  • Comparison of EPA 8260B to EPA TO-15

    Pearl J PereiraSenior Scientist

    IRIS EnvironmentalIrvine, California

    949-724-8900http://www.irisenv.com/

    IRIS EnvironmentalIRIS Environmental

    DTSC Soil Gas Advisory ForumMarch 4, 2008

    http://www.irisenv.com/

  • This presentation focuses on data quality challenges associated with differences in data sampling techniques and analytical methods

    IRIS Environmental

  • Understanding the problem

    IRIS Environmental

    Active Oil Well

    Potential Vapor Plume

    Road

    Not to scale

    Abandoned Oil Wells

  • Sampling Program

    A total of 209 soil gas samples were collected from 172 sample locations

    IRIS Environmental

  • Difference in Installation and Sampling Techniques

    IRIS Environmental

    Collection: Syringes Summa Canisters with flow controllers

    Method: On-Site Mobile Off-Site Fixed BasedLaboratory Laboratory

    Analysis: EPA 8260B EPA TO-15

    On-site Off-site

    172 Soil Gas Sample Locations

  • Field Observations inconsistent with current advisory

    No site-specific lithological information was collected prior to the soil gas investigation

    IRIS Environmental

    Three-stage wash and rinse decontamination of drilling rods and other re-usable components were NOT practiced

    Sample locations were installed along busy street at onefoot below ground surface

    No leak test or site-specific purge test was conducted

    Sampling equipment was stored on and near petroleum/diesel generators prior to installation

  • Results

    IRIS Environmental

    Tracer compound was detected in three of the on-site samples at values ranging from 17 to 7,200 parts per billion by volume (ppbv)

    Benzene was detected in blank samples by on-sitemobile laboratory at

  • Maximum Concentrations Comparison

    0

    500

    1000

    1500

    2000

    2500

    3000

    3500

    1,1,1-

    TRICH

    LORO

    ETHA

    NE

    1,1,2-

    TRIC

    HLOR

    O-1,2

    ,2-TR

    IFLUO

    ROET

    HAN

    1,2,4-

    TRIM

    ETHY

    LBEN

    ZENE

    1,2-D

    ICHLO

    ROET

    HANE

    1,3,5-

    TRIM

    ETHY

    LBEN

    ZENE

    1,3-B

    UTAD

    IENE

    2-BUT

    ANON

    E

    2-HEX

    ANON

    EAC

    ETON

    E

    4-ETH

    YLTO

    LUEN

    E

    4-MET

    HYL-2

    -PEN

    TANO

    NE (M

    IBK)

    BENZ

    ENE

    BROM

    ODIC

    HLOR

    OMET

    HANE

    BROM

    OMET

    HANE

    BUTA

    NE

    CARB

    ON DI

    SULF

    IDE

    CARB

    ON TE

    TRAC

    HLOR

    IDE

    CHLO

    ROFO

    RM

    CHLO

    ROET

    HANE

    CHLO

    ROME

    THAN

    E

    CIS-

    1,2-D

    ICHLO

    ROET

    HENE

    DICHL

    OROD

    IFLUO

    ROME

    THAN

    EET

    HANO

    L

    ETHY

    LBEN

    ZENE

    HEPT

    ANE

    N-HE

    XANE

    M,P-X

    YLEN

    EMT

    BE

    METH

    YLEN

    E CHL

    ORIDE

    O-XY

    LENE

    PROP

    YLEN

    EST

    YREN

    EPC

    E

    TETR

    AHYD

    ROFU

    RAN

    TOLU

    ENE

    TCE

    TRIC

    HLOR

    OFLU

    OROM

    ETHA

    NE

    TRIM

    ETHY

    L PEN

    TANE

    VINYL

    ACET

    ATE

    VINYL

    CHL

    ORID

    EM

    axim

    um C

    once

    ntra

    tion

    (ppb

    (v/v

    ))

    On-Site Off-Site

    Acetone

    Benzene

    m,p-xylene

    TolueneTCE

    PropyleneStyrene PCE

  • Minimum Concentrations Comparison

    0

    5

    10

    15

    20

    25

    30

    35

    40

    1,1,1-

    TRIC

    HLOR

    OETH

    ANE

    1,1,2-

    TRIC

    HLOR

    O-1,2

    ,2-TR

    IFLU

    OROE

    THAN

    E

    1,2,4-

    TRIM

    ETHY

    LBEN

    ZENE

    1,2-D

    ICHL

    OROE

    THAN

    E

    1,3,5-

    TRIM

    ETHY

    LBEN

    ZENE

    1,3-B

    UTAD

    IENE

    2-BU

    TANO

    NE

    2-HE

    XANO

    NEAC

    ETON

    E

    4-ET

    HYLT

    OLUE

    NE

    4-ME

    THYL

    -2-P

    ENTA

    NONE

    (MIB

    K)BE

    NZEN

    E

    BROM

    ODIC

    HLOR

    OMET

    HANE

    BROM

    OMET

    HANE

    BUTA

    NE

    CARB

    ON D

    ISUL

    FIDE

    CARB

    ON T

    ETRA

    CHLO

    RIDE

    CHLO

    ROFO

    RM

    CHLO

    ROET

    HANE

    CHLO

    ROME

    THAN

    E

    CIS-

    1,2-D

    ICHL

    OROE

    THEN

    E

    DICH

    LORO

    DIFL

    UORO

    METH

    ANE

    ETHA

    NOL

    ETHY

    LBEN

    ZENE

    HEPT

    ANE

    N-HE

    XANE

    M,P-

    XYLE

    NEMT

    BE

    METH

    YLEN

    E CH

    LORI

    DEO-

    XYLE

    NE

    PROP

    YLEN

    EST

    YREN

    EPC

    E

    TETR

    AHYD

    ROFU

    RAN

    TOLU

    ENE

    TCE

    TRIC

    HLOR

    OFLU

    OROM

    ETHA

    NE

    TRIM

    ETHY

    L PEN

    TANE

    VINY

    L ACE

    TATE

    VINY

    L CHL

    ORID

    E

    Min

    imum

    Con

    cent

    ratio

    n (p

    pb(v

    /v))

    On-Site Off-Site1,1,1-Trichloroethane

    cis-1,2-dichloroethene vinyl acetate

  • Conclusion

    Data quality objectives were not defined

    IRIS Environmental

    Recommended guidelines were not followed

    Variability in analytical and sampling techniques did not produce reliable and defensible data sets.

  • Discussion

    IRIS Environmental

    How significant are confirmation samples?

    Appropriate selection of sampling techniques and analytical methods as defined by data quality objectives is a MUST

    Need for protocols to address discrepancies with data sets collected with different analytical techniques

    This presentation focuses on data quality challenges associated with differences in data sampling techniques and analytical meResults Maximum Concentrations ComparisonConclusionDiscussion