12
10/15/09 1 Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy Winter 2009 Page 1 Topic 2b: X-ray Fluorescence Spectrometry Text: Chapter 12 Rouessac (1 week) 4.0 X-ray Fluorescence Topic 2: Atomic Spectroscopy Download, read and understand EPA method 6010C ICP-OES Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy Winter 2009 Page 2 Topic 2b: XRF Atomic X-ray Spectrometry Fundamental Principles Emission of X-rays Absorption Spectra Mass Absorption Coefficient X-Ray Fluorescence Instrumentation Sources Monochromators Transducers Applications X-ray Fluorescence Methods Qualitative Quantitative

Chemistry 311: Instrumentation Analysis · 2009. 10. 16. · X-ray Spectrometry Instrumentation: Topic 2b: XRF Sources: X-Ray Tube: These are the most commonly used sources for analytical

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

  • 10/15/09

    1

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 1

    Topic 2b: X-ray Fluorescence Spectrometry Text: Chapter 12 Rouessac (1 week)

    4.0 X-ray Fluorescence

    Topic 2: Atomic Spectroscopy

    Download, read and understand EPA method 6010C ICP-OES

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 2

    Topic 2b: XRF Atomic X-ray Spectrometry

    Fundamental Principles

    Emission of X-rays Absorption Spectra

    Mass Absorption Coefficient

    X-Ray Fluorescence

    Instrumentation Sources Monochromators Transducers

    Applications X-ray Fluorescence Methods

    Qualitative Quantitative

  • 10/15/09

    2

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 3

    Fundamental Principles: X-rays are short wavelength (10-5 Å to 100 Å) EM produced by the deceleration of high-energy electrons or by electronic transitions of electrons in the inner orbitals of atoms. In practice the wavelength range most often used for analytical purposes is 0.1 Å to 25 Å (0.01 nm to 2.5 nm).

    Emission of X-rays: There are 4 main sources of “analytical” X-rays; A) Bombardment of a metal target with a beam of high-energy electrons B) X-ray Fluorescence by a material irradiated by X-rays C) Use of an radioactive source D) Synchrotron radiation Source (highly specialized facility not discussed)

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 4

    Topic 2b: XRF A)  Bombardment of a metal target with a beam of high-energy electrons e- produced at cathode and accelerated toward a high potential anode (100 kV). Collision e- decelerated and X-ray spectrum is produced. This is dependent only on the accelerating voltage and is independent of target material. Energy of photon is equal to difference in KE before and after collision.

  • 10/15/09

    3

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 5

    A) Bombardment of a metal target with a beam of high-energy e- Maximum photon energy corresponds to the instantaneous and complete deceleration of the electron. Described mathematically by Duane-Hunt Law;

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 6

    A) Bombardment of a metal target with a beam of high-energy e- Emission behavior of Molybdenum is typical of all elements with atomic numbers (A#) greater than 23 X-ray line spectra relatively simple, with shorter wavelength being K series and longer L series. Elements with A#’s < 23 produce only K series. X-Ray line spectra have a specific “appearance energy” for Mo this is 20 KV. For Tungsten these lines appear at >70KV

  • 10/15/09

    4

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 7

    4d10 4p6 4s2

    3d10 3p6 3s2

    2p6 2s2

    1s2

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 8

    Topic 2b: XRF B) X-ray Fluorescence by a material irradiated by X-rays The absorption of X-rays produces electronically excited ions, when the ion returns to it’s ground electronic state, characteristic λ are produced. Cutoff λ from the primary X-ray source must be less (greater in energy) than the “absorption edge” of the analyte.

    C) Use of an radioactive source X-ray radiation can be produced by radioactive species. γ-rays are high energy EM that is indistinguishable from X-rays.

    Another radioactive process is electron capture in which the nucleus captures an electron to form a new atomic species (with lower atomic number). K electrons because of their proximity are captured must often, leaving the K-level electron hole needed for characteristic radiation. Common example is

    55Fe → 54Mn + h ν Mn Kα line at 2.1 Å results

  • 10/15/09

    5

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 9

    Topic 2b: XRF Absorption Spectra X-rays are absorbed by materials through an photoelectron effect process. Typical absorption spectra are presented below

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 10

    Topic 2b: XRF Absorption Spectra The absorption spectra of a given element is relatively simple Observed λ is characteristic of the element and is independent of it’s chemical state. Inner e- far removed from valence e-. Sharp discontinuities are called “absorption edges” The absorption edge for a given band ie., K reflects the difficulty in removing an electron from that orbital. It is more difficult to extract a 1s electron (e-) close to the a nucleus with 82 protons (+82 charge) than it is to extract a 1s electron (e-) close to the a nucleus with 47 protons (+47 charge) ∴ 82Pb has a much lower wavelength (higher energy) K band than does 47Ag.

    Mass Absorption Coefficient Beer’s law is also applicable to absorption of X-radiation.

  • 10/15/09

    6

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 11

    Topic 2b: XRF X-ray Spectrometry Instrumentation:

    Sources: X-Ray Tube: These are the most commonly used sources for analytical work. (see previous diagram, (Skoog Figure 12-7))

    Radioisotopes: The nature of the radiation used with these sources is completely dependent on the radioactive material used. Many produce line spectra. Since absorption sensitivity is related to the proximity to specific absorption edges, specific sources are more applicable to specific analysis.

    Secondary Fluorescence: This can be quite useful, as discrete lines are produced without the underlying continuum of X-ray tube. However, a primary X-ray tube or Radioisotope source is required to stimulate fluorescence.

    Source
 λ
selector
Sample
Holder
 Detector


    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 12

    λ Monochromators and Filters:

    Filters: Thin strips of metal can provide effective λ filters.

    λ Monochromators: Crystals can be used to produce monochromatic radiation via application of Bragg’s law

  • 10/15/09

    7

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 13

    Topic 2b: XRF

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 14

    X-ray Transducers: Usually the monitored signals in X-ray spectrometry are of low intensity and frequency, as a result transducers are often operated in a photon counting mode. Most of the detectors in X-ray spectrometry rely on the ionizing nature of X-ray radiation to produce measurable electronic signals. Gas-Filled Transducers: Inert gasses such as Argon, Xenon or Krypton are enclosed metal tube equipped with electrodes that have a high potential applied across them. When X-rays ionize the gas, a current is produced, the nature of which is dependent on the magnitude of the applied potential. 3 types of transducers are obtained (see next page).

    Geiger Tube Proportional Counters Ionization Chambers

    .

  • 10/15/09

    8

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 15

    Topic 2b: XRF X-ray Transducers (cont..):

    •  Scintillation Counters: Radiation striking a phosphor produces luminescence that can be monitored and amplified with a photomultiplier tube. •  Semi-conductor Transducers: semiconductor based detectors have a roughly analogous mode of operation to gas filled detectors.

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 16

    Gas-Filled Transducers:

    Geiger Tube: If the potential is > ~1000V significant amplification occurs (~109). Space charge effects cause a dead-time of 50-200 µsec for this device.

    Proportional Counters: Signal gains are less (500 - 10,000) and thus require additional amplification. Dead time is approximately 1 µsec. Signal intensity is dependent on the energy (frequency) of the incident radiation, thus if selected ranges of signals are counted in sequence a frequency domain spectra can be obtained.

    Ionization Chambers: Currents are small in this range and thus the sensitivity is also low. Not used in X-ray spectrometry.

  • 10/15/09

    9

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 17

    Topic 2b: XRF Signal Processors: Pulse-Height Selectors: Only signals with a preset range of intensities are collected. See Figure 12-13. Pulse-Height Analyzers: Signals with specific energy range have distinct energy ∴ scanning energy range is comparable to scanning frequency (or λ).

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 18

    Topic 2b: XRF Applications: X-Ray Fluorescence: The non-destructive nature of this technique makes it very popular especially for qualitative purposes. Semi-quantitative and even quantitative analyses are also possible although these are more difficult.

    X-Ray Fluorescence Instrumentation: There are 3 basic types. The later two listed below could be equipped with either a X-ray tube or radioactive source.

    Wavelength Dispersive: Since only a small fraction of incident radiation can be effectively dispersed into monochromatic radiation, an intense source is required. ∴ this type of instrument requires a X-ray tube (104 more intense than common Radioactive sources). These can be either sequential (~$60,000) or multi-channel (>$150,000).

    Energy Dispersive Non-Dispersive

  • 10/15/09

    10

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 19

    Topic 2b: XRF Energy Dispersive: A schematic for a typical energy dispersive instrument is adjacent. Since the source, sample and detector can be placed close to each other signal losses are significantly reduced. Much less expensive (~$15,000 – $20,000).

    Non-Dispersive: If a filter or series of filters are placed before the detector, only specific frequencies can be passed, producing a very simple low cost instrument.

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 20

  • 10/15/09

    11

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 21

    Qualitative and Semi-quantitative Analysis: Qualitative information is obtained by the observed frequency of the radiation. The observed relative intensity of the lines is a rough guide for quantitative determination. A better quick estimate is to use the following relationship;

    Px = Ps Wx Where; Px is the observed intensity: Ps is the intensity of pure material; and Wx is the weight fraction of x

    Quantitative Analysis: Reasonably accurate quantitative results can be obtained if standards with nearly identical matrices can be used for calibration.

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 22

    Topic 2b: XRF Matrix effects: Both bulk and surface elements can absorb X-rays and emit characteristic radiation. For those in the bulk material; the intensity of the excitation absorption is attenuated by the material radiation must pass through before reaching analyte. Furthermore, the fluorescence emitted by the analyte must pass back through material and therefore may also be absorbed. Furthermore, matrix material might also emit interfering radiation.

    Calibration: External calibration standards can be used in an identical manner. Dissolving or diluting sample (fusing) may also be used to create a constant matrix.

  • 10/15/09

    12

    Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

    Winter 2009 Page 23

    Topic 2b: XRF

    Disadvantages: •  Low sensitivity (0.01 to 100 %) •  Less applicable for lighter elements (elements below Vanadium A# 23)

    • Cost of $5000 to $500000

    Advantages and Disadvantages of X-Ray Fluorescence:

    Advantages: • Simple spectra • Spectral interferences limited • Non-destructive technique (for the most part) • Many sample types and sizes • Very rapid and convenient