. This can be seen in

Euler-Bernoulli’s beam equation for constant cross section beams:

EI

Ml (3.3)

It is then easy to see that as the length goes to zero, the rotational displacement for the rod

section will go to zero. However, the cross thickness Tc is still equivalent to the hinge thickness, t

during this distance. In this way it is possible to consider the hinges as a lumped mechanism. In

Phase B, the distance between the two hinges starts to have a more dominating effect on the

overall stiffness. Therefore, in Phase B the previous assumption is not valid and the section of the

rod in-between the hinges must be considered using a different function.

Page 30: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

21

Table 3-2. Coefficients for Empirical Stiffness Equations for d/t ratio.

Figure 3-4. Rotational Stiffness Empirical Model for d/t study.

0.00E+00

5.00E-09

1.00E-08

1.50E-08

2.00E-08

2.50E-08

0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00

Rota

tional

Sti

ffnes

s (Kϑ/E

w)

Non-Dimensional Design Parameter (d/t)

Phase BPhase A

n a b c d

7 1.032E-04 --- 6.331E-05 ---

6 -0.0031 6.139E-12 -0.002 9.829E-10

5 0.0386 -1.365E-10 0.0253 -2.098E-08

4 -0.2451 1.036E-09 -0.1679 1.450E-07

3 0.8491 -2.482E-09 0.6163 -2.099E-07

2 -1.5479 -4.694E-09 -1.2128 -1.517E-06

1 1.2451 2.590E-08 1.1034 5.376E-06

0 0.0140 -4.092E-09 -0.2597 -8.696E-07

Page 31: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

22

Figure 3-5. Rotation of the symmetric hinge into two aysmmetric hinges with short center element.

FEA Empirical Results for h/t study

In Figure 3-6 the rotational stiffness is shown to be nearly linear, R2 of 0.96, with the

change in length of the hinge. Table 3-3 provides the coefficients for a polynomial regression

with an R2 of 0.99. As the hinge length increases the rotational stiffness reduces. However, it

should be noted that the scale of the rotational stiffness does not change as much as any of the

other studies. It has a maximum stiffness of 5.6 Nm/rad and a minimum stiffness of roughly 4.8

Nm/rad. It can be seen that the length of the hinge has little to no effect on the overall stiffness of

the lumped hinge design. The reason for this can be seen in Castigliano’s theorem for bending:

l

dxEI

QMM

0

(3.4)

It can be seen that for a constant flexural rigidity, EI, a constant moment, M, for any arbitrary

load, Q, the deflection. Δ, will not increase sizably due to the short length and the even smaller

d

Tc

t

Short Rod Element

d

Increases

Page 32: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

23

increment. This would obviously not be the case for much larger lengths, but due to design

restrictions is outside the scope of this study due to long hinges having a potential of causing

unwanted tissue damage.

Table 3-3. Coefficients for Empirical Stiffness Equations for h/t ratio.

n a b c d

7 -776.197 --- --- ---

6 1.995E03 --- -1.147E03 ---

5 -2.053E03 -4.436E-08 2.0087E03 ---

4 1.0918E03 -2.867E-07 -1.383E03 -4.285E-05

3 -322.317 3.541E-07 475.895 4.594E-05

2 52.63 -1.258E-07 -86.1929 -1.433E-05

1 -4.4513 6.2121E-09 7.7681 -4.8345E-07

0 0.4168 2.2069E-08 -0.1837 4.0978E-06

Figure 3-6. Rotational Stiffness Empirical Model for h/t study.

1.85E-08

1.90E-08

1.95E-08

2.00E-08

2.05E-08

2.10E-08

2.15E-08

2.20E-08

0.10 0.20 0.30 0.40 0.50 0.60

Rota

tional

Sti

ffnes

s (K

ϑ/E

w)

Non-Dimensional Design Parameter (h/t)

Page 33: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

24

FEA Empirical Results for t/w Study

Rotational stiffness is shown to greatly increase as the hinge becomes thicker as seen in

Figure 3-7. There is an increase of over 500 times as the thickness increases from 10% up to 90%

of the total diameter of the rod. Again, referring back to Castigliano’s theorem it is apparent that

the deflection is dependent on the second moment of area, I:

R

dAI 2 (3.5)

Where ρ is the distance from the origin of the axis to the differential area. It is then

readily apparent that as the cross-sectional area becomes smaller, the deflection will increase.

However, for very thin hinges the stress experienced would cause failure in the part so care must

be taken. Table 3-4 provides the coefficients for the polynomial regression fits for Equations

3.2(a)-(d). All fits had an R2 greater than 0.97.

Table 3-4. Coefficients for Empirical Stiffness Equations for t/w ratio

n a b c d

7 --- --- 3.313E03 ---

6 52.107 --- -8.649E03 ---

5 -153.631 --- 8.390E03 ---

4 173.620 -3.511E-07 -3.651E03 -6.480E-05

3 -88.528 7.241E-07 627.645 1.335E-04

2 23.241 -3.397E-07 11.9023 -6.289E-05

1 -2.899 7.741E-08 -13.8646 1.445E-05

0 0.1325 -5.397E-09 0.9536 -1.0214E-06

Page 34: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

25

Figure 3-7. Rotational Stiffness Model for t/w study

Validation of FEA Based Empirical Equations

To validate the empirical compliant equations the end displacement and rotation angle of

several needle designs were calculated using the empirical equations. The results were then

compared to FEA results. Figure 3-8 shows the design parameters and loading condition for the

simulation. The values of parameters for each needle, N1-N4, can be found in Table 3-5. For

validation, only the distance with respect to hinge thickness is looked at. AISI 304 stainless steel

with a Modulus of Elasticity of 200 GPa, density of 8000 kg/m3 and a Poisson’s ratio of 0.29 was

used for the FEA simulation. For loading conditions an end moment, M, was incrementally

increased by 5 Nmm from 0 to 100 Nmm at the end of the needle. The left end of the needle was

fixed in all directions. The end displacement, as well as the angle of rotation at the right end of

the needle, were recorded.

0.00E+00

1.00E-08

2.00E-08

3.00E-08

4.00E-08

5.00E-08

6.00E-08

7.00E-08

8.00E-08

9.00E-08

1.00E-07

0 0.2 0.4 0.6 0.8 1

Rota

tional

Sti

ffnes

s (K

ϑ/E

w)

Non-Dimensional Design Parameter (t/w)

Page 35: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

26

Table 3-5. Needle Geometric Parameters for Verification.

D1 (mm) D2 (mm) r (mm) t (mm) h (µm)

N1 1 0.33 0.3175 0.3175 90

N2 1 0.5 0.3175 0.3175 90

N3 1 1.0 0.3175 0.3175 90

N4 1 1.5 0.3175 0.3175 90

Figure 3-8. Verification Simulation Set Parameters.

It can be seen in Figures 3-9 to 3-12 that the lumped model starts to deviate from the FEA

results as the distance between the hinges increases as well as when the load is increased. As the

distance, D2, is increased the empirical model over predicts the stiffness of the flexure hinge.

This is to be expected, due to the added deflection of the rod section in between the hinges starts

to become the dominating contributor to the overall rotational deflection. Table 3-6 shows the end

rotation angle percent error increases from an average of 4.7% for a ratio of 1.04 to an average of

438% for the ratio of 4.72. This increasing error highlights that there is a limited range that the

* D2

D1

Mz

y

x z

r t

Page 36: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

27

empirical compliance model can be used for. It is necessary to know when the transition from

Phase A to Phase B occurs in order to know when it is appropriate to use the empirical model.

Figure 3-9. Rotation Angle at End of Beam for d/t of 1.04.

Figure 3-10. Rotation Angle at End of Beam for d/t of 1.57.

0 20 40 60 80 100

-0.3

-0.25

-0.2

-0.15

-0.1

-0.05

0

End Moment (Nmm)

Angula

r R

ota

tion (

rad)

N1

FEA

Empircal

0 20 40 60 80 100

-0.2

-0.18

-0.16

-0.14

-0.12

-0.1

-0.08

-0.06

-0.04

-0.02

0

End Moment (Nmm

Angula

r R

ota

tion (

rad)

N2

FEA

Empirical

Page 37: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

28

Figure 3-11. Rotation Angle at End of Beam for d/t of 3.15.

Figure 3-12. Rotation Angle at End of Beam for d/t of 4.72.

0 20 40 60 80 100

-0.18

-0.16

-0.14

-0.12

-0.1

-0.08

-0.06

-0.04

-0.02

0

End Moment (Nmm

Angula

r R

ota

tion (

rad)

N3

Empircal

FEA

0 20 40 60 80 100

-0.25

-0.2

-0.15

-0.1

-0.05

0

End Moment (Nmm)

Angula

r R

ota

tion (

rad)

N4

FEA

Empirical

Page 38: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

29

Table 3-6. Verification Results. Rotation from M = -50 Nmm load

FEA – ϑ (rad) Empirical – ϑ (rad) Average Percent Error(%)

N1 -0.126703 -0.12077 4.97

N2 -0.097389 -0.10869 23.97

N3 -0.083813 -0.02878 164.6

N4 -0.121683 -0.02077 438

Limitation of Lumped Hinge Model

To determine when the transition from Phase A and Phase B occurs elastic strain energy

can be used. This allows the mechanism to be decomposed into its individual components then

the individual strain energy values can be compared to the total strain energy of the system. As

previously described, if the mechanism is considered to be a lumped model as the distance

between the centers of the hinges is increased, the new configuration can be considered a single

hinge that is rotating and increasing in thickness, Tc. At some distance, d, the lumped hinge is no

longer valid since the distance, t, will be much less than the Tc, and it will be more appropriate to

model it as two separate hinges. This is due to the fact that at some point, the solid rod sections

ability to store elastic strain energy will outpace that of the lumped hinge model and cannot be

neglected. By determining the strain energy in the total mechanism while the degree of

asymmetry is increased then comparing it to numerically determined strain energy of a single

hinge model versus two hinges and a rod model it will be possible to determine the transition.

To determine the total elastic strain energy of the mechanism an FEA simulation was

performed using ABAQUS software. To simplify the numerical calculations the hinges were

loaded under a pure end moment of 1 Nmm. As before quadratic hexahedral elements (C3D8R)

Page 39: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

30

were used to mesh the hinge geometry. The left end of the hinge was fixed in all directions while

the moment end was applied to the right end of the hinge. As for the design iteration, the degree

of asymmetry was increased by from 0.02 ≤ d/t ≤ 10. After the simulation completed the total

strain energy was recorded and can be seen in Figure 3-13. The phases can clearly be seen, with

Phase A transitioning to Phase B approximately around d/t of two.

Figure 3-13. AStrain Energy of compliant mechanism from FEA.

For the numerical calculations, the only loading condition for this set up is a moment

about the z axis. To calculate the strain energy for both models the strain from pure bending was

used.

L

bending dsEI

MU

2

(3.6)

As for the two hinges with a rod it is numerically solved as two constant hinges with an

increasing rod length starting from 0 ≤ L < 8 mm. Then using equation 3.6 the individual strain

1.00E-04

1.00E-03

1.00E-02

1.00E-01

0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00

Str

ain E

ner

gy (

Nm

m)

Distance Ratio (d/t)

Phase B

Phase A

Page 40: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

31

energy was calculated. For the cross hinge design a function for the second moment of area (I)

was generated by fitting a polynomial to several calculated geometry’s and can be seen in

equation 3.7.

10-4E - 09t(x)-3E

...09t(x)-6E - 09t(x)-6E 09t(x)--3E 234

I (3.7)

Where t(x) is defined as:

x)-x(2r-rt xt (3.8)

Figure 3-14. Analytical Strain Energy calculation to determine transition period.

The elastic strain energy for both models can be seen in Figure 3-14. It can be seen that at

a distance ratio of 1.68 the strain energy of a rod model starts to quickly outpace the lumped

hinge model. This is also supported by the FEA determined total strain energy seen previously in

Figure 3-13. It stands to reason then that when the distance between the hinges is less than 1.68

the cross hinge is a primary flexible member. Within this range, the empirical compliant models

allow for a simplified way to determine the transverse motion of the needle tip. For ranges greater

than the 1.68 a rod section, seen in Figure 3-5, can no longer be ignored or modeled as a rigid

1.00E-07

1.00E-06

1.00E-05

1.00E-04

1.00E-03

1.00E-02

1.00E-01

0.00 2.00 4.00 6.00 8.00 10.00

Str

ain E

ner

gy (

Nm

m)

Distance Ratio (d/t)

Cross Hinge

Constant Rod Element

Page 41: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

32

element since its contribution to the overall elastic strain energy quickly outpaces that of a cross

hinge configuration. After a ratio of five, there is a difference of an order of magnitude between

the two models. At this point, it is best to model the hinge mechanism as a multilink-hinge

mechanism using previously mention compliant equations.

Page 42: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

33

Chapter 4

Conclusions

Empirical compliance models for an asymmetric flexural hinge were presented in this

paper. It was shown that the asymmetric hinge design, modeled as a single lumped element, has a

limited scope. Since the length between the hinges will eventually become a dominating factor in

the overall stiffness. However, can greatly simplify the design of a compliant needle within a

given range. It was also shown that the cross-sectional area of the hinge, as well as the distance

between the hinges, affect the stiffness of the hinges greater than the length of the hinge. It is

recommended that if the distance between the flexure hinges is less than 1.68 times the thickness

of the hinge that the empirical compliance model can be used for the mechanism design.

However, if the distance is larger it should be modeled using non-symmetric compliance

equations, such as the ones developed by Lobontiu et al. [1], and a multilink approach.

Limitations and Future Work

This work presented an empirical compliance model for a hinge to be coupled with

ultrasonic axial vibration. Previous work demonstrated the feasibility of this type of hinge to

generate transverse cutting motion. With the current compliance models it will be possible to

optimize the needle design. This future work will have to incorporate stress considerations into

the optimization, due to the compliance model not taking into account fatigue failure from the

cyclic loading conditions. Further work should be done to improve the models prediction around

the transition between Phase A and Phase B.

Page 43: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

34

A PRB kinematic model can be used to analyze the motion of the cutting path of the

needle tip. It can be seen in Figure 4-1 that the needle can be modeled as a double link

manipulator with two torsion springs with a stiffness Kϑ connecting two rigid links or if link 2 is

ignored a single rigid link with a torsion spring. If a single lumped mechanism is used the

stiffness of the torsion springs can be determined by the empirical equations presented in chapter

3. The driving torque, M(t), can be applied to point 1 and should be a sinusoidal function of time

to mimic the loading conditions of the vibratory needle.

Figure 4-1. (a) Flexible hinge connecting two rigid links and (b) its equivalent discrete spring model

counterpart.

Y

X x

y

y2 x2

x1

y1

θ1

θ2

k1

k2

L2

L1

M(t)

Page 44: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

35

References

[1] N. Lobontiu, Compliant mechanisms: design of flexure hinges: CRC press,

2002.

[2] N. Abolhassani, R. Patel, and M. Moallem, "Needle insertion into soft

tissue: A survey," Medical Engineering & Physics, vol. 29, pp. 413-431,

May 2007.

[3] "Cancer Facts and Figures," American Cancer Society, 2013.

[4] J. C. Blasko, T. Mate, J. E. Sylvester, P. D. Grimm, and W. Cavanagh,

"Brachytherapy for carcinoma of the prostate: Techniques, patient

selection, and clinical outcomes," Seminars in Radiation Oncology, vol. 12,

pp. 81-94, Jan 2002.

[5] A. T. Porter, J. C. Blasko, P. D. Grimm, S. M. Reddy, and H. Ragde,

"Brachytherapy for Prostate-Cancer," Ca-a Cancer Journal for Clinicians,

vol. 45, pp. 165-178, May-Jun 1995.

[6] G. Strassmann, P. Olbert, A. Hegele, D. Richter, E. Fokas, N. Timmesfeld,

R. Hofmann, and R. Engenhart-Cabillic, "Advantage of robotic needle

placement on a prostate model in HDR brachytherapy," Strahlenther Onkol,

vol. 187, pp. 367-72, Jun 2011.

[7] G. Fichtinger, J. P. Fiene, C. W. Kennedy, G. Kronreif, I. Iordachita, D. Y.

Song, E. C. Burdette, and P. Kazanzides, "Robotic assistance for

ultrasound-guided prostate brachytherapy," Med Image Anal, vol. 12, pp.

535-45, Oct 2008.

[8] Y. Yu, L. L. Anderson, Z. Li, D. E. Mellenberg, R. Nath, M. C. Schell, F.

M. Waterman, A. Wu, and J. C. Blasko, "Permanent prostate seed implant

brachytherapy: report of the American Association of Physicists in

Medicine Task Group No. 64," Med Phys, vol. 26, pp. 2054-76, Oct 1999.

[9] N. N. Stone and R. G. Stock, "Complications following permanent prostate

brachytherapy," European Urology, vol. 41, pp. 427-433, Apr 2002.

[10] H. Egekvist, P. Bjerring, and L. Arendt-Nielsen, "Pain and mechanical

injury of human skin following needle insertions," European Journal of

Pain, vol. 3, pp. 41-49, 1999.

[11] D. Glozman and M. Shoham, "Image-guided robotic flexible needle

steering," IEEE Transactions on Robotics, vol. 23, pp. 459-467, 2007.

[12] B. T. Sitzman and D. R. Uncles, "The effects of needle type, gauge, and tip

bend on spinal needle deflection," Anesthesia and Analgesia, vol. 82, pp.

297-301, 1996.

Page 45: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

36

[13] V. Lagerburg, M. A. Moerland, J. J. Lagendijk, and J. J. Battermann,

"Measurement of prostate rotation during insertion of needles for

brachytherapy," Radiother Oncol, vol. 77, pp. 318-23, Dec 2005.

[14] S. Nath, Z. Chen, N. Yue, S. Trumpore, and R. Peschel, "Dosimetric effects

of needle divergence in prostate seed implant using I-125 and Pd-103

radioactive seeds," Med Phys, vol. 27, pp. 1058-1066, May 2000.

[15] S. P. DiMaio and S. E. Salcudean, "Needle insertion modeling and

simulation," Ieee Transactions on Robotics and Automation, vol. 19, pp.

864-875, Oct 2003.

[16] D. Glozman and M. Shoham, "Flexible needle steering and optimal

trajectory planning for percutaneous therapies," Medical Image Computing

and Computer-Assisted Intervention - Miccai 2004, Pt 2, Proceedings, vol.

3217, pp. 137-144, 2004.

[17] H. Kataoka, T. Washio, K. Chinzei, K. Mizuhara, C. Simone, and A. M.

Okamura, "Measurement of the tip and friction force acting on a needle

during penetration," Medical Image Computing and Computer-Assisted

Intervention-Miccai 2002, Pt 1, vol. 2488, pp. 216-223, 2002.

[18] A. M. Okamura, C. Simone, and M. D. O'Leary, "Force modeling for

needle insertion into soft tissue," Ieee Transactions on Biomedical

Engineering, vol. 51, pp. 1707-1716, Oct 2004.

[19] T. Podder, D. Clark, J. Sherman, D. Fuller, E. Messing, D. Rubens, J.

Strang, R. Brasacchio, L. Liao, W. S. Ng, and Y. Yu, "In vivo motion and

force measurement of surgical needle intervention during prostate

brachytherapy," Med Phys, vol. 33, pp. 2915-2922, Aug 2006.

[20] X. Qin, J. Z. Moore, and Y.-S. Lee, "Analytical Modeling of Long Flexible

Needle Insertion on Deformable Bio-Tissues for Biomedical Applications,"

presented at the 43rd International Conference on Computers and Industrial

Engineering (CIE43), Hong Kong, October 16-18, 2013.

[21] D. Bi and Y. Lin, "Vibrating Needle Insertion for Trajectory Optimization,"

2008 7th World Congress on Intelligent Control and Automation, Vols 1-

23, pp. 7444-7448, 2008.

[22] D. D. Gao, Y. Lei, and H. J. Zheng, "Needle Steering for Robot-Assisted

Insertion into Soft Tissue: A Survey," Chinese Journal of Mechanical

Engineering, vol. 25, pp. 629-638, Jul 2012.

[23] M. A. Meltsner, N. J. Ferrier, and B. R. Thomadsen, "Observations on

rotating needle insertions using a brachytherapy robot," Physics in

Medicine and Biology, vol. 52, pp. 6027-6037, Oct 7 2007.

[24] G. Wan, Z. P. Wei, L. Gardi, D. B. Downey, and A. Fenster,

"Brachytherapy needle deflection evaluation and correction," Med Phys,

vol. 32, pp. 902-909, Apr 2005.

Page 46: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

37

[25] T. Podder, D. Clark, J. Sherman, D. Fuller, E. Messing, D. Rubens, J.

Strang, Y. Zhang, W. O’Dell, and W. Ng, "Effects of tip geometry of

surgical needles: an assessment of force and deflection," in IFMBE Proc,

2005, pp. 1727-1983.

[26] M. A. Towler, W. McGregor, G. T. Rodeheaver, P. V. Cutler, R. F. Bond,

D. Phung, R. G. Morgan, J. G. Thacker, and R. F. Edlich, "Influence of

cutting edge configuration on surgical needle penetration forces," The

Journal of emergency medicine, vol. 6, pp. 475-481, 1988.

[27] Y. Wang, R. K. Chen, B. L. Tai, P. W. McLaughlin, and A. J. Shih,

"Optimal needle design for minimal insertion force and bevel length,"

Medical Engineering and Physics, vol. 36, pp. 1093-1100, 2014.

[28] Y. Wang, B. L. Tai, R. K. Chen, and A. J. Shih, "The needle with lancet

point: geometry for needle tip grinding and tissue insertion force," Journal

of Manufacturing Science and Engineering, vol. 135, p. 041010, 2013.

[29] J. L. Doyle and S. L. Koziol, "Five beveled point geometry for a

hypodermic needle," ed: Google Patents, 1998.

[30] L. Hirsch, M. Gibney, J. Berube, and J. Manocchio, "Impact of a modified

needle tip geometry on penetration force as well as acceptability,

preference, and perceived pain in subjects with diabetes," Journal of

diabetes science and technology, vol. 6, pp. 328-335, 2012.

[31] L. Vedrine, W. Prais, P. Laurent, C. Raynal-Olive, and M. Fantino,

"Improving needle-point sharpness in prefillable syringes," Medical Device

Technology, vol. 14, p. 32, 2003.

[32] S. P. Davis, B. J. Landis, Z. H. Adams, M. G. Allen, and M. R. Prausnitz,

"Insertion of microneedles into skin: measurement and prediction of

insertion force and needle fracture force," Journal of Biomechanics, vol. 37,

pp. 1155-1163, 2004.

[33] S. P. Davis, W. Martanto, M. G. Allen, and M. R. Prausnitz, "Hollow metal

microneedles for insulin delivery to diabetic rats," IEEE Transactions on

Biomedical Engineering, vol. 52, pp. 909-915, 2005.

[34] S. Kaushik, A. H. Hord, D. D. Denson, D. V. McAllister, S. Smitra, M. G.

Allen, and M. R. Prausnitz, "Lack of pain associated with microfabricated

microneedles," Anesthesia and Analgesia, vol. 92, pp. 502-504, 2001.

[35] J.-H. Park, Y.-K. Yoon, S.-O. Choi, M. R. Prausnitz, and M. G. Allen,

"Tapered conical polymer microneedles fabricated using an integrated lens

technique for transdermal drug delivery," IEEE Transactions on

Biomedical Engineering, vol. 54, pp. 903-913, 2007.

[36] Y.-C. Kim, J.-H. Park, and M. R. Prausnitz, "Microneedles for drug and

vaccine delivery," Advanced drug delivery reviews, vol. 64, pp. 1547-1568,

2012.

Page 47: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

38

[37] M. Heverly, P. Dupont, and J. Triedman, "Trajectory optimization for

dynamic needle insertion," in Robotics and Automation, 2005. ICRA 2005.

Proceedings of the 2005 IEEE International Conference on, 2005, pp.

1646-1651.

[38] Y. Kobayashi, T. Sato, and M. G. Fujie, "Modeling of friction force based

on relative velocity between liver tissue and needle for needle insertion

simulation," in Engineering in Medicine and Biology Society, 2009. EMBC

2009. Annual International Conference of the IEEE, 2009, pp. 5274-5278.

[39] N. Abolhassani, R. Patel, and M. Moallem, "Trajectory generation for

robotic needle insertion in soft tissue," in Engineering in Medicine and

Biology Society, 2004. IEMBS'04. 26th Annual International Conference of

the IEEE, 2004, pp. 2730-2733.

[40] W. Koelmans, G. Krishnamoorthy, A. Heskamp, J. Wissink, S. Misra, and

N. Tas, "Microneedle characterization using a double-layer skin simulant,"

Mechanical Engineering Research, vol. 3, p. 51, 2013.

[41] M. R. Sadiq, Y. Kuang, S. Cochran, Z. Huang, and G. Corner, "High

performance ultrasonic tool for tissue cutting," in Biomedical Engineering

and Informatics (BMEI), 2011 4th International Conference on, 2011, pp.

2339-2342.

[42] M. Yang and J. D. Zahn, "Microneedle insertion force reduction using

vibratory actuation," Biomed Microdevices, vol. 6, pp. 177-82, Sep 2004.

[43] K. G. Yan, W. S. Ng, K. V. Ling, T. I. Liu, Y. Yu, and T. Podder, "High

frequency translational oscillation & rotational drilling of the needle in

reducing target movement," 2005 IEEE International Symposium on

Computational Intelligence in Robotics and Automation, Proceedings, pp.

163-168, 2005.

[44] Y. C. Huang, M. C. Tsai, and C. H. Lin, "A piezoelectric vibration-based

syringe for reducing insertion force," presented at the International

Symposium on Ultrasound in the Control of Industrial Processes, Madrid,

Spain, 2012.

[45] M. Yang, and Zahn, J. D., "Microneedle Insertion Force Reduction Using

Vibratory Actuation,," Biomedical Microdevices, vol. 6, pp. 177-182, 2004.

[46] A. C. Barnett, J. A. Jones, Y.-S. Lee, and J. Z. Moore, "Compliant Needle

Vibration Cutting of Soft Tissue," Journal of Manufacturing Science and

Engineering, vol. 138, pp. 111011-111011, 2016.

[47] Z. Chen and Y. H. Chen, "Analysis of multi-hinge compliant needle

insertion," 2009 Ieee International Conference on Virtual Environments,

Human-Computer Interfaces and Measurement Systems, pp. 314-318,

2009.

[48] P. J. Swaney, Jessica Burgner, Hunter B. Gilbert, and Robert J. Webster. ,

"A Flexure-Based Steerable Needle: High Curvature With

Page 48: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

39

Reduced Tissue Damage," IEEE Transactions on Biomedical Engineering, vol.

60, pp. 906-909, 2013.

[49] J. E. Shigley, Uicker, J. J., Theory of Machines and Mechanisms. New

York: McGraw-Hill, 1995.

[50] A. G. Erdman, Sandor, G.N, Mechanism Design: Analysis and Synthesis

vol. 1. Upper Saddle River Prentice Hall, 1997.

[51] L. L. Howell, Compliant mechanisms: John Wiley & Sons, 2001.

[52] N. Lobontiu, J. S. N. Paine, E. Garcia, and M. Goldfarb, "Corner-Filleted

Flexure Hinges," Journal of Mechanical Design, vol. 123, p. 346, 2001.

[53] S. T. Smith, V. G. Badami, J. S. Dale, and Y. Xu, "Elliptical flexure

hinges," Review of Scientific Instruments, vol. 68, pp. 1474-1483, 1997.

[54] J. M. Paros, and Weisbord, L., "How to Design Flexure Hinges," Mach.

Des., pp. 151-156, 1965.

[55] N. Lobontiu, J. S. Paine, E. Garcia, and M. Goldfarb, "Corner-filleted

flexure hinges," Journal of Mechanical Design, vol. 123, pp. 346-352,

2001.

[56] N. Lobontiu, J. S. Paine, E. O’Malley, and M. Samuelson, "Parabolic and

hyperbolic flexure hinges: flexibility, motion precision and stress

characterization based on compliance closed-form equations," Precision

Engineering, vol. 26, pp. 183-192, 2002.

[57] N. Lobontiu, and Paine, J.S., "Design of circular cross-section corner-

filleted flexure hinges for three-dimensional compliant mechanisms.,"

Journal of Mechanical Design, vol. 124, pp. 479-484, 2002.

[58] L. L. Howell and A. Midha, "A method for the design of compliant

mechanisms with small-length flexural pivots," Journal of Mechanical

Design, vol. 116, pp. 280-290, 1994.

[59] A. MIdha, "Parametric deflection approximations for end-loaded, large-

deflection beams in compliant mechanisms," Journal of Mechanical

Design, vol. 117, p. 157, 1995.

[60] Y. K. Yong, T.-F. Lu, and D. C. Handley, "Review of circular flexure hinge

design equations and derivation of empirical formulations," Precision

Engineering, vol. 32, pp. 63-70, 2008.

[61] Q. Meng, Y. Li, and J. Xu, "New empirical stiffness equations for corner-

filleted flexure hinges," Mechanical Sciences, vol. 4, pp. 345-356, 2013.

Page 49: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

40

Appendix

Selected Compliant Equations

Lobontui et al. compliance equations

In-plane compliance for symmetric right-circular flexure hinge:

2

41arctan

4

221

t

r

trt

tr

EwFxxC

t

r

trt

trtrtr

trt

trttrtrrtrttr

trt

trtrrtr

trEwFyyC

41arctan

4

4628

4

214238248042

4

52844822

24

3

55

224

55

432234

22

223

t

rtrttrr

trtrtrt

trtrEwt

r

MzyC 4

1arctan426

464

42

242

22

33

2

r

MzyC

MzC

Page 50: COMPLIANT MEDICAL TOOL DESIGN FOR SOFT TISSUE CUTTING

41

Lobontui et al. compliance equations

In-plane compliance for nonsymmetric circular flexure hinge:

1

22arctan

2

221

trt

tr

trt

r

trt

tr

EwFxxC

22arctan

2

2432

2

312 223

22

3

trt

r

trt

trtrtr

tr

r

trt

trr

EwFyyC

22

arctan3

2243

)2(

122

22

55

2

trt

rtrr

trttrtr

trttrEw

r

MzyC

r

MzyC

MzC