4
See discussions, stats, and author profiles for this publication at: http://www.researchgate.net/publication/4351922 Fundamental differences of the PD measurement according to IEC 60270 and in UHF range CONFERENCE PAPER · MAY 2008 DOI: 10.1109/CMD.2008.4580234 · Source: IEEE Xplore CITATIONS 5 DOWNLOADS 52 VIEWS 308 4 AUTHORS, INCLUDING: S. Coenen Alstom 16 PUBLICATIONS 87 CITATIONS SEE PROFILE Stefan Tenbohlen Universität Stuttgart 199 PUBLICATIONS 652 CITATIONS SEE PROFILE Available from: Stefan Tenbohlen Retrieved on: 30 July 2015

Coenen CMD2008 Fundamental Differences of IEC and UHF

  • Upload
    alps

  • View
    10

  • Download
    1

Embed Size (px)

Citation preview

See discussions, stats, and author profiles for this publication at: http://www.researchgate.net/publication/4351922Fundamental differences of the PDmeasurement according to IEC 60270 and inUHF rangeCONFERENCE PAPER MAY 2008DOI: 10.1109/CMD.2008.4580234 Source: IEEE XploreCITATIONS5DOWNLOADS52VIEWS3084 AUTHORS, INCLUDING:S. CoenenAlstom16 PUBLICATIONS 87 CITATIONS SEE PROFILEStefan TenbohlenUniversitt Stuttgart199 PUBLICATIONS 652 CITATIONS SEE PROFILEAvailable from: Stefan TenbohlenRetrieved on: 30 July 20151

AbstractPDmeasurementsintheUHFfrequencyrangeare animportanttoolfordiagnosticatGIS.Ageneralcorrelation betweenthemeasurementaccordingtoIECstandard60270and measurementsintheUHFrangeisnotpossible.Bytheuseofa "sensitivitycheck"thequalityoftheUHFmeasuringsystemis verifiedatGIS.Inthiscontributiontheusualprocedureis pointedoutandtheboundaryconditionsandsimplificationsare shown.Theknowledgeofthisisimportantfortheinterpretation of the result of the sensitivity check. IndexTermspartialdischarge,gasinsulatedswitchgear (GIS),UHFmeasurement,IECstandard,IEC60270,verification of the sensitivity I.INTRODUCTION Ooptimiseprocessesinutilitiesaconditionbased maintenance strategy is essential. The liberalization of the energymarketforcesallutilitiestoreducecosts.Considering themaintenancestrategyofGIS,asensitivePDdetectionis important.ToidentifythetypeofthePD,variousproven methods are possible, like the analysesof thephased resolved partialdischarge(PRPD)patternorothertechniques.Foran on-lineandon-site,suitablePDmeasurementisitpossibleto usethe method withintheUHFrange,becausethismethodis more resistive to disturbances than the measurement according to the IEC 60270 standard. The sensitivity of this measurement methodisprovedwithaprocedurecalledverificationofthe sensitivityorsensitivitycheck,recommendedbyCIGRE. By the interpretation of this sensitivity, it is necessary tokeep the boundary conditions in mind. II.SENSITIVITY CHECK FOR GIS BecausethemeasurementintheUHFrangecannotbe calibrated, a two-step procedure isused to guarantee a certain sensitivityoftheUHFmeasurementsystemon-site.Forthe practical use of the UHF method, it is sufficient to guarantee a certainsensitivityoftheUHFmeasurementsystem.The procedureisbasedontheinjectionofaveryfastbroadband impulse, so that the measurable spectrum in theUHF range is comparabletothespectrumofamobileparticlewitha discharge level of 5 pC, according to the IEC standard [1]. The

moving particle is used as the reverence level because it is the most common failure type [2]. The value of 5 pC according to theIECstandardischosenbecausethissensitivityisneeded for on-site measurements [2]. A sufficient sensitivity is verified if the injected impulse is measurable at the adjacent sensors. A.Two step procedure for the verification of the sensitivity of a UHF-PD- measurement system 1)Step 1: Measurement in the laboratory Inthefirststeptheneededvalueoftheimpulsemagnitude is determined to create comparable PD spectrums in the lab. It is necessary to put a moving particle nearby an installed UHF-sensorinsidetheGIS.Thevoltageoftheinnerconductoris increaseduntil the PDmeasurementsystemaccordingtoIEC 60270,measuresavalueof5 pC.IfthePDvalueisreached, the UHF spectrum of the moving particleshould be measured attheadjacentUHFsensor.Nowthehighvoltage,whichis responsibleforthePDofthemovingparticle,mustbe switched off. The value of the impulse, which is injected at the sensornearbytheparticle,ischangeduntilthemeasurable spectrum at the adjacent UHF sensor is in best conformity with thespectrumofthemovingparticle,whichwasmeasured before. Besidesavisualcorrelationofthespectrums,itispossible tousedifferentmathematicalanalysismethodstofindthe impulsewiththebestcorrelationofthespectruminthe frequency range between 300 MHz and 1,4 GHz [3]. 2)Step 2: On-site measurement Toverifythesensitivityofanon-siteinstalledUHF-PD-measurementsystemitisnecessarytoinjecttheimpulseas definedinstep1atasensor.Itisimportanttousethesame sensor,cableconnectionsandmeasurementsystemasinthe lab.Also the impulseshape and magnitudemustbethesame. Ifthespectrum,stimulatedbytheinjectedimpulse,is measurable at an adjacent UHF sensor,the systemwill ensure asufficientsensitivityfortheareabetweenthesesensors. Furthermore,themaximumdistancebetweenarealmoving particle and a sensor approximates half of the distance of these twosensors.ThismeansthatthePDsignalhasasmaller damping as the injected impulse of the sensitivity check. S.M. Hoek*, S. Coenen, M.Bornowski and S. Tenbohlen Universitt Stuttgart IEH, Pfaffenwaldring 47, 70569 Stuttgart, Germany *E-mail : [email protected] Fundamental Differences of the PD Measure-ment according to IEC 60270 and in UHF range T2 B.Boundary conditions of the sensitivity check Byusingthisverification,itisimportanttoconsiderthe boundaryconditionsofthemethod.Thesensitivitycheck simplifies the transfer functions and uses boundary conditions, whichmustbecarefullyobservedfortheinterpretationofthe results [3]. 1)Disregard of the reflexion and resonance The measurable UHF-PD signal depends on the position of the PD and the sensor, caused by the complexity of resonance insideGIS[Feger].IfthepositionandorientationofthePD sourceisknown,itwillbepossibletocalculatethetransfer functionwithacomplexhighfrequencymodel[4]fora coaxialarrangement.Itisdifficulttocalculatethetransfer functionformthesensorstoalllocationsofpotentialPD sourcesinrespectofalldetailsofthegeometryoftheGIS. The sufficient transfer function from sensor to sensor, which is securedbythesensitivitycheck,mustnotbetheworst-case transferfunctionbetweenthePDsourceintheGISandthe sensor.Butitisagoodapproximationforlargedistances, because the damping effects dominate in this casethe transfer function.Influenceisthinkableincaseofstrongreflexionor resonance for example at T-junction or outgoing line. 2)Definition of a sufficient sensitivity Thesensitivitycheckisbasedontheuseofveryfast impulses,whichareinjectedataUHF-sensor.TheUHF-spectrum, which is stimulated by the impulse, is comparable to the apparent discharge of 5 pC of a moving particle. It must be pointedout,thatthevalueof5 pCapparentdischargefora movingparticleoranotherfailuretypecanbeconsidered already as critically. [5] 3)The reference failure type Theprocedureisbasedonthecomparisonofthe measurement according to the IEC andthe UHF methodfor a movingparticle.Forotherfailuretypes,forexamplepoint-plain,floatingelectrodeorotherarrangements,theratio betweenthelevelofIECmeasurementandthemeasurement spectrum (or impulse energy) of the UHF method is different.Reid[6]showsinhisworkthattheratiobetweenthese measuringprinciplesdependsonthefailuretype.Theratio between the measurement values according to the IEC and the UHF method of a moving particle is for example advantageous incomparisontoapoint-plainarrangement[6].Themoving particle,asareferencefailuretype,representsnotthehardly detectablefailure,butisthemostcommonfailuretypeina GIS. This effect is also recognizable in measurement of PRPD pattern. The characteristic pattern for different types of failures iscomparablefortheIECmethodandalsoinUHFmethod, howevertheratioofthesignalintensitydiffersdependingon the failure type. Fig. 1.Measurement of moving particle accordance with IEC Fig.2.Measurementofthesamefailure(simultaneousmeasurement/ moving particle) with the UHF method Figure 1 shows a PRPD pattern of a moving particle with a dischargeofapproximately2 pCaccordingtotheIEC standard.InFigure2isasimultaneousmeasurementofthe same failures measured inthe UHF range. Thetypicalpattern is measurable and recognizable in both figures. Fig.3.Measurementofaprotrusionattheinnerconductoratlowlevel according IEC 3 Fig. 4.Measurement of the same failure (protrusion on an inner conductor at low level / simultaneous measurements to Fig. 3.) with the UHF method Atthepoint-plainarrangementinFigure3,anapparent discharge of 5 pC according to the IEC standard is measurable. But the signal of the PDsource is not measurablein the UHF range (Fig 4.). The source is measurable in the UHF range not untiltheapparentdischargebecomesabouttentimeshigher. (Fig 5. + 6.). Fig.5.Measurementofaprotrusiononaninnerconductorathigherlevel according to IEC Fig.6.Measurementofthesamefailure(portiononaninnerconductorat higher level / simultaneous measurements to IEC Fig.5) with the UHF method AdditionallythePRPDpatternisnotdirectlycomparablefor this PD source. Fig. 6. shows higher detectable UHF signals in thesecondhalf-wavethaninthefirstone,whileaccordingto IEC the resultsin bothhalf-wavesseemstobeofcomparable amplitude. This is caused by the impulse less persistent corona ofthepoint-plainarrangementathighfieldstrength.The bandwidthof suchacoronaissmallandnotdetectableinthe UHFrange.I.e.thatthepoint-plainarrangement,especially during the first half-wave, seems to be better for performinga worst-case estimation of the sensitivity of UHF measurements. III.CONCLUSION ThisarticleshowsthesensitivitycheckatGISas recommendedbyCIGRE.Fortheinterpretationofthis procedureattentionisnecessaryfordifferentsimplification and special definitions.The meaning of the reference failure type is important. The movingparticleisnottheminimumdetectablecaseinthe UHFrange.Otherfailuretypescanhaveamuchsmaller intensity than the moving particle at similar value, according to theIECstandard.Themovingparticleisthemostcommon failureandsoitispossibletomeasureinmajorityofcases verysensitive.Foundedbytheon-linetheapplicabilityand becauseoftherobustnessagainstdisturbances,the measurementwiththeUHFmethodisausefulalternative, compared to other measurement principles. Additionally in all cases,suchasensitivitycheckisapossibilitytoverifythe function of the sensor, the connection cable, theamplifier and the measurement system.Inconnectionwithproceedings[7]toestablishnew verification of the sensitivity procedure at other equipment like powertransformers,itisnecessarytodiscussthetest conditions for eachdevice under test, especiallyregarding the failure type.IV.REFERENCES [1]CIGRE,JointTaskForce15/33.03.05(1999)"PartialDischarge Detection System for GIS: Sensitivity Verification for the UHF Method and the Acoustic Method", ELECTRA Nr. 183, April 1999, S. 75-87 [2]CIGRE, Joint Working Group 33/23.12 (1998) "Insulation Coordination ofGIS:ReturnofExperienceonsitetestsanddiagnostictechniques", ELECTRA Nr. 176, Feb. 1998, S. 67-97 [3]R. Feger, "Sensoren und Verfahren der UHF-Messtechnik an GIS" Ph.D. Thesis, IEH, Universitt Stuttgart, 2001. [4]Judd, M. D.; Farish, O.; Hampton, B. F. (1996) The Excitation of UHF SignalsbyPartial DischargesinGIS, IEEETransactionsonDielectrics and Electrical Insulation, Vol. 3 No. 2, April 1996,[5]Colombo, E., Kultonowicz, W., Pigini, A.: "Sensitivity of Electrical and AcousticalMethodsforGISDiagnosticswithParticularReferenceto On-Site Testing", CIGRE-Symp., Berlin, 1993, paper 130-13. [6]A.J.Reid,M.D.Judd,B.G.Stewart,R.A.FouracreandS. Venkatesan,"CorrelationbetweenRFenergyandIEC60270apparent charge for selected partial discharge source geometries", ISH Ljubljana, Slovenia, August 27-31, 2007, Proceedings of the 15th ISH, T7-531 [7]S.Coenen,S.Tenbohlen,S.Markalous,T.Strehl"PerformanceCheck andSensitivityVerificationforUHFPDmeasurementsonpower transformers",ISHLjubljana,Slovenia,August27-31,2007, Proceedings of the 15th ISH, T7-100