69
Chapter 32 Operating Room Fires and Personnel Injuries Related to Sources of Ignition P.908 Although flammable anesthetics have disappeared from operating rooms, perioperative fires continue to occur. They can have devastating consequences, precipitate legal action, and take a great psychological toll on everyone involved. They usually come as a complete surprise to the staff ( 1 , 2 , 3 , 4 ). Most operating room fires are of little consequence and are not reported, making the actual incidence difficult to determine ( 2 , 5 , 6 ). Approximately 10% to 20% of reported fires result in serious patient injury ( 7 , 8 ). Fuels present in the operating room include plastics that produce dense black smoke when ignited. The smoke may contain toxins and may hinder safe evacuation of the patient and staff from the room. Sudden ignition can present secondary problems ( 9 ). It can cause a startle reflex, causing the surgeon's hand to jerk and potentially cut into unintended tissue or set other areas of the surgical site on fire. The Fire Triangle For a fire to occur, there must be three factors (fire triangle or triad) present: an ignition source, a fuel, and an oxidizer to support combustion together in the proper proportions and under the right conditions ( Fig. 32.1 ).

Chapter 32. or Fires & Personnel Injuries

  • Upload
    adham08

  • View
    224

  • Download
    0

Embed Size (px)

Citation preview

Chapter 32 Operating Room Fires and Personnel Injuries Related to Sources of Ignition P.908

Al though f lammable anesthet ics have disappeared from operating rooms,

perioperat ive f i res cont inue to occur. They can have devastat ing consequences,

precipi ta te legal act ion, and take a great psychological to ll on everyone involved.

They usua lly come as a comple te surprise to the staff (1,2,3,4).

Most operat ing room fi res are o f l i t tle consequence and are not reported, making

the actual inc idence dif f icul t to determine (2,5 ,6). Approx imately 10% to 20% of

reported f i res resu lt in serious patient injury (7,8). Fuels present in the operat ing

room include plast ics tha t produce dense black smoke when ignited. The smoke

may conta in toxins and may h inder safe evacuation of the pat ient and staff f rom the

room. Sudden igni tion can present secondary problems (9). It can cause a start le

ref lex , caus ing the surgeon's hand to je rk and potentially cut into unin tended t issue

or set other a reas of the surgical s ite on fi re .

The Fire Triangle For a f i re to occur, there must be three fac tors (f ire tr iangle or triad) p resent: an

igni tion source, a fuel , and an oxid izer to support combustion together in the proper

proport ions and under the right condi t ions (Fig. 32.1).

View Figure

Figure 32.1 Three things are necessary for a fire to occur: an ignition source, fuel, and an oxidizer.

Ignition Source There are a number of igni tion sources in the operat ing room. Most are under the

surgeon's control.

Lasers The acronym laser ( l ight ampl if ication by s timulated emission of radiat ion) def ines

the process by which a form of energy is converted into l ight energy

(10,11,12,13,14,15,16). The term can also refer to the device tha t produces the

l ight or to the l ight i tself .

Lasers use a col imated, coherent, monochromatic , intense beam of electromagnetic

radiation to cut, coagulate, or vaporize tissue.

The rate that the laser energy is delivered is ca lled power and is measured in

watts . The wattage is equal to the amount of energy, measured in joules, d iv ided by

the duration of exposure measured in seconds. Laser power density is the amount

of power d istr ibu ted wi thin an area and is indicated in watts per square cent imeter.

Components of a Laser System The basic components of a laser system are two parallel mirrors encompassing the

laser medium and a power source (Fig. 32.2). In add it ion, there may be an aiming

beam for de livering laser energy tha t is outs ide the v isible range.

Laser Medium

The medium (head) holds the subs tance energized to produce laser l igh t. The

medium, wh ich may be a solid, liquid, o r gas, determines the wavelength of the

emitted radiation . The laser is named for the material used as the medium.

Power Source The power (pumping, pump, exci ta tion) source supplies energy to the laser med ium

to c reate the l ight. When the power source is act ivated, energy is absorbed by

electrons of the atoms in the laser medium, which are e levated to energy levels

above their ground state. They then decay to lower energy levels and emit photons

that are not in phase with one another and travel in all di rections.

Optical Cavity The optical (resonator) cavi ty p rov ides the envi ronment in which the laser medium

is conf ined. Energy released from the medium travels in al l di rect ions. Mirrors are

used to ref lec t and increase the energy of emission. One of the mirrors is no t 100%

ref lec tive and allows a small port ion of the l ight to escape.

Light Guide A light guide (delivery system) di rec ts the laser beam to the surgical s ite (Fig.

32.3). Fiber-optic bundles are a convenient f lex ible condui t for some wavelengths .

Other lasers ut i l ize a hol low tube wi th mirrors aligned to reflect the beam f rom i ts

source through the focusing lens .

Aiming Beam With certa in lasers, it is necessary to use a low-powered v is ib le beam as a marker.

This is usua lly a helium-neon gas laser. I t passes th rough the same optical pa th as

the laser beam.

P.909

View Figure

Figure 32.2 Components of a laser.

Laser Types Types of med ical lasers commonly used in surgery are shown in Table 32 .1.

Laser Hazards Lasers can cause s ignif icant damage to opera ting room personne l (inc lud ing the

anesthesia provider) and the pat ient. Hazards involv ing lasers include f ires ,

atmospheric contamination, eye damage, organ or vessel perforation, and embolism

(10). Figure 32.4 shows a s ign used to warn personne l that a laser is in use.

View Figure

Figure 32.3 This is a schematic representation of a CO2 laser guide as might be found in either an operating microscope or a handheld wand. The guide consists of hollow tubes with hinged, aligned mirrors that reflect the beam from its source through the focusing lens.

A recurrent problem wi th lasers is that they may be activated when not in use

(22,23). Often, a foot switch accidentally act iva tes the laser. The laser may be in a

posit ion where this is not no ticed for some time. When the laser is not actually in

use, i t should be p laced in the STAND-BY mode.

A f i re may resu lt when a laser beam hits a fuel or the laser f iber becomes

damaged. Ign ition can be almost ins tan taneous. Whi le mos t ignition sources must

be in contact with a material to cause ign it ion, a laser can supp ly heat to a fuel

f rom a few centimeters to severa l meters away, so i t can ignite material c lose to the

t ip, at some d istance from the tip , or under several layers of o ther material . A

surface drape can be penetrated by the laser but not igni te . Materials under the

drape may then igni te and burn wi thout being not iced for several minutes (24). The

laser beam can be ref lec ted f rom a metal surface, causing a burn or ign it ing

material in a remote loca tion.

Laser Risk Classification A classif ication system ref lecting laser risk to the patient and personnel has been

developed (25,26,27). The higher the c lass, the more stringent the protec tion

needed.

P.910

TABLE 32.1 Commonly Used Lasers and Associated Personnel Hazards

Laser Medium

Features Potential Hazards to the Eye

Special Considerations

CO2 Readily absorbed by all biologic materials, independent of pigmentation. Tissue destruction is proportional to its water content. Produces a very superficial tissue effect.

Injury to the eye will be confined to the cornea. There is no risk to the retina. Since the laser is absorbed by plastic and glass, ordinary eyeglasses with sideguards can be used for eye protection.

Fires involving both tracheal tubes and supraglottic devices have been reported with this laser.

Nd-YAG

Can be transmitted through fiber-optic fibers. Poorly absorbed by water but well absorbed by pigmented tissue (11).

Retinal damage can occur. Opaque green eyewear or eyewear with clear lenses with a special coating should be worn.

Because it is taken up by pigment, colored markings on tracheal tubes are more likely to be damaged than clear portions (12). Blood or mucus on or in the tracheal tube makes the tube less resistant to the laser beam. Fires have been reported from an Nd-YAG laser passed through the channel of a flexible bronchoscope (17,18). The rigid bronchoscope is recommended for use with this laser, although the flexible scope may be needed to treat hard-to-reach areas (19).

KTP Passes through clear substances but is absorbed by hemoglobin and other pigments (11,20,22).

Retinal damage may occur. Special eyewear with red filter should be worn.

Argon Beam is selectively absorbed by red, orange, and yellow pigment and strongly absorbed by hemoglobin and

Retinal damage may occur. Special opaque orange goggles/eyewear should be worn.

melanin. Fiber-optic bundles can be used to transmit the laser beam.

From Klarr P. Laser complications. In: Atlee J, ed. Complications in Anesthesia. Philadelphia: WB Saunders, 1999:588–590.

Class 1 Lasers Lasers that are totally enclosed or that emit extremely low output fall into c lass 1.

These are safe to v iew.

Class 2 Lasers Low-risk lasers are in c lass 2. Their r isk is approximate ly equivalent to staring at

the sun or o ther brigh t lights that can cause centra l re tinal injury. These are not

hazardous unless someone overcomes their natura l aversion response to bright

l ight.

Class 3 Lasers Class 3 lasers operate at a power level above 1 mW. This is a hazard even if

v iewed only momentari ly . Class 3 lasers are subdiv ided into a and b subclasses .

The 3a subclass cons is ts of lasers in the 1 to 5 mw power range. These lasers pose

a moderate ocular hazard. Class 3b inc ludes lasers that emit between 5 and 500

mW of output power. Even momentary v iewing of these lasers is potential ly

hazardous to the eye. They may also be a hazard to skin.

P.911

View Figure

Figure 32.4 Note that the laser class is on the sign.

Class 4 Lasers Any cont inuous wave laser wi th a power output above 500 mW is in c lass 4 . These

lasers pose serious skin, eye, and f i re hazards.

Igni tion risk is re lated to a number of fac tors beside the laser c lassif ication . Class

3b and 4 lasers are cons idered to be igni tion hazards. The igni t ion risk of other

c lasses of lasers wi l l depend on how they are focused, the t ime of exposure, and

the oxidizers and fuels tha t are present.

Metal instruments can become overheated with pro longed laser use and could

cause a burn (28).

Electrosurgery Unit Electrosurgery (radio frequency) apparatus is so widely used that there is of ten

complacency about reading the instruct ion manual or following safe pract ices

(29,30). Electrosurgery units are the mos t common igni tion source in surgica l f i res

(6). The instruction manua l should be s tudied at the beginning of a tra ining program

and reviewed periodica lly. A brief set of operat ing instruc tions should be readi ly

availab le on the instrument (31).

Terminology The vocabulary concerning electrosurgery is somewhat confusing (32). This chapter

wi l l def ine e lec trosurgery devices as those that employ a h igh-f requency (radio

f requency) electric current passing through t issues to cut, coagulate, or prov ide a

blend of cutting and coagu lat ion. Elec trocautery ut i l izes a heated wire or blade,

usual ly at the end of a probe fo r coagulation . Most elec trocautery devices are

battery-opera ted.

Physics There are three modes avai lable on modern electrosurgery devices .

Cutting Mode The cutt ing mode uses a continuous s ine wave. Heating is rap id and produces h igh

temperatures that explode cel ls that come in contact wi th the electrode. This

produces a cut in the tissue.

Coagulation Mode The coagu lat ion mode features short bursts of a dampened s ine wave. Less heat is

produced s ince the current is not cont inuous. Tissue desiccat ion occurs , and the

ends of blood vessels are thermally sealed .

Blend Mode The blend (cut-coagulation) mode employs a waveform and voltage between tha t of

the cutting and coagu la tion cycle . The rela tive amount of cutting or coagulat ion

depends on the lengths of the bursts of voltage. Severa l blend se tt ings may be

availab le on an elec trosurgical uni t (33).

Types Monopolar With monopolar (unipolar) electrosurgery un its , e lectrons f low f rom the genera tor to

the act ive electrode and through the patient to the dispers ive (return , inact ive)

electrode, sometimes incorrect ly called the grounding pad (Fig. 32.5). The curren t

disperses from high densi ty at the tip of the act ive electrode to a lower dens ity v ia

the lowest impedance pathway tha t it can f ind .

Generator The elec trically powered generator (power uni t) c rea tes a radio frequency current

that passes through the active electrode. The effect on tissue depends on the

current intensi ty, the type of current employed, the durat ion of exposure to the

current, the hand piece used, and the dispers ive electrode (34). Coagulat ion

requires longer contact wi th the probe than cutt ing. As electrica l energy is appl ied

to the t issue, char forms. Since char has h igher impedance, more current d iss ipates

di rectly to the dispers ive elec trode and the ef fect is reduced (34).

Active Electrode

The active elec trode has a handle and is connected to the generator by a f lex ible

wi re. It d irects current f low to the surgical si te. The tip of the act ive electrode may

be a blade, spatula, ball , lancet, loop, forceps , need le po int, or suct ion tip. I t may

be disposab le or reusable. The t ip

P.912

configuration determines i ts abil i ty to desiccate tissue. A needle-tip electrode

concentrates more power on t issue contact than does a paddle or ba ll -shaped

electrode (34). At a given electrode vol tage, the paddle-shaped t ip p resents a

higher power at the edge than on the flat port ion. The electrode may be activa ted

by a hand contro l or foot switch. The control may also be used to switch f rom one

mode to another.

View Figure

Figure 32.5 Monopolar electrosurgery unit.

Dispersive Electrode The dispers ive (inactive, return) electrode (grounding or ground electrode, pla te, or

pad) col lects the current from the patient and returns it to the generator to

comple te the c i rcu it . The large contact surface of the pad provides low impedance.

Most pads present ly in use are prejel led. Reusab le metal p lates used in conjunct ion

wi th conduc tive ge l are rarely used today. With some electrosurgery uni ts , two

dispers ive e lec trodes should be used (35).

Safety System

Modern electrosurgical uni ts possess a return e lec trode (contac t quali ty) moni to ring

system. This ensures tha t the patient is connec ted to the return e lec trode (34,36).

Current f lowing to the ac tive electrode is measured and compared with current

returning f rom the d ispers ive electrode. If the curren ts are not ba lanced, the uni t is

deactivated. A more recent innovation is active electrode monitoring (AEM)

(37,38,39,40). This continuously moni tors the electrical c ircui t and automatically

shuts down the generator and sounds an ala rm if dangerous e lectrical leakage

occurs. Newer generators can sense dramatic changes in t issue impedance or

temperature changes at the return electrode (33,41). Some modern electrosurgery

machines are equipped wi th “spl i t pads” (42). The uni t wi l l not func tion unless both

halves of the pad are in contact wi th the patient.

Bipolar The bipo lar electrosurgery unit is composed o f a generator and two electrodes

located wi th in mill imeters of each other. The energy f low is be tween the two

electrodes , and no current f lows through the patient 's body. This al lows the energy

to be local ized more precise ly. The current densi ty in the t issues surrounding the

ac tive electrode is substant ially less than that for monopolar elec trodes, and

deeper t issue layers are preserved wi th lower vol tage and power requirements. This

type of e lec trosurgery un it can coagulate even when the tip is immersed in b lood.

Bipolar units cannot be used fo r cutting .

Bipolar elec trodes are produced in a variety o f conf igurat ions: hooks , spatulas ,

sc issors, forceps , and needles (29 ,43). Of ten, these devices are batte ry-opera ted.

There are reports that the ON-OFF swi tch on one of these dev ices did not always

deactivate the dev ice when it was set aside (44 ,45). The hot t ip could then set

drapes on f i re. At least one device could be act ivated by another device's wire less

remote-control uni t or by f luorescent room l ights (46).

Hazards Associated with Electrosurgery Mishaps with the elec trosurgery uni t of ten involve faul ty re turn electrodes, improper

electrode placement, or al te rnate low-impedance outlets. Insuff ic ient contact wi th

the dispersive pad can resul t in pa tient burns and/or burns to others who are in

contact wi th the pat ient. Among the low-impedance al te rnatives are

electrocardiogram (ECG) monitoring elec trodes, temperature probes, urinary

ca theters, metall ic parts of surgica l tables, heating pads, and oximeters (47).

About 68% of reported surgical f i res involve e lectrosurgical equ ipment (48). Fires

associated wi th the electrosurgery unit are of ten associated with an oxidizer-

enriched atmosphere. When the electrosurgery uni t is used, heat vaporizes t issue

or expels tissue embers from the t ip. In room air, t issue vapors do not igni te, and

embers are quick ly extinguished. In an oxidizer-enriched atmosphere, the vapors

can ignite into a brief f lame, and embers can fly several inches and burn unt i l they

are consumed. A f la re of evolved gases can d irectly ignite any convenient fuel . The

electrical wires associa ted wi th the electrosurgery electrode may develop a short

c i rcui t, wh ich could resul t in a f ire (49).

During laparoscopic surgery, burns that are sometimes fatal can occur ou tside the

surgeon's v iew when using an elec trosurgery uni t (14,29,33,38,40,50,51,52). These

are usua lly caused by insu lat ion fai lure, di rect coupling, or capacit ive coup ling.

I f a defect occurs in the active electrode's insulation , current can f low through the

defec t. Defec ts can resul t f rom improper handling or damage to both reusable and

disposable electrodes . A number of devices are avai lable to detect insula tion

defec ts (33,40,51). Act ive electrode monitoring cont inuously monitors the elec trical

c i rcui t and automatical ly shuts down the generator and sounds an alarm if

dangerous elec trical leakage occurs (37,38,39,40).

Capaci ta tive coupling occurs when the electrosurgical current induces s tray

currents on other nearby conductors . It can occur even through intact insulat ion . It

occurs as a result of the active e lectrode contacting or coming into very c lose

proximity to a non-insula ted metal ins trument (such as a laparoscope) in the

surgical field. If the electrical ly-energized laparoscope has been placed through a

metal cannula, the current wi l l s imply f low to the pat ient's abdominal wal l . However,

i f the laparoscope has been placed th rough a nonconduct ive cannu la, the current

cannot f low to the abdomina l wal l. In th is case, any piece of bowel o r other internal

t issue touch ing the laparoscope can be burned. Capaci tative coupling is always

present and cannot be to tal ly eliminated.

Argon Beam Coagulator The argon beam (enhanced) coagulator (ABC) is solely a coagu la ting instrument.

Rad io f requency monopo lar curren t is delivered through a f low of ionized argon

gas.

P.913

The f low is al te red as the power is changed. The t ip does not touch the tissue. I f

the tip is greater than 1 cm from the t issue surface, only a gent le s tream of argon

wi l l f low. When the tip is 1 cm or less f rom the tissue surface, the ac tive mode

occurs. If the nozzle t ip glows red, the tip is too close to the t issue or the power

se tt ing is too high.

Fiber-optic Illumination System A f iber-optic i l luminat ion system consis ts of a l ight source and l ight-transmitting

cable . The cable is connected to an endoscope or headligh t. The term cold l ight ,

used to describe l ight f rom a f iber-opt ic source, is incorrec tly assumed by many to

mean that heat is not generated. Actual ly co ld l ight refers to l ight in which the

amount of inf ra red radiat ion has been reduced (53). These light sources can

provide several hundred watts of v is ible, infra red, and ul trav iolet light. Al though

some of these wavelengths can be f i l tered out, the power is typical ly focused into a

f iber-opt ic cable of smal l d iameter tha t can deliver a h igh-power densi ty .

A number of f ires have been igni ted when the unprotected end of the f iber-optic

cable was allowed to res t on a drape (6 ,49 ,53,54,55,56,57). The cable end can

reta in a signi ficant amount of heat a fter be ing d isconnected from the l ight source.

Defibrillator When a def ibri l la tor is act ivated, a spark may be generated if insuff icient force is

applied to the paddles; if the paddle pad is too smal l; if paddles are applied over an

i rregular surface or bony prominence or near an ECG electrode; when insuff ic ient,

excess or the wrong kind of gel is used; or if there is another conductive medium

between the paddles (6 ,58 ,59). I f disposable def ibri l la tion pads are used to

increase e lec trical conduction between the paddle and the patien t, an arc can occur

i f the surface of the paddle is not comple tely on the pad, if the pad is smal ler than

the padd le, if there is a fold in the pad, o r if the pad is dry.

Pressure Regulators When gas is al lowed to f low f rom a high-pressure to a low-pressure chamber,

recompression can cause a rap id rise in temperature. Materials that cannot

wi thstand both 100% oxygen and high temperatures wil l igni te . This hazard is

associated mostly with aluminum oxygen regula tors (60,61,62,63,64,65,66,67,68).

Another cause of heat in a pressure regu la tor is part ic le impact f rom contaminants.

Tef lon tape, chips f rom seal materials , or hydrocarbon contaminants may be

present (69).

Surgical Lights Surgical l ights can be a source of igni tion. I f the light does not have the proper

mechanism to diss ipa te inf rared radia tion or i f that mechanism fails , the patient o r

personnel envi ronment can be exposed to energy levels high enough to cause

burns or f i res (70,71). The heat f rom a surgical l ight can contac t and rupture a hose

f rom the medical gas pipel ine system (72).

Electrical Faults An electrica l f i re can occur in any env ironment where there is electrical equipment.

A short c i rcui t can occur in an anesthesia mach ine (73,74,75). In one reported

case, a short circu it in a laryngoscope wi th a rechargeable handle caused f lames to

shoot f rom the charging end of the handle (76). Elec trical arc ing in surg ical booms

has been reported to cause fi res (77,78).

Other Ignit ion Sources Other reported sources of ignit ion in opera ting room f i res include resectoscopes,

heat lamps, heated probes , pneumatic tourniquets, denta l and orthopedic burs and

dril ls , heated-wire breath ing tubes, and humid if ie rs (1,79,80,81,82,83,84).

Fuels Fuels abound in the operating room. A fuel is anything that can burn, inc luding

most things tha t come in contact with pat ients as we ll as the patients themselves.

Tracheal Tubes The risk and characteris t ics of a tracheal tube f ire wil l depend on the type of tube

used. In most cases , igni tion requires the tube to be penetra ted (85). The f ire

begins on the ins ide rim of the penetrat ion area and then spreads both wi th and

against the flow of ox id iz ing gas.

Polyvinyl Chloride Tubes Polyv inyl ch loride (PVC) tubes are combus tible in an ox idizer-enriched atmosphere

wi th a carbon d ioxide (CO2) laser. Once igni ted and penetrated, a PVC tube can

susta in a torchlike f lame. PVC tubes wi thout markings are re lat ively res is tant to the

Nd-YAG laser, but mark ings inc rease the risk (86). I f there is blood, mucus , or

sa liva on the tube, the risk of f i re is inc reased. The inju ries associated wi th PVC

tube f i res are more severe than wi th other types of tubes (87,88).

Red Rubber Tubes Red rubber tubes are combus tible with the CO2 , potassium t itanyl phosphate (KTP),

and Nd-YAG lasers (86,87,88,89,90). A major p rob lem is the inabil ity to see

through the tube. Should an intraluminal f i re develop, it may go undetec ted for a

longer period of time than with other tubes .

Damage to the tracheobronchial tree af te r an in traluminal f ire may be less severe

wi th a red rubber tube than a PVC tube (87). Red rubber tubes are less l ikely to

soften, deform, or fragment if ignited than are o ther tubes. Another advantage is

that they can be quickly

P.914

removed f rom the pat ient (91). The high pressure in the cuff causes it to def late

more quickly than is possible with a low-pressure cuff .

Silicone Tubes A sil icone tube is more resistant to penetration by a CO2 laser than other tubes

(87). If ign ited, a s i l icone tube rapidly becomes a bri tt le ash tha t crumbles eas ily

and may be aspira ted , ra is ing the possibil i ty of future problems with s il icosis

(87,90,92,93). However, the acute injuries are less severe than with red rubber and

PVC tubes (88).

Laser-resistant Tubes A number of ready-to-use laser-resistant tubes and tube wraps are avai lable and

are desc ribed in Chapter 19. They should meet the requirements of Inte rnational

Standards Organ ization (ISO) 14408 (93). It is importan t to remember tha t laser-

resistant does not mean laser-proof . Laser-resistant tubes can igni te, especial ly i f

manufac turer's warnings, precautions, or directions fo r use are not followed. If a

laser tube is used with a laser other than that for wh ich i t was designed or a

f lammable part such as the cuf f is exposed, it wi l l catch f ire (92 ,94). Laser-resistan t

tubes are usually not res is tan t to other heat sources such as an electrocautery

penci l.

Supraglottic Airway Devices The relative resistance o f suprag lott ic ai rway dev ices depends on their

composi tion . Those devices tha t are made of s i licone rubber such as the LMA-

Classic, LMA-Flexible , and LMA-ProSeal are s ign if icant ly more resis tant to the CO2

laser than are PVC tracheal tubes (95). Many disposable supraglottic dev ices are

made of PVC and are probably as suscept ible to fi res as tracheal tubes of the same

composi tion .

Surgical Products Surgical drapes, towe ls , and dressings are common fue ls for opera ting room f i res .

Of ten , the drape is igni ted f rom another f ire (24). Whi le many drapes are resis tan t

to ignit ion in room a ir, the oxidizer-enriched a tmosphere that is often present may

cause them to burn with v igor (49,96,97,98,99,100,101,102,103,104).

Disposable drapes may be part icularly diff icul t to deal wi th during a f i re because

they are water repellen t. Addi tion of water may even spread the f lame. Laser-

resistant drapes are avai lab le . Some synthetic drapes wi l l mel t away f rom the laser

rather than igni te (103).

There are numerous reports of surgical sponges, gauze pads, and swabs being

igni ted (49,105,106,107,108,109,110,111,112,113). Wet sponges may be used to

protect a tracheal tube from a laser beam. However, if they are allowed to dry out,

they become f lammable.

Adhesive Substances Adhesive tape can be the fuel for a fi re (114). Co llodion and benzo in are

f lammable .

Skin Preparatory Solutions Flammable volat i le organic so lut ions are often used to prepare the sk in prior to

surgery. If the l iqu id or vapor is contacted by an igni tion source , a flame can resu lt

(1,101,115,116,117,118,119,120,121). Alcohol-based solut ions are espec ia lly

f lammable . They can also in fi l t rate and pool under the drapes. When this happens,

they are slow to dissipate. Alcohol-based f ires have a f lame tha t is d if f icult to see.

Of ten , the result of the f lame is the f i rs t thing not iced.

Intestinal Gases Gases that accumulate in the bowel (especial ly hydrogen, hydrogen sulf ide , and

methane) can be the fuel for a fi re o r exp los ion if an elec trosurgery uni t is used to

open the bowel (105). Proper bowel preparation may prevent this (122). Suff ic ient

oxygen to support combustion is normally not present in intestina l gas . However,

ni trous oxide may diffuse into the intest inal lumen, c reating a f lammable mixture.

Oxygen Cannulas Oxygen cannulas are made of plas tic that can burn . They carry 100% oxygen. If

touched by a hot e lec trosurgery probe or a laser beam, they burn readily. Even if

not di rec tly contac ted, the area around them may be so oxygen-rich that a smal l

spark can turn in to a burning ember tha t can igni te the cannula.

Lubricants and Ointments

Petro leum-based ointments that are used in an oxidizer-enriched a tmosphere wi l l

igni te when enough heat is present to cause vaporizat ion (105,123,124). Water-

based lubricants wi l l not burn and can be used to coat ha ir to make i t f i re -resistant.

Body Hair Body hair, including eyelashes, moustaches, and beards can be involved in a f i re

(105,106,125,126). Coating the hair with a water-based lubricant wil l decrease the

risk .

Surface flame propagation occurs where there are fine surface f ibers of fabric o r

body hair (127). In the presence of an ox id izer, these f ine f ibers can be igni ted.

Of ten , the skin or underlying fabric is not burned. The surface f i re races in the

di rection of the oxygen source, where the oxygen supply tubing may be ignited

(58).

Other Combustible Substances A number of artic les used in or near the patient can serve as the f lammable

material . These include (but are not l imited to ) oxygen tubings, endoscopes, smoke

evacuator hoses, esophageal s te thoscopes, breath ing tubes, reservoir bags, eye

patches, s ten ts , masks, nasogastric tubes, en teric feeding tubes, rubber and plas tic

nasopharyngea l ai rways, covers, paper products, blood pressure cuffs , aeroso l

adhesives, tourniquets, gloves, s tethoscope tubing , th roat packs, eggcrate foam

P.915

mattresses, bandages, stockinettes, d ressings, pil lows, glue, gowns, straps, caps

and hoods , rubber elec trosurgica l uni t probe sheaths, shoe covers, loca l anesthetic

spray, and organic gas f rom a necro tic tumor

(106,116,128,129,130,131,132,133,134,135,136,137).

Alcohol-based hand sani tizers (gels, foams, and liqu ids ) are h ighly f lammable

(138). They shou ld be stored in a cabinet that is designed fo r flammable materials .

However, a study pub lished in 2003 found no fi res attribu tab le to hand sanit izers,

and the National Fire Protec tion Associat ion now a llows alcohol-based hand

sanit izer d ispensers in corridors and other public a reas, provided certain

restric t ions are observed (139). I f a dispenser is mounted in a hallway, the corridor

must be at leas t 6 fee t wide. The dispenser can project up to 6 inches f rom the wal l

and mus t be instal led above handrai l height.

Oxidizers

With an oxidizer-enriched atmosphere , a f i re igni tes easier, burns more v igorously,

spreads more rapidly, and is more d if f icu lt to extinguish . The oxidizers of g reates t

inte res t are oxygen and ni trous oxide. Ni trous oxide supports combustion and in the

process releases the energy of i ts formation, providing increased heat. Thus , any

mixture of oxygen and ni trous oxide wil l support combustion. Air wi l l a lso support

combustion, because it contains oxygen.

Because oxygen is heavier than ai r, i t collec ts in low-lying areas, inc lud ing drape

folds. Some materia ls such as drapes and towels absorb oxygen and reta in i t fo r

some t ime (1,127). Tenting drapes wi l l al low oxygen to dra in off toward the f loor

and be d iluted by ai r c i rculation.

Common Scenarios

Airway Fires During airway surgery, all of the th ree necessary components are in close

proximity: a combus tible subs tance (trachea l tube, gauze, etc.), an igni t ion source

(laser or e lectrosurgery apparatus), and gas to support combustion (oxygen wi th or

wi thout ni trous oxide). An ai rway f i re is part icularly serious because a considerab le

amount of heat is genera ted in a smal l area, and the smoke and gases f rom such

f ires can be b lown deep in to the pat ient's lungs .

Using Electrosurgery during Tracheostomy Many a irway f i res have occurred while using the electrosurgery uni t during

tracheos tomy

(9,105,128,130,140,141,142,143,144,145,146,147,148,149,150,151,152,153,154).

Of ten , the patien t is being given 100% oxygen in ant ic ipa tion of interrupted

venti lation and/or because of the underly ing c linical cond ition. If the tracheal tube

or cuff is contacted, an oxidizer wi l l be released. Current ly , there is no tracheal

tube that is safe for use wi th electrosurgical devices or electrocautery, al though

some are more resistant to igni tion than others (155).

Using Electrosurgery in the Mouth Using electrosurgery in the oral cavi ty may resul t in a f i re

(6,156,157,158,159,160,161,162). The tracheal tube is of ten the fuel , but o ther

i tems such as sponges or patient t issue may be involved. Of ten, wet gauze is used

in the throat to catch sec ret ions or to pro tect the tracheal tube. If there is a leak

around the tube, the gauze can becomes oxidizer-enriched and dry ou t more

quick ly.

Laser-induced Tracheal Tube Fires The l ikelihood that a laser wil l contact the tracheal tube during ai rway procedures

is high (163). Cuffed tubes are more l ikely to be contacted than noncuffed tubes.

The tube may be exposed to ei ther the di rec t or ref lec ted laser beam. Flaming

t issue in c lose proximity to the tube may cause it to ignite (164).

In addi tion to local damage in the larynx, injury can occur to the lower a irway and

the parenchymal t issue in the lung. The products of combustion may be b lown into

the lungs.

The cuff is the most vulnerable part of the tracheal tube. It cannot be wrapped, and

laser-resistan t tubes do not have laser-res is tan t cuffs .

Wet cotton sponges are of ten used around the tube to protect the cuff f rom the

laser beam. If these dry out, they can become fuel for a f ire (165). Tape that is

used to secure the tube can igni te.

During lower ai rway surgery using a laser, the cable may be passed through the

inside of a tracheal tube. The tracheal tube should be placed just below the vocal

cords so that the t ip is as fa r away f rom the operat ive s ite as possible.

Other Head and Neck Fires Head and neck f i res can occur during local anesthes ia procedures when

electrosurgery apparatus is used (6,49,56,126,166,167,168,169,170,171,172,173).

Of ten , oxygen is being administe red, and i t may diffuse into the surg ical area. Any

f lammable i tem in the v ic ini ty can catch f ire.

Fires Involving Pressure Regulators While no t a frequent occurrence, a f ire in a regulator can be d isastrous (174).

These fi res may resul t f rom adiabat ic compress ion or partic le ignit ion in which

debris is b lown f rom the cyl inder into the regulator wi th suffic ient energy to cause

igni tion. Improper pressure regulator assembly may also cause a f i re.

P.916

Regula tors wi th aluminum components are more l ike ly to burn than other

regulators . The Food and Drug Adminis trat ion (FDA) and the National Ins ti tute for

Occupational Safe ty and Heal th (NIOSH) now recommend that aluminum regulators

be replaced wi th ones wi th brass components (64,66). A number of aluminum

oxygen regulators have been recalled (67).

Intraperitoneal Fires Severa l explosions have been reported during laparoscopy when nitrous oxide was

used as the insuff lating gas (122,175,176,177). Ni trous oxide in the inspired gases

wi l l dif fuse into the peritoneal cavi ty and may reach a high enough concentrat ion to

support combustion (178,179). Fires have also resul ted f rom inf lat ing the peri tonea l

cavity wi th an oxygen-CO2 mixture instead of pure CO2 (180,181). The Pin Index

Safety System (Chapter 1) wi l l not p revent this error, because the index ho les for

tanks wi th 100% CO2 are the same as for any tank wi th greater than 7% CO2 .

Fires during Defibrillation Oxygen is always used during resusci tation, and the source is of ten disconnected

and p laced near the head (58,182,183). This may allow oxygen to f low over the

defibri l la tion electrodes (184). A spark caused by placement of the def ibri l lato r

paddle or electrode close to an ECG elec trode or poor contact be tween the

patient 's sk in and the def ibri l lator paddle or electrode may ignite hair or other

combustible materials in the area (183,185).

Fires in Canisters Fires in CO2 absorbent canisters have been reported

(186,187,188,189,190,191,192,193). These are d iscussed in more detail in Chapter

9.

Personnel Risks

Smoke Hazards Potent ial ly hazardous ai rborne contaminants are formed from tissue vaporization

when a laser is used. These by-products have the potent ial to be mutagenic,

carc inogenic, teratogenic, or a vector for v iral infect ion (10,194). Elec trosurgery

devices and bone saws also produce vaporized f ragments that have the same

hazards (25).

Inhaling potent ia lly hazardous ai rborne contaminants can be kept to a minimum by

scavenging the smoke. This wi l l help to give the surgeon a c learer v iew o f the

surgical field and removes o ffensive and irrita ting odors. Various smoke evacuators

are avai lable (103,195). To be most effect ive, the evacuator nozzle should be

posit ioned as c lose as possible to the operative f ie ld and be func tiona l before,

during and for 30 seconds after t issue is vaporized (196).

Personnel masks should be removed and discarded as soon as poss ib le af te r use

(196). Standard surgical masks f i l ter ou t partic les down to about 3 microns in

diameter. The laser plume has partic les as small as 0.31 microns . High-eff ic iency

masks can f il te r partic les down to 0.30 mic rons, bu t they are dif f icul t to breathe

through and lose their eff iciency when wet (10,103).

Ocular Injuries The laser beam may be di rected over some distance. The area of hazard may not

be l imi ted to the immediate surgical area and may present risk to operating room

personnel (25). The surgeon is not highly susceptible to injury because of the

safety dev ices bui lt into the ins trument. However, if the ins trument is accidental ly

ac tivated, he is exposed to the same hazards as other personnel .

The eye is especially vulnerable to injury. Because CO2 laser beams are absorbed

wi th in the fi rs t 200 µm of t issue, they are a hazard to the cornea (197). Argon, KTP,

and Nd-YAG lasers are more like ly to cause damage to the retina . The ex tent of

damage wil l depend on wh ich part of the retina is affected. Permanent v isual loss

may resul t (25).

All personnel in the area where a laser is used should wear appropriate protect ive

eyewear. This may inc lude goggles, face shields, spectacles , or presc ription

eyewear using special materials or reflec tive coatings (or a combination of both),

se lec ted to reduce the potent ial ocular exposure. The safety eyewear should be

marked according to the wavelength (laser type) (Fig. 32.6). I t should be noted that

laser safety eyewear is not des igned fo r looking di rectly at a laser beam (16). The

patient 's eyes should be c losed and covered wi th saline-soaked gauze or a

nonshiny metal shield . Awake patients should wear goggles specif ic for the laser

being used. Windows in the room should be covered and a warning sign placed

P.917

on the operat ing room door at eye level (Fig. 32.4). Spare eyeware should be

availab le at a ll entrances.

View Figure

Figure 32.6 All laser-protective eyewear should be clearly and permanently labeled.

Electrical Shock Another concern is electrical shock from the e lec trosurgery device. A common

pract ice among surgeons is to al low an assistant to touch the dista l end of a

forceps wi th the act ive electrode f rom the electrosurgery apparatus. This act ion

makes the forceps the active electrode. A glove may not offer suffic ient p rotec tion

f rom the electrical c ircui t . Holes appear in as many as 15% of g loves.

There are ways to minimize the shock potential wh ile using electrosurgery. The

ac tive electrode should be f irmly in contact wi th the forceps before i t is energized.

The person holding the forceps should try to contact i t in as large an area as

possible to spread the concentration of the charge.

Fire Prevention The best way to deal with fi res is to prevent them f rom start ing (127,198). Fi res can

be avoided by not allowing the th ree elements of the f ire tr iangle to come together

at the same t ime.

Controlling Ignition Sources One way to reduce f ires is to properly use equipment that might ac t as an igni tion

source. I t is important to read instruct ion manuals and to alter techniques to f i t

these instructions.

Lasers

Personnel who use lasers should take a cert if ied laser course (6,23,26,103). They

should practice their techn ique on inanimate ob jec ts and gain experience a t the

hands of an experienced c linic ian. Specif ic personnel should be designated to

moni tor and prevent laser hazards.

A laser protocol should be developed and fo llowed. Lasers should be test f ired onto

a safe surface before use to ensure tha t the beams are properly al igned. Lasers

should always be kept in the standby mode except when needed. The lowest power

densi ty and shortest pu lse dura tion that wi l l do the job should be used. The laser

should be activated only when the tip is under the surgeon's d irect v is ion. The laser

should be placed in standby mode before removing i t f rom the surg ical s i te.

Laser f ibers should not be c lipped to drapes because the f ibers may be broken. The

laser fiber should be careful ly passed through an endoscope before in troducing the

scope into the pat ient to minimize the risk of f iber damage. The f iber's funct ionali ty

should be verif ied before insert ing the endoscope into the pat ient.

I f a laser contact t ip is used, the t ip wil l remain hot for several seconds. Bl ind entry

into cavi ties wi th this tip should be avoided (199). The risk of tracheal tube igni tion

wi l l be lowered if the laser is activated during the expiratory pause (200).

Electrosurgery Unit The elec trosurgery uni t should not be used during tracheostomy

(31,34,50,111,171,201,202,203,204). Bleeding should not be trea ted wi th

electrosurgery af ter f lammable dressings have been appl ied.

A bipolar electrosurgery uni t should be used whenever possible . This wi l l reduce

the curren t densi ty in the t issues surrounding the ac tive electrode and minimize the

potential for di rect coupl ing. Bipolar units work at lower voltages and are not

associated wi th insulation fai lure .

Monopolar un its should have a means to moni tor the return electrode or ac tive

electrode. Us ing active e lec trode monitoring wi l l min imize the risk of s tray

electrosurgical burns (38). Before each use, the electrosurgery uni t and assoc ia ted

safety features should be inspected for s igns of damage and tested to ensure that

the uni t is functioning correct ly. If i t is not work ing properly or is damaged, the

electrosurgical apparatus should be immediately removed f rom serv ice. The uni t

should be pro tec ted f rom sp il ls . Unintentional act ivation may occur i f f lu ids enter

the generator. The alarm system should be checked prior to applying the dispers ive

electrode. The act ive electrode should have a t ip that is secure, because a loose

t ip may cause a spark. Prejel led dispers ive electrode pads should be checked for

uniform gel dis tr ibution and absence of exposed wires before being applied to the

patient. Outdated or previously opened but unused dispers ive e lectrode pads

should not be used, because the gel can undergo electrolys is and/or desicca tion. If

a dispers ive elec trode pad requiring gel is used, the pad shou ld be checked

careful ly to identify any dry spots on i ts surface before it is placed on the patient.

Af ter the patient has been posi tioned, the connect ion between the pat ient and the

unit should be establ ished by p lac ing the dispers ive electrode on a nonhairy a rea of

c lean, dry sk in. The dispers ive electrode should be the appropria te s ize for a

patient ( i.e. , neonate, infant, ped iatric , adult) and never be cu t to reduce its size.

The dispers ive electrode should be app lied to c lean, dry skin over a large, wel l-

perfused muscle mass as c lose to the operative s i te as practica l, avoiding bony

prominences, scar t issue, skin over a metal prosthesis, o r dis tal to a tourniquet.

Excessive hair shou ld be removed before apply ing the dispers ive e lec trode,

because hair wil l insu late the pad from the patient. The pad's ent ire surface area

should main tain uniform body contact. There should not be any ten ting, gaping, o r

moisture under the pad. This wil l interfere with adhesion to the sk in and decrease

the contac t surface. The dispers ive electrode

P.918

should not be moved to another s i te a fter ini tial appl ication. The status of the

dispers ive e lec trode should be checked if any tension is applied to the cord or i f

the pat ient is reposi tioned after the pad is appl ied. A used dispers ive electrode

should not be reapp lied.

Electrocardiographic electrodes should be placed as far as possible from the

operative si te to minimize the alternate f low o f current through the electrodes and

moni tor to g round.

I f insulated electrosurgical elec trode probes are required, only commercial ly

availab le insulated probes should be used (9). Insulating s leeves cut f rom catheters

or o ther materials shou ld no t be used to sheathe probes.

The lowes t power settings tha t are effective for the surgical procedure , as

determined by the surgeon in conjunc tion wi th the manufacturer's recommendation,

should be used. The ent i re ci rcui t should be checked if the operator requests higher

power sett ings because of ineffectual results .

The elec trosurgery uni t should be act ivated only when the tip is under the

surgeon's direc t v ision and only by the person using i t . Cords shou ld no t be

c lamped with or wound around any objec ts . The electrosurgery probe should be

kept c lean to minimize the risk of adherent t issue incandescing or f laming. The

electrosurgical penc il should be placed in a holster when i t is not in act ive use. The

electrosurgical uni t should be placed in s tandby mode whenever i t is no t in active

use. Unnecessary foo t swi tches should be removed so that they are not

accidentally ac tivated .

During laparoscopic surgery, all-metal or a ll-p lastic cannulas should be used and

not a hybrid cannula system (i.e. , a combination of plastic and metal cannulas )

(37). Electrosurgery elec trodes should not be used ins ide metal suct ion i rrigators.

The elec trosurgica l uni t should be deac tivated before removing it f rom the surgical

s i te. Even after deact ivation , the probe t ip may retain enough heat for a few

seconds to mel t plast ics or igni te some fuels, so it shou ld be placed ins ide a c lean,

wel l -insulated holder or broken off . In addi tion to protecting the t ip, this makes i t

more dif f icult to acc identally ac tivate the switch . The t ip should be broken, and a

safety cover should be placed over the tip and act ivation swi tch before the device

is discarded (205).

Argon-enhanced Coagulation All safety measures observed for the electrosurgical unit shou ld be observed for

argon-enhanced coagulation (31). The electrode should not be placed in di rect

contact wi th t issue. The hand piece should be moved away f rom the pat ient's t issue

af te r each activation. The argon gas f low and the argon coagu lator should be

ac tivated simul taneously. The argon gas l ine should be purged of ai r before each

procedure and by act ivat ing the system after moderate delays between act ivat ions.

Fiber-optic Light Sources All connec tions should be made before the l ight source is activated, because the

end of the cable can cause a f i re (53). An act ive f iber-optic cable should not be

placed on flammable material . The l ight source shou ld be turned OFF before

disconnecting the cab le. The end of a f iber-optic cable can retain a signif icant

amount of heat after be ing disconnected from the l ight source (206). Therefore, i t is

importan t to be careful where one places the end of a f iber-op tic cord.

Heated Humidifiers Only a breath ing tube and heat ing ci rcu it labeled for use with a spec if ic humid if ier

should be used (207). Heated breathing c i rcui ts should not be covered wi th sheets,

blankets, towels, clothing, o r other material. They should not rest on surfaces such

as the patient, operat ing table, blankets, o r medical equipment. Instead, a boom

arm or tube tree shou ld be used to support them. A heated-wire breathing ci rcui t

should not be turned ON before f low has been ini tiated. I f there is no f low (as

during cardiopulmonary bypass ), the heater should be turned OFF (208).

Defibrillators When us ing a defibri l lato r, care should be taken to hold the paddles f i rmly and

posit ion them properly. Disposable conduc tive pads should be larger than the meta l

surface of the paddle and be wi thin their expiration date. This wil l prevent arc ing

when the paddles are activa ted (58). ECG electrodes should be appl ied as far as

possible f rom the def ibri l lat ion pads.

Electrical Faults All electrica l cords should be regularly inspec ted for cuts and nicks in the

insulat ion, f rayed insulat ion, and loose connections at the plug or receptac le ends

(209). Al l plugs should be pushed completely in to the receptacles to prevent p rong-

to-prong arc ing (210).

Managing Fuels Tracheal Tubes

Non-intubation Techniques Since the tracheal tube is of ten the fuel for the f i re , anes thetic techn iques that do

not require a tracheal tube wi l l el iminate one component of the f i re triad (92).

These inc lude apnea, jet vent i lat ion, and spontaneous breathing.

Apnea With the apneic technique, the pat ient is vent i lated by us ing a mask or tracheal

tube, and these are withdrawn as the laser is used (92,93,211). Af ter a period of

t ime, vent i lat ion is reinstated. This is repeated as long as needed to perform the

surgery.

Insufflation Patients may be al lowed to spontaneously breathe a combinat ion of oxygen and

anesthet ic agents whi le laser procedures are undertaken. The gases are

administered through a catheter or hook in the

P.919

corner of the mouth tha t is wel l out of the way of the laser beam. The suc tion

channe l of a laryngoscope may be used to insuff late oxygen. Operat ing room

pollut ion wi th anesthet ic agents may be high wi th this technique. Oxidizers should

be used as sparingly as possible s ince the insuff lated gas cou ld support

combustion.

Jet Ventilation Jet vent i la tion techniques ei ther v ia a metal bronchoscope or a catheter inserted

through the crico thyroid membrane may be used (212). A h igh-veloci ty jet of oxygen

or helium and oxygen can be di rec ted into the ai rway above or below the glo tt is by

us ing various apparatus (21,92,93,196,213,214).

There has been a f i re reported with th is technique (215). An errant laser strike

igni ted the surgeon's glove. The resul tan t burning vapors were entra ined into the

patient 's airway. The patient's mustache was also igni ted , and the pat ient suffered

fac ial burns .

Filling the Cuff with Saline The cuff is the most vulnerable part of the tracheal tube, regardless of the tube or

cuff materia l. If a laser beam penetrates an ai r-f i l led cuff , gas can leak in to the

operative field, and if the oxygen concentrat ion is high, the risk of f i re is great.

Fluid in a cuf f ac ts as a heat s ink and makes the cuff less easy to perforate

(216,217). Fi l ling the tracheal tube cuff wi th a l idocaine jelly-plus -saline mixture no t

only prevents the cuff f rom being igni ted but may plug small holes in the cuff

resul ting f rom a laser h it (218).

I f the cuff is perfora ted, a je t of f lu id may ex tingu ish the f ire. If perfo rated, fluid-

f il led cuffs reta in their shape longer than a ir-f i l led cuffs . Care must be taken to

remove all ai r f rom the cuff , because any remaining ai r wil l sett le in the most

superior part of the cuff , which is the part most l ikely to be h it by the laser beam.

Addit ion of methylene blue or o ther biocompatible and highly v isible dye to the

sa line wi l l help the surgeon to recognize a perforated cuff . Some surgeons object to

the addi t ion of methylene blue because if the cuff is punc tured, the color of the

t issues is al tered, making it dif f icu lt to recognize pathology.

Protective Wrappings The tube can be covered wi th a protective wrapping. These are discussed in

greater de tai l in Chapter 19.

Merocel wrap (Laser-GuardTM) was found to be acceptable for surg ical levels of

CO2 , KTP, and Nd-YAG lasers (219,220,221,222). Merocel-wrapped tubes are not

more combus tible if they are coated with blood (223). Reflec ted laser beams have

not been a problem with th is wrap (224). This product is easier to apply than

metall ic tapes (225).

Aluminum and copper tapes are effective for use with the CO2 and Nd-YAG lasers if

the tube is meticulously wrapped (226,227,228). However, a f i re can occur if the

laser contacts part of the tube that is not wrapped (229,230). A wrapped red rubber

tube does not offer protection f rom the KTP laser (86). The adhesive backing or

surface coating of some tapes can be ign ited by laser beams (92). The presence of

blood on the surface of foi l-wrapped tubes makes combustion more l ikely to occur

(223). Tape-wrapped tubes may unravel, leaving portions of the tube exposed. I f

the wrapping is dislodged, i t could obstruct the airway. A wrapped tube may catch

f ire (231,232). Meta ll ic tapes may ref lect the laser to nontargeted areas (224).

Spiral wrappings may cause the tube to k ink . Any wrapping adds thickness to the

tracheal tube, which could be a problem for the surgeon.

Protecting the Cuff with Wet Covers As a fu rther p recaut ion, moist cottono ids, sponges, or pledgets can be placed on

the cuff (92,233,234). Cotton gauze is a good choice because i t s tays wet longer

than other covers and has low energy transmission (235). Wet gauze wi l l also

protect the shaft (236). These have been found to be especial ly help ful with the

CO2 laser (234). The Nd-YAG, KTP, and argon lasers may allow some energy to

penetrate the pledget and rup ture the cuff (235).

I t is important that wet covers be kept moist. Laser beam h its may dry them, and if

that happens, they lose their protect ive effect. Further hi ts can cause the

co ttonoids and/or cuff to ign ite. Al l these must be retrieved af ter surgery.

Special Tracheal Tubes Ready-to-use laser-resis tant tubes are discussed in Chapter 19. Laser res istant

means tha t the material can wi thstand l imi ted laser energy wi th l i tt le o r no damage.

The use of a special laser tube does not necessari ly prevent a f ire during laser

surgery (237). Even though the tube is labeled laser-resistant, the cuff wi l l not be

resistant, and special p recautions need to be taken to protect it . A fi re may occur i f

the pro tec tive wet gauze over the cuff dries or is displaced.

No laser-resistant tracheal tube is completely safe f rom all types of lasers under all

condi tions . All of these tubes can be damaged or igni ted by lasers for which they

are not intended or by high laser energies (238,239). Some are made more

combustible by blood (223). Tubes sold for use with lasers should indicate the type

of laser fo r which they are sui ted as we ll as the condi tions (power, power densi ty,

spot s ize, oxygen concentra tion) under which the tube is safe to use. Stric t

adherence to manufacturer's warnings and d irections is essential .

I f the Nd-YAG laser is used through a f iber-optic bronchoscope passed th rough a

tracheal tube, i t is best to use an unmarked PVC tube (240). The tracheal tube

should be placed just below the vocal cords so tha t the t ip is as far away from the

operative si te as possible. I t should be remembered tha t the cover on the f iber-

optic endoscope is plast ic and can burn.

Some laser-res is tan t tubes are more resistant to f i re than conventiona l tracheal

tubes when contacted by elec trosurgical instruments (155).

P.920

When laser contac t tips are used, there is an even greater opportuni ty for igni tion .

They can get hot enough to melt and burn plast ics even when the laser is inac tive

(241).

Double-cuff tubes are not more resistant to leaks that wil l occur af ter cuff puncture

but do al low a sea l to be maintained if one cuff remains intact. The second cuff is

s t i l l vulnerable to puncture (239).

There are disadvantages assoc ia ted wi th laser-resis tant tubes . They are more

expensive than PVC and red rubber tubes. Since these tubes are often used for

patients who have had previous laryngeal surgery, the anatomy may be dis to rted,

making intubat ion diff icult . Some laser-resistant tubes are so st if f that i t is dif f icult

to pass a stylet or use wi th a specia lized laryngoscope such as the Bu llard

laryngoscope (Chapter 18). Therefore , a diff icult intubation may become even more

challenging when these tubes are needed.

Using Smaller Tracheal Tubes Smal l -diameter trachea l tubes require higher power densit ies fo r ignit ion than

large-d iameter tubes because the higher gas f low cools smaller tubes more quick ly

than larger tubes. Also, the smal le r the tracheal tube, the less l ikely i t is to be hi t

by an igni t ion source.

Making the Tracheal Tube Easy to Remove The tube should be f ixed so that it can be removed rapidly, if necessary. If the tube

is to be removed, th is should be done immediately after igni tion to minimize

damage to the ai rway and lungs. If the tube continues to burn, i t may be very

diff icul t to remove. Some tracheal tubes break apart as they burn . If the tube is

wrapped, the wrapping may break in to pieces tha t could lodge in the airway.

Some tubes are easier to remove than others (91). PVC tubes wi th sal ine-f il led

cuffs are more diff icult to remove than red rubber tubes . The fastest way to deflate

a fluid-f i l led cuff is to remove the contents wi th a syringe rather than cu tt ing the

pi lot bal loon. Unclamping the inf lation tube on a red rubber tube wi l l result in rapid

release of the cuff contents.

Wetting Fuels The use of wet towe ls , packers, or sponges around the surgical s ite can prevent

igni tion of materials near the s i te . Gauze or sponges used with uncuffed tracheal

tubes to minimize gas leakage in to the pharynx and sponges , gauze, and pledges

(and their s tr ings) used to protec t the tracheal tube cuff should be moistened and

not allowed to dry.

Hair that is near the opera tive si te (e.g., eyebrows, beards , and moustaches)

should be made nonflammable by coating i t thoroughly wi th a water-soluble

lubricating jelly .

Proper Preparation Practices Water-based solut ions should be used to decontaminate the skin prior to surgery

whenever possible . If a lcohol-containing solut ions must be used, a minimum of the

agent should be used and appl ied l ike paint, no t in a thick, runny coat. The so lu tion

should not be applied in a manner tha t allows dripping, pooling, or wicking. I f

so lut ion drips away from the surgica l s i te , it should be immediate ly blotted with a

gauze sponge before it can soak into any absorbent material . Any soaked materials

should be removed. Daubing of p rep pooled on skin (in the umbil icus or c ricoid

notch) may be needed. Draping should be delayed to al low the solution to fully

vaporize and become d iluted in room air. Th is could take 10 minutes or longer

(1,117). Incise (adhesive, occ lus ive) drapes should be used, if possible, to iso late

head and neck inc is ions from oxygen-enriched atmospheres and from flammable

vapors beneath the drapes (120). I f the incise material does not adhere to the

patient, the prep is l ikely s ti ll wet, and the pat ien t should be redraped once the

prep is ful ly d ry.

Correct Product Choices I t is important to consider the f i re potent ial when choosing equipment. The

anesthesia provider should be aware of the c i rcumstances under which a f i re can

occur and strive to use the product that wi l l have the lowest flammabil ity under

condi tions where i t wi l l be used.

Fi re /laser-res is tan t drapes should be used when exposure to igni tion is poss ib le. It

is important that the drape is tested with the type of ignit ion source that is to be

used. A standard for testing drapes using the CO2 laser has been published (242).

Laser-resistant anesthes ia c i rcui t protectors and drapes are available. These are

aluminized to def lec t the laser beam. There is sti l l the possibi l i ty tha t the beam may

be reflec ted onto a flammable surface and start a f ire (243).

The use of a metall ic Y-piece and elbow wi l l ensure tha t tracheal tube combus tion

wi l l not spread to the anes thesia breathing system (244).

Other Measures A number of other measures may be benef ic ial in preventing a f ire. The tracheal

tube should be withdrawn to above the s i te of surgery during tracheotomy. The use

of posi tive end-expira tory pressure (PEEP) may decrease the risk of a irway f i re in

some cases (245,246). Nurs ing personnel can reduce the combustible load in the

room by removing disposable paper wrappers and covers before the start of the

case. Not only does this reduce the fuel in the room, bu t it a lso reduces the waste

that mus t be disposed as “red bag” wi th i ts higher disposal costs (2).

Minimizing Oxidizer Concentrations The f ire risk can be reduced by remov ing or iso la ting the oxidizer f rom the surgical

area or minimizing i ts concentrat ion.

P.921

During def ibri l lat ion, all sources of oxygen should be comple tely removed f rom the

area around the pat ient.

Oxygen should be adminis te red only when indicated and in no h igher concentration

than is needed (as gu ided by oxygen saturat ion monitoring).

There are several ways of supp ly ing a gas with an oxygen concentra tion greater

than 21% but less than 100% to the pat ient. An oxygen and an ai r f lowmeter can be

connec ted by using a Y-piece (Fig. 32 .7), or a blender can be used. If the machine

has an a ir f lowmeter, a mixture of oxygen in a ir can be set and del ivered to

common gas outlet (247). If the common gas outlet can be accessed, a nasal

cannula can be mated to the outlet by using a 5-mm tracheal tube connector. I f the

common gas outlet cannot be accessed, a ci rc le sys tem can be attached to the

machine, the adjustable pressure l imi ting (APL) va lve c losed, and the selector

swi tch set to BAG. A nasal cannula can then be attached to the Y-piece. This

arrangement may cause the cont inuous positive pressure alarm to be activated. A

forced-air convect ion machine can be used to prov ide a high f low o f ai r around the

patient 's head whi le a nasal cannu la delivers a low f low of 100% oxygen to the

patient.

View Figure

Figure 32.7 Gas from an oxygen flowmeter and an air flowmeter can be connected to produce a gas with an oxygen concentration of less than 100%.

I f possible , the oxygen flow should be discont inued for at leas t 1 minute before

heat-produc ing surg ical instruments are used (248,249). When the heat source is

no longer used and any t issue embers are ext inguished, oxygen administra tion can

be resumed.

During head and neck surgery, a barrier should be established between the oxygen-

enriched atmosphere beneath the drapes and the surgica l field, if possible. Th is

can often be accompl ished wi th an adhesive (incise, occlusive) drape (2,117).

When gas is used to venti late the area under the drapes and prevent CO2

accumulat ion, the lowest acceptab le concentrat ion of oxygen should be used (250).

Forming an open tent wi th the drapes wil l prevent gases from collecting and allow

oxygen to diss ipate (56 ,170). Since oxygen is s light ly heav ier than room air as long

as there is some way for the oxygen to get in and out of a space, i t wi l l tend to f low

toward the f loor. Actively scavenging the space beneath drapes wi th a suct ion

device wil l lower oxygen concentrat ions s ign if icant ly (251). Dev ices that combine

oxygen de livery wi th suction to prevent oxygen bui ldup under the drape have been

described (252,253). A fan can be used to blow the accumulated oxygen away f rom

the pat ient. Direct ing the f low o f nasa l oxygen through a nasal ai rway wil l lower the

oxygen concentration a t the sk in (254).

Hea lth care ins ti tu tions should consider purchasing anesthesia workstat ions tha t

include internal pipel ines and f lowmeters for ai r (248). This makes it easy to switch

patient venti lat ion beneath the drapes f rom oxygen to air before the use of heat-

producing surgical instruments. An ex ternal f lowmeter can be added to an

anesthesia mach ine if piped air is available (Fig. 32.7).

Ni trogen, ai r, o r hel ium may be used to reduce the inspired oxygen concentrat ion to

the lowest level that wi l l provide acceptable patien t oxygenation. Nitrous oxide

supports combus tion and should not be used as the di luent gas (255,256,257).

Helium is a good heat conduc tor and may offe r a s ligh t pro tective effect wi th the

CO2 bu t not the KTP laser (258,259). The use of hel ium may decrease the

resistance of the gases passing through the tracheal tube (260,261).

P.922

Ni trogen insuff lation can s ignif icantly reduce the inc idence of laser-induced

tracheal tube cuff f i res (262). This can be accompl ished by attaching a catheter

covered with a laser-resistant wrap to the tracheal tube. Ni trogen is then insuff lated

through the catheter. This technique should be undertaken wi th caut ion, as a

hypoxic mix ture may possibly be adminis te red.

Al though using lower inspired oxygen concentrat ions wi l l reduce the risk of igni t ion ,

i t wi l l not tota lly prevent it . If there is a s igni ficant leak around the tracheal tube

cuff , the anes thesia prov ider may f i l l the reservoir bag by pushing the oxygen f lush .

This wi l l result in an elevated oxygen concentra tion. A more appropria te response

wou ld be to inc rease the f resh gas f low wh ile main ta ining the same inspired oxygen

concentration or to replace the tracheal tube.

When elec trosurgery is used in the oral cavity , the use of a cu ffed tracheal tube wi l l

minimize the oxidizer level . Leaks around the tube may result in f ires (161,263). I f

an uncuf fed tube mus t be used, an occlusive pharyngeal pack moistened with a

nonf lammable l iquid wi l l reduce the gas f low into the oral cavi ty (264). Insuff la ting

the mouth with gases such as helium, CO2 , or n itrogen wi l l reduce the oxidizer

concentrat ion . The oropharynx can be scavenged wi th separate suct ion.

Pneumatic power tools shou ld be opera ted wi th n itrogen or ai r. Inf latable

tourniquets should be inf la ted wi th ai r or nitrogen but not oxygen or ni trous oxide

(56,79,265). Ni trous ox ide should not be used for insuff la ting the abdominal cav ity

during laparoscopy (266,267,268).

Preventing Pressure Regulator Fires Fires in regulators can be minimized by not al lowing them to become contaminated

wi th oi l , grease, or other combus tible materials o r c leaned with a f lammable agent

such as alcohol (122,269). Before a regula tor is f it ted to a cyl inder, partic les should

be cleared f rom the cylinder ou tlet by remov ing the protect ive cap or seal and

s lowly and briefly opening (“cracking”) the valve wi th the port pointed away f rom

the user and any other persons . A cyl inder valve should always be opened slowly to

al low heat to be dissipated as the gas is recompressed. Consideration should be

given to rep lacing aluminum regulators with those made f rom brass. Rules for

handling cylinders are discussed in Chapter 1.

Fire Plan Each heal th care fac il i ty should formulate a plan for deal ing quickly wi th an

operating room f ire (1 ,6,56,58,127,171,202,206,240,270,271,272,273,274). Surgica l

teams should be trained in and prac tice dri l ls for keeping minor f i res f rom gett ing

out of control and manag ing f ires that do get out of control . They should know the

locat ion and proper use of a larm boxes, gas shut-o ff valves, and fi re extinguishers.

Evacuation procedures should be reviewed periodically.

I t is important to immediately cal l for help, decide who is going to f ight the fi re ,

when i t wi l l be appropriate to leave the room, and how to care for the pat ient during

the f i re .

Burning material on or in the pat ient of ten can be ex tingu ished effec tively and

safely by hand or wi th a nonflammable liqu id (e .g ., saline f rom a bas in on the scrub

table) or wet c loth. With laser surgery in the a irway, a t least two syringes f il led with

sodium chloride should a lways be available to ex tingu ish the f ire (275). Larger

areas can be smothered with a blanket or wet towe l. Fi re b lankets are not

recommended (8 ,276). If the drape is water-resis tant, water poured on i t wi l l be

ineffective (277). The f ire wi l l burn on the unders ide. The only effect ive technique

is to pul l the burning materials from the patient.

Fires tha t involve electrica l components are best handled by disconnect ing the

device from i ts electric power supply and removing the dev ice f rom the room.

The f low o f oxygen, n itrous oxide, or ai r to any involved equipment should be

turned OFF, if this can be accompl ished wi thout injury to personnel or the pat ient.

Most pat ients can tolerate short periods of oxygen deprivation . The patient should

be venti lated with ai r and intravenous agents used to mainta in anesthesia unt i l all

possible sources of f i re or reigni t ion are suppressed. It may be necessary to c lose

the oxygen and ni trous oxide shut-off valves to the affected room (58,278,279).

Cyl inder valves should be c losed and all cylinders removed from the area. If

supplemental oxygen is necessary to oxygenate the pat ient, the patient should be

moved to the nearest area where oxygen can be safely used (56).

I f an ai rway f i re occurs, the breath ing system shou ld be disconnected from the

tracheal tube to stop gas f low. If this s i te is not easi ly accessible or the operator

may be burned in the process o f disconnecting the tube, the breathing system can

be disconnected at the absorber (280).

The trachea l tube should no t be unduly secured so that i t can be removed quick ly.

Al though immediate removal of the tracheal tube and protective devices is usua lly

recommended wi th an ai rway f i re , th is may not always be the bes t course of act ion

(148,281). The risk:benefi t ratio of extubat ion needs to be considered. A stylet or

ai rway exchange catheter can be used to subst i tute a new tube the burned tube for.

I f the decis ion to remove the tube has been made, sal ine-f i lled cuffs on plast ic

tubes can be def la ted faster by aspira ting the f luid than by cutt ing the p ilot tube

(91). With red rubber tubes, unclamping the pilot tube is fastest.

P.923

The airway should be re-establ ished and the patient vent ilated wi th ai r unt i l i t is

certain that nothing remains burning . Then, 100% oxygen should be administered.

A search for f ragments tha t remain in the trachea and assessment of damage to the

larynx and tracheobronchial tree shou ld be made.

Extinguishing a Fire Although mos t operat ing rooms have a sprink ler system, sprinklers are frequently

not effec tive wi th operat ing room f ires (206). This is because the sprink le rs are

rare ly located over the operating room table and are heat ac tivated. Operating room

f ires tend to give off a lot of smoke and tox ic produc ts bu t not necessarily enough

heat to immediately activate the sprinklers . Therefore, i t is important the all

operating room personnel are fami liar wi th f i re extinguishers .

Location Strategically placed portable fi re ext inguishers are an important part of any f ire

protect ion program (1 ,282,283). They shou ld be located in conspicuous, readi ly

access ib le loca tions that a re known by al l operat ing room staff (F ig. 32.8).

Hands-on experience with f i re ex tingu ishers is essent ial to a f ire safety education

program (284). Sett ing up real f i res in a park ing lot and prov iding practice wi th

ex tinguishers can have valuable payoffs during an actual f i re.

Use Use of any f ire extinguisher can be recalled by the acronym, pass . Pul l the p in. Aim

the horn or nozzle a t the base of the f ire. Squeeze the trigger o r handle wh ile

holding the ext inguisher upright. Sweep the ext inguisher from side to side, covering

the area of the f i re. Ins truct ions are found on the f ire ext inguisher (F ig. 32.9). The

pin and hose are shown in F igures 32.10 and 32.11.

Choice of Extinguisher The choice of ex tingu isher should be based on the pred ic ted f ire risk for that

locat ion (282,285,286). Fi res are c lassif ied according to their fuel. Class A f i res

include small masses of ordinary combus tibles—-c lo th, paper, and most plas tics.

Fi res involv ing f lammable l iqu ids or grease are in c lass B. Electrical ly energized

f ires are c lass C. Two standard sets of f i re c lass labels exist: a tradi tional le ttering

system and a newer universal pictograph system tha t does not require the user to

be famil iar wi th the alphabet f ire c lass codes (F ig. 32.9).

View Figure

Figure 32.8 Fire extinguishers should be in conspicuous locations, well known to all operating room staff.

Halogenated Agents Halogenated ext inguishing agents are l iquids that upon discharge under relatively

high pressure surround a f ire with a combinat ion of vapor and l iquid droplets (mist).

Halogenated agents are ideal ly sui ted for f ighting f ires involv ing electronic

equipment and f i res in areas where electrical or electronic equipment is nearby.

Halon was the agent of choice fo r operating room f ires fo r many years (202). I t has

the abi li ty to penetrate burning material and to smother and cool the chemical

react ion of the fi re . While usual ly marked for c lass B and C f ires, i t is also effec tive

against class A f ires found in the opera ting room. Unfortunately, because of

environmental concerns, halon ext inguishers are no longer avai lable . If present in

an opera ting room, they may be used but cannot be recharged. The National Fi re

Protect ion Association requires that nonrechargeable

P.924

P.925

ex tinguishers be removed from serv ice 12 years from the date of manufacture

(287). In response to the prohibi tions on the use of ha lon , a number of

ex tinguishing agents have been developed to rep lace it (282,288,289) (Fig. 32 .12).

View Figure

Figure 32.9 The instructions for using the fire extinguisher are printed on the extinguisher.

View Figure

Figure 32.10 The pin that needs to be pulled to activate the extinguisher is in the handle at the top of the extinguisher. The hose is normally attached to the side of the extinguisher. This figure also shows the pressure gauge, which indicates that the extinguisher is usable, and the monthly checkout list, which shows that the extinguisher has been checked, the extinguisher and hose is undamaged, and the pressure is in the use range.

View Figure

Figure 32.11 Close-up of pin and pressure gauge at top of extinguisher.

View Figure

Figure 32.12 There are several substitutes for halon available.

Carbon Dioxide CO2 ex tinguishers emit a fog of CO2 gas with l iquid and solid partic les that rapidly

vaporize to cool and smother a f i re. Although they carry a B and C rat ing, they can

be used to ex tinguish small class A fi res. CO2 f ire ext inguishers are heav ier and

bulk ier than halon-type devices . CO2 ext inguishers do not leave a residue but do

emit l iquid and sol id components that rap idly vaporize . This causes rapid cool ing

that might cause f reezing i f i t comes into contact wi th exposed skin or internal

organs (283,290).

Water Pressurized-water ext inguishers are chiefly ef fec tive against c lass A f ires . They are

more dif f icult to use than CO2 and halon-type extinguishers. To put ou t fi res on

water-repel lant d rapes, a f inger mus t be placed over the nozzle to produce a f ine

spray. A s tream or splash of water could fan the f i re (1). A non-magnetic water mis t

ex tinguisher is avai lable for use in the MRI unit magnet room.

View Figure

Figure 32.13 All operating room personnel must know where the fire alarm is located. The instructions for activating it are printed on the fire alarm.

I f the f i re starts to get out of control , the f i re alarm (Fig . 32.13) should be sounded,

and the f ire department should be not if ied to give f iref igh ters ample time to respond

(1). F iref igh ters would ra ther f ind that a small f i re has been ext inguished than see

smoke bil lowing out of a bui ld ing—especially in a health care faci l i ty .

The patient and staff should be evacuated if the f i re and smoke are excessive

(210). In some s i tua tions , however, it may be more hazardous to move the pat ient

than to a ttempt to ext inguish or contain the f i re. The attending physician must

determine which would present the lesser hazard .

The doors should be c losed to contain the smoke and isola te the f i re . Wet sheets or

towels can be stuffed under the doors to reduce the spread of smoke f rom the

room.

Whatever steps are necessary should be taken to protect or evacuate pat ients in

adjacent a reas.

Af ter the f ire, involved materials and dev ices should remain in place for the f ire

department invest igators . Notes should be made as soon as poss ible af te r the f ire

and an inc ident report comple ted.

P.926

References 1. Anonymous. Understanding the f i re hazard. Techno l Anes th 1992;12(9):1–6.

2. Lees DE. Opera ting room f i res : s t i l l a problem? ASA Newslett 2002;66:33–34.

3. Anonymous. JCAHO warns about surgical f ires . Outpat ien t Surgery Magazine

2003;4:8 .

4. Nationa l Fi re Protect ion Association. Fi re safety in heal th care fac il i t ies. Quincy,

MA: Author, 2003.

5. Rodriguez JG, Satt in RW. Injuries as an adverse react ion to c l inical ly used laser

devices. Lasers Surg Med 1987;7:457–460.

[CrossRef]

[Med line Link]

6. Anonymous. A c l inician's guide to surgica l f i res. How they occur, how to prevent

them, how to put them out. Hea lth Devices 2003;32:5–24.

7. Wolf GL. Danger from OR f ires st i l l a serious problem. J Clin Monit Comput

2000;16:237–238.

[CrossRef]

[Med line Link]

8. Wasek S. Preventing surgica l f i res. Outpat ient Surgery Magazine 2003;4:28–

35.

9. Anonymous. E lec trosurgica l fi res sti l l a hot topic. Technol Anesth 1996;17(4):1–

2.

10. Rampi l IJ . Anesthet ic considerat ions for laser surgery. Anesth Analg

1992;74:424–435.

[Fu ll text Link]

[CrossRef]

[Med line Link]

11. Garry B, Bivens HE. Laser safety in the operating room. I. Admin istering

anesthesia. Cancer Bul le tin 1989;441:219–223.

12. Sosis MB. Hazards of laser surgery. Seminars in Anesthesia 1990;9 :90–97.

13. Van Der Spek AFL, Spargo PM, Norton ML. The physics of lasers and

implica tions for their use during ai rway surgery. Br J Anaesth 1988;60:709–729.

[CrossRef]

[Med line Link]

14. Bal l KA. Lasers. The periopera tive challenge. St. Louis : Mosby, 1995.

15. Absten GT. The fundamentals of medical laser technology. Biomed Instrum

Technol 2002;36:203–207.

[Med line Link]

16. American National Standards Inst i tute. Safe use of lasers in health care

fac i li t ies (ANSI Z-136.3). New York: Author, 2005.

17. Denton RA, Dedhia HV, Abrons HL, e t al . Long-term surv ival af ter

endobronchial f i re during trea tment of severe malignant ai rway obs truct ion with the

Nd:YAG laser. Ches t 1988;94:1086–1088.

[CrossRef]

[Med line Link]

18. Krawtz S, Mehta AC, Wiedemann HP, et a l. Nd-YAG laser-induced

endobronchial burn. Management and long term follow-up. Chest 1989;95:916–

918.

[CrossRef]

[Med line Link]

19. Brut inel WM, Cortese DA Edel l ES, et al. Compl icat ions of Nd:YAG laser

therapy. Chest 1988;94:902–903.

[CrossRef]

[Med line Link]

20. Lobraico RV. Laser safe ty in heal th care faci l i ties. Am Col l Surg Bul l

1991;76:16–22.

[Med line Link]

21. Sosis M. Anesthesia for laser surgery. Clin Anes th Updates 1993;4(5):1–12.

22. Anonymous . The devastat ion of pat ient f ires. Heal th Dev ices 1992;21:3 .

23. Brodman M, Port M, Friedman F, et al. Opera ting room personne l morbidity from

carbon d ioxide laser use during preceptored surgery. Obstet Gyneco l 1993;81:607–

609.

[Fu ll text Link]

[Med line Link]

24. Anonymous . Laser igni tion of surgical drapes . Heal th Devices 1992;21:15–16.

[Med line Link]

25. Sliney DH. Laser safety. Lasers Surg Med 1995;16:215–225.

[CrossRef]

[Med line Link]

26. National Fi re Protection Assoc ia tion. Standard for laser f i re p rotec tion (NFPA

115). Qu incy, MA: Author, 2003.

27. National Fi re Protection Assoc ia tion. Standard for laser f i re p revent ion (NFPA

115). Qu incy, MA: Author, 2003.

28. Asai T. Burn in the mouth during oral laser surgery. Anaesthesia 1997;52:806–

807.

[Fu ll text Link]

[Med line Link]

29. Tucker RD, Voyles CR. Laparoscopic e lec trosurgical compl icat ions and their

prevention. AORN J 1995;62:52–78.

30. Pfenninger J. ABCs of electrosurgery. Outpat ient Surgery Magazine 2004;5:45–

50.

31. Anonymous . Recommended pract ices for electrosurgery. Ci ty, STATE:

Associat ion of Operat ing Room Nurses, Inc., 1999.

32. Kle in SL, Leonard PF. Buzz ‘em or burn ‘em. Anes th Ana lg 1991;73:358–359.

[CrossRef]

[Med line Link]

33. Lee J. Update on electrosurgery. Outpatient Surgery Magazine 2002;3:44–53.

34. Bel l AF, Shagets FW, Barrs DM. Princ iples and hazards of elec trocautery in

otolaryngology. Otolaryngol Head Neck Surg 1986;94:504–507.

[Med line Link]

35. Anonymous . Prevent ing pat ient burns at the return-electrode s i te during high-

current elec trosurgical procedures . Heal th Devices Alerts 2005;29:1–3.

36. Meltzer B. Issues in electrosurgery. Outpatient Surgery Magazine 2001;11:51–

61.

37 . Dennis V. Pref l ight ing your act ive e lectrodes. Outpat ient Surgery Magazine

2003;4:62–68.

38. Dennis V. Protect ing pat ien ts f rom laparoscopic burns. Outpatient Surgery

Magazine 2004;5:84–85.

39. Dennis V. 7 ho t spots in e lectrosurgery. Outpatien t Surgery Magazine

2005;6:49–53.

40. Dennis V. Toward safer laparoscopic electrosurgery. Outpat ient Surgery

Magazine 2006;7:38–40.

41. Anonymous . Return-elec trode-s i te burns assoc ia ted wi th Ri ta Medical Systems

Mode l 1500 and 1500X radio-frequency generators . Heal th Devices Alerts 2005;29:

3–4.

42. Gross JB. Less jo lts from your vol ts—elec trical safe ty in the opera ting room

(ASA Refresher Course #126). Park Ridge, IL: ASA, 2005.

43. Lohmann G. Faul t wi th an Ohmeda Excel 410 machine. Anaesthesia

1991;46:695.

[CrossRef]

[Med line Link]

44. Bortolussi ME, Hunter JG. Fi re hazard of the ophthalmic cautery. Plast

Reconstr Surg 1989;83:753.

[Fu ll text Link]

[CrossRef]

[Med line Link]

45. Curt in JW. The disposable cautery: a f ire hazard . Plast Recons tr Surg

1989;84:853.

[Fu ll text Link]

[CrossRef]

[Med line Link]

46. Anonymous . Hazard report. Fluorescent l ights can ac tiva te Val ley Forge

Scien tif ic Mal is CM CIII bipolar e lectrosurgica l uni t . Heal th Devices 1999;28(5–

6):230–231.

[Med line Link]

47. Ashman MN, Mathasko MJ. E lec trical and fi re safety in the operat ing room.

Seminars in Anesthesia 1993;12:276–281.

48. Anonymous . Tra ining paramount in reducing risk o f surgical f i res . Technol

Anesth 2003;24:5–7.

49. Mil l iken RA, Bizzarri DV. Flammable surgical drapes—a pat ient and personnel

hazard. Anesth Analg 1985;64:54–57.

[Fu ll text Link]

[CrossRef]

[Med line Link]

50. Tucker RD. Laparoscopic electrosurg ical injuries : survey resul ts and their

implica tions . Surg Laparosc Endosc 1995;5:311–317.

[Med line Link]

51. McKee K. A gu ide to tes ting your electrosurgery ins truments. Outpatien t

Surgery Magazine 2001;11:59–64.

52. Dennis V. Choosing your act ive electrodes wi th care. Outpatient Surgery

Magazine 2003;4:65–67.

53. Anonymous . OR f ires caused by f iberopt ic i l lumination sys tems. Heal th Devices

1982;11:148–149.

54. Taylor IR. Hazards of laparoscopic surgery. Br J Anaesth 1999;82:651.

[Med line Link]

55. Eggen MA, Brock-Utne JG. Fiberoptic i lluminat ion systems can serve as a

source o f smoldering f i res. J Cl in Moni t 1994;10:244–246.

[Med line Link]

56. Anonymous . OR f ires : preventing them and putting them out. Heal th Devices

1986;15:132–136.

57. Anonymous . Laparoscopy l ight cable ign ites opera ting room f ire. Biomed Safe

Stand 1994;24:162–163.

58. Anonymous . Fires f rom def ibri l la tion during oxygen administration. Techno l

Anesth 1994;15:1–2.

59. Anonymous . Safety notice . Anaesthesia 1995;50:581.

60. Anonymous . Oxygen regula tor fi re caused by use of two yoke washers. Technol

Anesth 1990;11:1–2.

61. French Minis try of Employment and Solidari ty . Withdrawal f rom the market of

some pressure regula tors fo r medical use. September 1 , 1998.

62. Anonymous . High-pressure gas regulators . Technol Anesth 1999;19(10):8–9.

63. Anonymous . High-pressure gas regulators . Technol Anesth 1999;19(11):6.

64. Anonymous . Are aluminum regulators safe? Technol Anesth 1999;11(11):4–5.

65. Explosions and f i res in aluminum oxygen regulators . FDA and NIOSH Publ ic

Hea lth Advisory, 1999.

66. Anonymous . Alert . www.hosmat.com, 2000.

67. Anonymous . Fires , explos ions lead to recall of aluminum oxygen regula tors.

Biomed Safe Stand 1999;29:57–58.

68. Anonymous . Withdrawal f rom the market of some pressure regulators for

medical use. www.hosmat.com, 2000.

69. Anonymous . “The patien t is on f i re! ” A surgical f i res primer. Heal th Dev ices

1992;21:19–23.

70. Eisenbaum SL. Facial burns as a complicat ion of of fice surgery l ighting . Plast

Reconst Surg 1989;83:155–159.

[Fu ll text Link]

[CrossRef]

[Med line Link]

71. Bourke DL, Yee K, Mark L . Severe burn caused by an opera ting room l ight.

Anesthesiology 1993;79:171–173.

[Fu ll text Link]

[CrossRef]

[Med line Link]

72. Anderson EF. A potent ia l ignit ion source in the operating room. Anesth Analg

1976;55:217–218.

[CrossRef]

[Med line Link]

73. Rogers S, Dav ies MW. My anaesthes ia machine's on f i re . Anaesthesia

1997;52:505.

[Fu ll text Link]

[Med line Link]

74. Anonymous . Anes thesia machine f i re hazard update . Biomed Safe S tand

2003;33: 73–74.

75. Usher AG, Cave DA, Finegan BA. Cri tical inc ident with Narkomed 6000

anesthesia mach ine. Anesthesiology 2003;99:762.

[Fu ll text Link]

[CrossRef]

[Med line Link]

76. Alexander PD, Meurer-Laban M. Rechargeable optima laryngoscopes. Br J

Anaesth 1995;74:724–725.

[CrossRef]

[Med line Link]

77. Anonymous . Risk of f ire in surgical booms. Heal th Devices Alerts 2005;29.

78. Anonymous . Risk of f ire in surgical booms. Technol Anesth 2005;25:5–6.

79. Anonymous . Overv iew of pneumatic tourniquets. Technol Anesth 2001;22(4):1–

3.

80. Rita L, Seleny F. Endotracheal tube ignition during la ryngea l surgery with

resec toscope. Anesthesiology 1982;56:60–61.

[Fu ll text Link]

[CrossRef]

[Med line Link]

81. Cel inski M. Explosions in anaesthesia. Br J Anaes th 2005;94:691–692.

[Fu ll text Link]

[CrossRef]

[Med line Link]

82. Anonymous . Incompatibi li ty of disposable heated-wire breathing ci rcu its and

heated-wire humid if ie rs. Technol Anesth 1993;14:4–5.

83. Anonymous . Anes thesia breathing ci rcuits any overheat & melt tub ing . Biomed

Safe Stand 1994;24:85.

84. Anonymous . Fisher & Paykel—Dual-heated adul t resp iratory vent i lator ci rcui ts :

r isk of fi re . Heal th Devices Alerts 2006;30:4.

85. Gaba DM. More on ni trous oxide and laser surgery. Anesth Analg 1988;67:488–

489.

[Fu ll text Link]

[CrossRef]

[Med line Link]

86. Geff in B, Shapshay SM, Bellack GS. Flammabi li ty of endotracheal tubes during

Nd-YAG laser appl ica tion in the ai rway. Anesthesiology 1986;65:511–515.

[Fu ll text Link]

[CrossRef]

[Med line Link]

87. Ossoff RH, Duncavage JA, Eisenman TS, et al . Comparison of tracheal damage

f rom laser igni ted endotracheal tube f ires. Ann Oto l Rhinol Laryngol 1983;92: 333–

336.

[Med line Link]

88. Duncavage JA, Ossoff RH, Rouman WC, et al. Injuries to the bronchi and lungs

caused by laser-igni ted endotracheal tube fires. Otola ryngol Head Neck Surg

1984;92:639–643.

[Med line Link]

89. Patel KF, H icks JN. Prevention of f i re hazards associated with use of carbon

dioxide lasers. Anesth Analg 1981;60:885–888.

[Fu ll text Link]

[CrossRef]

[Med line Link]

90. Fried MP, Mal lampati R, Caminear DS. Compara tive analysis of the safety of

endotracheal tubes with the KTP laser. Laryngoscope 1989;99:748–751.

[Fu ll text Link]

[CrossRef]

[Med line Link]

91. Sosis MB, Braverman B. Advantage of rubber over plas tic endotracheal tubes

for rapid extubat ion in a laser f ire. J Cl in Laser Med Surg 1996;14:93–95.

[Med line Link]

92. Anonymous . Guidance on airway management during laser surgery of upper

ai rway. Technical Report ISO TR 11991, 1995.

93. Anonymous . Upper airway management guide provided for laser ai rway surgery.

APSF News le tt 1993;8:13–16.

94. Anonymous . Endotracheal tube mel ts and ignites inside pat ient 's trachea.

Anesthesia Malpractice Prevention 2001;6:9–14.

95. Brimacombe J. The incendiary characteris tics of the laryngeal and reinforced

laryngea l mask ai rway to CO2 laser strike—a comparison wi th two polyv inyl chloride

tracheal tubes. Anaes th In tens Care 1994;22:694–697.

[Med line Link]

96. Galapo S, Wolf GL, Sidebotham GW, et al . Laser igni tion of surgical drapes in

an oxygen enriched atmosphere. Anesthes io logy 1998;89:A560.

97. Anonymous . Fire! Technol Anesth 1996;16:4–5.

98. Gibbs JM. Combustib le plast ic drape. Anaesth Intens Care 1983;11:176.

[Med line Link]

99. Lazard JL , Wolf GL, Charchofl ieh J, et al. Surg ical drape combustion in FIO2-

0.21, FIO2-0.50 and FIO2-0.95. Pos ter presentat ion, ASA Meeting, New Orleans,

2001.

100. Lazard JL, Wolf GL, Charchaf lich J, e t al . Surgical d rape combus tion in FIO2-

0.21, FIO2-0.50 and FIO2-0.95. Pos ter presentat ion, American Soc iety of Crit ical

Care Anes thesiologists , New Orleans, 2001.

101. Barker SJ, Polson JS. Fi re in the opera ting room: a case report and laboratory

study. Anes th Ana lg 2001;93:960–965.

[Fu ll text Link]

[CrossRef]

[Med line Link]

102. Ott AE. Disposable surgical d rapes—a potent ial f ire hazard. Obstet Gynecol

1983;61:667–668.

[Fu ll text Link]

[Med line Link]

103. Domin MA. Safety precaut ions fo r laser surgery. J Heal thc Mater Manage

1991(June);9:20–25.

104. Wolf GL, Sidebotham GW, Lazard JLP, e t al . Laser igni tion of surg ical drape

materials in air, 50% oxygen, and 95% oxygen. Anesthesiology 2004;100:1167–

1171.

[Fu ll text Link]

[CrossRef]

[Med line Link]

105. Anonymous . Surgical f i re case summaries. Hea lth Devices 1992;21(1):31–34.

P.927

106. Axe lrod EH, Kusnetz AB, Rosenberg MK. Operat ing room f ires init iated by hot

wi re cautery. Anes thesiology 1993;79:1123–1126.

[CrossRef]

[Med line Link]

107. Bennett JA, Agree M. Fi re in the ches t. Anesth Analg 1994;78:406.

[Med line Link]

108. Ortega RA. A rare cause o f fi re in the operating room. Anesthesiology

1998;89:1608.

[Fu ll text Link]

[CrossRef]

[Med line Link]

109. Lai A, Ng KP. Fi re during thoracic surgery. Anaesth Intens Care 2001;29:301–

303.

[Med line Link]

110. Blanchard DL. Fi re in surgery. Ophthal Plast Reconstr Surg 1987;3 :59.

[CrossRef]

[Med line Link]

111. Stouffer DJ. Fi res during surgery: two fatal incidents in Los Angeles. J Burn

Care Rehab il 1992;13:114–117.

[CrossRef]

[Med line Link]

112. Wood DK, Ho ll is R. Thermocautery causes a gauze pad f ire. JAMA

1993;270:2299–2300.

[CrossRef]

[Med line Link]

113. Katz JA, Campbel l L. F ire during thoracotomy: a need to control the inspired

oxygen concentration. Anesth Analg 2005;101:612.

[Fu ll text Link]

[CrossRef]

[Med line Link]

114. Santos P, Ayuso A, Luis M, et al . Ai rway ign it ion during CO2 laser la ryngeal

surgery and high f requency je t venti la tion. Eur J Anaesthesiol 2000;17:204–207.

[CrossRef]

[Med line Link]

115. Anonymous . $200,220 awarded in pat ient burn case. Biomed Safe Stand

1995;25:133.

116. Anonymous . Use of acetone & “eggcrate” mattress c ited in operating room fi re .

Indiana hospi ta ls advised to review in ternal pol ic ies. Biomed Safe Stand

1989;19:26.

117. Anonymous . Fi re hazard c reated by the misuse of DuraPrep solut ion . Heal th

Dev ices 1998;27(5):400–402.

[Med line Link]

118. Plumlee JE. Operat ing-room f lash f i re f rom use of cautery af ter aerosol spray:

a case report. Anesth Analg 1973;52:202–203.

[Fu ll text Link]

[CrossRef]

[Med line Link]

119. Sebben J. Fi re hazard and electrosurgery. J Dermatol Surg Oncol

1990;16:421–424.

[Med line Link]

120. Anonymous . Improper use of alcoho l-based skin preps can cause surgical

f ires. Tech Anesth 2003;24:3–6.

121. Prasad R, Quezado Z, St. Andre A, et al . Fi res in the operat ing room and

intensive care uni t : awareness is the key to prevention. Anesth Analg

2006;102:172–174.

[Fu ll text Link]

[CrossRef]

[Med line Link]

122. Macdonald AG. A brief his to rical rev iew of non-anaesthet ic causes of f i res and

explosions in the opera ting room. Br J Anaesth 1994;73:847–856.

[CrossRef]

[Med line Link]

123. Green JM, Sonbol ian N, Gonzalez RM, et al . CO2 laser res is tance of various

ointments and tapes. Anes thesiology 1991;74:964–965.

[Fu ll text Link]

[CrossRef]

[Med line Link]

124. Datta TD. Flash f i re hazard with eye ointment. Anesth Analg 1984;63:700–

701.

[Fu ll text Link]

[CrossRef]

[Med line Link]

125. Ches tler RJ , Lemke BN. Intraoperat ive f lash f ires associated wi th disposab le

cautery. Ophthal Pl Reconst Surg 1989;5 :194–195.

126. Martin L, Dolman P. Fi re! Can J Anaesth 1994;46:909.

127. de Richmond AL, Bruley ME. Head and neck surgica l f i res. In: Eisele DW, ed.

Compl icat ions in Head and Neck Surgery. St. Louis: Mosby–Year Book, 1993;311–

312.

128. Mandych A, Mickelson S, Amis R. Operat ing room f i re . Arch Otolaryngol Head

Neck Surg 1990;116:1452.

[Med line Link]

129. Anonymous . Infant dies af te r operat ing room flash f i re. Biomed Safe Stand

1988;18:154.

130. Bowdle TA, Glenn M, Cols ton H, et al . Fi re fol lowing use of electrocautery

during emergency percutaneous trans tracheal venti lat ion. Anesthesiology

1987;66:697–698.

[Fu ll text Link]

[CrossRef]

[Med line Link]

131. Anonymous . Laser-igni ted latex glove causes airway f ire. Biomed Safe Stand

1992;22:51.

132. Simpson JI, Wol f GL. F lammabili ty of esophageal s tethoscopes, nasogastric

tubes , feeding tubes, and nasopharyngeal ai rways in oxygen- and ni trous oxide-

enriched atmospheres. Anesth Analg 1988;67:1093–1095.

[Med line Link]

133. Demos JE. Fi re in the opera ting room. Plast Recons tr Surg 1995;95:419–

420.

[Fu ll text Link]

[CrossRef]

[Med line Link]

134. Bropuwers JW. Collodion and electrocautery do not mix . Anes th Analg

1998;67:718.

[Fu ll text Link]

[CrossRef]

[Med line Link]

135. Carter R, Smith JS. Oesophagea l prosthetic igni t ion following accidenta l

exposure to the Nd. YAG laser—a rea l or po tential hazard. Lasers in Medical

Science 1988;2:229–232.

[CrossRef]

136. Anonymous . Anesthesia f i re burns young boy. Biomed Safe Stand 2003;33:65–

66.

137. Kanno TA, Aso C, Saito S, et a l. A combustive des truct ion of exp irat ion valve

in an anes thetic ci rcu it . Anesthesiology 2003;98:577–579.

[Fu ll text Link]

[CrossRef]

[Med line Link]

138. Mayworm D. How to prevent infect ion from surg ica l instruments . Outpatient

Surgery Magazine 2002; Special edi tion:38–44.

139. Barlas S. TIA c leans up hand-sani tizer controversy. NFPA J 2004;98:36.

140. Anonymous . Venti la tor & ESU involved in electrosurgical f i re. Biomed Safe

Stand 1991;21:114–115.

141. Rogers SA, Mi l ls KG, Tufa il Z . Case report. Airway f ire due to d iathermy

during tracheostomy in an intens ive care pat ien t. Anaesthesia 2001;56:441–446.

[Fu ll text Link]

[CrossRef]

[Med line Link]

142. Lim HJ , Mil ler GM, Rainbird A. Airway f ire during e lec tive tracheotomy.

Anaesth Intens Care 1997;25:150–152.

[Med line Link]

143. Bai ley MK, Bromley HR, All ison JG, e t al . Electrocautery-induced ai rway f i re

during tracheostomy. Anesth Analg 1990;71:702–704.

[Fu ll text Link]

[CrossRef]

[Med line Link]

144. Aly A, McIlwa in M, Duncavage JA. Elec trosurgery-induced endotracheal tube

igni tion during tracheotomy. Ann Otol Rhinol Laryngol 1991;100:31–33.

[Med line Link]

145. Lew EO, Mitt leman RE, Murray D. Endotracheal tube igni tion by elec trocautery

during tracheostomy: case report wi th au topsy f ind ings. J Forensic Sci

1991;36:1586–1591.

[Med line Link]

146. LeClai r J , Gartner S, Halma G. Endotrachea l tube cuff igni ted by

electrocautery during tracheostomy. J Am Assoc Nurs Anesth 1990;58:259–261.

147. Thompson JW, Colin W, Snowden T, et a l. F ire in the operat ing room during

tracheos tomy. South Med J 1998;91:243–247.

[Fu ll text Link]

[Med line Link]

148. Chee WK, Benumof JL. Airway f i re during tracheos tomy: extubation may be

contraindicated. Anes thesiology 1998;89:1576–1578.

[Fu ll text Link]

[CrossRef]

[Med line Link]

149. Marsh B, Ri ley RH. Double-lumen tube f i re during tracheostomy.

Anesthesiology 1992;76:480–481.

[Fu ll text Link]

[CrossRef]

[Med line Link]

150. Michels AMJ, Stott S. Explosion of tracheal tube during tracheostomy.

Anaesthesia 1994;49:1104.

[CrossRef]

[Med line Link]

151. Wilson PTJ, Igbaseimokumo U, Martin J. Igni t ion of the tracheal tube during

tracheos tomy. Anaesthesia 1994;49:734–735.

[CrossRef]

[Med line Link]

152. Bowdle TA, Glenn M, Cols ton H, et al . Fi re fol lowing use of electrocautery

during emergency percutaneous trans tracheal venti lat ion. Anesthesiology

1987;66:697–698.

[Fu ll text Link]

[CrossRef]

[Med line Link]

153. Baur DA, But le r RC. Electrocautery-ign ited endotracheal tube f i re: case report.

Br J Oral Maxil lofac Surg 1999;37:142–143.

[CrossRef]

[Med line Link]

154. Awan MS, Ahmed I. Endotracheal tube f i re during tracheos tomy: a case report.

Ear Nose Throat J 2002;81:90–92.

[Med line Link]

155. Sos is MB, Braverman B. Prevent ion of cautery-induced a irway f ires wi th

special endotracheal tubes. Anesth Analg 1993;77:846–847.

[Fu ll text Link]

[CrossRef]

[Med line Link]

156. Gupte SR. Gauze f i re in the oral cavi ty: a case report. Anesth Analg

1972;51:645–646.

[Fu ll text Link]

[CrossRef]

[Med line Link]

157. Simpson JI, Wol f GL. Endotracheal tube fi re igni ted by pharyngeal

electrocautery. Anesthesio logy 1986;65:76–77.

[Fu ll text Link]

[CrossRef]

[Med line Link]

158. Kel ler C, El l iot t W, Hubbel l RN. Endotracheal tube safety during

electrodissection tonsi l lectomy. Arch Oto laryngol Head Neck Surg 1992;118:643–

645.

[Med line Link]

159. Collee GG. A f i re in the mouth. Anaesthesia 1984;39:936.

[CrossRef]

[Med line Link]

160. Macdonald MR, Wong A, Walker P, et a l. Elec trocautery-induced ign ition of

tonsi l la r packing. J Oto la ryngol 1994;23:426–429.

[Med line Link]

161. Kaddoum RN, Chidiac EJ, Zestos MM, et al. Elec trocautery-induced f i re during

adenotons il lectomy: report of two cases. J Cl in Anesth 2006;18:129–131.

[CrossRef]

[Med line Link]

162. Mattucci KF, Mil i tana CJ. The prevention of f i re during oropharyngeal

electrosurgery. Ear Nose Throat J 2003;82:107–109.

[Med line Link]

163. Pashayan AG, Gravens tein N. High incidence of CO2 laser beam contac t with

the tracheal tube during operat ions on the upper ai rway. J Clin Anes th 1989;1:354–

357.

[CrossRef]

[Med line Link]

164. Hirshman CA, Smith J. Ind irect igni t ion o f the endotracheal tube during carbon

dioxide laser surgery. Arch Oto laryngol 1980;106:639–641.

[Med line Link]

165. Cozine K, Rosenbaum LM, Askanazi J , et al . Laser-induced endotracheal tube

f ire. Anesthesiology 1981;55:583–585.

[Fu ll text Link]

[CrossRef]

[Med line Link]

166. Eade GG. Hazard of nasal oxygen during aesthet ic facial operations . Plast

Reconstr Surg 1986;78:539.

[Fu ll text Link]

[CrossRef]

[Med line Link]

167. Brechtelsbauer P, Carrol l W, Baker S. Intraopera tive f i re with electrocautery.

Otola ryngol Head Neck Surg 1996;114:328–331.

[CrossRef]

[Med line Link]

168. Chang BW, Petty P, Manson PN. Patient fi re safety in the operat ing room.

Plast Reconstr Surg 1994;93:519–521.

[Med line Link]

169. Reyes RJ, Smith AA, Mascaro JR, e t al . Supp lemental oxygen: ensuring i ts

safe del ivery during facial surgery. Plast Reconstr Surg 1995;95:924–928.

[CrossRef]

[Med line Link]

170. Anonymous . Fi res f rom oxygen use during head and neck surgery. Heal th

Dev ices 1995;24:155–157.

[Med line Link]

171. Anonymous . Preparing for and manag ing surgical f ires. Heal th Dev ices

1992;21(1):24–31.

[Med line Link]

172. Aston SJ, Bornstein SJ . An unusual compl icat ion associa ted wi th

blepharoplasty. Aes thetic Plast Surg 1978;2:451–453.

[CrossRef]

173. Lederman IR. Fi re hazard during ophthalmic surgery. Ophthalmic Surg

1985;16:577–578.

[Med line Link]

174. Garf ield JM, Al len GW, Si lverstein P, et al . Flash f ire in a reducing valve.

Anesthesiology 1971;34:578–579.

[Fu ll text Link]

[CrossRef]

[Med line Link]

175. Gunatilake DE. Case report: fata l intraperi toneal explosion during

electrocoagulat ion v ia laparoscopy. Int J Gynaecol Obste t 1978;15:353–357.

[Med line Link]

176. El-Kady AA, Abd-El Razek M. Intraperitonea l exp losion during female

steri l izat ion by laparoscopic elec trocoagulation. A case report. In t J Gynaecol

Obs tet 1976;14:487–488.

[Med line Link]

177. Robinson JS, Thompson JM, Wood AW. Fire and explosion hazards in

operating theatres : a reply and new evidence. Br J Anaesth 1979;51:908.

[CrossRef]

[Med line Link]

178. Diemunsch PA, Torp KD, Van Dorsselaer T, e t al . Ni trous ox ide f raction in the

carbon d ioxide pneumoperitoneum during laparoscopy under general inhaled

anesthesia in pigs . Anesth Analg 2000;90:951–953.

[Fu ll text Link]

[CrossRef]

[Med line Link]

179. Neuman GG, Sidebotham G, Negoianu E, et a l. Laparoscopy explosion

hazards with ni trous oxide. Anesthesiology 1993;78:875–879.

[Fu ll text Link]

[CrossRef]

[Med line Link]

180. Anonymous . Use of wrong gas in laparoscopic insuff lator causes f i re. Heal th

Dev ices 1994;23:55–56.

[Med line Link]

181. Gre il ich PE, Greil ich NB, Froel ich EG. Intraabdominal f i re during laparoscopic

cholecys tec tomy. Anesthesiology 1995;83:871–874.

[Fu ll text Link]

[CrossRef]

[Med line Link]

182. Anonymous . Def ib ri llator sparks f lames in ambulance. Hea lth Devices Alerts

2004;28:1–2.

183. Anonymous . Using ex ternal def ibri llators in oxygen-enriched atmospheres can

cause f ires. Heal th Dev ices Alerts 2006;30:1–2.

184. Robertshaw H, McAnul ty G. Ambient oxygen concentrat ions during s imulated

cardiopu lmonary resusci tation. Anaes thesia 1998;53:634–637.

[Fu ll text Link]

[CrossRef]

[Med line Link]

185. Anonymous . Pat ient se t af ire by def ib ri l lator paddles. Biomed Safe Stand

2004;34:25.

186. Cas tro BA, Freedman A, Craig WL, et a l. Explosion wi thin an anesthesia

machine: Baralyme, h igh f resh gas f lows and sevof lurane concentrat ion.

Anesthesiology 2004;101:537–539.

[Fu ll text Link]

[CrossRef]

[Med line Link]

187. McKeskey CH. Letter to hea lth care professionals. 2003.

188. Las ter M, Roth P, Eger EI. Fi res from the inte rac tion of anesthetics wi th

desiccated absorbent. Anesth Analg 2004;99:769–774.

[Fu ll text Link]

[CrossRef]

189. Wu J, Previte JP, Adler E, e t al . Spontaneous igni t ion , exp losion, and f i re wi th

sevof lu rane and barium hydroxide l ime. Anesthesiology 2004;101:534–537.

[Fu ll text Link]

[CrossRef]

[Med line Link]

190. Olympio MA, Morell RC. Canister f ires become a hot safety concern. APSF

News le tt 2004;18:45–48.

191. Anonymous . Anesthesia carbon dioxide absorber f ires . Technol Anesth

2003;24:1–3.

192. Fatheree RS, Leighton BL. Acute respiratory distress syndrome after an

exothermic Baralyme-sevoflurane react ion. Anesthesiology 2004;101:531–533.

[Fu ll text Link]

[CrossRef]

[Med line Link]

193. Holak EJ, Mei DA, Dunning MB, et al. Carbon monoxide production f rom

sevof lu rane breakdown: mode ling of exposures under c linical condit ions. Anes th

Analg 2003;96:757–764.

[Fu ll text Link]

[CrossRef]

[Med line Link]

194. Gatti JE, Bryant CJ, Noone RB, et a l. The mutagenici ty of electrocautery

smoke. Plast Reconstr Surg 1992;89:781–784.

[Fu ll text Link]

[CrossRef]

[Med line Link]

195. Bal l K. No smoking in the OR. Outpatient Surgery Magazine 2003;4:62–67.

196. American Society of Anesthesiologists. Recommendations for infection control

for the practice of anes thesiology. Park Ridge, IL : ASA, 1998.

197. Paes ML. General anaes thesia for carbon dioxide laser surgery wi thin the

ai rway. Br J Anaesth 1987;59:1610–1620.

[CrossRef]

[Med line Link]

198. Anonymous . The elements of surgical f ires . Technol Anesth 2003;23:1–7.

199. Anonymous . Laser contact t ips and tracheal tubes. Health Devices

1992;21:18.

[Med line Link]

200. Wolf GL, Sidebotham GW, Stern JB. Intraluminal f lame spread in tracheal

tubes . Laryngoscope 1994;104:874–879.

[Fu ll text Link]

[CrossRef]

[Med line Link]

201. Moak E. E lec trosurgica l uni t safety. AORN J 1991;53:744–752.

[CrossRef]

[Med line Link]

202. McCranie J. Fi re safe ty in the opera ting room. Today's OR Nurse 1994;16:33–

37.

[Med line Link]

203. Anonymous . Laser starts f ire in OR. Technol Anesth 1988;8:3–4.

204. Denn is V. Improving your electrosurgery safety. Outpat ient Surgery Magazine

2003;4:44–48.

205. Anonymous . Electrocautery uni ts : improper disposal can cause f i res. Health

Dev ices Alerts 2005;29.

P.928

206. Ehrenwerth J . A f i re in the opera ting room: i t can happen to you (ASA

Ref resher Course #144). Park Ridge, IL: ASA, 2005.

207. Anonymous . FDA issues safety a lert on heated-wire breath ing c ircui t hazards.

Biomed Safe Stand 1993;23(19):146–147.

208. Gowardman JR, Moriarty B. Explosion and f i re in the expiratory l imb of a

Fisher and Paykel “three in one” respiratory care system. Anaesth In tens Care

1998;26:427–430.

[Med line Link]

209. Sommers JR. Flammabil i ty s tandards for surg ical drapes and gowns. Pas t,

present, and fu ture. Surgical Serv ices Management 1998;4 :41–44.

210. McCranie J. Fi re safe ty in the opera ting room. Todays OR Nurse 1992;14:33–

37.

211. Hawk ins DB, Joseph MM. Avoiding a wrapped endotracheal tube in laser

laryngea l surgery: experiences with apne ic anesthes ia and metal Laser-Flex

endotracheal tubes. Laryngoscope 1990;100:1283–1287.

[Fu ll text Link]

[CrossRef]

[Med line Link]

212. Kwan A. The use of Storz bronchoscope in prevention of airway f ire. Anaesth

Intens Care 2004;32:720.

[Med line Link]

213. Kjeldsen L, Andersen APD, Hjorth A. Laser surgery in the ai rway. Anaesthes ia

1986;39:1146.

[CrossRef]

[Med line Link]

214. Hunsaker DH. Anesthesia for microlaryngeal surgery: the case fo r subglo ttic

jet venti lation . Laryngoscope 1994;104:1–30.

[Med line Link]

215. Wegrzynowicz ES, Jensen NF, Pearson KS, e t al . Ai rway f i re during jet

venti lation for laser excision of voca l cord papi l lomata . Anes thesiology

1992;76:468–469.

[Fu ll text Link]

[CrossRef]

[Med line Link]

216. Sos is MB, Di l lon FX. Sal ine-f i l led cuffs help prevent laser-induced

polyv inylchloride endotrachea l tube f i res. Anesth Analg 1991;72:187–189.

[Med line Link]

217. LeJeune FE, Guice C, LeTard F, e t al . Heat s ink pro tec tion against lasering

endotracheal cuffs . Ann Oto l Rhinol Laryngo l 1982;92:606–607.

[Med line Link]

218. Walsh M, Schubert A, Al Haddad S. The addi t ion of lidocaine je lly to saline in

the cuff of the endotracheal tube during laser surgery of the airway. Am J Anesth

1997;24:189–193.

219. Sos is MB, Di l lon F. Prevention of CO2 laser-induced wi th the Laser-Guard

protect ive coat ing. J Cl in Anesth 1992;4:25–27.

[CrossRef]

[Med line Link]

220. Sos is MB. Evaluat ion of foil coverings for protect ing p las tic endotracheal

tubes f rom the potassium-ti tanyl-phosphate laser. Anesth Analg 1993;77:589–591.

[Fu ll text Link]

[CrossRef]

[Med line Link]

221. Sos is MB. Why is the Laser-guard protec tive coating better than the foil

tapes? In response. Anesth Analg 1994;78:1030.

[Fu ll text Link]

[CrossRef]

222. Van De Merwe W. Silver p rotec tive tape resis ts burning better than aluminum.

Cl in Laser Mon 1989;7:21–22.

223. Sos is MB, Pri tikin JB, Caldarel l i DD. The effect of blood on laser-resistant

endotracheal tube combus tion. Laryngoscope 1994;104:829–831.

[Fu ll text Link]

[CrossRef]

[Med line Link]

224. Sos is M, Dil lon F. Reflec tion of CO2 laser radia tion f rom laser-resistant

endotracheal tubes. Anesth Analg 1991;73:338–340.

[CrossRef]

[Med line Link]

225. Sos is M, Dil lon F. Prevent ion of CO2 laser tracheal tube f i res with the Laser-

Guard protective coat ing. Can J Anaesth 1989;36:S88–S89.

226. Sos is M, Dil lon F. What is the safes t fo il tape for endotracheal tube protection

during Nd-YAG laser surgery? Anesthesiology 1990;72:553–555.

227. Sos is MB. Which is the safest endotracheal tube for use with the CO2 laser? A

compara tive study. J Clin Anesth 1992;4:217–219.

[CrossRef]

[Med line Link]

228. Walker P, Temperley A, Thelfol S, e t al . Avoidance of laser igni tion of

endotracheal tube by wrapping in aluminum foil tape. Anaesth Intens Care

2004;32:108–112.

[Med line Link]

229. Kuo CH, Tan PH, Chen JJ, et al. Endotrachea l f i res during carbon dioxide

laser surgery on the larynx—a case report. Ac ta Anaesthesiol Sin 2001;39:53–56.

[Med line Link]

230. DeVane CC. Laser ini tiated endotracheal tube explosion. AANA J

1996;58:188–192.

231. DeVane GG. Case report: laser ini tiated endotracheal tube explosion. J Am

Assoc Nurs Anesth 1990;58:188–192.

232. Ko C-H, Tan P-H, Chen J-J, e t al . Endotracheal tube f ires during carbon

dioxide laser surgery on the larynx—a case report. Ac ta Anaesthes iol Sin

2001;39:53–56.

[Med line Link]

233. Fontenot R, Bailey BJ, Stiernberg CM, et al . Endotracheal tube safety during

laser surgery. Laryngoscope 1987;97:919–921.

[Fu ll text Link]

[CrossRef]

[Med line Link]

234. Sos is MB. Sa line soaked pledgets prevent carbon dioxide laser-induced

endotracheal tube cuff inf lat ion. J Clin Anesth 1995;7:395–397.

[CrossRef]

[Med line Link]

235. Anonymous . Do pledgets protect the tracheal tube cuff f rom lasers? Heal th

Dev ices 1992;21:17.

236. Mil ler G, Limb R. The CO2 laser and latex armoured tube (le tter). Anaesth

Intens Care 1995;23:517–518.

[Med line Link]

237. Ses terhenn AM, Dunne AA, Braulke D, et al . Value of endotracheal tube safety

in laryngeal laser surgery. Lasers Surg Med 2003;32:384–390.

[CrossRef]

[Med line Link]

238. Anonymous . Laser-resistant trachea l tubes. Heal th Dev ices 1992;21:4–14.

[Med line Link]

239. Anonymous . Laser-resistant endotracheal tubes and wraps . Heal th Devices

1990;19:112–139.

240. Ehrenwerth J . Fi re safety in the operating room (ASA Refresher Course). Park

Ridge, IL: Author, 2002.

241. Anonymous . Airway f i res : reducing the risk during laser surgery. Technol

Anesth 1990;11:1–3.

242. Inte rnational Standards Organization. Lasers and laser-related equipment: tes t

method for determining the CO2 laser res istance of surgica l drapes and/or pat ient

protect ive covers (ISO 11810). Geneva, Swi tzerland: Author, 2002.

243. Sos is M. Evaluat ion of a new laser-resistant anesthesia ci rcui t protector.

drape and patien t eye shie ld. J Cl in Moni t 1991;7 :132.

244. Sos is M, Braverman B, Ivankov ich AD. Meta l anesthesia c i rcu it components

stop laser f i res . Anes thesiology 1991;75:A396.

[Fu ll text Link]

[CrossRef]

245. Anonymous . Increased PEEP levels may reduce laser-induced tracheal tube

f ire risks . Biomed Safe Stand 1993;23:123.

246. Pashayan AG, San Giovanni C, Dav is LE. Posi tive end-exp iratory pressure

lowers the risk of laser-induced po lyv inylchloride tracheal-tube f i res.

Anesthesiology 1993;79:83–87.

[Fu ll text Link]

[CrossRef]

[Med line Link]

247. Lampotang S, Gravenste in N, Paulus DA, et al. Reducing the incidence of

surgical fi res: supplying nasal cannulae wi th sub-100% O2 gas mixtures f rom

anesthesia mach ines. Anesth Analg 2005;101:1407–1412.

[Fu ll text Link]

[CrossRef]

[Med line Link]

248. Anonymous . Reducing the risk of f ires during head and neck surgery. Technol

Anesth 2002;22:1–2.

249. Lowry RK, Noone RB. Fi res and burns during plastic surgery. Ann Plast Surg

2001;46:72–76.

[Fu ll text Link]

[CrossRef]

[Med line Link]

250. Greco RJ, Gonzalez R, Johnson P, et a l. Potential dangers of oxygen

supplementat ion during fac ial surgery. Plast Recons tr Surg 1995;95:978–984.

[Fu ll text Link]

[CrossRef]

[Med line Link]

251. Barnes AM, Frantz RA. Do oxygen-enriched a tmospheres exis t beneath

surgical drapes and contribute to f ire potentia l in the operat ing room? AANA J

2000;68:153–161.

252. Ramanathan S, Capan L, Cha lon J , et al . Minienvi ronmental contro l under the

drapes during opera tions on the eyes of conscious patien ts. Anesthesiology

1978;48:286–288.

[Fu ll text Link]

[CrossRef]

[Med line Link]

253. Rinder CS, Dabu-Bondoc S, Salgar V. A dev ice to reduce O2 accumulat ion and

reduce f i re hazard during MAC anes thesia. Anesthesiology 2005;103:A804.

254. Borkowski RG, Meneghett i C, Morgan MM, e t al . A comparison of methods for

delivering supplementa l oxygen during plastic surgery. Anes thesiology

2005;103:A841.

255. Sos is M. Ni trous oxide is contraindicated in endoscop ic surgery. Can J

Anaesth 1987;34:539.

[Med line Link]

256. Sos is M. Ni trous oxide shou ld no t be used during laser endoscop ic surgery.

Anesth Analg 1987;66:1054–1055.

[Fu ll text Link]

[CrossRef]

[Med line Link]

257. Shapiro JD, El-Baz NM. N2O has no p lace during oropharyngeal and

laryngotracheal procedures. Anesthes iology 1987;66:447–448.

[Fu ll text Link]

[CrossRef]

[Med line Link]

258. Simpson JI, Schiff GA, Wolf GL. The effect of hel ium on endotracheal tube

f lammabili ty . Anes thesiology 1990;73:538–540.

[Full text Link]

[CrossRef]

[Med line Link]

259. Alhaddad S, Brenner J. Hel ium and lower oxygen concentrat ions do not

prolong tracheal tube ign it ion t ime during potassium t i tanyl phosophate laser use.

Anesthesiology 1994;80:936–938.

[Fu ll text Link]

[CrossRef]

[Med line Link]

260. Rampi l IJ . He lium and gas f low. Anesth Analg 1993;76:S453.

261. Eisenk raft JB, Barker SJ . Hel ium and gas flow. Anesth Analg 1993;76:452–

453.

[Med line Link]

262. Sos is MB. Evaluat ion of a technique of nitrogen insuff lat ion to decrease

plast ic endotracheal tube cuff f lammabi li ty during carbon d ioxide laser surgery. J

Cl in Anesth 1993;5:468–470.

[CrossRef]

[Med line Link]

263. Pashayan AG, Gravens tein JS. Airway f i res during surgery wi th the carbon

dioxide laser. Anesthes iology 1989;71:478.

[Fu ll text Link]

[CrossRef]

[Med line Link]

264. Sommer RM. Prevent ing endotracheal tube f i re during pharyngeal surgery.

Anesthesiology 1987;66:439.

[Fu ll text Link]

[CrossRef]

[Med line Link]

265. Inappropriate use of oxygen to inf late or power medical devices . Heal th

Dev ices 1983;3:3.

266. Coral l IM, E lias JA, Strunin L. Laparoscopy explosion wi th n itrous oxide. Br

Med J 1975;1 :288.

[Med line Link]

267. Robinson JF, Thompson JF, Wood AW. Laparoscopy explosion with ni trous

ox ide . Br Med J 1975;3 :764–765.

[Med line Link]

268. Robinson JS, Thompson JM, Wood AW. Laparoscopy explosion hazards wi th

ni trous oxide. Br Med J 1975;4:760–761.

[Med line Link]

269. MDA SN 2000(07), Medical gas cylinders : risk of f i re. MDA Adverse Incident

Center, aic@medica l-devices.gov.uk/sn2000(07).htm

270. Moxon MA, Reading ME, Ward MB. Fi re in the opera ting theatre. Evacuat ion

pre-planning may save l ives. Anaesthesia 1986;41:543–546.

[CrossRef]

[Med line Link]

271. Weston CJ. The operat ing theatre's on f i re . Heal th Serv Manage

1988;84(3):20–23.

[Med line Link]

272. Fennell M. A multidisc iplinary approach to intraoperat ive f i re safety. AORN J

1995;62:636–637.

[CrossRef]

[Med line Link]

273. Vidor K, Puterbaugh S, Will is C. Fire safety training: a program for the

operating room. AORN J 1989;49:1045–1049.

[CrossRef]

[Med line Link]

274. Anonymous . Recognizing and extingu ishing medical gas fi res. Health Devices

Alerts 2003;27:5.

275. Lierz P , Heinatz A, Gustorff B. Management o f in tratracheal f i re during laser

surgery. Anesth Analg 2002;93:502.

[Fu ll text Link]

[CrossRef]

[Med line Link]

276. Anonymous . Fi re blankets in the OR? Technol Anesth 2000;20:3.

277. Bruner JMR. Fire in the operat ing room. ASA Newslett 1990;54:22–25.

278. Anonymous . Responding to fi res in areas of oxygen use. Heal th Dev ices

1994;23:306–309.

[Med line Link]

279. Anonymous . Recognizing and extingu ishing medical gas fi res. Health Devices

Alerts 2003;27(A14):5.

280. Deboer GE. Endotracheal tube f i res . Can Anaesth Soc J 1993;40:1003–1004.

281. Wolf G, Sidebotham GW. Endotracheal tube f i re: comments on the advisab il i ty

of not ex tubating . Anes thesiology 1999;91:888.

[Fu ll text Link]

[CrossRef]

[Med line Link]

282. Nat ional F ire Protect ion Associat ion . NFPA guide to portable f i re

ex tinguishers . Quincy, MA: Author, 2003.

283. Anonymous . Talk to the specialis t . Techno l Anes th 1996;17(4):3–4.

284. Finnegan J. Staff educat ion can prevent OR f i res . Todays OR Nurse

1994;16(3):24–26.

[Med line Link]

285. Nat ional F ire Protect ion Associat ion . Standard fo r portab le f ire ext inguishers

(NFPA 10), 2002 ed. www.nfpacatalog.org. Quincy, MA: Author, 2002.

286. Nat ional F ire Protect ion Associat ion . Standard fo r portab le f ire ext inguishers

(NFPA 10), 2002 ed. Quincy, MA: NFPA, 2002.

287. Anonymous . 10 quest ions about NFPA 10. What you might want to know about

maintaining and using portable fi re ex tingu ishers. NFPA J 2001;95:26, 92 .

288. Freestone A. Finding a halon alternat ive. NFPA J 2002;96:68–73.

289. Gibson J. Finding a halon alternative. NFPA J 2002;96:18, 20.

290. Anonymous . Select ing fi re extinguishers for the opera ting room. Hea lth

Dev ices 1996;25:261, 263.

P.929

Questions For the fol lowing quest ions, select the correct answer

1. The percentage of operating room fires that result in patient injury is A. 10% to 20%

B. 20% to 30%

C. 30% to 40%

D. 40% to 50%

E. 50% to 75%

View Answer2. Which of the following is most commonly involved in operating room fires? A. Lasers

B. Elec trosurgery equipment

C. Argon beam coagula tors

D. Fiber-opt ic i l luminat ion systems

E. Def ibri l la tors

View Answer3. Which type of tracheal tube will cause the most severe damage to the tracheobronchia l tree if i t catches fire?

A. Red rubber

B. Si l icone

C. Laser resistant

D. PVC

View Answer4. In order to filter out potentia l airborne contaminants, the surgical mask must be able to fi lter particles of which size? A. 30 microns

B. 3 microns

C. 0.3 mic rons

D. 0. 03 mic rons

E. 0.003 microns

View Answer5. Wet cotton gauze will protect the shaft and cuff of a tube best w ith which laser?

A. Argon

B. KTP

C. Nd-YAG

D. CO2

View Answer6. A clear, unmarked PVC tube is relatively resistant to be used with which laser?

A. Argon

B. KTP

C. Nd-YAG

D. CO2

View AnswerFor the following quest ions, answer

• i f A, B, and C are correct

• i f A and C are correct

• i f B and D are correct

• i f D is correct

• i f A, B, C, and D are correct.

7. The factors necessary to start a fire include

A. Igni tion source

B. Fuel

C. Oxidizers

D. Lasers

View Answer8. The characteristics of laser l ight inc lude

A. Coherency

B. Monochromatic i ty

C. Coll imated

D. Ul trav iolen t

View Answer9. Which factors determine the ignition risk with a laser? A. Laser medium being used

B. How the laser is focused

C. Class of laser

D. Exposure t ime

View Answer10. Common errors when using e lectrosurgical equipment include A. Improper p lacement of dispers ive elec trode

B. Insuff ic ient dispers ive elec trode contact

C. Faulty re turn electrode

D. Insu lat ing the patien t from the opera ting room table

View Answer11. Measures to minimize the risk of a tracheal tube fire from a laser include A. Fil l ing the cuf f wi th saline

B. Minimiz ing the laser power densi ty and durat ion

C. Us ing a laser-resis tan t tube

D. Us ing low oxygen concentrations in the inspired gas

View Answer12. Concerning laser-induced tracheal tube fires A. The tube may be contacted by a ref lec ted laser beam

B. The cuff is laser-resis tant

C. Cotton sponges that p rotect the cuff may dry ou t

D. Use of an uncuffed tube increases the risk

View Answer13. Which laser(s) is (are) l ikely to damage the retina?

A. KTP

B. Nd-YAG

C. Argon

D. CO2

View Answer14. Which laser(s) is (are) l ikely to damage only the cornea? A. KTP

B. Nd-YAG

C. Argon

D. CO2

View Answer