4
Calibration Report Job number: IND201100148 Report number: IND200900105/4 Date: 06/07/2011 Pg 1 of 2 Customer Mr. Concrete Kodigamam road, Puloly South Point Pedro. Unit under Test Digital Caliper Make Serial no. /10 no. Range Division Precise Not Available / MT031 0-150 mm 0.01 mm Method of test Test was carried out refered to the international standard ISO 6906 : 1984. Uncertainty calculations were done Refered to EA-4/02, NABL 141 Standard Laboratory working standard Grade 1 gauge blocks set no 090013 traceable to NIST via 821/276375 - 08, PTB Via 4937 PTB 06 Temperature and Humidity meter: SN:1 0745936 Place of test performed : SGS Lanka Calibration Lab Results of Calibration Temperature during test Note: The calliper reading was checked at three different positions of the measuring jaw,viz. tip, mid portion, and the base. Gauge setting Caliper reading, mm mm Tip Mid portion Base Zero setting 0.00 20.00 20.00 20.00 20.01 40.00 40.00 40.00 40.00 50.00 50.00 50.01 50.00 70.00 69.99 69.99 70.00 100.00 99.99 99.99 100.00 120.00 119.99 120.00 120.00 150.00 150.00 150.00 150.00 (Company No. PV 19226) SGS Lanka 1M Ltd Industrial Services, AEC Building, 140, Vauxhall Street, Colombo, 2, Sri Lanka, t + 94 (0) 11 5395118 f + 94 (0) 11 4717878 e [email protected] www.sgs.com Head Office, 3< Floor, 140, Vauxhall Street, Colombo, 2, Sri Lanka, t + 94 (0) 11 5376280 f + 94 (0) 11 5376281 e [email protected] www.sgs.com Member of the SGS Group

Calibration & Measurement Report 0001

Embed Size (px)

Citation preview

Page 1: Calibration & Measurement Report 0001

Calibration Report

Job number: IND201100148 Report number: IND200900105/4 Date: 06/07/2011 Pg 1 of 2

Customer Mr. ConcreteKodigamam road, Puloly SouthPoint Pedro.

Unit under Test Digital CaliperMakeSerial no. /10 no.RangeDivision

PreciseNot Available / MT0310-150 mm0.01 mm

Method of testTest was carried out refered to the international standard ISO 6906 : 1984. Uncertainty calculations were doneRefered to EA-4/02, NABL 141

StandardLaboratory working standard Grade 1 gauge blocks set no 090013 traceable to NIST via 821/276375 - 08,PTB Via 4937 PTB 06

Temperature and Humidity meter: SN:1 0745936

Place of test performed : SGS Lanka Calibration Lab

Results of Calibration

Temperature during test

Note: The calliper reading was checked at three different positions of the measuring jaw,viz. tip, mid portion,

and the base.

Gauge setting Caliper reading, mmmm Tip Mid portion Base

Zero setting 0.00

20.00 20.00 20.00 20.01

40.00 40.00 40.00 40.00

50.00 50.00 50.01 50.00

70.00 69.99 69.99 70.00100.00 99.99 99.99 100.00

120.00 119.99 120.00 120.00150.00 150.00 150.00 150.00

(Company No. PV 19226)

SGS Lanka 1M Ltd Industrial Services, AEC Building, 140, Vauxhall Street, Colombo, 2, Sri Lanka, t + 94 (0) 11 5395118 f + 94 (0) 11 4717878 e [email protected] www.sgs.com

Head Office, 3< Floor, 140, Vauxhall Street, Colombo, 2, Sri Lanka, t + 94 (0) 11 5376280 f + 94 (0) 11 5376281 e [email protected] www.sgs.comMember of the SGS Group

Page 2: Calibration & Measurement Report 0001

Measurement ReportJob number: IND201100148 Report number: IND2011000143/3 Date: 06/0712011 Pg 1 of 2

Customer Mr. ConcreteKodigamam road, Puloly SouthPoint Pedro.

Unit under Test Steel RuleMake - EloraSerial no./ID no. : -/MT037

RangeDivision

0- 1000 m1 mm

Method of testThe accuracy of this instrument was checked in accordance with Linear measurements were madeby Linear Scale.

StandardThis certificate provides traceability of measurement to national standards through NIST Japan andIndustrial Technology Institute Sri Lanka

Place of test performed: Condition room at the SGS Lanka Pvt. Ltd.

Results of Calibration

Temperature during test 20 ·C

Zero Error: Present or Zero line missing

MeasuredRange 1 (mm) Value 1 (mm)

0-50 500-100 1000-150 1500-200 2000-250 2500-300 3000-350 3500-400 4000-450 4500-500 499

MeasuredRange 1 Value 1(mm) (mm)0-550 5490-600 5990-650 6490-700 6990-750 7490-800 7990-850 8490-900 8990-950 9490-1000 998

(Company No. PV 19226)

SCS Lanka IPvtI Uti Industrial Services, AEC Building, 140, Vauxhall Street, Colombo - 2, Sri Lanka, t + 94 (0) 11 5395118 f + 94 (0) 114717878 e [email protected] WWW.sgs.comHead Office, 3" Floor, 140, Vauxhall Street, Colombo - 2, Sri Lanka, t + 94 (0) 11 5376280 f + 94 (0) 115376281 e [email protected] www.sgs.com

Member of the SGS Group

Page 3: Calibration & Measurement Report 0001

Measurement ReportJob number: IND201100148 Date: 06/07/2011 Pg 1 of 2Report number: IND2011000143/2

Customer

Unit under Test

Mr. ConcreteKodigamam road, Puloly SouthPoint Pedro.

Steel TapeMake -JasonSerial no.llD no. : -/MT035

RangeDivision

0-7.5 m1 mm

Method of testThe accuracy of this instrument was checked in accordance with Linear measurements were madeby Linear Scale.

StandardThis certificate provides traceability of measurement to national standards through NIST Japan andIndustrial Technology Institute Sri Lanka

Place of test performed:

Results of Calibration

Condition room at the SGS Lanka Pvt. Ltd.

Temperature during test

Zero Error:

MeasuredRange 1 Value 1(mm) (mm)0-100 1000-200 2000-300 3000-400 4000-500 5000-600 6000-700 7000-800 8000-900 9000-1000 1000

Width ofthe Tape: 24 mm

20 DC

none

Rangel Measured(m) Value I (mm)

1 - 1.100 1001 - 1.200 2001 - 1.300 3001 - 1.400 4001 - 1.500 5001 - 1.600 6001 - 1.700 7001 -1.800 8001 -1.900 9001 - 2.000 1000

MeasuredRange I (m) Value 1 (m)

2-3 1.0003-4 1.0004-5 1.0005-6 1.0006-7 1.000

7 - 7.5 0.500

(Company No. PV 19226)

SGS Lanka [Pvtl Ltd Industrial Services, AEC Building, 140, Vauxhall Street, Colombo - 2, Sri Lanka, t + 94 (0) 11 5395118 f + 94 (0) 11 4717878 e sgs,lanka@sgs,com www.sgs.com

Head Office, 3" Floor, 140, Vauxhall Street, Colombo - 2, Sri Lanka, t + 94 (0) 11 5376280 f + 94 (0) 11 5376281 e sgs.lanka@sgs,com www.sgs.com

Member of the SGS Group

Page 4: Calibration & Measurement Report 0001

Measurement ReportJob number: IND201100148 Date: 06/07/2011 Pg 1 of 3Report number: IND2011000143/1

Customer

Unit under Test

Mr. ConcreteKodigamam road, Puloly SouthPoint Pedro.

Fiberglass TapeMake - TSB - 50 MSerial no./ID no. : -/MT034

RangeDivision

0- 50 m2mm

Method of test IThe accuracy of this irhlstrument was checked in accordance with Linear measurements were madeby Linear Scale. Weight is applide only to remove wrinkles

StandardThis certificate provides traceability of measurement to national standards through NIST Japan andIndustrial Technology Institute Sri Lanka

Place of test performed:

Results of Calibration

Condition room at the SGS Lanka Pvt. Ltd.

Temperature during test

Zero Error:

Range 1 Value 1(mm) (mm)0-10 1000-20 2000-30 3000-40 4000-50 4990-60 5990-70 6990-80 7990-90 8990-100 998

20 ·C

none

Range 1 Measured(m) Value 1 (mm)

1 - 1.100 1001 - 1.200 2001 - 1.300 3001 - 1.400 4001 - 1.500 5001 - 1600 6001 - 1.700 7001 - 1.800 8001 - 1.900 9001 - 2.000 1000

(Company No. PV 19226)

C I b 2 S . L nka t + 94 (0) 11 5395118 I + 94 (0) 114717878 e [email protected] www.sgs.comI d t' IS' es AEC Building 140 Vauxhall Street, oom 0 - , n a ,

SGS Lanka 1M) Ltd HnudsrOlaffiCee~~cFI~or 140 vau~all s;reet, Colombo _2. Sri Lanka, t + 94 (0) 115376280 f + 94 (0) 115376281 e [email protected].

ea , " Member of Ire SGS Group