48
Waters Column Bypass Module Installation and Maintenance Guide 34 Maple Street Milford, MA 01757 71500032104, Revision A

Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

Waters Column BypassModule

Installation and Maintenance Guide

34 Maple StreetMilford, MA 01757

71500032104, Revision A

Page 2: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

NOTICE

The information in this document is subject to change without notice and should not be construed as a commitment by Waters Corporation. Waters Corporation assumes no responsibility for any errors that may appear in this document. This document is believed to be complete and accurate at the time of publication. In no event shall Waters Corporation be liable for incidental or consequential damages in connection with, or arising from, the use of this document.

© 2002 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA. ALL RIGHTS RESERVED. THIS DOCUMENT OR PARTS THEREOF MAY NOT BE REPRODUCED IN ANY FORM WITHOUT THE WRITTEN PERMISSION OF THE PUBLISHER.

Integrity, Nova-Pak, and Waters are registered trademarks, and Empower and Thermabeam are registered trademarks of Waters Corporation.

Teflon, Tefzel, and Vespel are registered trademarks of E. I. du Pont de Nemours and Company.

All other trademarks or registered trademarks are the sole property of their respective owners.

Page 3: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

Note: When you use the instrument, follow generally accepted procedures for quality control and methods development.

If you observe a change in the retention of a particular compound, in the resolution between two compounds, or in peak shape, immediately determine the reason for the changes. Until you determine the cause of a change, do not rely on the separation results.

Note: The Installation Category (Overvoltage Category) for this instrument is Level II. The Level II Category pertains to equipment that receives its electrical power from a local level, such as an electrical wall outlet.

STOPAtención: Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.

Important : Toute modification sur cette unité n’ayant pas été expressément approuvée par l’autorité responsable de la conformité à la réglementation peut annuler le droit de l’utilisateur à exploiter l’équipement.

Achtung: Jedwede Änderungen oder Modifikationen an dem Gerät ohne die ausdrückliche Genehmigung der für die ordnungsgemäße Funktionstüchtigkeit verantwortlichen Personen kann zum Entzug der Bedienungsbefugnis des Systems führen.

Avvertenza: eventuali modifiche o alterazioni apportate a questa unità e non espressamente approvate da un ente responsabile per la conformità annulleranno l’autorità dell’utente ad operare l’apparecchiatura.

Atención: cualquier cambio o modificación realizado a esta unidad que no haya sido expresamente aprobado por la parte responsable del cumplimiento puede anular la autorización de la que goza el usuario para utilizar el equipo.

Page 4: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

Caution: Use caution when working with any polymer tubing under pressure:

• Always wear eye protection when near pressurized polymer tubing.

• Extinguish all nearby flames.

• Do not use Tefzel tubing that has been severely stressed or kinked.

• Do not use Tefzel tubing with tetrahydrofuran (THF) or concentrated nitric or sulfuric acids.

• Be aware that methylene chloride and dimethyl sulfoxide cause Tefzel tubing to swell, which greatly reduces the rupture pressure of the tubing.

Attention : soyez très prudent en travaillant avec des tuyaux de polymères sous pression :

• Portez toujours des lunettes de protection quand vous vous trouvez à proximité de tuyaux de polymères.

• Eteignez toutes les flammes se trouvant à proximité.

• N'utilisez pas de tuyau de Tefzel fortement abîmé ou déformé.

• N'utilisez pas de tuyau de Tefzel avec de l'acide sulfurique ou nitrique, ou du tétrahydrofurane (THT).

• Sachez que le chlorure de méthylène et le sulfoxyde de diméthyle peuvent provoquer le gonflement des tuyaux de Tefzel, diminuant ainsi fortement leur pression de rupture.

Vorsicht: Bei der Arbeit mit Polymerschläuchen unter Druck ist besondere Vorsicht angebracht:

• In der Nähe von unter Druck stehenden Polymerschläuchen stets Schutzbrille tragen.

• Alle offenen Flammen in der Nähe löschen.

• Keine Tefzel-Schläuche verwenden, die stark geknickt oder überbeansprucht sind.

• Tefzel-Schläuche nicht für Tetrahydrofuran (THF) oder konzentrierte Salpeter- oder Schwefelsäure verwenden.

• Durch Methylenchlorid und Dimethylsulfoxid können Tefzel-Schläuche quellen; dadurch wird der Berstdruck des Schlauches erheblich reduziert.

Page 5: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

Precauzione: prestare attenzione durante le operazioni con i tubi di polimero sotto pressione:

• Indossare sempre occhiali da lavoro protettivi nei pressi di tubi di polimero pressurizzati.

• Estinguere ogni fonte di ignizione circostante.

• Non utilizzare tubi Tefzel soggetti a sollecitazioni eccessive o incurvati.

• Non utilizzare tubi Tefzel contenenti tetraidrofurano (THF) o acido solforico o nitrico concentrato.

• Tenere presente che il cloruro di metilene e il dimetilsolfossido provocano rigonfiamento nei tubi Tefzel, che riducono notevolmente il limite di pressione di rottura dei tubi stessi.

Advertencia: manipular con precaución los tubos de polimero bajo presión:

• Protegerse siempre los ojos a proximidad de tubos de polimero bajo presión.

• Apagar todas las llamas que estén a proximidad.

• No utilizar tubos Tefzel que hayan sufrido tensiones extremas o hayan sido doblados.

• No utilizar tubos Tefzel con tetrahidrofurano o ácidos nítrico o sulfúrico concentrados.

• No olvidar que el cloruro de metileno y el óxido de azufre dimetilo inflan los tubos Tefzel lo que reduce en gran medida la presión de ruptura de los tubos.

Page 6: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas
Page 7: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

Caution: The user shall be made aware that if the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.

Attention : L’utilisateur doit être informé que si le matériel est utilisé d’une façon non spécifiée par le fabricant, la protection assurée par le matériel risque d’être défectueuses.

Vorsicht: Der Benutzer wird darauf aufmerksam gemacht, dass bei unsachgemäßer Verwenddung des Gerätes unter Umständen nicht ordnungsgemäß funktionieren.

Precauzione: l’utente deve essere al corrente del fatto che, se l’apparecchiatura viene usta in un modo specificato dal produttore, la protezione fornita dall’apparecchiatura potrà essere invalidata.

Advertencia: El usuario deberá saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas de protección del equipo podrían ser insuficientes.

Page 8: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

Caution: To protect against fire hazard, replace fuses with those of the same type and rating.

Attention : Remplacez toujours les fusibles par d’autres du même type et de la même puissance afin d’éviter tout risque d’incendie.

Vorsicht: Zum Schutz gegen Feuergefahr die Sicherungen nur mit Sicherungen des gleichen Typs und Nennwertes ersetzen.

Precauzione: per una buona protezione contro i rischi di incendio, sostituire i fusibili con altri dello stesso tipo e amperaggio.

Precaución: sustituya los fusibles por otros del mismo tipo y características para evitar el riesgo de incendio.

Page 9: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

Caution: To avoid possible electrical shock, power off the instrument and disconnect the power cord before servicing the instrument.

Attention : Afin d’éviter toute possibilité de commotion électrique, mettez hors tension l’instrument et débranchez le cordon d’alimentation de la prise avant d’effectuer la maintenance de l’instrument.

Vorsicht: Zur Vermeidung von Stromschlägen sollte das Gerät vor der Wartung abgeschaltet und vom Netz getrennt werden.

Precauzione: per evitare il rischio di scossa elettrica, spegnere lo strumento e scollegare il cavo di alimentazione prima di svolgere la manutenzione dello strumento.

Precaución: para evitar choques eléctricos, apague el instrumento y desenchufe el cable de alimentación antes de realizar cualquier reparación en el instrumento.

Page 10: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

Commonly Used Symbols

Direct currentCourant continuGleichstromCorrente continuaCorriente continua

Alternating currentCourant alternatifWechselstromCorrente alternataCorriente alterna

Protective conductor terminalBorne du conducteur de protectionSchutzleiteranschlussTerminale di conduttore con protezioneBorne del conductor de tierra

Page 11: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

Frame or chassis terminalBorne du cadre ou du châssisRahmen- oder ChassisanschlussTerminale di struttura o telaioBorne de la estructura o del chasis

Caution or refer to manualAttention ou reportez-vous au guideVorsicht, oder lesen Sie das HandbuchPrestare attenzione o fare riferimento alla guidaActúe con precaución o consulte la guía

Caution, hot surface or high temperatureAttention, surface chaude ou température élevéeVorsicht, heiße Oberfläche oder hohe TemperaturPrecauzione, superficie calda o elevata temperaturaPrecaución, superficie caliente o temperatura elevada

Commonly Used Symbols (Continued)

Page 12: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

Caution, risk of electric shock (high voltage)Attention, risque de commotion électrique (haute tension)Vorsicht, Elektroschockgefahr (Hochspannung)Precauzione, rischio di scossa elettrica (alta tensione)Precaución, peligro de descarga eléctrica (alta tensión)

Caution, risk of needle-stick punctureAttention, risques de perforation de la taille d’une aiguilleVorsicht, Gefahr einer SpritzenpunktierungPrecauzione, rischio di puntura con agoPrecaución, riesgo de punción con aguja

Caution, ultraviolet lightAttention, rayonnement ultrvioletVorsicht, Ultraviolettes LichtPrecauzione, luce ultraviolettaPrecaución, emisiones de luz ultravioleta

Commonly Used Symbols (Continued)

UV

Page 13: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

Column Bypass Module Information

Intended Use

The Waters® Column Bypass Module can be used for in-vitro diagnostic testing to analyze many compounds, including diagnostic indicators and therapeutically monitored compounds. When you develop methods, follow the “Protocol for the Adoption of Analytical Methods in the Clinical Chemistry Laboratory,” American Journal of Medical Technology, 44, 1, pages 30–37 (1978). This protocol covers good operating procedures and techniques necessary to validate system and method performance.

Biological Hazard

When you analyze physiological fluids, take all necessary precautions and treat all specimens as potentially infectious. Precautions are outlined in “CDC Guidelines on Specimen Handling,” CDC – NIH Manual, 1984.

Calibration

Follow acceptable methods of calibration with pure standards to calibrate methods. Use a minimum of five standards to generate a standard curve. The concentration range should cover the entire range of quality-control samples, typical specimens, and atypical specimens.

Quality Control

Routinely run three quality-control samples. Quality-control samples should represent subnormal, normal, and above-normal levels of a compound. Ensure that quality-control sample results are within an acceptable range, and evaluate precision from day to day and run to run. Data collected when quality-control samples are out of range may not be valid. Do not report this data until you ensure that chromatographic system performance is acceptable.

Page 14: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

Table of Contents

Table of Contents 14

Chapter 1 Introduction ...................................................................................... 21

1.1 Valve Control Connections.................................................... 23

1.2 Internal Electrical and Helium Connections .......................... 23

1.3 Making Fluidic Connections .................................................. 24

Chapter 2 Operation ......................................................................................... 27

2.1 Column Bypass Module Fluidics Overview........................... 27

2.2 Injecting Calibrants with the Sample Management System.................................................................................. 33

2.3 Injecting Calibrants Manually ................................................ 34

Chapter 3 Troubleshooting ............................................................................... 36

3.1 Inject/Load Valve Problems................................................... 36

3.1.1 Tightening the Pressure-Adjusting Screw.................. 37

3.1.2 Replacing the Rotor Seal........................................... 40

3.1.3 Tightening the Needle Seal ....................................... 41

3.1.4 Testing and Replacing the Position-Sensing Switch ........................................................................ 42

3.2 Control Signal Problems ....................................................... 43

Page 15: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

Appendix A Specifications ................................................................................... 44

Appendix B Spare Parts....................................................................................... 46

Index ....................................................................................... 47

Table of Contents 15

Page 16: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

List of Figures 16

1-1 2690/2695 Separations Module with Column Bypass Module ...... 201-2 Column Bypass Module and Column Heater................................. 211-3 Column Bypass Module Valves ..................................................... 211-4 Column Bypass Module Valve Control Connections...................... 221-5 Column Bypass Module Electrical and Helium Connections ......... 231-6 Installing a Column in the Column Heater Module ........................ 241-7 Installing the Waste Vial ................................................................ 25

2-1 Column Bypass Module - Column Mode ....................................... 272-2 Column Bypass Module - Bypass (Inject) Mode............................ 292-3 Column Bypass Module - Bypass (Load) Mode ............................ 302-4 Column Bypass Module - Sample Loop ........................................ 32

3-1 Rheodyne 7725i Injector, Exploded View ...................................... 383-2 Rotor Seal Orientation (Viewed from Stator) ................................. 40

List of Figures

Page 17: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

List of Tables 17

3-1 Rheodyne 7725i Injector Troubleshooting................................ 353-2 Control Signal Troubleshooting................................................ 42

A-1 Operational Specifications ............................................................ 43A-2 Physical Specifications............................................................ 43A-3 Environmental Specifications .................................................. 44A-4 Power Specifications ............................................................... 44A-5 Communications Specifications for 6-Port Switching

Valve ...................................................................................... 44

B-1 Spare Parts ................................................................................... 45

List of Tables

Page 18: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

Preface

The Waters Column Bypass Module Installation and Maintenance Guide is intended for use by personnel who need to maintain and troubleshoot the Waters® Column Bypass Module. This guide provides information on using, troubleshooting, and maintaining the Waters Column Bypass Module as a component of a Waters LC System.

Organization

This guide contains the following:

Chapter 1 describes the function of the Column Bypass Module.

Chapter 2 describes how to operate the Column Bypass Module.

Chapter 3 provides troubleshooting and maintenance procedures.

Appendix A lists specifications.

Appendix B lists spare parts.

Related Documentation

Waters Licenses, Warranties, and Support: Provides software license and warranty information, describes training and extended support, and tells how Waters handles shipments, damages, claims, and returns.

Online Documentation

Empower LIMS Help: Describes how to use the Empower LIMS Interface to export results and import worklists.

Printed Documentation for Software Options

Waters Integrity System Getting Started Guide: Describes features of the Waters Integrity® System and provides step-by-step tutorials that guide a user through the use of the Empower Mass Spectrometry (MS) option.

Documentation on the Web

Related product information and documentation can be found on the World Wide Web. Our address is http://www.waters.com.

Related Adobe Acrobat Reader Documentation

For detailed information about using Adobe® Acrobat® Reader, see the Adobe Acrobat Reader Online Guide. This guide covers procedures such as viewing, navigating, and printing electronic documentation from Adobe Acrobat Reader.

18

Page 19: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

Printing This Electronic Document

Adobe Acrobat Reader lets you easily print pages, page ranges, or the entire document by selecting File > Print. For optimum print quantity, Waters recommends that you specify a PostScript® printer driver for your printer. Ideally, use a printer that supports 600 dpi print resolution.

Documentation Conventions

The following conventions can be used in this guide:

Convention Usage

Purple Purple text indicates user action such as keys to press, menu selec-tions, and commands. For example, “Click Next to go to the next page.”

Italic Italic indicates information that you supply such as variables. It also indicates emphasis and document titles. For example, “Replace file_name with the actual name of your file.”

Courier Courier indicates examples of source code and system output. For example, “The SVRMGR> prompt appears.”

Courier Bold Courier bold indicates characters that you type or keys you press in examples of source code. For example, “At the LSNRCTL> prompt, enter set password oracle to access Oracle.”

Underlined Blue Indicates hypertext cross-references to a specific chapter, section, subsection, or sidehead. Clicking this topic using the hand symbol brings you to this topic within the document. Right-clicking and selecting Go Back from the shortcut menu returns you to the origi-nating topic. For example, “Repairable Column Bypass Module control signal problems involve the cable wiring interface between the 2690/2695 Separations Module and the Column Bypass Module as described in Section 3.2, Control Signal Problems.”

Keys The word key refers to a computer key on the keypad or keyboard. Screen keys refer to the keys on the instrument located immedi-ately below the screen. For example, “The A/B screen key on the 2414 Detector displays the selected channel.”

… Three periods indicate that more of the same type of item can optionally follow. For example, “You can store filename1, filename2, … in each folder.”

> A right arrow between menu options indicates you should choose each option in sequence. For example, “Select File > Exit” means you should select File from the menu bar, then select Exit from the File menu.

19

Page 20: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

Notes

Notes call out information that is helpful to the operator. For example:

Note: Record your result before you proceed to the next step.

Attentions

Attentions provide information about preventing damage to the system or equipment. For example:

Cautions

Cautions provide information essential to the safety of the operator. For example:

STOPAttention: To avoid damaging the detector flow cell, do not touch the flow cell window.

Caution: To avoid burns, turn off the lamp at least 30 minutes before removing it for replacement or adjustment.

Caution: To avoid electrical shock and injury, turn off the detector and unplug the power cord before performing maintenance procedures.

Caution: To avoid chemical or electrical hazards, observe safe laboratory practices when operating the system.

20

Page 21: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

1

Chapter 1Introduction

In the current Waters® Integrity® configuration, the Column Bypass Module is contained within the 2690/2695 Column Heater module (Figure 1-1 and Figure 1-2). The Column Bypass Module allows you to easily bypass the column in your chromatographic system. Bypassing the chromatography column is necessary when you inject calibrants or analytes that could damage the column packing, or if you desire Mass Spectral data without the need to develop the chromatography.

The Column Bypass Module contains a 6-port, 2-way, pneumatically operated (switching) valve that directs the fluid path to either include or bypass the column. The Column Bypass Module also has a 6-port, 2-way (inject/load) mechanical valve with a needle port for loading calibrants into a sample loop and injecting them into the fluidic path (Figure 1-3).

The Column Bypass Module connects to the 2690/2695 power supply, sample fluidic lines, and electronic control signals.

Figure 1-1 2690/2695 Separations Module with Column Bypass Module

ColumnBypassModule

ColumnHeater

21

Page 22: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

1

Figure 1-2 Column Bypass Module and Column Heater

Figure 1-3 Column Bypass Module Valves

TP01449

COLUMN

LOAD

INJECT

BYPASS

Column Bypass Module

(Slides Out for Access)

Column Heater(Door Opens for Access)

TP01447

Front Panel

Inject/LoadValve

Switching Valve

Introduction 22

Page 23: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

1

1.1 Valve Control Connections

Figure 1-4 illustrates the valve control connections between the Column Bypass Module and the 2690/2695 Separations Module. These connectors are located externally on the rear panel of both devices.

Figure 1-4 Column Bypass Module Valve Control Connections

1.2 Internal Electrical and Helium Connections

Figure 1-5 illustrates the internal electrical and helium connections between the Column Bypass Module, and the 2690/2695 Separations Module.

6 Switch 3

1 Switch 1

5 Switch 2

2 Switch 1

3 Ground

4 Switch 2

7 Switch 3

8 Switch 4

9 Switch 4

10 Ground

11 Run Stopped

12 Run Stopped

Pin Signal

1 +Valve In

24 Chassis Ground

–Valve In

Black

White

Clear

White

Black

Clear

Legend

2690/2695 SeparationsModule Rear PanelConnector A

CBM Screw Terminal Connector

Valve Control Connections 23

Page 24: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

1

Figure 1-5 Column Bypass Module Electrical and Helium Connections

Note: You need to make a helium connection to the ThermaBeam Mass Detector, as described in the Waters ThermaBeam Mass Detector Installation and Maintenance Guide, Section 3.7, Connecting the Helium Supply.

1.3 Making Fluidic Connections

This section describes how to make fluidic connections between the Column Bypass Module and other components in your LC system.

Note: For details on making connections to the ports on the Column Bypass Module valves, refer to Section 2.1, Column Bypass Module Fluidics Overview.

Installing a Column

Connect either a 2-mm I.D. microbore or 3.9-mm I.D. microbore NovaPak® (standard lengths of 15-cm or 30-cm) column to the Column Holder in the Column Heater module.

Electrical Connector

Helium Connection

Front of Unit

Control Lines toHelium Regulator

Introduction 24

Page 25: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

1

To install and connect a column in the Column Heater module:

1. Hold the column so that its outlet is on top (arrows pointing up) and secure the column using the column holder (Figure 1-6).

2. Connect the top of the column to the unconnected end of the factory-installed 0.005-inch I.D. tubing that connects to the From Column port.

3. Connect the bottom of the column to the unconnected end of the factory-installed 0.005-inch I.D. tubing that connects to the To Column port.

Figure 1-6 Installing a Column in the Column Heater Module

Connecting to the LC System

You connect the Column Bypass Module between the sample management system of the 2690/2695 Separations Module and the PDA detector. Use precut lengths of tubing obtained from Waters, or use your own accurately cut lengths of the same diameter tubing.

Note: For details on making connections to the ports on the Column Bypass Module valves, refer to Section 2.1, Column Bypass Module Fluidics Overview.

TP01448

Column Holder

Column

Making Fluidic Connections 25

Page 26: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

1

To make the connections:

1. Connect the tubing from the sample management system red line outlet to the port labelled From Sample Management on the Column Bypass Module valves. (See Figure 2-1 through Figure 2-4.)

2. Install a 30-inch length of 0.005-inch I.D. tubing between the To Detector outlet port on the Column Bypass Module valves and the PDA detector Inlet port. (See Figure 2-1 through Figure 2-4.)

Installing the Waste Vial

Position the waste vial under the ends of the two waste tubes and move the vial up and into the clip on the front panel (Figure 1-7).

Note: The Inject/Load valve should always be in the LOAD position when not in use.

• When the inject/load valve is in the Load position, any excess calibrant loaded into the sample loop is expelled from one of these waste tubes into the waste vial.

• When the inject/load valve is in the Inject position, any calibrant previously injected into the sample loop enters the fluid path and goes to the detector. Any flushing agent (or mobile phase) injected into the needle port is expelled through the second waste tube into the waste vial.

Figure 1-7 Installing the Waste Vial

Waste Tubes

Waste Vial

Introduction 26

Page 27: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

2

Chapter 2Operation

The Column Bypass Module receives commands from the 2690/2695 Separations Module to direct the fluid path in either of the following ways: through the column (Column mode), or around the column (Bypass mode). Front-panel indicators on the Column Bypass Module indicate the currently selected mode.

• During system calibration, the 2690/2695 Separations Module configures the Column Bypass Module 6-port switching valve to direct the fluid path around the column and into the detector. Calibrants can be either manually or automatically injected and results can be observed.

• During normal operation, the 2690/2695 Separations Module configures the Column Bypass Module 6-port switching valve to direct the fluid path through the column.

When the Column Bypass Module is included as part of the Integrity system, Empower software automatically instructs the Column Bypass Module to bypass the column from the flow path. Once Empower software optimizes the interface temperature, it flushes the system and reestablishes the original conditions.

This chapter describes:

• Column Bypass Module fluidics overview

• Injecting calibrants with the sample management system

• Injecting calibrants with a manual injector

2.1 Column Bypass Module Fluidics Overview

Figure 2-1 through Figure 2-4 are fluidics diagrams of the Column Bypass Module showing the 6-port, 2-way, pneumatically operated switching valve, and the 6-port, 2-way, mechanically operated inject/load valve (see Figure 1-3). The operational modes and fluid paths that are set by these valves are:

• Column mode

• Bypass mode

• Sample loop and needle port

Column Mode

In the Column mode, the fluid path is from the 2690/2695 Separations Module, through the column, and out to the detector. Fluid enters switching valve port 2, leaves through port 3,

Column Bypass Module Fluidics Overview 27

Page 28: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

2

then passes through the column. From the column, the fluid enters port 6, exits through port 5, and goes to the detector (Figure 2-1).

Figure 2-1 Column Bypass Module - Column Mode

Bypass Mode

In the Bypass mode (Figure 2-2 and Figure 2-3), the fluid path is from the 2690/2695 Separations Module, through the switching valve, through the inject/load valve, back through the switching valve, and out to the detector, completely bypassing the column. There are two sub-modes of operation associated with the Bypass mode:

• Bypass (inject) mode

• Bypass (load) mode

Bypass (Inject) Mode

In the Bypass (Inject) mode, the inject/load valve is rotated to the Inject position and the switching valve is in the bypass position. Any calibrant in the sample loop goes into the fluid stream and to the detector.

Fluid enters switching valve port 2, exits through port 1, enters inject/load valve port 2, exits through port 1 and then enters the sample loop. The calibrant in the sample loop enters inject/load valve port 4, exits through port 3, enters switching valve port 4, exits through port 5, and enters the detector. Any calibrant previously injected into the sample loop goes into the fluid stream and to the detector. (See Figure 2-2.)

TP01307

1

6

54

3

2

Column

SwitchingValve

Outlet(To Detector)

Inlet From 2690/2695Separations

Module

Operation 28

Page 29: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

2

Bypass (Load) Mode

In the Bypass (Load) mode, the inject/load valve is rotated to the Load position and the switching valve is in the bypass position. At this point, the sample loop is isolated from the flow path and can be loaded with calibrant, if desired.

Fluid from the 2690/2695 Separations Module enters switching valve port 2, exits through port 1, enters inject/load valve port 2, exits through port 3, enters switching valve port 4, exits through port 5, and enters the detector. (See Figure 2-3.)

Column Bypass Module Fluidics Overview 29

Page 30: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

2

Figure 2-2 Column Bypass Module - Bypass (Inject) Mode

TP01307A

Inject/LoadValve

1 2

3

45

6

To WasteVial

Needle Port(Flush)

Sa

mpl

e L

oop

Column

Outlet(To Detector)

ValveSwitching

Inlet From 2690/2695Separations

Module

2 1

6

54

3

Legend

Flow PathNon-flow Path

Operation 30

Page 31: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

2

Figure 2-3 Column Bypass Module - Bypass (Load) Mode

TP01307b

Legend

Flow PathNon-flow Path

Column

SwitchingValve

Outlet(To Detector)

Inject/LoadValve

Sam

ple

Lo

op

To WasteVial

Needle Port(Load)

1 2

3

45

6

2 1

6

54

3

Inlet From 2690/2695Separations

Module

Column Bypass Module Fluidics Overview 31

Page 32: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

2

Sample Loop and Needle Port

The sample loop and needle port fluid path is shown in Figure 2-4. The standard size loop for this configuration is 20 µl.

Loading the Sample Loop

When the inject/load valve is in the Load position, the sample loop is isolated from the fluid stream. When you inject calibrant into needle port 4, you fill the sample loop. Excess calibrant leaves the sample loop, enters port 1, exits through vent port 6 (which is now connected to port 1), and enters the front-panel waste vial. The desired amount of calibrant remains in the sample loop.

Injecting the Sample

Any calibrant in the sample loop becomes part of the fluid stream when the switching valve is in the Bypass mode and the inject/load valve is in the Inject position. In this configuration, inject/load valve port 1 is isolated from port 6, thereby removing the injection loop from atmospheric pressure. High-pressure fluid flowing into port 2 flows out of port 1 and forces the calibrant out of the sample loop and into the fluid stream at port 4.

Flushing the Needle Port

In the Inject position, the needle port is connected to port 5. At this time, any flushing agent (or mobile phase) injected into the needle port is expelled directly out of port 5, through a waste tube, and into the waste vial.

Caution: To prevent liquid from squirting back at you when you flush the needle port, slowly inject 1-mL of mobile phase using a Rheodyne (P/N 7125-054) needle port cleaner attached to the syringe barrel.

Operation 32

Page 33: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

2

Figure 2-4 Column Bypass Module - Sample Loop

Changing the Sample Loop

When you perform the Operational Qualification for the Integrity System 996/2996 PDA Detector, you need to change the sample loop in the Inject/Loop Valve from the standard 20-µL sample loop to a 1-mL sample loop to perform the wavelength accuracy test.

The sample loop is connected between port 1 and port 4 on the Inject/Load Valve (see Figure 1-3 and Figure 2-4). For more information about the Operational Qualification for the 996/2996 PDA Detector, refer to the Waters HPLC Systems Qualification Workbook for the 996/2996 PDA Detector: Part IV, Operational Qualification.

2.2 Injecting Calibrants with the Sample Management System

The sample management system mode of operation configures the Column Bypass Module to bypass the column while the sample management system automates the delivery of one or more calibration standards directly into the fluid path.

STOPAttention: Ensure that the eluent and sample are properly filtered to prevent additional precipitation. Flush the flow passages and vent lines with water after use of salt solutions. Do not allow salt solutions to sit stagnant in the valve overnight or for long periods of time.

Sam

ple

Lo

op

Inject/LoadValve

To WasteVial

Needle Port(Load)

1 2

3

45

6

Injecting Calibrants with the Sample Management System 33

Page 34: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

2

Using a 2690 Separations Module and Empower Software

Use the following procedure to control the Column Bypass Module from a 2690/2695 Separations Module that is operating under Empower software control:

1. Rotate the lever on the Column Bypass Module front panel to the Load position. This position isolates any calibrant in the Column Bypass Module sample loop from the injection stream and reduces band spreading.

2. Place the desired standards and samples into the sample management system.

3. In the Method tab of the Empower software Project window, double-click the appropriate instrument method. The Instrument Method window appears.

4. Access the 2690/2695 Separations Module. When using Empower Pro, double-click the icon for your 2690/2695 Separations Module. When using Empower QuickStart, single-click the icon for your 2690/2695 Separations Module. The 2690/2695 Instrument dialog box appears.

5. Set Switch 4 in the 2690/2695 Instrument dialog box according to your Column Bypass Module control requirements:

• For immediate event control, click Switch 4.

• For time-initiated event control, click the Events icon and specify the time you want the switch S4 event turned on and off.

For details on setting up a 2690/2695 Separations Module, refer to the Waters 2695 Separations Module Operator’s Guide.

6. Save the changes to your instrument method, then exit the window.

7. When using Empower Pro, access the Run Samples window. When using Empower QuickStart, access Sample Queue view. Refer to the Empower Software Getting Started Guide.

8. To initiate a run using Empower Pro, access the Run Samples window. To initiate a run using Empower QuickStart, access the Sample Queue view.

Note: If, during data acquisition, you want to change the Column Bypass Module to Column mode, using Empower Pro, press the Stop Flow key in the Run Samples window. When using Empower QuickStart, press the Stop Flow key in the Control Panel view.

2.3 Injecting Calibrants Manually

The manual injection mode allows you to:

• Manually load a calibrant or sample.

• Inject the calibrant into the fluid path.

• Observe the results of that injection. During this mode, the column is bypassed from the fluid path.

Operation 34

Page 35: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

2

Procedure

To inject calibrants with a manual injector:

1. Rotate the lever on the Column Bypass Module front panel to the Load position. This position isolates any calibrant in the sample loop from the fluid path.

2. Inject the desired calibrant (or sample) into the needle port on the Column Bypass Module front panel using a calibrated syringe. To obtain best reproducibility for fixed loop injections, overfill the loop with at least two to three loop volumes of sample.

Note: Five to ten loop volumes will provide even more precision. An excess of sample is needed because mobile phase near the wall of the loop is displaced slowly.

3. Use the 2690/2695 Separations Module under Empower software control, to initiate a timed run, without using the sample management system. A run time of 1 to 2 minutes is recommended.

4. Rotate the lever to the Inject position and observe the results of the calibrant injection.

Note: During the timed run, you can load and inject other calibrants and observe their results.

Injecting Calibrants Manually 35

Page 36: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

3

Chapter 3Troubleshooting

This chapter presents step-by-step troubleshooting procedures. The information is divided into the following topics:

• Inject/Load valve problems

• Control signal problems

3.1 Inject/Load Valve Problems

Repairable inject/load valve problems involve the Rheodyne 7725i Injector that is used to make manual injections. Table 3-1 lists potential Rheodyne injector symptoms that may occur during operation, along with possible causes and corrective actions.

Table 3-1 Rheodyne 7725i Injector Troubleshooting

Symptom Possible Cause Corrective Action

Siphoning (appears as leakage)

Tubes from ports 5 and 6 do not have outlet ends at the same level as needle port

Place the outlet ends of both tubes at the same horizontal level as that of the needle port.

Rotor seal leakage (liquid dripping between stator and stator ring, or from needle port or vent tube)

Pressure-adjusting screw loose

Tighten pressure-adjusting screw. Refer to Section 3.1.1, Tightening the Pres-sure-Adjusting Screw.

Scratches on rotor seal Replace rotor seal. Refer to Section 3.1.2, Replacing the Rotor Seal.

Needle seal leakage Syringe needle O.D. smaller than needle seal, reducing sample loading accuracy

Try another syringe needle. Gently push on plastic needle guide to deform Teflon® sleeve to provide a better needle fit. See Section 3.1.3, Tightening the Needle Seal.

Inject/Load Valve Problems 36

Page 37: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

3

3.1.1 Tightening the Pressure-Adjusting Screw

Required Tools

To tighten the pressure-adjusting screw, you need a Phillips-head screwdriver and a 5/64-inch Allen wrench.

Adjustment Procedure

To tighten the pressure-adjusting screw:

1. Turn off the 2690/2695 Separations Module and disconnect the power cord.

2. Slide the Column Bypass Module compartment open to access the pressure-adjusting screw.

3. Locate the pressure-adjusting screw on the shaft of the injector (Figure 3-1).

4. Loosen the two knob set screws.

Increase in system backpressure or band broadening

Plugged valve passages or vent lines due to unfiltered eluent or sample, or buffer solutions crystallized in vent lines

Remove the stator and clean the passages with a wire of 0.005-inch maximum diameter or sonicate in water.If the valve passage remains plugged, disconnect the tubing at the vent valve outlet. If this clears the passage, replace the tubing. If the passage remains blocked, disconnect the compression screw at the vent valve inlet. If this reduces the pressure, replace the vent valve. Refer to Section 3.1.2, Replacing the Rotor Seal.

Gradient does not start Position-sensing switch is not sending the inject signal to 2690/2695 Separations Module

Disconnect the switch from the inject start cable at the connector under the valve. Gently insert a small flat-blade screwdriver into the connector to short-circuit the contacts. If the gradient starts flowing, check the switch as described in Section 3.1.4, Testing and Replacing the Position-Sensing Switch.

Caution: To avoid chemical and electrical hazards, always wear safety glasses, turn off the 2690/2695 Separations Module, and disconnect the power cord before you perform this procedure.

Table 3-1 Rheodyne 7725i Injector Troubleshooting (Continued)

Symptom Possible Cause Corrective Action

Troubleshooting 37

Page 38: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

3

5. Let the handle slide down the shaft so that its two tabs fit into the slots on the adjusting screw.

6. Use the knob as a wrench to tighten the pressure-adjusting screw approximately 1/20th of a turn. Use the 20 dial markings on the body and the painted spot on the screw to gauge how far to tighten the screw.

7. If the new setting fails to accomplish leak-free operation, repeat step 6 only once. Avoid excessive tightening, which increases rotor seal wear.

8. Finish by retightening the knob set screws onto the flats of the shaft.

Note: If this procedure fails to stop the leak, replace the rotor seal as described in Section 3.1.2, Replacing the Rotor Seal.

Inject/Load Valve Problems 38

Page 39: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

3

Figure 3-1 Rheodyne 7725i Injector, Exploded View

Handle

Knob

Set Screws (2)

Needle Guide

Pressure Adjusting Screw

Body

Spring Washers (4)

Needle Port Tube

Needle Seal

60° Stop Ring

Seal Pins (4)

Bearing Ring

Isolation Seal

Thrust Bearing

Rotor Pin

Position Sensing Switch

Rotor Seal

Stator Ring

Stator Locating Pin

Stator Face Assembly

Stator Locating Hole

Handle Screw

Teflon Sleeve in Rotor Seal

Stator

Stator Screws (3)

Troubleshooting 39

Page 40: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

3

3.1.2 Replacing the Rotor Seal

Premature rotor seal failure can be caused by any of the following problems:

• Abrasive particles in the sample or mobile phase that scratch the rotor seal surface.

• A needle tip of the wrong type that chips the ceramic stator face, causing deep scratching of the rotor seal surface.

• Buffer or salt crystallization caused by failure to flush the flow passages and needle port with water after the use of aqueous buffers or salt solutions. The crystals scratch the rotor seal surface, resulting in leakage.

Required Tools

To replace the rotor seal (see Figure 3-1), you need the following tools:

• 9/64-inch Allen wrench

• 5/16-inch open-end wrench

• Flat-blade screwdriver

• Phillips-head screwdriver

Replacement Procedure

To replace the rotor seal:

1. Turn off the 2690/2695 Separations Module and disconnect the power cord.

2. Slide the Column Bypass Module compartment open.

3. Leave the injector attached to the front panel of the Column Bypass Module, and leave the knob on.

4. Disconnect the tubing from the back of the valve.

5. Remove the three stator screws.

6. Pull axially to remove the following items:

• Stator and stator face assembly (remove together). If the stator face assembly is damaged, replace it.

• Stator ring.

7. Use a flat-blade screwdriver to pry the rotor seal off the four seal pins. Leave the isolation seal and bearing ring in place.

Caution: To avoid chemical and electrical hazards, always wear safety glasses, turn off the 2690/2695 Separations Module, and disconnect the power cord before you perform this procedure.

STOPAttention: To avoid damaging the injector after you have used buffer solutions, flush the injector with HPLC-grade water.

Inject/Load Valve Problems 40

Page 41: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

3

Troubleshooting 41

Reassembling the Injector

To reassemble the injector:

1. Loosen the pressure-adjusting screw one half-turn.

2. Note the original position of the two red dots.

3. Orient the rotor seal as shown in Figure 3-2, with rotor seal slots facing the stator.

Figure 3-2 Rotor Seal Orientation (Viewed from Stator)

4. Replace the stator ring so that the pin in the 60° stop ring enters the mating hole in the stator ring.

5. Install the stator face assembly on the stator. The three pins on the stator face assembly fit only one way into the mating holes in the stator.

6. Install the stator and stator face assembly on the injector so that the pin in the stator ring enters the mating hole in the stator.

7. Tighten each of the three stator screws a little at a time to keep the stator surface parallel to the stator ring surface until all parts are held firmly in place.

8. Retighten the pressure-adjusting screw until the red dots are aligned as you noted them in step 2.

9. Replace the knob and tighten the two set screws against the two flat areas on the shaft.

3.1.3 Tightening the Needle Seal

The needle seal, a Teflon sleeve in the rotor seal, may not seal correctly around a thin needle. A poor seal reduces accuracy in sample loading.

Removing the Needle

To deform the Teflon sleeve to make a good seal:

1. Remove the needle from the needle port.

2. Push gently on the plastic needle guide with the eraser end of a pencil. Do not squash the Teflon sleeve.

3. Repeat if necessary.

Rotor PinNotch

Page 42: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

3

Inject/Load Valve Problems 42

Checking the Needle Seal

To check for a proper seal around the needle:

1. Turn the solvent management system off.

2. Fill the syringe with water.

3. Place the injector in the Load position and slowly discharge the water in the syringe into the injector.

4. Note the low resistance to syringe discharge.

5. Repeat steps 1 and 2 with the injector handle halfway between the Load and Inject positions.

6. Verify that the resistance to discharging the syringe is now much greater than in step 3.

Note: The needle seal holds only a few psi (mPa) of pressure, and does not completely prevent syringe discharge when the injector handle is in the halfway position.

3.1.4 Testing and Replacing the Position-Sensing Switchs

The position-sensing switch is a magnetic reed switch actuated by a magnet in the shaft. The switch sends an Inject Start signal to the controller, signalling the controller to begin a run. If a run does not begin after an injection, the switch may be faulty. (The switch is rated for 100 V at 200 mA.)

Required Tools

To test or replace the position-sensing switch, you need the following tools:

• 9/64-inch Allen wrench

• 5/16-inch open-ended wrench

• Phillips-head screwdriver

Testing the Switch

To test the position-sensing switch:

1. Turn off the 2690/2695 Separations Module and disconnect the power cord.

STOPAttention: To avoid damaging the rotor seal, always use a needle of the correct type and gauge to load the injector.

Caution: To avoid chemical and electrical hazards, always wear safety glasses, turn off the Column Bypass Module, and disconnect the power cord before you perform this procedure.

Page 43: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

3

Troubleshooting 43

2. Slide the Column Bypass Module compartment open.

3. Disconnect the switch cable from the inject start cable.

4. Using an ohmmeter with needle probes, insert the probes into the switch cable connector.

With the injector handle in the Load position, the resistance reading should be infinite. With the injector handle in the Inject position, the resistance reading should be less than 1 ohm.

Replacing the Switch

Refer to Figure 3-1 to replace the position-sensing switch:

1. Remove the stator and ring and 60° stop ring.

2. Pull the position-sensing switch out of the stop ring.

3. Insert a new position-sensing switch.

4. Reassemble the injector. Refer to Section 3.1.2, Replacing the Rotor Seal.

3.2 Control Signal Problems

Repairable Column Bypass Module control signal problems involve the cable wiring interface between the 2690/2695 Separations Module and the Column Bypass Module. Other possible causes involve 2690/2695 Separations Module malfunctions.

Table 3-2 lists the possible causes of control signal problems that may occur during operation, along with recommended corrective actions.

Table 3-2 Control Signal Troubleshooting

Possible Cause Corrective Action

2690 Separations Module defective

Troubleshoot the 2690/2695 Separations Module according to the related documentation for your LC system.

Empower soft-ware problems

Verify correct operation of the Empower software in accordance with the related documentation for your LC system.

Poor contact between the Column Bypass Module screw terminal connector and screw terminal socket

1. Remove the screw terminal connector from the terminal strip socket.

2. Examine the contacts and clean as necessary.3. Tighten each of the six connector screws as necessary.4. Insert the connector back into the socket.

Interface cable wiring problems

Check all cable connections to the 2690/2695 Separations Module and secure each connector screw.

Page 44: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

A

Appendix ASpecifications

This appendix includes information on:

• Operational specifications (Table A-1)

• Physical specifications (Table A-2)

• Environmental specifications (Table A-3)

• Power requirements (Table A-4)

• 6-Port Switching Valve communication requirements (Table A-5)

Table A-1 Operational Specifications

Parameter Specification

Pressure, maximum 5000 psi (35 MPa)

Wetted surfaces 316 stainless steel, Vespel® alumina ceramic, PEEK(Note: PEEK is not compatible with nitric acid, sulfuric acid and tetrahydrofuran {THF})

Table A-2 Physical Specifications

Parameter Specification

Height (including column heater)

18.3 in (46.5 cm)

Depth 11.8 in (30 cm)

Width 5.8 in (14.7 cm)

Weight 20 lbs (9.1 kg)

Specifications 44

Page 45: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

A

Table A-3 Environmental Specifications

Parameter Specification

Ambient temperature, operating 39 to 104 °F (4 to 40 °C)

Ambient humidity, operating 10 to 90%, noncondensing

Ambient temperature, storage −40 to 158 °F (−40 to 70 °C)

Ambient humidity, storage 0 to 100%, noncondensing

Table A-4 Power Specifications

Parameter Specification

DC voltage 24 Vdc

Power requirements 10 VA

Helium Inlet 80 to 85 psi

Table A-5 Communications Specifications for 6-Port Switching Valve

Parameter Specification

Control Input(+Valve In, – Valve In)

Bypass: <+1.8 Vdc (+1.8 to –30 Vdc max.)Column: >+3.0 Vdc (+3.0 to +30 Vdc max.)

Inject Output Inject: closedLoad: open

Maximum voltage: 10 VdcMaximum current: 10 mA

Specifications 45

Page 46: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

Spare Parts 46

B

Appendix BSpare Parts

Table B-1 lists the recommended spare parts for installation by the user. Any parts not listed in the table may require installation by a Waters® service representative.

Table B-1 Spare Parts

Item Quantity Part Number

Cable, Event 1 WAT020321

Cable, Shielded Output 1 WAT057235

Syringe, Rheodyne, 25 µl 1 WAT033381

Assembly, Tubing, 30 cm –– WAT055474

Assembly, Filter Element 2 WAT088084

Needle Guide, 7725i 1 WAT055459

Sample Loop, 1 mL (for 996/2996 PDA Detector Oper-ational Qualification testing)

1 WAT096226

Seal, Rotor, 7725i, Vespel 1 WAT055946

Seal, Rotor, 7725i, PEEKa

a. PEEK is not compatible with nitric acid, sulfuric acid, and tetrahydrofuran (THF).

–– WAT055458

Seal, Rotor, 7730/7716, Vespel

1 WAT0055457

Seal, Rotor, 7730/7716, PEEK –– WAT055406

Stator, 7730/7716 –– WAT055455

Page 47: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

I NDEX

Index

AAutoinjection mode 32

CColumn installation 23Connections

LC System 24Conventions, documentation 18

DDocumentation

conventions 18related 17

FFluidic connections

waste vial 25Fluidics description

bypass mode 27column mode 26

IInstallation

column 23waste vial 25

LLC System connections 24

MManual injection mode 33

NNeedle seal tightening 41

OOperation

autoinjection 32manual injection 33overview 26

PPosition sensing switch testing 41Pressure adjusting screw 36

RRelated documentation 17Rheodyne 7725i injector

needle seal 41position sensing switch 41pressure adjusting screw 36rotor seal 39troubleshooting 35, 42

Rotor seal replacement 39

SSpare parts 45Specifications 43

Index 47

Page 48: Column Bypass Module - Waters Corporation€¦ · Advertencia: El usuario deber á saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas

INDEX

TTroubleshooting

control signals 42fluid path 35Rheodyne 7725i injector 35, 42

VValve control connections 22

WWaste vial installation 25

I

Index 48