Upload
phamkhuong
View
215
Download
1
Embed Size (px)
Citation preview
103
CHAPTER-3
A VALIDATED STABILITY-INDICATING ANALYTICAL METHOD FOR THE
DETERMINATION OF IMPURITIES IN
PIOGLITAZONE HYDROCHLORIDE
104
3.1 Introduction on Pioglitazone hydrochloride and survey of
analytical methods
Pioglitazone hydrochloride is a prescription drug of the class
thiazolidinedione (TZD) with hypoglycemic (antihyperglycemic,
antidiabetic) action. Pioglitazone hydrochloride is an oral antidiabetic
agent that acts primarily by decreasing insulin resistance. It is used in
the management of type 2 diabetes mellitus (also known as non-
insulin-dependent diabetes mellitus [NIDDM] or adult-onset diabetes)
[1]. It is chemically designated as (±)-5-[4-(2-(5-Ethyl-2-
pyridinyl)ethoxy)benzyl]-2,4-thiazolidinedione hydrochloride.
Pioglitazone hydrochloride is an odorless white crystalline powder. It is
soluble in N,N-diethylformamide, slightly soluble in anhydrous
ethanol, very slightly soluble in acetone and acetonitrile, practically
insoluble in water, and insoluble in ether. The empirical formula is
C19H20N2O3S.HCl. The molecular weight of Pioglitazone hydrochloride
is 392.9.
Pioglitazone hydrochloride is marketed as trademarks ACTOS in
the USA, glustin in Europe and Zactos in Mexico by the
pharmaceutical company Takeda. Pharmacological studies indicate
that ACTOS improves sensitivity to insulin in muscle and adipose
tissue and inhibits hepatic gluconeogenesis. ACTOS improves
glycemic control while reducing circulating insulin levels.
105
Fig: 3.1 Chemical structure of Pioglitazone hydrochloride
N OS
NH H3C
O
O
. HCl
(±)-5-[4-(2-(5-Ethyl-2-pyridinyl)ethoxy)benzyl]-2,4-thiazolidinedione
hydrochloride
The empirical formula is C19H20N2O3S.HCl
The molecular weight is 392.9.
Several high performance liquid chromatographic (HPLC)
procedures for determination of Pioglitazone hydrochloride in body
fluids and pharmaceutical formulations have been reported in the
literature. HPLC with electrochemical detection has used for analysis
of the drug in human plasma [2-4], serum [5-7] and in pharmaceutical
formulations [8-10].
Organic impurities can arise during the manufacturing process
and storage of the drug substances and the criteria for their
acceptance up to certain limits are based on pharmaceutical studies
or known safety data [11]. Two unknown impurities were detected
consistently in HPLC along with four known impurities i.e. Imp-B, Imp
-C, Imp-D and Imp-E were reported in US Pharmacopeial Forum [12].
However, there was no report of these new impurities i.e., Imp-A and
Imp-F in the literature. As per regulatory guidelines, the
pharmaceutical studies using a sample of the isolated impurities can
106
be considered for safety assessment. It is, therefore, essential to
isolate and characterize unidentified impurities present in the drug
sample. Recently we have developed a process for the synthesis of
Pioglitazone hydrochloride in our laboratory. During the development
of an analytical procedure, the LC method was developed for the
determination of in-house synthesized Pioglitazone hydrochloride and
the impurities arising during its manufacturing. In the present study,
we describe a reverse phase column liquid chromatography method
for the separation and quantification of process related and
degradation impurities of Pioglitazone hydrochloride. The accuracy,
precision, limit of detection (LOD), limit of quantification (LOQ) and
robustness of the method were determined in accordance with ICH
guidelines [13]. The target is to develop a suitable stability-indicating
HPLC related substances method for Pioglitazone hydrochloride in this
chapter we describe a stability-indicating LC method for the
determination of Pioglitazone hydrochloride and its potential and
degradation impurities and also the method validation.
3.2 Development of a stability-indicating analytical method for
Pioglitazone hydrochloride
3.2.1 Materials
Reference standard of Pioglitazone hydrochloride and six
impurities namely, Imp-A, Imp-B, Imp-C, Imp-D, Imp-E and Imp-F
(Fig: 3.2) were synthesized and characterized by use of LC-MS, NMR
and IR in Aurobindo Pharma Ltd., Hyderabad, India. The commercial
107
samples of Pioglitazone hydrochloride are also manufactured by
Aurobindo Pharma Ltd. All reagents used were of analytical reagent
grade unless stated otherwise. Milli Q water, HPLC-grade acetonitrile,
HPLC-grade orthophosphoric acid (OPA) were purchased from Merck
(Darmstadt, Germany).
3.2.2 Equipment
The LC system was equipped with quaternary gradient pumps
with autosampler and auto injector (Alliance, Waters 2695, Milliford,
MA, USA) controlled with Empower software (Waters).
Fig: 3.2 Chemical structures of impurities of Pioglitazone
Hydrochloride
N O
H3C NH2
O
3[(5-Ethyl-2-pyridylethoxy)phenyl]propanamide (Imp-A)
Fig: 3.2 (a)
SNH
HOO
O
5-[4-Hydroxybenzyl]-2,4-thiazolidinedione (Imp-B)
Fig: 3.2 (b)
108
SNH
OO
O
N
H3C
5-[4-[2-(5-Ethyl-2-pyridinyl) ethoxy] benzylidine]-2,4- thiazolidinedione
(Imp-C)
Fig: 3.2 (c)
SN
OO
O
N
H3C
CH2CH3
3(N)-Ethyl-5-[4-[2-(5-ethyl-2-pyridinyl)ethoxy]-benzyl]-2,4-
thiazolidinedione (Imp-D)
Fig: 3.2 (d)
SN
OO
O
N
H3C
CH3
3(N)-Methyl-5-[4-[2-(5-ethyl-2-pyridinyl)ethoxy]-benzyl]-2,4-
thiazolidinedione (Imp-E)
Fig: 3.2 (e)
C
ON
H3C H
S
COOCH3
C
ON
H3C H
COOCH3
S
2,2'-Bis[4-[2-(5-ethylpyridine-2-yl) ethoxy]-phenyl]-2,2'-dithiobisacetic acid, dimethylester (Imp-F)
Fig: 3.2 (f)
109
3.2.3 Sample Preparation
The stock solutions of Pioglitazone hydrochloride (1.0 mg/ml)
and spiked with 0.15% of Imp-A of Imp-B, Imp-D, Imp-E, Imp-F and
0.2% of Imp-C with respect to the Pioglitazone hydrochloride analyte
concentration. The stock solutions were further diluted with diluent to
obtain a standard solution of 0.0015 mg/ml (1.5 µg/ml) for related
substances determination and 0.1mg/ml (100 µg/ml) for assay
determination.
3.2.4 Generation of stress samples
One lot of Pioglitazone hydrochloride drug substance selected for
stress testing. From the ICH stability guideline: “Stress testing likely
to be carried out on a single batch of sample [14]”. Different kinds of
stress conditions (i.e., acid hydrolysis, base hydrolysis, oxidative
stress, heat, humidity and light) were employed on one lot of
Pioglitazone hydrochloride drug substance based on the guidance
available from ICH stability guideline (Q1AR2). The details of the
stress conditions are as follows:
a) Acid hydrolysis: drug in 5.0 M HCl solution was kept at 85°c for
120 mins.
b) Base hydrolysis: drug in 1 M NaOH solution was kept 85°c for
30 mins.
c) Oxidative stress: drug in 30% H2O2 solution was kept at 85°c
for 240 mins.
110
d) Thermal stress: drug was subjected to dry heat at 105°C for 120
hrs.
e) Phtolytic degradation: drug was subjected to UV at 254 nm (10
K Lux ) for 48 hrs.
3.2.5 Optimization of chromatographic conditions
The main objective of the chromatographic method was to
seperate Pioglitazone hydrochloride from Imp-A, Imp-B, Imp-C, Imp-D,
Imp-E and Imp-F impurities were coeluted using different stationary
phases such as C8, phenyl and cyano as well as different mobile
phases containing buffers like phosphate, sulfate and acetate with
different pH and using organic modifiers like acetonitrile and
methanol in the mobile phase. Apart from the co-elution of impurities,
we have also observed poor peak shapes for Pioglitazone, some
impurities and degradants. The chromatographic separation was
achieved on a Inertsil ODS-3V (150 x 4.6 mm), 5 µm particle size. The
gradient LC method employs solution A and B as mobile phase. The
solution A contains phosphate buffer pH 3.1 and acetonitrile as
solution B. The flow rate of the mobile phase was 1.5 ml/min. The
HPLC gradient program was set as: time% solution B: 0.01/25,
20/35, 30/50, 60/50, 62/25, 70/25 with a post run time of 10 min.
The column temperature was maintained at 30°C and the detection
was monitored at a wavelength of 225 nm. The injection volume was
20 µl. Standard and test solutions were prepared in mixture of 0.1%
OPA solution : acetonitrile (50:50, v/v) was used as diluent. In the
111
optimized chromatographic conditions of Pioglitazone hydrochloride,
Imp-A, Imp-B, Imp-C, Imp-D, Imp-E and Imp-F were separated with a
resolution greater than 3, typical relative retention times were
approximately 0.31, 0.58, 1.62, 1.85, 2.61, 5.21 with respect to
Pioglitazone hydrochloride eluted at 7.415 min.
No considerable degradation was observed in Pioglitazone
hydrochloride bulk samples under stress conditions such as acid
hydrolysis, photolytic, thermal conditions (Fig: 3.3, Fig: 3.6 and Fig:
3.7). The degradation of drug substance was observed during base
hydrolysis and oxidative stress condition. Pioglitazone hydrochloride
was degraded to Imp-A (1.5%) under Base conditions (1M
NaOH/85°C/90 min) and it was confirmed by co-injection with a
qualified Imp-A standard and some unknown degradation peaks
observed (12.5%). Mild degradation was observed under oxidative
environment (treated with 30% H2O2/85°C/240 min) leads to the
formation of some unknown degradation peaks (1.2%).
Peak purity test results obtained by using a PDA detector
confirmed that the Pioglitazone hydrochloride peak is homogenous
and pure in all the analyzed stress samples. The mass balance of
Pioglitazone hydrochloride in all stress samples was close to 99.5%
(%Assay + %Degradation). This clearly demonstrates that the
developed HPLC method was found to be specific for Pioglitazone
hydrochloride in presence of its impurities (Imp-A, Imp-B, Imp-C, Imp-
D, Imp-E & Imp-F) and degradation compounds.
112
Optimized liquid chromatographic conditions:
Column : Inertsil ODS-3V (150 x 4.6 mm),
5 µm particle size.
Mobile phase : The solution A contains phosphate
buffer pH 3.1 and acetonitrile as
Solution B
Pump mode : Gradient
Flow rate : 1.5 ml/min
Column oven temperature : 30°C
UV detection : 225 nm
Injection volume : 20 l
Run time : 60 min
Retention time : 7.415 min
Relative Retention Time (RRT) : Imp-A about 0.31
Imp-B about 0.58
Imp-C about 1.62
Imp-D about 1.85
Imp-E about 2.61
Imp-F about 5.21
Diluent : A mixture of 0.1% OPA solution :
acetonitrile (50:50, v/v) was used as
diluent.
113
Figures:
Fig: 3.3 to Fig 3.7 is the typical HPLC chromatograms showing
the degradation of Pioglitazone hydrochloride in various forced
degradation studies and also the corresponding peak purityplots.
Fig: 3.3 Typical HPLC chromatograms of Acid hydrolysis
Fig: 3.3 (a)
Fig: 3.3 (b)
Blank Chromatogram of Acid hydrolysis (5N HCl)
Pioglitazone HCl stressed with 5N HCl at 85°C for 240 mins
114
Fig: 3.3 (c): Peak purity plot of Acid hydrolysis
Purity Angle Purity Threshold Purity Flag Peak Purity
0.018
0.260 No Pass
Fig: 3.3 (c)
115
Fig: 3.4 Typical HPLC chromatograms of Base hydrolysis
Fig: 3.4 (a)
Fig: 3.4 (b)
Blank Chromatogram of Base hydrolysis ( 5N NaOH )
Pioglitazone HCl stressed with 5N NaOH at 85°C for 90 mins
116
Fig: 3.4 (c): Peak purity plot of Base hydrolysis
Purity Angle Purity
Threshold Purity Flag Peak Purity
0.015
0.267 No Pass
Fig: 3.4 (c)
117
Fig: 3.5 Typical HPLC chromatograms of Peroxide Degradation
Fig: 3.5 (a)
Fig: 3.5 (b)
Blank Chromatogram of Peroxide Degradation ( 30% H2O2 )
Pioglitazone HCl stressed with 30%H2O2 at 85°Cfor 240 mins
118
Fig: 3.5 (c): Peak purity plot of Peroxide Degradation
Purity Angle
Purity Threshold Purity Flag Peak Purity
0.023
0.285 No Pass
Fig: 3.5 (c)
119
Fig: 3.6 Typical HPLC chromatograms of Thermal Degradation
Fig: 3.6 (a)
Fig: 3.6 (b)
Pioglitazone HCl stressed at 105°C for 120 hours
120
Fig: 3.6 (c) Peak purity plot of Thermal Degradation
Purity Angle
Purity Threshold Purity Flag Peak Purity
0.016
0.256 No Pass
Fig: 3.6 (c)
121
Fig: 3.7 Typical HPLC chromatograms of Photolytic Degradation
Fig: 3.7 (a)
Fig: 3.7 (b)
Blank
Pioglitazone HCl stressed with 10K Lux for120 hours
122
Fig: 3.7(c) Peak purity of Photolytic Degradation
Purity Angle
Purity Threshold Purity Flag Peak Purity
0.016
0.256 No Pass
Fig: 3.7 (c)
123
3.2.6 Validation of Analytical method and its results:
The developed and optimized HPLC method was taken up to
validation. The analytical method validation was carried out is
accordance with ICH guideline [15].
3.2.6.1 System suitability : A mixture of Pioglitazone hydrochloride
reference standard, Imp-A, Imp-B, Imp-C, Imp-D, Imp-E and Imp-F
were injected into HPLC system and good resolution was obtained
between impurities and Pioglitazone hydrochloride [Fig: 3.8], and the
system suitability results are tabulated (Table: 3.1).
Fig: 3.8 Typical Blank, Pioglitazone HCl Sample and SST
Chromatograms
Fig: 3.8 (a)
Blank
124
Fig: 3.8 (b)
Fig: 3.8 (c)
Pioglitazone HCl sample spiked with impurities
Pioglitazone HCl Sample
125
Table: 3.1 System suitability results
Compound
(n=3)
No. of
theoretical plates (N)
USP Tailing factor
(T) USP Resolution(Rs)
Imp-A 12326 1.07 -
Imp-B 13668 1.11 3.02
Pioglitazone 216870 1.15 6.02
Imp-C 17630 1.39 4.36
Imp-D 26040 1.07 2.86
Imp-E 60019 1.10 4.05
Imp-F 108491 0.97 4.91
3.2.6.2 Precision:
The precision of an analytical process experiment the closeness
of agreement between a series of measurements obtained from
multiple sampling of the some homogeneous same under prescribed
conditions.
It may be considered at three levels: System precision, Method
precision and Intermediate Precision. Assay method precision study
was evaluated by carrying out six independent assays of Pioglitazone
hydrochloride test sample against qualified reference standard and
RSD of six consecutive assays was 0.6% (Table: 3.2).
The results showed insignificant variation in measured
response. Which demonstrated that the assay method was repeatable
with RSD’s below 0.4%.
126
Table: 3.2 System Precision results of the Assay method
Injection ID Pioglitazone Area
1 2425249
2 2421767
3 2427876
4 2433405
5 2444339
6 2445469
Mean 2433018
SD 9967
% RSD 0.4
95% Confidence
Interval ± 10462
Table: 3.3 Method Precision results of the Assay method
Sample ID Assay (% w/w)
1 99.0
2 99.1
3 99.9
4 99.8
5 99.5
6 99.9
Mean 99.5
SD 0.4
% RSD 0.4
95% Confidence
Interval ± 0.4
127
Table: 3.4 Intermediate Precision results of the Assay method
Sample ID Assay (% w/w)
1 99.8
2 99.6
3 99.3
4 99.6
5 99.8
6 99.7
Mean 99.6
SD 0.19
% RSD 0.2
95% Confidence Interval
± 0.2
The precision of the related substance method was checked by
injecting six individual preparations of Pioglitazone hydrochloride
(1.00 mg/ml) spiked with 0.15% of Imp-A, Imp-C, Imp-D, Imp-E, Imp-
F and 0.2% of Imp-B with respect to the Pioglitazone hydrochloride
analyte concentration. The % RSD of the area percentage of each
impurity (impurities-A, -B, -C, -D, -E and -F) for six consecutive
determinations was respectively as below (Table: 3.5 to Table: 3.7).
The results showed insignificant variation in measured
response. Which demonstrated that the related substance method was
repeatable with RSD’s below 1.3%.
128
Table: 3.5 System Precision results of the Related substances
method
Injection ID Pioglitazone Area
1 44290
2 44578
3 44307
4 44430
5 44579
6 44352
Mean 44290
SD 44423
% RSD 130
95% Confidence
Interval 0.3
Table: 3.6 Method Precision results of the Related Substance
method
Preparation Imp-A Imp-B Imp-C Imp-D Imp-E Imp-F
1 0.150 0.174 0.248 0.164 0.150 0.240
2 0.147 0.175 0.247 0.160 0.152 0.238
3 0.152 0.175 0.248 0.163 0.152 0.236
4 0.151 0.174 0.247 0.164 0.150 0.241
5 0.152 0.172 0.247 0.166 0.151 0.241
6 0.150 0.174 0.245 0.164 0.152 0.240
Mean 0.150 0.174 0.247 0.164 0.151 0.239
SD 0.0019 0.001 0.001 0.002 0.001 0.002
%RSD 1.24 0.6 0.4 1.2 0.7 0.8
95% Confidence
interval ±0.001 ±0.001 ±0.001 ±0.002 ±0.001 ±0.002
129
Table: 3.7 Intermediate Precision results of the Related substance
method
Preparation Imp-A Imp-B Imp-C Imp-D Imp-E Imp-F
1 0.152 0.174 0.260 0.163 0.146 0.237
2 0.145 0.175 0.258 0.162 0.148 0.237
3 0.151 0.174 0.258 0.161 0.150 0.238
4 0.152 0.175 0.260 0.162 0.149 0.237
5 0.150 0.176 0.258 0.162 0.151 0.238
6 0.152 0.176 0.258 0.164 0.149 0.239
Mean 0.150 0.175 0.254 0.162 0.149 0.238
SD 0.0019 0.001 0.001 0.001 0.002 0.002
%RSD 1.24 0.6 0.4 0.6 1.3 0.4
95%
Confidence interval
±0.001 ±0.001 ±0.001 ±0.002 ±0.001 ±0.002
3.2.6.3 Limit of Detection (LOD):
The detection limit of an individual analytical procedure is the
lowest amount of analyte is a sample, which can be detected but not
necessarily quantitated as an exact value (Table: 3.8).
130
Table: 3.8 LOD values of the Pioglitazone and its impurities
Injection ID
Area
Imp-A Imp-B Pioglitazone Imp-C Imp-D Imp-E Imp-F
1 1691 1772 1281 1796 1745 1808 2182
2 1587 2127 1722 1403 1661 1393 2677
3 2013 1696 1611 1813 1656 1865 2050
4 2089 1702 1614 1794 1285 1341 2134
5 2310 1471 1619 1480 1736 1416 2185
6 1646 1638 1461 2048 1434 1646 2024
Mean 1889 1734 156 1722 1586 1578 2209
SD 290 218 10.1 239 185 227 239
% RSD 15.4 12.6 0.056 13.9 11.7 14.4 10.8
Conc. (µg/mL)
0.058 0.049 0.056 0.083 0.060 0.066 0.075
Conc. (% w/w)
0.005 0.005 0.006 0.008 0.006 0.007 0.008
131
3.2.6.4 Limit of Quantification (LOQ)
The quantitation limit (LOQ) of an analytical procedure is the
lowest amount of analyte in a sample, which can be quantitatively
determined with suitable precision and accuracy. The quantitative
limit is a parameter of quantitative assays for low levels of compounds
in sample matrices, and is used particularly for the determination of
impurities and/or degradation products (Table: 3.9).
Table: 3.9 LOQ values of the Pioglitazone and its impurities
Injection ID Area
Imp-A Imp-B Pioglitazone Imp-C Imp-D Imp-E Imp-F
1 6956 5381 5104 5680 5268 5651 6747
2 6148 5308 4984 5594 5286 5009 6764
3 7080 5642 4993 5512 5144 5185 6665
4 7045 5423 4917 5644 5287 5052 6700
5 6630 5659 5050 5752 5475 5193 6872
6 6933 5575 5010 5509 5294 5288 6962
Mean 6799 5498 5010 5615 5292 5230 6785
SD 356 147 63 96 106 230 112
% RSD 5.2 2.7 1.3 1.7 2.0 4.4 1.7
Conc. (µg/mL)
0.175 0.150 0.170 0.251 0.181 0.200 0.226
Conc. (% w/w)
0.020 0.015 0.017 0.025 0.018 0.020 0.023
132
3.2.6.5 Linearity
Linearity of the Assay method
The linearity of an analytical procedure is its ability to obtain
test results, which are directly proportional to the concentration of
analyte in the test sample. The linearity of the assay method was
established by injecting test sample at 80%, 90%, 100%, 110% and
120% of Pioglitazone hydrochloride assay concentration (i.e.100
µg/ml). Each solution injected twice (n=2) into HPLC and the average
area at each concentration calculated (Table: 3.10).
Calibration curve drawn by plotting average area on the Y-axis and
concentration on the X-axis (Fig: 3.9).
133
Table: 3.10 Linearity results of the Assay method
%
Concentration Average area
80 1946315
90 2180238
100 2440221
110 2689004
120 2910526
Slope 24372
Intercept -3927
% Y - Intercept -0.2
Residual Sum of
Squares 11304
Correlation Coefficient 0.9996
Linearity Plot (Concentration Vs Response)
Fig: 3.9 Linearity Plot for Assay method
80.00 90.00 100.00 110.00 120.00
Aver
age
Are
a
Conc.(µg/mL)
134
Linearity of the Related Substance method
Linearity experiments were carried out by preparing the
Pioglitazone hydrochloride sample solutions containing Imp-A, Imp-B,
Imp-C, Imp-D, Imp-E, Imp-F and Imp-G from LOQ to 150% (i.e. LOQ,
10%, 15%, 20%, 25%, 50%, 75%, 100%, 125%, 150%) with respect to
their specification limit (0.15%). Calibration curve was drawn by
plotting average area of the impurity Imp-A, Imp-B, Imp-C, Imp-D,
Imp-E, Imp-F and Imp-G) on the Y-axis and concentration on the X-
axis (Fig: 3.10 to Fig: 3.16).
135
Linearity results of the related substance method
Table: 3.11 Linearity results of the Imp-A
Imp-A
Concentration (µg/ml)
Area Statistical Analysis
0.176 5970 Slope 32494
0.249 8194 Intercept 277
0.498 16627 Residual Sum of
Squares 142
0.748 24616
0.997 32745 Correlation Coefficient
0.9999 1.246 40856
1.495 42164
1.870 53010 Response factor 1.08
2.244 63993
Linearity Plot (Concentration Vs Area)
Fig: 3.10 Linearity Plot for Imp-A
5555
17555
29555
41555
53555
65555
77555
0.151 0.501 0.851 1.201 1.551 1.901 2.251
Are
a
Con. (µg/mL)
136
Table: 3.12 Linearity results of the Imp-B
Imp-B
Concentration
(µg/ml) Area Statistical Analysis
0.151 5555 Slope 35638
0.226 8241 Intercept 321
0.302 10837 Residual Sum of
Squares 264
0.377 14102
0.755 27359 Correlation
Coefficient 0.9999 1.132 40635
1.510 54568
1.887 67373 Response factor 0.86
2.265 80884
Linearity Plot (Concentration Vs Area)
Fig: 3.11 Linearity Plot for Imp-B
5555
17555
29555
41555
53555
65555
77555
0.151 0.501 0.851 1.201 1.551 1.901 2.251
Are
a
Con. (µg/mL)
137
Table: 3.13 Linearity results of the Pioglitazone hydrochloride
Pioglitazone
Concentration
(µg/ml) Area Statistical Analysis
0.170 5010 Slope 30481
0.200 5867 Intercept -315
0.300 8951
0.400 11794
Residual Sum of
Squares 416 0.500 14542
1.000 30883
1.501 44899
Correlation
Coefficient 0.9999
2.001 60361
2.501 75688
3.001 91639
Linearity Plot (Concentration Vs Area)
Fig: 3.12 Linearity Plot for Pioglitazone hydrochloride
5010
18510
32010
45510
59010
72510
86010
0.170 0.620 1.070 1.520 1.970 2.420 2.870
Are
a
Con. (µg/mL)
138
Table: 3.14 Linearity results of the Imp-C
Imp-C
Concentration
(µg/ml) Area Statistical Analysis
0.251 5615 Slope 21953
0.299 6575 Intercept -165
0.399 8484 Residual Sum of
Squares 180
0.499 10558
0.998 21571
Correlation
Coefficient 0.9999 1.497 32609
1.996 43783
2.495 54591 Response factor 1.39
2.994 65602
Linearity Plot (Concentration Vs Area)
Fig: 3.13 Linearity Plot for Imp-C
5615
15615
25615
35615
45615
55615
0.251 0.701 1.151 1.601 2.051 2.501 2.951
Are
a
Con. (µg/mL)
139
Table: 3.15 Linearity results of the Imp-D
Imp-D
Concentration
(µg/ml) Area Statistical Analysis
0.181 5292 Slope 28436
0.224 6371 Intercept -77
0.299 8551
Residual Sum of
Squares 229 0.374 10302
0.748 20990
1.122 31987
Correlation Coefficient
0.9999 1.496 42164
1.870 53010
2.244 63993 Response factor 1.07
Linearity Plot (Concentration Vs Area)
Fig: 3.14 Linearity Plot for Imp-D
5292
14292
23292
32292
41292
50292
59292
0.181 0.501 0.821 1.141 1.461 1.781 2.101
Are
a
Con. (µg/mL)
140
Table: 3.16 Linearity results of the Imp-E
Imp-E
Concentration
(µg/ml) Area Statistical Analysis
0.200 5230 Slope 26436
0.226 5941 Intercept -196
0.302 7858 Residual Sum of
Squares 229
0.377 9764
0.754 19644
Correlation Coefficient
0.9999 1.131 29266
1.508 39470
1.885 49732 Response factor 1.15
2.262 59865
Linearity Plot (Concentration Vs Area)
Fig: 3.15 Linearity Plot for Imp-E
5230
14230
23230
32230
41230
50230
59230
0.200 0.550 0.900 1.250 1.600 1.950
Are
a
Con. (µg/mL)
141
Table: 3.17 Linearity results of the Imp-F
Imp-F
Concentration (µg/ml)
Area Statistical Analysis
0.225 6695 Slope 30225
0.300 8917 Intercept -260
0.375 11163 Residual Sum of Squares
346 0.749 22471
1.124 33313 Correlation
Coefficient 0.9999
1.498 44439
1.873 56499 Response factor 1.01
2.248 68074
Linearity Plot (Concentration Vs Area)
Fig: 3.16 Linearity Plot for Imp-F
6695
16695
26695
36695
46695
56695
66695
0.225 0.575 0.925 1.275 1.625 1.975
Are
a
Con. (µg/mL)
142
3.2.6.6 Accuracy/Recovery
The accuracy of an analytical procedure expresses the closeness
of agreement between the value, which is accepted either as a
conventional true value or an accepted reference value and the value
found.
Accuracy of the assay method
Accuracy of the assay method was developed by injecting three
preparations of test sample at 80%, 90%, 100% 110%, 120% and
150% of analyte concentration (i.e.100 µg/ml). Each solution was
injected twice (n=2) into HPLC and the mean peak area of Pioglitazone
hydrochloride peak was calculated. Assay (%w/w) of test solution was
determined against three injections (n=3) of qualified Pioglitazone
hydrochloride reference standard (Table: 3.18).
The method showed consistent and high absolute recoveries at
all three concentration (80%, 90%, 100% 110%, 120% and 150%)
levels with mean absolute recovery ranging from 99.3 % to 100.5%.
The obtained obsolute recoveries were normally distributed around
the mean with uniform RSD values. The method was found to be
accurate with low % bias (<1.0).
143
Table: 3.18 Acuracy results of the Assay method
S.NO % Concentration Mean recovery (%)
(n=3) %RSD
1 80 99.3 0.06
2 90 99.5 0.15
3 100 99.8 0.10
4 110 100.0 0.15
5 120 100.6 0.10
Accuracy of the related substances method established at 50% 100%
and 150% of the impurities specification limit (0.15%).
Accuracy at 50% impurity specification level:
Test solutions prepared in triplicate (n=3) with impurities (Imp-
A, B, C, D, E and F) at 0.075% (Imp-A, B, D, E, F) and 0.1% (Imp-C)
level w.r.s analyte concentration (i.e 1.00 mg/ml). Each solution was
injected thrice into HPLC (Table: 3.19).
144
Table: 3.19 Accuracy at 50% level
S.NO Impurity
name
Mean
recovery(%) SD %RSD
1 Imp-A 102.0 0.00 0.0
2 Imp-B 98.2 0.75 0.8
3 Imp-C 100.0 1.00 1.0
4 Imp-D 99.1 0.75 0.8
5 Imp-E 97.7 1.59 1.6
6 Imp-F 100.9 2.03 2.0
Accuracy at 100% impurity specification level:
Test solution prepared in triplicate (n=3) with impurities (Imp-A,
B, C, D, E , F ) at 0.15% ( Imp-A, B, D, E, F) and 0.2% (Imp-C) level
w.r.s analyte concentration (i.e 1.00 mg/ml). Each solution was
injected thrice into HPLC (Table: 3.20).
145
Table: 3.20 Accuracy at 100% level
S.NO Impurity
name
Mean
recovery(%) SD %RSD
1 Imp-A 96.6 0.65 0.7
2 Imp-B 96.6 0.65 0.7
3 Imp-C 98.8 0.29 0.3
4 Imp-D 96.1 0.65 0.7
5 Imp-E 97.6 1.40 1.4
6 Imp-F 101.7 0.75 0.7
Accuracy at 150% impurity specification level:
Test solution prepared in triplicate (n=3) with impurities (Imp-A,
B, C, D, E , F and G) at 0.12% ( Imp-A, B, D, E, F) and 0.3% (Imp-C)
level w.r.s analyte concentration (i.e 1.00 mg/m l). Each solution was
injected thrice into HPLC (Table: 3.21).
146
Table: 3.21 Accuracy at 150% level
S.NO Impurity
name
Mean
recovery(%) SD %RSD
1 Imp-A 99.7 1.40 1.4
2 Imp-B 96.6 0.67 0.7
3 Imp-C 98.9 1.33 1.3
4 Imp-D 97.8 0.45 0.5
5 Imp-E 97.3 1.17 1.2
6 Imp-F 102.1 0.23 0.2
The related substance method showed consistent and high
accurate recoveries of all six impurities at all three different
concentrations (50%, 100%, 150%) levels in drug substance.
3.2.6.7 Solution state stability
The solution state stability of Pioglitazone Hydrochloride in
diluent in the assay method was carried out by leaving both the test
solutions of sample and reference standard in tightly capped
volumetric flasks at room temperature for two days. The same sample
solutions were assayed for every one hour interval up to the study
period. The % RSD of assay of Pioglitazone during solution stability
experiments was with in 1.0%.
The solution state stability of Pioglitazone hydrochloride related
substance method was carried out by leaving sample solution in
tightly capped volumetric flask at room temperature for two days.
Content of Imp A, B, C, D, E, and F were checked for every six hours
internal up to the study period. No significant change was observed in
147
the content of all six impurities drug solution stability experiments up
to the study period. Hence Poiglitazone hydrochloride sample
solutions are stable for atleast 48 hours in the developed method.
In assay method the standard and sample solution injected at
each 0h, 1h, 2h, 3h, 4h, 5h, 6h, 7h, 8h, 9h, 10h, 11h, 12hr (Table:
3.22).
Table: 3.22 Solution stability results of the assay method
S.No Time in Hours Assay (% w/w)
1 initial 99.8
2 1 99.7
3 2 99.2
4 3 99.4
5 4 99.5
6 5 99.6
7 6 99.7
8 7 99.4
9 8 99.3
10 9 99.1
11 10 99.4
12 11 99.6
13 12 99.5
% RSD 0.21
In related substances method the stability of Pioglitazone
hydrochloride in diluent was established for 15 hr by injecting test
solution for every one hour interval up to the study period. The
impurity profiles obtained at different interval were very consistent
and matched with initial value.
148
3.2.6.8 Robustness
The robustness of an analytical procedure is a measure of its
capacity to remain unaffected by small, but deliberate variations in
method parameters and provides an indication of its reliability during
normal usage. To determine the robustness of the developed method
experimental conditions were purposely altered and the resolution
between Imp-C and Imp-D was evaluated. In each of the deliberately
altered chromatographic condition (flow rate 1.3 ml/min and
1.7ml/min, acetonitrile 23% and 27% in the mobile phase, column
temperature 25C and 35C) the resolution between Imp-B, Imp-C and
Imp-D, Imp-E and Imp-F was greater than 2.0, illustrating the
robustness of the method.
3.2.6.9 Mass balance of the analytical method
The mass balance is a process of adding together the assay
value and the levels of degradation products to see how closely these
add up to 100% of the initial value, with due consideration of the
margin of analytical error. Its establishment hence is a regulatory
requirement. The mass balance is very closely linked to the
development of stability-indicating assay method as it acts as an
approach to establish its validity. The stressed samples of Pioglitazone
hydrochloride bulk drug were assayed against the qualified reference
standard and the results of mass balance obtained were very close to
99.8%. The results of mass balance obtained in each condition is
presented below [Table 3.23).
149
Table: 3.23 Mass balance of the Assay method
Degradation
Mechanism
Degradation
Condition
% Assay of
active substance
Mass balance
(% Assay+ %
impurities+ %
degradants)
Remarks
Acid 5M HCl / 85°C
/ 240 min 99.5 99.8
No
degradation
observed
Base 1M NaOH /
85°C /90min 85.9 99.9
Degraded to Imp-A and
some
unknown degradants
observed
Peroxide
30% H2O2 /
85°C / 240
min
98.6 99.8
some
unknown degradants
observed
Thermal 105°C / 120
Hours 99.5 99.7
No
degradation observed
Photolytic 10K Lux / 120
Hours 99.2 99.5
No
degradation
observed
3.3 Analysis of Pioglitazone hydrochloride drug substance
stability samples
One manufacturing lot of Pioglitazone hydrochloride drug
substance was placed on stability study in chambers maintained at
ICH set conditions. The analysis of stability samples were carried up
to 24 months period using the above optimized method. The stability
data results obtained are presented in Table: 3.24 & Table: 3.25. The
developed HPLC method performed satisfactorily for the quantitative
evaluation of stability samples.
150
Table: 3.24 Accelerated stability data ( Storage conditions:
40°C/75%RH)
Batch No: MHK(684)90 Packing & storage conditions: Each sample packed in a polyethylene bag in a triple
laminated bag and kept in a HDPE drum
Stability study duration: 6 months Temperature
%Relative humidity 40°C/75%RH
Tests Description Loss on drying
(%w/w)
Identification
Assay
(By HPLC,
%w/w, on dried basis,)
Specifications
A white to almost
white
powder
NMT 1.0
IR spectrum
should
concordant with that of
standard
NLT 98.5 and
NMT 101.5
Initial A white
powder 0.13 Complies 99.8
1M A white powder
0.12 Complies 100.4
2M A white
powder 0.14 Complies 100.3
3M A white
powder 0.16 Complies 100.0
6M A white powder
0.19 Complies 100.1
Related substances details on next page.
151
Related
Substances
LOQ
(%w/w)
LOD
(%w/w)
Related Substances (By HPLC, %w/w)
Initial 1M 2M 3M 6M
Imp-A 0.015 0.005 Below
LOQ ND ND ND ND
Imp-B 0.014 0.005 Below
LOQ
Below
LOQ
Below
LOQ
Below
LOQ
Below
LOQ
Imp-C 0.025 0.008 0.06 0.04 0.04 0.04 0.04
Imp-D 0.018 0.006 Below LOQ
Below LOQ
Below LOQ
Below LOQ
Below LOQ
Imp-E 0.020 0.007 Below
LOQ
Below
LOQ
Below
LOQ
Below
LOQ
Below
LOQ
Imp-F 0.023 0.008 0.03 0.03 0.03 0.02 0.02
Highest
unknown - - ND ND ND ND ND
Total unknown
- - NA NA NA NA NA
Total RS - - 0.09 0.07 0.07 0.06 0.06
ND: Not detected NA: Not applicable
152
Table: 3.25 Long term stability data ( Storage conditions
25°C/60%RH)
Batch No: MHK(684)90 Packing & storage conditions: Each sample packed in a polyethylene bag in a
triple laminated bag and kept in a HDPE drum
Stability study duration: 24 months Temperature %Relative humidity
25°C/60%RH
Tests Description
Loss on
drying (%w/w)
Identification
Assay (By HPLC,
%w/w, on
dried basis,)
Specifications
A white to
almost white
powder
NMT 1.0
IR spectrum should
concordant
with that of standard
NLT 8.5 and NMT 101.5
Initial A white
powder 0.13 Complies 99.8
3M A white powder
0.12 Complies 100.0
6M A white
powder 0.19 Complies 100.1
9M
A white
powder 0.21 Complies 100.2
12M
A white powder
0.21 Complies 99.8
24M A white
powder 0.20 Complies 99.8
Related substances details on next page.
153
Related Substances
LOQ (%w/w)
LOD
(%w/w)
Related Substances (By HPLC, %w/w)
Initial 3M 6M 9M 12M 24M
Imp-A 0.015 0.005 Below LOQ
ND ND ND ND ND
Imp-B 0.014 0.005 Below
LOQ
Below
LOQ
Below
LOQ
Below
LOQ
Below
LOQ
Below
LOQ
Imp-C 0.025 0.008 0.06 0.04 0.04 0.05 0.05 0.05
Imp-D 0.018 0.006 Below
LOQ
Below
LOQ
Below
LOQ
Below
LOQ
Below
LOQ
Below
LOQ
Imp-E 0.020 0.007 Below LOQ
Below LOQ
Below LOQ
Below LOQ
Below LOQ
Below LOQ
Imp-F 0.023 0.008 0.03 0.03 Below
LOQ
Below
LOQ
Below
LOQ
Below
LOQ
Highest
unknown - - ND ND ND ND ND ND
Total unknown
- - NA NA NA NA NA NA
Total RS - - 0.09 0.07 0.06 0.05 0.05 0.05
ND: Not detected NA: Not applicable
154
3.4 Summary and conclusions
Validated stability-indicating HPLC method was developed for
Pioglitazone hydrochloride after subjecting the samples to stress
testing under ICH recommendes conditions. The RPLC method
developed for quantitative and related substance determination of
Pioglitazone hydrochloride is rapid precise, accurate linear and
selective. The method was completely validated showing satisfactory
data for all the method validation parameters tested. The developed
method was found ‘specific’ to the drug, as the peaks of the
degradation products did not interfere with the degradation peak.
Thus the proposed method can be employed for assessing the stability
of Pioglitazone hydrochloride bulk drug samples.
155
Table: 3.26 Summary of Analytical method validation data
Test
Parameter
Related Substances method Assay
method
Imp-A Imp-B Imp-C Imp-D Imp-E Imp-F
Precision
(RSD) 0.6 0.4 1.2 0.7 0.8 0.6 0.4
LOD(µg/ml) 0.049 0.033 0.034 0.033 0.033 0.034 N/A
LOQ(µg/ml) 0.150 0.101 0.103 0.100 0.100 0.102
N/A
Linearity
(corre
coefficient
0.9999 0.9999 0.9999 0.9999 0.9999 0.9999
0.9999
Accuracy (%) 96.6-98.2 96.6-98.2 98.8-100.0 96.1-99.1 97.3-97.7 100.9-102.1 99.3-100.5
Robustness Resolution
b/w Imp-A&
Imp-B>2
Resolution b/w Imp-A&
Imp-B>2
Resolution b/w Imp-A&
Imp-B>2
Resolution b/w Imp-A&
Imp-B>2
Resolution b/w Imp-A&
Imp-B>2
Resolution b/w Imp-A&
Imp-B>2
Resolution b/w Imp-A&
Imp-B>2
Solution
stability
Stable up to
15hr
Stable up to
15hr
Stable up to
15hr
Stable up to
15hr
Stable up to
15hr
Stable up to
15hr
Stable up to
15hr
Mobile phase stability
Stable up to 15hr
Stable up to 15hr
Stable up to 15hr
Stable up to 15hr
Stable up to 15hr
Stable up to 15hr
Stable up to 15hr
156
3.5 References:
1. Sean. C. S.; Martindale-The complete drug reference,
35th edition, 2007, Vol.2, p.414.
2. Lin, Z. J.; Ji, W.; Desai-Karieger, D.; Shum, L.; J. Pharm.
Biomed. Anal., 2003, 33, 101.
3. Souri, E.; Jalalizadeh, H.; Saremi, S.; J. Chromatogr. Sci. 2008,
46, 80.
4. Sripalakit, P.; Neamhom, P.; Saraphanchotiwittaya, A.;
J.Chromatogr. B, 2006, 843, 164.
5. Zhong, W. Z.; Williams, M. G.; J. Pharm. Biomed. Anal., 1996,
14, 465.
6. Xue, Y. J.; Turner, C. K.; Meeker, B. J.; Pursley, J.; Arnold, M.;
Unger, S.; J. Chromatogr.B, 2003, 795, 215.
7. Yamashita, K.; Murakani, H.; Okuda, T.; Motohashi, M.;
J. Chromatogr B, 1996, 677, 141.
8. Mehta, R. S.; Patel, D. M.; Bhatt, K. K.; Shankar, M. B.;
Ind. J. Pharm Sci., 2005, 67, 487.
9. Jedlicka, A.; Kimes, J.; Grafnetterova, T.; Pharmazie. 2004, 59,
178.
10. Radhakrishna, T.; Sreenivas Rao, D.; Om Reddy, G.; J. Pharm.
Biomed. Anal., 2002, 29, 593.
11. International Conferences on Hormonization Q3A (R2): Draft
Revised Guidance on impurities in New Drug Substances,
2006.
12. US Pharmacopeial Forum, 2010, 36, 126.
157
13. ICH, Validation of analytical Procedures: Text and methodology
Q2 (R1), International Conferences on Hormonization, IFPMA,
Geneva, 2005.
14. Steven, W. Baertschi Pharmaceutical Stress Testing Predicting
Drug Degradation.
15. Validation of Analytical Procedures: Methodology Q2B – ICH
Guidelines.