Upload
others
View
10
Download
0
Embed Size (px)
Citation preview
KHP
CHARACTERIZATION
of NANOMATERIALS
KHP
Overview of the most common nano-
characterization techniques
MAIN CHARACTERIZATION TECHNIQUES:
General Techniques
NMR, IR, UV, CV, etc.,
1.Transmission Electron Microscope (TEM)
2. Scanning Electron Microscope (SEM)
3. Scanning Probe Microscope (SPM)
4. Elemental Analysis (EDS, XPS, ICP)
5. X-ray Powder Diffractometer (XRD)
KHP
1.Transmission Electron
Microscope (TEM)
KHP5
Transmission electron microscopy (TEM) is a microscopy
technique in which a beam of electrons is transmitted through
an ultra-thin specimen, interacting with the specimen as it
passes through. An image is formed from the interaction of the
electrons transmitted through the specimen; the image is
magnified and focused onto an imaging device, such as a
fluorescent screen, on a layer of photographic film, or to be
detected by a sensor such as a CCD camera.
The first TEM was built by Max Knoll and Ernst Ruska in 1931,
with this group developing the first TEM with resolution greater
than that of light in 1933 and the first commercial TEM in 1939.
KHP
Energy Filtering TEM(EF-TEM)
High Voltage TEM(HVEM)
Field Emission TEM(FE-TEM)
KHP
KHP
High voltage Tank
Field emission Gun
Specimen Holder
KHP9
KHP
KHP11
KHP12
KHP
Different contrast of sample
KHP14
Electron diffraction is most frequently used in solid state physics and chemistry to
study the crystal structure of solids. Experiments are usually performed in a
transmission electron microscope (TEM), or a scanning electron microscope (SEM
). Selected area (electron) diffraction (abbreviated as SAD or SAED), is a
crystallographic experimental technique that can be performed inside a transmissi
on electron microscope (TEM).
In a TEM, a thin crystalline specimen is subjected to a parallel beam of high
energy electrons. As TEM specimens are typically ~100 nm thick, and the
electrons typically have an energy of 100–400 kiloelectron volts, the electrons
pass through the sample easily.
In this case, electrons are treated as wave-like, rather than particle-like. Because
the wavelength of high-energy electrons is a few thousandths of a nanometer, and
the spacing between atoms in a solid is about a hundred times larger, the atoms
act as a diffraction grating to the electrons, which are diffracted. That is, some
fraction of them will be scattered to particular angles, determined by the crystal
structure of the sample, while others continue to pass through the sample without
deflection. As a result, the image on the screen of the TEM will be a series of
spots—the selected area diffraction pattern, SADP, each spot corresponding to a
satisfied diffraction condition of the sample's crystal structure. If the sample is tilted
, the same crystal will stay under illumination, but different diffraction conditions
will be activated, and different diffraction spots will appear or disappear.
SAED
KHP
Example
KHP
KHP
HVEM
KHP
2. Scanning Electron
Microscope (SEM)A scanning electron microscope (SEM) is a type of electron microscope that
produces images of a sample by scanning it with a focused beam of electrons.
The electrons interact with atoms in the sample, producing various signals that
can be detected and that contain information about the sample's surface
topography and composition. The electron beam is generally scanned in a
raster scan pattern, and the beam's position is combined with the detected
signal to produce an image. SEM can achieve resolution better than 1
nanometer. Specimens can be observed in high vacuum, in low vacuum, in dry
conditions (in environmental SEM), and at a wide range of cryogenic or
elevated temperatures.
The most common mode of detection is by secondary electrons emitted by
atoms excited by the electron beam. On a flat surface, the plume of secondary
electrons is mostly contained by the sample, but on a tilted surface, the plume
is partially exposed and more electrons are emitted. By scanning the sample
and detecting the secondary electrons, an image displaying the topography of
the surface is created. Since the detector is not a camera, there is no diffraction
limit for resolution as in optical microscopes and telescopes.
KHP
KHP
KHP
3. Scanning Probe
Microscope (SPM)
Scanning probe microscopy (SPM) is a branch of microscopy that forms
images of surfaces using a physical probe that scans the specimen. SPM
was founded with the invention of the scanning tunneling microscope in
1981.
Many scanning probe microscopes can image several interactions
simultaneously. The manner of using these interactions to obtain an image
is generally called a mode.
KHP
SPM(Scanning Probe
Microscope)
AFM(Atomic Force
Microscope, SFM)
STM(Scanning Tunneling
Microscope)
(Scanning Force Microsccope)
KHP27
Atomic force microscopy (AFM) or scanning force microscopy (SFM) is a very
high-resolution type of scanning probe microscopy (SPM), with demonstrated
resolution on the order of fractions of a nanometer, more than 1000 times better
than the optical diffraction limit. The precursor to the AFM, the scanning
tunneling microscope (STM), was developed by Gerd Binnig and Heinrich
Rohrer in the early 1980s at IBM Research - Zurich, a development that earned
them the Nobel Prize for Physics in 1986.
The AFM consists of a cantilever with a sharp tip (probe) at its end that is used
to scan the specimen surface. The cantilever is typically silicon or silicon nitride
with a tip radius of curvature on the order of nanometers. When the tip is
brought into proximity of a sample surface, forces between the tip and the
sample lead to a deflection of the cantilever according to Hooke's law.
Depending on the situation, forces that are measured in AFM include
mechanical contact force, van der Waals forces, capillary forces, chemical
bonding, electrostatic forces, magnetic forces, etc.
Along with force, additional quantities may simultaneously be measured
through the use of specialized types of probes. Typically, the deflection is
measured using a laser spot reflected from the top surface of the cantilever into
an array of photodiodes.
Atomic force microscopy
KHP30
KHP
KHP
4. Elemental
Analysis
(EDS, XPS, ICP)
KHP
1) X-ray photoelectron spectroscopy (XPS)
ESCA(Electron Spectroscopy for
Chemical Analysis)XPS is also known as ESCA (Electron
Spectroscopy for Chemical Analysis), an
abbreviation introduced by Kai Siegbahn's
research group to emphasize the chemical
(rather than merely elemental) information
that the technique provides.
X-ray photoelectron spectroscopy (XPS) is a surface-sensitive
quantitative spectroscopic technique that measures the elemental
composition at the parts per thousand range, empirical formula,
chemical state and electronic state of the elements that exist within a
material. XPS spectra are obtained by irradiating a material with a
beam of X-rays while simultaneously measuring the kinetic energy
and number of electrons that escape from the top 0 to 10 nm of the
material being analyzed.
KHP36
XPS is used to measure:
elemental composition of the surface (top 0–10 nm usually)
empirical formula of pure materials
elements that contaminate a surface
chemical or electronic state of each element in the surface
uniformity of elemental composition across the top surface (or
line profiling or mapping)
uniformity of elemental composition as a function of ion beam
etching (or depth profiling)
KHP37
2) Energy Dispersive Spectrometer(EDS)
Energy Dispersive X-ray Analysis(EDX, EDXS, XEDS, etc.)
Energy-dispersive X-ray spectroscopy (EDS, EDX, or XEDS), sometimes
called energy dispersive X-ray analysis (EDXA) or energy dispersive X-ray
microanalysis (EDXMA), is an analytical technique used for the elemental
analysis or chemical characterization of a sample. It relies on an
interaction of some source of X-ray excitation and a sample. Its
characterization capabilities are due in large part to the fundamental
principle that each element has a unique atomic structure allowing unique
set of peaks on its X-ray emission spectrum.
KHP38
To stimulate the emission of characteristic X-rays from a specimen, a high-energy beam
of charged particles such as electrons or protons (see PIXE), or a beam of X-rays, is
focused into the sample being studied. At rest, an atom within the sample contains
ground state (or unexcited) electrons in discrete energy levels or electron shells bound
to the nucleus. The incident beam may excite an electron in an inner shell, ejecting it
from the shell while creating an electron hole where the electron was. An electron from
an outer, higher-energy shell then fills the hole, and the difference in energy between the
higher-energy shell and the lower energy shell may be released in the form of an X-ray.
The number and energy of the X-rays emitted from a specimen can be measured by an
energy-dispersive spectrometer. As the energy of the X-rays are characteristic of the
difference in energy between the two shells, and of the atomic structure of the element
from which they were emitted, this allows the elemental composition of the specimen to
be measured
KHP39
3) Inductively Coupled Plasma (ICP) Emission Spectroscopy
Inductively coupled plasma atomic emission spectroscopy (ICP-AES), also
referred to as inductively coupled plasma optical emission spectrometry (ICP-
OES), is an analytical technique used for the detection of trace metals. It is a
type of emission spectroscopy that uses the inductively coupled plasma to
produce excited atoms and ions that emit electromagnetic radiation at
wavelengths characteristic of a particular element. The intensity of this
emission is indicative of the concentration of the element within the sample.
KHP
5. X-ray Powder Diffractometer
(XRD)
Powder diffraction is a scientific technique using X-ray, neutron, or electron
diffraction on powder or microcrystalline samples for structural characterization of
materials
KHP47
KHP
KHP
Bragg's law (equation)
nl = 2dhkl sinq
KHP
KHP
JCPDS : Joint Committee on Powder Diffraction Standards
ICDD : International Center for Diffraction Data
Powder Diffraction File (PDF) #
Program: EVA, ICSD, Pcpdfwin ….
KHP
PCPDFWIN: JCPDS No. or PDF No.
KHP
Example
KHP
KHP
KHP
Example 4
Hexagonal In2S3
Cubic InSe
KHP
B is the full-width at half maximum