Upload
777894292
View
217
Download
0
Embed Size (px)
Citation preview
7/27/2019 Chapter 3 Error
1/11
Week 2: chapter three
assessing and presenting experimental
data
Dr. Belal Gharaibeh
6/3/2011
7/27/2019 Chapter 3 Error
2/11
Ask your self, how good are the data?
Determine the quality of the data measured
before using it and making engineering decision
We can compare if the data are good by
comparing to theory derived results. Theory is also derived for a physical system and
it is also like data but we call it a model
(Newtons law)
Measurement should NOT be compared to a
theory to assess its quality
7/27/2019 Chapter 3 Error
3/11
We are trying to measure actual value of physical quantity beingmeasured and that is our Standard
The error is defined as the difference between the measured andtrue physical value of the quantity, we should ask, what is the errorof the data?
True value is something we can never know exactly because wehave to measure it as the first step and process of measurementwill have errors
We can estimate the possible amount of error, example: 95% ofreading from one flowmeter will have an error of less than 1 L/s, sowe say with 95% certainty (19 times of 20) that the meter has anerror of 1 L/s or less.
The reading has accuracy of 1 L/s at odds of 19 to 1. And you canfind a theory within this accuracy
7/27/2019 Chapter 3 Error
4/11
Types of error
Error = = xm-xtrue We want to minimize error in the experiment design
step, but we also need to estimate the boundon ,
This bound is in the form of:
Where u is the uncertainty estimated at odds of n:1only one measurement in n will have an error whose
magnitude is greater than u
7/27/2019 Chapter 3 Error
5/11
Specific cause of error varies from experiment to experiment
Or within the same experiment
Two general classes of error , see figures
Bias error: also called systematic errors are those
happening the same way each time ameasurement is made, a scale reads 5% high, then
every time you measure with it the reading will be
+5% higher than true value
7/27/2019 Chapter 3 Error
6/11
Precision error: also called random error are different foreach successive measurement but have an average valueof zero, for example: errors from mechanical friction orvibration may cause the reading to fluctuate about thetrue value
If enough measurements are taken then the precisionerror will be clear
Readings will cluster about the true value; therefore wecan use statistical analysis to estimate the size of the error.
Bias errors cannot be treated using statistical analysisbecause they are fixed and do not show a distribution
Bias errors are estimated by comparison of the instrumentto a more accurate standard, or by knowledge of how theinstrument was calibrated
7/27/2019 Chapter 3 Error
7/11
Classification of errors
Bias or systematic errors :
Calibration errors (most common) see figure
the most common bias errors is from calibration, calibration is theadjusting the equipment to read the measured values in the right way
zero offset error: causes all readings to be offset by a constant amount (xoffset)
scale errors: change in the slope of the output relative to the input , causes allreadings to change by a fixed percentage , see figure
Certain recurring human errors: when a human reads high values every time Certain errors caused by defective equipment: equipment sometimes have
built-in errors resulting from incorrect design, manufacturing andmaintenance. These errors are constant and can be solved by calibration ifthey dont change with time.
Loading error: the effect of the measurement procedure on the system beingtested. The measuring process changes the characteristics of both source of
the measured quantity and the measuring system. For example: the soundpressure level sensed by a microphone is not the same level if the microphoneis not in the room we want to measure the sound level for
Limitations of system resolution
7/27/2019 Chapter 3 Error
8/11
Classification of errors (continue)
Precision or random error Certain human errors: human is inconsistent in taking the
reading
Errors caused by disturbances to the equipment:
Errors caused by fluctuating experimental conditions:usually coming from outside interference like vibration ortemperature
Errors derived from not measuring the system sensitivity:the sensitivity comes from the design or the
manufacturing process of the instrument, example:instrument designed to measure constant speed will notmeasure any changes in speed you should not use it, or inthe process of making light bulbs not every bulb will be thesame exactly. Such errors are NOT a measurement errorsbut they look like precision errors from measurements and
can be estimated in similar statistical methods
7/27/2019 Chapter 3 Error
9/11
Classification of errors (continue)
Illegitimate errors, errors before or after
making measurements
Mistakes during an experiment: human is not
trained to use the instrument
Calculation errors after an experiment
7/27/2019 Chapter 3 Error
10/11
Classification of errors (continue) Errors that are sometimes bias and sometimes precision errors
Instrument backlash, friction, and hysteresis (path dependence,figure): example is the friction of a scale indicator of an instrument.The reading is low when the measured variable is increasing andreading low when the measured variable is decreasing. figure
Errors from calibration drift and variation in test or environmentalconditions: happens when the response is varying with time and
usually from the sensitivity to temperature and humidity this will bebias error
If the test time is long, errors will fluctuate during that time causingdifferent calibration errors for each time you make long tests. This isprecision error Errors from variations in procedure or definition among experiments:
when the experiment is done with more than one instrument or bydifferent people. In this case each time the test is made it has differentbias which means you have precision error from all the tests
7/27/2019 Chapter 3 Error
11/11
Terms used in rating instrument performance
Accuracy: the difference between the measured and
true values. The manufacturer will specify a maximumerror as the accuracy, what about the odds?
Precision: the difference between the instrumentsdata values during repeated measurements of thesame quantity.
Resolution: the smallest increment or change in themeasured value that can be determined from theinstrument readout scale.
Sensitivity: the change of an instrument output per
unit change in the measured quantity. Reading error: errors when reading a number from the
display scale. The display screen might round a numberor truncate numbers