Upload
shuai-zhang
View
214
Download
1
Embed Size (px)
Citation preview
FULL SHORT COMMUNICATION
Comparison of SPE-TD-GC-FID with UPLC-PDA and GC-MSMethods for Analysis of Benzene, Toluene and Xylene Isomersin Solid-Liquid Mixing Paints
Shuai Zhang • Tianbo Zhao • Haiwang Wang •
Jia Wang
Received: 21 December 2010 / Revised: 10 February 2011 / Accepted: 22 March 2011 / Published online: 22 April 2011
� Springer-Verlag 2011
Abstract Three methods were developed for analysis of
benzene, toluene, p-xylene, m-xylene and o-xylene (BTX)
in a solid-liquid mixing paint. These methods were based
on solid phase extraction-thermal desorption-gas chroma-
tography with a flame ionization detector (SPE-TD-GC-
FID), ultra performance liquid chromatography-photodiode
array detector (UPLC-PDA) and gas chromatography-mass
spectrometry (GC-MS). At their optimum conditions of
operation, the developed methods were compared in terms
of recovery, sensitivity, selectivity and universality.
Although the time required for GC-MS analysis was
shorter than for GC-FID and UPLC-PDA, it offered rela-
tively poorer recoveries and suffered from matrix inter-
ferences. All of the mentioned methods were proven to be
ideal for the analysis of targeted analytes; SPE-TD-GC-
FID was particularly fit for the determination of trace level
BTX residues present in the complex matrix. As one of the
sample pretreatment techniques, the novel SPE also
showed some selectivity towards BTX and was found to be
superior to the conventional SPE.
Keywords Gas chromatography-mass spectrometry �Solid phase extraction-thermal desorption � Gas
chromatographic-flame ionization detector � Ultra
performance liquid chromatography-photodiode array
detector � BTX in paint
Introduction
Chronic exposure to benzene, toluene, p-xylene, m-xylene
and o-xylene (collectively BTX) through vapor inhalation
causes disorders of the central nervous system and gas-
trointestinal tract. The most relevant consequences of
chronic poisoning are plastic anemia and leukemia [1].
BTX has already been confirmed to be a strong carcinogen
by the World Health Organization (WHO) and classified as
a priority pollutant by the US Environmental Protection
Agency (EPA) [2]. At present, the use of BTX is restricted
in the manufacturing of primers, top-finish paints, barrier
paints, fire-retardant paints, decoration adhesives and
diluents.
Widely used paints are known to contain a considerable
number of organic compounds and solid additives. For
example, BTX is present in complex matrices at a wide
range of concentrations [3]. From an analytical point of
view, the composition is responsible for the difficulties in
the analysis of trace level BTX residues present in solid-
liquid mixing paint. There are various options for chro-
matographic detection that depend upon the phase and/or
the concentration range of samples. For instance, direct
injection (DI) with a microsyringe can be adopted for
highly concentrated BTX samples [4]. However, in the
case of low concentration paint samples, i.e., trace level
BTX in solid-liquid mixing paint, the adoption of certain
preconcentration treatments is inevitable. The combination
of GC with the pretreatment stage (e.g., the use of SPE-TD)
is efficient for extending the detectability of complex
matrix samples [5]. In this work, we used Tenax as an
absorbent to accomplish selective pre-separation of BTX
and solvents.
Although there is a considerable amount of literature
[1, 6–14] on the analysis of BTX in different matrices
S. Zhang (&) � T. Zhao � H. Wang � J. Wang
Key Laboratory of Cluster Science of Ministry of Education,
Beijing Institute of Technology, Beijing 100081,
People’s Republic of China
e-mail: [email protected]
123
Chromatographia (2011) 74:163–169
DOI 10.1007/s10337-011-2029-z
(e.g., gaseous, aqueous and solid sediment), we could find
hardly any papers dealing with the application of UPLC-
PDA for the determination of BTX in solid-liquid mixing
paint samples. UPLC, as its first producer Waters claims,
means ‘speed, resolution and sensitivity’ [15], and it can be
regarded as a new direction for liquid chromatography.
Recent technological advances have resulted in the avail-
ability of reversed-phase chromatographic media of 1.7-lm
particle size and a liquid-handling system that can be
applied at much higher pressures. UPLC has significant
additional advantages in time saving and solvent con-
sumption as well [16–20].
The aim of this paper was to investigate and compare
the performance of SPE-TD-GC-FID with UPLC-PDA and
GC-MS. We tried to determine which method is the most
reliable solution for the determination of BTX compounds
in solid-liquid mixing paint.
Experimental
Chemicals and Reagents
Standards of benzene, toluene, p-xylene, m-xylene and
o-xylene were purchased from the State Center for Stan-
dard Matter (concentration levels: 0.97, 1.01, 0.98, 0.99
and 1.00 mg mL-1, respectively; Beijing, China). All of
the above standards were prepared in methanol. Methanol
of HPLC grade was obtained from Merck (Darmstadt,
Germany). Ultrapure water (18.2 MX cm, 25 �C, produced
by Millipore, USA) was used throughout the study. High-
purity nitrogen and helium (99.999%) were purchased from
South Asia Air Products (Beijing, China). The adsorption
tube was fabricated by filling Tenax 60–80 mesh (Beifen
Tianpu Instrument Technology Co. Ltd., Beijing, China)
into a silanized glass tube (150 9 6 mm o.d. 9 4 mm
I.D.). The Tenax adsorbent was kept in place by a piece of
stainless-steel wire.
Instrumentation
The Waters ACQUITYTM Ultra-Performance Liquid Chro-
matography system (UPLC, Waters, Milford, MA) is com-
posed of a photodiode array detector (PDA), sample manager,
column heater and binary solvent manager. The guard col-
umn (3.9 9 20 mm) was packed with Symmetry C18 (5-lm
particle size, Waters, USA), and the analytical column was
an ACQUITYTM BEH C18 column (100 9 2.1 mm I.D.,
1.7-lm particle size, Waters, Milford, MA).
The GC-MS (HP6890 N GC Series combined with the
5975I MS, Agilent, USA) was equipped with an Agilent
7863 autosampler (with an Agilent 10-lL syringe) and a
30-m HP-5MS capillary column with an I.D. of 0.25 mm
and a film thickness of 0.1 lm (J&W Scientific, Folsom,
CA).
Gas chromatographic analysis was performed by using
a GC1100, equipped with a flame ionization detector
(GC-FID, Purkinje General Instrument Co. Ltd., Beijing,
China). Hydrogen was supplied by a DH-500 high-purity
hydrogen generator (BCHP Analytical Technology Insti-
tute, Beijing, China). Chromatographic separation was
carried out on a stainless-steel packed column
(300 9 3 mm I.D.) containing 5% DNP ? 3.5% Ben-
tone34/Chromsorb WAW (80–100 mesh, Lanzhou Atech
Technologies Co. Ltd.). For the desorption of analytes from
the adsorption tube, a TP-2050 thermal desorption unit
(Beifen Tianpu Instrument Technology Co. Ltd., Beijing,
China) coupled to a GC-FID was adopted [3].
A glass insert for GC-MS was silane-treated by di-
chlorodimethylsilane in order to diminish adsorption of
compounds at the injection port. All of the glassware for
the preparation of samples and standard solutions was
cleaned with deionized water and acetone, and then dried at
room temperature.
Preparation of Standard Solutions
A stock solution containing benzene, toluene, p-, m- and
o-xylene standards was prepared by mixing the commercial
standards to a final concentration of 0.194, 0.202, 0.196,
0.198 and 0.200 mg mL-1, respectively. Later, the stock
solution mixture was diluted step by step with methanol to
appropriate concentrations for calibration curve construc-
tions and spiking tests. All these solutions were used for
peak identification and quantitative analysis as well and
stored at 4 �C in a refrigerator before use.
Preliminary Procedures for BTX Extraction from Paint
(1) A 0.01–0.1-g (accurate to 0.0001 g) commercial paint
sample was weighed into an extraction vessel;
(2) The analytes were extracted in 3 mL methanol
(ultrasonic-associated extraction, 40 kHz working
frequency and 100 W power, 40 �C, 3 min);
(3) The extract was collected in a 5-mL volumetric flask,
and the volume was made up to the mark with
methanol;
(4) The solution was filtered through a 0.45-lm micro-
syringe filter into 2-mL brown high-airtight vials and
stored at 4 �C in a refrigerator before analysis.
SPE-TD-GC-FID Analysis
To conduct a GC-FID-based analysis of standard solutions
and extracts, a GC system was equipped with a thermal
164 S. Zhang et al.
123
desorption (TD) unit as well as the solutions needed to be
further pretreated by the novel solid phase extraction (SPE)
technique.
SPE Procedures
As shown in the schematic diagram of Fig. 1, the SPE was
carried out in several steps as follows:
(1) Standard solution or extract (10–50 lL) was injected
through the head gasket seal into a Tenax adsorption
tube under 20 mL min-1 nitrogen flow. (One end
connected to the nitrogen flow while the other end
was exposed to the atmosphere.)
(2) Most of the solvents were eliminated by nitrogen
flow, and the BTX was selectively enriched in the
absorbent.
(3) Both ends of the Tenax adsorption tube were sealed
with silicone caps; the analytes were then desorbed
and analyzed as soon as possible after the preparation.
TD Operation
The TD unit was operated in the following sequences. At
the beginning, the Tenax adsorption tube was inserted into
the tube furnace, and the stainless-steel injection needle,
which was connected with the Tenax tube, directly pene-
trated through the septum for an injection port. Then, the
Tenax adsorption tube was rapidly heated up to 200 �C and
held for 5 min. Finally, the tube was purged by
30 mL min-1 nitrogen flow, and thermally desorbed ana-
lytes were subsequently delivered into the stainless-steel
packed column. The desorbed analytes were separated and
detected by GC-FID [3].
GC-FID Analysis
The GC-FID system was operated at a flow rate of
30 mL min-1 (high-purity nitrogen was used as carrier
gas). The flow rate of hydrogen was 55 mL min-1,
whereas the air flow rate was 400 mL min-1. The column
temperature was maintained at 110 �C. The injection port
temperature and the FID detector temperature were set at
250 and 260 �C, respectively.
UPLC-PDA Analysis
For UPLC-PDA analysis, all of the extracts were filtered
through a filtering cartridge with a 0.2-lm nylon mem-
brane. The injection (volume: 3.0 lL) was performed by an
autosampler with an injection needle. Other detail analyt-
ical conditions were as follows: needle wash, strong wash
(MeOH/H2O = 8/2): weak wash (MeOH/H2O = 1/9) =
600:300 lL; seal wash, MeOH/H2O = 1/10; re-equilibra-
tion time: 3 min; the mobile phase consisted of methanol
(A, 70%) and ultrapure water (B, 30%); the analysis were
achieved at a flow rate of 0.1 mL min-1 and the column
temperature fixed at 45.0 �C; the wavelength of PDA was
set at 203 nm for analysis; the data were collected and
processed by Waters Empower system.
GC-MS Analysis
For GC-MS analysis, the extract was dried with Na2SO4.
The temperature of the injection port, interface and qua-
druple as well as ion source was constantly maintained at
250, 260, 120 and 250 �C, respectively. The GC oven was
temperature programmed as follows: the column was ini-
tially maintained at 40 �C for 1 min. Subsequently, the
temperature was increased up to 120 �C at a rate of
40 �C min-1. Then, it was maintained at 120 �C for 2 min.
The total time for each GC run was 5 min. High-purity
helium (purity 5.0) was used as a carrier gas, and the
column flow rate was 1.2 mL min-1. Injection of 1 lL in
the splitless mode was performed by using an Agilent 7863
autosampler equipped with an Agilent 10-lL syringe. The
mass spectrometer setup was as follows: electron impact
mode, EI; ionization energy, 70 eV in the positive-ion
mode; repel voltage, 25 V; analytical mode, full scan
(mass range of m/z 50–150 a.m.u. with mass accuracy of
0.1 a.m.u.).
Fig. 1 Schematic diagram of the solid phase extraction device.
A assembled SPE device; B SPE decomposition diagram; a head
gasket seal, b three-way piece, c and d graphite seal, e tail pipe,
f cylinder, g stainless-steel injection needle, h nitrogen pipe, i silicon
rubber cap, j stainless-steel wire, k Tenax adsorbent
Comparison of SPE-TD-GC-FID with UPLC-PDA and GC-MS Methods 165
123
Results and Discussion
Conventional SPE and Novel SPE
SPE is a sample preparation method that is well adapted
especially to the handling of liquid samples. Trace ana-
lytes are trapped by a suitable sorbent packed in a car-
tridge through which the liquid is percolated and then
eluted with a small volume of organic solvent. Extraction
and concentration are performed simultaneously [21].
Although SPE is an effective technique for the extraction
and pre-concentration of analytes from gaseous or aque-
ous matrices, it tends to use expensive and toxic elution
solvents and suffers from problems of cartridge plugging
and significant background interferences, which have an
adverse affect on repeatability [22]. In this study, we tried
to improve the conventional SPE technique by injecting
solutions onto the Tenax absorbent under nitrogen flow.
Most of the solvents were successfully purged by nitrogen
flow, whereas BTX was still retained in the absorbent.
The significant advantage of this novel SPE technique is
the replacement of expensive and toxic elution solvents
by nitrogen. The result has shown good reproducibility
with very little matrix dependency. The results of recov-
eries and limits of detection (LOD) would strengthen our
opinion.
K. Dettmer and W. Engewald [23] reported that Tenax is
a very hydrophobic material characterized by high thermal
stability. Because of its low specific surface area
(30 m2 g-1), it is not suitable for the sampling of highly
volatile organics. Referring to hydrocarbons, Tenax was
used for compounds with carbon numbers higher than four
[24]. However, because of the hydrophilic characteristic of
methanol with a carbon number of only one, it is not easily
adsorbed by Tenax, whereas the situation was the reverse
for BTX. Some studies on Tenax [25–28] have primarily
explained the reasons why BTX was still retained in the
absorbent while most of the solvents were removed by
nitrogen flow at room temperature. Our present study on a
novel SPE technique has exhibited good agreement with
the previous works.
Comparison of SPE-TD-GC-FID with UPLC-PDA
and GC-MS
Under optimum conditions, the SPE-TD-GC-FID, UPLC-
PDA and GC-MS methods were applied to standard solu-
tions and real commercial paint samples in order to
determine their recoveries and LOD. The results are sum-
marized in Table 1, and the advantages and drawbacks of
all three techniques are also presented in Table 1.
Time Demand
The sample preparations for UPLC-PDA and GC-MS were
more straightforward than what was required for SPE-TD-
GC-FID analysis, i.e., only dilution and filtration were
required after the ultrasonic-associated extraction. For
GC-FID analysis, the SPE procedure and TD operation
increased the sample preparation time significantly. The
analysis time of GC-MS was relatively shorter (twice as
fast as previous reports) compared with previous experi-
ments reported in the literature [29, 30]. Although the
GC-MS method was able to enhance the rapidity of
analysis, the injection volume limitation due to column
capacity presented a major problem. In addition, the
detection of a trace amount BTX in paint samples was
severely affected by the tailing area of the solvent (espe-
cially for benzene). In the case of GC-FID and UPLC, the
analysis time was commonly longer than 20 min. Never-
theless, it could be shortened when the flow rate of the
mobile phase was increased.
Table 1 Comparison of the
application of SPE-TD-GC-
FID, UPLC-PDA and GC-MS in
the analysis of benzene, toluene
and xylene isomers in solid-
liquid mixing paint
? Satisfactory; ?? normal;
??? good; ???? excellent
SPE-TD-GC-FID UPLC-PDA GC-MS
Solvent demand Little Too much Little
Time for sample preparation (min) 12–15 7–10 3–5
Analysis time (min) 20 15 5
Peak separation ???? ??? ??
Identification ??? ?? ????
Sensitivity ???? ??? ???
Ruggedness of the system ??? ???? ?
Recovery % 92.3–97.4 95.1–100.7 79.8–89.3
LOD range (lg mL-1) 0.021–0.16 0.23–0.79 5.6–13
Automation ?? ???? ????
Universality ???? ??? ???
Cost Low Expensive High
166 S. Zhang et al.
123
Peak Separation
Figures 2, 3 and 4 show the chromatograms corresponding
to GC-FID, UPLC-PDA and GC-MS. Simultaneous
determinations of p-xylene, m-xylene and o-xylene were
initially attempted, but they were successfully separated
only by using a Bentone34/Chromsorb column. The iden-
tification of xylene isomers was a great challenge with the
UPLC-PDA and GC-MS methods. For both methods,
standard solutions were used to qualitatively and quanti-
tatively analyze BTX in paint samples, while in GC-MS,
qualitative analysis was based on library spectra (NIST/
EPA/NIH Mass Spectral Database) and fragmentation
patterns.
Although we need further analysis to confirm the com-
position of the mixed peaks in UPLC chromatograms for
the assignments of p- and m-, p- and o-, or m- and o-xylene,
our preliminary study has readily shown that the total
quantitative analysis of xylene isomers could be achieved.
Recovery Test
The accuracies of all developed methods were confirmed
by the following recovery test. Toluene was chosen as the
spiked analyte for the recovery test because there were no
interfering peaks presented in its chromatogram. The
spiked samples were extracted and analyzed under the
conditions as described above. The overall average
recoveries for solid-liquid mixing paint spiked at levels of
0.1, 0.3 and 0.5 mg mL-1 ranging from 79.8 to 89.3% for
GC-MS, 92.3 to 97.4% for GC-FID and 95.1 to 100.7%
for UPLC-PDA, as listed in Table 1. The results indicated
that all methods were satisfyingly accurate for the deter-
mination of the targeted analytes in solid-liquid mixing
paint.
Apparently, the recovery of GC-MS was remarkably
lower than those of GC-FID and UPLC-PDA. One reason
for the bad values by GC/MS could be the condition of the
MS instrument (impurities in the ion source, calibration of
mass scales, etc.). In case of a low concentration solution,
quantitative analysis of GC-MS can be affected by the
tailing area of the solvent, and loss of targeted analytes was
possible during sample injection. The splitless injection
mode is not absolutely splitless in reality; only over 95% of
analytes could be ensured to enter the chromatographic
column. UPLC-PDA does not behave in this way, since its
mobile phase and solution are using the same solvent, and
methanol has no strong adsorption at 203 nm. It became
obvious that the recovery of UPLC appeared to be superior
to GC methods.
LOD
The LODs were calculated on the basis that concentration
gives a signal three times greater than background noise.
The GC-FID and UPLC-PDA method exhibited excellent
LOD. As shown in Table 1, it is obvious that the LODs for
GC-FID were 2–4 orders of magnitude lower than those for
Fig. 2 GC-FID chromatograms obtained from the analysis of stan-
dard solution and real commercial paint sample. a BTX standards in
methanol, the concentration levels (lg mL-1) of the solutions were:
1.94, 2.02, 1.96, 1.98 and 2.00 for benzene, toluene, p-xylene,
m-xylene and o-xylene, respectively; b real commercial sample,
injection volume: 15 lL
Fig. 3 GC-MS chromatograms obtained from the analysis of stan-
dard solution and real commercial paint sample. a BTX standards in
methanol, concentration levels (mg mL-1) of the solutions were:
0.194, 0.202, 0.196, 0.198 and 0.200 for benzene, toluene, p-xylene,
m-xylene and o-xylene, respectively; b real commercial sample,
injection volume: 1.0 lL; SD solvent delay
Comparison of SPE-TD-GC-FID with UPLC-PDA and GC-MS Methods 167
123
GC-MS. There are two possible reasons for this
phenomenon:
(1) The injection volume of GC-FID was dozens of times
larger than that of GC-MS. Meanwhile, drawbacks
like the tailing area of the solvent could be overcome
by the SPE-TD pretreatment technique and
(2) MS and PDA were much less sensitive than FID for
BTX analysis. This could also be seen in the
repeatability of the peak area: better results were
obtained in GC-FID analysis, even after more com-
plicated SPE-TD pretreatment. In case of GC/MS, the
sensitivity can be increased using the SIM technique.
Conclusions
A new SPE technique comprised by a stainless-steel tube
packed with Tenax adsorbent material has enabled highly
efficient and repeatable adsorption and desorption of trace
quantities of BTX from the solid-liquid mixing paint
sample. The main advantages of SPE-TD-GC-FID in
comparison with UPLC-PDA and GC-MS included an
excellent matrix effect elimination capacity, very low
detection limit, high reproducibility, simplicity and sol-
vent-free elution. It is also noteworthy that the excellent
lifetime of the Tenax adsorption tube enables at least 150
desorption operations at 200 �C to be carried out besides
maintaining excellent stability and reproducibility. It is
suitable to analyze BTX in complex matrix in an accurate
manner.
Acknowledgments The authors would like to thank the China 111
project no. B07012 and the National Natural Science Funds no.
20973022 for supporting the research.
References
1. Mangani F, Lattanzi L, Rezai MA, Cecchetti G (2000) Chro-
matographia 52:217–220
2. Florez Menendez JC, Fernandez Sanchez ML, Sanchez Urıa JE,
Fernandez Martınez EF, Sanz-Medel A (2000) Anal Chim Acta
415:9–20
3. Zhang S, Zhao TB, Xu X, Wang HW, Miao C (2010) Chroma-
tographia 71:1131–1135
4. Pandey SK, Kim KH (2007) J Chromatogr A 1216:5436–5444
5. Kim KH, Pal R (2010) Environ Monit Assess 161:295–299
6. Campos-Candel A, Llobat-Estelles M, Mauri-Aucejo A (2009)
Talanta 78:1286–1292
7. Djozan D, Assadi Y (1997) Chromatographia 45:183–189
8. Mangani G, Berloni A, Maione M (2003) Chromatographia
58:115–117
9. Prikryl P, Kubinec R, Jurdakova H, Sevcık J, Ostrovsky I, Sliak
L, Berezkin V (2006) Chromatographia 64:65–70
10. Wang ST, Wang Y, You H, Yao J (2006) Chromatographia
63:365–371
11. Sarafraz-Yazdi A, Amiri AH, Es’haghi Z (2008) Chemosphere
71:671–676
12. Sarafraz-Yazdi A, Amiri AH, Es’haghi Z (2009) Talanta
78:936–941
Fig. 4 UPLC-PDA
chromatograms obtained from
the analysis of standard solution
and real commercial paint
sample. a Real commercial
sample, b BTX standards in
methanol (same standard
solution as for GC-MS was
used); injection volume: 3.0 lL
168 S. Zhang et al.
123
13. Sieg K, Fries E, Puttmann W (2008) J Chromatogr A
1178:178–186
14. Godoi AFL, Sawada EY, Marchi MRR, Grieken RV, Godoi
RHM (2009) Water Air Soil Pollut Focus 9:163–169
15. Waters Corporation (2004) Ultra performance LC by design.
Waters Corporation USA, 720000880EN:LL. &LW-UL
16. Wu T, Wang C, Wang X, Xiao HQ, Ma Q, Zhang Q (2008)
Chromatographia 68:803–806
17. Guvenc A, Satir E, Coskum M (2007) Chromatographia
68:S141–S145
18. Li X, Xiong Z, Ying X, Cui L, Zhu W, Li F (2006) Anal Chem
Acta 580:170–180
19. Guan J, Lai CM, Li SP (2007) J Pharm Biomed Anal
44:996–1000
20. Qin LX, Chao J, Wei Y, Yum L, Li YM, Gang CX (2008)
Chromatographia 68:57–63
21. Rivasseau C, Caude M (1995) Chromatographia 41:462–470
22. Magdic S, Pawliszyn J (1996) J Chromatogr A 723:111–122
23. Dettmer K, Engewald W (2002) Anal Bioanal Chem
373:490–500
24. Yang KL, Lo JG (1997) Chromatographia 44:405–410
25. Sakodynskii K, Panina L, Klinskaya N (1974) Chromatographia
7:339–344
26. Agelopoulos NY, Pickett JA (1998) J Chemical Ecology
24:1161–1172
27. Briand O, Millet M, Bertrand F, Clement M, Seux R (2002) Anal
Bioanal Chem 374:848–857
28. Zhao DY, Pignatello JJ (2004) Environ Toxicol Chem
23:1592–1599
29. Liu JF, Li N, Jiang GB, Liu JM, Jonsson JA, Wen MJ (2005) J
Chromatogr A 1066:27–32
30. Carrion CC, Lucena R, Cardenas S, Valcarcel M (2007) J
Chromatogr A 1171:1–7
Comparison of SPE-TD-GC-FID with UPLC-PDA and GC-MS Methods 169
123