Instructions for Use

Colibri. Universal battery power toolsystem for use in traumatology, spineand maxillofacial.

Page 2: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic
Page 3: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 1

Contents

Introduction

Using the Colibri

Care and Maintenance

Troubleshooting 36

Ordering Information 38

General Information 2

System Specifications 3

Electromagnetic Compatibility 4

Handpiece 8

Use 10

Attachments 14

General Information 27

Preparation for Cleaning and Disinfection 28

Cleaning and Disinfection 29

Lubrication 31

Inspection and Sterilization 33

Repairs and Technical Service 34

Disposal of Waste 35

Page 4: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Cd

Ni-Cd

2 Synthes Colibri Instructions for Use

Introduction

Intended UseThe Colibri is a battery-driven power tool for use in trauma-tology, maxillofacial and spine surgery. To ensure proper operation of the Colibri, use only Synthes original acces-sories.

Patient safetyThe Colibri may only be used on patients after the medicalpersonnel have read the instructions. Since it is impossible to fully exclude the possibility of technical problems, always ensure that an alternative system is ready when using theunit on patients.

Starting the systemNew power tools and their accessories must undergo the entire reprocessing process before being used. Completelyremove protective caps and films.

Caution – The batteries may never be sterilized. Explosion hazard!– To prevent injury, the power tool must be locked with the

safety system when coupling and removing attachmentsand tools, and before laying it down (see page 8).

– Properly operating cutting tools are essential to the successof an operation. For this reason, cutting tools must bechecked for wear and/or damage after each use and replaced if necessary. For each operation, we recommendusing a new Synthes original cutting tool.

MaintenanceThe life of the equipment can be substantially extended by following the service instructions.

For the devices to function properly, Synthes recommendsthat they be serviced annually by Synthes, or by exclusiveSynthes service centers.

Synthes is not responsible for damage arising from improperuse or technical service rendered by unauthorized parties.

Additional information on the use and preparation of theproducts can be obtained from a Synthes representative.

Particular attention should be given to the section “Care andMaintenance“ starting on page 27.

Liability of the userThe user of the product is liable for proper use of the equip-ment during surgery.

Explanation of symbols usedCaution. Read the provided instructions for use before operating the device.

Read the provided instructions for use before operat-ing the device.

This symbol indicates that the corresponding devicemay not be immersed in liquids.

The device is classified as type BF against electricalshock and leakage current. The device is suitable foruse on patients according to the standards defined by CSA 601.1, IEC 60601-1 and UL 2601.

This symbol indicates that the battery containsrechargeable NiCd (nickel cadmium) cells. The batteryis recyclable (observe the following instructions).

Since they contain cadmium, batteries may not be disposed with household garbage. Dispose or recyclethe batteries in accordance with local and nationalregulations.

The device meets the requirements of directive93/42/EEC for medical devices. It is authorizedby an independent notified body for which it bearsthe CE symbol.

General Information

0123

Page 5: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 3

Technical data

Continuously adjustable speed: 0 –3200 rpm

Weight 810 g

Operating voltage: 12 VDC 14.4 VDC

Battery capacity: 0.5 Ah

Battery type: NiCd

Cannulation: � 3.2 mm

Empty battery charging time: max. 60 min.

Degree of protection against electrical shock: BF

Protection against the penetration of water: IP X4

Noise level in operating position: Approx. 65 dB(A)

Duty Cycle: Intermittent operation

Xmin on Ymin off

Drilling and tapping threads: 8 45

Reaming: 8 45

Sawing: 8 45

Kirschner Wire setting: 8 45

Generally, electrical systems can heat up if in constant use.For this reason the handpiece and the attachment should beallowed to cool for at least 45 minutes following 8 minutesof constant use. If this is observed the system will be pre-vented from overheating and possibly harming the patient oruser. The user is responsible for the application and for turn-ing off the system as prescribed. If longer periods of constantuse are required, an additional handpiece and/or attachmentshould be used.

Subject to technical modifications.

The device meets the following standardsEN 60601-1 / IEC 601-1EN 60601-1-2 / DIN EN 60601-1-2, (VDE 0750 part 1–2) EN 55011 / DIN VDE 0875 part 11, class B, group 1 thresholdsMedical electrical devices

UL 2601-1

Only in conformance with UL 2601-1 and CAN/CSA C22.2No. 601.1 in reference to electrical shock, fire, mechanicaldanger.

Environmental conditions

Operation Transportation and storage

Temperature: 10–40 °C –20–50 °C

Relative humidity: 30–90% 10–90%

Atmospheric pressure: 500–1060 hPA 500–1060 hPA

May not be stored or operated in explosive atmospheres.

System Specifications

Page 6: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

4 Synthes Colibri Instructions for Use

Electromagnetic Compatibility

Accompanying Documents According toIEC 60601-1-2, 2004, Clause 6

Table 201: Emission

Guidance and manufacturer’s declaration – electromagnetic emissions

The Synthes Colibri System is intended for use in the electromagnetic environment specified below. The customer or the user of the Synthes Colibri System should assure that it is used in such an environment.

Emission test Compliance Electromagnetic environment – guidance

RF emissionsCISPR 11

Group 1 The Synthes Colibri System uses RF energy onlyfor its internal function. Therefore, its RF emis-sions are very low and are not likely to cause anyinterference in nearby electronic equipment.

RF emissionsCISPR 11

Class B The Synthes Colibri System is suitable for use in allestablishments, including domestic establishmentsand those directly connected to the public low-voltage power supply network that supplies build-ings used for domestic purposes.

Harmonic emissionsIEC 61000-3-2

Not applicable

Voltage fluctuations / flicker emissionsIEC 61000-3-3

Not applicable

Page 7: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 5

Table 202: Immunity (all devices)

Guidance and manufacturer’s declaration – electromagnetic immunity

The Synthes Colibri System is intended for use in the electromagnetic environment specified below. The customer or the user of the Synthes Colibri System should assure that it is used in such an environment.

Immunity test standard

IEC 60601test level

Compliancelevel

Electromagnetic environment – guidance

Electrostatic discharge (ESD)IEC 61000-4-2

±6 kV contact±8 kV air

±6 kV contact±8 kV air

Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.

Electrical fast transient / burstIEC 61000-4-4

±2 kV for power supply lines

±1 kV for signal lines

Not applicable Mains power quality should be that of a typicalcommercial or hospital environment.

SurgeIEC 61000-4-5

±1 kV line to line

±2 kV line to earth

Not applicable Mains power quality should be that of a typicalcommercial or hospital environment.

Voltage dips, short interrup-tions and voltage variations on power supply linesIEC 61000-4-11

<5% UT

(0.5 cycle)

40% UT (5 cycles)

70% UT

(25 cycles)

<5% UT for 5 s

Not applicable Mains power quality should be that of a typicalcommercial or hospital environment. If the user of the Synthes Colibri System requires continued operation during power mains interruptions, it isrecommended that the Synthes Colibri System ispowered from an UPS.

Note: UT is the AC. mains voltage prior to application of the test level.

Power frequency (50/60 Hz)magnetic fieldIEC 61000-4-8

3 A/m 30 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

Page 8: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

6 Synthes Colibri Instructions for Use

Electromagnetic Compatibility

Table 204: Immunity (not life-supporting devices)

Guidance and manufacturer’s declaration – electromagnetic immunity

The Synthes Colibri System is intended for use in the electromagnetic environment specified below. The customer or the user of the Synthes Colibri System should assure that it is used in such an environment.

Electromagnetic environment – guidance

Portable and mobile RF communications equipment should be used no closer to any part of the Synthes Colibri System, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Immunity teststandard

IEC 60601test level

Compliancelevel

Recommended separation distance

Conducted RFIEC 61000-4-6

3 Vrms150 kHz to 80 MHz

V1 = 10 Vrms150 kHz to 80 MHz

d = 0.35 ��P150 kHz to 80 MHz

Radiated RFIEC 61000-4-3

3 V/m80 MHz to 800 MHz

E1 = 10 V/m80 MHz to 800 MHz

d = 0.35 ��P80 MHz to 800 MHz

Radiated RFIEC 61000-4-3

3 V/m800 MHz to 2.5 GHz

E2 = 7 V/m800 MHz to 2.7 GHz

d = ��P800 MHz to 2.7 GHz

where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).

Field strengths from fixed RF transmitters as determined by an electromagnetic site survey, a

should be less than the compliance level in each frequency range. b

Interference may occur in the vicinity of equipment marked with the following symbol:

Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcastand TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site surveyshould be considered. If the measured field strength in the location in which the Synthes Colibri System is used exceeds the applicable RF compliance level above, theSynthes Colibri System should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orientingor relocating the Synthes Colibri System.

b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.

Page 9: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 7

Table 206: Recommended separation distances

Recommended separation distances between portable and mobile RF communications equipment and the Synthes Colibri System

The Synthes Colibri System is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Synthes Colibri System can help prevent electromagnetic interference by maintain-ing a minimum distance between portable and mobile RF communications equipment (transmitters) and the Synthes ColibriSystem as recommended below, according to the maximum output power of the communication equipment.

Rated maximum output power of transmitterW

Separation distance according to frequency of transmitterz

m

150 kHz to 80 MHzd = 0,35 ��P

80 MHz to 800 MHzd = 0,35 ��P

800 MHz to 2700 MHzd = ��P

0,01 3.5 cm 3.5 cm 10 cm

0,1 11 cm 11 cm 32 cm

1 35 cm 35 cm 1 m

10 1.1 m 1.1 m 3.5 m

100 3.5 m 3.5 m 10 m

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m)can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output powerrating of the transmitter in watts (W) according to the transmitter manufacturer.

Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Page 10: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

� �

Using the Colibri

� Attachment coupling � Trigger for speed regulation� Trigger for switching to reverse / oscillating drilling� Mode selector switch � Battery pack (battery casing with inserted battery) � Release buttons for attachment Release buttons for battery casing� Knob for the battery casing cover

Safety system The Colibri is equipped with a safety system that preventsthe machine from being unintentionally started. To lockand unlock the tool, turn the mode selector switch � to the appropriate setting on the front plate of the handpiece:“OFF”, “ ” or “ON” position.

Protective systemsThe Colibri is equipped with two safety systems:– A thermal overload safety system shuts off the tool if it be-

comes too hot during use. After it cools, the tool can beused again.

– An exhaustive discharge protection ensures that the bat-tery does not completely discharge. This protects the battery and extends its life.

Speed and rotational direction control

Mode selector switch in “ON” positionThe bottom trigger � gradually controls the forward speed(up to approx. 1200 rpm with the AO/ASIF quick coupling).When the bottom and top triggers � are pressed at thesame time, the tool immediately switches to reverse. Whenthe bottom trigger is released, the tool im mediately stops.

Mode selector switch in “ ” positionWhen the bottom and top triggers � are pressed at thesame time, the tool immediately switches to oscillating rota-tion. When the top trigger is released, the tool returns tonormal forward.

8 Synthes Colibri Instructions for Use

Handpiece

Page 11: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 9

Important notes– To prevent injury, the machine must be locked with the

mode selector switch � when coupling and removing attachments and tools, and before laying it down (see page 8).

– Never place the Colibri on a magnetic surface since themachine might start unintentionally.

– Pay particular attention to all instructions in the individualsections that are identified with “Caution”.

– Components that are no longer useful must be disposedof in accordance with local and national regulations.

Page 12: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Fig. 5

Fig. 3 Fig. 4

Fig. 2Fig. 1

10 Synthes Colibri Instructions for Use

Using the Colibri

Use

Before initial use, brand-new tools and accessories shouldundergo the entire reprocessing process, and the batteriesshould be charged. Completely remove protective capsand films.

Inserting the battery in the battery casingTo ensure sterility, the battery is inserted into the battery cas-ing by two people, one of whom is wearing sterile garments:

1. The person with the sterile garments holds the sterile battery casing and opens the cover (Fig. 1). Leave the lock-ing mechanism swung outward.

2. The person wearing the sterile garments places the sterilecover on the battery casing (Fig. 2) and checks if it isseated correctly. The sterile cover ensures that the unsterilebattery does not contact the outside of the sterile casing.

3. The person not wearing sterile garments carefully guidesthe unsterile battery through the sterile cover and pressesit completely into the battery casing to ensure a correctseat (Fig. 3 and 4). This person may not contact the out-side of the battery casing. The shape of the battery ensures that it is inserted with the correct pole alignment.

4. The person not wearing sterile garments grasps theflanges on the sterile cover and removes it from the bat-tery casing (Fig. 5).

5. The person wearing the sterile garments closes the casingcover from the outside without contacting the battery orthe inside of the casing.

Normally, one battery is sufficient for one surgery. For safety,two battery packs (battery casing with the battery) should bekept ready ensuring fast intraoperative changing of batteriesunder sterile conditions if necessary.

Page 13: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 11

Fig. 6

Fig. 7

Caution– If the unsterile battery contacts the outside of the casing,

the casing must be resterilised before being used in theOR.

– To close the casing cover, press it firmly to ensure that it iscompletely closed (Fig. 6 and 7) so that the locking mecha-nism properly engages.

– Sterilize the sterile cover after each use to ensure asepticconditions when inserting the unsterile battery into thesterile battery casing.

Inserting the battery pack into the power toolGuide the battery pack (battery casing with inserted battery)from below into the shaft of the handpiece (Fig. 8). Theshape of the battery casing prevents the battery from beinginserted incorrectly. Check if the battery pack is seated cor-rectly by gently pulling on it.

Removing the battery pack from the power toolSimultaneously press the release buttons for the battery casing with one hand (Fig. 9), and use the other hand to remove the battery pack from the handpiece.

Two batteries availableThe small 12 VDC battery pack of the Colibri (532.003 withBattery Casing 532.002) is powerful enough for many appli-cations and offers a better handling since it is smaller andlighter. The large 14.4 VDC battery pack (532.033 with Bat-tery Casing 532.032) has mainly been designed for heavierindications such as intramedullary reaming and drilling forDHS/DCS, but can be used with all Colibri attachments. TheColibri must not be used for e.g. acetabular reaming or hipand knee replacements.

Fig. 8 Fig. 9

Page 14: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

12 Synthes Colibri Instructions for Use

Using the Colibri

Charging, storing and using batteries

Charging– The batteries should always be charged before use.– Only use chargers recommended by Synthes to charge the

batteries. Using a charger that does not originate from Synthes can damage the battery.

– Charge the batteries within an ambient temperature rangeof 0 °C to 40 °C.

Storage– When the battery is not being used, store it in the appro-

priate Synthes charger. This will keep it from discharging,and the battery will be fully charged and ready to use.

– Before first use or after storing the battery outside of thecharger for more than one month, up to five charging/dis-charging cycles with the Synthes Universal Battery Charger(530.600, 530.601) or a refreshing-cycle with the SynthesUniversal Battery Charger II (05.001.204) may be necessaryfor the battery to be fully charged.

– Before first use or after storing the battery outside of thecharger for more than one month, up to five charging/discharging cycles may be necessary for the battery to befully charged.

– The charging station should always be turned on when abattery is in the charging base. This ensures availability andprevents discharge.

Use– Only insert the battery pack directly before using the

Colibri. This saves battery energy and prevents having tochange it during surgery.

– Never open the battery casing during surgery.– Only use batteries for the indicated purpose. – Do not transport or store the batteries together with mate-

rials that conduct electricity and can cause a short. This candamage the battery and generate heat which can causeburns.

– Never expose batteries to temperatures above 60 °C.– Do not apply force to the batteries, and do not let them

fall.– Never use damaged batteries; they can damage the power

tool.– Follow the additional information in the section “Care and

Maintenance“ starting on page 27 as well as the Instruc-tions for Use of the Synthes Universal Battery Charger II(036.000.500).

Page 15: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 13

Fig. 2 Fig. 1

Using the Colibri with mains currentThe Colibri can also be operated with mains current. Whenthe Colibri is operated with mains current it is classified astype B against electrical shock and leakage current. Whenoperated with mains current use the Colibri adapter(Art. 05.001.108) and one of the electrical consoles belong-ing to the Electric Pen Drive (Art. 05.001.000, 05.001.001or 05.001.002). The adapter can be inserted into the Colibrihandpiece and removed like a battery pack (see Fig. 1 and 2).Also follow the Electric Pen Drive’s Instructions for Use(036.000.800).

“Oscillating drilling” modeTo protect soft tissue when drilling and inserting Kirschnerwires, the Colibri has an electronically controlled oscillatingmode.

To preset the oscillating mode, switch the mode selectorswitch to “ “ position.

Pressing the bottom trigger by itself causes the tool to rotateclockwise as usual. Simultaneously pressing the top and bottom triggers causes the tool to immediately switch to os-cillating mode. The clamped tool oscillates clockwise/anti-clockwise. The speed can be changed by means of the bot-tom trigger. After the top trigger is released, the tool returnsto normal clockwise rotation.

Caution– Oscillating mode may only be used with the following

attachments:– AO/ASIF Quick Coupling (532.013, 05.001.250) – Chuck (05.001.252, 05.001.253) – Quick Coupling for Kirschner Wires (532.022)

– Do not use the “oscillating drilling” mode with the oscillat-ing saw attachments!

– You can only switch to reverse by turning the mode selector switch to “ON” position.

– The maximum cutting speed of a cutting tool is slightly lessin oscillation mode than in normal mode.

Fig. 1 Fig. 2

Page 16: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Using the Colibri

14 Synthes Colibri Instructions for Use

Attachments

Mounting the attachmentsInsert the attachment into the attachment coupling (Fig. 1).If the positioning pins do not lock into place right away, twistthe attachment a bit to the right or left until it locks into thecorrect position. Check if the attachment is seated correctlyby gently pulling on it.

Removing the attachments Press the attachment release buttons � (see Fig. on page 8) simultaneously, and remove the attachment from the coupling.

Caution – To prevent injury, the tool must be locked with the safety

system (see page 8) during each manipulation. – Only use original attachments and tools from Synthes.

Damage that arises from using attachments and tools byother manufacturers is not covered by the warranty.

Fig. 1

Attachment 12 VDC battery 14.4 VDC battery Color coding (532.003) (532.033)

532.011 Drill Attachment Mini Quick Approx. 3200 rpm Approx. 3840 rpm None532.012 Drill Attachment J-Latch

05.001.250 AO/ASIF Quick Coupling Approx. 1200 rpm Approx. 1440 rpm Blue05.001.252 Chuck 4.0 mm05.001.253 Chuck 6.5 mm

532.015 DHS Quick Coupling Approx. 320 rpm Approx. 384 rpm Red532.017 AO/ASIF Reamer Attachment532.018 Hudson Reamer Attachment532.019 Trinkle Reamer Attachment 532.020 Modified Trinkle Reamer Attachment05.001.251 Screw Attachment AO/ASIF05.001.254 Chuck 6.5 mm05.001.255 Chuck 6.5 mm, without reverse motion

532.022 Attachment for Kirschner wires Approx. 800 rpm Approx. 960 rpm None

A wide range of rotating attachments have color-codedrings, so that they can easily be identified. The table belowlists the speed for each attachment, the size of battery andthe color code:

Page 17: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 15

Quick Coupling for Kirschner Wires (532.022) Kirschner Wires of any length with a diameter of 0.6–3.2 mm can be used with the Quick Coupling forKirschner Wires.

1. Adjust the Kirschner Wire diameter according to the labelon the adjusting sleeve �. Slightly press the adjustingsleeve axially against the tool and rotate the sleeve.

2. Apply a slight amount of pressure to insert the KirschnerWire from the front or back into the cannulation �.The wire is held automatically.

3. Adjust the working length by pulling on the wire. 4. To affix the wire, pull the tension lever � against the tool

with your little finger and ring finger. Only pull the tensionlever against the tool as much as necessary. The clampingforce can be varied by pulling and releasing the clampinglever.

5. Insert the wire into the bone. Apply the clamping forceas long as the wire is advanced.

6. To adjust the grip on the wire, reduce the clamping forceand move the tool to the desired length. Reclamp the wireby pulling on the tension lever.

Page 18: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Using the Colibri

16 Synthes Colibri Instructions for Use

Mini Quick Coupling (532.011)J-Latch Coupling (532.012) The Mini Quick Coupling and the J-Latch Coupling are com-patible with all protection sleeves for the MicroDrive plus(512.101) and Mini Air Drill (512.100) systems by Synthes.

Mounting and removing the toolsTo connect the tool, pull the coupling sleeve back and thenintroduce the tool while turning slightly. To disconnect,push the coupling sleeve of the attachment back and removethe tool.

Quick Coupling for DHS/DCS Triple Reamers (532.015)* For DHS/DCS Triple Reamers; can also be used to open themedullary cavity with most of the Synthes nailing systems.

Mounting and removing the toolsTo connect the tool, push the coupling sleeve forward andthen introduce the tool while turning slightly. To disconnect,push the coupling sleeve of the attachment forward and remove the tool.

AO/ASIF Quick Coupling (05.001.250) For tools with an AO/ASIF coupling shaft

Mounting and removing the toolsInsert the tool into the attachment from the front applyingslight pressure and turning slightly. It is not necessary to operate the coupling sleeve of the attachment.

To disconnect, push the coupling sleeve of the attachmentback and remove the tool.

Page 19: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 17

ChucksThere are four Chucks available for the Colibri system. For a more detailed overview, please see the table on thebottom of this page.

Inserting cutting toolsOpen the jaws of the Chuck using the appropriate key or byhand. Insert the shaft of the tool into the open drill chuck,and close it by twisting the chuck. Make sure that the shaft lies central to the three jaws. Tighten the drill chuckwith the key. Make sure that the teeth of the key engagecorrectly in the toothed rim of the chuck.

Removing cutting toolsOpen the Chuck with the key and remove the tool.

Caution– Do not use the colibri for acetabular reaming.– Check the cutting tool for wear and/or damage after each

use, and replace if necessary.– To ensure a good fixation of the tools, make sure the

toothed rims on the drill chuck and key are not worn.– Use the Chuck with Reverse (05.001.255) only with tools

that are approved for such use. Otherwise the tool maybreak with consequential damage.

The following Chucks are available for the Colibri system:

Article number Clamping range Spare key Comments

05.001.252 0–4.0 mm 310.932 For drilling

05.001.253 0–6.5 mm 510.191 For drilling

05.001.254* 0–6.5 mm 510.191 For drilling and medullary reaming, with reverse

05.001.255* 0–6.5 mm 510.191 For drilling and medullary reaming, without reverse

*For these attachments, it is recommended to use the 14.4 VDC Colibri battery (532.033) with the corresponding battery casing (532.032) – called “Power Pack”.

Page 20: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

18 Synthes Colibri Instructions for Use

Quick Couplings for Medullary Reaming

AO/ASIF Quick Coupling (532.017)*Hudson Quick Coupling (532.018)*Trinkle Quick Coupling (532.019)*Trinkle Quick Coupling, modified (532.020)*The Quick Couplings for Medullary Reaming enable the useof flexible shafts with the appropriate coupling geometry. Reverse, which can damage the flexible shafts, is preventedby a special mechanical system.

* For these attachments, it is recommended to use the14.4 VDC Colibri battery (532.033) with the correspondingbattery casing (532.032) – called “Power Pack”.

Inserting cutting tools into the couplingManeuver the unlocking ring on the attachment backward,and insert the tool (such as a drill bit) while ro tating it slightlyuntil it locks into place. Release the ring. Check if the tool isseated correctly in the coupling by gently pulling on it.

Removing cutting toolsPush the unlocking ring on the attachment backward, andremove the tool.

Caution– Do not use Quick Couplings for Medullary Reaming for

acetabular reaming. They may only be used for medullaryreaming.

– Check the cutting tool for wear and/or damage after eachuse, and replace if necessary.

Using the Colibri

Page 21: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 19

Screw Attachment, with AO/ASIF Quick Coupling(05.001.251)

Mounting and removing the toolsInsert the tool into the attachment from the front applyingslight pressure and turning slightly. It is not necessary to operate the coupling sleeve of the attachment.

To disconnect, push the coupling sleeve of the attachmentback and remove the tool.

Caution– Care should be taken when inserting screws with the drive

unit.– Never fully insert screws with the drive unit. The last turns

or locking should always be done manually.– Always use an appropriate torque limiting attachment

when putting locking screws into a locking plate.– Theoretically it is also possible to use the AO/ASIF Quick

Coupling (05.001.250) to insert screws. The Screw Attachment (05.001.251) does, however, have a lowerrpm speed and a higher torque and is therefore more suitable. Screws with a large diameter may not be able tobe inserted with the AO/ASIF Quick Coupling as the torquemay not suffice.

– The attachment is also suitable for application at lowerrpm and/or higher torque. 

Page 22: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

20 Synthes Colibri Instructions for Use

Using the Colibri

Torque Limiter 1.5 Nm (511.773)Torque Limiter 0.8 Nm (511.776) Torque Limiter 0.4 Nm (511.777)

Coupling the Torque Limiter to the power toolTorque Limiters are connected to the Colibri using theAO/ASIF Quick Coupling (05.001.251).

Mounting and removing a screwdriver shaft:Insert the screwdriver shaft while rotating it slightly until itlocks into place. To remove it, pull back the unlocking ring,and pull out the screwdriver shaft.

Using the Torque Limiter:Pick up a screw from the corresponding LCP system (LockingCompression Plate) with the screwdriver shaft, and insert it inthe desired plate hole. To insert the screw, start the powertool slowly, increase the speed and then reduce it again be-fore the screw is fully tightened. The torque is automaticallylimited to 1.5, 0.8, or or 0.4 Nm. When this limit is reached,you will hear a distinct clicking. Stop the tool immediately,and pull the tool away from the screw.

Follow the surgical technique of the respective LCP system.

Caution:The Torque Limiter must be annually serviced and recali-brated by Synthes. Note the information on the test certificate in the packaging. The user is responsible for following the calibration schedule.

Page 23: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 21

(511.300)(532.031)(05.001.250)

Radiolucent Drive (511.300) The Radiolucent Drive can be used with the Colibri in combi-nation with the AO/ASIF Quick Coupling (05.001.250) andthe adapter for Radiolucent Drive (532.031).

Coupling the Radiolucent Drive to the power toolConnect the AO/ASIF Quick Coupling to the Colibri and theadapter to the Quick Coupling. Position the Radio lucentDrive as far as it will go over the Quick Coupling and theadapter, and rotate it into the desired working position. Sup-port the drive with your free hand.

Removing:Follow the same procedure in reverse.

Inserting the drill bitPull the ring on the attachment forward, and position thedrill bit into the coupling as far as it can go while rotating itslightly. Engage the ring on the attachment back to fix thedrill. Check if the drill bit is seated correctly by gently pullingon it.

Removing:Follow the same procedure in reverse.

Page 24: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

22 Synthes Colibri Instructions for Use

Using the Colibri

Using the Radiolucent Drive:Before positioning the Radiolucent Drive, align the image intensifier until the distal locking hole of the medullary nail isround and easily visible.

After the incision, position the Radiolucent Drive and centerthe drill bit tip over the locking hole. On the monitor of theimage intensifier, you can see both the drill bit and the targetrings of the drive.

Swing the drive up and center it precisely so that the drill bitappears as a round point and the locking hole is visiblearound it. The target rings also assist centering. The lockinghole can now be drilled directly.

Page 25: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 23

Caution– Grip the coupled Radiolucent Drive tightly when switching

on the power tool, particularly if the power tool is heldface down.

– Only special 3-flute spiral drill bits can be used. Your Synthes representative will be glad to provide you with additional information on which drill bits can be used.

– Handle the Radiolucent Drive with great care. Do not allowcontact between the drill bit and the medullary nail.

– Depending on the setting of the image intensifier, a zonemay appear in the rear of the Radiolucent Drive that is notradiolucent. However, this does not inhibit aiming andworking with the device.

– To protect the gears, the Radiolucent Drive is equippedwith a slip clutch that disengages in case of an overloadand emits an audible rattling.

– The following procedures can cause an overload:– Correcting the drilling angle when the cutting edges of

the drill bit are completely in the bone. – Running into the nail with the drill bit.

– Drilling can continue after making the following corrections:– Correcting the drilling angle: Remove the drill bit until

the flutes are visible and restart drilling. – Hitting a nail: Remove the drill bit until the flutes are

visible, and re-aim the drill bit or exchange the drill bit ifnecessary.

Page 26: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

��

Fig. 2

Fig. 1

24 Synthes Colibri Instructions for Use

Using the Colibri

Oscillating Saw Attachment (532.021)

Positioning the saw attachmentThe attachment can be locked in eight different positions(45° steps) when coupled: Lock the machine, shove the sliding sleeve � toward the saw blade coupling, and rotatethe attachment into the desired position (Fig. 1).

Caution: To prevent injury, always grip the saw attachmentwith the inserted saw blade from the direction of the machine.

Changing saw blades (Fig. 2). 1. Lock the machine. 2. Pull the locking knob � down and turn it counterclock-

wise. 3. Lift and remove the saw blade. 4. Use a slight amount of pressure to insert the new saw

blade, and turn it to the desired position. The desired posi-tions can be offset from each other at 45º angles.

5. Place your thumb on the saw blade coupling to hold thesaw blade, and turn the locking mechanism clockwise until the saw blade is fixed.

6. Unlock the power tool.

Caution: Saw blades labeled “Single Use” should not beused repeatedly because of cleaning problems.

Page 27: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 25

Fig. 3

Positioning the saw bladesThe saw blades can be adjusted in the desired position verti-cally and horizontally at an angle of 45º (See the prior sections: “Positioning the saw attachment“ and “Exchangingsaw blades“).

Using the oscillating saw attachment The saw blade must already be oscillating when the saw isapplied to the bone. Do not apply strong pressure to the sawblade; this will delay the cutting process, and the saw teethwill catch in the bone. The best sawing performance isachieved by moving the power tool slightly back and forth inthe plane of the saw blade so that the blade oscillates be-yond the bone on both sides. Very precise cuts can be madewhen the saw blade is guided steadily. Imprecise cuts arisedue to used blades, excess pressure, or jamming the sawblade.

Instructions for handling saw bladesSynthes recommends using a new blade for each operationto ensure that the saw blade is optimally sharpened andclean. The following risks are associated with used blades: – Necrosis caused by excessive heat build-up– Infection caused by residue– Extended cutting time from poor sawing performance

Page 28: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

70 mm

26 Synthes Colibri Instructions for Use

Using the Colibri

Oscillating Saw Attachment II (Crescentic Technique)(532.023)The Oscillating Saw Attachment II is essentially designed foruse with semicircular saw blades (for example 03.000.313S)guided by a 1.6 mm Kirschner Wire. It can also be used withsaw blades with a shaft extension (for example 03.000.340S)for reaching difficult-to-access sites (such as intraoral).

Inserting saw bladesPull the saw blade coupling toward the handpiece, and insertthe saw blade while rotating it slightly until it locks into thesaw attachment coupling. Release the saw blade coupling,and check if the saw blade is correctly fixed by gently pullingon the saw blade.

Removing the saw bladesPull the saw blade coupling toward the handpiece to releasethe saw blade.

Caution:– The saw attachments may only be used with the hand

piece in the “ON” mode. Never use the attachments in oscillation mode (“ “).

– The appropriate surgical technique (036.000.907) shouldbe observed to ensure safe and successful applicationof the Crescentic Technique.

Page 29: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 27

Care and Maintenance

Regular care and maintenance according to the following instructions (fulfilling DIN EN ISO 17664) can substantially increase the reliability and life of the system components.Synthes recommends annual servicing and inspection by theoriginal manufacturer or its exclusive service centers. The manufacturer assumes no warranty for damages arising fromimproper use or unauthorized servicing.

If not stated otherwise, the following reprocessing steps apply to the entire Synthes Colibri product line. Detergentsused on the products will contact the following materials:stain less steel, aluminum, plastic, and rubber seals.

Caution– Cannulations, unlocking sleeves and other narrow sites

require special attention during cleaning.– Synthes recommends cleaners with pH 7–9.5. The use of

cleaners with higher pH-values can – depending on thecleaner – cause dissolution of the surface of aluminum, titanium, its alloys, plastics or compound materials. At pH-values higher than 11 also the surfaces of stainless steelcan be affected.

– Lubricating regularly with Synthes special oil, especiallywhen mechanical cleaning is performed, will reduce wearand can substantially extend the service life of the product.

Restriction regarding reprocessingFrequent reprocessing does not have a great effect on theservice life of the unit and attachments. The expiration of theproduct’s service life is normally determined by wear anddamage from use and can be determined in a timely mannerthrough annual servicing. Synthes always recommends thatcutting tools such as saw blades be only used once to ensureoptimum patient treatment.

General Information

Page 30: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Preparation for Cleaning and Disinfection

In the operating roomRemove surface soiling with a disposable cloth or papertowel.

Storage and transportNo special requirements. Reprocess an instrument directly after it is used so that blood does not dry on it.

Preparation for cleaning– Reprocessing must be carried out immediately after each

use.– Before disinfection and cleaning, all attachments and

instruments must be removed from the machine.– Remove the battery pack from the power tool and take the

battery out of the housing. Immediately charge the batter-ies after each use in the Synthes battery charger. Wipe offthe batteries with a clean, soft and lint-free cloth.

– Power tool hand pieces and attachments should not be im-mersed in water or cleaning solution. Power equipmentmay not be cleaned ultrasonically.

– Do not use pointed objects for cleaning.

28 Synthes Colibri Instructions for Use

Care and Maintenance

Page 31: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 29

Fig. 2

Fig. 1

Manual cleaning and disinfection1. Wipe or spray the outside of the unit, battery housing and

accessories with disinfectant and let the disinfectant workaccording to the instructions of the disinfectant’s manu-facturer.

2. Remove all residues from the disinfected machine and ac-cessories under running water using a soft-bristled brushor a clean, soft and lint-free cloth.

3. Clean cannulations of the power tool and attachmentswith the provided cleaning brush. Fig. 1 shows an exampleof how to clean the cannulation in the power tool.

4. Coupling sleeves, sliding sleeves and similar difficult-to- access locations require special attention.

5. Rinse the unit and accessories with clear water and thendry with a clean, soft and lint-free cloth.

6. Open the saw blade coupling with the locking mechanismto clean the Oscillating Saw Attachment (532.021). Cleanall sides of the positioning pins and fixing cap with a soft-bristled brush under running water. Press the fixing capdownward with your thumb so that the O-ring below onthe locking mechanism is made visible. Make sure that theO-ring is not totally exposed (See Fig. 2) since it could oth-erwise be ejected and damaged while cleaning. Also cleanthe locking knob and its shaft up to the O-ring with a soft-bristled brush under running water.

7. Dry the power tool and accessories with a clean, soft andlint-free cloth.

Caution– The battery contacts must not come into contact with

water or solvents (short-circuit hazard).– The batteries may not be disinfected. – Reprocessing must be carried out immediately after each

use.– Let the disinfectant act for the period indicated in the

respective instructions for use.– Power tool hand pieces and attachments should not be im-

mersed in water or cleaning solution.– Do not use pointed objects for cleaning.– Make sure that the triggers, release buttons and lever of

the safety system is free of all residues and can be easilyactuated.

Cleaning and Disinfection

Page 32: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

30 Synthes Colibri Instructions for Use

Care and Maintenance

Automated cleaning and disinfection1. Manually clean critical device features such as cannula-

tions, unlocking sleeves and other difficult-to-access loca-tions.

2. Place all articles in the washing tray in a way that an effec-tive washing/disinfection can be performed. Ensure thatthe water can flow off any surface. Place the handpiece insuch a way that the attachment end (the tool couplingend) points diagonally downward. We recommend usingthe Holder for Washing Machine Baskets (No. 68.000.100)as shown in Fig. 1 on the right.

3. Prewash with cold water.4. Wash with detergent according to the instructions of the

detergent manufacturer.5. Rinse with tap water and then with cold demineralized

water.6. Thermally disinfect with demineralized water at 92°7. Dry with hot air at 90°.8. Upon removing the devices, inspect the cannulations, cou-

pling sleeves, etc. for visible soiling. If necessary, repeat thecycle or clean manually.

Caution– For automated cleaning, also consult the brochure

“Synthes Power Tools – Machine Washing Recommen -dations” (036.000.101).

– Power tools (especially seals and bearings) are exposed togreater stress in machine washing than when they arecleaned manually. For this reason, power tools must be lubricated (see page 31) and regularly sent to be serviced(at least once per year) when they are machine washed.

– Other cleaning and disinfecting methods are possible.Have an expert check the effectiveness under local condi-tions.

– Other detergents and disinfectants are possible. Clarify thematerial compatibility and effectiveness beforehand withthe detergent or disinfectant manufacturer.

Fig. 1

Page 33: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 31

2

2

2

2

2

2

2

2

2

Fig. 1

Fig. 3

Lubrication

The power tool and attachments should be regularly lubri-cated to ensure a long service life. It is recommended to oilthe accessible moving parts of the unit, battery casing andattachments with 2–3 drops of Synthes special oil (519.970)and distribute the oil by moving the components. Wipe offexcess oil with a cloth.

The poster „Care and Maintenance“ (Art. 038.000.017),has clear pictures showing which items need lubrication. Toorder this, please contact your local Synthes representative.

Lubricating the power tool (Fig. 1 and 2)– Lubricate the trigger shafts and then press the triggers

several times.– Lubricate the attachment release buttons and then press

the buttons several times.– Lubricate the battery casing release buttons from both the

outside and inside (see Fig. 2) and then press the buttonsseveral times.

– Lubricate the mode selector switch and then move it sev-eral times.

– Lubricate the attachment coupling.

Lubricating the battery casing (Fig. 3)– Place oil on the seal of the cover, and then evenly distrib-

ute the oil on the seal.– Lubricate the lock, hinge and knob, then actuate it several

times.

Fig. 2

Page 34: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Fig. 2

Fig. 1

32 Synthes Colibri Instructions for Use

Lubricating the attachments

Chuck (05.001.252–05.001.255)Lubricating jaws and toothed rim.Open and close the drill chuck several times.

Quick Coupling for Kirschner Wires (532.022) Lubricating tension lever and clamping mechanism.Hold the Quick Coupling up and add two drops of oil intothe attachement hole and on the holder of the lever (Fig. 2).Move the tension lever several times.

Mini Quick Coupling (532.011),J-Latch Coupling (532.012), AO/ASIF Quick Coupling (05.001.250/05.001.251), Quick Coupling for DHS/DCS Triple Reamers (532.015), Quick Coupling for Medullary Reaming(532.017/532.018/532.019/532.020)Lubricating unlocking ring.Move back and forth several times.

Oscillating Saw Attachment (532.021)Lubricating locking mechanism and saw blade coupling.Open and close the locking mechanism several times.

Oscillating Saw Attachment II (532.023)Lubricating unlocking sleeve, tool holder and attachmentcoupling.Move back and forth several times.

Caution– To ensure a long service life and reduce repairs, the power

tool and all attachments must be lubricated after each use.Exception: The Radiolucent Drive (511.300) may not be lubricated.

– The power tool and accessories may only be lubricatedwith Synthes special oil (519.970). The composition of thevapor-permeable and biocompatible oil is optimized for thespecific requirements of the power tool. Lubricants withother compositions can cause the power tool to jam andbe toxic.

– Only lubricate the power tool and attachments whenclean.

Care and Maintenance

Page 35: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 33

Inspection and function test– Visually inspect for damage and wear.– Check the handpiece controls for smooth operation and

function.– Check the coupling and unlocking sleeves of the hand-

piece and attachments for smooth operation, and checkfor function together with instruments such as cuttingtools.

Packaging– We recommend the Synthes Vario Case (68.001.250) or

sterilizing trays in combination with double-layer steriliza-tion wrap.

– Sterile containers can also be used. Please follow the manufacturer’s instructions.

Steam sterilizationFor steam sterilization, we recommend a fractionated pre -vacuum method with the following parameters: – Temperature: 134 °C– Plateau time: 5 minutes– After sterilization, the handpiece may only be used when it

has cooled to room temperature. Cooling may not be accelerated. When the machine has cooled and the batterypack is attached, the power tool is ready to use.

Inspection and Sterilization

Caution– The following maximum values may not be exceeded:

143°C over a maximum of 22 minutes. Higher values candamage the sterilized products.

– The batteries may not be sterilized: Explosion hazard!– The battery housing must be sterilized when open.– The machine and battery housing may not be joined when

they are sterilized.– Hot air sterilization, ethylene oxide and formaldehyde

sterilization are not suitable for reasons of material incom-patibility.

StorageNo special requirements (environmental conditions: see page 3). Follow the instructions of the sterilization wrapand sterile container manufacturers.

Contact ManufacturerPlease consult a local Synthes representative if you need anyadditional information.

Page 36: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Repairs and Technical Service

The tool should be sent to the Synthes office for repair if it isfaulty or malfunctions.

Faulty devices may not be used. If it is no longer possible orfeasible to repair the tool it should be disposed of, cf. thefollowing chapter “Disposal”.

Other than the above mentioned care and maintenance nofurther maintenance work can be carried out independentlyor by third parties.

Synthes recommends that the device and accessories such asattachments are regularly (once a year) serviced by the origi-nal manufacturer or an authorized site.

Caution: The manufacturer shall assume no responsibilityfor damage resulting from unauthorized maintenance.

34 Synthes Colibri Instructions for Use

Care and Maintenance

Page 37: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 35

Disposal

In most cases faulty tools can be repaired (cf. previous chap-ter “Repairs and Technical Service”).

This tool contains NiCd (nickel cadmium) batteriesthat should be disposed of in an environmentallyfriendly manner. The European Battery Directive2006/66/EC applies to this tool.

Caution: Risk of fire, explosion and burns. Do not disassem-ble, crush, heat above 100 °C, or incinerate the battery cells.

Please send tools that are no longer used to the local Synthesrepresentative. This ensures that they are disposed of in accordance with the national application of the respective di-rective. The tool may not be disposed of with householdwaste.

Contaminated components should be decontaminated be-fore return or disposal in order to prevent possible infectionof third parties.

Ni-Cd

Page 38: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

36 Synthes Colibri Instructions for Use

Troubleshooting

Problem Possible causes Remedy

Tool does not start. Battery is dead. Charge the battery or replace it with a charged battery.

The tool was not cooled off after sterilization.

Let the tool cool to room temperature.

Safety system is activated. Turn the mode selector switch to ”ON“ or “ ”.

No contact between the tool and battery pack.

Reinsert the battery pack or replace it.

Tool does not have enough power. Battery is dead. Charge the battery or replace it with a charged battery.

The machine stops suddenly. The tool is overheated (overload protection is activated).

Let the tool cool to room temperature.

Battery is dead. Charge the battery or replace it with a charged battery.

Attachments cannot be coupled to the tool.

The tool coupling is plugged from deposits.

Lock the machine. Remove particleswith blunt tweezers.

Tool is difficult to couple or cannot be coupled.

The attachment or tool is deformedfrom wear.

Replace the attachment or tool, or send it to a Synthes service center.

The Kirschner wire is inserted in thedrill and cannot be moved forward.

The Kirschner wire was inserted from the rear.

Lock the machine. Remove the attach-ment, hold the drive shaft opening down,and shake out the Kirschner wire.

Bone and tool heat up during surgery. The cutting tool is blunt. Replace the tool.

Oscillating saw attachment vibrates too much.

The saw blade locking mechanism is not tight.

Tighten the locking knob of the sawblade coupling.

The mode selector switch is set to “ “

Turn the selector switch to “ON”.

Page 39: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 37

If the recommended solutions do not work, send the powertool to your local Synthes service center.

For further technical questions or information on our serv-ices, please contact your Synthes representative.

Problem Possible causes Remedy

It is difficult to close the battery casing.

The battery casing seal has becomebrittle from repeated cleaning.

Lubricate the seal as described on page 31.

The battery casing knob is difficult to turn.

The locking mechanism needs to be lubricated.

Lubricate the locking mechanism as described on page 31.

The knob mechanism needs to be lubricated.

Lubricate the knob as described on page 31.

The triggers are difficult to move. The trigger shafts need to be lubricated. Lubricate the trigger shafts as described on page 31.

It is difficult to couple the battery casing to the machine.

The battery casing release buttons needto be lubricated.

Lubricate the release buttons as described on page 31.

Page 40: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Accessories

68.001.250 Vario Case for Nos. 532.001, 532.002 and 532.032, height 67 mm, without Lid, without Contents

689.507 Lid (stainless steel), size 1/1, for Vario Case

519.400 Cleaning Brush

519.970 Oil Dispenser with Synthes Special Oil

532.024 Cleaning Brush for Oscillating Saw Attachment II (532.023)

310.932 Spare Key for Drill Chuck, clamping range up to � 4.0 mm

510.191 Spare Key for Drill Chuck, clamping range up to � 6.5 mm

Saw blades (for 532.021)

Art. No. Art. No Usable Width Cut thicknessnon-sterile sterile length (mm) (mm) (mm)

532.041 532.041S 15 6 0.4

532.042 532.042S 15 10 0.4

532.043 532.043S 15 16 0.4

532.044 532.044S 18 4 0.4

532.045 532.045S 22 8 0.4

532.046 532.046S 22 12 0.4

532.047 532.047S 31 6 0.4

532.048 532.048S 31 10 0.4

532.061 532.061S 27 6 0.6

532.062 532.062S 27 10 0.6

532.063 532.063S 27 14 0.6

532.064 532.064S 50 10 0.6

532.065 532.065S 50 14 0.6

532.066 532.066S 50 20 0.6

532.067 532.067S 50 27 0.6

Art. No. Length Usable Width Cutsterile (mm) length (mm) (mm)

532.081S 47.1 25 12 90° cut, 2.0 mm shortening

532.082S 47.1 25 12 90° cut, 2.5 mm shortening

532.083S 47.1 25 12 90° cut, 3.0 mm shortening

532.084S 47.1 25 12 90° cut, 4.0 mm shortening

532.085S 47.1 25 12 90° cut, 5.0 mm shortening

532.091S 47.1 25 12 45° cut, 2.0 mm shortening

532.092S 47.1 25 12 45° cut, 2.5 mm shortening

532.093S 47.1 25 12 45° cut, 3.0 mm shortening

532.094S 47.1 25 12 45° cut, 4.0 mm shortening

532.095S 47.1 25 12 45° cut, 5.0 mm shortening

Drive unit

532.001 Colibri

Charger, Battery and Accessories for Battery

532.002 Battery Casing for Nos. 532.001 and 532.010, Standard

532.003 Battery for Colibri, 12 VDC, Standard

532.032 Battery Casing for 14.4 VDC battery, for Colibri

532.033 Battery for Colibri, 14.4 VDC

532.004 Sterile Cover for Colibri

05.001.204 Universal Battery Charger II

Attachments

532.011 Mini Quick Coupling, for Colibri

532.012 J-Latch Coupling for Colibri

05.001.250 AO/ASIF Quick Coupling for Nos. 532.001 and 532.010

05.001.251 Screw Attachment with AO/ASIF Quick Coupling, for Nos. 532.001 and 532.0101

05.001.252 Chuck (Drilling Speed), with Key, for Nos. 532.001 and 532.010, clamping range up to � 4.0 mm

05.001.253 Chuck (Drilling Speed), with Key, for Nos. 532.001 and 532.010, clamping range up to � 6.5 mm

05.001.254 Chuck (Reaming Speed), with Key, for Nos. 532.001 and 532.010, clamping range up to � 6.5 mm, with reverse motion*

05.001.255 Chuck (Reaming Speed), with Key, for Nos. 532.001 and 532.010, clamping range up to � 6.5 mm, without reverse motion*

532.015 Quick Coupling for DHS/DCS triple reamers, for Colibri*

532.017 AO/ASIF Quick Coupling for Medullary Reaming, for Colibri*

532.018 HUDSON Quick Coupling for Medullary Reaming, for Colibri*

532.019 Trinkle Quick Coupling for Medullary Reaming, for Colibri*

532.020 Trinkle Quick Coupling, modified, for Medullary Reaming, for Colibri*

532.021 Oscillating Saw Attachment, for Colibri

532.022 Quick Coupling for Kirschner Wires � 0.6 to 3.2 mm, for Colibri

532.023 Oscillating Saw Attachment II (Crescentic Technique), for Colibri

532.026 Large Oscillating Saw Attachment, for Colibri*

532.031 Adapter for Radiolucent Drive, for Colibri

511.300 Radiolucent Drive

511.773 Torque Limiter, 1.5 Nm

511.776 Torque Limiter, 0.8 Nm

511.777 Torque Limiter, 0.4 Nm

Power supply, power adapter and accessories

05.001.002 Basic Console, for Electric Pen Drive

05.001.108 Light Adapter, for Colibri

38 Synthes Colibri Instructions for Use

Ordering Information

* It is recommendable to use the more powerful battery for these attachments(532.033)

Page 41: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

Synthes 39

Saw blades (for 532.023)

Art. No Usable Blade width Cut thickness Diametersterile length (mm) (mm) (mm) (mm)

03.000.311S 26.5 13.3 0.60 18

03.000.312S 25 16.3 0.60 22

03.000.313S 30 17.9 0.60 18

03.000.316S 30 21.9 0.60 22

Art. No Usable Blade width Cut thickness Shaft length sterile length (mm) (mm) (mm) (mm)

03.000.340S 7 12.0 0.38 70

03.000.341S 12 11.5 0.38 70

03.000.342S 12 9.5 0.38 70

03.000.343S 12 4.5 0.38 70

Cutting toolsDetailed ordering information on the cuttings tools for theColibri machine system with original-size pictures can befound in the brochure on saw blades (Art. No. 036.000.172).

Page 42: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

40 Synthes Colibri Instructions for Use

Page 43: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic
Page 44: Colibri. Universal battery power tool system for use in ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · The Synthes Colibri System is intended for use in the electromagnetic

036.

000.

173

501

8226

4 S

E_06

6005

AD

40

1000

11

© 0

2/20

10 S

ynth

es, I

nc. o

r its

aff

iliat

es

All

right

s re

serv

ed

Synt

hes

and

DH

S/D

CS

are

trad

emar

ks o

f Sy

nthe

s, In

c. o

r its

aff

iliat

es

Ö036.000.173öADJä

All technique guides are available as PDF files at www.synthes.com/lit