Upload
others
View
4
Download
2
Embed Size (px)
Citation preview
Chapter 10
On-Machine Metrology for Hybrid Machining Processes
Dr. J. Ramkumar1 and Vijay Mandal2
1Professor and 2Research Student
Department of Mechanical Engineering
Micromanufacturing Lab, I.I.T. Kanpur
Micromanufacturing Lab, I.I.T. Kanpur
1. Introduction.
(Concept of surface metrology.)
2. Optical methods for On-Machine surface Metrology .
(Criteria for the development of OMM system, advantages of optical techniques.)
3. Challenges from offline to On-machine Metrology.
(Overview of various optical methods and their suitability for application to OMM system.)
4. Case study: On-Machine Measurement using Dispersed Reference Interferometry.
(The working principle and hardware design of the DRI system)
5. DRI on-machine calibration.
(On-machine vibration test, machine kinematic error compensation and linearity error
calibration.)
6. Application of DRI to an ultra precision diamond turning machine.
7. Conclusions
8. References
Organization of the presentation
Micromanufacturing Lab, I.I.T. Kanpur
1. Introduction
Surface metrology refers to the measurement that describes the deviation of a
structured surface from its ideal shape. It is the science that deals with the
measurement of topography, i.e. the characterization of a surface by the
amplitude, spacing and shape of its features [2].
Surface measurement is fundamental to further enhance machining accuracy
and efficiency in manufacturing.
On-machine metrology (OMM) has become very popular as it can avoid
time-consuming realignment operations and possible damage during the
transportation between machine and measurement platform.
3 terms commonly used in surface characterization:-
i) Form
ii) Waviness
iii) Roughness 1 Micromanufacturing Lab, I.I.T. Kanpur
Form refers to the deviations caused bysuch factors as errors of a machine tool,distortions such as gravity and thermaleffects, spindles, or incorrect alignmentof the workpiece. It has low frequencycomponents on a surface.
Waviness has medium frequencycomponents and is more strictly definedas the irregularities whose spacing isgreater than the roughness samplinglength. It results from the machinedeflections and vibration.
Roughness has high frequency or shortwavelength components and representsthe fine irregularities normally causedby the manufacturing process includingthe impression left by grinding orpolishing.
1. Introduction
Fig. Roughness, waviness, and form of a
measured surface [3-6].2 Micromanufacturing Lab, I.I.T. Kanpur
Criteria For The Development Of On-machine Optical Surface Metrology System
Implementing On-machine metrology(OMM) requires a measurement
instrument capable of integrating into a platform executing a manufacturing
process.
Instrumentation for this task is ideally highly automated and requires minimal
operator interaction, while at the same time coping with the challenges posed
by the localized environment.
“On-machine” metrology is distinct from “in-process” metrology as the
former considers only measurement with the manufacturing process at a halt.
The latter involves active measurement during a machining process which
brings substantial extra challenge in terms of dealing with environmental and
dynamic effects.
2. Optical methods for On-Machine surface Metrology [1]
3 Micromanufacturing Lab, I.I.T. Kanpur
Advantages of Optical Techniques for Measurement:-
• Optical methods of metrology are the most realistic solution to the vastmajority of OMM applications.
• They allow fast, noncontact measurement of surface topography anddimensions.
• Optical techniques can operate effectively over scales ranging fromnanometers to meters meaning there is a high level of applicability acrossmany types of manufacturing.
• For instance, structured light (fringe projection) and deflectometrytechniques can measure large parts such as car body panels to micronlevels of form accuracy while phase shifting interferometry can measureform error in optics in nanometers.
2. Optical methods for On-Machine surface Metrology [1]
4 Micromanufacturing Lab, I.I.T. Kanpur
Measurement instrumentation must meet some or all of the following
criteria in order to be suitable for implementing OMM, depending on the
specific application:
• Small physical size
• Low cost
• High speed
• High dynamic range
• Highly automated
• High vibration tolerance
• Environmental tolerance
• Robust to contaminants
• Robust to electromagnetic interference
• Measurement flexibility
2. Optical methods for On-Machine surface Metrology [1]
5 Micromanufacturing Lab, I.I.T. Kanpur
No single optical measurement technology is currently suitable for all
measurand types or machine environments.
Commonly used commercial technologies are:-
• Coherence Scanning Interferometry (CSI)
• Focus Variation (FV)
• Laser Scanning Confocal Microscopy (LSCM)
All these provide areal surface topography measurement.
3. Challenges from offline to on-machine metrology [1]
6 Micromanufacturing Lab, I.I.T. Kanpur
FV LSCM CSI
Working
Principle
Focus-Variation integrates
the small depth of focus of
an optical system with
vertical scanning to give
color and topographical
data from the variation of
focus. It relies upon
analyzing contrast changes
between adjacent pixels as
the focus position changes
during axial scanning.
In LSCM a finely
focused beam of laser
light is scanned across
a sample. Additionally,
by varying the stage
height, a series of
sequential images
through the thickness
of a single sample can
be collected and used
to project a 3D image.
Coherence scanning
interferometry (CSI) is a 3D
imaging technique used to
measure areal surface
topography. It combines the
vertical resolution of an
interferometer with the lateral
resolution of a high-power
microscope and provides a fast,
non-contacting alternative to
contact stylus profilometers.
Characte
ristics
It is only effective on
textured surfaces but is
able to measure very high
slopes due to its ability to
incorporate external
illumination.
LSCM can measure
both rough and smooth
surfaces and provides
the highest possible
lateral resolutions for a
given lateral
magnification.
CSI provides a consistent axial
measurement resolution across
all lateral magnifications, a
distinct advantage over FV and
LSCM.CSI is suited for the
measurement of both low and
moderate roughness surfaces.
7 Micromanufacturing Lab, I.I.T. Kanpur
FV, CSI, and LSCM require the physical scanning of the objective lens to
obtain surface topography.
The cost and size of this mechanical scanning apparatus which involves
microscope or dedicated interference objectives as well as a camera are
significant drawbacks.
In addition the measurement process, while providing areal datasets and
thus data rich, is slow because of the requirement to capture many frames at
varying positions.
Wavelength Scanning Interferometry (WSI) dispenses with mechanical
scanning in favor of scanning the illumination wavelength. It provides a
higher measurement but the depth of field is fixed and limited, especially at
high magnifications.
3. Challenges from offline to on-machine metrology [1]
8 Micromanufacturing Lab, I.I.T. Kanpur
For any measurement technique requiring data acquisition over a fixed period,vibration can cause movement of the measurand in between camera framesleading to measurement errors.
One solution for interferometric methods is the stabilization of the measurementsystem through translation of the reference mirror position using a closed-loopfeedback system to match the vibration of the measurand.
Commercially a similar technique has been shown by Zygo with use of model-based phase shifting interferometry (MPSI) at a high frequency to stabilize theDynafiz and Verifire range of interferometers.
A beneficial side-effect of this is the video-rate areal measurement feedback thatthe MPSI provides to operators which increases ease of use and decreasessetup/sample alignment time.
A limitation of approaches such as these is that the stabilization interferometercan only correct on-axis vibration errors.
3. Challenges from offline to on-machine metrology [1]
9 Micromanufacturing Lab, I.I.T. Kanpur
Therefore the major challenges are:-
i) overcoming the low measurement rate
ii) susceptibility to vibration of metrology instruments
These have been the drivers for the development of single-shot methods whichcan broadly be defined as either spatial phase shifting interferometry, orinstantaneous interferometry.
Division of the wave front into multiple reference and measurement pathswhich experience unique phase shifts, usually by polarization, allowsencoding of the phase information spatially instead of temporally.
Deflectometry and fringe projection provide measurement through theprojection of a pattern of known shape on specular and diffuse surfacesrespectively.
3. Challenges from offline to on-machine metrology [1]
10 Micromanufacturing Lab, I.I.T. Kanpur
While of lower resolution than interferometric methods, microdeflectometry hasbeen demonstrated for OMM of specular diamond turned samples.
While these methods may be unsuitable for measurement of roughness and smallfeatures they have potential to offer excellent value for money for formmeasurement of tens of microns and upwards due to the simplicity of the apparatusused.
The requirement for compact measurement instrumentation in many on-machineapplications often forces compromise of instrument function to facilitate OMM.
Limited work volumes associated with hybrid machining centers may in somecases exclude the use of areal surface topography measuring instruments altogetherdue to their large physical size.
3. Challenges from offline to on-machine metrology [1]
11 Micromanufacturing Lab, I.I.T. Kanpur
More limited measurement sensors may still provide useful metrology forcontrolling processes.
Moving toward single-point measurement, the availability of miniature fiber-linked probes, interrogated by remotely located optical instrumentationbecomes attractive when the limited machining volumes excludes use ofbulkier areal methods of metrology.
The availability of precision motion stages for machining encourages use ofmethods such as
i) MWLI (Multi Wavelength Interferometry)
ii) CCS (Confocal Chromatic Sensors)
iii) LCI (Low-coherence interferometry)
iv) DRI (Dispersed Reference Interferometry)
for surface topography and dimensional assessment of manufactured products.
3. Challenges from offline to on-machine metrology [1]
12 Micromanufacturing Lab, I.I.T. Kanpur
The last decade in techniques surrounding Spectral Interferometry (SI) for
displacement and surface topography.
The primary advantage of SI techniques is that they do not require the
mechanical scanning of an objective lens unlike for instance, Scanning
White Light Interferometry, Focus Variation (FV), And Laser Scanning
Confocal Microscopy (LSCM) techniques.
All SI techniques work by analyzing sets of spectral interferograms, each of
which is generated at a distinct wavelength of light. The method of generating
spectral interferograms is apparatus dependent but generally involves either:
i) Changing the source wavelength over time, while capturing a number of
interferograms sequentially; or
ii) Using a spectrometer/monochromator to separate broadband interfered
light into discrete components at discrete wavelengths.
4. Case study: on-machine measurement using Dispersed Reference Interferometry [1]
13 Micromanufacturing Lab, I.I.T. Kanpur
Advantages of SI:-
• The dynamics of methods of sweeping source wavelengths and/orreading out from detector arrays are such that it is possible to acquire therequired data relatively quickly compared to those techniques employingmechanical scanning.
• This can be a useful attribute because it enables scope for higher speedmeasurement; not only attractive for many general measurementapplications, but also a key factor for implementing effective OMMbecause of the requirement to minimize the interruption of themanufacturing process.
• Furthermore, it can allow for the reduction in the overall footprint whichis of importance in applications having constrained working volumes.
4. Case study: on-machine measurement using Dispersed Reference Interferometry [1]
14 Micromanufacturing Lab, I.I.T. Kanpur
Fig. Idealized spectral interferograms generated by a DRI having a chromatic
constant dispersion with respect to wavenumber [7-10].
DRI is a form of SI.
In DRI, spectral interferograms are generated that feature a stationaryphase point which occurs at an equalization wavelength located wherethe interferometer optical path difference (OPD) is zero.
Analysis of the spectral interferograms from DRI allows thedetermination of the OPD in absolute terms, with previously reportedworks across a variety of implementations demonstrating submicronresolutions [7, 9].
4. Case study: on-machine measurement Using Dispersed Reference Interferometry
15 Micromanufacturing Lab, I.I.T. Kanpur
Figure shows a simplified Michelson interferometer having dispersion applied
in the reference arm. The DRI system which implements the necessary
chromatic dispersion by using a pair of matched diffraction gratings. The above
Michelson interferometer has dispersion applied by two matched transmission
gratings (G1, G2) in the reference arm formed between the beam splitter (BS)
and mirror, M2.
The gratings are identical, separated by a
distance, l and aligned to ensure parallelism. As
the beam exits G1 it is angularly dispersed,
resulting in a wavenumber dependent path length
r(k) resulting from the diffracted angle,
θ = arcsin(2π/kD)
Here D is the grating period Fig. Schematic of the DRI structure [1]
Principle of DRI
4. Case study: on-machine measurement using Dispersed Reference Interferometry
16 Micromanufacturing Lab, I.I.T. Kanpur
The wavenumber dependent path length in air produced by the grating pair isthen
After passage though grating G2, the beam is re-collimated and
retroreflected by the reference mirror, M2. The total dispersion produced is
doubled by the round trip of the reference beam through the grating pair. The
overall OPD in the interferometer is then simply the difference between the
OPDs in the reference and measurement arms.
The wavenumber dependent interference intensity, neglecting any spatialdependence in the axial plane for simplicity, is then,
where the interferometer phase is φ(k)=k*OPD and V is the fringevisibility[1].
4. Case study: on-machine measurement using Dispersed Reference Interferometry
17 Micromanufacturing Lab, I.I.T. Kanpur
In the case where the dispersion produced by the gratings G1 and G2 is
approximately linear the resulting phase function is parabolic with a vertex
appearing at the wavenumber, at which the OPD is zero.
The occurrence of this vertex, or stationary phase point, manifests as a point of
symmetry in the resulting interferograms as shown in Fig.
It is possible to measure the modulation of OPD by surface topography if a
suitable objective lens is used to focus the measurement beam onto the
measurand.
Fig. Idealized spectral interferogram (blue) and normalized
auto convolution result (red) [1].
4. Case study: on-machine measurement Using Dispersed Reference Interferometry
18 Micromanufacturing Lab, I.I.T. Kanpur
Fig. Idealized spectral interferograms showing fringe evolution for consecutive OPD
increases of several nm [1].
Fig. Signal processing steps for spectral interferograms detection using template
matching [1].
4. Case study: on-machine measurement Using Dispersed Reference Interferometry
19 Micromanufacturing Lab, I.I.T. Kanpur
DRI system working
Figure in next slide shows the arrangement of a DRI instrument using thediscussed "remote probe” topology.
The instrument comprises an interrogation interferometer, which consists of themajority of the optical components, and a simple probe that is situated in theworking volume close to the measurand.
The instrument operates as a “single-point” measurement system, i.e., thedatasets built up sequentially with kinematics being provided by the motionstages in the machine tool.
Fig. Structure of the fiber-linked probe [1].
4. Case study: on-machine measurement Using Dispersed Reference Interferometry
20 Micromanufacturing Lab, I.I.T. Kanpur
DRI System Design
Fig. Schematic representation of the DRI with fiber-linked probe [1].
4. Case study: on-machine measurement Using Dispersed Reference Interferometry
21 Micromanufacturing Lab, I.I.T. Kanpur
Light from a fiber-coupled super-luminescent diode (center wavelength 853 nm,
linewidth 57.5 nm, optical power 5 5 mW) is incident on an evanescent coupler
and propagates through to the remote fiber-linked probe.
At the point the light exits the fiber Fresnel reflection, due to the index mismatch
at the fiber/air interface, results in approximately 4% of the light power being
retro-reflected back into the fiber.
The rest of the light enters free space where it is first collimated and then incident
on an aspheric objective lens (effective focal length 56.24, effective NA 0.32).
On reflection from the measurand the returning light is collected by the objective
and then propagates back to the interrogation interferometer; this forms the
effective DRI measurement arm.
4. Case study: on-machine measurement Using Dispersed Reference Interferometry
22 Micromanufacturing Lab, I.I.T. Kanpur
Once the reference and measurement light returns to the interrogationinterferometer it enters a Michelson interferometer with two armsformed between the BS and the mirrors M1 and M2 respectively.
One of these arms contains the disperser formed from two gratingswhich is essential to the operation of DRI.
The resulting spectral interferogram is then captured by a spectrometerformed from the grating (G3), spherical mirror, and an 8192 pixelCMOS line array detector.
Fig. Spectral interferogram captured from the DRI instrument with fiber-linked probe
[1].
4. Case study: on-machine measurement Using Dispersed Reference Interferometry
23 Micromanufacturing Lab, I.I.T. Kanpur
DRI System Specification
4. Case study: on-machine measurement Using Dispersed Reference Interferometry
24
• The maximum axial measurement resolution is based upon noise
floor of the instrument which was determined experimentally to be
1.6 nm.
• The measurement rate possible is defined by the maximum readout
rate for the Charge-coupled device (CCD) line array (12 kHz).
• This assumes sufficient computational power is available to execute
real-time signal processing and that in the described apparatus the
optical power is sufficient to maintain this measurement rate.
Micromanufacturing Lab, I.I.T. Kanpur
Machine Tool Calibration For On-machine metrology(OMM), the measurement probe is carried by the
machine tool axes to cover the inspection area. Due to mechanical imperfections,
wear of machine tool elements, and stage misalignments, the deviation from the
programmed scanning path will induce additional measurement errors [11]. The
influence of machine tool kinematic errors on measurement results needs to be
modelled, measured, and compensated.
Fig. Error kinematic chain for on-machine
measurement system [1].
4. Case study: on-machine measurement Using Dispersed Reference Interferometry
25
Kinematic error modelling in machine
tools is based on rigid body kinematic
[12] and multibody system theory. For
the three-axis turning configuration in
the current work, there are two
kinematic error chains shown in Figure.
Micromanufacturing Lab, I.I.T. Kanpur
Based on rigid body kinematics, transformation matrix describes the coordinate transformation from coordinate k to coordinate j, which comprises four component matrices and can be formulated as:
The overall configuration of the machine tool coordinate systems is shown in next slide.
location
transformation
matrixlocation error
transformation matrix
motion (translation or
rotation) transformation
matrix
motion (translation or
rotation) error
transformation matrix
4. Case study: on-machine measurement Using Dispersed Reference Interferometry
26 Micromanufacturing Lab, I.I.T. Kanpur
Fig. Configuration of the machine tool coordinate systems [1]
4. Case study: on-machine measurement Using Dispersed Reference Interferometry
27 Micromanufacturing Lab, I.I.T. Kanpur
Fig. Machine kinematic error map [1]
Based on the error
measurement and kinematic
model established, the
machine tool kinematic error
as shown in Fig
Four selected error components
are considered as primary factors
affecting the OMM results in the
sensitive Z direction. The X axis
straightness in the Z direction EZX,
squareness error between X axis
and C axis EBOC, C axis axial error
EZC, and C axis tilt error EBC.
4. Case study: on-machine measurement Using Dispersed Reference Interferometry
28 Micromanufacturing Lab, I.I.T. Kanpur
Measuring conditions vary with machine configuration, probing system
setup, and measurement task.
Calibration of the OMM system is thus considered to be a task-specific
process [13].
Three aspects of this calibration process are taken into consideration and
discussed in the following sections: on-machine vibration test, machine
kinematic error compensation, and linearity error calibration.
5. DRI On-Machine Calibration
29 Micromanufacturing Lab, I.I.T. Kanpur
30
[1]
Micromanufacturing Lab, I.I.T. Kanpur
On-Machine Vibration Test It is necessary to conduct on-machine vibration testing and analysis to assess its
relationship with the sampling frequency, scanning parameters, and filtration
operations in post processing.
The induced vibration components onto the OMM result should be filtered out for
accurate characterization of the surface form and topography.
The relationship between λtopo and Ftopo is described as follows:
λTopo=𝐹𝑒𝑒𝑑𝑟𝑎𝑡𝑒
𝐹𝑇𝑜𝑝𝑜
where λtopo is the wavelength of the surface topography of interest and Ftopo is the
corresponding frequency
5. DRI On-Machine Calibration
31 Micromanufacturing Lab, I.I.T. Kanpur
According to the topography band of interest and vibration test results, afrequency decision graph is plotted , providing guidance in selection of theproper scanning parameters and sampling frequency.
Fig. Sampling frequency decision graph [1].
5. DRI On-Machine Calibration
32
For a given scanning
feedrate, the topography
frequency of interest
should be lower than the
vibration frequency shown
in the hatched region.
To meet the requirement
for avoiding signal
aliasing, lower scanning
speed and higher sampling
frequency.Micromanufacturing Lab, I.I.T. Kanpur
Fig. Vibration Test Results
The vibration measurement results under different test modes are summarized in
Table.
The static vibration test was performed when the machine is in static condition,
while the scanning vibration test was performed when the machine axes moves
simultaneously to measure the sample surface.
In the table above, static vibration on the machine is nearly four times the DRI
internal noise in the laboratory environment, indicating the machine tool
environmental effect on the measurement.
5. DRI On-Machine Calibration
33 Micromanufacturing Lab, I.I.T. Kanpur
[14]
The multiple radial scanning
vibration results and frequency
analysis are shown respectively in
Fig. A and B.
Selection of the DRI sampling
frequency should meet the
requirement for inspection of the
bandwidth of interest on the scale
limited surface and avoid signal
aliasing.
The spectrum analysis in figure B
indicates the primary vibration
components are less than 100 Hz
and the sampling frequency of DRI
probe is consequently set to be 200
Hz.
5. DRI On-Machine Calibration
Fig.A & B. Scanning vibration test [1].
34 Micromanufacturing Lab, I.I.T. Kanpur
Machine Tool Kinematic Error Compensation
Fig. DRI measurement (A), scanning error map (B), and Fisba measurement (C), of optical flat [1].
To validate the proposed kinematics error compensation method, a commercial
optical flat was adopted and measured on-machine.
The probe was scanned over the sample in a spiral path with C axis rotational
speed of 1 rpm and X axis feed rate of 2 mm/min.
The flat was measured offline on a calibrated Twyman-Green interferometer
and this offline result was regarded as the accurate representation of the flat
surface form. The measurement results and scanning error map are shown in
Fig.
5. DRI On-Machine Calibration
35 Micromanufacturing Lab, I.I.T. Kanpur
Fig. DRI measurement (A), versus combination of scanning error
and Fisba measurement (B)[1].
The similarity of two results in Figure below indicates that DRI OMM is the
superposition of machine kinematic error and flat form error.
Using this approach the characterized flatness error from OMM reduced from
17.3 to 11.4 nm, compared with results of the calibrated offline measurement of
8.7 nm.
The optical flat measurement by DRI on-machine and offline Twyman-Green
interferometer indicates that the kinematics error compensation can effectively
increase the accuracy of the OMM system.
5. DRI On-Machine Calibration
36 Micromanufacturing Lab, I.I.T. Kanpur
Amplification Coefficient and Linearity Error Correction
To further analyze and improve the OMM performance it is necessary to
calibrate the response curve of the instrument in the machine tool environment.
The linearity error is defined as the maximum deviation of the instrument
response curve from the linear fitted curve where the slope is the
amplification coefficient [15].
The artefact is designed with four nominal step heights (1, 2, 4, and 8 μm) to
cover the necessary working range in the Z direction.
By fitting a first order polynomial curve to the characterization results of the
different step heights, the linearity errors and amplification coefficient are
consequently derived.
5. DRI On-Machine Calibration
37 Micromanufacturing Lab, I.I.T. Kanpur
Mean step height and repeatability is reported over all radial profiles with threerepeated measurements. Measurement error δerror is defined as the differencebetween multiple step height value of OMM and that of offline calibrated whitelight interferometer (Talysurf CCI 3000, Taylor Hobson).
The CCI result was also employed as the calibrated values to correct the DRIlinearity error.
Fig. Flow chart of step height characterization [1].
5. DRI On-Machine Calibration
38 Micromanufacturing Lab, I.I.T. Kanpur
Fig.(A) and (B) show the uncorrected and corrected error plot for the step height
measurement.
The error bars represent the measurement repeatability calculated as the standard
deviation of the mean values.
After calibration, slope correction coefficient was 1.0123 and the linearity error
was reduced from 93 to 14 nm.
5. DRI On-Machine Calibration
Fig. Uncorrected (A) and corrected (B) error plot of the step height measurement [1]
39 Micromanufacturing Lab, I.I.T. Kanpur
6. Application of DRI to Diamond Turning Machine
Fig. OMM system setup [1].
• The DRI probe is integrated onto an ultra precision diamond turning
machine, equipped with two linear hydrostatic axes and a high precision air
bearing spindle.
• The DRI fiber-linked probe is mounted on to a manual adjustment stage for
the purpose of the alignment process and the DRI measurement setup is
installed beside the diamond tool holder on the Z axis.
40 Micromanufacturing Lab, I.I.T. Kanpur
Before OMM operation, the DRI probe is aligned coaxially to the spindle
rotational axis by means of multiple scanning of a convex sphere sample.
The fitted apex point can be considered as the coaxial position and is saved
in the machine tool coordinate system.
Three measurement paths (multiple radial, multiple circular, and spiral)
with corresponding applicable surfaces are illustrated in Fig.
Fig. Multiple radial, multiple circular, spiral measurement paths [1].
6. Application of DRI to Diamond Turning Machine
41 Micromanufacturing Lab, I.I.T. Kanpur
To evaluate the performance of DRI OMM and effectiveness of the machining-
measurement closed loop process, experimental work is carried out.
A 2D cosine curve (Z =Acos 2π/λX ) with A=5 μm and λ=2.5 mm was fabricated
on an aluminum sample, followed by the DRI OMM.
Six measurement profiles were spaced across the surface at equal angles (30
degrees) as shown in figure.
Fig. Multiple radial scanning measurement [1].
6. Application of DRI to Diamond Turning Machine
42 Micromanufacturing Lab, I.I.T. Kanpur
In order to find out the correlation between the online and offline measurements,
offline measurements of the machined sample were carried out using a calibrated
stylus PGI profilometer (Talysurf PGI, Taylor Hobson). For comparison, the 0o
profile of DRI OMM was extracted.
Fig A and B in next slide shows the DRI on-machine and PGI offline
measurements, respectively.
It is observed that the DRI OMM agrees well with the PGI offline measurement
in terms of form evaluation.
The derived form error also has similar shape and the characterization
parameters difference is less than 10%, described in Table.
6. Application of DRI to Diamond Turning Machine
43 Micromanufacturing Lab, I.I.T. Kanpur
Fig. Measurement results and error analysis of DRI OMM (A) and PGI
offline measurement (B) [1].
Fig. Comparison of Form Error Characterization [1]
6. Application of DRI to Diamond Turning Machine
44 Micromanufacturing Lab, I.I.T. Kanpur
For the purpose of similarity quantification, Pearson’s correlationcoefficient was employed as a measure of correlation between the twoprofiles measurements, which is described as [16]
Where Sx and Sy are the sample standard deviation. The calculatedcoefficient between DRI OMM and PGI offline measurement is P50.991(close to 1), which indicates the DRI OMM result agrees well with theresult of PGI offline measurement.
6. Application of DRI to Diamond Turning Machine
45 Micromanufacturing Lab, I.I.T. Kanpur
Procedure of corrective machining
The proposed procedure of corrective machining is illustrated in figure in next
slide.
After the initial cutting process, the sample surface was measured on-machine
by DRI probe scanning.
Subsequently, multiple radial measurement results were compared with CAD
design model and 2D compensation tool path was subsequently generated based
on the measured error profiles.
The corrective machining was then executed until the error was less than the
tolerance value.
6. Application of DRI to Diamond Turning Machine
47 Micromanufacturing Lab, I.I.T. Kanpur
Fig. Flow chart of profile error correction process [1].
6. Application of DRI to Diamond Turning Machine
48 Micromanufacturing Lab, I.I.T. Kanpur
Profile error before and after
correction process was derived
and compared, as shown Fig.
The experiment results in
Table, show that the profile
accuracy was improved from
104.7 nm (RMS) and 495.2
nm (PV), to 58.6 nm (RMS)
and 257.6 nm (PV).
Table: Profile Error Correction Results [1]Fig. Profile error correction results[1]
6. Application of DRI to Diamond Turning Machine
49 Micromanufacturing Lab, I.I.T. Kanpur
7. Conclusions
➢ The concept of surface metrology and optical methods for OMM was
reviewed.
➢ The applications of optical methods for OMM and the calibration of machine
tools was discussed. From the case study of the application of DRI for OMM
on a diamond turning machine the differences between OMM and measurement
in an optical lab were presented.
➢ OMM allows us to avoid time-consuming realignment operations and possible
damage during the transportation between machine and measurement platform.
➢ For OMM, the machine static and motion vibration tend to induce additional
error of measurement results.
➢ Proper selection of the sampling frequency and scanning parameters is
important to achieve a better on-machine measurement result.
➢ Other techniques besides DRI such as MWLI, CCS, LCI can also be studied
for application to OMM.
50 Micromanufacturing Lab, I.I.T. Kanpur
1
[1] Luo, Xichun, and Yi Qin. "Hybrid Machining: Theory, Methods, and Case
Studies." (2018).
[2] D, W., Surfaces and their Measurement Surfaces and their Measurement, 2004
[3] L. Blunt, X. Jiang, Advanced Techniques for Assessment Surface Topography:
Development of a Basis for 3D Surface Texture Standards “Surfstand”., Elsevier,
Amsterdam, 2003.
[4] ISO, Geometrical Product Specifications (GPS)--Surface Texture: Profile
Method--Terms, Definitions and Surface Texture Parameters. 1997: International
Organization for Standardization.
[5] J. Raja, B. Muralikrishnan, S. Fu, Recent advances in separation of roughness,
waviness and form, Precision Eng. 26 (2) (2002) 222235.
[6] S. Texture, Surface roughness, waviness and lay, Am. Soc. Mech. Eng. Standard
B 46 (1995) 1.
[7] P. Hlubina, Dispersive white-light spectral interferometry to measure distances
and displacements, Opt. Communicat. 212 (1-3) (2002) 6570.
8. References
Micromanufacturing Lab, I.I.T. Kanpur
1
8. References
[8] P. Hlubina, I. Gurov, Spectral interferograms including the equalization
wavelengths processed by autoconvolution method, SPIE Proc. 5064 (2002)
198205.
[9] P. Pavlı´cek, G. Ha ¨usler, White-light interferometer with dispersion: an
accurate fiberoptic sensor for the measurement of distance, Appl. Opt. 44 (15)
(2005) 29782983.
[10] H. Martin, X. Jiang, Dispersed reference interferometry, CIRP Ann. Manuf.
Technol. (2013)
[11] R. Ramesh, M. Mannan, A. Poo, Error compensation in machine tools—a
review: part I:geometric, cutting-force induced and fixture-dependent errors, Int. J.
Mach. Tools Manuf.40 (9) (2000) 12351256.
[12] A. Okafor, Y.M. Ertekin, Derivation of machine tool error models and error
compensationprocedure for three axes vertical machining center using rigid body
kinematics, Int. J.Mach. Tools Manuf. 40 (8) (2000) 11991213.
Micromanufacturing Lab, I.I.T. Kanpur
1
[13] R. Wilhelm, R. Hocken, H. Schwenke, Task specific uncertainty in
coordinate measurement, CIRP Ann. Manuf. Technol. 50 (2) (2001) 553563.
[14] J. Williamson, H. Martin, X. Jiang, High resolution position measurement
from dispersed reference interferometry using template matching, Opt. Exp. 24
(9) (2016) 1010310114.
[15] C.L. Giusca, R.K. Leach, F. Helery, Calibration of the scales of areal
surface topography
measuring instruments: part 2. Amplification, linearity and squareness, Meas.
Sci. Technol. 23 (6) (2012) 065005
[16] A. Stuart, Kendall, M.G., The advanced theory of statistics, Charles
Griffin, 1968
8. References
Micromanufacturing Lab, I.I.T. Kanpur
55
Thank You
Micromanufacturing Lab, I.I.T. Kanpur