Chapter 1&2 Introduction & Centrifugation

Embed Size (px)

Citation preview

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    1/39

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    2/39

    ASSESSMENTS :

    5 QUIZZES 10%

    2 ASSIGNMENTS 20%

    MID SEM EXAM 30%

    FINAL EXAM 40%

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    3/39

    CHAPTER 1 INTRODUCTION

    CHAPTER 2 CENTRIFUGATION

    CHAPTER 3 CHROMATOGRAPHY

    CHAPTER 4 ELECTROCHEMISTRY

    CHAPTER 5 MICROSCOPE

    CHAPTER 6 SPECTROMETRYCHAPTER 7 ELECTROPHORESIS

    CHAPTER 8 HEMATOLOGY ANALYSIS

    CHAPTER 9 IMMUNOCHEMISTRY ANALYSIS

    CHAPTER 10 FORENSIC ANALYSIS

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    4/39

    CHAPTER 1 : INTRODUCTION

    BIOMEDICAL INSTRUMENTATION

    Devices that can be used to make measurements of biologic or

    medical quantities and give quantitative results

    BASIC BIOMEDICAL INSTRUMENT

    PHYSIOLOGIC SENSOR PROCESSOR DISPLAY OBSERVER

    SYSTEM STORAGE

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    5/39

    FUNDAMENTAL RULES OF BIOMEDICALINSTRUMENTATION

    Minimum disturbance to physiologic

    Sensor must be at physiologic variable value

    Maintain simplicity

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    6/39

    IMPORTANT INSTRUMENTATION TERMS

    SENSITIVITY - Change in output as a function of achange in input

    STABILITY

    Consistency in output for a constantinput

    SPECIFICITY Ability to distinguish desired variable

    from other competing variables

    ACCURACY the closeness of the instrumentoutput to the true value of the measured quantity

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    7/39

    PRECISION The smallest change in a variables thatcan be correctly measured

    RESOLUTION The degree of fineness with which ameasurement can be made

    REPRODUCIBILITY Same output for the same input

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    8/39

    TRANSFER CHARACTERISTIC

    SENSITIVITY- change in output as a function of a changein input

    Instrument

    Output

    Variable Measured

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    9/39

    STABILITY Consistency in output for a constantinput

    Variable

    Time

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    10/39

    SPECIFICITY Ability to distinguish desired variablefrom other competing variables

    Instrument

    Output

    Variable Measured

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    11/39

    ACCURACY the closeness of the instrument output tothe true value of the measured quantity

    Instrument

    Output

    Variable Measured

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    12/39

    ACCURACY

    The accuracy of the instruments can be measured by usingthis calculation:

    Percentage of true value = measured value true value x 100true value

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    13/39

    ACCURACY vs PRECISION

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    14/39

    POSSIBLE TYPES OF TRANSDUCERS

    To convert the input signal to a form which can beconveniently processed

    Mechanical Thermal

    Electrical Acoustic

    Mechanical Hydraulic

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    15/39

    TYPES OF SENSOR

    PHYSICAL-THERMOMETER

    BIOANALYTICAL

    -GLUCOSE SENSOR

    CHEMICAL-OXYGEN

    ELECTRODE

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    16/39

    SINGLE CONVERSION SENSOR

    PHYSIOLOGIC SENSOR ELECTRICAL

    VARIABLE SIGNAL

    PARTIAL CLARK ELECTRICAL

    PRESSURE ELECTRODE CURRENT

    OF OXYGEN

    BODY THERMISTOR ELECTRICAL TEMPERATURERESISTANCE

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    17/39

    MULTIPLE CONVERSION SENSOR

    PHYSIOLOGIC INTERMEDIATE SENSOR ELECTRICAL

    SENSOR VARIABLE SIGNAL

    BLOOD DIAPHRAGM DISPLACEMENT ELECTRICAL

    PRESSURE DISPLACEMENT SENSOR SIGNAL

    SERUM HYDROGEN H2O2 SENSOR ELECTRICAL

    GLUCOSE PEROXIDE SIGNAL

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    18/39

    EXAMPLES OF FAMILIAR BIOMEDICALINSTRUMENTATION

    http://rds.yahoo.com/_ylt=A0S020oJhERK.8UAFwWjzbkF/SIG=127iohu06/EXP=1246090633/**http:/www.valuemed.co.uk/acatalog/on-call_plus.gifhttp://rds.yahoo.com/_ylt=A0S020ykgkRKlX4AtyyjzbkF/SIG=127h98g72/EXP=1246090276/**http:/www.flickr.com/photos/tessawatson/383858436/http://rds.yahoo.com/_ylt=A0S020ykgkRKlX4ApiyjzbkF/SIG=12qa8o9sv/EXP=1246090276/**http:/img.alibaba.com/photo/50071008/Digital_Clinical_Thermometer.jpghttp://rds.yahoo.com/_ylt=A0S020xCgkRK.4YAdaCjzbkF/SIG=126a5o8s0/EXP=1246090178/**http:/www.flickr.com/photos/jessmercer/358719372/
  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    19/39

    STANDARDS OF MEASUREMENTS

    -Defined as the physical representation of the unit ofmeasurement (S.I. unit)

    -For example : kilogram was originally defined as the

    mass of a cubic decimetre of water at its temperature ofmaximum density at 4 C

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    20/39

    CALIBRATION

    -Is the act or result of quantitative comparison between

    a known standard and the output of the measuring

    system measuring the same quantity

    -Calibration curve is a references standards withknown values to cover the range of interest

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    21/39

    ERROR

    -Defined as the differences between the measured and thetrue value

    ERROR

    systematic/cumulative Random error

    error - arises from our

    - can in principle be ability to makediscovered and corrected physical measurement

    - always present and

    cannot be corrected

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    22/39

    Systematic/cumulative Error

    Instrument Environmental Zero error

    Error Error

    Loading

    Error

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    23/39

    Instrument error

    -Poor design of instrument-Can be avoided by selecting a suitable instrument

    for a given application and calibrating instrument

    using a suitable standard

    Environmental error

    -The condition of environment such as windy, rainy,hot .

    -Can be avoided by doing the measurement in close

    room.

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    24/39

    CHAPTER 2 : CENTRIFUGATION

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    25/39

    CENTRIFUGATION

    -Device for separating particles from a solutionaccording to their size, shape, density,viscosity ofthe medium and rotor speed-Centrifuge have been used since middle 1800-Hand-driven (only after 1912, it is electricallydriven)

    THEORY OF CENTRIFUGATION

    -Sedimentation of a molecule is influenced by :a) properties of the molecule (size,shape,density)b) properties of the solvent (density, temperature)c) interactions between solute molecules & solvent

    material

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    26/39

    FORCES IN CENTRIFUGATION

    As the rotor spins in a centrifuge, centrifugal force isapplied to each molecule in the sample :

    CF = Mw2r

    M = mass (particle weight/molecular weight)

    w = angular velocity

    r = distance from the axis of rotation

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    27/39

    -The larger the molecule or the faster thecentrifugation or the longer the axis of rotation,

    the greater the centrifugal force and the rate ofsedimentation

    -Two forces act to counteract the centrifugal force are

    bouyant force and frictional force

    -A sedimenting molecule moves faster and faster in a

    centrifugal field until :

    Centrifugal force = bouyant force + frictional force

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    28/39

    TYPES OF CENTRIFUGES

    Desktop clinical centrifuges

    -Simplest & least expensive

    -Max speed of most desktop centrifuges is below 3000 rpm

    -Most operate at all temperature-Typical application : rapid sedimentation of blood samples

    Highspeed centrifuges

    -Operating up to speeds of 20,000 to

    25,000 rpm

    -Typical application : collect larger organelles

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    29/39

    Ultracentrifuges

    -Typical application : separation of the major

    lipoprotein fractions from plasma

    -4 principles :

    a)Drive and speed control : flexible drive shaft system,an overspeed system

    b)Temperature control : infrared temp. sensor

    c)Vacuum system

    d)Rotor : swinging bucket rotor & fixed

    angle rotor

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    30/39

    TYPES OF CENTRIFUGATION

    -Analytical-Preparative

    ANALYTICAL

    -measuring physical properties of sedimenting particlesuch as molecular weight

    PREPARATIVE-To isolate specific particles which can be reused

    -Three types : rate zonal, differential, isopycnic

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    31/39

    Analytical

    Uses small sample (less than 1 ml)

    Built in optical system to analyze progress of molecules during

    centrifugation Uses relatively pure sample

    Used to precisely determine sedimentation coefficient and MWof molecules

    Preparative

    Larger sample size can be used

    No optical read out collect fractions and analyze them after therun

    Less pure sample can be used Can be used to estimate sedimentation coefficient and MW

    Generally used to separate organelles and molecules. Mostcentrifugation work done using preparative ultracentrifuge

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    32/39

    Density gradient centrifugation

    1) Rate zonal centrifugation

    -Separates molecules based on molecular weight-Used to separate various types of macromolecules suchas different types of RNA, DNA

    2) Isopycnic centrifugation-separate molecules based on density

    -used to separate cell organelles

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    33/39

    Differential centrifugation

    -Start with a uniform mixture of sample

    -Centrifuge to obtain pellet of heavier molecules andsupernatant

    -Take supernatant & recentrifuge to pellet the next

    largest particle-Commonly used for separation of organelles

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    34/39

    CLASSIFICATION OF CENTRIFUGATION BASED ONRATE/SPEED

    Ultracentrifugation - x> 20,000 rpm

    Super speed ultracentrifugation 10,000

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    35/39

    ROTOR

    -Two types of rotor : swinging-bucket rotor andfixed-angle rotor

    SWINGING-BUCKET ROTOR

    Consist of a rotor from which

    hang three to six free moving

    bucket that hang vertically when

    the rotor is at rest

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    36/39

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    37/39

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    38/39

    ASSIGNMENT 1

    TITLE : INTRODUCTION TO LABORATORY

    INSTRUMENTS

    Font : 11, Arial, 1.5 spacing

    Pages : Maximum 10

    Cover : Green (Name, ID Num., Lecturers name,Group, Dateline)

    Dateline : 06 JULY 2011

  • 7/29/2019 Chapter 1&2 Introduction & Centrifugation

    39/39

    SAMPLE

    INSTRUMENTATIONSGS 1102

    ASSIGNMENT 1

    INTRODUCTION TO LABORATORY INSTRUMENTS

    NAME :MATRIC NO :LECTURER :GROUP :SUBMISSION DATE :