CASE BOOK:ENDOANCHORTM

FIXATION

UC201700368a EN

Page 2: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PURPOSE OF THIS MATERIAL

The Heli-FXTM EndoAnchorTM systems have been used in over 10,000 patients worldwide, with over 50,000 EndoAnchorTM implants deployed to-date. Over 900 patients have been enrolled in IDE and post-market registry studies.

Medtronic is providing this case book to help physicians with making stronger informed decisions for their patients with abdominal aortic aneurysms (AAA) or thoracic aortic aneurysms (TAA), for whom they have or may be considering the Heli-FXTM EndoAnchorTM systems.

Medtronic's Mission is to contribute to human welfare by application of biomedical engineering in the research, design, manufacture, and sale of instruments or appliances that alleviate pain, restore health, and extend life. This technology reflects our commitment to our Mission and improving patient outcomes.

Questions or comments should be directed to Randy Bassett, PhD, Sr. Manager, Medical Affairs, Tel +1.707.543.2259, Fax +1.651.367.7050, [email protected]

2

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 3: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

• The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

• Indicated for use in patients whose endovascular grafts have exhibited migration or endoleak, or are at risk of such complications , in whom augmented radial fixation and/or sealing is required to regain or maintain adequate aneurysm exclusion

• The EndoAnchorTM implant may be implanted at the time of the initial endograft placement, or during a secondary (i.e. repair) procedure

• Evaluated and determined to be compatible with1,2

– Medtronic EndurantTM, ValiantTM, TalentTM and AneuRxTM endografts

– Gore ExcluderTM* and Gore TAGTM* endografts

– Cook ZenithTM* and Zenith TX2TM* endografts

FDA INDICATIONS FOR USE

Cook ZenithTM* Gore ExcluderTM*Medtronic AneuRxTM Medtronic EndurantTMMedtronic TalentTM

No damage post 400M cycles fatigue testing1

(equivalent to 10 years in vivo)

1Per data on file at Medtronic2Per Instructions for Use, the EndoAnchorTM implant should be used with caution in TalentTM and ValiantTM endografts.

33

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 4: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

4

OFF LABEL PROMOTION PROHIBITED

▪Off label promotion of any endograft is strictly prohibited by Medtronic

4

1Per manufacturers' IFU as of May 20162Other contraindication for EndoAnchorTM fixation is in patients with known allergies to the EndoAnchorTM implant material (MP35N-LT)

Infrarenal Neck Parameters for Various Commercially Available AAA Endografts1

Stent Graft Neck Length Neck Angle

EndurantTM ≥10mm ≤60⁰

ZenithTM* Fenestrated ≥4mm ≤45⁰

ZenithTM* Flex ≥15mm ≤60⁰

ExcluderTM* ≥15mm ≤60⁰

OvationTM* Not tested with EndoAnchorTM implants

AorfixTM* Not tested with EndoAnchorTM implants

Endologix AFXTM* Not tested with EndoAnchorTM implants

Endologix PowerlinkTM* Contraindicated with EndoAnchorTM fixation2

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 5: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

Prophylactic Applications

CASE BOOK MAPHOW LEADING PHYSICIANS USE ENDOANCHORTM FIXATION TO

ENHANCE OUTCOMES AND DURABILITY

Conical necks

Multiple variables

Therapeutic Applications

Intra-op type I

endoleak

Late type I

endoleak

Graft migration

Wide necks

Angulated

necks

Short necks

Specialty Applications

Ruptures FEVAR

5

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 6: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

CONICAL NECKS

UC201700368a EN

Page 7: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – CONICAL NECKSHEPPARD MONDY III, MD, SAVANNAH, GA

HOME

7

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 8: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – CONICAL NECKPETER SCHNEIDER, MD, HONOLULU, HAWAII

HOME

8

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 9: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – CONICAL NECKFRANK ARKO, MD, CHARLOTTE, NORTH CAROLINA

HOME

9

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Pre op CT No endoleak at 1-month post op

No endoleak at 1-year post op

Page 10: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY TEVAR – CONICAL NECKALAN LUMSDEN, MD, HOUSTON, TX

HOME

10

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 11: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY TEVAR – CONICAL NECKGRAYSON WHEATLEY, MD, TEMPLE UNIVERSITY, PENNSYLANIA

HOME

11

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 12: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

SHORTNECKS

UC201700368 a EN

Page 13: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – SHORT, CONICAL NECKPETER SCHNEIDER, MD, HONOLULU, HAWAII

HOME

13

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 14: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – SHORT NECKJAMES JOHNSON, DO, CLINTON TOWNSHIP, MICHIGAN

HOME

14

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 15: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY TEVAR – SHORT PROXIMAL/DISTAL NECKSWILLIAM JORDAN, MD, UNIVERSITY OF ALABAMA

HOME

15

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 16: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY TEVAR – SHORT PROXIMAL/DISTAL NECKSPIOTR KASPRZAK, MD, REGENSBURG, GERMANY

Final Angio, leak resolved

Type 1a leak after

graft implantation

4 EndoAnchorTM implants

deployed

Initial angio

HOME

16

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 17: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY TEVAR – SHORT PROXIMAL/DISTAL NECKSJEAN M. PANNETON, MD, EASTERN VIRGINA MEDICAL

HOME

17

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 18: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

WIDE NECKS

UC201700368a EN

Page 19: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – WIDE NECKELIE SEMAAN, MD & MARCUS D’AYALA, MD, BROOKLYN, NY

Post balloon

Type 1

HOME

19

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 20: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

ANGULATED NECKS

UC201700368a EN

Page 21: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – ANGULATIONFRANK ARKO, MD, CHARLOTTE, NORTH CAROLINA

HOME

21

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 22: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – ANGULATIONMAZIN FOTEH, MD, AUSTIN, TEXAS

HOME

22

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 23: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – ANGULATION, INTRA-OP TYPE I ENDOLEAKCOLIN D. BICKNELL, MD, IMPERIAL COLLEGE, LONDON, UK

EndoAnchorTM implants resolve endoleak

1 year post-op follow-up shows no complications

Type 1a endoleak in angulated aortic neck

HOME

23

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 24: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

REVISION EVAR – ANGULATION, LATE TYPE I ENDOLEAKKRASSI IVANCEV, MD, LONDON, UK

HOME

24

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 25: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

MULTI-VARIABLE COMPLEX NECKS

UC201700368a EN

Page 26: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – SHORT / CONICAL / ANGULATEDCLIFF LYND, MD

HOME

26

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 27: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

INTRA-OP TYPE I ENDOLEAKS

UC201700368a EN

Page 28: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – INTRA-OP TYPE IA ENDOLEAKELIE SEMAAN, MD & MARCUS D’AYALA, MD, BROOKLYN, NY

Post balloon

Type 1

HOME

28

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 29: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – INTRA-OP TYPE IA ENDOLEAKRIPAL GANDHI, MD, BARRY KATZEN, MD, MIAMI, FL

Pre-op 2 months post opPersisting type 1 after

repeat ballooning 6 EndoAnchorTM implants

HOME

29

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 30: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – INTRA-OP TYPE IA ENDOLEAKCOLIN D. BICKNELL, MD, LONDON, UK

EndoAnchorTM fixation resolves endoleak

1 year post-op follow-up shows no complications

Type 1a endoleak in angulated aortic neck

HOME

30

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 31: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – INTRA-OP TYPE IA ENDOLEAKJAMES JOYE, DO, MOUNTAIN VIEW, CA

HOME

31

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 32: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – INTRA-OP TYPE IA ENDOLEAKMANISH MEHTA, MD, PHILIP PATY, MD, NEW YORK

HOME

32

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 33: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – INTRA-OP TYPE IA ENDOLEAKMANISH MEHTA, MD, PHILIP PATY, MD, NEW YORK

HOME

33

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 34: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – INTRA-OP TYPE IA ENDOLEAKMAZIN FOTEH MD, AUSTIN, TEXAS

HOME

34

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Short, conical proximal neck highlights potential for complications

Intra-operative type Iaendoleak observed

EndoAnchor™ fixation improves graft alignment and seals endoleak

Post-op CT confirms aneurysm exclusion

Page 35: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – INTRA-OP TYPE IA ENDOLEAKJEAN-PAUL DE VRIES MD, PHD, NIEUWEGEIN, HOLLAND

HOME

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Short, conical proximal neck highlights potential for complications

Intra operative type 1a endoleak identified

EndoAnchorTM implants seal endoleak and enhance durability of proximal seal

1 month post op CT confirms aneurysm exclusion

Page 36: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY TEVAR – SHORT PROXIMAL/DISTAL NECKSPIOTR KASPRZAK MD, REGENSBURG, GERMANY

Final Angio, leak resolved

Type 1a leak after

graft implantation

4 EndoAnchorTM implants

deployed

Initial angioHOME

36

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 37: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY TEVAR – INTRA-OP TYPE IA ENDOLEAKFIRAS MUSSA, MD, NYU LANGONE, NEW YORK, NEW YORK

HOME

37

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 38: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

LATE TYPE I ENDOLEAKS

UC201700368 a EN

Page 39: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

REVISION EVAR – LATE TYPE IA ENDOLEAKERIC VERHOEVEN, MD, PHD, NUREMBERG, GERMANY

Type 1a Endoleak Endoleak persists after cuff and 4 EndoAnchorTM implants

2 more EndoAnchorTM implants and endoleak is resolved Post op CT

HOME

39

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 40: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

REVISION EVAR – LATE TYPE IA ENDOLEAKFRED BEAVERS, MD, CLINTON, MARYLAND

HOME

40

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 41: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

REVISION EVAR – LATE TYPE IA ENDOLEAKMOHAMMAD RAZA, MD, CHESTER, PENNSYLVANIA

HOME

41

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 42: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

REVISION EVAR – LATE TYPE IA ENDOLEAKKRASSI IVANCEV, MD, LONDON, UK

HOME

42

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 43: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

REVISION EVAR – LATE TYPE IA ENDOLEAKDITTMAR BÖCKLER, MD, MHBA, HEIDELBERG, GERMANY

EndoAnchorTM fixation after cuff implantation

Endoleak resolvedLate type I endoleak

HOME

43

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 44: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

REVISION EVAR – LATE TYPE IA ENDOLEAKJEAN-PAUL DE VRIES MD, PHD, NIEUWEGEIN, HOLLAND

HOME

44

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Late type I endoleak EndoAnchorTM fixation after cuff implantation

Endoleak resolved

Page 45: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

REVISION TEVAR – LATE TYPE IA/IB ENDOLEAKSNEIL HALIN, DO, TUFTS UNIVERSITY

HOME

45

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 46: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

REVISION TEVAR – LATE TYPE IB ENDOLEAKJOSH BERNHEIM, MD, DANIEL CHAR, MD, RIDGEWOOD, NJ

HOME

46

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 47: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

REVISION TEVAR – LATE TYPE IB ENDOLEAKFRANK ARKO, MD, CHARLOTTE, NORTH CAROLINA

HOME

47

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 48: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

ENDOGRAFT MIGRATION

UC201700368a EN

Page 49: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

REVISION EVAR – ENDOGRAFT MIGRATIONERIC VERHOEVEN, MD, PHD, NUREMBERG, GERMANY

HOME

49

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 50: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

REVISION EVAR – ENDOGRAFT MIGRATIONERIC VERHOEVEN, MD, PHD, NUREMBERG, GERMANY

Cuff and EndoAnchorTM

implants deployedEndoleak resolvedLate endograft migration

with Type Ia endoleak

HOME

50

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 51: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

REVISION TEVAR – ENDOGRAFT MIGRATIONCOLIN BICKNELL, MD, MOHAMAD HAMADY, MD, IMPERIAL COLLEGE, LONDON, UK

Initial Angio Post graft deployment Post deployment of 4 EndoAnchorTM implants

HOME

51

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 52: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

RUPTURES

UC201700368a EN

Page 53: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY EVAR – RUPTURED AAATIMOTHY NYPAVER, DETROIT, MICHIGAN

Type Ia endoleak upon graft implantation

EndoAnchorTM fixation resolves endoleak

Ruptured 10cm diameter AAA

HOME

53

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 54: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

REVISION EVAR – TYPE I ENDOLEAK IN RUPTURED AAABART MUHS, MD, PHD, NEW HAVEN, CONNECTICUT

9 EndoAnchorTM implants deployed in proximal neck

Endoleak resolvedPersisting type Ia endoleak post rEVAR

HOME

54

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 55: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

FEVAR

UC201700368 a EN

Page 56: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY FEVARDARREN SCHNEIDER, MD, NEW YORK PRESBYTERIAN

HOME

56

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 57: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

PRIMARY FEVAR – INTRA-OP TYPE IA ENDOLEAKMAZIN FOTEH, MD, AUSTIN, TEXAS

HOME

57

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

FEVAR chosen for patient with 10mm long conical proximal neck

Intra-operative type Ia endoleak observed, likely due to non-alignment of graft with aorta

EndoAnchor™ fixation employed to target source of type Ia endoleak and establish aneurysm exclusion

Page 58: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

58

Indications for Use: The Heli-FX™ EndoAnchor™ system is intended to provide fixation and sealing between endovascular aortic grafts and the native artery. The Heli-FXTM EndoAnchorTM system is indicated for use in patients whose endovascular grafts have exhibited migration or endoleak, or are at risk of such complications, in whom augmented radial fixation and/or sealing is required to regain or maintain adequate aneurysm exclusion. The EndoAnchorTM implant may be implanted at the time of the initial endograft placement, or during a secondary (i.e. repair) procedure.

Contraindications: Treatment with the Heli-FX™ EndoAnchor™ system is contraindicated for use in the following circumstances:• In patients with known allergies to the EndoAnchor™ implant material (MP35N-LT)• In conjunction with the Endologix Powerlink™* endograft

Warnings:• The long term performance of the EndoAnchor™ implant has not been established. All patients should be advised

endovascular aneurysm treatment requires long-term, regular follow-up to assess the patient’s health status and endograft performance, and the EndoAnchor™ implant does not reduce this requirement.

• The EndoAnchor™ implant, Heli-FX™ EndoAnchor™ system and Heli-FX™ Thoracic EndoAnchor™ system have been evaluated via in vitro testing and determined to be compatible with the Cook Zenith™*, Cook Zenith TX2™*, Gore Excluder™*, Gore TAG™, Jotec E™-vita abdominal, Jotec E™-vita thoracic, Medtronic AneuRx™, Medtronic Endurant™, Medtronic Talent™ and Medtronic Valiant™ endografts. Use with endografts other than those listed above has not been evaluated.

• The performance of the EndoAnchor™ implant has not been evaluated for securing multiple endograft components to one another. Not securing EndoAnchor™ implants into aortic tissue could result in graft fabric damage, component separation, and resultant Type III endoleaks.

• The performance of the EndoAnchor™ implant has not been evaluated in vessels other than the aorta. Use of the EndoAnchor™ implant to secure endografts to other vessels may result in adverse patient consequences such as vascular perforation, bleeding, or damage to adjacent structures.

• The performance of the EndoAnchor™ implant has not been evaluated for securing multiple anatomical structures together. Such use could result in adverse patient consequences such as vascular perforation, bleeding, or embolic events.

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 59: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

59

MRI Safety and Compatibility:• The EndoAnchor™ implants have been determined to be MR Conditional at 3T or less when the scanner is in Normal Operating Mode with whole body averaged SAR of 2 W/kg, or in First Level Controlled Mode with a maximum whole body averaged SAR of 4 W/kg.

• Please refer to documentation provided by the endograft system manufacturer for MR safety status of the endograft system with which the EndoAnchor™ implants are being used.

Potential Adverse Events: Possible adverse events associated with the Heli-FX™ EndoAnchor™ system include, but are not limited to:• Aneurysm rupture • Death • EndoAnchor embolization • Endoleaks (Type III) • Enteric fistula • Failure to correct/prevent Type I endoleak• Failure to prevent endograft migration • Infection • Renal complications (renal artery occlusion/dissection or contrast-induced AKI) • Stroke • Surgical conversion to open repair • Vascular access complications, including infection, pain, hematoma, pseudoaneurysm, arteriovenous fistula • Vessel damage, including dissection, perforation, and spasm

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17

Page 60: CASE BOOK: ENDOANCHORTM FIXATION · • The Heli-FX™ EndoAnchor™ systems are intended to provide fixation and sealing between endovascular aortic grafts and the native artery

60

Please reference product Instructions for Use for more information regarding indications, warnings, precautions, contraindications and adverse events.

CAUTION: Federal (USA) law restricts these devices to sale by or on the order of a licensed healthcare practitioner. See package inserts for full product information.

CAUTION: EndoAnchor™ implant locations should be based upon a detailed examination of the preoperative CT imaging in cases involving irregular or eccentric plaque in the intended sealing zone(s). EndoAnchor™ implants should be implanted only into areas of aortic tissue free of calcified plaque or thrombus, or where such pathology is diffuse and less than 2mm in thickness. Attempting to place EndoAnchor™ implants into more severe plaque or thrombus may be associated with implantation difficulty and suboptimal endograft fixation and/or sealing.

This therapy is not for everyone. Please consult your physician. A prescription is required. For further information, please call Medtronic at +1.888.283.7868.

UC201700368a EN ©2017 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further.Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. For distribution in the USA only. 04/17