This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2017. This accreditation demonstrates technical competence for a defined scope and the operation of a laboratory quality management system (refer to joint ISO-ILAC-IAF Communiqué dated April 2017). CERTIFICATE OF ACCREDITATION ANSI National Accreditation Board 11617 Coldwater Road, Fort Wayne, IN 46845 USA This is to certify that Keysight Technologies Taiwan LTD. Calibration Laboratory No. 20, Kao-Shuang Road Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard ISO/IEC 17025:2017 and national standards ANSI/NCSL Z540-1-1994 (R2002) and ANSI/NCSL Z540.3-2006 (R2013) while demonstrating technical competence in the field of CALIBRATION Refer to the accompanying Scope of Accreditation for information regarding the types of activities to which this accreditation applies AC-1928.01 Certificate Number Certificate Valid Through: 03/11/2022 Version No. 006 Issued: 02/27/2020

CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2017. This accreditation demonstrates technical competence for a defined scope and the operation of a laboratory

quality management system (refer to joint ISO-ILAC-IAF Communiqué dated April 2017).

CERTIFICATE OF ACCREDITATION ANSI National Accreditation Board

11617 Coldwater Road, Fort Wayne, IN 46845 USA

This is to certify that

Keysight Technologies Taiwan LTD. Calibration Laboratory No. 20, Kao-Shuang Road

Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

ISO/IEC 17025:2017 and national standards

ANSI/NCSL Z540-1-1994 (R2002) and ANSI/NCSL Z540.3-2006 (R2013)

while demonstrating technical competence in the field of

CALIBRATION Refer to the accompanying Scope of Accreditation for information regarding the types of

activities to which this accreditation applies

AC-1928.01 Certificate Number

Certificate Valid Through: 03/11/2022

Version No. 006 Issued: 02/27/2020

Page 2: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 1 of 29

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017, ANSI/NCSL Z540-1-1994 (R2002) AND ANSI/NCSL Z540.3-2006 (R2013)

Keysight Technologies Taiwan LTD. Calibration Laboratory

No. 20, Kao-Shuang Road Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C.

Freddy Hsu Phone: +886 3 4959389 [email protected] www.keysight.com

CALIBRATION

Valid to: March 11, 2022 Certificate Number: AC-1928.01

Electrical – RF/Microwave

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

Signal Analyzers – Relative RF Power

in Coaxial Line

(0 to 70) dB (70 to 110) dB

Max power level: 10 dBm 50 MHz to 2 GHz

0.013 dB 0.04 dB

Signal Source E8257D, Step Attenuators

Direct measurement

Signal Analyzers – Phase Noise

Carrier frequency: 1 GHz Offset frequency: 100 Hz to 10 MHz

0.36 dB

Signal Source 500-13438

Source phase noise: -107 to -167 dBc/Hz Direct measurement

Power Sensor – Calibration Factors

9 kHz to < 100 MHz 100 MHz

(>0.1 to 4.2) GHz (>4.2 to 8) GHz (>8 to 17) GHz

(>17 to 26.5) GHz (>26.5 to 33) GHz (>33 to 37) GHz (>37 to 40) GHz (>40 to 44) GHz (>44 to 50) GHz

0.7 % cal factor 0.6 % cal factor 0.7 % cal factor 1 % cal factor

1.1 % cal factor 1.5 % cal factor 1.6 % cal factor 2.4 % cal factor 2.3 % cal factor 3.5 % cal factor 3.3 % cal factor

Power Sensor 8481A, 8482A, 8485A, 8487A and

E9304A

Passive Devices Scattering Parameter: Reflection

coefficient (linear) 9 kHz to 50 GHz (See Matrix A)

Measure using E5071C and N5230C Network Analyzer

Direct Measurement

Page 3: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 2 of 29

Electrical – RF/Microwave

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

Passive Devices Scattering Parameter: Transmission

coefficient 9 kHz to 50 GHz (See Matrix B)

Measure using E5071C, N5230C, N5245A, and

E8363C Network Analyzer Direct Measurement

Signal Sources – Absolute RF Power in

Coaxial 9 kHz to 50 GHz (See Matrix C)

Direct measurement Power Sensor Power Meter

Absolute RF Power in Coaxial Line 9 kHz to 50 GHz (See Matrix D)

Signal Sources, Power Splitters, Power Meters,

Power Sensors Direct measurement

Signal Sources – Harmonic Content

Fundamental Frequency 1 MHz to 25 GHz

(0 to 10) dBm Harmonic Frequency

2 MHz to 50 GHz (-110 to 0) dBm

0.51 dB

yzer E4448A Direct measurement

Signal Sources – Pulse Time Parameters

Rise/Fall Time

(0 to 10) dBm 10 MHz to 50 GHz

18 ps to 10 ns 86 ps

Oscilloscope mainframe 86100C, Electrical module

86117A. Direct measurement

Signal Sources –Modulation

Frequency Modulation

Peak Frequency Deviation

(-18 to 30) dBm fc: 250 kHz to 10 MHz

fm: 20 Hz to 10 kHz Δf: 200 Hz to 40 kHz

β> 0.2 fc: 250 kHz to 10 MHz

fm: 20 Hz to 10 kHz Δf: 200 Hz to 40 kHz

β> 1.2 fc: 10 MHz to 6.6 GHz fm: 50 Hz to 200 kHz Δf: 250 Hz to 400 kHz

β> 0.2 fc: 10 MHz to 6.6 GHz fm: 50 Hz to 200 kHz Δf: 250 Hz to 400 kHz

β> 0.45

0.015 Hz/Hz

0.01 Hz/Hz

0.015 Hz/Hz

0.01 Hz/Hz

Signal analyzer E4448A fc = Carrier Frequency fm = Modulation Rate Δf=Peak Deviation

β = Δf / fm Direct measurement

Page 4: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 3 of 29

Electrical – RF/Microwave

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

Signal Sources –Modulation

Frequency Modulation

Peak Frequency Deviation

(-18 to 30) dBm fc: 6.6 GHz to 13.2 GHz

fm: 50 Hz to 200 kHz Δf: 250 Hz to 400 kHz

β> 0.2

fc: 6.6 GHz to 13.2 GHz fm: 50 Hz to 200 kHz

Δf: 250 Hz to 400 kHz β> 8

fc: (13.2 to 31.15) GHz fm: 50 Hz to 200 kHz

Δf: 250 Hz to 400 kHz β> 0.2

fc: (13.2 to 31.15) GHz fm: 50 Hz to 200 kHz

Δf: 250 Hz to 400 kHz β> 16

fc: (31.15 to 50) GHz fm: 50 Hz to 200 kHz

Δf: 250 Hz to 400 kHz β> 0.2

fc: (31.15 to 50) GHz fm: 50 Hz to 200 kHz

Δf: 250 Hz to 400 kHz β> 32

0.025 Hz/Hz

0.01 Hz/Hz

0.038 Hz/Hz

0.01 Hz/Hz

0.85 Hz/Hz

0.01 Hz/Hz

Signal analyzer E4448A fc = Carrier Frequency fm = Modulation Rate Δf=Peak Deviation

β = Δf / fm Direct measurement

Page 5: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 4 of 29

Electrical – RF/Microwave

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

Signal Sources –

Amplitude Modulation

Depth

-18 to 30 dBm fm: 50 Hz to 100 kHz

fc: 100 kHz to 10 MHz

Depth: 5 % to 99 %

fc: 10 MHz to 3 GHz Depth: 20 % to 99 %

fc: 10 MHz to 3 GHz Depth: 5 % to 20 %

fc: (3 to 26.5) GHz

Depth: 20 % to 99 %

-18 to 30 dBm fm: 50 Hz to 100 kHz

fc: (3 to 26.5) GHz Depth: 5 % to 20 %

fc: (26.5 to 31.15) GHz Depth: 20 % to 99 %

fc: (26.5 to 31.15) GHz

Depth: 5 % to 20 %

fc: (31.15 to 50) GHz Depth: 5 % to 20 %

fc: (31.15 to 50) GHz Depth: 5 % to 20 %

0.007 5 %/% Depth

0.005 %/% Depth

0.025 %/% Depth

0.015 %/% Depth

0.045 %/% Depth

0.019 %/% Depth

0.068 %/% Depth

0.06 %/% Depth

0.26 %/% Depth

Signal analyzer E4448A fc = Carrier Frequency fm = Modulation Rate Direct measurement

Page 6: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 5 of 29

Electrical – RF/Microwave

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

Phase Modulation

Peak phase deviation

-18 to 30 dBm fc: 100 kHz to 6.6 GHz

ΔΦ > 0.7 rad

fc: 100 kHz to 6.6 GHz ΔΦ > 0.3 rad

fc: (6.6 to 13.2) GHz

ΔΦ > 2.0 rad

fc: (6.6 to 13.2) GHz ΔΦ > 0.6 rad

fc: (13.2 to 26.5) GHz

ΔΦ > 4.0 rad

fc: (13.2 to 26.5) GHz ΔΦ > 1.2 rad

fc: (26.5 to 31.15) GHz

ΔΦ > 4.0 rad

-18 to 30 dBm

fc: (26.5 to 31.15) GHz ΔΦ > 1.3 rad

fc: (31.15 to 50) GHz

ΔΦ > 8 rad

fc: (31.15 to 50) GHz ΔΦ > 2.4 rad

0.01 rad/rad

0.03 rad/rad

0.01 rad/rad

0.03 rad/rad

0.01 rad/rad

0.03 rad/rad

0.01 rad/rad

0.03 rad/rad

0.01 rad/rad

0.01 rad/rad

Signal analyzer E4448A fc = Carrier Frequency ΔΦ = Phase Deviation

Direct measurement

Page 7: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 6 of 29

Electrical – RF/Microwave

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment Phase Modulation

Modulation Rate

Amplitude Modulation Rate

Frequency Modulation Rate

Phase Modulation Rate

-18 to 30 dBm 100 kHz ≤ fc< 50 GHz

Depth ≥ 20% fm ≤ 100 kHz

β ≥ 0.01

fm ≤ 200 kHz

β ≥ 0.01 fm ≤ 20 kHz

0.062 Hz

0.062 Hz

0.062 Hz

Signal analyzer E4448A β = Δf / fm

fm = Modulation Rate Direct measurement

Signal Sources – Modulation Distortion

Amplitude Modulation

Distortion

0.01% to 100%

fm: 20 Hz to 1 kHz fc: 100 kHz to 10 MHz

Depth: > 1 % Depth: > 3 %

fm: 20 Hz to 1 kHz

fc: 10 MHz to 26.5 GHz Depth: > 1 % Depth: > 3 %

fm: 20 Hz to 1 kHz

fc: (26.5 to 50) GHz Depth: > 1 % Depth: > 3 % Depth: > 5 %

0.8 % Depth 0.3 % Depth

1 % Depth 0.4 % Depth

6.2 % Depth 2 % Depth

1.5 % Depth

Signal analyzer E4448A fc =Carrier Frequency fm = Modulation Rate Direct measurement

Signal Sources –

Phase Modulation Distortion

0.01% to 100%

fc: 1 MHz to 6.6 GHz fm: (20 to 500) Hz

ΔΦ > 0.8 rad ΔΦ ≥ 2.5 rad

fc: 1 MHz to 6.6 GHz fm: 500 Hz to 1 kHz

ΔΦ> 0.4 rad ΔΦ ≥ 1.0 rad

0.3 % of reading 0.1 % of reading

0.3 % of reading 0.1 % of reading

Signal analyzer E4448A fc =Carrier Frequency fm = Modulation Rate Direct measurement

Page 8: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 7 of 29

Electrical – RF/Microwave

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

Signal Sources –

Phase Modulation Distortion

0.01% to 100%

fc: (6.6 to 13.2) GHz fm: (20 to 500) Hz

ΔΦ > 1.8 rad ΔΦ ≥ 5.5 rad

fc: (13.2 to 31.15) GHz

fm: (20 to 500) Hz ΔΦ > 3.5 rad ΔΦ ≥ 10.0 rad

0.01% to 100%

fc: (13.2 to 31.15) GHz

fm: 500 Hz to 1 kHz ΔΦ > 1.2 rad ΔΦ ≥ 4.0 rad

fc: (31.15 to 50) GHz

fm: (20 to 500) Hz ΔΦ > 7.5 rad ΔΦ ≥ 19.0 rad

fc: (31.15 to 50) GHz fm: 500 Hz to 1 kHz

ΔΦ > 3.0 rad ΔΦ ≥ 8.0 rad

0.3 % of reading 0.1 % of reading

0.3 % of reading 0.1 % of reading

0.3 % of reading 0.1 % of reading

0.3 % of reading 0.1 % of reading

0.3 % of reading 0.1 % of reading

Signal analyzer E4448A fc =Carrier Frequency fm = Modulation Rate Direct measurement

PARAMETER/ EQUIPMENT MODULATION TYPES FREQUENCY RANGE

Digital Modulation RF Quality1

Measure – Carrier 2 MHz to 44 GHz

TETRA, PDC, NADC, PHS, EDGE, CDMA 2000A/C, WCDMA, 3GPP, QPSK, BPSK, PI/4

DQPSK, 16QAM, 256QAM, DECT, PHP, GSM, 2FSK, 4FSK, GMSK, MSK, DQPSK, 8PSK,

32QAM FSK

2 MHz to 2.65 GHz using the VSA directly (2.65 to 44) GHz.

The digitally modulated RF signal needs to be down-converted with an external Mixer

and a Local Oscillator L.O. center frequency = 150 MHz

Page 9: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 8 of 29

Electrical – RF/Microwave

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

Signal Sources –

Digital Modulation Carrier: 2 MHz to 2.65 GHz

Error Vector Magnitude

(EVM)

Phase Error

Frequency Error

Mod Frequency Span: f ≤ 100kHz

100kHz ≤ f ≤ 1MHz f > 1MHz

Up to 180 °

Mod Frequency Span: f ≤ 100kHz

100kHz ≤ f ≤ 1MHz f > 1MHz

Mod Frequency

1 GHz 2 GHz 3 GHz 4 GHz 5 GHz 6 GHz

0.43 % of reading 0.48 % of reading 0.82 % of reading

0.17 º rms 0.34 º rms 0.57 º rms

0.063 % of reading 0.068 % of reading 0.079 % of reading 0.099 % of reading 0.33 % of reading 0.39 % of reading

HP 89441A Vector Signal Analyzer

Modulation Types:

TETRA, PDC, NADC, PHS, EDGE, CDMA

2000A/C, WCDMA, 3GPP, QPSK, BPSK, PI/4 DQPSK, 16QAM,

256QAM, DECT, PHP, GSM,

2FSK, 4FSK, GMSK, MSK, DQPSK, 8PSK,

32QAM FSK

Frequency Range: 2 MHz to 2.65 GHz using

the VSA directly (2.65 to 44)

GHz. The digitally modulated RF

signal needs to be down-converted with an

external Mixer and a Local Oscillator

L.O. center frequency = (RF-150 MHz)

Page 10: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 9 of 29

Electrical – RF/Microwave

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

Modulation Accuracy (Rho)

Mod Frequency Span: f ≤ 100kHz

0.9999 ≤ ρ ≤ 1 0.9975 ≤ρ < 0.9999 0.9936 ≤ρ < 0.9975 0.99 ≤ρ < 0.9936 0.978 ≤ρ < 0.99 0.96 ≤ρ < 0.978

Mod Frequency Span: 100 kHz ≤ f ≤ 1 MHz

0.9999 ≤ρ ≤ 1 0.9975 ≤ρ < 0.9999 0.9936 ≤ρ 0.9975 0.99 ≤ρ < 0.9936 0.978 ≤ρ < 0.99 0.96 ≤ρ < 0.978

Mod Frequency Span:

f > 1MHz 0.9999 ≤ρ ≤ 1

0.9975 ≤ρ < 0.9999 0.9936 ≤ρ < 0.9975 0.99 ≤ρ < 0.9936 0.978 ≤ρ < 0.99 0.96 ≤ρ < 0.978

8.6E-5 ρ 0.000 43 ρ 0.000 68 ρ 0.000 84 ρ 0.001 2 ρ 0.001 6 ρ

9.6E-5 ρ 0.000 48 ρ 0.000 76 ρ 0.000 94 ρ 0.001 4 ρ 0.001 8 ρ

1.6E-4 ρ 0.000 82 ρ 0.001 3 ρ 0.001 6 ρ 0.002 4 ρ 0.003 ρ

HP 89441A Vector Signal Analyzer

Modulation Types:

TETRA, PDC, NADC, PHS, EDGE, CDMA

2000A/C, WCDMA, 3GPP, QPSK, BPSK, PI/4 DQPSK, 16QAM,

256QAM, DECT, PHP, GSM,

2FSK, 4FSK, GMSK, MSK, DQPSK, 8PSK,

32QAM FSK

Frequency Range: 2 MHz to 2.65 GHz using

the VSA directly (2.65 to 44)

GHz. The digitally modulated RF

signal needs to be down-converted with an

external Mixer and a Local Oscillator

L.O. center frequency = (RF-150 MHz)

Signal Sources –

Frequency Modulation Distortion

(0.01 to 100) % fc: (6.6 to 13.2) GHz fm: 20 Hz to 1 kHz

Δf> 2.3 kHz Δf ≥ 4.5 kHz

fc: (13.2 to 31.15) GHz fm: 20 Hz to 1 kHz

Δf> 2.7 kHz Δf ≥ 6.0 kHz

fc: (31.15 to 50) GHz fm: 20 Hz to 1 kHz

Δf> 4.0 kHz Δf ≥ 12.0 kHz

0.3 % of reading 0.1 % of reading

0.3 % of reading 0.1 % of reading

0.3 % of reading 0.1 % of reading

Signal analyzer E4448A fc =Carrier Frequency fm = Modulation Rate Direct measurement

Page 11: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 10 of 29

Electrical – RF/Microwave

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

Signal Sources –

Frequency Modulation Distortion

(0.01 to 100) % fc: 1 MHz to 6.6 GHz

fm: 20 Hz to 1 kHz Δf: 500 Hz to 2 kHz

Δf ≥ 2.0 kHz

0.3 % of reading 0.1 % of reading

Signal analyzer E4448A fc =Carrier Frequency fm = Modulation Rate Direct measurement

Phase Noise for Signal Sources

Offset Frequency ≤ 100 kHz ≤ 100 kHz ≤ 1 MHz ≤ 10 MHz

< 100 MHz

Offset Frequency ≤ 100 kHz ≤ 100 kHz ≤ 1 MHz ≤ 10 MHz

< 100 MHz

Offset Frequency ≤ 100 kHz ≤ 100 kHz ≤ 1 MHz ≤ 10 MHz

< 100 MHz

Offset Frequency ≤ 100 kHz ≤ 100 kHz ≤ 1 MHz ≤ 10 MHz

< 100 MHz

LREF - LDUT) ≥ 10dB ≤ 100 MHz

100 MHz < f ≤ 26.5 GHz 50 kHz < f ≤ 26.5 GHz 50 kHz < f ≤ 26.5 GHz 50 kHz < f ≤ 26.5 GHz

10dB > (LREF - LDUT) ≥

5dB ≤ 100 MHz

100 MHz < f ≤ 26.5 MHz 50 kHz < f ≤ 26.5 GHz 50 kHz < f ≤ 26.5 GHz 50 kHz < f ≤ 26.5 GHz

5dB > (LREF - LDUT) ≥

3dB ≤ 100 MHz

100 MHz < f ≤ 26.5G Hz 50 kHz < f ≤ 26.5 GHz 50 kHz < f ≤ 26.5 GHz 50 kHz < f ≤ 26.5 GHz

3dB > (LREF - LDUT) ≥

0dB ≤ 100 MHz

100 MHz < f ≤ 26.5 GHz 50 kHz < f ≤ 26.5 GHz 50 kHz < f ≤ 26.5 GHz 50 kHz < f ≤ 26.5 GHz

2.3 dB 2.3 dB 2.3 dB 4.6 dB 4.6 dB

2.8 dB 2.9 dB 2.9 dB 5.2 dB 5.3 dB

3.2 dB 3.3 dB 3.3 dB 5.4 dB 5.5 dB

4.3 dB 4.3 dB 4.3 dB 6.1 dB 6.2 dB

Keysight E5505A Phase Noise System

Page 12: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 11 of 29

Electrical – RF/Microwave

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment Phase Noise for Signal

Sources Offset Frequency

≤ 100 kHz ≤ 100 kHz ≤ 100 kHz ≤ 100 kHz ≤ 100 kHz ≤ 100 kHz ≤ 100 kHz ≤ 100 kHz ≤ 1 MHz ≤ 10 MHz

< 100 MHz

dB > (LREF - LDUT) ≥ 0

dB ≤ 100 MHz

100 MHz < f ≤ 255 MHz 255 MHz < f ≤ 600 MHz 600 MHz < f ≤ 1.8 GHz 1.8 GHz < f ≤ 3.2 GHz 3.2 GHz < f ≤ 10 GHz 10 GHz < f ≤ 20 GHz

20 GHz < f ≤ 26.5 GHz 50 kHz < f ≤ 26.5 GHz 50 kHz < f ≤ 26.5 GHz 50 kHz < f ≤ 26.5 GHz

4.3 dB 4.6 dB 4.6 dB 4.5 dB 4.5 dB 4.8 dB .8 dB 4.5 dB 4.7 dB 6.2 dB 6.2 dB

Keysight E5505A Phase Noise System

Power Sources Fitted with female Type-N

connectors 1 mW at 50 MHz 3.1 µW

Thermistor Mount 478A Direct measurement

(1 mW 50 MHz Reference Source Calibration)

Page 13: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 12 of 29

Matrix A Passive Devices Scattering Parameter: Reflection coefficient

Frequency (S11 - Reflection) Measured Magnitude (+/- dB)

< 0.1 > 0.1 to ≤ 0.2

> 0.2 to ≤ 0.3

> 0.3 to ≤ 0.4

> 0.4 to ≤ 0.5

> 0.5 to ≤ 0.6

> 0.6 to ≤ 0.7

> 0.7 to ≤ 0.8

> 0.8 to ≤ 0.9

9 kHz to 10 MHz 0.003 8 0.003 9 0.004 0.004 3 0.004 8 0.005 4 0.005 8 0.007 4 0.008 7 Using E5071C with 85032F Calibration

Kit

(0.01 to 3) GHz 0.005 2 0.005 3 0.005 4 0.005 7 0.006 1 0.006 8 0.007 2 0.008 8 0.01 (3 to 6) GHz 0.01 0.01 0.01 0.011 0.011 0.012 0.013 0.015 0.017

(6 to 8.5) GHz 0.013 0.013 0.013 0.013 0.014 0.015 0.015 0.018 0.02 (0.045 to1) GHz 0.004 2 0.004 3 0.004 4 0.004 6 0.004 8 0.005 1 0.005 4 0.006 2 0.006 9 Using N5230C,

E8364B with 85054B Calibration Kit

(1 to 6) GHz 0.004 2 0.004 3 0.004 4 0.004 6 0.004 8 0.005 2 0.005 4 0.006 2 0.006 9 (6 to 18) GHz 0.008 0.008 1 0.008 3 0.008 6 0.009 1 0.01 0.011 0.013 0.015

(0.045 to 6) GHz 0.004 2 0.004 2 0.004 4 0.004 7 0.005 2 0.005 9 0.006 3 0.008 0.009 5 Using N5230C, E8364B with 85052B

Calibration Kit (6 to 20) GHz 0.006 4 0.006 5 0.006 8 0.007 5 0.008 7 0.011 0.012 0.016 0.019

(0 to 26.5) GHz 0.006 5 0.006 7 0.007 2 0.008 2 0.009 9 0.012 0.014 0.019 0.024 (0.045 to 20) GHz 0.008 1 0.008 2 0.008 2 0.008 4 0.008 6 0.008 9 0.009 1 0.01 0.011 Using N5230C,

E8364B with 85056A Calibration Kit

(20 to 40) GHz 0.013 0.013 0.013 0.014 0.014 0.016 0.016 0.02 0.022 (40 to 50) GHz 0.016 0.016 0.017 0.017 0.018 0.019 0.02 0.024 0.028

Frequency (S11 - Reflection) Measured Phase (+/- Degree)

< 0.1 > 0.1 to ≤ 0.2

> 0.2 to ≤ 0.3

> 0.3 to ≤ 0.4

> 0.4 to ≤ 0.5

> 0.5 to ≤ 0.6

> 0.6 to ≤ 0.7

> 0.7 to ≤ 0.8

> 0.8 to ≤ 0.9

9 kHz to 10 MHz 2.1 1.1 0.75 0.61 0.54 0.51 0.51 0.53 0.55 Using E5071C with 85032F Calibration

Kit

(0.01 to 3) GHz 2.9 1.5 1 0.81 0.7 0.65 0.63 0.63 0.65 (3 to 6) GHz 5.8 2.9 2 1.5 1.3 1.1 1.1 1.1 1.1

(6 to 8.5) GHz 7.3 3.7 2.5 1.9 1.6 1.4 1.37 1.3 1.3 (0.045 to1) GHz 1.6 0.97 0.72 0.59 0.52 0.47 0.46 0.44 0.44 Using N5230C,

E8364B with 85054B Calibration Kit

(1 to 6) GHz 1.6 0.96 0.71 0.58 0.51 0.46 0.45 0.44 0.44

(6 to 18) GHz 3.1 1.9 1.4 1.1 0.99 0.93 0.92 0.94 0.97 (0.045 to 6) GHz 2.3 1.2 0.82 0.66 0.58 0.56 0.55 0.57 0.61 Using N5230C,

E8364B with 85052B Calibration Kit

(6 to 20) GHz 3.6 1.6 1.3 1.1 1 1 13 1.1 1.2 (0 to 26.5) GHz 3.7 1.9 1.4 1.2 1.1 1.2 1.2 1.4 1.5

(0.045 to 20) GHz 4.6 2.3 1.6 1.2 0.98 0.85 0.81 0.72 0.7 Using N5230C, E8364B with 85056A

Calibration Kit (20 to 40) GHz 7.4 3.7 2.5 1.9 1.6 1.5 1.4 1.4 1.4

(40 to 50) GHz 9.3 4.7 3.1 2.4 2.1 1.9 1.8 1.8 1.8

Page 14: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 13 of 29

Matrix B Passive Devices Scattering Parameter: Reflection coefficient

Frequency (S21 - Transmission) Measured Magnitude (+/- dB)

0-10 10-20 20-30 30-40 40-50 50-60 60-70 70-80 80-90 9 kHz to 10 MHz 0.029 0.039 0.048 0.064 0.086 0.111 0.147 0.219 0.355

Using E5071C with 85032F Calibration Kit

(0.01 to 3) GHz 0.031 0.04 0.049 0.065 0.086 0.11 0.144 0.199 0.3 (3 to 6) GHz 0.062 0.066 0.072 0.084 0.101 0.122 0.153 0.206 0.304

(6 to 8.5) GHz 0.092 0.096 0.1 0.109 0.122 0.14 0.168 0.217 0.313 (0.045 to 2) GHz 0.024 0.035 0.046 0.059 0.077 0.106 0.137 0.186 0.243 Using N5230C, N5245A,

E8363C with 85054B Calibration Kit

(2 to 4) GHz 0.044 0.058 0.065 0.073 0.084 0.106 0.14 0.18 0.244 (4 to 18) GHz 0.079 0.087 0.097 0.105 0.114 0.133 0.157 0.193 0.253

(0.045 to 6) GHz 0.024 0.035 0.046 0.06 0.077 0.106 0.137 0.186 0.243 Using N5230C, N5245A, E8363C with 85052B

Calibration Kit (6 to 20) GHz 0.052 0.064 0.071 0.078 0.089 0.11 0.143 0.181 0.242

(0 to 26.5) GHz 0.096 0.102 0.111 0.123 0.137 0.157 0.186 0.224 0.296 (0.045 to 20) GHz 0.024 0.035 0.047 0.061 0.078 0.106 0.137 0.186 0.243 Using N5230C, N5245A,

E8363C with 85056A Calibration Kit

(20 to 40) GHz 0.108 0.129 0.147 0.153 0.161 0.188 0.213 0.25 0.327 (40 to 50) GHz 0.184 0.193 0.2 0.208 0.219 0.235 0.27 0.381 0.566

Frequency (S21 - Transmission) Measured Phase (+/- Degree)

0-10 10-20 20-30 30-40 40-50 50-60 60-70 70-80 80-90 9 kHz to 10 MHz 0.193 0.254 0.317 0.424 0.571 0.756 1.011 1.543 2.868

Using E5071C with 85032F Calibration Kit

(0.01 to 3) GHz 0.205 0.263 0.325 0.429 0.573 0.755 0.979 1.402 2.292 (3 to 6) GHz 0.407 0.439 0.478 0.555 0.672 0.833 1.041 1.445 2.319

(6 to 8.5) GHz 0.613 0.634 0.662 0.719 0.813 0.95 1.138 1.52 2.386

(0.045 to 2) GHz 0.157 0.229 0.302 0.374 0.462 0.672 0.89 1.231 1.605 Using N5230C, N5245A, E8363C with 85054B

Calibration Kit (2 to 4) GHz 0.157 0.239 0.3 0.368 0.456 0.567 0.814 1.152 1.604

(4 to 18) GHz 0.289 0.375 0.42 0.474 0.548 0.679 0.919 1.173 1.611 (0.045 to 6) GHz 0.156 0.229 0.303 0.379 0.465 0.674 0.892 1.231 1.605 Using N5230C, N5245A,

E8363C with 85052B Calibration Kit

(6 to 20) GHz 0.189 0.421 0.461 0.51 0.58 0.706 0.939 1.189 1.622 (0 to 26.5) GHz 0.326 0.68 0.739 0.816 0.915 1.049 1.251 1.491 1.989

(0.045 to 20) GHz 0.161 0.232 0.306 0.386 0.471 0.678 0.895 1.231 1.605 Using N5230C, N5245A, E8363C with 85056A

Calibration Kit (20 to 40) GHz 0.639 0.679 0.738 0.815 0.914 1.051 1.252 1.491 1.99 (40 to 50) GHz 1.226 1.292 1.342 1.4 1.474 1.584 1.825 2.59 4.009

Page 15: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 14 of 29

Matrix C RF Absolute Power Measure ( 9 kHz to 12.4 GHz)

REFERENCE STANDARD OR

EQUIPMENT

Power Sensors E9304A, 8481A, 8482A, 8485A, 8487A, E9300A, N8482A, N8485A, N8488A

Frequency Ranges (uncertainties in dB)

Frequency Range

9 kHz ≤ f <

100 kHz

100 kHz ≤ f <

10 MHz

10 MHz ≤ f <

50 MHz

50 MHz ≤ f <

100 MHz

0.1 GHz ≤ f <

2 GHz

2 GHz ≤ f <

4.2 GHz

4.2 GHz ≤ f <

6 GHz

6 GHz ≤ f <

8 GHz

8 GHz ≤ f <

12.4 GHz -140 dBm ≤ P < -130 dBm 0.167 0.073 0.072 0.070 0.072 0.073 0.074 0.078 0.079

-130 dBm ≤ P < -110 dBm 0.16 0.067 0.064 0.061 0.063 0.065 0.068 0.071 0.072

-110 dBm ≤ P < -90 dBm 0.16 0.05 0.05 0.045 0.049 0.051 0.051 0.055 0.056

-90 dBm ≤ P < -30 dBm 0.16 0.041 0.04 0.035 0.04 0.042 0.042 0.046 0.047

-30 dBm ≤ P < -20 dBm 0.16 0.04 0.039 0.034 0.038 0.041 0.041 0.045 0.046

-20 dBm ≤ P < -10 dBm 0.13 0.04 0.038 0.032 0.037 0.04 0.04 0.044 0.045

-10 dBm ≤ P < -3 dBm 0.13 0.039 0.038 0.032 0.037 0.039 0.04 0.044 0.045

-3 dBm ≤ P < 3 dBm 0.027 0.027 0.028 0.02 0.024 0.024 0.031 0.036 0.037

3 dBm ≤ P < 10 dBm 0.058 0.035 0.031 0.024 0.03 0.033 0.033 0.038 0.039

10 dBm ≤ P < 15 dBm 0.097 0.04 0.036 0.03 0.035 0.038 0.038 0.042 0.044

15 dBm ≤ P < 20 dBm 0.1 0.046 0.046 0.041 0.046 0.048 0.049 0.051 0.061

20 dBm ≤ P < 30 dBm 0.156 0.16 0.16 0.15 0.16 0.16 0.16 0.161 0.16

Matrix C (cont) RF Absolute Power Measure (12.4 GHz to 67 GHz) REFERENCE

STANDARD OR EQUIPMENT

Power Sensors E9304A, 8481A, 8482A, 8485A, 8487A, E9300A, N8482A, N8485A, N8488A

Frequency Ranges (uncertainties in dB)

Frequency Range

12.4 GHz ≤ f <

14 GHz

14 GHz ≤ f <

18 GHz

18 GHz ≤ f ≤

24 GHz

24 GHz ≤ f <

26.5 GHz

26.5 GHz ≤ f <

33 GHz

33 GHz ≤ f ≤

40 GHz

40 GHz ≤ f ≤

50 GHz

50 GHz ≤ f ≤

60 GHz

60 GHz ≤ f ≤

67 GHz -140 dBm ≤ P < -130 dBm 0.077 0.078 0.078 0.096 0.1 0.11 0.12 0.14

-130 dBm ≤ P < -110 dBm 0.071 0.072 0.072 0.09 0.097 0.1 0.12 0.13

-110 dBm ≤ P < -90 dBm 0.056 0.057 0.057 0.077 0.085 0.093 0.11 0.13

-90 dBm ≤ P < -30 dBm 0.047 0.048 0.048 0.07 0.079 0.088 0.1 0.11 0.19

-30 dBm ≤ P < -20 dBm 0.046 0.047 0.047 0.069 0.079 0.087 0.1 0.11 0.1

-20 dBm ≤ P < -10 dBm 0.045 0.046 0.046 0.069 0.078 0.086 0.1 0.11 0.1

-10 dBm ≤ P < -3 dBm 0.045 0.046 0.046 0.068 0.078 0.086 0.1 0.11 0.1

-3 dBm ≤ P < 3 dBm 0.037 0.039 0.039 0.064 0.074 0.083 0.097 0.11 0.1

3 dBm ≤ P < 10 dBm 0.039 0.041 0.041 0.065 0.075 0.084 0.098 0.11 0.1

10 dBm ≤ P < 15 dBm 0.043 0.045 0.045 0.068 0.077 0.086 0.1 0.11 0.11

15 dBm ≤ P < 20 dBm 0.065 0.068 0.073 0.075 0.09 0.11 0.12 0.13 0.11

20 dBm ≤ P < 30 dBm 0.16 0.17 0.17 0.17

Page 16: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 15 of 29

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

DC voltmeter DCV source

(0 to 220) mV (0.22 to 2.2) V (2.2 to 11) V (11 to 22) V (22 to 220) V

(220 to 1 100) V

7.5 µV/V + 0.4 µV 5 µV/V + 0.7 µV

3.5 µV/V + 2.5 µV 3.5 µV/V + 4 µV 5 µV/V + 40 µV

6.5 µV/V + 400 µV

Multifunction calibrator Fluke 5730A

Direct Measurement

DC current meter DC current source

(0 to 220) μA (0.22 to 2.2) mA (2.2 to 22) mA (22 to 100) mA

(100 to 220) mA (0.22 to 1) A (1 to 2.2) A (2.2 to 5) A (5 to 10) A

40 µA/A + 6 nA 35 µA/A + 7 nA

35 µA/A + 40 nA 45 µA/A + 0.7 µA 50 uA/A - 1.2 uA 80 µA/A + 12 µA 120 uA/A - 42uA

360 µA/A + 480 µA 360 µA/A + 580 µA

Multifunction calibrator Fluke 5730A

Direct Measurement

Matrix D Signal Analyzers –Absolute RF Power in 50 Ω Coaxial Line

Frequency Range RF Power Level

(Uncertainty in mW/W) Power Sensor 10µW to 1nW 1nW to 500nW 500nW to 1µW

10 MHz to 18 GHz 20 9.8 9.8 8481D 50 MHz to 26.5 GHz 39 18 18 8485D 50 MHz to 50 GHz 38 15 15 8487D

Frequency Range

RF Power Level (Uncertainty in mW/W)

Power Sensor 1mW to

1µW

1µW to

100µW

100µW to

500µW

500µW to

700µW

700µW to

3mW

3mW to

6.31mW

100 kHz to 3 GHz 14 14 14 14 14 14 75 Ω Type N 8483A

100 kHz to 4.2 GHz 9.3 9.7 9.2 9.2 9.2 9.2 8482A 10 MHz to 18 GHz 9.4 10 9.4 9.4 9.4 9.2 8481A

50 MHz to 26.5 GHz 17 17 17 17 17 17 8485A 50 MHz to 50 GHz 16 16 16 16 16 16 8487A

Page 17: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 16 of 29

AC voltmeter AC volt source

(0.22 to 2.2) mV (10 to 20) Hz (20 to 40) Hz

40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz

(100 to 300) kHz (300 to 500) kHz

500 kHz to 1 MHz (2.2 to 22) mV

(10 to 20) Hz (20 to 40) Hz

40 Hz to 20 kHz (20 to 50) kHz

(50 to 100) kHz (100 to 300) kHz (300 to 500) kHz

500 kHz to 1 MHz (22 to 220) mV

(10 to 20) Hz (20 to 40) Hz

40 Hz to 20 kHz (20 to 50) kHz

(50 to 100) kHz (100 to 300) kHz (300 to 500) kHz

500 kHz to 1 MHz (0.22 to 2.2) V

(10 to 20) Hz (20 to 40) Hz

40 Hz to 20 kHz (20 to 50) kHz

(50 to 100) kHz (100 to 300) kHz (300 to 500) kHz

500 kHz to 1 MHz (2.2 to 22) V

(10 to 20) Hz (20 to 40) Hz

40 Hz to 20 kHz (20 to 50) kHz

(50 to 100) kHz (100 to 300) kHz (300 to 500) kHz

500 kHz to 1 MHz

240 µV/V + 4 µV 90 µV/V + 4 µV 80 µV/V + 4 µV 200 µV/V + 4 µV 500 µV/V + 5 µV 1.1 mV/V + 10 µV 1.4 mV/V + 20 µV 2.7 mV/V + 20 µV

240 µV/V + 4 µV 90 µV/V + 4 µV 80 µV/V + 4 µV 200 µV/V + 4 µV 500 µV/V + 5 µV 1.1 mV/V + 10 µV 1.4 mV/V + 20 µV 2.7 mV/V + 20 µV

240 µV/V + 12 µV 90 µV/V + 7 µV 57 µV/V + 7 µV 120 µV/V + 7 µV

310 µV/V + 17 µV 655 µV/V + 20 µV 1.4 mV/V + 25 µV 2.7 mV/V + 45 µV

240 µV/V + 40 µV 90 µV/V + 15 µV 42 µV/V + 8 µV 67 µV/V + 10 µV 85 µV/V + 30 µV

336 µV/V + 80 µV 1 mV/V + 200 µV

1.7 mV/V + 300 µV

240 µV/V + 400 µV 90 µV/V + 150 µV 42 µV/V + 50 µV

67 µV/V + 100 µV 83 µV/V + 200 µV

254 µV/V + 600 µV 1.0 mV/V + 2 mV

1.5 mV/V + 3.2 mV

Multifunction calibrator Fluke 5730A

Direct Measurement

Page 18: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 17 of 29

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

AC volt meter AC volt source

(22 to 120) V (10 to 20) Hz (20 to 40) Hz

40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz

(100 to 300) kHz (300 to 500) kHz

500 kHz to 1 MHz (120 to 220) V

(10 to 20) Hz (20 to 40) Hz

40 Hz to 20 kHz (20 to 50) kHz

(50 to 100) kHz (220 to 250) V

(15 to 50) Hz

240 µV/V + 4 mV 90 µV/V + 1.5 mV 52 µV/V + 0.6 mV 80 µV/V + 1 mV

150 µV/V + 2.5 mV 900 µV/V + 16mV 4.4 mV/V + 40 mV 8 mV/V to 80 mV

240 µV/V + 4 mV 90 µV/V + 1.5 mV 52 µV/V + 0.6 mV 80 µV/V + 1 mV

150 µV/V + 2.5 mV

300 µV/V + 16 mV

Multifunction calibrator Fluke 5730A

Direct Measurement

AC voltmeter AC volt source

(220 to 700) V (30 to 50) kHz

(50 to 100) kHz

(220 to 1000) V (40 to 50) Hz

50 Hz to 1 kHz (1 to 20) kHz

(20 to 30) kHz

600 µV/V + 11 mV 2.3 mV/V + 45 mV

90 µV/V + 4 mV 70 µV/V + 3.5 mV 165 µV/V + 6 mV

600 µV/V + 11 mV

Multifunction calibrator Fluke 5730A, 5725A Direct Measurement

AC current meter AC current source

(9 to 220) µA (10 to 20) Hz (20 to 40) Hz

40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz

(0.22 to 2.2) mA (10 to 20) Hz (20 to 40) Hz

40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz

250 µA/A + 16 nA 160 µA/A + 10 nA 120 µA/A + 8 nA

280 µA/A + 12 nA 1.1 mA/A + 65 nA

250 µA/A + 40 nA 160 µA/A + 35 nA 120 µA/A + 35 nA

200 µA/A + 110 nA 1.1 mA/A + 650 µA

Multifunction calibrator Fluke 5730A

Direct Measurement

Page 19: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 18 of 29

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

AC current meter AC current source

(2.2 to 22) mA (10 to 20) Hz (20 to 40) Hz

40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz

(22 to 220) mA (10 to 20) Hz (20 to 40) Hz

40 Hz to 1 kHz (1 to 5) kHz

(5 to 10) kHz (0.22to 2.2) A

20 Hz to 1 kHz (1 to 5) kHz

(5 to 10) kHz

250 µA/A + 400 nA 160 µA/A + 350 nA 120 µA/A + 350 nA 200 µA/A + 550 nA

1.1 mA + 5 µA

250 µA/A + 4 µA 160 µA/A + 3.5 µA 103 µA/A + 2.5 µA 200 µA/A + 3.5 µA 1.1 mA/A + 10 µA

244 µA/A + 35 µA 450 µA/A + 80 µA 7 mA/A + 160 µA

Multifunction calibrator Fluke 5730A

Direct Measurement

AC current meter AC current source

(2.2 to 5.0) A 40 Hz to 1 kHz

(1 to 5) kHz (5 to 10) kHz

(5.0 to 10) A 40 Hz to 1 kHz

(1 to 5) kHz (5 to 10) kHz

460 µA/A + 170 µA 950 µA/A + 380 µA 3.6 mA/A + 750 µA

460 µA/A + 260 µA 950 µA/A + 420 µA 3.6 mA/A + 750 µA

Multifunction calibrator Fluke 5730A, 5725A

Ohm Meter Resistance Source (Specific Value)

1 Ω 1.9 Ω 10 Ω 19 Ω

100 Ω 190 Ω 1 kΩ

1.9 kΩ 10 kΩ 19 kΩ 100 kΩ 190 kΩ 1 MΩ

1.9 MΩ 10MΩ 19 MΩ 100 MΩ

0.1 mΩ 0.18 mΩ 0.23 mΩ 0.44 mΩ

1 mΩ 1.9 mΩ 6.5 mΩ 12 mΩ 65 mΩ 0.12 Ω 0.86 Ω 1.63 Ω 13 Ω

34.2 Ω 0.4 kΩ

0.89 kΩ 10 kΩ

Multifunction calibrator Fluke 5730A

Direct Measurement

Page 20: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 19 of 29

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

DC Dissipated Power

(1 to 11.5) µA Shunt 100 kΩ

(0.1 to 1.15) µA Shunt 1 MΩ

(10 to 115) nA Shunt 10 MΩ

(1 to 11.5) nA Shunt 10 MΩ

(1 to 11.5) nA Shunt 100 MΩ

(0.1 to 1.15) nA Shunt 1 GΩ

(10 to 115) pA Shunt 1 GΩ

(1 to 115) pA Shunt 10 GΩ

(0.1 to 1.15) pA Shunt 100 GΩ

0.013 mA/A + 0.028 nA

0.02 mA/A + 1.7 pA

0.051 mA/A + 0.54 pA

0.026 mA/A + 0.25 pA

0.14 mA/A + 0.11 pA

0.24 mA/A + 0.021 pA

0.24 µA/A + 1.4 fA

0.92 mA/A + 0.38 fA

0.83 mA/A + 0.59 fA

Keysight 16353A, 16353H Shunt, Keysight 3458A

Multimeter

DC Dissipated Power (1 to 3) A (3 to 20) A

0.017 mA/A + 0.12 mA 0.22 mA/A + 0.29 mA

Keysight 16353J Shunt + Keysight 3458A

Multimeter

DC Voltage Measure

(0 to 100) mV (0.1 to 1) V (1 to 10) V

(10 to 100) V (100 to 1 000) V

6.5 µV/V+ 0.42 µV 6 µV/V + 0.42 µV

5.4 µV/V + 0.93 µV 7.6 µV/V + 42 µV 7.6 µV/V + 150 µV

Keysight 3458A Multimeter

DC Current Measure

(0 to 100) µA (0.1 to 1.0) mA (1 to 10) mA

(10 to 100) mA (0.1 to 1) A

24 µA/A + 1.1 nA 24 µA/A + 7.5 nA

24 µA/A + 0.075 µA 41 µA/A + 0.75 µA 120 µA/A + 14µA

Keysight 3458A Multimeter

AC Voltage Meters

1.0 to 10) mV (10 to 40) Hz

40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz

(50 to 100) kHz (100 to 300) kHz

0.035 % of reading + 3.5 µV 0.023 % of reading + 1.3 µV 0.035 % of reading + 1.3 µV 0.12 % of reading + 1.3 µV 0.58 % of reading +1.3 µV 4.6 % of reading + 2.3 µV

Generate using Fluke 5520A Multifunction Calibrator with Multimeter 3458A Direct and Indirect

Measurement

Page 21: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 20 of 29

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

AC Voltage Meters

(10 to 100) mV (10 to 40) Hz

40 Hz to 1 kHz (1 to 20) kHz

(20 to 50) kHz (50 to 100) kHz

(100 to 300) kHz 300 kHz to 1 MHz

100 mV to 1 V (10 to 40) Hz

40 Hz to 1 kHz (1 to 20) kHz

(20 to 50) kHz (50 to 100) kHz

(100 to 300) kHz 300 kHz to 1 MHz

(1 to 10) V (10 to 40) Hz

40 Hz to 1 kHz (1 to 20) kHz

(20 to 50) kHz (50 to 100) kHz

(100 to 300) kHz 300 kHz to 1 MHz

(10 to 100) V (10 to 40) Hz

40 Hz to 1 kHz (1 to 20) kHz

(20 to 50) kHz (50 to 100) kHz

(100 to 200) kHz (10 to 20) V

200 kHz to 1MHz (100 to 700) V

40 Hz to 1 kHz 1 kHz to 20 kHz 20 kHz to 50 kHz 50 kHz to 100 kHz

0.008 4 % of reading + 4.7 µV 0.008 4 % of reading + 2.3 µV 0.017 % of reading + 2.3 µV 0.035 % of reading + 2.3 µV 0.093 % of reading + 2.3 µV 0.35 % of reading + 12 µV 1.2 % of reading + 12 µV

0.008 4 % of reading + 46 µV 0.008 4 % of reading + 23 µV 0.017 % of reading + 23 µV 0.035 % of reading + 23 µV 0.093 % of reading + 23 µV 0.35 % of reading + 0.12 mV 1.2 % of reading + 0.12 mV

0.008 4 % of reading + 0.46 mV 0.008 4 % of reading + 0.23 mV 0.017 % of reading + 0.23 mV 0.035 % of reading + 0.23 mV 0.093 % of reading + 0.23 mV

0.35 % of reading + 1.2 mV 1.2 % of reading + 1.2 mV

0.024 % of reading +4.6 mV 0.024 % of reading + 2.3 mV 0.024 % of reading + 2.3 mV 0.041 % of reading + 2.3 mV 0.14 % of reading + 2.3 mV 0.46 % of reading + 12 mV

1.73 % of reading + 11.6 mV

0.047 % of reading + 16.2 mV 0.07 % of reading + 16.2 mV 0.14 % of reading + 16.2 mV 0.35 % of reading + 16.2 mV

Generate using Fluke 5520A Multifunction Calibrator with Multimeter 3458A Direct and Indirect

Measurement

Page 22: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 21 of 29

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

AC Voltage Meters

(1.0 to 10) mV (1 to 4) MHz (4 to 8) MHz

(10 to 100) mV (1 to 2) MHz (2 to 4) MHz (4 to 8) MHz

(8 to 10) MHz (0.1 to 1) V

(1 to 2) MHz (2 to 4) MHz (4 to 8) MHz

(8 to 10) MHz (1 to 10) V

(1 to 2) MHz (2 to 4) MHz (4 to 8) MHz (8 to 10) MHz

8.1 % of reading + 8.1 µV 23 % of reading + 9.3 µV

17.3 % of reading + 11.6 µV

4.6 % of reading + 81 µV 4.6 % of reading + 92 µV

17.3 % of reading + 11.6 µV

1.73 % of reading + 0.12 mV 4.6 % of reading + 0.81 mV 4.6 % of reading + 0.92 mV

1.73 % of reading + 0.12 mV

17.3 % of reading + 1.16 mV 4.6 % of reading + 8.1 mV 4.6 % of reading + 9.2 mV

17.3 % of reading + 1.16 mV

Generate using Fluke 5520A Multifunction Calibrator with Multimeter 3458A Direct and Indirect

Measurement

AC Current Meters

(10 to 100) µA (10 to 20) Hz (20 to 45) Hz

45 Hz to 1 kHz 100 µA to 1 mA

(10 to 20) Hz (20 to 45) Hz

(45 to 100) Hz 100 Hz to 5 kHz

(1 to 10) mA (10 to 20) Hz (20 to 45) Hz

(45 to 100) Hz 100 Hz to 5 kHz

(10 to 100) mA (10 to 20) Hz (20 to 45) Hz

(45 to 100) Hz 100 Hz to 5 kHz

0.47 % of reading + 0.035 µA 0.18 % of reading + 0.035 µA 0.07 % of reading + 0.035 µA

0.47 % of reading + 0.23 µA 0.18 % of reading + 0.23 µA 0.07 % of reading + 0.23 µA

0.036 % of reading + 0.23 µA

0.47 % of reading + 2.3 µA 0.18 % of reading + 2.3 µA 0.07 % of reading + 2.3 µA 0.036 % of reading + 2.3 µA

0.47 % of reading + 23 µA 0.18 % of reading + 23 µA 0.07 % of reading + 23 µA

0.036 % of reading + 23 µA

Generate using Fluke 5520A Multifunction Calibrator with Multimeter 3458A Direct and Indirect

Measurement

Page 23: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 22 of 29

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

AC Current Meters

100 mA to 1 A (10 to 20) Hz (20 to 45) Hz

(45 to 100) Hz 100 Hz to 5 kHz

0.47 % of reading + 23 mA 0.19 % of reading + 23 mA 0.093 % of reading + 23 mA 0.12 % of reading + 23 mA

Generate using Fluke 5520A Multifunction Calibrator with Multimeter 3458A Direct and Indirect

Measurement

Resistance Measure

(1 to 10) Ω (10 to 100) Ω (0.1 to 1) kΩ

(1 to 10) kΩ

(10 to 100) kΩ (0.1 to 1) MΩ

(1 to 10) MΩ

(10 to 100) MΩ (0.1 to 1) GΩ

21 µΩ/Ω + 8 µΩ/Ω 15 µΩ/Ω + 8 µΩ/Ω

13 µΩ/Ω + 0.79 µΩ/Ω

13 µΩ/Ω + 0.79 µΩ/Ω 13 µΩ/Ω + 0.79 µΩ/Ω 18 µΩ/Ω + 2.6 µΩ/Ω

62 µΩ/Ω + 12 µΩ/Ω

580 µΩ/Ω + 37 µΩ/Ω 5.8 mΩ/Ω + 260 µΩ/Ω

Keysight 3458A

Resistance Meters

(0 to 11) Ω (11 to 33) Ω (33 to 110) Ω

(0.11 to 1.1) kΩ (1.1 to 11) kΩ (11 to 110) kΩ

(0.11to 1.1) MΩ (1.1 to 3.3) MΩ (3.3 to 11) MΩ (11 to 33) MΩ (33 to 110) MΩ (110 to 330) MΩ (0.33 to 1) GΩ

31.2 µΩ/Ω + 0.78 mΩ 23.6 µΩ/Ω + 1.16 mΩ 21.8 µΩ/Ω + 1.17 mΩ 21.7 µΩ/Ω + 1.55 mΩ 21.7 µΩ/Ω + 15.6 mΩ 21.7 µΩ/Ω + 0.16 Ω 24.8 µΩ/Ω + 1.55 Ω 46.6 µΩ/Ω + 23.3 Ω 0.1 mΩ/Ω + 38.8 Ω

0.19 mΩ/Ω + 1.94 kΩ 0.38 mΩ/Ω + 2.32 kΩ 2.33 mΩ/Ω + 77.6 kΩ 11.6 mΩ/Ω + 0.39 MΩ

Multiproduct Calibrator Fluke 5520A,

Direct Measurement

Capacitance - Source

(0.19 to 3.29) nF (3.3 to 10.99) nF (11 to 109.99) nF

(110 to 329.99) nF (0.33 to 1.099) µF (1.1 to 3.299) µF (3.3 to 10.99) µF (11 to 32.99) µF (33 to 109.99) µF (110 to 329.99) µF (0.33 to 1.099) mF

3.2 mF/F + 8.3 pF 1.6 mF/F + 8.3 pF 1.6 mF/F + 83 pF

1.6 mF/F + 0.25 nF 1.6 mF/F + 0.83 nF 1.6 mF/F + 2.5 nF 1.6 mF/F + 8.3 nF 2.5 mF/F + 25 nF 2.8 mF/F + 84 nF

2.8 mF/F + 0.25 µF 4 mF/F + 1.2 µF

Multiproduct Calibrator Fluke 5520A,

Direct Measurement

Page 24: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 23 of 29

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

Capacitance - Source

(1.1 to 3.299) mF (3.3 to 10.99) mF (11 to 32.99) mF (33 to 110) mF

4 mF/F + 3.6 µF 4 mF/F + 12 µF

8.3 mF/F + 36 µF 12 mF/F + 0.11 mF

Multiproduct Calibrator Fluke 5520A,

Direct Measurement

Electrical Simulation of Thermocouple Indicators

Type B (600 to 800) °C

(800 to 1 000) °C (1 000 to 1 550) °C (1 550 to 1 820) °C

Type C (0 to 150) °C

(150 to 650) °C (650 to 1 000) °C

(1 000 to 1 800) °C (1 800 to 2 316) °C

Type E (-250 to -100) °C (-100 to -25) °C (-25 to 350) °C (350 to 650) °C

(650 to 1 000) °C Type J

(-210 to -100) °C (-100 to -30) °C (-30 to 150) °C (150 to 760) °C

(760 to 1 200) °C Type K

(-200 to -100) °C (-100 to -25) °C (-25 to 120) °C

(120 to 1 000) °C (1 000 to 1 372) °C

Type L (-200 to -100) °C (-100 to 800) °C (800 to 900) °C

0.47 °C 0.36 °C 0.32 °C 0.35 °C

0.32 °C 0.28 °C 0.33 °C 0.53 °C 0.88 °C

0.53 °C 0.18 °C 0.16 °C 0.18 °C 0.23 °C

0.29 °C 0.18 °C 0.16 °C 0.19 °C 0.25 °C

0.35 °C 0.20 °C 0.18 °C 0.28 °C 0.42 °C

0.39 °C 0.28 °C 0.19 °C

Multiproduct Calibrator Fluke 5520A

Page 25: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 24 of 29

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

Electrical Simulation of Thermocouple Indicators

Type N (-200 to -100) °C (-100 to -25) °C (-25 to 120) °C (120 to 410) °C

(410 to 1 300) °C Type R

(0 to 250) °C (250 to 400) °C

(400 to 1 000) °C (1 000 to 1 767) °C

Type S (0 to 250) °C

(250 to 1 000) °C (1 000 to 1 400) °C (1 400 to 1 767) °C

Type T (-250 to -150) °C

(-150 to 0) °C (0 to 120) °C

(120 to 400) °C Type U

(-200 to 0) °C (0 to 600) °C

0.42 °C 0.24 °C 0.21 °C 0.2 °C 0.29 °C

0.6 °C 0.37 °C 0.35 °C 0.42 °C

0.5 °C 0.38 °C 0.39 °C 0.49 °C

0.66 °C 0.26 °C 0.18 °C 0.16 °C

0.59 °C 0.29 °C

Multiproduct Calibrator Fluke 5520A

4T Capacitance Meter

1 pF (20 to 100) Hz (0.1 to 1) kHz

1 kHz to 1 MHz (1 to 3) MHz (3 to 5) MHz

(5 to 13) MHz 10 pF

(20 to 100) Hz (0.1 to 1) kHz

1 kHz to 1 MHz (1 to 3) MHz (3 to 5) MHz (5 to 13) MHz

15 fF

0.73 fF 3.1E-08 fF/ Hz + 0.09 fF 1.6E-07 fF/ Hz - 0.05 fF 2.3E-07 fF/ Hz - 0.27 fF 3.6E-07 fF/ Hz - 0.96 fF

19 fF

0.82 fF 0.63 fF

2.5E-08 fF/ Hz + 0.6 fF 8.5E-08 fF/ Hz + 0.44 fF 1.3E-07 fF/ Hz + 0.28 fF

Measure using 16380A /16380C Capacitor Set Measure using 16380A /16380C Capacitor Set

Page 26: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 25 of 29

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

4T Capacitance Meter

100 pF (20 to 100) Hz (0.1 to 1) kHz

1 kHz to 1 MHz (1 to 3) MHz (3 to 5) MHz

(5 to 13) MHz 1000 pF

(20 to 100) Hz (0.1 to 1) kHz

1 kHz to 1 MHz (1 to 3) MHz (3 to 5) MHz

(5 to 13) MHz 10 nF

(20 to 100k) Hz 100 nF

(20 to 100k) Hz 1 µF

(20 to 120) Hz 1 kHz 10 kHz

100 kHz 10 µF

(20 to 120) Hz 1 kHz 10 kHz

100 kHz

18 fF 5.2 fF

2.1E-07 fF/ Hz + 5.2 fF 1.2E-06 fF/ Hz + 4.1 fF 3.2E-06 fF/ Hz - 2.1 fF 4.7E-06 fF/ Hz - 10 fF

0.16 pF 0.11 pF

1.1E-05 fF/ Hz + 0.11 pF 9.5E-05 fF/ Hz + 0.02 pF 1.7E-04 fF/ Hz - 0.21 pF 2.9E-04 fF/ Hz - 0.84 pF

0.52 pF

4.4 pF

55 pF 47 pF 48 pF 82 pF

0.78 nF 0.77 nF 1.8 nF 7.7 nF

Measure using 16380A /16380C Capacitor Set Measure using 16380A /16380C Capacitor Set

4T Dissipation Factor2

1 pF (20 to 100) Hz

100 Hz to 1 kHz 1 kHz to 1 MHz

(1 to 4) MHz (4 to 13) MHz

10 pF (20 to 100) Hz

100 Hz to 1 kHz 1 kHz to 1 MHz

(1 to 4) MHz (4 to 13) MHz

9.4E-03 pF 9.20E-04 pF 2.1E-05 pF 3.1E-05 pF 1.6E-04 pF

1.3E-03 pF 8.7E-05 pF 2E-05 pF 2E-05 pF

4.7E-05 pF

Measure using 16380A /16380C Capacitor Set Measure using 16380A /16380C Capacitor Set

Page 27: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 26 of 29

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

4T Dissipation Factor2

100 pF (20 to 100) Hz

100 Hz to 1 kHz 1 kHz to 1 MHz

(1 to 4) MHz (4 to 13) MHz

1 000 pF (20 to 100) Hz

100 Hz to 1 kHz 1 kHz to 1 MHz

(1 to 4) MHz (4 to 13) MHz

10 nF (20 to 120) Hz

1 kHz 10 kHz

100 kHz 100 nF

(20 to 120) Hz 1 kHz 10 kHz

100 kHz 1 µF

(20 to 120) Hz 1 kHz 10 kHz

100 kHz 10 µF

(20 to 120) Hz 1 kHz 10 kHz 100 kHz

1.6E-04 2.1E-05 2E-05 2E-05

5.5E-05

1.9E-04 2E-05 2E-05 3E-05

1.6E-04

2.2E-05 2E-05

2.2E-05 2.2E-05

3.4E-05 2E-05

2.6E-05 3E-05

4.3E-05 2E-05

3.9E-05 4.6E-05

4.3E-05 3E-05

2.8E-04 7.2E-04

Measure using 16380A /16380C Capacitor Set Measure using 16380A /16380C Capacitor Set

Page 28: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 27 of 29

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

4T AC Resistance Meter

0.01 Ω DC

1kHz 0.1 Ω

DC 1kHz

1 Ω DC

1kHz 10 Ω

DC (20 to 100) Hz

100 Hz to 1 MHz (1 to 3) MHz (3 to 5) MHz

(5 to 13) MHz 100 Ω

DC (20 to 100) Hz

100 Hz to 1 MHz (1 to 3) MHz (3 to 5) MHz

(5 to 13) MHz 1 kΩ

DC (20 to 100) Hz

100 Hz to 100 kHz 100 kHz

100 kHz to 1 MHz (1 to 3) MHz (3 to 5) MHz

(5 to 13) MHz 10 kΩ

DC (20 to 100) Hz

100 kHz 100 kHz to 1 MHz

1 MHz

9.9E-04 Ω

1.9E-03 mΩ

2.0E-03 Ω 1.9E-02 mΩ

1.1E-03 Ω 2E-04 Ω

4.1E-03 Ω 6.3E-04 Ω 1.2E-03 Ω

1.4E-09 Ω/ Hz + 2.07E-03 Ω 2.0E-09 Ω/ Hz - 6.7E-05 Ω 6.3E-09 Ω /Hz - 2.2E-2 Ω

1.7E-02 Ω 6E-03 Ω

5.9E-03 Ω 1.0E-08 Ω/ Hz + 2.1E-02 Ω

5.1E-02 Ω 3.2E-08 Ω/ Hz - 1.1E-01 Ω

1.3E-01 Ω 6.1E-02 Ω 5.9E-02 Ω 3.1E-01 Ω 3.1E-01 Ω 3.1E-01 Ω

1.0E-07 Ω/ Hz - 1.7E-03 Ω 3.2E-07 Ω/ Hz - 1.1 Ω

5.2 Ω 2.3 Ω 2.1 Ω 3.2 Ω 3.1 Ω

Measure using 42030A Resistor Set

Page 29: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 28 of 29

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

4T AC Resistance Meter

100 kΩ DC

(20 to 100) Hz 100 kHz

100 kHz to 1 MHz 1 MHz

6.3 Ω 22 Ω 12 Ω 24 Ω 6.6 Ω

Measure using 42030A Resistor Set

DC Power (Power Meters)

3 µW 10 µW 30 µW 100 µW 300 µW 1 mW 3 mW 10 mW 30 mW 100 mW

34 nW 27 nW 42 nW 31 nW

0.23 µW 0.41 µW 1.7 µW 2.8 µW 24 µW

110 µW

Fluke 57X0A Calibrator, Keysight 3458A, HP

11683A

Time and Frequency

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

Frequency Sources 10 MHz 5 pHz/Hz

Measure using 5071A Frequency

Standard, 53132A Counter Differential measurement

Signal Sources – Frequency 100 MHz to 45 GHz 0.7 Hz

Signal Analyzer E4448A Frequency Standard 5071A

Frequency Standard Differential measurement

Frequency (Source)

(0.1 to 1) Hz 1 Hz to 250 kHz (0.25 to 1) MHz

1 MHz to 40 GHz

2 µHz/Hz + 0.43 µHz 2.4 µHz/Hz

5.2 nHz/Hz + 0.14 mHz 5.3 nHz/Hz

Function Generators 33250A, Signal Generator

E8257D

Frequency (Measure)

(0.1 to 1) Hz (1 to 10) Hz

(10 to 100) Hz 100 Hz to 12.4 GHz

(12.4 to 40) GHz

67 pHz 0.88 nHz 4.7 nHz

24 pHz/Hz 23 pHz/Hz + 1.2 Hz

53132A opt 124 Counter and CW Microwave

Counter 5352B

Page 30: CERTIFICATE OF ACCREDITATION...2020/02/27  · Ping-Chen district 32450, Tao-Yuan City, Taiwan R.O.C. has been assessed by ANAB and meets the requirements of international standard

Version 006

Issued: February 27, 2020 www.anab.org

Page 29 of 29

Time and Frequency

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)

Reference Standard, Method, and/or

Equipment

Signal Analyzers, Frequency Counters – Frequency 100 MHz to 50 GHz 1.4 × 10-11 Hz/Hz

Signal Source E8257D, Frequency Standard 5071A Differential measurement

Calibration and Measurement Capability (CMC) is expressed in terms of the measurement parameter, measurement range, expanded uncertainty of measurement and reference standard, method, and/or equipment. The expanded uncertainty of measurement is expressed as the standard uncertainty of the measurement multiplied by a coverage factor of 2 (k=2), corresponding to a confidence level of approximately 95%. Notes: 1. On-site calibration service is available for this parameter, since on-site conditions are typically more variable than those in the laboratory, larger measurement

uncertainties are expected on-site than what is reported on the accredited scope. 2. Unitless linear measure. 3. This scope is formatted as part of a single document including Certificate of Accreditation No. AC-1928.01.