25
TESTNCAL LABORATORY has been assessed and accredited in accordance with the standard ISO/IEC 17025:2017 ''General Requirements for the Competence of Testing & Calibration Laboratories'' for its facilities at PLOT NO-113, POCKET-A2, SECTOR-17,DWARKA, NEW DELHI, DELHI, INDIA in the field of CALIBRATION Certificate Number: CC-2055 Issue Date: 12/04/2019 Valid Until: 11/04/2021 This certificate remains valid for the Scope of Accreditation as specified in the annexure subject to continued satisfactory compliance to the above standard & the relevant requirements of NABL. (To see the scope of accreditation of this laboratory, you may also visit NABL website www.nabl-india.org) Signed for and on behalf of NABL N. Venkateswaran Chief Executive Officer

CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

TESTNCAL LABORATORYhas been assessed and accredited in accordance with the standard

ISO/IEC 17025:2017''General Requirements for the Competence of Testing &

Calibration Laboratories''for its facilities at

PLOT NO-113, POCKET-A2, SECTOR-17,DWARKA, NEW DELHI, DELHI, INDIA

in the field of

CALIBRATIONCertificate Number: CC-2055

Issue Date: 12/04/2019 Valid Until: 11/04/2021

This certificate remains valid for the Scope of Accreditation as specified in the annexure subject to continuedsatisfactory compliance to the above standard & the relevant requirements of NABL.

(To see the scope of accreditation of this laboratory, you may also visit NABL website www.nabl-india.org)

Signed for and on behalf of NABL

N. VenkateswaranChief Executive Officer

Page 2: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

Permanent Facility

1 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current @ 50Hz 1 A to 10 A 0.065% to 0.055% Using 61/2 Digit DMMby Direct/ ComparisonMethod

2 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current @ 50Hz 1 mA to 400 mA 0.16% to 0.77% Using 61/2 Digit DMMby Direct/ ComparisonMethod

3 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current @ 50Hz 10 µA to 100 µA 1.0% to 0.63% Using 61/2 Digit DMMby Direct/ ComparisonMethod

4 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current @ 50Hz 100 µA to 1mA 0.093% to 0.09% Using 61/2 Digit DMMby Direct/ ComparisonMethod

5 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current @ 50Hz 400 mA to 1 A 0.77% to 0.065% Using 61/2 Digit DMMby Direct/ ComparisonMethod

6 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Resistance 1 ohm to 100 kohm 1.04% to 1.01% Using LCR-Q Meter byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 1 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 3: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

7 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Voltage @ 50Hz 1 mV to 100 mV 0.17% to 0.1% Using 61/2 Digit DMMby Direct/ ComparisonMethod

8 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Voltage @ 50Hz 1 V to 100 V 0.1% to 0.1% Using 61/2 Digit DMMby Direct/ ComparisonMethod

9 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Voltage @ 50Hz 100 mV to 1 V 0.1% to 0.1% Using 61/2 Digit DMMby Direct/ ComparisonMethod

10 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Voltage@ 50Hz 100 V to 1000 V 0.1% to 0.1% Using 61/2 Digit DMMby Direct/ ComparisonMethod

11 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

Capacitance 1µF to 100µF 4.13% to 2.01% Using LCR-Q Meter byDirect Method

12 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

Inductance 1 mH to 10 H 2.16% to 0.68% Using LCR-Q Meter byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 2 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 4: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

13 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC High Current @50Hz

100 A to 900 A 3.55% Using Zeal MultiFunction Calibrator,51/2 Digit with CurrentCoil by Direct Method

14 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

DC High Current 100 A to 900 A 2.35% to 1.74% Using Zeal MultiFunction Calibrator,51/2 Digit with CurrentCoil by Direct Method

15 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

Resistance (2-Wire) 200 M Ohm to 200 GOhm

1.42% to 5.15% Using High ResistanceBox, by Direct Method

16 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

Resistance (4-Wire) 0.001 Ohm to 10 Ohm 2.4% Using Low ResistanceBox, by Direct Method

17 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 1 A to 10 A 0.55% to 0.26% Using 61/2 Digit DMMby Direct/ ComparisonMethod

18 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 1 mA to 400 mA 0.21% to 0.07% Using 61/2 Digit DMMby Direct/ ComparisonMethod

19 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 10µA to 100µA 1.21% to 0.1% Using 61/2 Digit DMMby Direct/ ComparisonMethod

20 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 100µA to 1mA 0.1% to 0.21% Using 61/2 Digit DMMby Direct/ ComparisonMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 3 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 5: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

21 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 400 mA to 1 A 0.07% to 0.055% Using 61/2 Digit DMMby Direct/ ComparisonMethod

22 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Voltage 1 mV to 100 mV 0.40% to 0.40% Using 61/2 Digit DMMby Direct/ ComparisonMethod

23 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Voltage 100 mV to 1000 V 0.21% Using 61/2 Digit DMMby Direct/ ComparisonMethod

24 ELECTRO-TECHNICAL-MISCELLANEOUS(Measure)

Resistance 0.001 Ohm to 1 Ohm 1.86% to 1.12% Using 61/2 Digit DMMby Direct/ ComparisonMethod

25 ELECTRO-TECHNICAL-MISCELLANEOUS(Measure)

Resistance 1 ohm to 100 ohm 1.12% to 0.41% Using 61/2 Digit DMMby Direct/ ComparisonMethod

26 ELECTRO-TECHNICAL-MISCELLANEOUS(Measure)

Resistance (2 wire) 1 MOhm to 100 MOhm 0.31% to 0.095% Using 61/2 Digit DMMby Direct/ ComparisonMethod

27 ELECTRO-TECHNICAL-MISCELLANEOUS(Measure)

Resistance (2 wire) 100 ohm to 1 Mohm 0.41% to 0.31% Using 61/2 Digit DMMby Direct/ ComparisonMethod

28 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

J Type ThermocoupleIndicator Controller/Recorder

-200°C to 1050°C 1.08°C Using ProcessCalibrator by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 4 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 6: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

29 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

K Type ThermocoupleIndicator Controller/Recorder

0°C to 1350°C 1.4°C Using ProcessCalibrator by DirectMethod

30 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

RTD (PT-100) IndicatorController/ Recorder

-200°C to 700°C 1.4°C Using ProcessCalibrator by DirectMethod

31 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

RType ThermocoupleIndicator Controller/Recorder

0°C to 1750°C 1.42°C Using ProcessCalibrator by DirectMethod

32 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

B Type ThermocoupleIndicatorcontroller/Recorder

100°Cto 1000°C

1.79 °C Using ProcessCalibrator by DirectMethod

33 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

J Type ThermocoupleIndicatorcontroller/Recorder

-200°C to 1050°C 1.44°C Using ProcessCalibrator by DirectMethod

34 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

K Type ThermocoupleIndicatorcontroller/Recorder

0°C to 1350°C 1.19°C Using ProcessCalibrator by DirectMethod

35 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

RTD (PT-100) Indicatorcontroller/Recorder

-200°C to 700°C 1.38°C Using ProcessCalibrator by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 5 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 7: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

36 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Frequency 10 Hz to 1 kHz 0.82% to 0.053% Using 61/2 Digit DMMby Direct/ ComparisonMethod

37 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Time 1 s to 3600 s 0.42s to 5s Using Time IntervalMeter by ComparisonMethod

38 MECHANICAL-ACCELERATION ANDSPEED

RPM Meter,Stroboscope, DigitalTachometer, Centrifuse(Non Contact Type)LC:- 0.1RPM

10 RPM to 1000 RPM 0.45% Using DigitalTachometer byComparison Method

39 MECHANICAL-ACCELERATION ANDSPEED

RPM Meter,Stroboscope, DigitalTachometer, Centrifuse(Non Contact Type)LC:- 1RPM

1000 RPM to 35000RPM

0.19% Using DigitalTachometer byComparison Method

40 MECHANICAL-ACOUSTICS

Sound Level Meter 94 dB to 114 dB 1.34dB Using Sound LevelCalibrator byComparison Method

41 MECHANICAL-DENSITY ANDVISCOSITY

Hydrometers , SpecificGravity Hydrometers,Lactometers, BrixHydrometers, SoilHydrometer

0.650 g/ml to 1.200g/ml

0.0011g/ml Using StandardHydrometer & Liquidsof known Densities

42 MECHANICAL-DENSITY ANDVISCOSITY

Hydrometers , SpecificGravity Hydrometers,Lactometers, BrixHydrometers, SoilHydrometer

1.200 g/ml to 1.800g/ml

0.0011g/ml Using StandardHydrometer & Liquidsof known Densities

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 6 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 8: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

43 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Depth Gauge L.C: 0.01mm

0 mm to 150 mm 11.6µm Using “0” Grade GaugeBlock Set &accessories byComparison method

44 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Depth Micro Meter L.C:0.01 mm

0 mm to 150 mm 11.6µm Using “0” Grade GaugeBlock Set &accessories byComparison method

45 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Dial Thickness GaugeL.C: 0.01 mm

0 mm to 25 mm 8.0µm Using “0” Grade GaugeBlock Set &accessories byComparison method

46 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

External Micro MeterLC: 0.001mm

0 mm to 100 mm 2.0µm Using “0” Grade GaugeBlock Set &accessories byComparison method

47 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

External Micro MeterLC: 0.01mm

100 mm to 300 mm 9.0µm Using “0” Grade GaugeBlock Set &accessories byComparison method

48 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Feeler Gauge 0 mm to 2 mm 11.0µm Using Digital ExternalMicrometer byComparison method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 7 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 9: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

49 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Height Gauge LC:0.01mm

0 mm to 300 mm 12µm Using “0” Grade GaugeBlock Set &accessories ,Surfaceplate by Comparisonmethod

50 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Snap Gauge 3 mm to 100 mm 3.0µm Using “0” Grade GaugeBlock Set &accessories byComparison method

51 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Vernier/Dial/DigitalCalipers LC: 0.01mm

0 mm to 300 mm 10µm Using “0” Grade GaugeBlock Set &accessories byComparison method

52 MECHANICAL-MOBILE FORCEMEASURING SYSTEM

Digital / Analog PushPull Gauge , ForceGauge

100 N to 1000 N 0.51% Using NewtonianWeights & FrameFixture & Hangers

53 MECHANICAL-MOBILE FORCEMEASURING SYSTEM

Digital / Analog PushPull Gauge, ForceGauge

1 N to 100 N 0.61% Using NewtonianWeights & FrameFixture & Hangers

54 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure(Dial / Digital PressureGauge, PressureTransmitter, PressureSwitch, PressureValue, Transducer

>70 bar to 700 bar 0.41bar Using Digital PressureGauge & PressureComparator byComparison Method asper DKD-R6-1

55 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure(Dial / Digital PressureGauge, PressureTransmitter, PressureSwitch, PressureValue, Transducer

0 bar to 70 bar 0.09bar Using Digital PressureGauge & PressureComparator byComparison Method asper DKD-R6-1

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 8 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 10: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

56 MECHANICAL-PRESSUREINDICATING DEVICES

Pneumatic Pressure(Dial / Digital PressureGauge, PressureTransmitter,Transducer

0 bar to 10 bar 0.011bar Using Digital PressureGauge & PressureComparator byComparison Method asper DKD-R6-1

57 MECHANICAL-PRESSUREINDICATING DEVICES

Vacuum Transmitter,Gauge (Digital / Dial)Vacuum Gauge,VacuumSwitch,Transducer

(-) 0.95 bar to 0 bar 0.022bar Using Digital PressureGauge by ComparisonMethod as per DKD-R6-1

58 MECHANICAL-VOLUME

Glassware, GlassPipette, Burette,Measuring Cylinder,Volume Flask,Beakers, MeasuringDispenser

1 ml to 25 ml 0.012 ml Using StandardWeights of Accuracyclass E2 and PrecisionBalances with andDistilled Water ofknown Density byGravimetric Method

59 MECHANICAL-VOLUME

Glassware, GlassPipette, Burette,Measuring Cylinder,Volume Flask,Beakers, MeasuringDispenser

25 ml to 150 ml 0.012 ml Using StandardWeights of Accuracyclass E2 and PrecisionBalances with andDistilled Water ofknown Density byGravimetric Method

60 MECHANICAL-VOLUME

Micropipette 100 µl to 10 ml 9µl Using StandardWeights of Accuracyclass E2 and PrecisionBalances with andDistilled Water ofknown Density byGravimetric Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 9 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 11: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

61 MECHANICAL-VOLUME

Micropipette 10µl to 100µl 0.3µl Using StandardWeights of Accuracyclass E2 and PrecisionBalances with andDistilled Water ofknown Density byGravimetric Method

62 MECHANICAL-VOLUME

Pipettes / Burrets /Measuring Flasks /Cylinders/Beakers ,Volumetric Measures/Glassware

150 ml to 1000 ml 1ml STD Weights &Electronics Balance

63 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing BalanceReadability 0.01mg &Coarser

1 g to 80 g 0.18mg Using StandardWeights of AccuracyClass E2

64 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing BalanceReadability 0.1g &Coarser

1 kg to 10 kg 0.71g Using StandardWeights of AccuracyClass F1 & F2 as perOIML R-76

65 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing BalanceReadability 0.1mg &Coarser

80 g to 200 g 0.60mg Using StandardWeights of AccuracyClass E2,

66 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing BalanceReadability 10 mg &Coarser

200 g to 5000 g 30mg Using StandardWeights of AccuracyClass E2 & F1

67 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing BalanceReadability 1g &Coarser

10 kg to 100 kg 3.7g Using StandardWeights of AccuracyClass F1 and F2

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 10 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 12: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

68 MECHANICAL-WEIGHTS

Mass (Accuracy classF2 & coarser)

10 g 0.04 mg Using E2 AccuracyClass StandardWeights with WeighingBalance (Readability:0.01 mg & 0.1mg) bysubstitution method(ABBA cycle) as perOIML R–111: 2004

69 MECHANICAL-WEIGHTS

Mass (Accuracy classF2 & coarser)

1 g 0.03mg Using E2 AccuracyClass StandardWeights with WeighingBalance (Readability:0.01 mg & 0.1mg) bysubstitution method(ABBA cycle) as perOIML R–111: 2004

70 MECHANICAL-WEIGHTS

Mass (Accuracy classF2 & coarser)

1 mg 0.03mg Using E2 AccuracyClass StandardWeights with WeighingBalance (Readability:0.01 mg & 0.1mg) bysubstitution method(ABBA cycle) as perOIML R–111: 2004

71 MECHANICAL-WEIGHTS

Mass (Accuracy classF2 & coarser)

10 mg 0.03mg Using E2 AccuracyClass StandardWeights with WeighingBalance (Readability:0.01 mg & 0.1mg) bysubstitution method(ABBA cycle) as perOIML R–111: 2004

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 11 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 13: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

72 MECHANICAL-WEIGHTS

Mass (Accuracy classF2 & coarser)

100 g 0.15mg Using E2 AccuracyClass StandardWeights with WeighingBalance (Readability:0.01 mg & 0.1mg) bysubstitution method(ABBA cycle) as perOIML R–111: 2004

73 MECHANICAL-WEIGHTS

Mass (Accuracy classF2 & coarser)

100 mg 0.03mg Using E2 AccuracyClass StandardWeights with WeighingBalance (Readability:0.01 mg & 0.1mg) bysubstitution method(ABBA cycle) as perOIML R–111: 2004

74 MECHANICAL-WEIGHTS

Mass (Accuracy classF2 & coarser)

2 g 0.03mg Using E2 AccuracyClass StandardWeights with WeighingBalance (Readability:0.01 mg & 0.1mg) bysubstitution method(ABBA cycle) as perOIML R–111: 2004

75 MECHANICAL-WEIGHTS

Mass (Accuracy classF2 & coarser)

2 mg 0.03mg Using E2 AccuracyClass StandardWeights with WeighingBalance (Readability:0.01 mg & 0.1mg) bysubstitution method(ABBA cycle) as perOIML R–111: 2004

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 12 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 14: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

76 MECHANICAL-WEIGHTS

Mass (Accuracy classF2 & coarser)

20 g 0.04mg Using E2 AccuracyClass StandardWeights with WeighingBalance (Readability:0.01 mg & 0.1mg) bysubstitution method(ABBA cycle) as perOIML R–111: 2004

77 MECHANICAL-WEIGHTS

Mass (Accuracy classF2 & coarser)

20 mg 0.03mg Using E2 AccuracyClass StandardWeights with WeighingBalance (Readability:0.01 mg & 0.1mg) bysubstitution method(ABBA cycle) as perOIML R–111: 2004

78 MECHANICAL-WEIGHTS

Mass (Accuracy classF2 & coarser)

200 g 0.20mg Using E2 AccuracyClass StandardWeights with WeighingBalance (Readability:0.01 mg & 0.1mg) bysubstitution method(ABBA cycle) as perOIML R–111: 2004

79 MECHANICAL-WEIGHTS

Mass (Accuracy classF2 & coarser)

200 mg 0.03mg Using E2 AccuracyClass StandardWeights with WeighingBalance (Readability:0.01 mg & 0.1mg) bysubstitution method(ABBA cycle) as perOIML R–111: 2004

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 13 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 15: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

80 MECHANICAL-WEIGHTS

Mass (Accuracy classF2 & coarser)

5 g 0.03mg Using E2 AccuracyClass StandardWeights with WeighingBalance (Readability:0.01 mg & 0.1mg) bysubstitution method(ABBA cycle) as perOIML R–111: 2004

81 MECHANICAL-WEIGHTS

Mass (Accuracy classF2 & coarser)

5 mg 0.03mg Using E2 AccuracyClass StandardWeights with WeighingBalance (Readability:0.01 mg & 0.1mg) bysubstitution method(ABBA cycle) as perOIML R–111: 2004

82 MECHANICAL-WEIGHTS

Mass (Accuracy classF2 & coarser)

50 g 0.10mg Using E2 AccuracyClass StandardWeights with WeighingBalance (Readability:0.01 mg & 0.1mg) bysubstitution method(ABBA cycle) as perOIML R–111: 2004

83 MECHANICAL-WEIGHTS

Mass (Accuracy classF2 & coarser)

50 mg 0.03mg Using E2 AccuracyClass StandardWeights with WeighingBalance (Readability:0.01 mg & 0.1mg) bysubstitution method(ABBA cycle) as perOIML R–111: 2004

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 14 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 16: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

84 MECHANICAL-WEIGHTS

Mass (Accuracy classF2 & coarser)

500 mg 0.03mg Using E2 AccuracyClass StandardWeights with WeighingBalance (Readability:0.01 mg & 0.1mg) bysubstitution method(ABBA cycle) as perOIML R–111: 2004

85 MECHANICAL-WEIGHTS

Weights 1 kg to 5 kg 80 mg std. weights &Electronic Balance

86 MECHANICAL-WEIGHTS

Weights 2 kg to 5 kg 26mg Using std. Weights &Electronic Balance

87 MECHANICAL-WEIGHTS

Weights 500 g to 1 kg 2mg std weights &Electronic Balance

88 THERMAL-TEMPERATURE

RTD , Thermocouplewith or withoutindicator, DigitalThermometer,Transmitter , DataLogger

20°C to 600°C 1.78°C Using 4 wire RTD with6.5 DMM byComparison Methodusing Dry BlockCalibrator

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 15 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 17: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

Site Facility

1 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current @ 50Hz 1 A to 10 A 0.065% to 0.055% Using 61/2 Digit DMMby Direct/ ComparisonMethod

2 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current @ 50Hz 1 mA to 400 mA 0.16% to 0.77% Using 61/2 Digit DMMby Direct/ ComparisonMethod

3 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current @ 50Hz 10 µA to 100 µA 1.0% to 0.63% Using 61/2 Digit DMMby Direct/ ComparisonMethod

4 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current @ 50Hz 100 µA to 1mA 0.093% to 0.09% Using 61/2 Digit DMMby Direct/ ComparisonMethod

5 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current @ 50Hz 400 mA to 1 A 0.77% to 0.065% Using 61/2 Digit DMMby Direct/ ComparisonMethod

6 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC High Voltage @ 50Hz

>1 kV to 5 kV 5.86% Using HV Probe withDMM by Direct Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 16 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 18: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

7 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Resistance 1 ohm to 100 kohm 1.04% to 1.01% Using LCR-Q Meter byDirect Method

8 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Voltage @ 50Hz 1 mV to 100 mV 0.17% to 0.1% Using 61/2 Digit DMMby Direct/ ComparisonMethod

9 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Voltage @ 50Hz 1 V to 100 V 0.1% to 0.1% Using 61/2 Digit DMMby Direct/ ComparisonMethod

10 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Voltage @ 50Hz 100 mV to 1 V 0.1% to 0.1% Using 61/2 Digit DMMby Direct/ ComparisonMethod

11 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Voltage@ 50Hz 100 V to 1000 V 0.1% to 0.1% Using 61/2 Digit DMMby Direct/ ComparisonMethod

12 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

Capacitance 1µF to 100µF 4.13% to 2.01% Using LCR-Q Meter byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 17 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 19: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

13 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

Inductance 1 mH to 10 H 2.16% to 0.68% Using LCR-Q Meter byDirect Method

14 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC High Current @50Hz

100 A to 900 A 3.55% Using Zeal MultiFunction Calibrator,51/2 Digit with CurrentCoil by Direct Method

15 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

DC High Current 100 A to 900 A 2.35% to 1.74% Using Zeal MultiFunction Calibrator,51/2 Digit with CurrentCoil by Direct Method

16 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

Resistance (2-Wire) 200 M Ohm to 200 GOhm

1.42% to 5.15% Using High ResistanceBox, by Direct Method

17 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

Resistance (4-Wire) 0.001 Ohm to 10 Ohm 2.4% Using Low ResistanceBox, by Direct Method

18 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 1 A to 10 A 0.55% to 0.26% Using 61/2 Digit DMMby Direct/ ComparisonMethod

19 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 1 mA to 400 mA 0.21% to 0.07% Using 61/2 Digit DMMby Direct/ ComparisonMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 18 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 20: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

20 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 10µA to 100µA 1.21% to 0.1% Using 61/2 Digit DMMby Direct/ ComparisonMethod

21 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 100µA to 1mA 0.1% to 0.21% Using 61/2 Digit DMMby Direct/ ComparisonMethod

22 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 400 mA to 1 A 0.07% to 0.055% Using 61/2 Digit DMMby Direct/ ComparisonMethod

23 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Voltage 1 mV to 100 mV 0.40% to 0.40% Using 61/2 Digit DMMby Direct/ ComparisonMethod

24 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Voltage 100 mV to 1000 V 0.21% Using 61/2 Digit DMMby Direct/ ComparisonMethod

25 ELECTRO-TECHNICAL-MISCELLANEOUS(Measure)

Resistance 0.001 Ohm to 1 Ohm 1.86% to 1.12% Using 61/2 Digit DMMby Direct/ ComparisonMethod

26 ELECTRO-TECHNICAL-MISCELLANEOUS(Measure)

Resistance 1 ohm to 100 ohm 1.12% to 0.41% Using 61/2 Digit DMMby Direct/ ComparisonMethod

27 ELECTRO-TECHNICAL-MISCELLANEOUS(Measure)

Resistance (2 wire) 1 MOhm to 100 MOhm 0.31% to 0.095% Using 61/2 Digit DMMby Direct/ ComparisonMethod

28 ELECTRO-TECHNICAL-MISCELLANEOUS(Measure)

Resistance (2 wire) 100 Mohm to 1 Gohm 0.095% to 1.89% Using 61/2 Digit DMMby Direct/ ComparisonMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 19 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 21: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

29 ELECTRO-TECHNICAL-MISCELLANEOUS(Measure)

Resistance (2 wire) 100 ohm to 1 Mohm 0.41% to 0.31% Using 61/2 Digit DMMby Direct/ ComparisonMethod

30 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

J Type ThermocoupleIndicator Controller/Recorder

-200°C to 1050°C 1.08°C Using ProcessCalibrator by DirectMethod

31 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

K Type ThermocoupleIndicator Controller/Recorder

0°C to 1350°C 1.4°C Using ProcessCalibrator by DirectMethod

32 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

RTD (PT-100) IndicatorController/ Recorder

-200°C to 700°C 1.4°C Using ProcessCalibrator by DirectMethod

33 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

RType ThermocoupleIndicator Controller/Recorder

0°C to 1750°C 1.42°C Using ProcessCalibrator by DirectMethod

34 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

B Type ThermocoupleIndicatorcontroller/Recorder

100°Cto 1000°C

1.79 °C Using ProcessCalibrator by DirectMethod

35 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

J Type ThermocoupleIndicatorcontroller/Recorder

-200°C to 1050°C 1.44°C Using ProcessCalibrator by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 20 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 22: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

36 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

K Type ThermocoupleIndicatorcontroller/Recorder

0°C to 1350°C 1.19°C Using ProcessCalibrator by DirectMethod

37 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

RTD (PT-100) Indicatorcontroller/Recorder

-200°C to 700°C 1.38°C Using ProcessCalibrator by DirectMethod

38 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Frequency 10 Hz to 1 kHz 0.82% to 0.053% Using 61/2 Digit DMMby Direct/ ComparisonMethod

39 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Time 1 s to 3600 s 0.42s to 5s Using Time IntervalMeter by ComparisonMethod

40 MECHANICAL-ACCELERATION ANDSPEED

RPM Meter,Stroboscope, DigitalTachometer, Centrifuse(Non Contact Type)LC:- 0.1RPM

10 RPM to 1000 RPM 0.45% Using DigitalTachometer byComparison Method

41 MECHANICAL-ACCELERATION ANDSPEED

RPM Meter,Stroboscope, DigitalTachometer, Centrifuse(Non Contact Type)LC:- 1RPM

1000 RPM to 35000RPM

0.19% Using DigitalTachometer byComparison Method

42 MECHANICAL-ACOUSTICS

Sound Level Meter 94 dB to 114 dB 1.34dB Using Sound LevelCalibrator byComparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 21 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 23: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

43 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure(Dial / Digital PressureGauge, PressureTransmitter, PressureSwitch, PressureValue, Transducer

>70 bar to 700 bar 0.41bar Using Digital PressureGauge & PressureComparator byComparison Method asper DKD-R6-1

44 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure(Dial / Digital PressureGauge, PressureTransmitter, PressureSwitch, PressureValue, Transducer

0 bar to 70 bar 0.09bar Using Digital PressureGauge & PressureComparator byComparison Method asper DKD-R6-1

45 MECHANICAL- UTM,TENSION CREEPAND TORSIONTESTING MACHINE

Uniaxial TestingMachine ,TensileTesting Machine,Compression TestingMachineCompressionCompression Mode

20 N to 1 kN 0.6% Using Force ProvingInstrument Class-1

46 MECHANICAL- UTM,TENSION CREEPAND TORSIONTESTING MACHINE

Uniaxial TestingMachine ,TensileTesting Machine,Compression TestingMachineCompressionTensionMode

20 N to 1 kN 1.02% Using Force ProvingInstrument Class-1

47 MECHANICAL- UTM,TENSION CREEPAND TORSIONTESTING MACHINE

Uniaxial TestingMachine ,TensileTesting Machine,Compression TestingMachineCompression

1 kN to 50 kN 0.82% Using Force ProvingInstrument Class-1 ofUTM.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 22 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 24: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

48 MECHANICAL- UTM,TENSION CREEPAND TORSIONTESTING MACHINE

Uniaxial TestingMachine ,TensileTesting Machine,Compression TestingMachineCompression

100 kN to 1000 kN 1.00% Using Force ProvingInstrument Class-1 ofUTM

49 MECHANICAL- UTM,TENSION CREEPAND TORSIONTESTING MACHINE

Uniaxial TestingMachine ,TensileTesting Machine,Compression TestingMachineTension

1 kN to 50 kN 0.88% Using Force ProvingInstrument Class-1 ofUTM.

50 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing BalanceReadability 0.01mg &Coarser

1 g to 80 g 0.18mg Using StandardWeights of AccuracyClass E2

51 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing BalanceReadability 0.1g &Coarser

1 kg to 10 kg 0.71g Using StandardWeights of AccuracyClass F1 & F2 as perOIML R-76

52 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing BalanceReadability 0.1mg &Coarser

80 g to 200 g 0.60mg Using StandardWeights of AccuracyClass E2,

53 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing BalanceReadability 10 mg &Coarser

200 g to 5000 g 30mg Using StandardWeights of AccuracyClass E2 & F1

54 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing BalanceReadability 1g &Coarser

10 kg to 100 kg 3.7g Using StandardWeights of AccuracyClass F1 and F2

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 23 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019

Page 25: CALIBRATION ISO/IEC 17025:2017 TESTNCAL LABORATORY...Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI, DELHI, INDIA Accreditation Standard ISO/IEC

S.No Discipline / Group Quantity Measured/Instrument

Range / Frequency * CalibrationMeasurementCapability(±)

Remarks

55 MECHANICAL-WEIGHTS

Mass (Accuracy classF2 & coarser)

10 g 0.04 mg Using E2 AccuracyClass StandardWeights with WeighingBalance (Readability:0.01 mg & 0.1mg) bysubstitution method(ABBA cycle) as perOIML R–111: 2004

56 THERMAL- SPECIFICHEAT & HUMIDITY

Humidity Indicator OfHumidityCalibrator/Generator,Humidity Chamber

10 %RH to 95 %RH@25°C

1.7% Using Digital RHIndicator with SensorSingle PositionCalibration atmeasuring location inDUC

57 THERMAL-TEMPERATURE

Temperature IndicatorOf Dry Block Furnace/Muffle Furnace

0 to 1200 °C 2.24 °C Using Digital R-TypeThermocouple with61/2 digit multimeterSingle PositionCalibration atmeasuring location inDUC

58 THERMAL-TEMPERATURE

Temperature IndicatorOf Liquid Bath, Oven,Autoclave(IndustrialPurpose only ), BODIncubator (IndustrialPurpose only ),EnvironmentalChamber, Refrigerator,Deep Freezer

-80°C to 350°C 0.48°C Using RTD Sensor with6 ½ Digit MultimeterSingle PositionCalibration atmeasuring location inDUC

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name TESTNCAL LABORATORY, PLOT NO-113, POCKET-A2, SECTOR-17,, NEW DELHI,DELHI, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2055 Page No. : 24 / 24

Validity 12/04/2019 to 11/04/2021 Last Amended on 07/06/2019