Page 1 of 12 Manufacturer: Agilent Technologies, Inc. Model Number: 33120A Serial Number: US12345678 Description: 15 MHz Waveform Generator Options Installed: Customer Asset No: Customer: Agilent Technologies Sales & Services GmbH & Co. KG Herrenberger Str. 130 71034 Böblingen GERMANY Location of Calibration: Procedure: STE-50111088-A.02.26 Date of Calibration: 2 Mar 2011 Temperature: (23±5) °C Product Firmware Revision: NA Customer PO Number: Humidity: (50±30) %RH Conformance with the requirements of ISO/IEC 17025:2005 for testing and calibration laboratories is proven by the accreditation of our calibration laboratory. Our quality management system is ISO 9001:2008 certified. As Received Condition Initial testing found the equipment to be IN SPECIFICATION at the points tested including the expanded measurement uncertainty based on the stated coverage probability. As Completed Condition The equipment was adjusted to OPTIMISE the performance as recommended by Agilent Technologies. At the completion of the calibration, the measured values of the equipment including the expanded measurement uncertainty were IN SPECIFICATION at the points tested based on the stated coverage probability. Remarks or Special Requirements: The test limits stated in the measurement report correspond to the published specifications of the equipment, at the points tested. The calibration document is retained for a period of six years. Agilent Technologies UK Ltd 610 Wharfedale Road Winnersh Triangle WOKINGHAM BERK RG415TP UK Issue date: 2 Mar 2011 Peter Spokes Approved Signatory ……….. ……….. Agilent Technologies UK Ltd 610 Wharfedale Road Winnersh Triangle WOKINGHAM BERK RG415TP UK Certificate of Calibration Accredited Calibration Certificate Number: 1-1234567890-1

Certificate of Calibration Certificate of Calibration Page ... · Conformance with the requirements of ISO/IEC 17025:2005 for testing and calibration laboratories is proven by the

  • Upload
    others

  • View
    25

  • Download
    1

Embed Size (px)

Citation preview

Page 1: Certificate of Calibration Certificate of Calibration Page ... · Conformance with the requirements of ISO/IEC 17025:2005 for testing and calibration laboratories is proven by the

Certificate of CalibrationAccredited Calibration

Certificate Number: 1-3127808697-1

Page 1 of 12

Manufacturer: Agilent Technologies, Inc.Model Number: 33120ASerial Number: US12345678

Description: 15 MHz Waveform GeneratorOptions Installed:Customer Asset No:

Customer:Agilent Technologies Sales & Services GmbH & Co. KGHerrenberger Str. 13071034 BöblingenGERMANY

Location of Calibration:Agilent Technologies Agilent Tech. Sales & Serv. GmbH & Co.KG Herrenberger Str. 130, Geb. 4 D-71034 Boeblingen - Germany Tel:+49 (0) 7031-464 6333

Procedure: STE-50111088-A.02.26Date of Calibration: 2 Mar 2011Temperature: (23±5) °C

Product Firmware Revision: NACustomer PO Number:Humidity: (50±30) %RH

Conformance with the requirements of ISO/IEC 17025:2005 for testing and calibration laboratories is proven by the accreditation of our calibration laboratory. Our quality management system is ISO 9001:2008 certified.

As Received ConditionInitial testing found the equipment to be IN SPECIFICATION at the points tested including the expanded measurement uncertainty based on the stated coverage probability.

As Completed ConditionThe equipment was adjusted to OPTIMISE the performance as recommended by Agilent Technologies. At the completion of the calibration, the measured values of the equipment including the expanded measurement uncertainty were IN SPECIFICATION at the points tested based on the stated coverage probability.

Remarks or Special Requirements:The test limits stated in the measurement report correspond to the published specifications of the equipment, at the points tested.The calibration document is retained for a period of six years.

Agilent Technologies UK Ltd610 Wharfedale RoadWinnersh TriangleWOKINGHAM BERK RG415TPUK

Issue date: 2 Mar 2011

Peter SpokesApproved Signatory

………..

………..

Agilent Technologies UK Ltd610 Wharfedale RoadWinnersh TriangleWOKINGHAM BERK RG415TPUK

Certificate of CalibrationAccredited Calibration

Certificate Number: 1-1234567890-1

Page 2: Certificate of Calibration Certificate of Calibration Page ... · Conformance with the requirements of ISO/IEC 17025:2005 for testing and calibration laboratories is proven by the

Certification ReportAgilent Calibration

Certificate Number: 1-1234567890-1

Page 2 of 12

Traceability Information

Technician ID Number: 123456 Traceability is to national standards Euramet members (NPL, PTB, LNE, etc), or other recognized standards laboratories. Supporting documentation relative to traceability is available for review by appointment. This certificate shall not be reproduced, except in full, without prior written approval of the laboratory.

Calibration Equipment Used

Model Number Model Description Trace Number Cal Due Date Certificate Number11049A Converter DE2821 25 Feb 2012 1-3104673391-13458A Digital Multimeter, 8.5 Digit DE2778 29 Apr 2011 1-2513706471-153132A Universal Counter DE2585 30 Apr 2011 1-2303328290-1E4440A PSA Spectrum Analyzer 3 Hz - 26.5 GHz DE2360 12 Feb 2012 1-2779111774-1

Traceability Table

Model Model Description ID Number Certificate Number Trace ValueW 11049A Converter DE2821 1-3104673391-1

R 11049A Thermal Converter DE921 1-2358275070-1-UKAS:C 0147 AC VoltageRF Voltage

W 3458A Digital Multimeter, 8.5 Digit DE2778 1-2513706471-1R 3458A System Multimeter DE943 1-2462132436-1-UKAS:C 0147 AC Voltage

DC Voltage

W 53132A Universal Counter DE2585 1-2303328290-1R 5065A Frequency Standard DE648 1-1825493436-1-UKAS:C 0147 Frequency

W E4440A PSA Spectrum Analyzer 3 Hz - 26.5GHz

DE2360 1-2779111774-1

R 5065A Frequency Standard DE648 1-2552526447-1-UKAS:C 0147 FrequencyR 8482A Power Sensor DE3062 1-2516730581-1-A2LA:1920.01 RF Power

Legend:

W - Working standard. Measurement standard that is used routinely to calibrate or verify measuring instruments or measuringsystems.R - Reference standard. Measurement standard designated for the calibration of working standards for quantities of a given kindin a given organization or at a given location.

nesaydin
Text Box
Certificate of Calibration
nesaydin
Text Box
Accredited Calibration Certificate Number: 1-1234567890-1
Page 3: Certificate of Calibration Certificate of Calibration Page ... · Conformance with the requirements of ISO/IEC 17025:2005 for testing and calibration laboratories is proven by the

Certification ReportAgilent Calibration

Certificate Number: 1-3127808697-1

Page 3 of 12

Compliance with specificationIn the assessment of compliance with specification, the uncertainty of measurement has been taken into account. If the uncertainty ofmeasurement overlaps the specification limit (upper limit or lower limit), it is not possible to state compliance/non-compliance basedon a 95% level of confidence. However, where a confidence level less than 95% is acceptable, a compliance/non-compliance statementmay be possible. The status of compliance with the acceptance criteria is reported as:

• Pass (P): Compliance with specification.• Pass# (P#): Compliance with specification providing a lower level of confidence at one or more tested points is acceptable• Fail# (F#): Non-compliance with specification providing a lower level of confidence at one or more tested points is acceptable• Fail (F): Non-compliance with specification.• Not tested (NT): Not measured and not tested for compliance with specification.

Upper Limit

Nominal

Lower LimitPass (P) Pass# (P#) Fail# (F#) Fail (F)

A compilation for all performed tests of the status as received, i.e. before any adjustment/repair, and the status as completed, i.e. afterany adjustment/repair, is reported on the first page. The compliance with typical, i.e. non-warranted specifications will not affect thestatus as received or the status as completed reported on the first page. The status summaries relate to the tested item only. A final decision about whether the item's performance actually satisfiesrequirements of the user can only be made by the user. Uncertainty of MeasurementThe reported expanded uncertainty of measurement is the standard uncertainty multiplied by the coverage factor k=2 (for a normaldistribution) or k=1.65 (for a uniform distribution), which corresponds to a coverage probability of approximately 95%. Where this isnot the case, the distribution, coverage factor, k, effective degrees of freedom, veff, and coverage probability, p, are stated. Any quoted uncertainty measurement applies only to the measured value and does not imply anything regarding the long-term stabilityof the device under test.

nesaydin
Text Box
Accredited Calibration Certificate Number: 1-1234567890-1
nesaydin
Text Box
Certificate of Calibration
Page 4: Certificate of Calibration Certificate of Calibration Page ... · Conformance with the requirements of ISO/IEC 17025:2005 for testing and calibration laboratories is proven by the

Certification ReportAgilent Calibration

Certificate Number: 1-3127808697-1

Page 4 of 12

Performance Test Results Summary

Test Name As Received Status As Completed StatusSELF TEST PASSED PASSEDFREQUENCY PASSED PASSEDHARMONICS_LOW PASSED PASSEDHARMONICS PASSED PASSEDSQ DUTY CYCLE PASSED PASSEDAC/DC VOLTS PASSED PASSEDAMPLITUDE FLATNESS PASSED PASSED

nesaydin
Text Box
Certificate of Calibration
nesaydin
Text Box
Accredited Calibration Certificate Number: 1-1234567890-1
Page 5: Certificate of Calibration Certificate of Calibration Page ... · Conformance with the requirements of ISO/IEC 17025:2005 for testing and calibration laboratories is proven by the

Measurement Report Certificate Number: 1-1234567890-1

Page 5 of 12

Model Number: 33120A Serial Number: US12345678Product Firmware Revision: NA Options Installed:

Date of Calibration: 2 Mar 2011Condition: As Received

SELF TEST PASSED

TESTCONDITIONS

STAT

SELF TEST +0

FREQUENCY PASSED

TEST CONDITIONS MINIMUM MEASURED MAXIMUM UNCERT. 1 kHz SINE WAVE 999.980 Hz 1000.000 Hz 1000.020 Hz 0.0050 Hz P

BURST MODE, 500 Hz 495.000 Hz 500.200 Hz 505.000 Hz 0.0030 Hz P

HARMONICS_LOW PASSED

TEST CONDITIONS MEASURED MAXIMUM UNCERT. REFERENCE 20 kHz, 1.1 Vrms -16.3 dBm

2nd Harmonic, 40 kHz -74.8 dBc -70.0 dBc 1.2 dB P

3rd Harmonic, 60 kHz -81.6 dBc -70.0 dBc 1.4 dB P

4th Harmonic, 80 kHz -88.4 dBc -70.0 dBc 1.8 dB P

5th Harmonic, 100 kHz -93.9 dBc -70.0 dBc 1.2 dB P

REFERENCE 100 kHz, 1.1 Vrms -16.3 dBm

2nd Harmonic, 200 kHz -74.4 dBc -60.0 dBc 1.5 dB P

3rd Harmonic, 300 kHz -73.1 dBc -60.0 dBc 1.2 dB P

4th Harmonic, 400 kHz -86.1 dBc -60.0 dBc 1.5 dB P

5th Harmonic, 500 kHz -98.3 dBc -60.0 dBc 1.7 dB P

REFERENCE 1 MHz, 1.1 Vrms -16.3 dBm

2nd Harmonic, 2 MHz -75.1 dBc -45.0 dBc 1.1 dB P

3rd Harmonic, 3 MHz -60.6 dBc -45.0 dBc 1.0 dB P

4th Harmonic, 4 MHz -83.1 dBc -45.0 dBc 1.0 dB P

5th Harmonic, 5 MHz -81.4 dBc -45.0 dBc 1.0 dB P

HARMONICS PASSED

TEST CONDITIONS MEASURED MAXIMUM UNCERT. REFERENCE 15 MHz, 1.1 Vrms -16.3 dBm

2nd Harmonic, 30 MHz -49.2 dBc -35.0 dBc 1.7 dB P

3rd Harmonic, 45 MHz -52.1 dBc -35.0 dBc 1.7 dB P

4th Harmonic, 60 MHz -59.2 dBc -35.0 dBc 1.7 dB P

5th Harmonic, 75 MHz -68.8 dBc -35.0 dBc 1.7 dB P

Page 6: Certificate of Calibration Certificate of Calibration Page ... · Conformance with the requirements of ISO/IEC 17025:2005 for testing and calibration laboratories is proven by the

Measurement Report Certificate Number: 1-1234567890-1

Page 6 of 12

Model Number: 33120A Serial Number: US12345678Product Firmware Revision: NA Options Installed:

Date of Calibration: 2 Mar 2011Condition: As Received

SQ DUTY CYCLE PASSEDSquare Wave, 1.0 Vrms at 300 Hz

TESTCONDITIONS

MINIMUM MEASURED MAXIMUM UNCERT.

50% DUTY CYCLE

-0.010 Vdc 0.000 Vdc 0.010 Vdc 0.0030 V P

25% DUTY CYCLE

-0.510 Vdc -0.503 Vdc -0.490 Vdc 0.0030 V P

75% DUTY CYCLE

0.490 Vdc 0.499 Vdc 0.510 Vdc 0.0020 V P

AC/DC VOLTS PASSED

TESTCONDITIONS

MINIMUM MEASURED MAXIMUM UNCERT.

Function Gain and Linearity Verification SIN 7.0 Vrms 6.930 Vrms 7.009 Vrms 7.070 Vrms 0.0028 V P

SIN 5.7 Vrms 5.643 Vrms 5.709 Vrms 5.757 Vrms 0.0025 V P

TRI 5.7 Vrms 5.643 Vrms 5.708 Vrms 5.757 Vrms 0.0027 V P

RAMP 5.7 Vrms 5.643 Vrms 5.708 Vrms 5.757 Vrms 0.0026 V P

SQU 10.0 Vrms 9.900 Vrms 10.012 Vrms 10.100 Vrms 0.0035 V P

SQU 8.0 Vrms 7.920 Vrms 8.013 Vrms 8.080 Vrms 0.0031 V P

DC Function Offset Verification DC 10 V 9.800 Vdc 9.997 Vdc 10.200 Vdc 0.59 mV P

DC -10 V -10.200 Vdc -10.006 Vdc -9.800 Vdc 0.59 mV P

AM Modulation Depth Verification AM MOD, 0% 0.495 Vrms 0.500 Vrms 0.505 Vrms 0.62 mV P

AM MOD 100% 0.6039 Vrms 0.6134 Vrms 0.6161 Vrms 0.00027 V P

AC Amplitude Verification High Impedance Load SIN 7.0 Vrms 6.930 Vrms 7.009 Vrms 7.070 Vrms 0.0028 V P

SIN 5.7 Vrms 5.643 Vrms 5.708 Vrms 5.757 Vrms 0.0026 V P

SIN 5.5 Vrms 5.445 Vrms 5.508 Vrms 5.555 Vrms 0.0024 V P

SIN 4.4 Vrms 4.356 Vrms 4.407 Vrms 4.444 Vrms 0.0023 V P

SIN 3.5 Vrms 3.465 Vrms 3.505 Vrms 3.535 Vrms 0.0020 V P

SIN 2.8 Vrms 2.772 Vrms 2.804 Vrms 2.828 Vrms 0.0019 V P

SIN 2.2 Vrms 2.178 Vrms 2.204 Vrms 2.222 Vrms 0.0017 V P

SIN 1.7 Vrms 1.683 Vrms 1.703 Vrms 1.717 Vrms 0.0017 V P

SIN 1.4 Vrms 1.386 Vrms 1.402 Vrms 1.414 Vrms 0.0016 V P

SIN 1.1 Vrms 1.089 Vrms 1.102 Vrms 1.111 Vrms 0.0014 V P

SIN 0.88 Vrms 0.8712 Vrms 0.8813 Vrms 0.8888 Vrms 0.00066 V P

SIN 0.70 Vrms 0.6930 Vrms 0.7011 Vrms 0.7070 Vrms 0.00030 V P

SIN 0.55 Vrms 0.5445 Vrms 0.5509 Vrms 0.5555 Vrms 0.00063 V P

SIN 0.44 Vrms 0.4356 Vrms 0.4406 Vrms 0.4444 Vrms 0.00062 V P

Page 7: Certificate of Calibration Certificate of Calibration Page ... · Conformance with the requirements of ISO/IEC 17025:2005 for testing and calibration laboratories is proven by the

Measurement Report Certificate Number: 1-1234567890-1

Page 7 of 12

Model Number: 33120A Serial Number: US12345678Product Firmware Revision: NA Options Installed:

Date of Calibration: 2 Mar 2011Condition: As Received

AC/DC VOLTS (cont.)TEST

CONDITIONSMINIMUM MEASURED MAXIMUM UNCERT.

SIN 0.35 Vrms 0.3465 Vrms 0.3506 Vrms 0.3535 Vrms 0.00020 V P

SIN 0.28 Vrms 0.2772 Vrms 0.2805 Vrms 0.2828 Vrms 0.00061 V P

SIN 0.22 Vrms 0.2178 Vrms 0.2203 Vrms 0.2222 Vrms 0.00020 V P

SIN 0.17 Vrms 0.1683 Vrms 0.1702 Vrms 0.1717 Vrms 0.00016 V P

SIN 0.14 Vrms 0.1386 Vrms 0.1402 Vrms 0.1414 Vrms 0.00060 V P

SIN 0.11 Vrms 0.1089 Vrms 0.1102 Vrms 0.1111 Vrms 0.00060 V P

SIN 0.088 Vrms 0.08712 Vrms 0.08818 Vrms 0.08888 Vrms 0.000032 V P

SIN 0.070 Vrms 0.06930 Vrms 0.07014 Vrms 0.07070 Vrms 0.000064 V P

SIN 0.055 Vrms 0.05445 Vrms 0.05511 Vrms 0.05555 Vrms 0.000024 V P

SIN 0.044 Vrms 0.04356 Vrms 0.04408 Vrms 0.04444 Vrms 0.000022 V P

SIN 0.036 Vrms 0.03564 Vrms 0.03607 Vrms 0.03636 Vrms 0.000061 V P

50 Ohm Load SIN 3.5 Vrms 3.465 Vrms 3.506 Vrms 3.535 Vrms 0.0046 V P

SIN 2.8 Vrms 2.772 Vrms 2.806 Vrms 2.828 Vrms 0.0037 V P

SIN 2.2 Vrms 2.178 Vrms 2.205 Vrms 2.222 Vrms 0.0031 V P

SIN 1.7 Vrms 1.683 Vrms 1.703 Vrms 1.717 Vrms 0.0026 V P

SIN 1.4 Vrms 1.386 Vrms 1.403 Vrms 1.414 Vrms 0.0023 V P

SIN 1.1 Vrms 1.089 Vrms 1.103 Vrms 1.111 Vrms 0.0020 V P

SIN 0.88 Vrms 0.8712 Vrms 0.8816 Vrms 0.8888 Vrms 0.0011 V P

SIN 0.70 Vrms 0.6930 Vrms 0.7013 Vrms 0.7070 Vrms 0.00086 V P

SIN 0.55 Vrms 0.5445 Vrms 0.5511 Vrms 0.5555 Vrms 0.00068 V P

SIN 0.44 Vrms 0.4356 Vrms 0.4408 Vrms 0.4444 Vrms 0.00056 V P

SIN 0.35 Vrms 0.3465 Vrms 0.3506 Vrms 0.3535 Vrms 0.00045 V P

SIN 0.28 Vrms 0.2772 Vrms 0.2805 Vrms 0.2828 Vrms 0.00037 V P

SIN 0.22 Vrms 0.2178 Vrms 0.2204 Vrms 0.2222 Vrms 0.00031 V P

SIN 0.17 Vrms 0.1683 Vrms 0.1704 Vrms 0.1717 Vrms 0.00026 V P

SIN 0.14 Vrms 0.1386 Vrms 0.1403 Vrms 0.1414 Vrms 0.00023 V P

SIN 0.11 Vrms 0.1089 Vrms 0.1103 Vrms 0.1111 Vrms 0.00020 V P

SIN 0.088 Vrms 0.08712 Vrms 0.08822 Vrms 0.08888 Vrms 0.00011 V P

SIN 0.070 Vrms 0.06930 Vrms 0.07018 Vrms 0.07070 Vrms 0.000086 V P

SIN 0.055 Vrms 0.05445 Vrms 0.05514 Vrms 0.05555 Vrms 0.000068 V P

SIN 0.044 Vrms 0.04356 Vrms 0.04411 Vrms 0.04444 Vrms 0.000056 V P

SIN 0.035 Vrms 0.03465 Vrms 0.03509 Vrms 0.03535 Vrms 0.000045 V P

SIN 0.028 Vrms 0.02772 Vrms 0.02807 Vrms 0.02828 Vrms 0.000037 V P

SIN 0.022 Vrms 0.02178 Vrms 0.02205 Vrms 0.02222 Vrms 0.000031 V P

SIN 0.018 Vrms 0.01782 Vrms 0.01805 Vrms 0.01818 Vrms 0.000027 V P

AMPLITUDE FLATNESS PASSEDReference condition: Sinewave, 1 kHz at 3.0 Vrms

TESTCONDITIONS

MINIMUM MEASURED MAXIMUM UNCERT.

100 kHz -0.030 Vrms 0.002 Vrms 0.030 Vrms 0.0060 V P

500 kHz -0.045 Vrms 0.000 Vrms 0.045 Vrms 0.0070 V P

1 MHz -0.045 Vrms 0.000 Vrms 0.045 Vrms 0.0060 V P

3 MHz -0.060 Vrms 0.004 Vrms 0.060 Vrms 0.0070 V P

5 MHz -0.060 Vrms 0.005 Vrms 0.060 Vrms 0.0070 V P

7 MHz -0.060 Vrms 0.004 Vrms 0.060 Vrms 0.0070 V P

9 MHz -0.060 Vrms 0.002 Vrms 0.060 Vrms 0.0070 V P

11 MHz -0.060 Vrms 0.003 Vrms 0.060 Vrms 0.0070 V P

13 MHz -0.060 Vrms 0.003 Vrms 0.060 Vrms 0.0070 V P

Page 8: Certificate of Calibration Certificate of Calibration Page ... · Conformance with the requirements of ISO/IEC 17025:2005 for testing and calibration laboratories is proven by the

Measurement Report Certificate Number: 1-1234567890-1

Page 8 of 12

Model Number: 33120A Serial Number: US12345678Product Firmware Revision: NA Options Installed:

Date of Calibration: 2 Mar 2011Condition: As Received

AMPLITUDE FLATNESS (cont.)TEST

CONDITIONSMINIMUM MEASURED MAXIMUM UNCERT.

15 MHz -0.060 Vrms 0.006 Vrms 0.060 Vrms 0.0080 V P

Page 9: Certificate of Calibration Certificate of Calibration Page ... · Conformance with the requirements of ISO/IEC 17025:2005 for testing and calibration laboratories is proven by the

Measurement Report Certificate Number: 1-1234567890-1

Page 9 of 12

Model Number: 33120A Serial Number: US12345678Product Firmware Revision: NA Options Installed:

Date of Calibration: 2 Mar 2011Condition: As Completed

SELF TEST PASSED

TESTCONDITIONS

STAT

SELF TEST +0

FREQUENCY PASSED

TEST CONDITIONS MINIMUM MEASURED MAXIMUM UNCERT. 1 kHz SINE WAVE 999.980 Hz 1000.000 Hz 1000.020 Hz 0.0050 Hz P

BURST MODE, 500 Hz 495.000 Hz 500.189 Hz 505.000 Hz 0.0030 Hz P

HARMONICS_LOW PASSED

TEST CONDITIONS MEASURED MAXIMUM UNCERT. REFERENCE 20 kHz, 1.1 Vrms -16.3 dBm

2nd Harmonic, 40 kHz -74.1 dBc -70.0 dBc 1.2 dB P

3rd Harmonic, 60 kHz -83.2 dBc -70.0 dBc 1.4 dB P

4th Harmonic, 80 kHz -88.6 dBc -70.0 dBc 1.8 dB P

5th Harmonic, 100 kHz -95.0 dBc -70.0 dBc 1.2 dB P

REFERENCE 100 kHz, 1.1 Vrms -16.3 dBm

2nd Harmonic, 200 kHz -74.7 dBc -60.0 dBc 1.5 dB P

3rd Harmonic, 300 kHz -73.2 dBc -60.0 dBc 1.2 dB P

4th Harmonic, 400 kHz -85.7 dBc -60.0 dBc 1.5 dB P

5th Harmonic, 500 kHz -98.1 dBc -60.0 dBc 1.7 dB P

REFERENCE 1 MHz, 1.1 Vrms -16.3 dBm

2nd Harmonic, 2 MHz -75.5 dBc -45.0 dBc 1.1 dB P

3rd Harmonic, 3 MHz -60.7 dBc -45.0 dBc 1.0 dB P

4th Harmonic, 4 MHz -83.3 dBc -45.0 dBc 1.0 dB P

5th Harmonic, 5 MHz -80.8 dBc -45.0 dBc 1.0 dB P

HARMONICS PASSED

TEST CONDITIONS MEASURED MAXIMUM UNCERT. REFERENCE 15 MHz, 1.1 Vrms -16.3 dBm

2nd Harmonic, 30 MHz -48.8 dBc -35.0 dBc 1.7 dB P

3rd Harmonic, 45 MHz -52.3 dBc -35.0 dBc 1.7 dB P

4th Harmonic, 60 MHz -59.2 dBc -35.0 dBc 1.7 dB P

5th Harmonic, 75 MHz -68.7 dBc -35.0 dBc 1.7 dB P

Page 10: Certificate of Calibration Certificate of Calibration Page ... · Conformance with the requirements of ISO/IEC 17025:2005 for testing and calibration laboratories is proven by the

Measurement Report Certificate Number: 1-1234567890-1

Page 10 of 12

Model Number: 33120A Serial Number: US12345678Product Firmware Revision: NA Options Installed:

Date of Calibration: 2 Mar 2011Condition: As Completed

SQ DUTY CYCLE PASSEDSquare Wave, 1.0 Vrms at 300 Hz

TESTCONDITIONS

MINIMUM MEASURED MAXIMUM UNCERT.

50% DUTY CYCLE

-0.010 Vdc -0.001 Vdc 0.010 Vdc 0.0030 V P

25% DUTY CYCLE

-0.510 Vdc -0.503 Vdc -0.490 Vdc 0.0030 V P

75% DUTY CYCLE

0.490 Vdc 0.498 Vdc 0.510 Vdc 0.0020 V P

AC/DC VOLTS PASSED

TESTCONDITIONS

MINIMUM MEASURED MAXIMUM UNCERT.

Function Gain and Linearity Verification SIN 7.0 Vrms 6.930 Vrms 6.997 Vrms 7.070 Vrms 0.0028 V P

SIN 5.7 Vrms 5.643 Vrms 5.698 Vrms 5.757 Vrms 0.0025 V P

TRI 5.7 Vrms 5.643 Vrms 5.698 Vrms 5.757 Vrms 0.0027 V P

RAMP 5.7 Vrms 5.643 Vrms 5.699 Vrms 5.757 Vrms 0.0026 V P

SQU 10.0 Vrms 9.900 Vrms 9.997 Vrms 10.100 Vrms 0.0035 V P

SQU 8.0 Vrms 7.920 Vrms 7.999 Vrms 8.080 Vrms 0.0031 V P

DC Function Offset Verification DC 10 V 9.800 Vdc 10.000 Vdc 10.200 Vdc 0.59 mV P

DC -10 V -10.200 Vdc -9.999 Vdc -9.800 Vdc 0.59 mV P

AM Modulation Depth Verification AM MOD, 0% 0.495 Vrms 0.499 Vrms 0.505 Vrms 0.62 mV P

AM MOD 100% 0.6039 Vrms 0.6117 Vrms 0.6161 Vrms 0.00027 V P

AC Amplitude Verification High Impedance Load SIN 7.0 Vrms 6.930 Vrms 6.998 Vrms 7.070 Vrms 0.0028 V P

SIN 5.7 Vrms 5.643 Vrms 5.698 Vrms 5.757 Vrms 0.0026 V P

SIN 5.5 Vrms 5.445 Vrms 5.499 Vrms 5.555 Vrms 0.0024 V P

SIN 4.4 Vrms 4.356 Vrms 4.399 Vrms 4.444 Vrms 0.0023 V P

SIN 3.5 Vrms 3.465 Vrms 3.499 Vrms 3.535 Vrms 0.0020 V P

SIN 2.8 Vrms 2.772 Vrms 2.800 Vrms 2.828 Vrms 0.0019 V P

SIN 2.2 Vrms 2.178 Vrms 2.200 Vrms 2.222 Vrms 0.0017 V P

SIN 1.7 Vrms 1.683 Vrms 1.700 Vrms 1.717 Vrms 0.0017 V P

SIN 1.4 Vrms 1.386 Vrms 1.400 Vrms 1.414 Vrms 0.0016 V P

SIN 1.1 Vrms 1.089 Vrms 1.100 Vrms 1.111 Vrms 0.0014 V P

SIN 0.88 Vrms 0.8712 Vrms 0.8798 Vrms 0.8888 Vrms 0.00066 V P

SIN 0.70 Vrms 0.6930 Vrms 0.6999 Vrms 0.7070 Vrms 0.00030 V P

SIN 0.55 Vrms 0.5445 Vrms 0.5499 Vrms 0.5555 Vrms 0.00063 V P

SIN 0.44 Vrms 0.4356 Vrms 0.4399 Vrms 0.4444 Vrms 0.00062 V P

Page 11: Certificate of Calibration Certificate of Calibration Page ... · Conformance with the requirements of ISO/IEC 17025:2005 for testing and calibration laboratories is proven by the

Measurement Report Certificate Number: 1-1234567890-1

Page 11 of 12

Model Number: 33120A Serial Number: US12345678Product Firmware Revision: NA Options Installed:

Date of Calibration: 2 Mar 2011Condition: As Completed

AC/DC VOLTS (cont.)TEST

CONDITIONSMINIMUM MEASURED MAXIMUM UNCERT.

SIN 0.35 Vrms 0.3465 Vrms 0.3499 Vrms 0.3535 Vrms 0.00020 V P

SIN 0.28 Vrms 0.2772 Vrms 0.2800 Vrms 0.2828 Vrms 0.00061 V P

SIN 0.22 Vrms 0.2178 Vrms 0.2199 Vrms 0.2222 Vrms 0.00020 V P

SIN 0.17 Vrms 0.1683 Vrms 0.1700 Vrms 0.1717 Vrms 0.00016 V P

SIN 0.14 Vrms 0.1386 Vrms 0.1400 Vrms 0.1414 Vrms 0.00060 V P

SIN 0.11 Vrms 0.1089 Vrms 0.1100 Vrms 0.1111 Vrms 0.00060 V P

SIN 0.088 Vrms 0.08712 Vrms 0.08802 Vrms 0.08888 Vrms 0.000032 V P

SIN 0.070 Vrms 0.06930 Vrms 0.07001 Vrms 0.07070 Vrms 0.000064 V P

SIN 0.055 Vrms 0.05445 Vrms 0.05501 Vrms 0.05555 Vrms 0.000024 V P

SIN 0.044 Vrms 0.04356 Vrms 0.04400 Vrms 0.04444 Vrms 0.000022 V P

SIN 0.036 Vrms 0.03564 Vrms 0.03600 Vrms 0.03636 Vrms 0.000061 V P

50 Ohm Load SIN 3.5 Vrms 3.465 Vrms 3.500 Vrms 3.535 Vrms 0.0046 V P

SIN 2.8 Vrms 2.772 Vrms 2.800 Vrms 2.828 Vrms 0.0037 V P

SIN 2.2 Vrms 2.178 Vrms 2.201 Vrms 2.222 Vrms 0.0031 V P

SIN 1.7 Vrms 1.683 Vrms 1.700 Vrms 1.717 Vrms 0.0026 V P

SIN 1.4 Vrms 1.386 Vrms 1.400 Vrms 1.414 Vrms 0.0023 V P

SIN 1.1 Vrms 1.089 Vrms 1.101 Vrms 1.111 Vrms 0.0020 V P

SIN 0.88 Vrms 0.8712 Vrms 0.8801 Vrms 0.8888 Vrms 0.0011 V P

SIN 0.70 Vrms 0.6930 Vrms 0.7001 Vrms 0.7070 Vrms 0.00086 V P

SIN 0.55 Vrms 0.5445 Vrms 0.5501 Vrms 0.5555 Vrms 0.00068 V P

SIN 0.44 Vrms 0.4356 Vrms 0.4400 Vrms 0.4444 Vrms 0.00056 V P

SIN 0.35 Vrms 0.3465 Vrms 0.3500 Vrms 0.3535 Vrms 0.00045 V P

SIN 0.28 Vrms 0.2772 Vrms 0.2800 Vrms 0.2828 Vrms 0.00037 V P

SIN 0.22 Vrms 0.2178 Vrms 0.2200 Vrms 0.2222 Vrms 0.00031 V P

SIN 0.17 Vrms 0.1683 Vrms 0.1700 Vrms 0.1717 Vrms 0.00026 V P

SIN 0.14 Vrms 0.1386 Vrms 0.1400 Vrms 0.1414 Vrms 0.00023 V P

SIN 0.11 Vrms 0.1089 Vrms 0.1100 Vrms 0.1111 Vrms 0.00020 V P

SIN 0.088 Vrms 0.08712 Vrms 0.08802 Vrms 0.08888 Vrms 0.00011 V P

SIN 0.070 Vrms 0.06930 Vrms 0.07002 Vrms 0.07070 Vrms 0.000086 V P

SIN 0.055 Vrms 0.05445 Vrms 0.05501 Vrms 0.05555 Vrms 0.000068 V P

SIN 0.044 Vrms 0.04356 Vrms 0.04401 Vrms 0.04444 Vrms 0.000056 V P

SIN 0.035 Vrms 0.03465 Vrms 0.03501 Vrms 0.03535 Vrms 0.000045 V P

SIN 0.028 Vrms 0.02772 Vrms 0.02800 Vrms 0.02828 Vrms 0.000037 V P

SIN 0.022 Vrms 0.02178 Vrms 0.02200 Vrms 0.02222 Vrms 0.000031 V P

SIN 0.018 Vrms 0.01782 Vrms 0.01800 Vrms 0.01818 Vrms 0.000027 V P

AMPLITUDE FLATNESS PASSEDReference condition: Sinewave, 1 kHz at 3.0 Vrms

TESTCONDITIONS

MINIMUM MEASURED MAXIMUM UNCERT.

100 kHz -0.030 Vrms 0.002 Vrms 0.030 Vrms 0.0060 V P

500 kHz -0.045 Vrms -0.001 Vrms 0.045 Vrms 0.0070 V P

1 MHz -0.045 Vrms -0.001 Vrms 0.045 Vrms 0.0060 V P

3 MHz -0.060 Vrms 0.001 Vrms 0.060 Vrms 0.0070 V P

5 MHz -0.060 Vrms 0.000 Vrms 0.060 Vrms 0.0070 V P

7 MHz -0.060 Vrms -0.002 Vrms 0.060 Vrms 0.0070 V P

9 MHz -0.060 Vrms -0.004 Vrms 0.060 Vrms 0.0070 V P

11 MHz -0.060 Vrms -0.005 Vrms 0.060 Vrms 0.0070 V P

13 MHz -0.060 Vrms -0.005 Vrms 0.060 Vrms 0.0070 V P

Page 12: Certificate of Calibration Certificate of Calibration Page ... · Conformance with the requirements of ISO/IEC 17025:2005 for testing and calibration laboratories is proven by the

Measurement Report Certificate Number: 1-1234567890-1

Page 12 of 12

Model Number: 33120A Serial Number: US12345678Product Firmware Revision: NA Options Installed:

Date of Calibration: 2 Mar 2011Condition: As Completed

AMPLITUDE FLATNESS (cont.)TEST

CONDITIONSMINIMUM MEASURED MAXIMUM UNCERT.

15 MHz -0.060 Vrms -0.006 Vrms 0.060 Vrms 0.0080 V P