Upload
vucong
View
214
Download
0
Embed Size (px)
Citation preview
CLIOCLIO application: application:Room and Building AcousticsRoom and Building Acoustics
AudiomaticaAudiomatica
2
Room and Building Acoustics
● Normative references in Room and Building acoustics● Applications:
● Measurement of impulse response● Evaluation of acoustical parameters● Speech transmission index● Wavelet analysis to room acoustics● ETF (Energy-Time-Frequency) to room acoustics● Measuring airborne sound insulation● Rating R' airborne sound insulation
● Normative related to measurement equipment
3
Room and Building Acoustics● Normative references:
● ISO 18233:2006. Acoustics — Application of new measurement methods in building and room acoustics
● ISO 3382. Acoustics — Measurement of room acoustic parametersPart 1:2009. Performance spacesPart 2:2008. Reverberation time in ordinary rooms
● IEC 60268. Sound system equipmentPart 16:2003. Objective rating of speech intelligibility by speech transmission index
● ISO 140 series. Acoustics — Measurement of sound isulation in buildings and of building elements
● ISO 717 series. Acoustics — Rating of sound isulation in buildings and of building elements
4
Room Acoustics● The acoustical properties of a room can be described
by means of a set of Acoustical Parameters derived from the measurement of its impulse response
● These parameters are defined in the ISO 3382 standard:● Reverberation time (EDT, RT20, RT30, RTU)● Early/late energy ratios (C50, C80, D, ts)● Sound pressure levels and Strength (G)● Background noise levels
● An important factor describing the quality related to human voice is the speech transmission index STI
5
Room Impulse Response
● The room can be seen as a linear system where the Impulse Response between a sound source and a microphone can be measured using different tecniques:
● Maximum Length Sequences MLS (ISO3382 and ISO18233)
● Sine Sweeps SS (ISO3382 and ISO18233)● Impulse source (ISO3382)
7
MLS and LogChirp
● The room impulse response can be measured using a loudspeaker as source and a MLS or Sine Sweep technique
sourcemic
CLIOfw
A IN B A OUT B
room
8
Measuring Impulse Responses with CLIO 10● MLS & LogChirp Analysis menu:
● size up to 512k, up to 10.8 seconds at 48 kHz● MLS:
– White stimulus– STI calculation
9
Measuring Impulse Responses with CLIO 10● MLS & LogChirp Analysis menu:
● size up to 512k, up to 10.8 seconds at 48 kHz● LogChirp
– Pink stimulus– Distortion separation
11
Impulse Source
● The room impulse response can be directly measured recording the sound produced by a impulsive source as:● pistol shot● balloon popping
CLIOfw
A IN B A OUT B
sourcemic
room
12
Recording of impulses with CLIO 10● The recording can be done:
● directly from the CLIO input using Leq menu● with an external recording device● the impulse recorded must be saved as a .wav file
13
From Impulse to Parameters
● The Acoustical Parameters are calculated using the measured impulse responses
● The decay curve for each frequency band (octave or 1/3 of octave) of interest is generated by backward integration of the impulse response
● Background noise compensation is performed● Precise knowledge of signal level is achieved by CLIO
hardware capabilities
16
STI (Speech Transmission Index)● If the impulse response is measured using the MLS
tecnique it is possible also to calculate the Speech Transmission Index (D. Rife, "Modulation Transfer Function Measurement with Maximum-Length Sequences", JAES Volume 40 Issue 10 pp. 779-790; October 1992)
-----------------------------------------------------STI index-----------------------------------------------------Oct.Band 125 250 500 1k 2k 4k 8kf1=0.63 0.745 0.811 0.761 0.813 0.829 0.880 0.972f2=0.80 0.686 0.739 0.688 0.747 0.755 0.801 0.890f3=1.00 0.644 0.686 0.633 0.699 0.699 0.741 0.827f4=1.25 0.579 0.604 0.556 0.635 0.616 0.651 0.733f5=1.60 0.523 0.536 0.510 0.592 0.553 0.581 0.661f6=2.00 0.477 0.478 0.485 0.558 0.501 0.521 0.602f7=2.50 0.415 0.404 0.458 0.525 0.431 0.438 0.527f8=3.15 0.345 0.338 0.434 0.504 0.362 0.353 0.461f9=4.00 0.298 0.302 0.427 0.455 0.235 0.166 0.373f10=5.00 0.353 0.260 0.421 0.405 0.204 0.090 0.361f11=6.30 0.442 0.144 0.368 0.415 0.317 0.363 0.462f12=8.00 0.563 0.493 0.447 0.493 0.450 0.548 0.650f13=10.00 0.400 0.392 0.379 0.495 0.403 0.505 0.687f14=12.50 0.372 0.298 0.340 0.477 0.272 0.379 0.507-----------------------------------------------------MTI 0.489 0.463 0.493 0.558 0.473 0.501 0.623STI=0.512 ALcons=10.6% rated Fair
17
Wavelet Analysis
● The wavelet analisys can be used as powerful tool for room acoustics
● The measured impulse response can be inspected in the time-frequency plane
● In the following example we see the effects of controlling reverberation and resonant modes by means of absorption and diffusion
18
Wavelet Analysis0
-5
-10
-15
-20
-25
dB
0 32 64 96 128 160 192 224 256 288 320 ms
100
1k
10k
20k
40
Hz
CLIO
Wavelet Analysis 9-20-2008 4.31.33 PM
Time-Frequency Energy Normalized Q 3.000 BW 0.333 octaves
File:
320cm – nessun trattamento
19
Wavelet Analysis0
-5
-10
-15
-20
-25
dB
0 32 64 96 128 160 192 224 256 288 320 ms
100
1k
10k
20k
40
Hz
CLIO
Wavelet Analysis 9-20-2008 4.30.18 PM
Time-Frequency Energy Normalized Q 3.000 BW 0.333 octaves
File:
320cm – full
Diminuzione riverberazioneDiminuzione riverberazionee controllo modalee controllo modale
tramite assorbimento tramite assorbimento e diffusionee diffusione
23
Building AcousticsMeasurement of the Apparent Sound Reduction Index R'
R' measurements depend on the following factors:● L1 : Emitting SPL ● L2 : Receiving SPL● Background noise● S : Dividing surface● A: Receiving absorption coefficient
24
Building AcousticsMeasurement of the Apparent Sound Reduction Index R'
100 1000 100000
20
40
60
80
100
120
SPL
BackgroundEmittingReceiving
100 1000 100000
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
RT 60
AveragePos 1Pos 2Pos 3Pos 4Pos 5Pos 6
26
Room and Building AcousticsMeasurement Equipment
● Normative references:● IEC 61672:2003. Electroacoustics — Sound level
meters. (replacing former IEC 651, Sound level meters and IEC 804, Integrating-averaging sound level meters)
● IEC 61260:2001. Electroacoustics — Octave-band and fractional-octave-band filters