Cardiac applications of 16 slice MDCT
:Initial experience
Samsung Medical CenterDong-Joon Ahn
Purpose
• Applications of 16 slice MDCT
in cardiac disease
• To evaluate the effectiveness :
CAC , CTA , Functional analysis
Materials & Equipments
• Lightspeed ultra 16 (GE)
• Advantage workstation version 4.0 (GE)
• Millenia 3500 CT ECG monitoring system
(Invivo rsearch Inc.)
• Envision CT (Medrad)
• Iomeron 300 (Braco co.)
• CAC : 8 , CTA : 28
Method
– No contrast material
– 2.5mm x 8i scan (cine mode)
– Gantry rotation time : 0.5sec
– Prospective ECG gating : 75%
Calcium scoring
II. Evaluation
I. Image acquisition
– Agatstone score
Coronary arteries angiography
– Non-ionic CM 120 - 150ml, 4ml/s
– 0.625mm scan
– Gantry rotation time : 0.5sec
– Retrospective ECG gating
– Image reconstruction according to
cardiac
phases : 5 - 95% of RR interval
Method
I. Image acquisition
II. Evaluation
– Compare 45%, 55% & 75% of RR interval
Reconstruction method & heart rate
H R Acquisition
Mode
Temporal
Resolution
40~60 Snapshot
Segment
250ms, 1cycle
60~75 Snapshot Burst 125ms, 2cycles
75~90 Snapshot Burst
Plus
65ms, 4cycles
Reconstruction method
Snapshot segment mode
Snapshot burst mode
Snapshot burst plus mode
Post processing methods
• Volume rendering
• Maximum intensity projection (MIP)
• Reformation in cardiac short and long axis
• Vessel analysis
• Endoscopic view (fly-through)
• Curved reformation
Method
Results
Agatscore
0 in 4 14 in 1 55 in 1 821 in 1 928 in 1
I. Calcium scoring
• Best phase locations for image quality
– 45% : RCA in 7, distal LAD in 5
– 55% : RCA in 1
– 75% : LAD in most patients and RCA in 2
• good image quality in 33/36 (91%)
II.Coronary arteries angiographyResults
LAD: 75% = Best
45%
55%
75%
Phase location and image quality
45% 55%
75%65%
Short and long axis of heart
Volume renderingCoronary vessel tree
view
LIMA - SVG Artifact
Automatic MIP MPVR
Curved reformation
Consideration
• Improved image reconstruction
algorithm and software
• Prospective or retrospective ECG gating
• Better temporal resolution
• Image acquisition of isotropic resolution
Limitations of coronary CTA
• Extensive calcifications
• Stents : spatial resolution
• Variable heart rate : poor image quality
• Radiation dose
• Small branches / septal branches
– Screening asymptomatic high risk population (CAC, CTA)
– Exclusion of stenosis in patients with low likelihood of extensive disease
– Diagnosis in pts with atypical angina
– Post-procedural evaluation (CABG, stent)
– Plaque characterization
– Follow-up after drug treatment
Conclusion (I)
Conclusion (II)
• The fast volume coverage of ECG gated
16 slice CT enables acquisition of the
entire heart volume with nearly
isotropic resolution within a single
breath hold.
Hr 100-105
Cardiac CT
• EKG EKG 동조화동조화
• Temporal resolution Temporal resolution 향상향상
• Isotropic resolutionIsotropic resolution영상획득영상획득
• SoftwareSoftware
Cardiac functionCardiac function
Short and long axis of heart
Before Before AfterAfter
Phase registrationPhase registration
Vessel Vessel analysisanalysis
Volume Volume renderingrendering
Isotropic resolutionIsotropic resolution영상획득영상획득
RCA Stent HR = 75-47-75RCA Stent HR = 75-47-75
Y- graft :RITA-LAD, Y- graft :RITA-LAD,
RITA-OM1-OM2-PLRITA-OM1-OM2-PL
45% volume rendering, HR: 45% volume rendering, HR: 75-8475-84