Cell Density Measurements for Algae Cultures

Embed Size (px)

Citation preview

Instructions For Preparing Broth For Chlorella vulgaris and Measuring Cell Density

For Measuring Cell Density for Algae Cultures

Detach flowtube from the top of algae reactor.

Take the syringe and attach it to the inlet of the top of the algae reactor and pull it out a sufficient amount of the algae. (Rinse with DI water three times and spray the tip of the syringe each time to ensure sterility.)

Fill 1 ml cuvettes with the solution with the algae to the marked line on the cuvette and place them in their respective locations on the table to ensure the right cuvette is taken for the OD reading on the spectrometer. Picture shows an example on how to best organize this...

Once the cuvettes are placed, cover the tops of the cuvettes with little pieces of parafilm to prevent contamination.

Then go to the spectrometer, as seen from the picture below, and turn it on.

After turning on the spectrometer, press the second top button under set to adjust to desired wavelength and press Enter.

Fill an empty 1 ml cuvette with DI water and use as the blank to calibrate spectrometer and press the button where it says, measure blank.

Once it is calibrated, you may now insert the cuvettes and take OD readings one at a time.

(Remember, each time when you need to do measurements at a different wavelength, the spectrometer needs to be calibrated again using the blank cuvette with DI water.When measuring OD readings, record the results found in the laboratory notebook that can be found in the lab titled, Algae Growth Systems 002 SDSU 2013 and sign once the measurements or whatever experiments related to algae are completed.