Chapter 3-System Overview
The Oasis MRI System is designed for use as a diagnostic imaging device. It
uses a superconductive magnet to produce cross-sectional images that display
the internal structures of the head and body. This system configuration is highly
reliable and has many features such as:
• Patient-friendly open MRJ magnet
• High power and high slew rate gradient system
• Windows XP-based platform workstation
• Large color LCD monitor
• Rich variety of image analysis programs
• High-speed processing capability
This chapter contains the fo llowing information about the Oasis MRJ System:
• Descriptions of hardware and software components and their basic functions
• Procedures for starting up and shutting down the system
• General overview of Windows functions
• Descriptions of the Launcher buttons, wh ich are used to access the windows,
and features of the Oasis M RI System
System Components
The system components of the Oasis MRI System are described in this section.
QS4·53287v4 3. 1
0AS Is~ MRI System Reference Manual T'MhtJent MqHt
Definition: CPU is an acronym for central processing unit.
Note: If you have purchased the optional UPS (uninterruptible power supply) and a power failure occurs, perform the following steps:
1. Close all open windows.
2. Cancel all storage and restore jobs in the Job Queue window.
3. Cancel all printing jobs.
4. Cancel all DICOM transfer jobs.
Data that is in the save process will finish and be saved.
After power has been restored, wait approximately five minutes before rebooting the system.
3-2
Workstation
You will operate the Oasis MRI System from a workstation, as shown in figure
3-1 , which will be located outside of the scan room that contains the Oasis MRJ System magnet. The workstation will include a console unit with the fol lowi ng
components:
• Control unit witl1 a microphone
• CPU w ith a DVD drive
• Keyboard and mouse
• Computer monitor
Microphone
CPU with DVD drive
Figure 3-1. Components of the workstation.
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
Q54-53287v4
System Overview 0 Control Unit
The control unit, as shown in the illustration below, contains controls to operate
the Oasis MRI System and a microphone that can be used to communicate with
patients before, during, and after scans. The contro ls and their functions are
described in this section.
System error LED ........_
Emergency LED
ECG volume
Table set LED CPU ready LED
Abort Talk
Back of control unit
Auto Voice volume
Figure 3-2. Controls on control unit.
MR ready LED
Microphone volume (gantry}
Start
Microphone volume (control unit)
3-3
0AS/S~ MRI System RefeTence Manual n..Pau ... .._
Control Unit Function Description
Button/Control
@ 6 CPU on I off Turns power on or off for control console and system.
@ 6 MR on I off Turns power on or off for units other than contro l console.
~ Emergency Emergency power shutdown of RF (radio frequency) control unit, gradient coils power unit, and patient table.
8 Abort Stops scan.
8 Talk A llows communication to patient on table using microphone.
8 Start s scan. The START button can also be used to pause a
Start scan that is currently running. The scan can be resumed by pressing the START button a second time.
0 Microphone volume (gantry) Adjusts volume of gantry microphone.
Controls on Back Function Description
of Control Unit
0 Microphone volume (control unit) Adj usts volume of contro l unit microphone. MICVOL
0 Auto Voice volume Adjusts volume of Auto Voice commands. AUTO VOICE
0 ECG vo lume Adjusts volume of ECG. ECGVOL
3-4 Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
DVD Drive Control
~ c:::::::::J
=
QS4-53287v4
Eject
System Overview 0 DVD Drive
The DVD drive portion of the workstation a llows you to store and access
information using CD a nd DVD discs, as shown in figure 3-3.
I I DVD multi-drive - · ........_ -F:: ~
a 0 0
Eject
Ready LED
0
0
Figure 3-3. Controls on the DVD drive.
Function Description
Ejects disk.
Ready LED Displays status of DVD drive by fl ashing on and off when drive is storing or restoring files.
To protect a C D or DVD disc from possible corruption, you should remove the
disk from the drive before any system staitups o r shutdowns. This includes daily
system sta1tups, forced syste m restarts (or reboots), daily system shutdowns, and
forced system shutdowns.
You sho uld take the fo llow ing precautions whe n using the DVD drive:
• Do not look inside the DVD drive. Looking directly at the laser light inside
can cause visual impairment.
• Do not move or raise the main unit whi le the disk rotates; such handling can
damage the disk.
• Do not use a cracked or damaged disk. The disk can fly off ins ide the unit
caus ing injury.
• Do not inse1t your fingers into the DVD tray; this can cause injury.
3·5
0AS Is~ MRI System Reference Manual T'MhtJent MqHt
Gantry control panel
3-6
MRI Gantry
The gantry and patient table leading into the Oasis MRI System magnet are
shown in figure 3-4.
Gantry control panel
Patient table
Figure 3-4. Gantry and patient table.
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
Gantry Control Panel
Lateral move counter
Set mode transit/set move
~ n/up move
System Overview 0
Two control panels are located on the gantry. Scans can be started, aborted,
paused, and restarted frorn ins ide the scan roorn by pressing the appropriate
button on either control pane l.
The buttons and indicators on the gantry control pane ls are shown in figure 3-5
and their functions are described in this section. The button lights indicate
operation status.
Longitudinal move counter
Clears the longitudinal
move counter
CLEAR
Ba vail li(J = ~o~oto 0
~BB
lateral move position
Up/down move position
Emergency patient table release
Emergency patient table stop
Move left--+-.-On/off illumination
Out/down move --+----+\
Q54-53287v4
Move right
Scan status display
Laser localizer
on/off
Scan start
Scan stop
Scan suspend/resume
Figure 3-5. Gantry control panel detail.
3-7
0AS/S~ MRI System RefeTence Manual n..Pau ... .._
Gantry Control Panel Indicator
=BBBB
0 <JO I>
D
, .. - ..... ( SCAN': ' ' ........ ,.'
Gantry Control Panel Button
lv STOP~
CLEAR
0
®
3-8
Function
Longitudinal move counter
Lateral move counter
Lateral patient table position indicator
Up I down patient table position indicator
Scan status display
Function
Emergency patient table stop
Emergency patient table release
C lears the longitudinal move counter
Brings the patient table to the up pos ition
Moves the patient table into the magnet bore
Forward (slow)
Set
Speed fixed
Description
Indicates the position of longitudinal direction (forward and backward). When the orig in (the back end of longitudinal direction) has not been detected right after the contro l program has been sta1ted, it is unlit.
Indicates the position of lateral direction (right and left). When the origin (the center of the lateral direction) has not been detected right after starting the control program, it is unlit.
<J: Left 0 : Center r> : Right
: Highest and backmost position : Position other than the highest position.
Lit during the scan .
Description
Quickly stops operation of the patient table and keeps the state of emergency stop until the RELEASE button is pressed. In the state of emergency stop, all control panel buttons are turned off.
Releases the patient table from the state of emergency stop.
Clears the longitud inal move counter to 0. The CLEAR button does not clear the lateral move counter. If it is pressed during the set mode, the set mode is released.
When the patient table is not at the highest position, the IN button functions as an ascending button. As soon as the patient table reaches the highest position, the laser localizer is lit.
When the patient table is at the highest position, the IN button functions as an advance button.
If the IN and OUT buttons are pressed at the same time during a forward move, the patient table moves forward at slow speed.
Moves the laser-localized area of interest to the isocenter of the magnet.
If the IN button is pressed again (released once and pressed again right away), the current speed is maintained.
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
System Overview 0 Gantry Control
Function Description Panel Button
Out When the patient table is not at the back end of the longitudinal move, the OUT button functions as a backward button.
If the OUT and IN buttons are pressed at the same time during Out (slow) a backward move, the patient table moves backward at slow
® speed.
When the patient table is at the back end, the OUT button Down functions as a descending button. As soon as the lowest
position is reached, the laser localizer goes out.
Speed fixed If the O UT button is pressed again (re leased once and pressed again right away), the c urrent speed is maintained.
@) Left move Moves the patient table toward the le ft.
Left move (Slow) If the LEFT and RIGHT buttons are pressed at the same time while moving left, the patient table moves left s lowly.
@ Right move Moves the patient table toward the right.
Right move (Slow) If the RIG HT and LEFT buttons are pressed at the same time while moving right, the patient table moves right s lowly.
Moves the patient table to position the area of interest at the
B center of the magnetic fi eld. The patient table move is carried
Set mode transit I set out whi le the SET button is kept pressed. During the set mode, move the SET button LED blinks and the longitudinal move counter
shows the remaining distance. When the patient table move is completed, the laser localizer goes out.
Speed fixed If the SET button is pressed again (released once and pressed aga in right away), the current speed is maintained.
lill Laser localizer on I off While the LASER button is lit, the localizer laser is on. Th is laser is used for aligning the area of interest. -
I START I Scan start Starts the scan.
EJ] Scan suspend I resume Suspends or resumes the scan.
8 Scan stop Stops the scan.
I LLUMI I On I off illumination Ilium inates the gantry.
QS4-53287v4 3-9
0AS Is-MRI System Reference Manual ...... u .. , .._..
Special State Description
The emergency patient table STOP button or the Emergency button on the
All buttons off control unit has been pressed. After checking the patient table to determine that there is no abnormality, press the RELEASE button to release the emergency stop state.
The patient table control program has restarted and the system is in the
All buttons blinking state of emergency patient table stop. After checking the patient table to determine that there is no abnom1ality, press the RELEASE button to release the emergency stop state.
Button blinking The patient table control program has restatted and the blinking button is on. Press the blinking button to turn it off.
Longitudinal move position When the patient table control program has restatted, this indicator remains indicator off unlit until the origin (the back end ofthe patient table) is detected. When
the origin has not been detected, the patient table can only move backward.
IN button off Move the patient table backward until the system can detect the origin for
(in the state other than operation longitudinal direction.
limit)
Lateral move position indicator When the patient table control program has restatted, this indicator remains off unlit until the origin (the center of lateral direction) is detected. When the
origin has not been detected, the patient table can only move to the center.
LEFT/RIGHT button off Move the patient table to the center until the system can detect the origin
(in the state other than operation for lateral direction.
limit)
3- 10 Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
Note: For more information about the placement of coils, refer to the Oasis MRI System Patient Positioning and Coils Manual.
QS4-53287v4
System Overview 0 Coil Connector
Use the fo llowing procedure to connect and d isconnect coils from the gantry:
I. Open the slip cover on the patient table and connect the coi l cable as shown
in figure 3-6.
Figure 3-6. Connecting the coil cable to the gantry.
2. Aft:er a scan is completed, press and hold the OUT/DOWN button, which
moves the patient table to the backward limit position, then lowers the table
to a position where the patient can get down from the table safely.
3. Disconnect the coi l cable.
4. Remove the receiving coil from the patient.
• Do not allow the coil cable to be caught in the table while the table is
moving.
• Do not al low the coil cable to wrap around the patient. Place a towel or mat
under the coil cable so that it does not directly touch the patient's skin.
• Make sure the coil cable is not tangled o r looped.
• Do not use a coil cable with a damaged cable sheath o r metallic part
exposed by a worn-out sheath.
• Do not leave the receiving coil and cable inside the gantry or on the table
when not in use.
• Move the patient table with care.
3. 11
0AS Is-MRI System Reference Manual Thehtltnl ...,....
System Specifications
3. 12
Magnet System
The Oasis MR1 System is based on a high-performance superconductive
magnet. it delivers high-field clinical uti lity in a patient-friendly design.
Magnet Type- Superconductive
Magnetic Field - Vertical
Field Strength - I .2T
Homogeneity-< 0.3 ppm@35 em DSV (VRMS)
Shimming- Passive iron shim Per-patient higher-order active shim (HOAST)
Shielding-Active magnetic shielding
5-Gauss • Horizontal- 4.2m • Vert ical- 3 .2m
Cryogen System- Helium cryogen
Refill Interval-Annually with continuous power and Hitachi-approved maintenance
Gradient System
The powerful gradient system prov ides for rapid acquisition of high-resolution
images. The high peak strength and slew rate a llow the user to select short echo times, small fields-of-view, high matrices, and thin sl ices.
Peak Strength - 33mT/m
Slew Rate-] 00 T/m/s
Shielding-Active
Cooling Method - Water cooling
Max. Avg. Acoustic Noise- < 99 dB
SoftSound™-Mechanical gradient noise dampening • Noise Reduction- Low noise sequence mode
SoftSound™ is a trademark of Hitachi.
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
Patient Access Area
Non-patient Access Area
Patient Access Area
on-patient Access Area
QS4-53287v4
System Overview 0 Patient Management System
The short bore and wide aperture create a comfortable patient environment
without compromising cl inical utility. The large, high-capacity patient table
easily supports a broad range of patient body types.
Magnet Gap- 17.32 in. (44 em)
Weight Limit- 660 lbs. (300 kg)
Table Width-32 in. (82 em)
Total Longitudinal Travel-> 7 ft.
Vertical Range-49 - 88 em
Class 11 Laser-+/- I mm accuracy
• Positioning - Automatic movement to isocenter
Spatial Gradient of the Main Magnetic Field
Oasis Serial Numbers M001 through M1 09 and M951
Position Field (T) G r adient (T/m) Product (T2/m)
R(mm) Z ' (mm) BO Grad BO BO x Grad BO
800 220 1.89 21.5 40.6
800 223 1.93 22.4 43.3
Oasis Serial Numbers M11 0 and above
Position F ield (T) Gradient (T/m) Product (T2/m)
R(mm) Z 1 (mm) BO Grad BO BO x Grad BO
800 220 1.96 24.2 47.4
800 223 2.06 25.2 51.0
1 Refer to figure 1-1 and 1-2 for x, y, and z axes.
3. 13
0AS Is~ MRI System Reference Manual TM h:Uent M..qiMI
Starting Up and Shutting Down the System
3. 14
This section describes the procedures for starting up and shutting down the
Oasis MRI System.
System Startup
If the system is in Monitor Service mode, click the Shutdown button to end
Monitor Service mode, as shown in figure 3-7.
5~ Monitor Service
ff~JO'" wrotWwM ~ .... (. ~-UIII:lf: dawloa..•utD.INtlC m.d nab'I"IORof'IIV'Wim notr6crs '-"'---l
~n~cktl ·~· bua.on bttowto etld ~s.-v-
[C.utlon]
Corl.rrn. th.f ndwortl.:CII'YIKbon 6wM:ew ~ 1C1 .net the lnt.n.t. ~ ~ c.n t- aoc4"'1Hd 2Hloii"C(J~Miaf'l ~ IY.t.nl l•I'IM •Y••I.ibl•l( U. trwn.t, (bro.dbend CCif1nM:tlonl. CIM'W10t IN ~ci
Figure 3-7. Monitor Service mode screen display.
If the system is not in Monitor Service mode, use the following procedure.
To power up the system from shutdown, fo llow these instructions :
1. Make sure the patient table is at the backmost position. If the table is in the
magnet, use the release bar to manually pull the table out of the magnet.
2. Press the C PU on/off button (figure 3-8) to power up the control unit.
3. Press the MR on/off button (figure 3-8) on the control unit to power up the
MRI system unit and patient table.
CPU on I off
MR on I oft
Figure 3-8. The CPU and MR on/off buttons on the control unit.
Copyright@ 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
Note: When a CD or DVD is ejected immediately after system booting. the system occasionally displays the TriaiOptlonManage.exe -No Disk window. Click Continue when this window is displayed to continue normal system booting.
Note: The User name and Password can only be changed by the Account Administrator initially established by the Hitachi field service engineer.
OS4·53287v4
Syslem Overview 0 4. After about three minutes, the Log On to System window will open, as
shown in figure 3-9.
OK J [ Sh.Jtdown J
Figure 3-9. The Log On to System window.
Logging On
To log on to the Oasis MRI System, follow these steps:
I. Type your User name and Password.
2. C lick OK.
The Launcher w indow opens, as shown in figure 3-10.
Figure 3·1 0. The Launcher window.
The Launcher Toolbar at the top of the window contains buttons that allow you
to access and perform the functions of the Oasis MRl System. The Launcher
Toolbar and its functions are explained in detail later in this chapter.
3. 15
0AS Is-MRI System Reference Manual YbehtJ ... tlilqHt
Note: The software is able to reconstruct images from the data it has acquired up to the point at which a scan is stopped, provided that at least half the data needed has been acquired. This reconstruction cannot be done when a Time of Flight (TOF) sequence is used, or when the Half Scan or Half Echo options are turned ON.
3. 16
Stopping or Starting a Scan That Is in Progress
You can use the online features of the Oasis MRJ System or the buttons of the
control unit or gantry control panel to pause or stop a scan that is in progress.
Pause During Prescan
When the START button is clicked in the lower right corner of the Exam
window, a prescan operation begins and the START button changes state to
become the PAUSE button.
While the system is performing the prescan operation, clicking the PAUSE
button has no effect until the prescan is complete. The entire prescan continues
to completion.
Pausing a Scan
To pause a scan that is currently running, click the PAUSE button in the lower
right corner of the active Exam window (refer to Chapter 5, Exam WTndow
Scanning Functions, for more details).
The scan also can be paused by pressing the PAUSE button on one of the gantry
control panels.
The START button on the workstation control unit can also be used to pause a
scan that is currently running. The scan can be resumed by pressing the START
button a second time.
Restarting a Paused Scan
To restart a scan after it has been paused, click the CONTINUE button in
the lower right corner of the active Exam window (refer to Chapter 5, Exam
Window Scanning Functions, for more details).
You can also restart a scan by pressing the START button on the workstation
control unit or on one of the gantry control panels.
Aborting a Scan
To abort a scan that is currently running, click the STOP button in the lower
right corner of the active Exam window (refer to Chapter 5, Exam Window
Scanning Functions, for more details).
You can also abo11 a scan by pressing the ABORT button on the workstation
control unit or on one of the gantry control panels.
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
Note: Remove any DVDs from the DVD drive before system startup.
Note: After the remote desktop connection is established (by selecting Monitor service), the workstation computer monitor switches to "personal information protection mode". The following restrictions exist for personal information protection mode:
• The Patient Registration window closes
• The standard patient registration process becomes unavailable. Use the Rapid Registration process to register new patients.
Q54-53287v4
System Overview 0 Normal Logging Off I Shutdown
To log o ff o r shut down the system, cl ic k the Shutdown button on the Launc her
Tool bar.
The Shutdown System window appears. C lick the down arrow to choose an
option, as shown in figure 3- 11.
Shutdown System
F.iitactii MRI System
Please choose lhe operation to perform.
Monll!lr ser.-l::e
~a Lock
b====~~CK _ _JI [ earcec:J
Figure 3-11. The Shutdown System window.
You can then select the options described be low.
Monitor service - This is the default option in the Shutdown System window
and the selection recommended by Hitachi. This setting (figure 3- 12) a llows the
Service department to continua lly monitor the system status (for example,
cryogen sta tus).
Monitor Service
"~-....,....v.6eveoe..,...~•..,.. ..... •~--4 "'llbb'u\IOR~,...,....,.,t""...,
,._ c:lidt. ·~· tll.lttorl ~toeNNomor s--.
leo...-) ~ thlltN't-'o ........et'IWI ................. ,. ""' ..... (~ ...-n.-.N.~Me«~ 2'f....,....,.._MOntorillt~'•I'WI<t......_.lf ....................... ~ .... -. ........
Figure 3-12. The Monitor Service option.
3. 17
0AS Is~ MRI System Reference Manual 1'1MI hUnt llqnet
Note: To close the Shutdown System window without choosing an option, click the Cancel button instead of the OK button.
Note: Remove any DVDs from the DVD drive before performing forced shutdowns or restarts.
3- 18
Shutdown-To shut down the system, follow these steps:
l. Press the MR on/off button on the workstation control unit.
2. C lick Shutdown from the Shutdown System window list.
3. Click OK
The system will shut down and the power supply will be shut off.
Logoff- To log off the system but leave it running, follow these steps:
I. Click Logoff from the Shutdown System window list.
2. Click OK
This ends your session, lea\'ing the system running on fu ll power. The system
will display the Log On to System window. To begin a new session, a user must
enter a User name and Password.
Lock- To lock the system for security purposes (for example, if you need to
step away for a few minutes), follow these steps:
I. C lick Lock from the Shutdown System window list.
2. C lick OK
The system is locked until you re-enter your password.
Forced System Shutdown I Restart
You may not be able to use the normal system shutdown procedure if the system
is " locked up" and not respondi11g. In these instances, you may be able to use
the forced system shutdown I resta1t methods described in this section.
!ll Caution Forcing a shutdown or restart may interfere with saving information. Whenever
possible, use the Shutdown button on the Launcher Tool bar to perform system
shutdowns as described in the "Normal Logging Off I Shutdown" section of this
chapter.
Before performing a forced shutdown or restart operation, determine that the
fol lowing cond itions have been met:
• The system is unable to perform any operations.
• All windows that can be closed have been c losed and Exam windows have
been c losed for at least one minute.
Copyright@ 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
Q54-53287v4
System Overview 0 • If possible, the following processes have been completed:
- Scanning/reconstruction
- Post processing
- DVD storing/restoring
- Transmission/reception of network messages
The Forced System Shutdown button is located at the bottom right of the screen.
Right-click the button and select an option from the menu that appears, as
shown in tigure 3-13.
Figure 3-13. Forced options.
Forced Shutdown of the system-Select this option when the nom1al
shutdown process is not working properly (refe r to the Normal Logging Off I
Shutdown section in th is chapter for more deta ils). Use the fo llowing procedure:
I. C lick Forced Shutdown of the system.
2. A Confirmation window opens. as shown in figure 3- 14. C lick Yes to
continue or No to stop the forced shutdown.
~innation ]
Do )'OU fO<ce lhe ] .,.tem<!Vdown?
~et _j 1 tio 1
Figure 3-14. Shutdown Confirmation window.
3. If Yes was c licked, a message and progress bar is displayed while the Oasis
MRJ System shuts down.
4. A message is displayed stating that Windows is shutting down.
5. The CPU REA DY and MR READY LEOs on the workstation control unit
go out and the monitor screen goes black.
6. Wait at least 60 seconds after the CPU READY and MR READY LEOs
have gone off before restarting the CPU. Refer to the System Startup section
of this chapter for normal startup instructions.
3. 19
0AS Is~ MRI System Reference Manual T'MhtJent MqHt
Note: During operation, the End Program I This program is not responding dialog box may be displayed. If this happens, click the Cancel button to continue.
3-20
Forced resta rt of the Application -Select this option if a single application
(for example, a post-processing task) is not functioning properly. A forced restart
of the application takes the least amount of time. Use the fo llowing procedure;
I. Close all windows that can be c losed.
2. C lick F orced resta rt of the Application.
3. A Confirmation window opens, as shown in figure 3-1 4. C lick Yes to
continue or No to stop the forced resta11.
y~ I I NQ
Figure 3-15. Application restart Confirmation window.
4. lfYes was cl icked, an information w indow is displayed in the lower right
portion of the screen to advise you that the application has been stopped
safely. Another information window will advise you that the application is
being resta1ted.
5. If more than one application is open during the forced restart, a ll windows
will be safely c losed and restarted.
Once these forced shutdown/restart operations are complete, you must log on
to the system again. The Log On to System window is displayed (refer to the
Logging On section of this chapter for more information).
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
System Overview 0 Basic Windows Operations
Q54-53287v4
The Oas is MRI System is based on Microsoft Windows and uses many standard
Windows functions and commands. lfyou are familiar with Windows, you
will find that the Oasis MRI System windows and menus operate in a way
similar to those of other Windows-based applications. If you are not fam iliar
with Windows, make sure that you review the information in this section about
general Windows operations.
Mouse Operations
To move around the Oasis MRI System windows, you will use a mouse. The
mouse operates the pointer that appears on your computer screen, as shown in
figure 3-16.
Pointer-Marker that moves on the screen when the mouse is moved. Normally, it is an arrow. However, its shape may change depending on the operation.
Figure 3-16. Mouse terminology.
Screen
Wheel button
Mouse
3. 21
0AS Is~ MRI System Reference Manual T'MhtJent MqHt
Note: The directions in this manual have been written for a standard right-handed mouse button.
3-22
Mouse Buttons and Scroll Wheel Functions
You will use the left mouse button, the wheel button (between the left and right
mouse buttons), and the right mouse button to perform various functions as
described below.
To Click an Object on Your Screen
Point to the target object (such as a button or folder displayed in the window).
Press and release the left button once so that a click sound is heard. This
operation is referred to as a "click" and is used to select buttons, menus, and
other items.
Clicking with the right mouse button is referred to as a " right-click" .
Right-clicking a target object (a series, a n image, or similar item) may display
a sho1tcut menu o r submenu of tasks or commands. These menus wi ll vruy,
depending on the target object, and are useful for completing tasks quickly.
To Double-Click an Object on Your Screen
Point to the target object, then quickly press and re lease the left nnouse button
twice in rapid succession so that a click-click sound is heard. This operation is
referred to as a "double-c lick" . Double-c licking can be used to open folders and
perform other operations.
To Drag an Object on Your Screen
Point to the target object, hold down the left mouse button, move the mouse to a
new location and then release the mouse button. This operation is refened to as
a "drag", and is used for operations such as moving items to different positions
o r loading output series to be used as input images. Perfo rming the same
sequence of steps, but using the right mouse button is referred to as a "right
drag" .
To Move Up and Down in the Windows
The wheel button is used for moving up and down (scrolling) in the various
computer windows. Using the wheel button is particula1·1y he lpful when viewing
images in the viewports. Rotate the wheel button forward to scroll up in the
document o r window. Rotate the wheel button backward to scroll down in the
docume nt o r window.
Changing Mouse Functionality
If you need to chru1ge the functionality of your mouse (such as the button
configuration or c lick speed), you can do this in the Mouse Properties window.
This window is described in the System Settings Windows section of Chapter 9,
Additional Launcher Functions.
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
Q54-53287v4
System Overview 0 Window Features
The Oasis MRT System software provides numerous windows to perform
operations. These windows have some standard features, which are described as
follows.
Title Bar
The name of the window appears in the title bar (figure 3-1 7). The right corner
of the title bar may display some or all of these buttons:
Minimize- Shrinks the window from view, but sti ll a llows access from tbe
task bar (does not close the window).
Maximize- When not already fu ll size, en larges the window to full s ize.
Restore - Restores a maximized window to its previous size (not shown in
figure 3-17).
Close-Closes the window completely.
/Window title
~Patient .{eg!;tratlon Setting
Minimize Maximize Close
~~ ----
Figure 3-17. Title bar and buttons.
3· 23
0AS Is-MRI System Reference Manual YbehtJ ... tlilqHt
3 • .24
Menus and Toolbars
Many windows will have a menu under the title bar. This menu is a list of
various operations that can be performed on that window. When you click a
menu (thus "opening" the menu), the commands that can be selected are listed
below that menu. Once a menu is opened, you can point to other menus to
display their contents without actually clicking them.
J n figure 3- 18, if you open the Main menu, Edit Patient, Save Image, and other
commands are displayed. Click the desired command in the list. For example,
Click Exit to close the window.
Figure 3-18. Typical Edit menu commands.
Under the menu bar, there may be a toolbar displaying buttons representing
other functions that can be performed (figure 3-19). When pointing to a tool bar
button, the name of the button will be displayed in most cases.
When the tool bar button is cl icked, you activate it (the background color of the
button may change to indicate that it has been selected). Cl icking a toolbar
button accomplishes the same result as selecting a command from a menu.
Figure 3-19. Typical too/bar button detail.
Tool bar settings in the Exam window, Review Task and Post Processing tasks
can be applied globally.
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
Q54-53287v4
System Overview 0 Window Buttons
Many windows have buttons in the bottom right corner that operate as fo llows:
OK- Saves selections or changes made in the window and closes the window.
Apply- Saves selections or changes made in the window, but the window
remains open.
Cancel-Closes the window without saving your changes.
Buttons that are unavailable for use are shown by the OK and Apply buttons in
figure 3-20.
Figure 3-20. Unavailable button detail.
Taskbar
The taskbar is located at the bottom of the screen. It displays buttons labeled
with the names of the windows, tasks, and I or Launcher buttons that are open
or runn ing. The taskbar button corresponding to the currently active window
or task will be displayed with a dark gray background, while inactive taskbar
buttons will be displayed with a I ight g ray background (figure 3-21 ).
When a window is minimized (by c licking the minimize button in the top right
corner of the window), the window name appears as an inactive task bar button.
Wl1en a window is max imized (by c licking the taskbar button to bring the
window back up for display), the window name appears as the active taskbar
button.
Figure 3-21. Typical taskbar detail.
3· 25
0AS Is~ MRI System Reference Manual T'MhtJent MqHt
3-26
Basic Window Procedures
This section contains general procedures for working with the Oasis MRI
System windows.
Selecting an Item in a Window
To select an item in a window, click the item with your mouse pointer. In
genera l, the item wi II change color to indicate that it has been selected, as shown
in figure 3-22.
Basic I AI
''·" II 1¥
[ Reset n ~ :B ~ Click to select 144 :B j30 3
Figure 3-22. Selecting an item in a window.
Selecting Multiple Lines /Items
You can select multiple items using the SHIFT key and CTRL key as follows.
Use the SHIFT key to select multiple items (all adjacent) in a data list or group:
I. Click the first row in the desired group.
2. Hold down the SHIFT key.
3. Click the last row in the desired group.
4. Release the SHIFT key. The first, last, and all rows in between are selected.
Use the CTRL key to select multiple items in a data list or group, but not
adjacent to each other:
I. Hold down the CTRL key.
2. Click cl ick each desired row or item.
3. Release the CTRL key. All items that were clicked are selected.
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
Q54-53287v4
System Overview 0 Entering Words or Numbers
Often you will need to enter or change words o r numbers in d1e system (for
example, the patient's ID number or name). The fo llowing example shows how
to enter or change information:
I. C lick the box in which you want to make an entry or change ( in figure 3-23,
the Patient ID box).
Click
Figure 3-23. Making an entry or a change.
The area may be highlighted in blue, ind icating that it is ready to receive
input. If the area is not blue, drag across the current entry to highlight it
(figure 3-24 ).
• Patient Registration
Patient Information GetFrom MWM
PaMnt to · ~689
Figure 3-24. Text box contents highlighted in blue, ready to be changed.
2. Enter the new input. The previous contents will be replaced.
3 . Press ENTER to set the input, or press the TAB key to move to the next
parameter, or click outside of the text box to reposition the pointe r.
Using List Box Controls
The Oas is MRI System often provides list box contro ls (figure 3-25) that allow
you to easily se lect information from a list. When an arrow is visible, it indicates
a list the user can view by c licking the arrow:
Scan Control
Figure 3-25. List box control examples.
3· 27
0AS Is-MRI System Reference Manual YbehtJ ... tlilqHt
3 . .28
The list is displayed with the currently-selected option high lighted, as shown in
figure 3-26. To select a different option from the list, cl ick the desired option.
The new selection is displayed in the box and the list closes.
Scan Control
Ql£111 • .1.1.1 Evel)l [v •• ;, Normal Tv
.. Normal ~narric FW!o
G
'" p
Scan Control ti;n;,;.;m liffiil(j§=l,ldl.!.@
.. . Evel)l
~010 off
nn m n
Figure 3-26. Selecting a different option from the list.
vJ .... , .... ,
c
N p
Some selections will have a submenu, indicated by a small right arrow as shown
in figure 3-27. To open the submenu, point to the command with the small right
arrow on the main menu, which causes the submenu to open, and then click the
desired submenu command.
Figure 3-27. Submenu example.
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
6o "' ~--,. :10 .. ~ 130 .. .... ~ 132
& ... ... ... ~ 11'11 6, 1311 ... ... ~ ... E >01 6o ,. ~ .. & 100 & .. ~ 142
'· ,.,
p.,_. .. _
System Overview 0 Using Right·Ciick Menus
In some areas of the system, a menu will appear when you c lick the right mouse
button, as shown in figure 3-28. The right-click menu commands are context
sensitive, which means they will change depending on w here you are in the
system. T he commands available will be those that are most likely needed when
working in that area.
c-. ... .,. 0.1~ P.t"'"' II) ,.,.,_" L•t•n"" OASIS A800tA:N /LIVER 8ARJATllJC 0>/lt/2009 :144200 L'- -0AS1$ A800IEN I ~p 01/211'2010 01\Sl$ Nll<l.( 01/2712010 • 12a4 OM13MMII ---~~~~ - . """"' OASJS 8R.AIH --OASIS 8RAIH I PIT\.ITAAV .... ........,. OASIS 8ftAlN ~~~~ Sll"ttM~Dtli (O&QN.,MlA OA$1$ QR(AST - ""'-""" OASIS 8RtASl I lto.l>l.NfT& OASIS CAAOIAC - • K21AVo1001 OASIS CAROnO t.I\A - .S«i2088 O~SCAAOTID~ ---- -100 OASIS CAAOnD MAA •W900(10()2818:P OASTS CEfMCAL ---- tl::tJO?IO OASlS CER\I!CAL """'- 10161814 o~s crAV~CAL ---OIISJ$C(fMCAI. 04107/2010 -o!l-'!939 OASIS CLRVICAl I OWl C./07/2010 t X21B-G13
L'-...... -..... ..._.., On• ...... .._, .,...._ v..c.o.r -·· Nod es-c-s..,. c-s-C- ScMw c-._
-Riaht --..... ..... ..... ..... ..... -..... -..... ..... ..... --•OJJ10411 OASIS 8R.NN 08122121110 14 23 Z3 8HAIN -,. s...-._ ... _ s.,_ l:b•• n,... - ... ~-•• AlMI OWl osJn/201o ••~e'49 1'00Wf.l~ ., -~- 015/~11010 ··~· .. tOOWCPI
•• Axllti i)IM: 08/nnoto , • .,e 49 toow r~ .. s.1 n FU Olt/2212010 1428 12 , .. "" • 10 A:ocilll ~~" ootn no•o ••~:zoo t~GC ... .-.I fLAIR 08/HnotO 14.3$ • ..,.m ... Axi.l Tl FUM 08/22/2010 14a9c41 . .... ••• Cor FL.-;IR osm1201o 144334 . .... .I. Axiitl Tl Fl .AIR • 08/H/1'010 150409 20fJR ... C.r1 n FLAIR • 0 os/nhO•o 1::10148 fi)JJR
Q54-53287v4
..... """ loDC
... _ rov TR
AX >00 ..... , AX >00 ...... AX 200 .... 2
SN3 2:10 .. , AX 220 .... AX 220 .... AX ,. ,., COR 2>0 1111 AX
,. ,,., COR 220 ....
Tl"
"" 105 100
•• 120 100 ,. .. ,. ••
"""' • c-· ...... "' .. .. •• •• •• , . ..
....... :r• ...... ..
Figure 3-28. Right-click menu example.
Generally, the commands on right-click menus can also be accessed through
various other menus in the system. Right-click menus are provided as a
convenient way to quickly access the functions you need the most, particularly
as you become more fam iliar with the system.
3-29
0AS Is-MRI System Reference Manual YbehtJ ... tlilqHt
3-30
Moving or Resizing Windows
A window can be moved on your screen by dragging the title bar to a new
location.
Some windows can be resized by positioning the pointer on the edge of the
window, at which time the pointer changes to a double-arrow symbol (+-+).You
can drag the edge of the window to en large or reduce the window size.
Scroll Bar
Some windows (and viewports) have a scro ll bar that is used for easy movement
through the area. It is located in a vertica l position a long the right side of the
window or in a horizontal position at the bottom of a window. The scroll bar can
be used for movement in the following ways:
• To move one line (or image) at a time, click the up scro ll arrow at the top of
the scroll bar or the down scro ll arrow at the bottom (figure 3-29).
• To move one screen at a time, c lick the scroll bar above or below the scroll
box.
• To move up or down more freely in a window or viewport, drag the scroll box.
Eli Automatic Protocol
v~ Scroll arrow
CKJ (c-.1)
Figure 3-29. Scroll bar details.
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
Note: Pressing the ALT key acts as a toggle, allowing you to activate and deactivate the shortcut keys.
Q54-53287v4
System Overview 0 Using Shortcut Keys
Some users prefer to use the keyboard (via shortcut keystrokes) instead of
selecting items with the mouse. This is an optional approach to selecting menu
commands, you are never requi red to use this approach.
If you choose to use shortcut keys, follow the instructions below:
I. Press the ALT key. Menu items will be displayed with one letter underlined,
as shown in figure 3-30 (if underlined letters are not displayed, click
somewhere in the window to activate them).
2. Press the letter key corresponding to the underlined letter in the menu you
would like to select. The commands for the menu wil l be displayed, as shown in figure 3-30.
h Patient Director S.tudy E,d1t l{,lew
Add Another Study
B.egistBr Patient
Start !;xamlnatlon
p
Figure 3-30. Menu commands with ALT key shortcuts.
3. Press the letter key corresponding to the underlined letter in the command to
be performed. In figure 3-31 , X cou ld be pressed for the Exit command.
h Patient Director Edit l{,lew _
Add Another Study St . &egistBr Patient
Start t;xamnatlon orl
EKit
Patient ID r
Figure 3-31. Exit command ALT key example.
4. After a command is performed, the command menu closes and underlined
letters are no longer accessible. Press the ALT key again to conti nue using
the sh011cut keys.
3. 31
0AS Is~ MRI System Reference Manual T'MhtJent MqHt
Note: In some instances. menu commands will not be displayed with underlined letters, but will still respond to the use of the arrow keys.
Menu commands can also be selected by using the UP ARROW, DOWN
ARROW, LEFT ARROW, and RIGHT ARROW keys on the keyboard. Follow
the instructions below:
I. Press the ALT key. Menu items will be displayed with one letter underlined,
and the menu item furthest to the left wi ll be highlighted, as when selected.
2. Use the right and left arrow keys to navigate across the menu bar.
3. Use UP ARROW and DOWN ARROW to open a selected menu and
navigate through the menu commands. Use LEFT ARROW and RIGHT
ARROW to open and close any additional fly-out menus that appear beside
a menu choice, indicated by an arrow symbol next to the menu choice.
4. Press ENTER to perform a menu command.
Launcher Buttons
3-32
Launcher buttons are used to select the various operations available on the Oasis
MRl System. These buttons are located on the Launcher Tool bar, which can be
docked in different locations on the screen. There are twelve Launcher buttons
set up by applications during initial training. Additional Launcher buttons
and various associated menu commands can be added or deleted using the Customize feature on the Launcher Tool bar menu.
Function Keys
The F9, Fl 0, F 11 and F12 funct ion keys can be ass igned shor1cuts to specific window width (WW) and window level (WL) settings. To assign these
shortcuts, proceed as follows:
I. Right-cl ick the Launcher Tool bar and c lick C ustomize from the menu that
appears (figure 3-32).
Figure 3-32. Accessing the Customize command.
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
Q54-53287v4
System Overview 0 2. In the Launcher Customization window (figure 3-33), click Function Key
Settings from the Available launcher buttons list, c lick System Settings
from the Current launcher buttons/Drop-Down menus list and cl ick the
right arrow button ( ~ ) .
.,. Launcher Cus tomlzatlo n
1. Click
I o ..... ...;o
Figure 3-33. The Launcher Customization window.
3. Cl ick OK to save this customization and c lose the Launcher Customization window.
2. Select System Settings from this list
3. Click
4. C lick the System Settings launcher button ( t · ) and c lick Function Key Settings from the list that appears (figure 3-34).
Display Properties
GSCFTool
r-.............. Au1D Archive Setthg
~~----~--~~·~ ~~Setthg
·~---~~ Lode: Settng
Figure 3-34. Accessing the Function Key Settings command.
3-33
0AS Is~ MRI System Reference Manual 11141 htlet~t olllqHt
3-34
5. In the Function Key Settings window, enter the desired Window Width and
Window Level settings for each of the function keys (figure 3-35).
~ Function Key Settings ~ 'WWIY/L ~hor!cull:ey se«1ng'
Window Width Window Level
F9 ~ ~ 1500 :±J
F10 J1200 ffi 1600 :±1
F11 11400 :H poo :±1
rsoo ill Click to load
F12 default values
OK I I Cancel
Figure 3-35. The Launcher Customization window.
To use the system default settings, cl ick Set to defaults. The actual default
values are displayed in the F unction Key Settings window in figure 3-35.
6. Cl ick OK to save your WW and WL shortcut settings and close the
Function Key Settings window.
Launcher Buttons
The 12 Launcher buttons in the Launcher Tool bar are shown in figure 3-36:
Figure 3-36. Launcher Too/bar buttons.
These buttons were chosen to be default buttons because they are frequently
needed by most users. The functions of the buttons are described as follows:
l e Patient Directory- Displays the Patient Directory window, wh ich is used for
managing patient information.
Patient Regist•·ation -Displays the Patient Registration window, which is
used for registering patient and study information.
tl\ P rotocol/Task L ibrary-Protocols and/or tasks can be added or ed ited in this
window.
Copyright© 2012 by Hitachi Medical Systems America. Inc. All rights reserved.
Q54-53287v4
System Overview 0 Post Processing-Opens a menu that provides access to post-processing tasks
such as MIP, MPR, diffusion analysis, and other similar tasks.
Review Task - Displays the Review Task window, where images can be
reviewed, manipulated, and filmed.
~ Filming Tool - Opens the Filming Tool window, where images can be
manipulated before filming.
Ea Waveform - Opens the WaveForm window, where you can observe
waveforms of electrodes or sensors that are attached to the patient.
~ Stopwatch - Opens the StopWatch window for activation of the StopWatch
settings.
I · System Settings- Opens a menu that provides access to a variety of windows
used in the setup of various system functions (such as Patient Registration
Setting window, Mouse Propea·ties window, and other similar functions).
~ Maintena nce- Provides access to system check and control functions, as well
as to a menu for tests to be performed by the Hitachi field service engineer.
G. Help-Opens a menu that provides access to Version Information, Sentinel
Messenger, and Internet Browser (default menu), which are described later in
this chapter.
Shutdown- Opens the Shutdown System window, with various shutdown
modes available.
3· 35
0AS Is· MRI System Reference Manual YbehU ... tM.acHt
3-36
Docking and Displaying the Launcher Toolbar
You can choose where you want to display the Launcher Toolbar on the screen.
Right-click the Launcher Toolbar to display a menu, as shown in figure 3-37.
Dock Left
~ Dock Top
Dock Bottom
Auto Hide
Always on Top
Customize
Figure 3-37. Docking and displaying options for the Launcher Too/bar.
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
QS4-53287v4
System Overview 0 The menu offers the fol lowing options, which wi ll appear with a check mark
when selected:
Dock Right- Displays the Launcher Toolbar long the right side of the screen.
Dock Left - Displays the Launcher Toolbar along the left side o f the screen, as
shown in figure 3-38.
Figure 3-38. Launcher Too/bar docked left.
3·37
0AS Is-MRI System Reference Manual YbehtJ ... t ........
Dock Top - Displays the Launcher Toolbar across the top of the screen.
Dock Bottom - Displays the Launcher Toolbar across the bottom of the screen.
Auto Hide-Hides the Launcher Toolbar, as shown in figure 3-39. The
Launcher Tool bar can be viewed by pointing close to the edge of the screen
where the Launcher Tool bar is docked.
Figure 3-39. Launcher Too/bar with Auto Hide selected. Too/bar is hidden until user takes action to display it.
3-38 Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
----------------------------------------3 System Overview 0 Always on Top-Displays the Launcher Toolbar on top (or in the forefront) of
all other windows that are opened for display, regardless of where the toolbar is
docked. Figure 3-40 shows the Launcher Tool bar, docked at the top of the
screen, displayed on top of the Protocolffas k L ibrary window .
ht S . 0 I!; E& ~ ... , . ~ 0 . £J ..... "" ....... , Oor:lo f-00
61 1!1 ·--.lo .. :;o .. & ... ... ... & 1:12
& ... ... & ,.,, "' 1211 & ... ~ ... ~ ""' 6o .,. 6o " {,j ... "' .. t .... .. ,
"03110411 OASIS BRAIN .. S.riMN.:t-•• -·I)WI ., .... ,I)WI •• -·I)WI •• S.a 11 r!'.l
••o Aldlll T2 FSE ... AJO.I Fl..AIA
••• A*IT1 FLAIR
••• CorrLAIA
••• Alcilll II fLAIR • . ., Corl Tl HAIM • 0
QS4-53287v4
rJIPlwNw l• ::-
''•tloHtiH. .. S1~0.•• P.tienT OASIS A800UfN I LIVER BARIATRIC rxi/12/2Ca ...,.,. OAs.S AOI)C)U[N I ~p <an./2010 ""...,.,.,... .. OASISANU..[ o•n1no1o •lr.t4
OASI§- oem1to10 111(1)110411 OASIS BRAIN 11/lonGII t1)7f OASIS BRAIN I PfT\.ITARY 01~/1'010 o(IOOOII ... OASIS BANN ~A r./1012010 •VliiW OA~S BREASt 01101/2010 •1181)3 OASIS Dft£AST / IM>lAHTS <11/22:/2010 _,. OAS.S CAAOtAC Olnt/2010 •X21AVoiXIG OASIS CAAOn o p,ARA US/04/1:010 -2'0811 OASaS CAAOTIO MRA 0311011'010 ...... OASIS CAROTID ~A (XJ~IO t'MJOOOOOZI)" OASIS C£fMCN 00/'11/tOII 11~10
OASIS CtRVICAL 04/1112011 IOifH/4 OASIS CUr~-. Ol/0812011 I OASIS CERVICAl. 04/07/2010 .... ... OASIS CDMCAL I OIM 01/0l/7010 • )(:IIB-QI3
0612212010 14 ~ 23 ORNN .,, .... "J ll014P~ l!.00/0110
s.n- O.t• n.. . .... ~~ """"' ,, .. ,. rov IR oet.nno1o 14 H.., >DOW (PI AX ""' ..... osmno•o 141'848 tDOWIJ'I ""' AX - ..... 08/n/2010 14"1841 >DOW 9'1 """ AX - ..... oamncno 14.21!112 ,,. .. SAO t30 847 08/22/2010 14.32.05 ,,. .. AX 220 6>01 OllmnG10 14Zot >OAR AX "" -08/22/1010 14~:41 20fiR AX .... . ,. . OM»/1010 14 43'34 >on• COR '"' 7111 oemnoto t:.Otc» 20fiR AX .... . ,.. 011/»/1'0'0 1301 ... 70F-1R COR ... ,...
o~'" Ul'ltln PIIW~ry
·~ ... ··-._ .. c.~.:
Vuc.,. V•tciiAII,. -· c-c-s..,.. c-._ C- Soine c-._
T1 "lhlc: .. C OI"'tr
"" • ,., • ,., • 10 • ... , 100 s •• .. •• ... _ ••
... _
l nu1.-m .. 28 ?3 24
•• • • •• • • 28
L••.,.IFtr --------..... ---------"' (I!
,j; IS; &
II !;\ ~
• 'I! .., & s.
off
• 0
Figure 3-40. Launcher Too/bar displayed on top of other window and docked at top of screen.
3· 39
0AS Is-MRI System Reference Manual YbehtJ ... tlilqHt
3-40
C ustomize-Opens the Launcher Customization window (figure 3-42),
where you can customize the Launcher Toolbar by add ing or deleting Launcher
buttons. You can a lso create Launcher buttons with your choice of I ist options.
These procedures are explained in the next section.
Figure 3-41 shows a Launcher Tool bar to which additional buttons have been
added through the Customize function (the circled buttons).
Figure 3-41. Launcher Too/bar buttons added through customization are circled.
Customizing the Launcher Toolbar
You may wish to customize the Launcher Tool bar by adding frequently used
buttons or menu choices, deleting buttons or menu choices you rare ly use, or
changing the order of items according to persona l preference. To change the
toolbar, use the Launcher Customization window (figure 3-42).
The Launcher Customization window is accessed by right-clicking the
Launcher Tool bar and then cl icking Customize from the menu that appears. The
Launcher Customization window opens as shown in figure 3-42.
~ ,.,. 11 P.,ientReooso<llion lh p.,...,o.cetoov ll'J Pool Recon (jMIP tSl) MPR S VcbneRonde<W>g ij] AddJStb
II!JT11T2 tJ.ilollusOO~ :llb C£ PeriUSIQn"""""'
~ Veloc(y""""s# ~o,.,..,..,~ .. ., ... to!~ s-·-AAalysiJ I! I~Mge""""'" Js~ fi R-~·~
o;r P • ..: or
c..,nlot..nchlf Wton'ID'~ menu$ L- s.-.....
~ '""' IJw p.,..,..o<ec~ .. v !! p.,..... R091balton ill PtotoediT osk Li>IOI)' t$) P01t Ptocenng lilJ R....,... ,..,._ & Fi...-.gT ... ei Wevefotm &;;?, StOj:l#.,ch loll SydeMS":.:} Maintonanee €) H.-, 13 Slu do•m (jMIP ~ JobQ .....
I OeldSettin!l' I
Figure 3-42. The Launcher Customization window.
Deletes Launcher
items from
Creates a Launcher button with a menu
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
Q54-53287v4
System Overview 0 The window is divided into the two areas described below:
Available la unchet· buttons - Displays available choices of button images and
names and I ist options.
C urrent la uncher buttons/Drop-Down menus-Shows the contents of the
current Launcher Tool bar. The list box at the top of this area contains a list of
options that determine what will be displayed in the area below. Select from the
fol lowing:
• Launcher Buttons-Shows the current Launcher buttons on the toolbar.
• System Settings- Shows the current list options for the System Settings
button on the toolbar.
• Help-Shows the current list options for the Help button on the toolbar.
The items wi ll be displayed in the same order in which they appear on the
Launcher Toolbar.
The folder button in this area (figure 3-42) is used to c reate a Launcher button
with a menu. The red X button (figure 3-42) is used to delete items from the
toolbar or menu commands.
Adding an Item to the Launcher Toolbar
To add a button or a list item to the Launcher Toolbar:
I. ln the Current launcher buttons/Drop-Down menus list, click the option
that specifies what to add and where to add it: Launcher Buttons (button),
Post Processing (l ist item), Waveform (l ist item), System Settings (list
item), or Help (list item). In figure 3-43. Launcher Buttons has been
selected, so a button wi ll be added to the Launcher Tool bar.
' Launcher Customlzatlon
Select what to add I where to add it
Figure 3-43. Adding a launcher button to the Launcher Too/bar.
3. 41
0AS Is~ MRI System Reference Manual T'MhtJent MqHt
2. In the Available launcher buttons list, click the item you want to add
(figure 3-44).
3. Click the right arrow button to add the item to t he Current launcher
buttons/Drop-Down menus box. The added item will be shown at the
bottom of the list, and will still be listed in the Available lal!lncher buttons
list. Use the up or down arrows (figure 3-44) to change the display order of
the items on the Launcher Tool bat·.
•..- launcher Customlzatlon
U Text
lf Pal;.,tAegod!"""" [h Po1101110110Clmy ~ Pot:~Aeccn 0 MIP c$J MPA til Vcbno Ron<lomo m AddiS..., II!JT1112
C...,. lal.n=hef bU:torwtllop-0CMn menus L....,.,Bcttom v {]~
2. Click
"""'""" Aev~ew Task
from this list IIi D~""'"An¥~ct:PetflnKwl~*' ~ vo~ocoy!\t>ol)ltis ~0_.-.:~ .Flt .. lil!]Specbo._,~
& fmnglool 6i 'W-O<m if& Stop\llalch
Definition: ~. DICOM refers to Digital «r
Imaging and Communications in Medicine. fi\
3-42
a ·--~ J Solchng li'ii.Jl...;..·~
0 Show uted Modo opooe
Ell s,...em Sethi'J' a ..... enance i)H~ a s~-0 MIP ~ JcbQueue
OK Cane~
Figure 3-44. Adding an item to the Launcher Too/bar.
Use these buttons to change the display order of Launcher Toolbar items
4. Click OK to apply your changes and close the Launcher Customization
window. The added item appears on the Launcher Tool bar.
The functions of some additionally-available Launcher buttons a re described as
fo llows:
Exam Window-Displays the Exam window, where studies are set up and
performed.
J ob Queue- Displays the Job Queue window where you can track the progress
offi lm ing, storing, DlCOM, or post-processingjobs.
Pati.ent Table Information-Displays the Patient Table Information window
for tracking of patient table movement and coil mode settings.
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
Q54-53287v4
System Overview 0 The various system settings and post-processing tasks, as shown in figure 3-45,
can be added to the Launcher Toolbar as individual buttons or as options in one
of the list items. All of the system settings have the same default button image.
The post-processing tasks have different button images associated with each
task. •• Launcher Customlzatlon
System settings
OK I l Ctncol
·~ Launcher Customlzatlon
Post-processing tasks
Figure 3-45. Examples of button images.
3· 43
0AS Is~ MRI System Reference Manual T'MhtJent MqHt
Note: Default Launcher buttons and I or list items can be deleted from your Launcher too/bar, but will be restored when the Default Settings button is clicked.
3-44
Deleting an Item From the Launcher Toolbar
To remove a button or list item from the Launcher Tool bar, perform the
following steps:
I. In the Current launcher buttons/Drop-Down menus box, click to select
the item to delete (figure 3-46).
2. Do one of the following:
Click the left arrow button. This deletes the selected item from the
Cur rent launcher buttons/Drop-Down menus box.
OR
• C lick the red X button.
3. Click OK to apply your changes and close the Launcher Customization window. The deleted button or list item is no longer displayed on the
Launcher Tool bar.
2. Click
' '"" EJ' P-R..,otr-
1~:~0>
I<PR Vd~Mre.R~
AddiS.b lli]Tln2
~0-~C!OP-ljv_.,.,_ ~o,.....,.,.,_ . .... \.;.!) Spec;tr~ ~
r!l ......
. ,.,. !J. P-0-ooy Jl.P_R __ ~ Pldoedfl .... Lb ...
mn::::.7.:" & Ffto,glod
.( l;';lw ... -.. ~s.__ .. ,s_s_ ,a ... ..._... €}H$ D stvdoowl
I o~...os-,. I
@) @]
Figure 3-46. Deleting an item from the Launcher Too/bar.
1. Select what to delete from this list
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
Q54-53287v4
System Overview 0 Changing the Location of a Toolbar Button or Menu Option
To change the location (display order) of a button or list item on the Launcher
Toolbar, petform the following steps:
I. In the C urrent la uncher buttons/Drop-Down menus box, c lick to select
the item to move (figure 3-47).
2. Click the up or down atTow buttons on the right side of the window to
change the location of the item in the box.
3. Click OK to apply your changes and close the Launcher C ustomization
window. The button o r menu option you moved wi ll now appear in its new
location on the Launcher Toolbar.
•• Launcher Customlzatlo n
9 s-uood Loe.i o.oe- oooco
as--~o<lo--
Figure 3-47. Deleting an item from the Launcher Too/bar.
1. Select what to move from this list
3· 45
0AS Is-MRI System Reference Manual YbehtJ ... tlilqHt
3· 46
Creating a Launcher Button With a Drop-Down Menu
To create a Launcher Tool bar button, fo llow the procedure below:
1. In the Current launcher buttons/Drop-Down menus list, cl ick the
Launcher Buttons option.
2. Click the fo lder button (this button is on ly accessible when the Launcher
Buttons option has been selected). The Add Drop-Down Menu window
(figure 3-48) opens.
3. In the Add Drop-Down Menu window, enter a name for the Launcher
Toolbar button in the Name box, as shown in figure 3-48 for Post
Processing.
Add Drop·Down Menu
Name Post Processing
Image lroeM !Sial
OK Jl Cancel
Figure 3-48. Entering a name for the launcher button.
4. Cl ick the Image Index button to open the Launcher Image List window,
as shown in figure 3-49.
Launche r Image Us t l-JP fa
mJ ~[1J[IJ~fDJ~ ~ ~~~~~ml~~ [l) g l!j] ~~~ [§i] ~ ~~~~~~~m B;~ (gJ
Figure 3-49. The Launcher Image List window.
5. Click an image for the Launcher button. The Launcher Image List window
closes.
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
Q54-53287v4
System Overview 0 6. The Add Drop-Down Menu window shows the name of the button and the
image se lected, as shown in figure 3-50. Cl ic k OK to c lose the Add DropDown Menu window.
Add Drop-Down Menu
OK II e-el
Figure 3-50. The Add Drop-Down Menu window showing selections made.
O n the Launcher C ustomization window (figure 3-5 1 ), the newly-created
Launcher Tool bar button is added to the bottom of the C urrent launcher
buttons/Drop-Down menus list and to the bottom of the Current la uncher buttons/Drop-Down menus box .
... Launcher Customlzatlon
Figure 3-51. The Launcher Customizatlon window showing the new launcher button.
To create a menu for the newly-created Launcher button, fo llow the procedure
below:
1. O n the Launcher Customization window, in the Current launcher buttons/Drop-Down menus list, c lick the newly-created Launcher button
from the list.
3·47
0AS Is-MRI System Reference Manual YbehtJ ... tlilqHt
2. Select the desired menu commands from the Available launcher· buttons box and click the right arrow button to add the menu commands to the list of
the newly-created Launcher button. Figure 3-52 illustrates the process of
adding menu commands to the newly-created Post Processing Launcher
button.
1. Click the new Launcher button from this list
2. Select the commands here that you want in the menu
Note: Created Launcher buttons cannot be added to the Available Launcher buttons box on the Launcher Customization window.
3-48
3. Click E)S-onedt..c.ID"..,_....,.
E) S-onedMedu _.
Figure 3-52. Menu commands being added to the newly-created button.
3. C lick OK to apply your changes and close the Launcher Customization window.
The newly-created Launcher button is displayed on the Launcher Toolbar with a
menu containing the added commands (figure 3-53). Point to the newly-created
Launcher button to display the name given to the Launcher button.
Figure 3-53. The newly-created Launcher Too/bar button.
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
Q54-53287v4
System Overview 0 Returning the Launcher Toolbar to Default Settings
To return the Launcher Toolbar to its default settings:
I. C lick Default Settings (figure 3-54).
'• Launcher Customlzatlon
____--Click
~ .. ~ "
Figure 3-54. Returning the Launcher Too/bar to its default settings.
The Current launcher buttons/Drop-Down menus box wi ll display on ly
the defau lt toolbar options, as shown in figure 3-55. The Current la uncher
buttons/Drop-Down menus list displays only the System Settings and
Help list options. Any created Launcher buttons or menus will be deleted.
•• Launcher Customlzatlon
1 OddS.. I
Figure 3-55. The default too/bar options.
3· 49
0AS Is~ MRI System Reference Manual T'MhtJent MqHt
3-50
2. Click OK to apply your changes and c lose the Launcher Customization
window.
Disk Storage Space Indicators
Loca l Database and Media Usage icons that indicate storage space usage can be
added to the Launcher Toolbar. In the Launcher Customization window, under
the Show space indicator area, click the Sliiow used Loca.l Database space or
the S how used Media space check boxes (figure 3-56) to display the icons.
Select either or both check boxes to display the icons
'• Launcher Customlzatlon
~~~===~-=~ II Text
liJ PocieniD<"''""'
A
p __ _
Plo<oooi/T .. LbiW)I .. ,p
$ Post Ptoeesq lil_,.,. & FlmngTod lli wovo~ ... Eo?, ~IOPW•~h Ul~,-~ ,:a ..... ...,.... 0~ c s .... -
Figure 3-56. The Show space indicator check boxes.
When the Launcher Toolbar is docked to the top or bottom of the monitor
screen, disk space usage is indicated by a horizontal bar displayed below the
device icon (figure 3-57). The horizonta l bar grows longer as disk space is used.
To display used disk space as a percentage(%), move the pointer (mouse cursor)
over the horizontal bar and pause there briefly.
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
QS4-53287v4
System Overview 0
Figure 3-57. Disk space usage indicators.
When the Launcher Toolbar is docked to the left o r right side of the monitor
screen, used disk space is indicated as a percentage(%) displayed below the
device icon (figure 3-58).
Figure 3-58. Disk space usage indicated as a percentage.
3. 51
0AS Is-MRI System Reference Manual YbehtJ ... tlilqHt
Help Feature
G . The Launcher Toolbar contains a Help button.
Note: Only servers and pages within the hitachimed.com domain, including the home page (http:/! www.hitachimed.com), are accessible using the Internet Browser command..
Note: The following operations and fuctions are unavailalbe with the Internet Browser command:
• File downloads, execution and installation
• Java Applet I Active X component execution
• Searches using the search bar
• Right-click context menus
• Opening other windows
• Too/bar changes
• PDF displays
3-52
Clicking the Help button arrow displays a list of help-related options as shown
in figure 3-59.
Sentinel Messenger
Internet Browser
Figure 3-59. Help button options.
You may select from the fo llowing features:
Version Info•·mation - Opens the Version Information window, which
prov ides the version number of your site's Oasis MR I System software, as
shown in figure 3-60.
•:t Version Information
Hitachi M'<I System
OASIS
Ver•"n Number : X3.0A Q;/1$/2011)
Senol Number . 22951
Copyrict>t (C) H~ochi Medicol Corporol"n 2007- 2011 All Richts Re:erved
Figure 3-60. The Version Information window.
Sentinel Messenger -Provides a blank email message that you can complete
and send to a Hitachi field serv ice engineer to report a system problem or ask
a question. However, do not send the email while actively scanning. Certain
maintenance functions can also be carried out by remote access using the
Sentinel feature. When these maintenance tasks are started or completed, a
dialog box is displayed with a message notifying you of the function status.
Internet Browser-Opens the home page of the Hitach i Medical Systems
America, Inc., website.
Copyright© 2012 by Hitachi Medical Systems America, Inc. All rights reserved.
Q54-53287v4
System Overview 0 Launcher Buttons Displayed in Other Locations
When the Oasis MRI System is operating, the following Launcher buttons
may also be displayed in locations other than on the Launcher Tool bar:
-o- J ob Q ueue - This button contains a green circle when active and a red line
through the green c ircle when stopped. It is displayed at the right end of the
Launcher Tool bar. C licking the Job Queue button opens the J ob Q ueue window,
and the selected tab wi II correspond with the job operation currently being
performed.
02:07 Scan Remaining T ime indicator - This progress bar is displayed when scans
are running. It is located at the bottom of the monitor screen when the Launcher
Toolbar is docked to the right or left, or on the right side of the monitor when
the Launcher Toolbar is docked at the top or bottom. C lick the progress bar to
open the Scan Remaining T ime window.
System Date Change Confirmation
A prompt appears whenever an attempt is made to change the system date
(figure 3-61 ).
f Conflrm41t lon for changing date
n o.oe .. ,r be ...,.,lied.
fl.i.h" ron n w Ailor 03o(812C ID
·r~bJmd.-I!J ·~rontJ:tsd -..iyr.lt~:M..nd rr.q iltlJJIUI(J1 dntwttbt- n:orJ~Jtti!IJ
Fttr..c ~ ~ 1"e moor-aacn ag3n. [OC] Aal!(l<..., ..-alfbtm [C~I]C.. toli~ttmt:xMc....t~:r
Figure 3-61. The Confinnation for changing date window.
When a change to the system date is attempted, the Confirmation for changing
date window is displayed, cautioning the user that the study date and image
acquis ition date will be incorrect if the system date is changed. The system date
can still be changed by clicking OK in the Confirmation for cha nging d ate
window, but the user is advised that the system date wil l be inconsistent with the
study and image acquisition dates.
Changes to the system date are initiated by double-clicking the time display in
the system tray (the bottom right corner of the screen). This d isplays the Date
and T ime P roper ties window, where the system date and time can be changed.
3· 53
0AS Is-MRI System Reference Manual YbehtJ ... tlilqHt
This page intentionally left blank.