Collaborative Studies: XRF Component of Round 1 Study Current Status of PQRI Round 2 Study Donna Seibert on behalf of the Technical/Analytical Challenges Sub-team to the Coalition for Rational Implementation November 9, 2016

Collaborative Studies: XRF Component of Round 1 Study Current … · 2016. 11. 2. · • WDXRF - ARL Thermo PERFORM’X • WDXRF – Rigaku Primus II • EDXRF – PANalytical Epsilon

  • Upload
    others

  • View
    9

  • Download
    0

Embed Size (px)

Citation preview

  • Collaborative Studies: XRF Component of Round 1 Study Current Status of PQRI Round 2 Study Donna Seibert on behalf of the Technical/Analytical Challenges Sub-team to the Coalition for Rational Implementation November 9, 2016

  • X-Ray Fluorescence Analyses of Elemental Impurities: Round 1 Experimental and Results Experimental slides courtesy of Thanh Nguyen

  • Subject Comments

    Objective Follow-on study to compare data generated with XRF to previously generated ICP-MS data for standardized evaluation samples.

    Participating Laboratories

    Four (4) voluntary XRF groups participated in this round robin testing.

    Sample Distribution Raw materials and unknown samples (tablet form) were distributed to all testing labs for analysis.

    Sample Preparation Consistency between standards and samples and between labs

    Analysis Standardized calibration ranges

    Objective and Outline

  • Preparation of Combined Standard Solutions

    Certified Metal ICP Standards 5-100 ppm Combined Standards

    (Solutions for Spiking) 5X Concentration from Target Level

  • Preparation of Solid Matrix for Standards

    A blend of the round robin tablet formulation was prepared with the above composition in order to match the native and enhanced tablet matrix

    Samples were prepared by grinding the native and enhanced tablets and preparing pellets for analysis

  • Spike and Mix Standards in Matrix

    1 mL volume of 5X standard solution was added to 5 g matrix in disposable acrylic mix vials

    Spiked matrix dried overnight in oven at 600C

    Clamp vial into mixer/mill and mix for ~5 min

    Spex CertiPrep 8000 Mixer/Mill

  • Preparation of Solid Standards

    Pellet Ready for Analysis

    Calibration/Analysis of the “ENTIRE” pellet

    Transfer “ENTIRE” sample to die assembly

    Pressed Pellet

  • Final Solid Standard Concentration Ranges

    Cd (ppm)

    Pb (ppm)

    As (ppm)

    Hg (ppm)

    Co (ppm)

    V (ppm)

    Mo (ppm)

    Se (ppm)

    0 0 0 0 0 0 0 0

    0.25 0.25 0.25 0.25 0.25 0.25 0.25 0.25

    0.50 0.50 0.50 0.50 0.50 0.50 0.50 0.50

    1.0 1.0 1.0 1.0 1.0 1.0 1.0 1.0

    2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5

    5.0 5.0 5.0 5.0 5.0 5.0 5.0 5.0

    Standards were prepared by serial dilution or through randomized concentrations

  • Calibration Curves

    Element Lowest R2 Highest R2

    Hg 0.9800 0.9990

    Cd 0.9869 0.9986

    Pb 0.9904 0.9998

    As 0.9918 1.0000

    Co 0.9815 0.9992

    V 0.9812 0.9999

    Mo 0.9958 0.9994

    Se 0.9881 0.9999

  • Instrumental Analysis Variations

    INSTRUMENTS XRF Instrument Models • WDXRF - ARL Thermo PERFORM’X • WDXRF – Rigaku Primus II • EDXRF – PANalytical Epsilon 3XLE Mixer/Mills • Spex 8000 • Retsch PM400

    Presses • Angstrom Briquet 445AE-1 • Atlas Autotouch Specac 25T • Atlas Autotouch Specac 40T

    VARIABLE PARAMETERS Analysis Atmosphere • Ambient air, He, Vacuum Analysis Time (per analysis) • 10 min – 45 min Size of Pellet • 32-40 mm Pressure of Pellet • 10-25 Ton

  • XRF Results by laboratory

    Native Tablets (ppm) Hg Cd Pb As Co V Mo Se

    Lab A (n=3) (0.1) ND ND ND (0.2) ND 0.3 (0.2) Lab B (n=3) ND ND ND ND ND (0.1) 0.5 ND Lab C (n=3) ND (0.1) ND (0.1) (0.1) ND (0.2) ND

    Lab D (n=10) (0.2) ND (0.1) ND ND ND (0.0) (0.1) XRF mean (n=12) (0.2) (0.1) (0.1) (0.1) (0.2) 0.3 0.3 (0.2)

    Enhanced Tablets (ppm) Hg Cd Pb As Co V Mo Se

    Lab A (n=3) (0.2) 0.4 2.5 1.7 0.9 1.6 1.8 0.5 Lab B (n=3) ND 0.3 1.4 1.4 0.4 1.7 2.1 0.3 Lab C (n=3) (0.1) 0.5 1.3 1.1 0.7 1.0 2.1 0.5

    Lab D (n=10) 0.6 0.4 2.2 1.6 0.6 1.2 1.6 0.5 XRF mean (n=12) 0.3 0.4 1.8 1.5 0.6 1.5 1.9 0.5

    Values in parentheses are below LOQ (0.25 ppm)

  • XRF Results by laboratory

    Native Tablets (ppm) Hg Cd Pb As Co V Mo Se

    Lab A (n=3) (0.1) ND ND ND (0.2) ND 0.3 (0.2) Lab B (n=3) ND ND ND ND ND (0.1) 0.5 ND Lab C (n=3) ND (0.1) ND (0.1) (0.1) ND (0.2) ND

    Lab D (n=10) (0.2) ND (0.1) ND ND ND (0.0) (0.1) XRF mean (n=12) (0.2) (0.1) (0.1) (0.1) (0.2) 0.3 0.3 (0.2)

    Enhanced Tablets (ppm) Hg Cd Pb As Co V Mo Se

    Lab A (n=3) (0.2) 0.4 2.5 1.7 0.9 1.6 1.8 0.5 Lab B (n=3) ND 0.3 1.4 1.4 0.4 1.7 2.1 0.3 Lab C (n=3) (0.1) 0.5 1.3 1.1 0.7 1.0 2.1 0.5

    Lab D (n=10) 0.6 0.4 2.2 1.6 0.6 1.2 1.6 0.5 XRF mean (n=12) 0.3 0.4 1.8 1.5 0.6 1.5 1.9 0.5

    Values in parentheses are below LOQ (0.25 ppm)

  • XRF Results Compared to ICP/MS Means

    Native Tablets (ppm) Hg Cd Pb As Co V Mo Se

    Lab A (n=3) (0.1) ND ND ND (0.2) ND 0.3 (0.2) Lab B (n=3) ND ND ND ND ND (0.1) 0.5 ND Lab C (n=3) ND (0.1) ND (0.1) (0.1) ND (0.2) ND

    Lab D (n=10) (0.2) ND (0.1) ND ND ND (0.0) (0.1) XRF mean (n=12) (0.2) (0.1) (0.1) (0.1) (0.2) 0.3 0.3 (0.2)

    ICP/MS mean 0.5 0.0 1.1 0.6 0.2 1.9 0.2 0.2

    Enhanced Tablets (ppm) Hg Cd Pb As Co V Mo Se

    Lab A (n=3) (0.2) 0.4 2.5 1.7 0.9 1.6 1.8 0.5 Lab B (n=3) ND 0.3 1.4 1.4 0.4 1.7 2.1 0.3 Lab C (n=3) (0.1) 0.5 1.3 1.1 0.7 1.0 2.1 0.5

    Lab D (n=10) 0.6 0.4 2.2 1.6 0.6 1.2 1.6 0.5 XRF mean (n=12) 0.3 0.4 1.8 1.5 0.6 1.5 1.9 0.5

    ICP/MS mean 0.4 0.4 3.1 1.5 0.7 2.7 1.5 0.6

    Values in parentheses are below LOQ (0.25 ppm)

  • Arsenic

    Tablet Type XRF ICP/MS Open circles denote outliers x indicates point below LOQ (XRF only) Error bars set at 1.5 x Interquartile Range

  • Cadmium

    Tablet Type XRF ICP/MS Open circles denote outliers x indicates point below LOQ (XRF only) Error bars set at 1.5 x Interquartile Range

  • Lead

    Tablet Type XRF ICP/MS Open circles denote outliers x indicates point below LOQ (XRF only) Error bars set at 1.5 x Interquartile Range

  • Mercury

    Tablet Type XRF ICP/MS Open circles denote outliers x indicates point below LOQ (XRF only) Error bars set at 1.5 x Interquartile Range

  • Cobalt

    Tablet Type XRF ICP/MS Open circles denote outliers x indicates point below LOQ (XRF only) Error bars set at 1.5 x Interquartile Range

  • Molybdenum

    Tablet Type XRF ICP/MS Open circles denote outliers x indicates point below LOQ (XRF only) Error bars set at 1.5 x Interquartile Range

  • Selenium

    Tablet Type XRF ICP/MS Open circles denote outliers x indicates point below LOQ (XRF only) Error bars set at 1.5 x Interquartile Range

  • Vanadium

    Tablet Type XRF ICP/MS Open circles denote outliers x indicates point below LOQ (XRF only) Error bars set at 1.5 x Interquartile Range

  • Conclusions

    • Results generally comparable to ICP-MS • Bias from background levels in materials used to

    make standards ▫ Apparent in Pb and V results ▫ Standard additions in calibration setup could

    correct for this bias • When sample is not limited, XRF offers quick

    and economical analysis ▫ Once calibration is established, only check

    standards need to be analyzed

  • Acknowledgements

    Company Contributors

    Rigaku Thanh Nyugen, Glenn Williams

    PANalytical Andrew Fussell

    Dow Corning Tom Case, Jason Sturm

    RTI Andrea McWilliams, Frank Weber, James Harrington

    Perrigo Josh Foote, Tanya Davis

  • PQRI Sponsored Study Concept Elemental Impurities in Pharmaceuticals Second Round Experimental Outline

    Slides courtesy of James Harrington

  • Project Objectives

    • Address some of the key technical challenges faced by industry in preparation for compliance to ICH Q3D and USP /

    • Provide a data-driven way to discuss technical aspects and expected variation of ICP-MS analysis of elemental impurities

    • More specific objectives: ▫ Inter-laboratory data comparison for standardized samples ▫ Inter-laboratory evaluation of effectiveness of microwave digestion ▫ Comparison of acid leach/extraction techniques to total metal extraction ▫ Examination of the correlation (good or bad) between the analysis of

    individual components (summation) vs. the formulated tablet analysis ▫ Comparison of ICP-MS and alternative techniques (ICP-OES and XRF)

    • PQRI Sponsorship allows wider participation

    • Study administrator—RTI International

  • Second Round Design Improvements & Best Practices

    General • Consistency among alternative techniques and digestion methods to

    ensure adequate data for comparison • ICP-OES (14) and XRF (6) analysis considered proactively • Raw materials distributed widely for summation approach comparison

    Uniform Sample Preparation • Specify parameters such as sample size, sampling technique, acid

    mixtures, and digestion temperature/pressure • Document type of digestion vessels and microwave model used

    Uniform Analysis • Define procedures around LOQs, calibration, and data reporting • Document interference management (reaction/collision gases,

    correction equations, etc.) and internal standards

  • Evaluation Samples & Analysis

    Second Round Improvements—

    Evaluation Samples

    • Liquid sample to assess instrumental variation

    • Evaluation samples with higher EI levels overall

    • Multiple powdered or tableted evaluation samples targeting different levels

    • EI source from pharma materials wherever possible

    Powdered or tableted material at three concentrations and liquid sample are aliquoted and shipped to each participating laboratory.

    Samples extracted in triplicate using the uniform method. Alternate extraction methods may be used if available in

    addition to the uniform method.

    Uniform method sample extracts/liquid

    sample must be analyzed by ICP-MS.

    Optional alternate method sample

    extracts are analyzed by ICP-MS.

    Replicate results and average/SD reported by extraction method and analysis method. Total samples for ICP-MS analysis are 12, alternative

    digestion and analysis techniques will be in addition to basic results.Minimum samples: 12, maximum: 36.

    Uniform method sample extracts, liquid may be analyzed by ICP-OES or other

    instrument.

    Samples extracted in triplicate using the uniform

    acid leaching method.

    Samples extracted by acid leaching method must be analyzed by

    ICP-MS.

    Results for replicates and average/SD are reported. Total samples for acid leach analysis

    are 12.

  • Formulation Challenges

    • Ideal solid formulation is tableted ▫ To preserve homogeneity

    • Pharmaceutical materials that contain significant, known levels have been elusive ▫ Few materials contain significant As & Hg

    • Combination of materials must have ▫ Favorable mixing & flow properties ▫ Compressibility

    • Current path for solid formulations ▫ Tablets similar to the first round tablets ▫ Include small amounts of XRF standards ▫ Three levels of elemental impurities

  • Second Round Recruiting

    • Distributed participant questionnaire for analytical laboratories in early August.

    • Response as of 11/01/16: ▫ 28 laboratories submitted forms + 1 verbal commitment

    Pharma manufacturers: 15 laboratories Contract/CRO: 9 laboratories Instrument manufacturers: 2 laboratories Government: 2 laboratories

    1 lab

    11 labs

    17 labs

  • Next steps:

    • Finalize formulations • Produce tablets and liquid sample

    ▫ Partnering with Liverpool John Moore’s University

    • Uniform method and acid leach method development ▫ To be developed with finalized tablets prior to distribution

    • Still seeking participant laboratories! ▫ Can accommodate up to 50 labs

  • Acknowledgements

    • James Harrington, RTI • Frank Weber, RTI • Phil Riby, Liverpool John Moore’s University • Dave Schoneker, Colorcon • PQRI • TAC Team members • All participating labs

  • Detail Slides

  • Second Round Participating Laboratories

    • J.M. Huber • Chemical Solutions • LGC, Ltd • MSD Oss • AstraZeneca (x2) • Colorcon • Dow Corning • Eurofins Frontier Global Sciences • ELS, Ltd • PANalytical • Pfizer (x2) • Evonik • Perrigo

    • Rigaku • Ineos Oxide • Perkin Elmer Health Sciences • Reading Scientific Service, Ltd • Solvias AG • Health and Safety Laboratory, UK • Procter and Gamble • SGS Life Science Services • RTI International • USP India (verbal) • Vertex Pharmaceuticals • AbbVie • GlaxoSmithKline • Intertek Pharmaceutical Services

  • Second Round Equipment

    • Available equipment ▫ ICP-MS: 23 laboratories ▫ ICP-OES: 14 laboratories ▫ XRF: 6 laboratories ▫ Microwave digestion: 21 laboratories ▫ Alternative digestion: 11 laboratories

    • Raw Materials analysis interest ▫ 20 laboratories

    • Acid Leaching interest ▫ 10 laboratories

    • HF digestion available ▫ 14 laboratories

    Collaborative Studies:�XRF Component of Round 1 Study �Current Status of PQRI Round 2 Study�X-Ray Fluorescence Analyses of Elemental Impurities:�Round 1 Experimental and Results�Slide Number 3Slide Number 4Slide Number 5Slide Number 6Slide Number 7Slide Number 8Slide Number 9Slide Number 10XRF Results by laboratoryXRF Results by laboratoryXRF Results Compared to ICP/MS MeansArsenicSlide Number 15Slide Number 16Slide Number 17Slide Number 18Slide Number 19Slide Number 20Slide Number 21ConclusionsAcknowledgements���PQRI Sponsored Study Concept��Elemental Impurities in Pharmaceuticals �Second Round Experimental OutlineProject ObjectivesSecond Round �Design Improvements & Best PracticesEvaluation Samples & AnalysisFormulation ChallengesSecond Round Recruiting Next steps:AcknowledgementsDetail SlidesSlide Number 33Second Round Equipment