Tino Noll, Manuel Fohler, Durgamadhab Mishra, Jan Geilhufe, Dieter Engel, Peter Bischoff and

Stefan Eisebitt

Institut für Optik und atomare Physik der TU-Berlin /

Helmholtz-Zentrum-Berlin für Materialien und Energie

New mechanics

for the

Coherent X-ray scattering endstation

at

BESSY

Page 2: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

Functionalities of the new endstation

Dedicated coherent scattering chamber for users

High flux beam line at UE 49 SGM

Reflection, Transmission

Microscopy, Scattering (static and dynamic (time, temperature, field etc))

Holography, Ptychography, Coherent Diffraction Imaging, Tomo-holography…

Retractable for using other endstations like solid state RICXS

Page 3: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

ID aperture

Wall

WAU1 Grating

Exit slit

WAU2 WAU3

Filters

Focus

Schematic view of the beam line (not up to scale)

Beamline

Page 4: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

Design

Liquid helium dewar

Support frame

CXS Chamber

Sample transfer

Focusing mirror

X-ray

Page 5: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

The experiment patform at 1.6 m

Page 6: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

Opening and retracting the chamber

Page 7: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

Scattering geometry

Incoming x-ray

Scattered x-ray

Vector magnet

CCD detector

Slider with beam stop

y

x

- z

Page 8: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,
Page 9: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

Inner mechanics without magnet

Sample holder

Sample manipulator

CCD

Page 10: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

Inner part without magnet

Page 11: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

Seite 11

The detector manipulator inside

the transport frame

Page 12: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

CCD movements

Page 13: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

Sample manipulator

sample holder

piezo linear motors

capacitive encoders

movable platform

Page 14: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

Sample manipulator

X

Z Y

light direction

Page 15: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

Sample manipulator

X

Z Y

pitch

roll

jaw

light direction

Page 16: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

Präsentationstitel Blindtext Lorem ipsum dolores | M. Mustermann | Anlass der Präsentation

Seite 16

The sample manipulator

Page 17: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

MEDSI14 Tino Noll et. al. Coherent X-ray scattering endstation at BESSY Page 17

moving the sample platform 50 microns

in beam direction and return to origin

Page 18: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

MEDSI14 Tino Noll et. al. Coherent X-ray scattering endstation at BESSY Page 18

moving the sample platform 1 micron

lateral to the beam

Page 19: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

MEDSI14 Tino Noll et. al. Coherent X-ray scattering endstation at BESSY Page 19

moving the platform 2mm

vertical up and down

Page 20: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

MEDSI14 Tino Noll et. al. Coherent X-ray scattering endstation at BESSY Page 20

Page 21: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

Sample transfer

coronal section of the endstation

sample

magazine

for 30

samples

transfer arm

with grabber

vector magnet

path to the

spectrometer

sample

6 DOF

platform for

sample

manipulation

Page 22: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

Sample transfer unit

Magazine

Grabber with three fingers

Page 23: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

Sample exchange inside the chamber

Page 24: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

MEDSI14 Tino Noll et. al. Coherent X-ray scattering endstation at BESSY Page 24

chuck mechanism

Page 25: Coherent X-ray Scattering endstation at BESSY...BESSY Functionalities of the new endstation Dedicated coherent scattering chamber for users High flux beam line at UE 49 SGM Reflection,

Thank you!