LABSTAT INTERNATIONAL ULC. 262 Manitou Drive Kitchener, Ontario, Canada N2C 1L3 Phone: (519) 748-5409 Fax: (519) 748-1654 Web: www.labstat.com Peter JOZA , David LI, Andrew MASTERS, Bill RICKERT 1 Determination of Pyrazine Flavourants in Electronic Cigarette Cartridges and Liquids by Headspace Solid Phase Microextraction - Gas Chromatography- Mass Spectrometry (HS-SPME-GCMS) 2014 CORESTA CONGRESS October 12-16, 2014 Québec City, Canada 2014_ST47_Joza.pdf Congress2014 - Document not peer-reviewed by CORESTA

Cigarette Cartridges and Liquids by Headspace Solid Phase

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Cigarette Cartridges and Liquids by Headspace Solid Phase

LABSTAT INTERNATIONAL ULC. 262 Manitou Drive

Kitchener, Ontario, Canada N2C 1L3 Phone: (519) 748-5409 Fax: (519) 748-1654 Web: www.labstat.com

Peter JOZA , David LI, Andrew MASTERS, Bill RICKERT

1

Determination of Pyrazine Flavourants in Electronic

Cigarette Cartridges and Liquids by Headspace

Solid Phase Microextraction - Gas Chromatography-Mass Spectrometry (HS-SPME-GCMS)

2014 CORESTA CONGRESS October 12-16, 2014 Québec City, Canada

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 2: Cigarette Cartridges and Liquids by Headspace Solid Phase

Objectives

2

Apply the HS-SPME-GCMS used for profiling volatile and semi-volatile compounds in tobacco products to e-liquids and e-cigarette aerosol

Quantify four targeted compounds:

2,3,5-Trimethylpyrazine (TriMPZ), 2,3,5,6-Tetramethylpyrazine (TetraMPZ), 2-Acetylpyrrole (APL) 2-Acetylpyrazine (APZ)

Provide some measure of the changes in the delivery of flavour components (targeted and non-targeted) during the “life” of an e-cigarette

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 3: Cigarette Cartridges and Liquids by Headspace Solid Phase

Background

67th TSRC: A rapid profiling method which can be applied to the detection of flavorings in various tobacco products

HS-SPME-GCMS Profiles Flavoring ID

Cigarettes, Cigar/Cigarillos

RYO, Pipe, Chew, Snus,

Snuff, Shisha, Dissovable Wiley, NIST/EPA/NIH, FFNSC

3

Can the same process be used for the analysis of e-liquids and e-cig aerosols?

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 4: Cigarette Cartridges and Liquids by Headspace Solid Phase

Considerations

4

The impact of the e-liquid matrix on the quantification and detection limits of flavour components using HS-SPME?

The type(s) and amount of flavour components

• Are these reflective of tobacco products?

The transfer of flavour components of an e-liquid

• Are these transferred to the e-cigarette aerosol?

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 5: Cigarette Cartridges and Liquids by Headspace Solid Phase

Sample Preparation for HS-SPME-GCMS Analysis

Base Method vs. Emissions / e-Liquids

ISTD/FPCS:

d6-Benzene

d5-Acetophenone

d5-Benzophenone

Add 20 μL ISTD/FPCS

HS-SPME-GCMS

Add 2 mL 3.0 M KClaq

0.5 g of tobacco filler in 10 mL HS vial

Cap the vial and keep at room temp for 2 hrs

Spike FLs std solution

5

Collection Pad

(up to 250mg)

Add 10 μL ISTD

ISTD/FPCS:

d5-2-ethylpyrazine

d7-quinoline

Chromatographic

Differences

Standard

Preparation

Only

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 6: Cigarette Cartridges and Liquids by Headspace Solid Phase

Instrument Parameters

Instrument Parameters

SPME Fiber 65 µm PDMS/DVB

Sample Ads Temp/Time 50 C/20 min

Sample Des Temp/Time 260 C/5 min

Fiber Condition Temp/Time 260 C/26 min

GC Column DB-WAXETR (30 m x 0.25 mm x 0.25 µm)

GC Injector Splitless mode, inlet temp 260 C.

MS Mode Full Scan (35 - 400 m/z)

GC Run Time 50 min (program 40 to 260°C)

6

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 7: Cigarette Cartridges and Liquids by Headspace Solid Phase

Instrument Parameters

Compound RT (min) Quantifier Qualifier

d5-2-ethylpyrazine 15.6 112 113

2,3,5-Trimethylpyrazine

(TriMPZ) 17.5 122 42

2,3,5,6-Tetramethylpyrazine

(TetraMPZ) 19.4 136 54

d7-quinoline 30.1 136 108

2-Acetylpyrrole (APL) 23.4 122 80

2-Acetylpyrazine (APZ) 30.9 94 109

7

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 8: Cigarette Cartridges and Liquids by Headspace Solid Phase

Instrument Parameters

8

Nicotine

Propylene Glycol

Nicotyrine

Fibre Bleed Residual Solvent

~ 4ng TriMPZ

and TetraMPZ

~ 20ng APL

and APZ

Calibration Standard

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 9: Cigarette Cartridges and Liquids by Headspace Solid Phase

Instrument Parameters

9

d5-2-ethylpyrazine

TriMPZ TetraMPZ

d7-quinoline

APL APZ

Nicotine

Propylene Glycol

Nicotyrine

Fibre Bleed Residual Solvent

~ 4ng TriMPZ

and TetraMPZ

~ 20ng APL

and APZ

Calibration Standard

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 10: Cigarette Cartridges and Liquids by Headspace Solid Phase

Determination of Matrix Effect

10

Impact of Propylene Glycol and Glycerin

0

200000

400000

600000

800000

1000000

1200000

0 20 40 60 80 100

Res

po

nse

(A

rea)

Propylene Glycol (%)

TriMPZ

0

1000000

2000000

3000000

4000000

5000000

6000000

7000000

0 20 40 60 80 100

Res

po

nse

(A

rea)

Propylene Glycol (%)

APL

0.05

0.06

0.07

0.08

0.09

0.1

0.11

0 20 40 60 80 100

Res

po

nse

Rat

io

Propylene Glycol (%)

TetraMPZ/d5-2-EthylPZ

Calibration Standards Matrix

50:50 (v/v) Propylene Glycol to Glycerin solution

8 mg/mL Nicotine 250mg matrix at each level

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 11: Cigarette Cartridges and Liquids by Headspace Solid Phase

r2=0.997

Quantification of Target Analytes

LOD Determination - Root Mean Square Error (RMSE) *

Generation of a 4 or 5 point calibration curve

Instrument response (area counts) vs. standard concentration

Standards of concentration within an order of magnitude of estimated LOD (e.g. TetraMPZ - 1.10ng to 10.6ng)

* Handbook of Residue Analytical Methods for Agrochemicals, p68-70, Editor-in-Chief Dr Philip W Lee, DuPont Crop Protection,

USA; Published by John Wiley & Sons Ltd in 2003. 11

RMSE

LOD

LOQ

Compound LOQ (ng)

TriMPZ 4.27

TetraMPZ 2.75

APL 107

APZ 26.1

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 12: Cigarette Cartridges and Liquids by Headspace Solid Phase

Quantification of Target Analytes

12

Compound Range

(ng)

TriMPZ 1 - 750

TetraMPZ 1 - 850

APL 10 - 8000

APZ 10 - 8200

d5-2-ethylpyrazine - 194ng d7-quinoline - 112ng

(ISTD for APZ)

All r2 > 0.99

TIC Response (semi-quantitative)

TriMPZ / TetraMPZ: 1 to 4µg

APL and APZ: 10 to 40µg

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 13: Cigarette Cartridges and Liquids by Headspace Solid Phase

Analysis of E-liquids – Experimental

13

18 different formulations tested

• 13 e-liquids (1 advertized as “tobacco flavour)

• 1 e-liquid originating from a tobacco extract • 4 e-cigarette cartomizers (fresh)

• 4 variants of same product line (1 mentholated) • Centrifuge cartomizer at 1500 rpm for 5 min.

• 1 e-cigarette cartomizer (after use – first 20 puffs)

Single replicate for each formulation

• 0.25g sample size

Opportunity to identify other flavour components

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 14: Cigarette Cartridges and Liquids by Headspace Solid Phase

Analysis of E-liquids - Chromatography

14

Sample 1

Tobacco Flavour

Sample 2

No Description

Sample 5

Americano

Sample 1

TriMPZ

Saturated Signal

Sample 2

Requires a more dilute sample for analysis

Sample 5

APZ

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 15: Cigarette Cartridges and Liquids by Headspace Solid Phase

Analysis of E-liquids - Results

15

Sample No Matrix InfoTriMPZ

(ng/g)

TetraMPZ

(ng/g)

APL

(ng/g)

APZ

(ng/g)

1 E-Liquid Tobacco Flavour > UCL 197 - NQ

2 E-Liquid - 5.33 > UCL > UCL

3 E-Liquid Forest Fruit Mix 61.4 NQ - -

4 E-Liquid 7 Leaves 2601 > UCL - -

5 E-Liquid Americano 290 - NQ 8.75

6 E-Liquid Strawberry 5.13 NQ - -

7 Cubana 544 NQ - 195

8 E-Liquid 6.02 4.08 8767 1166

9 E-Liquid 6.21 4.94 - 31166

10 E-Liquid 45.9 7.07 NQ 402

11 E-Liquid 5.06 4.04 - NQ

12 E-Liquid 4.37 4.44 - 57.3

13 Tobacco Extract E-Liquid 52.7 NQ 129 -

14 Cartomizer Liquid 1 > UCL 3.7 NQ -

15 Cartomizer Liquid 2 > UCL NQ 106 -

16 Cartomizer Liquid 3 > UCL - - -

17 Cartomizer Liquid 4M > UCL NQ 117 -

18 Cartomizer Liquid (Used) > UCL - - -

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 16: Cigarette Cartridges and Liquids by Headspace Solid Phase

Analysis of E-liquids - Results

16

Sample No Matrix InfoTriMPZ

(ng/g)

TetraMPZ

(ng/g)

APL

(ng/g)

APZ

(ng/g)

1 E-Liquid Tobacco Flavour > 16 µg 197 - NQ

2 E-Liquid - 5.33 68795 * 47656 *

3 E-Liquid Forest Fruit Mix 61.4 NQ - -

4 E-Liquid 7 Leaves 2601 3638 * - -

5 E-Liquid Americano 290 - NQ 8.75

6 E-Liquid Strawberry 5.13 NQ - -

7 Cubana 544 NQ - 195

8 E-Liquid 6.02 4.08 8767 1166

9 E-Liquid 6.21 4.94 - 31166

10 E-Liquid 45.9 7.07 NQ 402

11 E-Liquid 5.06 4.04 - NQ

12 E-Liquid 4.37 4.44 - 57.3

13 Tobacco Extract E-Liquid 52.7 NQ 129 -

14 Cartomizer Liquid 1 4625 * 3.7 NQ -

15 Cartomizer Liquid 2 4089 * NQ 106 -

16 Cartomizer Liquid 3 5605 * - - -

17 Cartomizer Liquid 4M 10716 * NQ 117 -

18 Cartomizer Liquid (Used) 3842 * - - -

* Estimated using TIC area count

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 17: Cigarette Cartridges and Liquids by Headspace Solid Phase

E – Cig Aerosol – Experimental

17

1 brand (cartomizer 1 from E-liquids) • 4 replicates • 5 cartridges per replicate • Collection in 25 puff increments (5 increments) • Same cartomizers used for all 5 increments (20 total) • Freshly charged battery every 50 puffs • 44mm pad collection (½ pad SPME, ½ pad liquid extract)

Puffing Regimen • 70mL puff volume • 3.5 second puff duration (square wave profile) • 30 second puff frequency

Identify changes in delivery of flavour components (targeted and non-targeted)

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 18: Cigarette Cartridges and Liquids by Headspace Solid Phase

E – Cig Aerosol – Results

18

Puff Segment TPM

(mg/cart)

Cartridge Wt. Loss (mg/cart)

Propylene Glycol

(mg/cart)

Nicotine (mg/cart)

Glycerin (mg/cart)

1 – 25

69.6

± 2.0

(2.9%)

67.5

± 1.8

(2.7%)

4.50

± 0.15

(3.4%)

3.02

± 0.10

(3.3%)

51.6

± 1.3

(2.5%)

26 – 50

65.1

± 2.0

(3.1%)

63.3

± 2.2

(3.5%)

4.27

± 0.10

(2.3%)

2.84

± 0.07

(2.5%)

48.5

± 1.6

(3.4%)

51 – 75

61.7

± 1.0

(1.6%)

60.2

± 1.1

(1.8%)

4.13

± 0.18

(4.4%)

2.72

± 0.11

(4.1%)

46.1

± 1.4

(3.0%)

76 – 100

58.2

± 1.2

(2.1%)

56.8

± 1.2

(2.1%)

4.06

± 0.06

(1.5%)

2.65

± 0.03

(1.2%)

44.3

± 0.7

(1.6%)

101 - 125

54.9

± 1.9

(3.4%)

54.1

± 1.9

(3.4%)

3.70

± 0.14

(3.7%)

2.42

± 0.09

(3.8%)

40.5

± 1.8

(4.4%)

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 19: Cigarette Cartridges and Liquids by Headspace Solid Phase

E – Cig Aerosol – Results

19

Consistent loss in delivery of major aerosol components

Delivery of 5th segment is 80% that of the 1st segment (average)

75.0

80.0

85.0

90.0

95.0

100.0

2 3 4 5

Per

cen

tage

(%

) o

f Fi

rst

Pu

ff S

egm

ent

Puff Segment

Changes in Aerosol Delivery

MS TPM Cartridge Wt. Loss Propylene Glycol Nicotine Glycerol

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 20: Cigarette Cartridges and Liquids by Headspace Solid Phase

E – Cig Aerosol – Chromatography

20

E-liquid

Collection Segment 1

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 21: Cigarette Cartridges and Liquids by Headspace Solid Phase

E – Cig Aerosol – Chromatography

21

Only 2,3,5-Trimethylpyrazine (TriMPZ) found in the emmission

E-liquid

Collection Segment 1

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 22: Cigarette Cartridges and Liquids by Headspace Solid Phase

HS-SPME Results – Targeted

22

Puff Segment Nicotine

(mg/cart) TriMPZ * (ng/cart)

1 – 25

3.02

± 0.10

(3.3%)

472

± 19

(4.0%)

26 – 50

2.84

± 0.07

(2.5%)

435

± 14

(3.3%)

51 – 75

2.72

± 0.11

(4.1%)

427

± 16

(3.7%)

76 – 100

2.65

± 0.03

(1.2%)

412

± 18

(4.0%)

101 - 125

2.42

± 0.09

(3.8%)

386

± 7

(1.8%)

* TriMPZ results extrapolated using TIC response

Delivery of TriMPZ in the 5th segment is 82% that of the 1st segment

Results are consistent with major components of e-cigarette aerosol

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 23: Cigarette Cartridges and Liquids by Headspace Solid Phase

E – Cig Aerosol – Chromatography

23

No observed changes in TIC ratios amongst non-targeted compounds

Collection Segment 5

Collection Segment 4

Collection Segment 3

Collection Segment 2

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 24: Cigarette Cartridges and Liquids by Headspace Solid Phase

HS-SPME Aerosol – Non-Targeted

24

0.0%

20.0%

40.0%

60.0%

80.0%

100.0%

120.0%

ETHYL ACETATE

ETHANOL

2-METHOXY-3-METHYLPYRAZINE

2,3,5-TRIMETHYLPYRAZINE

PROPYLENE GLYCOL

OXOPHORONE

BETA-DEMASCONEISOMER

2-HYDROXY-3-METHYL-2-CYCLOPENTEN-1-ONE

NICOTYRINE

ETHYLVANILLIN

Segment 2 Segment 5

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 25: Cigarette Cartridges and Liquids by Headspace Solid Phase

Summary

25

HS-SPME-GCMS can be used for the determination of flavour components for both e-liquids and e-cigarette aerosol

• Optimization/Compromises dependent on levels

Matrix matching (glycerin / propylene glycol) need be considered for trace level quantitation

Screening of e-liquid is useful for identification of targeted compounds

This method can be used to assess the transfer of flavour components of an e-liquid to the E-cig aerosol.

• Targeted and non-targeted compounds

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA

Page 26: Cigarette Cartridges and Liquids by Headspace Solid Phase

Acknowledgements

26

Mr. MingZhong Cui

Mrs. Helena Coetzer

Thank You !

2014

_ST

47_J

oza.

pdf

Con

gres

s201

4 -

Doc

umen

t not

pee

r-re

view

ed b

y C

OR

ES

TA