Upload
jorgeluisunknownman667
View
218
Download
0
Embed Size (px)
Citation preview
8/12/2019 ch068
1/14
D.C. Barber. Electrical Impedance Tomography.The Biomedical Engineering Handbook: Second Edition.
Ed. Joseph D. BronzinoBoca Raton: CRC Press LLC, 2000
8/12/2019 ch068
2/14
2000 by CRC Press LLC
68Electrical ImpedanceTomography68.1 The Electrical Impedance of Tissue68.2 Conduction in Human Tissues
68.3 Determination of the Impedance DistributionData Collection Image Reconstruction Multifrequency
Measurements
68.4 Areas of Clinical ApplicationPossible Biomedical Applications
68.5 Summary and Future Developments
68.1 The Electrical Impedance of Tissue
The specific conductance (conductivity) of human tissues varies from 15.4 mS/cm for cerebrospinal fluid
to 0.06 mS/cm for bone. The difference in the value of conductivity is large between different tissues(Table 68.1). Cross-sectional images of the distribution of conductivity, or alternatively specific resistance
(resistivity), should show good contrast. The aim of electrical impedance tomography (EIT) is to produce
such images. It has been shown [Kohn and Vogelius, 1984a, 1984b; Sylvester and Uhlmann, 1986] that
for reasonable isotropic distributions of conductivity it is possible in principle to reconstruct conductivity
images from electrical measurements made on the surface of an object. Electrical impedance tomography
(EIT) is the technique of producing these images. In fact, human tissue is not simply conductive. There
is evidence that many tissues also demonstrate a capacitive component of current flow and therefore, it
is appropriate to speak of the specific admittance (admittivity) or specific impedance (impedivity) of
tissue rather than the conductivity; hence the use of the world impedance in electrical impedance
tomography.
68.2 Conduction in Human Tissues
Tissue consists of cells with conducting contents surrounded by insulating membranes embedded in a
conducting medium. Inside and outside the cell wall is conducting fluid. At low frequencies of applied
current, the current cannot pass through the membranes, and conduction is through the extracellular
space. At high frequencies, current can flow through the membranes, which act as capacitors. A simple
model of bulk tissue impedance based on this structure, which was proposed by Cole and Cole [1941],
is shown byFig. 68.1.
Clearly, this model as it stands is too simple, since an actual tissue sample would be better represented
as a large network of interconnected modules of this form. However, it has been shown that this modelfits experimental data if the values of the components, especially the capacitance, are made a power
D.C. BarberUniversity of Sheffield
8/12/2019 ch068
3/14 2000 by CRC Press LLC
function of the applied frequency!. An equation which
describes the behavior of tissue impedance as a function
of frequency reasonably well is
(68.1)
where Z0 and Z" are the (complex) limiting values of
tissue impedance low and high frequency and fc is a
characteristic frequency. The value of #allows for the
frequency dependency of the components of the model
and is tissue dependent. Numerical values for in-vivo
human tissues are not well established.
Making measurements of the real and imaginary com-
ponents of tissue impedivity over a range of frequencies
will allow the components in this model to be extracted. Since it is known that tissue structure alters in
disease and that R, S, C are dependent on structure, it should be possible to use such measurements to
distinguish different types of tissue and different disease conditions. It is worth noting that although maxi-
mum accuracy in the determination of the model components can be obtained if both real and imaginary
components are available, in principle, knowledge of the resistive component alone should enable the values
to be determined, provided an adequate range of frequencies is used. This can have practical consequences
for data collection, since accurate measurement of the capacitive component can prove difficult.
Although on a microscopic scale tissue is almost certainly
electrically isotropic, on a macroscopic scale this is not so forsome tissues because of their anisotropic physical structure.
Muscle tissue is a prime example (see Table 68.1), where the bulk
conductivity along the direction of the fibers is significantly
higher than across the fibers. Although unique solutions for
conductivity are possible for isotropic conductors, it can be
shown that for anisotropic conductors unique solutions for con-
ductivity do not exist. There are sets of different anisotropic
conductivity distributions that give the same surface voltage dis-
tributions and which therefore cannot be distinguished by these
measurements. It is not yet clear how limiting anisotropy is to
electrical impedance tomography. Clearly, if sufficient data could
be obtained to resolve down to the microscopic level (this is not
possible practically), then tissue becomes isotropic. Moreover, the tissue distribution of conductivity,
including anisotropy, often can be modeled as a network of conductors, and it is known that a unique
solution will always exist for such a network. In practice, use of some prior knowledge about the
anisotropy of tissue may remove the ambiguities of conductivity distribution associated with anisotropy.
The degree to which anisotropy might inhibit useful image reconstruction is still an open question.
68.3 Determination of the Impedance Distribution
The distribution of electrical potential within an isotropic conducting object through which a low-frequency current is flowing is given by
(68.2)
TABLE 68.1 Values of Specific Conductance
for Human Tissues
Tissue Conductivity, mS/cm
Cerebrospinal fluid 15.4
Blood 6.7Liver 2.8
Skeletal muscle 8.0 (longitudinal)
0.6 (transverse)
Cardiac muscle 6.3 (longitudinal
2.3 (transverse)
Neural tissue 1.7
Gray matter 3.5
White matter 1.5
Lung 1.0 (expiration)
0.4 (inspiration)
Fat 0.36
Bone 0.06
Z Z Z Z
j
= +
+
0
1f
fc
FIGURE 68.1 The Cole-Cole model of
tissue impedance.
( ) = 0
8/12/2019 ch068
4/14
2000 by CRC Press LLC
where $is the potential distribution within the object and %is the distribution of conductivity (generally
admittivity) within the object. If the conductivity is uniform, this reduces to Laplaces equation. Strictly
speaking, this equation is only correct for direct current, but for the frequencies of alternating current
used in EIT (up to 1 MHz) and the sizes of objects being imaged, it can be assumed that this equation
continues to describe the instantaneous distribution of potential within the conducting object. If this
equation is solved for a given conductivity distribution and current distribution through the surface of
the object, the potential distribution developed on the surface of the object may be determined. The
distribution of potential will depend on several things. It will depend on the pattern of current applied
and the shape of the object. It will also depend on the internal conductivity of the object, and it is this
that needs to be determined. In theory, the current may be applied in a continuous and nonuniform
pattern at every point across the surface. In practice, current is applied to an object through electrodes
attached to the surface of the object. Theoretically, potential may be measured at every point on the
surface of the object. Again, voltage on the surface of the object is measured in practice using electrodes
(possibly different from those used to apply current) attached to the surface of the object. There will be
a relationship, the forward solution, between an applied current patternji, the conductivity distribution
%, and the surface potential distribution $iwhich can be formally represented as
. (68.3)
If %andjiare known, $ican be computed. For one current patternji, knowledge of $iis not in general
sufficient to uniquely determine %. However, by applying a complete set of independent current patterns,
it becomes possible to obtain sufficient information to determine %, at least in the isotropic case. This
is the inverse solution.In practice, measurements of surface potential or voltage can only be made at a
finite number of positions, corresponding to electrodes placed on the surface of the object. This also
means that only a finite number of independent current patterns can be applied. For N electrodes, N-1
independent current patterns can be defined and N(N-1)/2 independent measurements made. This latternumber determines the limit of image resolution achievable with N electrodes. In practice, it may not
be possible to collect all possible independent measurements. Since only a finite number of current
patterns and measurements is available, the set of equations represented by Eq. (68.3) can be rewritten as
(68.4)
where v is now a concatenated vector of all voltage values for all current patterns, c is a vector of
conductivity values, representing the conductivity distribution divided into uniform image pixels, and
Aca matrix representing the transformation of this conductivity vector into the voltage vector. Since Ac
depends on the conductivity distribution, this equation is nonlinear. Although formally the precedingequation can be solved for cby inverting Ac, the nonlinear nature of this equation means that this cannot
be done in a single step. An iterative procedure will therefore be needed to obtain c.
Examination of the physics of current flow shows that current tends to take the easiest path possible
in its passage through the object. If the conductivity at some point is changed, the current path redis-
tributes in such a way that the effects of this change are minimized. The practical effect of this is that it
is possible to have fairly large changes in conductivity within the object which only produce relatively
small changes in voltage at the surface of the object. The converse of this is that when reconstructing the
conductivity distribution, small errors on the measured voltage data, both random and systematic, can
translate into large errors in the estimate of the conductivity distribution. This effect forms, and will
continue to form, a limit to the quality of reconstructed conductivity images in terms of resolution,
accuracy, and sensitivity.
Any measurement of voltage must always be referred to a reference point. Usually this is one of the
electrodes, which is given the nominal value of 0 V. The voltage on all other electrodes is determined by
measuring the voltage difference between each electrode and the reference electrode. Alternatively, voltage
i iR j= ( ),
v A cc=
8/12/2019 ch068
5/14 2000 by CRC Press LLC
differences may be measured between pairs of electrodes. A common approach is to measure the voltage
between adjacent pairs of electrodes (Fig. 68.2). Clearly, the measurement scheme affects the form of Ac.
Choice of the pattern of applied currents and the voltage measurement scheme used can affect the
accuracy with which images of conductivity can be reconstructed.
Electrical impedance tomography (EIT) is not a mature technology. However, it has been the subject
of intensive research over the past few years, and this work is still continuing. Nearly all the research
effort has been devoted to exploring the different possible ways of collecting data and producing imagesof tissue resistivity, with the aim of optimizing image reconstruction in terms of image accuracy, spatial
resolution, and sensitivity.
Very few areas of medical application have been explored in any great depth, although in a number of
cases preliminary work has been carried out. Although most current interest is in the use of EIT for medical
imaging, there is also some interest in its use in geophysical measurements and some industrial uses. A
recent detailed review of the state of Electrical Impedance Tomography is given in Boone et al. [1997].
Data Collection
Basic Requirements
Data are collected by applying a current to the object through electrodes connected to the surface of theobject and then making measurements of the voltage on the object surface through the same or other
electrodes. Although conceptually simple, technically this can be difficult. Great attention must be paid
to the reduction of noise and the elimination of any voltage offsets on the measurements. The currents
applied are alternating currents usually in the range 10 kHz to 1 MHz. Since tissue has a complex
impedance, the voltage signals will contain in-phase and out-of-phase components. In principle, both
of these can be measured. In practice, measurement of the out-of-phase (the capacitive) component is
significantly more difficult because of the presence of unwanted (stray) capacitances between various
parts of the voltage measurement system, including the leads from the data-collection apparatus to the
electrodes. These stray capacitances can lead to appreciable leakage currents, especially at the higher
frequencies, which translate into systematic errors on the voltage measurements. The signal measured
on an electrode, or between a pair of electrodes, oscillates at the same frequency as the applied current.
The magnitude of this signal (usually separated into real and imaginary components) is determined,
typically by demodulation and integration. The frequency of the demodulated signal is much less than
the frequency of the applied signal, and the effects of stray capacitances on this signal are generally
FIGURE 68.2 Idealized electrode positions around a conducting object with typical drive and measurement elec-
trode pairs indicated.
8/12/2019 ch068
6/14
2000 by CRC Press LLC
negligible. This realization has led some workers to propose that the signal demodulation and detection
system be mounted as close to the electrodes as possible, ideally at the electrode site itself, and some
systems have been developed that use this approach, although none with sufficient miniaturization of
the electronics to be practical in a clinical setting. This solution is not in itself free of problems, but this
approach is likely to be of increasing importance if the frequency range of applied currents is to be
extended beyond 1 MHz, necessary if the value of the complex impedance is to be adequately explored
as a function of frequency.
Various data-collection schemes have been proposed. Most data are collected from a two-dimensional
(2D) configuration of electrodes, either from 2D objects or around the border of a plane normal to the
principal axis of a cylindrical (in the general sense) object where that plane intersects the object surface.
The simplest data-collection protocol is to apply a current between a pair of electrodes (often an adjacent
pair) and measure the voltage difference between other adjacent pairs (see Fig. 68.2). Although in
principle voltage could be measured on electrodes though which current is simultaneously flowing, the
presence of an electrode impedance, generally unknown, between the electrode and the body surface
means that the voltage measured is not actually that on the body surface. Various means have been
suggested for either measuring the electrode impedance in some way or including it as an unknown inthe image-reconstruction process. However, in many systems, measurements from electrodes through
which current is flowing are simply ignored. Electrode impedance is generally not considered to be a
problem when making voltage measurements on electrodes through which current is not flowing, pro-
vided a voltmeter with sufficiently high input impedance is used, although, since the input impedance
is always finite, every attempt should be made to keep the electrode impedance as low as possible. Using
the same electrode for driving current and making voltage measurements, even at different times in the
data collection cycle, means that at some point in the data-collection apparatus wires carrying current
and wires carrying voltage signals will be brought close together in a switching system, leading to the
possibility of leakage currents. There is a good argument for using separate sets of electrodes for driving
and measuring to reduce this problem. Paulson et al. [1992] have also proposed this approach and also
have noted that it can aid in the modeling of the forward solution (see Image Reconstruction). Brown
et al. [1994] have used this approach in making multi-frequency measurements.
Clearly, the magnitude of the voltage measured will depend on the magnitude of the current applied.
If a constant-current drive is used, this must be able to deliver a known current to a variety of input
impedances with a stability of better than 0.1%. This is technically demanding. The best approach to
this problem is to measure the current being applied, which can easily be done to this accuracy. These
measurements are then used to normalize the voltage data.
The current application and data-collection regime will depend on the reconstruction algorithm used.
Several EIT systems apply current in a distributed manner, with currents of various magnitudes being
applied to several or all of the electrodes. These optimal currents (see Image Reconstruction) must be
specified accurately, and again, it is technically difficult to ensure that the correct current is applied ateach electrode. Although there are significant theoretical advantages to using distributed current patterns,
the increased technical problems associated with this approach, and the higher noise levels associated
with the increase in electronic complexity, may outweigh these advantages.
Although most EIT at present is 2D in the sense given above, it is intrinsically a three-dimensional
(3D) imaging procedure, since current cannot be constrained to flow in a plane through a 3D object.
3D data collection does not pose any further problems apart from increased complexity due to the need
for more electrodes. Whereas most data-collection systems to date have been based on 16 or 32 electrodes,
3D systems will require four times or more electrodes distributed over the surface of the object if adequate
resolution is to be maintained. Technically, this will require belts or vests of electrodes that can be
rapidly applied [McAdams et al., 1994]. Some of these are already available, and the application of an
adequate number of electrodes should not prove insuperable provided electrode-mounted electronics
are not required. Metherell et al. [1996] describe a three-dimensional data collection system and recon-
struction algorithm and note the improved accuracy of three-dimensional images compared to two-
dimensional images constructed using data collected from three-dimensional objects.
8/12/2019 ch068
7/14 2000 by CRC Press LLC
Performance of Existing Systems
Several research groups have produced EIT systems for laboratory use [Brown and Seagar, 1987; Smith
et al., 1990; Rigaud et al., 1990; Jossinet and Trillaud, 1992; Lidgey et al., 1992; Riu et al., 1992; Gisser
et al., 1991; Cook et al., 1994; Brown et al., 1994; Zhu et al., 1994; Cusick et al., 1994] and some clinical
trials. The complexity of the systems largely depends on whether current is applied via a pair of electrodes(usually adjacent) or whether through many electrodes simultaneously. The former systems are much
simpler in design and construction and can deliver higher signal-to-noise ratios. The Sheffield Mark II
system (Smith et al., 1990) used 16 electrodes and was capable of providing signal-to-noise ratios of up
to 68 dB at 25 datasets per second. The average image resolution achievable across the image was 15%
of the image diameter. In general, multi-frequency systems have not yet delivered similar performance,
but are being continuously improved.
Spatial resolution and noise levels are the most important constraints on possible clinical applications
of EIT. As images are formed through a reconstruction process the values of these parameters will depend
critically on the quality of the data collected and the reconstruction process used. However practical
limitations to the number of electrodes which can be used and the ill-posed nature of the reconstruction
problem make it unlikely that high quality images, comparable to other medical imaging modalities, canbe produced. The impact of this on diagnostic performance still needs to be evaluated.
Image Reconstruction
Basics of Reconstruction
Although several different approaches to image reconstruction have been tried [Wexler, 1985; Yorkey, et
al., 1986; Barber and Seagar, 1987, Kim et al., 1987; Hua et al., 1988; Cheny et al., 1990; Breckon, 1990;
Zadecoochak et al., 1991; Kotre et al., 1992; Bayford et al., 1994; Morucci et al., 1994], the most accurate
approaches are based broadly on the following algorithm. For a given set of current patterns, a forward
transform is set up for determining the voltages v produced form the conductivity distribution c(Eq. 68.4). Acis dependent on c,so it is necessary to assume an initial starting conductivity distribution c0.
This is usually taken to be uniform. Using Ac, the expected voltages v0are calculated and compared with
the actual measured voltages vm . Unless c0is correct (which it will not be initially), v0and vmwill differ.
It can be shown that an improved estimate of cis given by
(68.5)
(68.6)
where Scis the differential of Acwith respect to c, the sensitivity matrix and Stc is the transpose of Sc. The
improved value of cis then used in the next iteration to compute an improved estimate of vm, i.e., v1.
This iterative process is continued until some appropriate endpoint is reached. Although convergence is
not guaranteed, in practice, convergence to the correct cin the absence of noise can be expected, provided
a good starting value is chosen. Uniform conductivity seems to be a reasonable choice. In the presence
of noise on the measurements, iteration is stopped when the difference between vand vmis within the
margin of error set by the known noise on the data.
There are some practical difficulties associated with this approach. One is that large changes in cmay
only produce small changes inv,and this will be reflected in the structure of Sc, making Stc Scvery difficult
to invert reliably. Various methods of regularization have been used, with varying degrees of success, to
achieve stable inversion of this matrix although the greater the regularization applied the poorer theresolution that can be achieved. A more difficult practical problem is that for convergence to be possible
the computed voltagesvmust be equal to the measured voltagesvmwhen the correct conductivity values
are used in the forward calculation. Although in a few idealized cases analytical solutions of the forward
c S S S v v ct
c ct= ( ) ( )
1
0 m
c c c1 0
= +
8/12/2019 ch068
8/14
8/12/2019 ch068
9/14 2000 by CRC Press LLC
Three-Dimensional Imaging
Most published work so far on image reconstruction has concentrated on solving the 2D problem.
However, real medical objects, i.e., patients, are three-dimensional. Theoretically, as the dimensionality
of the object increases, reconstruction should become better conditioned. However, unlike 3D x-ray
images, which can be constructed from a set of independent 2D images, EIT data from 3D objects cannotbe so decomposed and data from over the whole surface of the object is required for 3D reconstruction.
The principles of reconstruction in 3D are identical to the 2D situation although practically the problem
is quite formidable, principally because of the need to solve the forward problem in three dimensions.
Some early work on 3D imaging was presented by Goble and Isaacson [1990]. More recently Metherall
et al. [1996] have shown images using data collected from human subjects.
Single-Step Reconstruction
The complete reconstruction problem is nonlinear and requires iteration. However, each step in the
iterative process is linear. Images reconstructed using only the first step of iteration effectively treat image
formation as a liner process, an assumption approximately justified for small changes in conductivity
from uniform. In the case the functions Acand Scoften can be precomputed with reasonable accuracybecause they usually are computed for the case of uniform conductivity. Although the solution cannot
be correct, since the nonlinearity is not taken into account, it may be useful, and first-step linear
approximations have gained some popularity. Cheney et al. [1990] have published some results from a
first-step process using optimal currents. Most, if not all, of the clinical images produced to date have
used a single-step reconstruction algorithm [Barber and Seagar, 1987; Barber and Brown, 1990]. Although
this algorithm uses very nonoptimal current patterns, this has not so far been a limitation because of
the high quality of data collected and the limited number of electrodes used (16). With larger numbers
of electrodes, this conclusion may need to be revised.
Differential Imaging
Ideally, the aim of EIT is to reconstruct images of the absolute distribution of conductivity (or admit-tivity). These images are known as absolute (or static) images. However, this requires that the forward
problem can be solved to an high degree of accuracy, and this can be difficult. The magnitude of the
voltage signal measured on an electrode or between electrodes will depend on the body shape, the
electrode shape and position, and the internal conductivity distribution. The signal magnitude is in fact
dominated by the first two effects rather than by conductivity. However, if a changein conductivity occurs
within the object, then it can often be assumed that the changein surface voltage is dominated by this
conductivity change. In differential (or dynamic) imaging, the aim is to image changes in conductivity
rather than absolute values. If the voltage difference between a pair of (usually adjacent) electrodes before
a conductivity change occurs is g1and the value after change occurs is g2, then a normalized data value
is defined as
(68.7)
Many of the effects of body shape (including interpreting the data as coming from a 2D object when in
fact it is from a 3D object) and electrode placing at least partially cancel out in this definition. The values
of the normalized data are determined largely by the conductivity changes. It can be argued that the
relationship between the (normalized) changes in conductivity &cn and the normalized changes in
boundary data &gnis given by
(68.8)
where Fis a sensitivity matrix which it can be shown is much less sensitive to object shape end electrode
positions than the sensitivity matrix of Eq. 68.5. Although images produced using this algorithm are not
gg g
g g
g
gn=
+
=2 1 21 2 mean
g F cn n=
8/12/2019 ch068
10/14 2000 by CRC Press LLC
completely free of artifact that is the only algorithm which has reliably produced images using data taken
from human subjects. Metherall et al. [1996] used a version of this algorithm for 3D imaging and Brown
et al. [1994] for multi-frequency imaging, in this case imaging changes in conductivity with frequency.
The principle disadvantage of this algorithm is that it can only image changes in conductivity, which
must be either natural or induced. Figure 68.3 shows a typical differential image. This represents the
changes in the conductivity of the lungs between expiration and inspiration.
Multifrequency Measurements
Differential algorithms can only image changes in conductivity. Absolute distributions of conductivity
cannot be produced using these methods. In addition, any gross movement of the electrodes, either
because they have to be removed and replaced or even because of significant patient movement, make
the use of this technique difficult for long-term measurements of changes. As an alternative to changes
in time, differential algorithms can image changes in conductivity with frequency. Brown et al. [1994]
have shown that if measurements are made over a range of frequencies and differential images produced
using data from the lowest frequency and the other frequencies in turn, these images can be used to
compute parametric images representing the distribution of combinations of the circuit values inFig. 68.1.For example, images representing the ratio of Sto R,a measure of the ratio of intracellular to extracellular
volume, can be produced, as well as images of fo= 1/2'(RC+ SC), the tissue characteristic frequency.
Although not images of the absolute distribution of conductivity, they are images of absolute tissue
properties. Since these properties are related to tissue structure, they should produce images with useful
contrast. Data sufficient to reconstruct an image can be collected in a time short enough to preclude
significant patient movement, which means these images are robust against movement artifacts. Changes
of these parameters with time can still be observed.
68.4 Areas of Clinical Application
There is no doubt that the clinical strengths of EIT relate to its ability to be considered as a functionalimaging modality that carries no hazard and can therefore be used for monitoring purposes. The best
spatial resolution that might become available will still be much worse than anatomic imaging methods
such as magnetic resonance imaging and x-ray computed tomography. However, EIT is able to image
small changes in tissue conductivity such as those associated with blood perfusion, lung ventilation, and
FIGURE 68.3 A differential conductivity image representing the changes in conductivity in going from maximum
inspiration (breathing in) to maximum expiration (breathing out). Increasing blackness represents increasing conduc-
tivity.
8/12/2019 ch068
11/14 2000 by CRC Press LLC
fluid shifts. Clinical applications seek to take advantage of the ability of EIT to follow rapid changes in
physiologic function.
There are several areas in clinical medicine where electrical impedance tomography might provide
advantages over existing techniques. These have been received elsewhere [Brown et al., 1985; Dawids,
1987; Holder and Brown, 1990; Boone et al., 1997].
Possible Biomedical Applications
Gastrointestinal System
A priori,it seems likely that EIT could be applied usefully to measurement of motor activity in the gut.
Electrodes can be applied with ease around the abdomen, and there are no large bony structures likely
to seriously violate the assumption of constant initial conductivity. During motor activity, such as gastric
emptying or peristalsis, there are relatively large movements of the conducting fluids within the bowel.
The quantity of interest is the timing of activity, e.g., the rate at which the stomach empties, and the
absolute impedance change and its exact location in space area of secondary importance. The principal
limitations of EIT of poor spatial resolution and amplitude measurement are largely circumvented inthis application [Avill et al., 1987]. There has also been some interest in the measurement of other aspects
of gastro-intestinal function such as oesophageal activity [Erol et al., 1995].
Respiratory System
Lung pathology can be imaged by conventional radiography, x-ray computed tomography, or magnetic
resonance imaging, but there are clinical situations where it would be desirable to have a portable means
of imaging regional lung ventilation which could, if necessary, generate repeated images over time.
Validation studies have shown that overall ventilation can be measured with good accuracy [Harris et
al., 1988]. More recent work [Hampshire et al., 1995] suggests that multifrequency measurements may
be useful in the diagnosis of some lung disorders.
EIT Imaging of Changes in Blood Volume
The conductivity of blood is about 6.7 mS/cm, which is approximately three times that of most intratho-
racic tissues. It therefore seems possible that EIT images related to blood flow could be accomplished.
The imaged quantity will be the change in conductivity due to replacement of tissue by blood (or vice
versa) as a result of the pulsatile flow through the thorax. This may be relatively large in the cardiac
ventricles but will be smaller in the peripheral lung fields.
The most interesting possibility is that of detecting pulmonary embolus (PE). If a blood clot is present
in the lung, the lung beyond the clot will not be perfused, and under favorable circumstances, this may
be visualized using a gated blood volume image of the lung. In combination with a ventilation image,
which should show normal ventilation in this region, pulmonary embolism could be diagnosed. Some
data [Leathard et al., 1994] indicate that PE can be visualized in human subjects. Although more sensitive
methods already exist for detecting PE, the noninvasiveness and bedside availability of EIT mean that
treatment of the patient could be monitored over the period following the occurrence of the embolism,
an important aim, since the use of anticoagulants, the principal treatment for PE, needs to be minimized
in postoperative patients, a common class of patients presenting with this complication.
68.5 Summary and Future Developments
EIT is still an emerging technology. In its development, several novel and difficult measurement and
image-reconstruction problems have had to be addressed. Most of these have been satisfactorily solved.
The current generation of EIT imaging systems are multifrequency, with some capable of 3D imaging.These should be capable of greater quantitative accuracy and be less prone to image artifact and are likely
to find a practical role in clinical diagnosis. Although there are still many technical problems to be
answered and many clinical applications to be addressed, the technology may be close to coming of age.
8/12/2019 ch068
12/14
2000 by CRC Press LLC
Defining Terms
Absolute imaging: Imaging the actual distribution of conductivity.
Admittivity: The specific admittance of an electrically conducting material. For simple biomedical
materials such as saline with no reactive component of resistance, this is the same as conductivity.
Anisotropic conductor: A material in which the conductivity is dependent on the direction in whichit is measured through the material.
Applied current pattern: In EIT, the electric current is applied to the surface of the conducting object
via electrodes placed on the surface of the object. The spatial distribution of current flow through
the surface of the object is the applied current pattern.
Bipolar current pattern: A current pattern applied between a single pair of electrodes.
Conductivity: The specific conductance of an electrically conducting material. The inverse of resistivity.
Differential imaging: An EIT imaging technique that specifically images changes in conductivity.
Distributed current: A current pattern applied through more than two electrodes.
Dynamic imaging: The same as differential imaging.
EIT: Electrical impedance tomography.Forward transform or problem or solution: The operation, real or computational, that maps or
transforms the conductivity distribution to surface voltages.
Impedivity: The specific impedance of an electrically conducting material. The inverse of admittivity.
For simple biomedical materials such as saline with no reactive component of resistance, this is
the same as resistivity.
Inverse transform or problem or solution: The computational operation that maps voltage measure-
ments on the surface of the object to the conductivity distribution.
Optimal current: One of a set of a current patterns computed for a particular conductivity distribution
that produce data with maximum possible SNR.
Pixel: The conductivity distribution is usually represented as a set of connected piecewise uniform
patches. Each of these patches is a pixel. The pixel may take any shape, but square or triangularshapes are most common.
Resistivity: The specific electrical resistance of an electrical conducting material. The inverse of con-
ductivity.
Static imaging: The same as absolute imaging.
References
Avill RF, Mangnall RF, Bird NC, et al. 1987. Applied potential tomography: A new non-invasive technique
for measuring gastric emptying. Gastroenterology 92:1019.
Barber DC, Brown BH. 1990. Progress in electrical impedance tomography. In D Colton, R Ewing,
W Rundell (eds), Inverse Problems in Partial Differential Equations, pp 149162. New York, SIAM.Barber DC, Seagar AD. 1987. Fast reconstruction of resistive images. Clin Phys Physiol Meas 8(A):47.
Bayford R. 1994. PhD thesis. Middlesex University, U.K.
Boone K, Barber D, Brown B. 1992. Imaging with electricity: Report of the European Concerted Action
on Impedance Tomography. J Med Eng & Tech 21:6 pp 201232.
Breckon WR. 1990. Image Reconstruction in Electrical Impedance Tomography. PhD thesis, School of
Computing and Mathematical Sciences, Oxford Polytechnic, Oxford, U.K.
Brown BH, Barber DC, Seagar AD. 1985. Applied potential tomography: Possible clinical applications.
Clin Phys Physiol Meas 6:109.
Brown BH, Barber DC, Wang W, et al. 1994. Multifrequency imaging and modelling of respiratory related
electrical impedance changes. Physiol Meas 15:A1.Brown DC, Seagar AD. 1987. The Sheffield data collection system. Clinical Physics and Physiological
Measurement 8(suppl A):9198.
Cheney MD, Isaacson D, Newell J, et al. 1990. Noser: An algorithm for solving the inverse conductivity
problem. Int J Imag Sys Tech 2:60.
8/12/2019 ch068
13/14 2000 by CRC Press LLC
Cole KS, Cole RH. 1941. Dispersion and absorption in dielectrics: I. Alternating current characteristics.
J Chem Phys 9:431.
Cook RD, Saulnier GJ, Gisser DG, Goble J, Newell JC, Isaacson D. 1994. ACT3: A high-speed high-
precision electrical impedance tomograph. IEEE Transactions on Biomedical Engineering,
41:713722.
Cusick G, Holder DS, Birkett A, Boone KG. 1994. A system for impedance imaging of epilepsy in ambu-
latory human subjects. Innovation and Technology in Biology and Medicine 15(suppl 1):3439.
Dawids SG. 1987. Evaluation of applied potential tomography: A clinicians view. Clin Phys Physiol Meas
8(A):175.
Erol RA, Smallwood RH, Brown BH, Cherian P, Bardham KD. 1995. Detecting oesophageal-related changes
using electrical impedance tomography. Physiological Measurement 16(suppl 3A):143152.
Gisser DG, Newell JC, Salunier G, Hochgraf C, Cook RD, Goble JC. 1991. Analog electronics for a high-
speed high-precision electrical impedance tomograph. Proceedings of the IEEE EMBS 13:2324.
Goble J, Isaacson D. 1990. Fast reconstruction algorithms for three-dimensional electrical tomography.
In IEEE EMBS Proceedings of the 12th Annual International Conference, Philadelphia, pp 285286.
Hampshire AR, Smallwood RH, Brown BH, Primhak RA. 1995. Multifrequency and parametric EITimages of neonatal lungs. Physiological Measurement 16(suppl 3A):175189.
Harris ND, Sugget AJ, Barber DC, Brown BH. 1988. Applied potential tomography: A new technique for
monitoring pulmonary function. Clin Phys Physiol Meas 9(A):79.
Holder DS, Brown BH. 1990. Biomedical applications of EIT: A critical review. In D Holder (ed), Clinical
and Physiological Applications of Electrical Impedance Tomography, pp 641. London, UCL Press.
Hua P, Webster JG, Tompkins WJ. 1988. A regularized electrical impedance tomography reconstruction
algorithm. Clinical Physics and Physiological Measurement 9(suppl A):137141.
Isaacson D. 1986. Distinguishability of conductivities by electric current computed tomography. IEEE
Trans Med Imaging 5:91.
Jossinet J, Trillaud C. 1992. Imaging the complex impedance in electrical impedance tomography. Clin
Phys Physiol Meas 13(A):47.
Kim H, Woo HW. 1987. A prototype system and reconstruction algorithms for electrical impedance
technique in medical imaging. Clin Phys Physiol Meas 8(A):63.
Kohn RV, Vogelius M. 1984a. Determining the conductivity by boundary measurement. Comm Pure
Appl Math 37:289.
Kohn RV, Vogelius M. 1984b. Identification of an unknown conductivity by means of the boundary.
SIAM-AMS Proc 14:113.
Koire CJ. 1992. EIT image reconstruction using sensitivity coefficient weighted backprojection. Physio-
logical Measurement 15(suppl 2A):125136.
Leathard AD, Brown BH, Campbell J, et al. 1994. A comparison of ventilatory and cardiac related changes
in EIT images of normal human lungs and of lungs with pulmonary embolism. Physiol Meas15:A137.
Lidgey FJ, Zhu QS, McLeod CN, Breckon W. 1992. Electrode current determination from programmable
current sources. Clinical Physics and Physiological Measurement 13(suppl A):4346.
McAdams ET, McLaughlin JA, Anderson JMcC. 1994. Multielectrode systems for electrical impedance
tomography. Physiol Meas 15:A101.
Metherall P, Barber DC, Smallwood RH, Brown BH. 1996. Three-dimensional electrical impedance
tomography. Physiol Meas 15:A101.
Morucci JP, Marsili PM, Granie M, Dai WW, Shi Y. 1994. Direct sensitivity matrix approach for fast
reconstruction in electrical impedance tomography. Physiological Measurement 15(suppl
2A):107114.
Riu PJ, Rosell J, Lozano A, Pallas-Areny RA. 1992. Broadband system for multi-frequency static imaging
in electrical impedance tomography. Clin Phys Physiol Meas 13(A):61.
Smith RWM. 1990. Design of a Real-Time Impedance Imaging System for Medical Applications. Ph.D.
thesis, University of Sheffield, U.K.
8/12/2019 ch068
14/14
Sylvester J, Uhlmann G. 1986. A uniqueness theorem for an inverse boundary value problem in electrical
prospection. Comm Pure Appl Math 39:91.
Wexler A, Fry B, Neuman MR. 1985. Impedance-computed tomography: Algorithm and system. Appl
Optics 24:3985.
Yorkey TJ. 1986. Comparing Reconstruction Algorithms for Electrical Impedance Imaging. Ph.D. thesis,
University of Wisconsin, Madison, Wisc.
Zadehkoochak M, Blott BH, Hames TK, George RE. 1991. Special expansion in electrical impedance
tomography. Journal of Physics D: Applied Physics 24:19111916.
Zhu QS, McLeod CN, Denyer CW, Lidgey FL, Lionheart WRB. 1994. Development of a real-time adaptive
current tomograph. Clinical Physics and Physiological Measurement 15(suppl 2A):3743.
Further Information
All the following conferences were funded by the European commission under the biomedical engineering
program. The first two were directly funded as exploratory workshops, the remainder as part of a
Concerted Action on Electrical Impedance Tomography. Electrical Impedance TomographyAppliedPotential Tomography. 1987. Proceedings of a conference held in Sheffield, U.K., 1986. Published in Clin
Phys Physiol Meas 8:Suppl.A. Electrical Impedance TomographyApplied Potential Tomography. 1988.
Proceedings of a conference held in Lyon, France, November 1987. Published in Clin Phys Physiol Meas
9:Suppl.A. Electrical Impedance Tomography. 1991. Proceedings of a conference held in Copenhagen,
Denmark, July 1990. Published by Medical Physics, University of Sheffield, U.K. Electrical Impedance
Tomography. 1992. Proceedings of a conference held in York, U.K., July 1991. Published in Clin Phys
Physiol Meas 13:Suppl.A. Clinical and Physiologic Applications of Electrical Impedance Tomography.
1993. Proceedings of a conference held at the Royal Society, London, U.K. April 1992. Ed. D.S. Holder,
UCL Press, London. Electrical Impedance Tomography. 1994. Proceedings of a conference held in Bar-
celona, Spain, 1993. Published in Clin Phys Physiol Meas 15:Suppl.A.