CARDIAC production tests G. Auriemma, M. Martino & C. Satriano

CARDIAC production tests G. Auriemma, M. Martino & C. Satriano

Embed Size (px)

Citation preview

Page 1: CARDIAC production tests G. Auriemma, M. Martino & C. Satriano

CARDIAC production tests

G. Auriemma, M. Martino & C. Satriano

Page 2: CARDIAC production tests G. Auriemma, M. Martino & C. Satriano

G. Auriemma PRR FEB Roma 15/7/2005 2

Objectives

• Prepare an “user friendly” automatic test bench to be used at the Cardiac production site

• Check functionality of the boards• Operate a quality selection of FEB

(if necessary)• Generate a database for setup &

future reference

Page 3: CARDIAC production tests G. Auriemma, M. Martino & C. Satriano

G. Auriemma PRR FEB Roma 15/7/2005 3

Type of tests which can be performed

• Check of the current absorption• Check of the connectivity from the

signal inputs to the LVDS digital output• Measurement of sensitivity, noise and

bias at different Cdet• Measurement of the time jitter• Check of the functionality of the on

board test system

Page 4: CARDIAC production tests G. Auriemma, M. Martino & C. Satriano

G. Auriemma PRR FEB Roma 15/7/2005 4

Schematic of the test bench

Page 5: CARDIAC production tests G. Auriemma, M. Martino & C. Satriano

G. Auriemma PRR FEB Roma 15/7/2005 5

The test bench

Page 6: CARDIAC production tests G. Auriemma, M. Martino & C. Satriano

G. Auriemma PRR FEB Roma 15/7/2005 6

The injector

Page 7: CARDIAC production tests G. Auriemma, M. Martino & C. Satriano

G. Auriemma PRR FEB Roma 15/7/2005 7

A sample S-curve

•Raw data are obtained as S-curves at fixed threshold•The value of the Qin @ Vth is evaluated by a gaussian fit of this curve •Noise is evaluated from the variance of the gaussian

Page 8: CARDIAC production tests G. Auriemma, M. Martino & C. Satriano

G. Auriemma PRR FEB Roma 15/7/2005 8

The panel of the virtual test instrument

Page 9: CARDIAC production tests G. Auriemma, M. Martino & C. Satriano

G. Auriemma PRR FEB Roma 15/7/2005 9

Preliminary results

• We have tested 100 Cardiac in 8 hours (51 positives and 49 negatives)

• 20 of this were draining an excessive current from the LV supply (>1 Amp)

Page 10: CARDIAC production tests G. Auriemma, M. Martino & C. Satriano

G. Auriemma PRR FEB Roma 15/7/2005 10

Sensitivity @ Cdet=15 pF

Negative Positive

Page 11: CARDIAC production tests G. Auriemma, M. Martino & C. Satriano

G. Auriemma PRR FEB Roma 15/7/2005 11

Noise @ Cdet=15 pF

Negative Positive

Page 12: CARDIAC production tests G. Auriemma, M. Martino & C. Satriano

G. Auriemma PRR FEB Roma 15/7/2005 12

Stability of the measurements

Page 13: CARDIAC production tests G. Auriemma, M. Martino & C. Satriano

G. Auriemma PRR FEB Roma 15/7/2005 13

Conclusions

• The test bench has been demonstrated to be fast, reliable and accurate

• A new rugged injiector board with 4 card slots is under construction

• The testing of 100 boards in a normal workshifts is feasible

• Limits on the acceptable parameter’s range will be implemented in the software

• Test data will be saved in a database similar to the one used for the MWPC