19
© Fraunhofer IOF Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert 1 , Frank Wippermann 1 , Sarah Walther 1 , Thomas Burkhardt 1,2 , Bernhard Messerschmidt 3 , Franz Bechtold 4 , Torsten Vahrenkamp 5 , Ramona Eberhardt 1 , Daniel Gäbler 6 , Andreas Tünnermann 1,2 1 : Fraunhofer-Institute for Applied Optics and Precision Engineering, Jena, Germany 2 : Institute of Applied Physics, Friedrich Schiller University, Jena, Germany 3 : GRINTECH GmbH, Jena, Germany 4 : VIA electronic GmbH, Hermsdorf, Germany 5 : ficonTEC Service GmbH, Achim, Germany 6 : X-FAB Semiconductor Foundries AG, Erfurt, Germany

Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

  • Upload
    donhan

  • View
    220

  • Download
    5

Embed Size (px)

Citation preview

Page 1: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

© Fraunhofer IOF

Chip in the Tip ultra-compact cameras for endoscopic applications

Erik Beckert1, Frank Wippermann1, Sarah Walther1, Thomas Burkhardt1,2, Bernhard

Messerschmidt3, Franz Bechtold4, Torsten Vahrenkamp5, Ramona Eberhardt1, Daniel Gäbler6, Andreas Tünnermann1,2

1: Fraunhofer-Institute for Applied Optics and Precision Engineering, Jena, Germany2: Institute of Applied Physics, Friedrich Schiller University, Jena, Germany

3: GRINTECH GmbH, Jena, Germany4: VIA electronic GmbH, Hermsdorf, Germany5: ficonTEC Service GmbH, Achim, Germany

6: X-FAB Semiconductor Foundries AG, Erfurt, Germany

Page 2: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

Slide 2 of 19

© Fraunhofer IOF

Chip in the Tip Mini Endoscope – DesignOpticsOptomechanics

CMOS-Imager and LTCC PCBSpecifications and TechnologyThrough Silicon Via & Freeform DicingLTCC PCB Design and Packaging

Optics Manufacturing and AssemblySPDT PrototypingFlip-Chip / Pick&Place AssemblyCharacteristics

Summary

Outline

Page 3: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

Slide 3 of 19

© Fraunhofer IOF

Chip in the Tip – Motivation

Miniaturized Video Endoscope

best possible resolution

Image Akquision and ProcessingCapacity in the endoscope Head

Wire or Wireless Connection to the Environment

compact Designs, round

flexible Layouts and Assembly

cheap Manufacturing

fast Prototyping

Stable or disposable Systems 30 ° View, 70 ° FOV

Straight View, 110 ° FOV

Objective Imager

Page 4: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

Slide 4 of 19

© Fraunhofer IOF

Chip in the Tip – Optical Designs

Straight View

2.8 µm Pixel

Nyquist 178 LP / mm

FOV 75 ° and 110 °

F-Number 4.6

Object Distance 8 mm

Front Plano-Asphere

GRIN Lens with back side Aperture

Back Sphere

Page 5: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

Slide 5 of 19

© Fraunhofer IOF

Chip in the Tip – Optomechanical Design

Stainless Steel Tube Ø 3 mmObjective Ø 2.8 mm

SPDT / Injection moulded LensesBack Sphere: Drop-on-Demand possible

stacked ObjectiveImager Anisotropic Conductive

Adhesive mounted on LTCC Printed Circuit Board

LTCC PCB carries ClockImager connected during Assembly & actively alignedexternal Illumination

CMOS-ImagerLTCC PCB

Oscillator50 MHz

4 WiresDistal

Interface

Front Window

Back Sphere

GRIN Lens

Front Asphere

Objective:

Imager Sub-Assembly:

Page 6: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

Slide 6 of 19

© Fraunhofer IOF

VGA CMOS-Imager – Specifications

…because the world is analog.

Design by:>VGA(640×480) Resolution - 680x680Pixel Size 2.78 µmImage Circle Ø 2.1 mmOuter Ø 2.7 mm50 Frames per SecondOn-Chip Pre-Processing6 Connectors

Page 7: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

Slide 7 of 19

© Fraunhofer IOF

VGA CMOS-Imager – Through Silicon Via and FreeForm Dicing

…because the world is analog.

FreeForm Dicing - FFD

fully flexible chip shape

no unfunctional Si left

Through Silicon Via - TSV

for 400 µm wafers

TSV lands on 60x60 µm² metal1 areas in CMOS Backend

Via filling by Cu electroplating

Backside pads and routingBackside Al-Routing on FFD Imager Chip

Page 8: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

Slide 8 of 19

© Fraunhofer IOF

Ceramic Carrier – LTCC PCB

Round LTCC Carrier∅ 2.8 mm, 0.8 mm thick4 Layers DP951Rerouting Front (Sensor) -> Back (Interface) with Oscillator SG-150 (20..50 MHz) and C-IntegrationOut: Data +/-, 3.3 V, GND

4“ Multi-PCB Coupon49 PCB per CouponPre-singulated by cutting / laseringFinal Singulation after Reflow by Breaking

Back Side

LTCC Coupon 4“

∅ 2.8 mm

Front (Sensor) Side

Page 9: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

Slide 9 of 19

© Fraunhofer IOF

Ceramic Carrier – LTCC PCB

Front (Sensor) SideBack (Component and Interface) Side

Page 10: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

Slide 10 of 19

© Fraunhofer IOF

LTCC PCB Packaging

Two Stage Flip-Chip

Sensor: Solder Bumps 200 µm, 80Au20Sn on LTCC

370°C / 100°C Sensor / LTCC Reflow

Components. Solder Bumps 400 µm SAC305

240°C / 100°C Oscillator, C / LTCC

Reflow by FiconTec BL2000

SG-150

Sensor

LTCC

X-Sec Package

Reflow Sensor

Reflow Oscillator

Reflow Capacity

Page 11: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

Slide 11 of 19

© Fraunhofer IOF

Optics Manufacturing – Prototype Lenses SPDT

Single Point Diamond Turning for Rapid Prototyping in Polymer

Material: ZEONEX

PV: <300 nm, rms ca. 50 nm

Single Point Diamond Turning for Mould Manufacturing

Mould Material: electro-plated NiP

Ultrasonic Sound enhanced SPDT -> Stainless Steel Moulds

Replication by Injection Moulding

SPDT Machining

Prototype Asphere

Lens Surface Location / mm

Form

Dev

iatio

n / µ

m

Page 12: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

Slide 12 of 19

© Fraunhofer IOF

Optics Assembly – Automated Pick&Place for Stacking

Objective Stackingpassive Concentric AlignmentDecenter < 10 µmTip / Tilt < 1‘

active Alignment (Tip/Tilt/ Focus) Objective <-> Imager-Assembly

Imager pre-Assembly on LTCCHandling / Connector frame on LTCCSingulation @ predetermined break point

automated in FlipChip Bonder BL2000 (ficonTEC)

75 ° FOV Components Set ObjectiveStacking

LTCC PCB

Automated Assembly with BL2000

Page 13: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

Slide 13 of 19

© Fraunhofer IOF

Optics Assembly – Front Window Soldering

Long term stable soldered Front Window

Solderjet Bumping1 – overlapping discrete Solder Droplets

liquid Solder Application, no flux, local thermal Impact

300 µm Ø Spheres AuSn used

Time ca. 1 min (Bumping Fre-quency 1 Hz, 10 Hz possible)

Autoclave Cycles 5x @ 134 °C, 98 % rel. Humidity, 5 min

Hermeticity before and after Autoclave Cycles: 4e-9 mbar*l*s-1

X-SecWindow / Solder /

Tube

3D View

Jet Processing

1: Patented: DE102007002436 & WO03/006197

Page 14: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

Slide 14 of 19

© Fraunhofer IOF

Front Window Soldering

First Step: Fixation Second Step: Hermetic Sealing

Page 15: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

Slide 15 of 19

© Fraunhofer IOF

Optics Evaluation – Modulation Transfer Function

0

0,1

0,2

0,3

0,4

0,5

0,6

0,7

0,8

0,9

1

0 20 40 60 80 100 120 140 160 180 200 220 2400

0,1

0,2

0,3

0,4

0,5

0,6

0,7

0,8

0,9

1

0 20 40 60 80 100 120 140 160 180 200 220 240

MTF FOV 75 ° MTF FOV 110 °

LP / mmLP / mm

rel.

Con

trast

rel.

Con

trast

Method: Slanted Edge*, Goal(Nyquist/2): 90 LP / mm @ 30 % contrast

FOV 75 °: > 100 LP / mm @ 30 % contrast

FOV 110 °: > 100 LP / mm @ 30 % contrast*: ISO/TC42N 4230 (1997), Electronic Still Picture Imaging. Spatial Frequency Response (SFR) Measurements

Page 16: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

Slide 16 of 19

© Fraunhofer IOF

0

0,1

0,2

0,3

0,4

0,5

0,6

0,7

0,8

0,9

1

0 20 40 60 80 100 120 140 160 180 2000

0,1

0,2

0,3

0,4

0,5

0,6

0,7

0,8

0,9

1

0 20 40 60 80 100 120 140 160 180 200

0

0,1

0,2

0,3

0,4

0,5

0,6

0,7

0,8

0,9

1

0 20 40 60 80 100 120 140 160 180 2000

0,1

0,2

0,3

0,4

0,5

0,6

0,7

0,8

0,9

1

0 20 40 60 80 100 120 140 160 180 200

MTF @ -6 mm,FOV 75 °

MTF @ -6 mm,FOV 110 °

MTF @ +12 mm,FOV 75 °

MTF @ +12 mm,FOV 110 °

Optics Evaluation – Depth of Fokus

LP / mmLP / mm

rel.

Con

trast

rel.

Con

trast

LP / mmLP / mm

rel.

Con

trast

rel.

Con

trast

distal proximalGoal: 60 LP / mm @ 20 % contrast

Page 17: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

Slide 17 of 19

© Fraunhofer IOF

Optics Evaluation – Distortion

scary „real life“ object

FOV 75 °

-18 % max.

FOV 110 °

-40 % max.

Page 18: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

Slide 18 of 19

© Fraunhofer IOF

Summary

SPDT manufactured Prototype Lenses

Central GRINS lens for stacking

FFD round CMOS-Sensor with TSV

Pick& Place for simplified Assembly

long term stable soldered Front Window

automated mid-Volume Production

Flexible Integration

Page 19: Chip in the Tip ultra-compact cameras for endoscopic ... - Beckert.pdf · Chip in the Tip ultra-compact cameras for endoscopic applications Erik Beckert1, ... X-Sec Package Reflow

Slide 19 of 19

© Fraunhofer IOF

Acknowledgement

Thank you for your Attention.

The work presented in this presentation is based on the project HoKa (miniaturized high resolution camera systems). Funding by the Thuringian Government and the European Regional Development Fund (ERDF) as well as project management by the Thueringer Aufbaubank (project number 2008 VF 0057) is gratefully acknowledged.