Upload
dyna-mo
View
218
Download
0
Embed Size (px)
Citation preview
8/11/2019 COAS_P2_Ch16_it
1/12
238
Diagnostic methods in medicine
Chapter 16
As we saw in the last chapter, X-rays are used to
diagnose problems in a patient because they can
be generated outsidethe body and then directed
through the patients body to produce an image. An
alternative approach is to place a source of radiation
insidethe patient and use the radiation that emerges
from inside the body to produce an image.
An example of this approach is where a patient
undergoes a bone scan (to see whether there iscancerous tissue in their bones). The patient is
given an injection containing the radioactive
substance. They then have a few hours in which
to relax while the material circulates around their
body. Next, they are placed in a machine called
a gamma camera (Figure 16.1) which detects
-rays coming from inside their body. The result
is an image showing points in the body where the
radioisotope has accumulated (Figure 16.2). After
the scan, the patient must take care to ush the
toilet twice after use and to avoid kissing otherpeople this is because the radioisotope is still
Using radioisotopes
active for several hours and it is present in their
saliva and urine.
In this chapter, we will look in detail at some
ways in which radioisotopes (also known as
radionuclides) are used for diagnosis. Radioisotopes
are also used in treatment, where the radiation
they produce is used to destroy harmful tissue,
particularly cancerous tumours.
Figure 16.1 A female patient undergoing a bone
scan in a gamma camera. There are two detectors,
one above the patient and one below.
Figure 16.2 Bone scan images of a healthy patient.
The radionuclide has been taken up by the patients
bones; there is a lot of it in the bladder also.
e-Learning
Objectives
8/11/2019 COAS_P2_Ch16_it
2/12
Chapter 16: Diagnostic methods in medicine
Figure 16.3shows the system used for the
extraction (elution) of Tc-99m from the container
of Mo-99. A saline solution is passed through the
container, and this dissolves out the Tc-99m.
Radiopharmaceuticals
To ensure that the radioisotope reaches the
correct organ, it must be converted into aradiopharmaceutical. This means that it is
chemically combined with other elements to produce
a substance which will be taken up by the tissue of
interest. For example, for a bone scan, Tc-99 m is
combined with a phosphorus-containing compound
and the patient injected with a dose of activity about
600 MBq. This is taken up by bone tissue, particularly
cancerous tissue where there is a high rate of
metabolism as new cells are being formed.
Because radiopharmaceuticals are designed to
target particular organs or tissues in the body, they are
often described as tracers.
A summary of a few radioisotopes used by
hospitals and their diagnostic use is given in
Table 16.1.
Choosing a radionuclideAll radioactive substances decay, some more quickly
than others. There are several hundred different
radionuclides which are found in Nature or which
can be created in a nuclear reactor or a small linear
particle accelerator (linac). So which ones are
suitable for medical purposes?
The radionuclide is put into the patients body and
its radiation detected. This requires that the substance
chosen should be a gamma emitter. (An alpha or beta
source is not suitable because the body will absorb
the - or -particles. These types of radiation are
also extremely lethal inside the body because of their
strong ionising properties.)
The radionuclide should also have a short half-life.
There are two reasons for this.
It will give out its radiation quickly, so that only1 a small amount is needed to form an image in the
gamma camera.
Any radionuclide that remains in the patient2
will soon decay away, ensuring that they are not
exposed to hazardous levels of radiation.
The problem then is that, if a hospital buys a batch of
a short-lived radioisotope, it has bought something
that is rapidly decaying away. What can be done to
stop it decaying before the hospital has time to make
use of it?
One solution is illustrated by a radioisotope calledtechnetium-99m. This is an isotope of the element
technetium (Tc) with nucleon number 99. Tc-99m is
produced when molybdenum-99 undergoes decay.
This happens in two stages:
9942 Mo
9943 Tc
m+ 01 e + half-life 67 h
9943 Tc
m 9943 Tc + half-life 6 h
9943 Tc decays by emission half-life 2.110
5years
The m indicates that 9943 Tcmis metastable, that is,
it remains in an energetic state for some time before
decaying by emission. Each -ray photon has an
energy of 140 keV. How does this solve the hospitals
problem? The nuclear medicine department of the
hospital buys a supply of Mo-99, which is produced
in a nuclear reactor. The Mo-99 then produces
Tc-99m at a predictable rate, and this can then be
extracted for use with patients.
reduced
pressure
elution vial
sodiumpertechnetate
solution
terminalfilter
lead shielding
salinereservoir
one wayair filter
Mo adsorbed on toalumina column
99
Figure 16.3 A simplied diagram of a technetium-
99m generator. It is designed to minimise the risk
that the technician will be exposed to radiation. An
evacuated collection vial on the output side draws
saline from the reservoir through the column. Here
the saline dissolves the technetium to form a solution
of sodium pertechnetate.
8/11/2019 COAS_P2_Ch16_it
3/12
Chapter 16: Diagnostic methods in medicine
240
SAQ
1 Mo-99 is used to produce Tc-99m. The half-life of
molybdenum-99 is 67 h.
a Explain why a hospital will require supplies of
this substance to be delivered each week.b Explain why it would be inconvenient if the
half-life of Mo-99 was much longer than
67 h, and if it was much shorter
than 67 h.
The gamma cameraIt is over 50 years since the gamma camera
(Figure 16.1) was rst invented, and it is now the
major imaging device used in diagnostic nuclear
medicine. It detects -ray photons coming fromsources such as technetium-99m inside the patient.
Inside the camera is a single, very large crystal of
sodium iodide with about 0.5% of thallium iodide,
typically between 400 and 500 mm in diameter and
912 mm thick. This crystal is a scintillator; that is, a
gamma photon incident on this material may produce
a ash of visible light in the crystal.
Figure 16.4shows how the gamma camera
constructs an image of the patients insides from these
ashes of light.
The gamma photons pass upwards through the collimator. The collimator consists of ahoneycomb of cylindrical tubes in a lead plate.
The scintillator detects only photons travelling
along the axis of these tubes. It therefore cuts
out any -rays travelling at an angle to the
scintillator.
Radioisotope Uses
uorine-18 (189 F) bone imaging
technetium-99m (9943 Tcm) bone growth
blood circulation in lung, brain and liver
function of heart and liver
iodine-123 (12353 I) function of thyroid
function of kidney
xenon-133 (13354 Xe) function of lung
Table 16.1 Some radioisotopes used in hospitals and their uses.
computer
hexagonal array
photomultipliertubes
lightguide
scintillator
collimator
display
y
a
b
x
Figure 16.4 aThe structure of a gamma camera.
bThe arrangement of photomultipliers.
Answer
8/11/2019 COAS_P2_Ch16_it
4/12
Chapter 16: Diagnostic methods in medicine
avalanche of electrons. These eventually give rise
to an electrical pulse at the last electrode. Thus
a single electrical pulse is produced by a single
photon of light incident on the photocathode.
The electrical signals from the photomultipliersare processed electronically by a computer to
produce a high-quality image on a screen. The
output from each photomultiplier corresponds to a
single point or pixel on the screen.
Uses of the gamma camera
A gamma camera is used in a bone scan. This is an
example of a static study, in which a single image is
produced a suitable time after the injection of
the tracer.
Another use is in situations where it is desired
to see the progress of the tracer through the body.An example is a kidney scan (a renogram
see Figure 16.6). The patient is given a
radiopharmaceutical which will pass through their
system and be excreted by the kidneys. A series of
images of the kidneys are made over a period of time
to see the process of excretion as it happens. This is
an example of a dynamic study.
The beam of gamma photons then strikes thescintillator crystal, where each photon produces
a ash of light. About 10% of the incident -ray
photon energy is converted into visible light.
The light is detected by one or more of thephotomultiplier tubeswhich produce an
electrical pulse for each photon of light they
receive. The photomultipliers are arranged in a
hexagonal array over the surface of the crystal.
Figure 16.5shows a single photomultiplier tube.
The incident gamma photon strikes the scintillatorcrystal to produce photons of visible light. A
single light photon releases a single electron from
the photocathode by the process of photoelectric
effect. This electron is accelerated to the +100 V
electrode (dynode) and on impact releases two
or three secondary electrons. This process isrepeated at each electrode and soon there is an
scintillation crystal
+600 V
+800 V
+400 V
+200 V
0 V
incident gamma-ray photon
+500 V
+700 V
+300 V
+100 V
single electron
dynode
photocathode
Figure 16.5 Details of a photomultiplier tube.
Figure 16.6 A gamma camera image of a patients
kidneys. The image, called a scintigram, has been
coloured to show how the intensity of the -rays
varies. The kidney on the left is functioning normally
while the one on the right has very limited blood ow
through it.
8/11/2019 COAS_P2_Ch16_it
5/12
Chapter 16: Diagnostic methods in medicine
242
However, it is not the gamma photon emitted in
this decay that we are interested in. Rather, it is the
positron 0+1 e that is useful. Once emitted, it soon
collides with an electron and the two are annihilated.
Their mass is released as energy in the form of two
gamma photons. These are emitted at 180 to each
other (see Figure 16.7). In PET scanning, it is these
two gamma photons that are detected.
PET scanner
A PET scanner looks similar to a CAT scanner but,
of course, it is detecting -rays, not X-rays. The
principle is illustrated in Figure 16.8.
The patient is injected with a positron-emitting
radiopharmaceutical, in this case a form of glucose
(sugar) tagged with uorine-18. This tends to
accumulate in tissues with a high rate of respiration.
In this case, we imagine that the doctors are looking
at brain function, so the tracer will be taken up most
by active cells in the brain.
The patient is surrounded by a ring of gamma
detectors (similar to those in a gamma camera). These
detect pairs of -rays coming from inside the patient
and travelling in oppositedirections. The times at
which they arrive at the detectors are compared and
SAQ
2 A gamma camera can be adjusted by changing
the collimator.
a The collimator is changed to one with lead
tubes of larger diameter. Explain why this will
allow a shorter exposure time, but will give a
less well-dened image.
b If the collimator is changed to one with lead
tubes of longer length, more -rays will be
cut out as they cannot pass through to the
scintillator crystal. How will this affect the
exposure time and the
denition of the image?
Positron emission tomography
PET scanning is another technique which uses thefact that gamma rays can emerge from a source inside
the body. The name is an abbreviation of positron
emission tomography. As with CAT scanning,
the word tomographyimplies that images of slices
through the body can be obtained (using computer
manipulation of the image data).
The radiopharmaceuticals used in a PET scan
contain radioisotopes that emit positrons; that is, they
are +emitters. An example is uorine-18, which
decays like this:
189 F
188 O +
0+1 e + +
gamma
detectors
e+
e
Figure 16.8 A patient undergoing a PET scan. In
this case, -rays emitted by the tracer in the brain
are being detected.
Figure 16.7 Positronelectron annihilation. The
masses of the positron e+and electron eappear as
the energy of the two gamma photons.
Answer
8/11/2019 COAS_P2_Ch16_it
6/12
Chapter 16: Diagnostic methods in medicine
SAQ
3 Fluorine-18 is a +-emitting radioisotope
commonly used in PET scanning. Its half-life
is 110 minutes. Suggest why this
makes it a good choice for
this purpose.
4 Tomography is a type of imaging.
a Describe how tomography differs from
conventional imaging.
b Describe how a PET scan
forms a tomographic image.
Magnetic resonance imagingMagnetic resonance imaging, or MRI, is another
technique from nuclear medicine. However, it does
not rely on nuclides that are radioactive; rather, it
relies on the fact that some atomic nuclei behave like
tiny magnets in an external magnetic eld.
(MRI was originally known as nuclear magnetic
resonance imaging, but the word nuclear was
dropped because it was associated in patients minds
with bombs and power stations. To emphasise: MRI
does not involve radioactive decay, ssion or fusion.)As in CAT scanning, PET scanning and the gamma
camera, MRI scanning involves electromagnetic
radiation, in this case radio frequency (RF)
electromagnetic waves. The patient lies on a bed in
a strong magnetic eld (Figure 16.10), RF waves are
sent into their body, and the RF waves that emerge are
detected. From this, a picture of the patients insides
can be built up by computer. As we will see, MRI gives
rather different information from that obtained by the
other non-invasive techniques we have been looking at.
from this the position at which they were emitted
can be determined. Because gamma photons travel
at the speed of light, the time interval that must be
measured is very small.
Gradually, a three-dimensional image of the
distribution of radioactive tracer in the patient is built
up and, from this, any abnormal functioning can be
deduced. An image of a slice through the patient can
be viewed on a computer screen.
Uses of PET scanning
PET scanning is an important diagnostic tool, in
particular for showing up cancerous tissue. However,
it has also proved very useful in showing up aspects
of normal bodily functions, such as brain activity.
Figure 16.9shows scans of a persons brain when
they were reading aloud and then silently. Differentareas of the brain are active in these apparently
similar situations.
Figure 16.9 Articially coloured PET scans of a
human brain when the subject was reading aloud
(top) and silently (bottom). Reading aloud requires
extra areas of the brain to be active in order to
control the mouth and tongue and to listen to the
sounds produced.
Answer
Answer
Extension
8/11/2019 COAS_P2_Ch16_it
7/12
Chapter 16: Diagnostic methods in medicine
244
of hydrogen atoms that are studied, since hydrogen
atoms are present in all tissues. A hydrogen nucleus is
a proton, so we will consider protons from now on.
A proton has positive charge. Because it spins, it
behaves like a tiny magnet with N and S poles. Figure
16.11ashows a number of protons aligned randomly.
When a very strong external magnetic eld is
applied, the protons respond by lining up in the eld
(just as plotting compasses line up to show the direction
of a magnetic eld). Most line up with their N poles
facing the S pole of the external eld, a low energy
state; a few line up the other way round, which is an
unstable, higher energy state (Figure 16.11b).
A proton does not align itself directly along the
external eld. In practice, its magnetic axis rotates
around the direction of the external eld (Figure
16.12), just like the axis of a spinning top. Thisrotation or gyration action is known as precession.Principles of nuclear magnetic resonanceThe nuclei of certain atoms have a property called
spin, and this causes them to behave as tiny magnets
in a magnetic eld. In MRI, it is usually the nuclei
N
SN
S
N
S
NS
NS
S
N
N
S
S
N
N
S
S
N
N
S N
SS
N
S
N
S
N
S
N
S
N
S
N
S
N
S
N
S
N
N
S
S
N
N
SN
S
NS
NS
NS
NS
NS
NS
NS N
S
N S
N S
N
S
b
a
electromagnet
external
field
N
S
Figure 16.11 How protons behave in a strong
magnetic eld. aProtons are randomly directed when
there is no external magnetic eld. bBecause protons
are magnetic, a strong external magnetic eld causes
most of them to align themselves with the eld.
gravitational field
spin
spin
axis of
spin
path of
precession
path of
precession
magnetic field
axis of spin
Figure 16.12 A spinning top (left) rotates about its
axis; at the same time, its axis precesses about the
vertical, which is the direction of the gravitational
eld. In a similar way, a proton (right) spins and its
axis of rotation precesses about the direction of the
external magnetic eld.
Figure 16.10 A patient undergoing an MRI scan of
the brain. This is a form of tomography; the display
shows different slices through the patients brain.
8/11/2019 COAS_P2_Ch16_it
8/12
Chapter 16: Diagnostic methods in medicine
In Figure 16.13, you can see that the relaxation
of the protons follows an exponential decay pattern.
Curves like this are characterised by two relaxation
times:
T 1,the spinlattice relaxation time, where the
energy of the spinning nuclei is transferred to thesurrounding lattice of nearby atoms;
T 2,the spinspin relaxation time, where the energy
is transferred to other spinning nuclei.
These relaxation times depend on the environment
of the nuclei. For biological materials, it depends on
their water content:
Water and watery tissues (e.g. cerebrospinal uid)have relaxation times of several seconds.
Fatty tissues (e.g. white matter in the brain)have shorter relaxation times, several hundred
milliseconds.
Cancerous tissues have intermediate relaxationtimes.
This means that different tissues can be distinguished
by the different rates at which they release energy
after they have been forced to resonate. That is the
basis of medical applications of nuclear magnetic
resonance.
The angular frequency of precession is called
the Larmor frequency0, and depends on the
individual nucleus and the magnetic ux densityB0
of the magnetic eld:
0= B0
So, the stronger the external eld, the faster the
protons precess about it. The quantity is called the
gyromagnetic ratio for the nucleus in question and is
a measure of its magnetism. (Note that the Larmor
frequency is measured in radians per second. This
means that, strictly speaking, it is not a frequency.)
For protons, has the approximate value
2.68108rad s1T1. To determine the frequencyf0
of the precessing nuclei, we can use the equation
0= 2f0
Therefore:
f0=B0
2
In an MRI scanner, the external magnetic eld is
very strong, of the order of 1.5 T (thousands of times
the strength of the Earths eld). The precession
frequencyf0is
f0=2.68 1081.5
2= 6.4 107Hz = 64 MHz
This frequency lies in the radio frequency (RF) region
of the electromagnetic spectrum.
You should recall that resonancerequires a
system with a natural frequency of vibration; when
it is stimulated with energy of the same frequency, it
absorbs energy. In MRI, protons precessing about the
strong external eld are exposed to a burst or pulse
of RF waves whose frequency equals the frequency
of precession. Each proton absorbs a photon of RF
energy and ips up into the higher energy state; this is
nuclear magnetic resonance (Figure 16.13).
Now we come to the useful bit. The RF waves are
switched off and the protons gradually relax into their
lower energy state. As they do so, they release their
excess energy in the form of RF waves. These can be
detected, and the rate of relaxationtells us something
about the environment of the protons.
absorption
Time00
Energy
ofprotons
relaxation
Figure 16.13 In nuclear magnetic resonance, a
spinning nucleus is ipped into a higher energy state
when it absorbs a photon of RF energy; then it relaxes
back to its lower energy state.
8/11/2019 COAS_P2_Ch16_it
9/12
Chapter 16: Diagnostic methods in medicine
246
produce an additional external magnetic eld that
varies across the patients body. These coils are
arranged such that they alter the magnitude of the
magnetic ux density across the length, depth and
width of the patient. This ensures that the Larmor
frequency of the nuclei within the patient will be
slightly different for each part of the body. This
means that only a small volume of the body is at
exactly the right eld value for resonance and so
the computer can precisely locate the source of theRF signal within the patients body and construct
an image.
A computer that controls the gradient coils and RFpulses, and which stores and analyses the received
data, producing and displaying images.
Procedure
The patient lies on a bed which is moved into the bore
of the electromagnet. The central imaging section is
about 0.9 m long and 0.6 m in diameter. The magnetic
eld is very uniform, with variations smaller than 50
parts per million in its strength. The gradient eld
is superimposed on this xed eld. An RF pulse is
then transmitted into the body, causing protons to
ip (resonate). Then the receiving coils pick up the
relaxation signal and pass it to the computer.
SAQ
5 Protons precess at a frequency of 42.6 MHz in an
external eld of magnetic ux density 1.0 T.
a Determine the frequency at
which will they precess in a
eld of magnetic ux density 2.5 T.
b State the frequency of RF radiation that will
cause the protons to resonate
in this stronger magnetic eld.
6 Figure 16.14shows how the amplitude of RF
waves coming from watery tissue varies after
resonance. Copy the graph and add lines and
labels to show the graphs you would expect
to see for cancerous and
fatty tissues.
Figure 16.14 See SAQ 6.
MRI scanner
Figure 16.15shows the main components of MRI
scanner. The main features are:
A large superconducting magnet which producesthe external magnetic eld (up to 2.0 T) needed to
align the protons. Superconducting magnets are
cooled to 4.2 K (269 C) using liquid helium.
An RF coil that transmits RF pulses into the body.An RF coil that detects the signal emitted by therelaxing protons.
A set of gradient coils. (For clarity, only one pairof gradient coils in shown in Figure 16.14.) These
Time
watery tissue
00
Amplitude
large
external
magnet
RF receiving coil
RF transmitting coil
computer
longitudinal
gradient coil
Figure 16.15 The main components of an MRI
scanner.
Hint
Answer
Answer
8/11/2019 COAS_P2_Ch16_it
10/12
Chapter 16: Diagnostic methods in medicine
One disadvantage of MRI is that any metallic objects
in the patient, such as surgical pins, can become
heated. Also, heart pacemakers can be affected, so
patients with such items cannot undergo MRI scans.
Loose steel objects must not be left in the room as
these will be attracted to the magnet, and the room
must be shielded from external radio elds.
Figure 16.17shows how an MRI scan can becombined with a CAT scan to show detail of both
bone and soft tissue, allowing medical staff to see
how the two are related. Compare this with
Figure 16.16.
SAQ
7 An MRI scan might be considered a safer
procedure than a CAT scan.
a Explain why it might be considered to be safer.
b Why might a CAT scan be chosen in preference
to an MRI scan?
c Explain why MRI is described
as non-invasive.
The result is an image like the one shown in
Figure 16.16. This image has been coloured to show
up the different tissues, which are identied by their
different relaxation times.
Advantages and disadvantages of MRI
MRI has several advantages compared to other
scanning techniques:It does not use ionising radiation which causes ahazard to patients and staff.
There are no moving mechanisms, just changingcurrents and magnetic elds.
The patient feels nothing during a scan (althoughthe gradient coils are noisy as they are switched),
and there are no after-effects.
MRI gives better soft-tissue contrast than a CATscan, although it does not show bone as clearly.
Computer images can be generated showing any
section through the volume scanned, or as a three-dimensional image.
Figure 16.16 MRI scan through a healthy human
head. Different tissues, identied by their differentrelaxation times, are coloured differently.
Figure 16.17 A combined CAT scan and MRI scan,
showing how the tissues revealed by MRI relate tothe bone structure shown by X-rays.
Answer
8/11/2019 COAS_P2_Ch16_it
11/12
Chapter 16: Diagnostic methods in medicine
248
Summary
Medical tracers such as technetium-99m are used to diagnose the function of organs.A gamma camera detects gamma radiation coming from the medical tracer in the body and deducesits position.
The main components of a gamma camera are: collimator, scintillator crystal, photomultiplier tubesand computer.
In PET scanning, a +emitter is used as a tracer; -rays produced as the positrons annihilate with electronsare used to determine the position of the tracer.
In MRI scanning, spinning, precessing protons are forced to resonate using radio frequency pulses. RFradiation from relaxing protons is used to obtain diagnostic information about internal organs, particularly
soft tissues.
The main components of an MRI scanner are: superconducting magnet, RF transmitter coil, RF receivercoil, set of gradient coils and computer.
Questions
a1 State one application of technetium-99m as a tracer. [1]
Technetium-99m nuclei are produced when radioactive nuclei of molybdenum-99b
emit -particles.
Complete the nuclear reactions below:i
9942 Mo99
? Tcm+ 01e + ?
9943 Tcm 9943 Tc + ? [2]
Suggest why technetium-99m is suitable as a tracer. [2]ii
Molybdenum-99 has a half-life of 67 h. The initial activity of a sample ofiii
molybdenum is 600 MBq. For this sample of molybdenum, calculate:
the decay constant in s1 1 [1]
the initial mass in grams of the sample [3]2
the activity of the sample after 30 h. [2]3
[Total 11]
2 This question is about magnetic resonance imaging (MRI).
Explain what is meant by:a
the Larmor frequencyi
0 of nuclei [2] relaxation time of nuclei. [3]ii
Describeb two advantages of MRI compared with a CAT scan. [2]
Outline some of the main components of an MRI scanner. [5]c
[Total 12]
continued
Glossary
Hint
Hint
Answer
Answer
8/11/2019 COAS_P2_Ch16_it
12/12
Chapter 16: Diagnostic methods in medicine
a3 The diagram below shows the key components of a gamma camera.
A-ray
photonsB C
photomultiplier
tubes
Name each component A, B and C and state its function. [6]
A photomultiplier tube has 10 electrodes known as dynodes. An electron emittedb
from the photocathode is accelerated towards the rst dynode. On impact, it
produces, on average, three secondary electrons. These are accelerated towards
the second dynode and the whole process is repeated. An electrical pulse lasting
for 2.0 ns at the tenth dynode is produced. Calculate:
the total number of electrons at the tenth dynode as a result of a singlei
electron impacting the rst dynode [2]
the average current from the last dynode. [3]ii [Total 11]
Hint
Hint
Answer