80
GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the standard ISO/IEC 17025:2017 ''General Requirements for the Competence of Testing & Calibration Laboratories'' for its facilities at PLOT NO. 126, POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA in the field of CALIBRATION Certificate Number: CC-2205 Issue Date: 17/01/2020 Valid Until: 17/05/2021 This certificate remains valid for the Scope of Accreditation as specified in the annexure subject to continued satisfactory compliance to the above standard & the relevant requirements of NABL. (To see the scope of accreditation of this laboratory, you may also visit NABL website www.nabl-india.org) Signed for and on behalf of NABL N. Venkateswaran Chief Executive Officer

CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

  • Upload
    others

  • View
    32

  • Download
    2

Embed Size (px)

Citation preview

Page 1: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORSGROUP

has been assessed and accredited in accordance with the standard

ISO/IEC 17025:2017''General Requirements for the Competence of Testing &

Calibration Laboratories''for its facilities at

PLOT NO. 126, POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

in the field of

CALIBRATIONCertificate Number: CC-2205

Issue Date: 17/01/2020 Valid Until: 17/05/2021

This certificate remains valid for the Scope of Accreditation as specified in the annexure subject to continuedsatisfactory compliance to the above standard & the relevant requirements of NABL.

(To see the scope of accreditation of this laboratory, you may also visit NABL website www.nabl-india.org)

Signed for and on behalf of NABL

N. VenkateswaranChief Executive Officer

Page 2: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

Permanent Facility

1 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current 1 kHz 100 µA to 200 µA 0.26% to 0.39% Using 6 ½ DigitMultimeter 8846A byDirect Method

2 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current 50 Hz 10 A to 3200 A 0.4% Using Standard CurrentTransformer with 6 ½Digit Multimeter 8846Aby Direct Method

3 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current 50 Hz to 1kHz

0.2 mA to 100 mA 0.39% to 0.18% Using 6 ½ DigitMultimeter 8846A byDirect Method

4 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current 50 Hz to 1kHz

100 mA to 10 A 0.18% to 0.29% Using 6 ½ DigitMultimeter 8846A byDirect Method

5 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 1 kHz to 5kHz

3 mA to 3 A 0.16% to 3.54% Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 1 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 3: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

6 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 50 Hz 20 A to 1000 A 0.79% Using MPF 5502 A withCurrent Coil by DirectMethod

7 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 50 Hz to 1kHz

10 A to 20 A 0.17% to 0.36% Using Multi ProductCalibrator 5502A byDirect Method

8 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 50 Hz to 1kHz

3 A to 10 A 0.16% to 0.17% Using Multi ProductCalibrator 5502A byDirect Method

9 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 50 Hz to 1KHz

30 uA to 3 mA 1.44% to 0.16% Using Multi ProductCalibrator 5502A byDirect Method

10 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 10 uA to 100 mA 0.36% to 0.064% Using 6 ½ DigitMultimeter 8846A byDirect Method

11 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 100 mA to 3 A 0.064% to 0.19% Using 6 ½ DigitMultimeter 8846A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 2 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 4: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

12 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 3 A to 10 A 0.19% Using 6 ½ DigitMultimeter 8846A byDirect Method

13 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 10 µA to 100 µA 0.26% to 0.044% Using Multi ProductCalibrator 5502A byDirect Method

14 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 100 uA to 3 mA 0.044% to 0.031% Using Multi ProductCalibrator 5502A byDirect Method

15 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 20 A to 1000 A 0.64% Using MPF 5502 AWith Current Coil byDirect Method

16 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 3 A to 20 A 0.11% to 0.38% Using Multi ProductCalibrator 5502A byDirect Method

17 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 3 mA to 3 A 0.031% to 0.11% Using Multi ProductCalibrator 5502A byDirect Method

18 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

CurrentTransformer(PhaseError)5A (Secondary)1% to 120% of ratedcurrent

1 A to 3200 A 3.34min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

19 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

CurrentTransformer(RatioError)5A (Secondary)1% to 120% of ratedcurrent

1 A to 3200 A 0.047% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 3 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 5: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

20 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC HIGH VOLTAGE 1 kV to 40 kV 3.2% Using HV Divider withKV Meter

21 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC High Voltage 50 Hz 1 kV to 5 kV 3.2% to 3.8% Using Udeyraj makeHV Divider withindicator by DirectMethod

22 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC High Voltage 50 Hz 5 kV to 25 kV 3.8% to 3.2% Using Udeyraj makeHV Divider withindicator by DirectMethod

23 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC Voltage 1 kHz 1 mV to 20 mV 4.72% to 0.33% Using 6 ½ DigitMultimeter 8846A byDirect Method

24 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC Voltage 50 Hz to 1kHz

20 mV to 1000 V 0.54% to 0.10% Using 6 ½ DigitMultimeter 8846A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 4 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 6: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

25 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Phase Error)1A(Secondary)From 1%to 120% of ratedcurrent

1 A to 3200 A 3.36min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

26 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Phase Error)1A/5ASecondary from 1% to120% of rated current

2000 A to 7500 A 3.4min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

27 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Phase Error)5A (Secondary)From 5% to120% of rated current

3000 A to 6000 A 3.36min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

28 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Phase Error)5A(Secondary)From 1%to 5% of rated current

3000 A to 6000 A 3.36min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

29 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Ratio Error)1A(Secondary)From 1%to 120% of ratedcurrent

1 A to 3200 A 0.05% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 5 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 7: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

30 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Ratio Error)1A/5ASecondary from 1% to120% of Rated Current

2000 A to 7500 A 0.085% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

31 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Ratio Error)5A (Secondary)From 5% to120% of rated current

3000 A to 6000 A 0.1% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

32 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Ratio Error)5A(Secondary)From 1%to 5% of rated current

3000 A to 6000 A 0.1% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

33 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC High Voltage 1 kV to 5 kV 3.2% Using Udeyraj makeHV Divider withindicator by DirectMethod

34 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC High Voltage 5 kV to 25 kV 3.2% Using Udeyraj makeHV Divider withindicator by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 6 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 8: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

35 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Resistance 1 mOhm to 1 Ohm 0.42% to 0.36% Using 6 ½ DigitMultimeter 8846A byV/I Method

36 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Resistance 1 MOhm to 1000MOhm

0.013% to 2.31% Using 6 ½ DigitMultimeter 8846A byDirect Method

37 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Resistance 1 Ohm to 1 MOhm 0.36% to 0.013% Using 6 ½ DigitMultimeter 8846A byDirect Method andVoltage/ CurrentMethod

38 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Voltage 1 mV to 100 mV 0.42% to 0.085% Using 6 ½ DigitMultimeter 8846A byDirect Method

39 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Voltage 100 mV to 1000 V 0.085% to 0.007% Using 6 ½ DigitMultimeter 8846A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 7 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 9: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

40 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Potential Transformer(Phase Error)

11000/110 4.03min Using StandardPotential transformer &Instrument transformertest set up ByComparison Method

41 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Potential Transformer(Phase Error)

11000/63.5 3.16min Using StandardPotential transformer &Instrument transformertest set up ByComparison Method

42 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Potential Transformer(Ratio Error)

11000/110 0.16% Using StandardPotential transformer &Instrument transformertest set up ByComparison Method

43 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Potential Transformer(Ratio Error)

11000/63.5 0.16% Using StandardPotential transformer &Instrument transformertest set up ByComparison Method

44 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Voltage Ratio 0.8 to 2021 ratio 0.16% Using 6 ½ DigitMultimeter 8846A andFluke 5502A by DirectMethod

45 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Power 50 [email protected] UPF 120/2240 V 0.1A to 20 A

1.2 W to 4800 W 3.47% to 0.23% Using Multi ProductCalibrator 5502A withPower Option by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 8 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 10: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

46 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Vertical deflectionHorizontal Deflection 1kHz

5m V to 50 V 2.64% to 0.40% Using Multi ProductCalibrator 5502A withScope Option by DirectMethod

47 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 1 kHz 2 mV to 10 mV 1.76% to 0.41% Using Multi ProductCalibrator 5502A byDirect Method

48 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 1 kHz to 10kHz

30 V to 1000 V 0.1% Using Multi ProductCalibrator 5502A byDirect Method

49 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 50 Hz to 1kHz

10 mV to 300 mV 0.41% to 0.044% Using Multi ProductCalibrator 5502A byDirect Method

50 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 50 Hz to 1kHz

300 mV to 300 V 0.044% to 0.1% Using Multi ProductCalibrator 5502A byDirect Method

51 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 50 Hz to 1kHz

300 V to 1000 V 0.1% Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 9 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 11: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

52 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Capacitance @ 1 kHz 10 pF to 1 nF 2.98% to 1.2% Using DecadeCapacitance Box &Multi Product Calibrator5502A by DirectMethod

53 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Capacitance @1 kHz 9 nF to 1 mF 1.26% to 0.80% Using DecadeCapacitance Box &Multi Product Calibrator5502A by DirectMethod

54 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Capacitance @1kHz 1 nF to 9 nF 1.7% to 1.26% Using DecadeCapacitance Box &Multi Product Calibrator5502A by DirectMethod

55 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Power @ 1 V to600 V, 1A to 20 A

1 W to 12 KW 0.37% to 0.25% Using MultiproductCalibrator 5502A withPower Option by DirectMethod.

56 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Resistance 1 mohm to 1 mohm 0.6% to 0.6% Using Shunt By DirectMethod

57 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Resistance 1 Ohm to 100 Ohm 0.25% to 0.072% Using Multi ProductCalibrator 5502A byDirect Method.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 10 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 12: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

58 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Resistance 100 kOhm to 100MOhm

0.021% to 0.58% Using Multi ProductCalibrator 5502A byDirect Method

59 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Resistance 100 ohm to 100 kOhm 0.072% to 0.021% Using Multi ProductCalibrator 5502A byDirect Method

60 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Voltage 1 mV to 100 mV 0.36% to 0.013% Using Multi ProductCalibrator 5502A byDirect Method.

61 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Voltage 100 mV to 30 V 0.013% to 0.0064% Using Multi ProductCalibrator 5502A byDirect Method.

62 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Voltage 30 V to 1000 V 0.0064% to 0.007% Using Multi ProductCalibrator 5502A byDirect Method.

63 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

High Resistance 100 kohm to 1 Tohm 1.2% to 2.4% Using DecadeResistance Box byDirect Method

64 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Inductance @1 kHz 1 mH to 10 H 1.53% to 2.03% Using DecadeInductance Box byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 11 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 13: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

65 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

ODD Harmonics(Current) upto 41st

100 mA to 6 A 0.75% Using FlukeMultiProduct Calibrator5502A

66 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

ODD Harmonics upto41st Voltage

100 mV to 60 V 0.75% Using FlukeMultiproduct Calibrator5502A

67 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Oscilloscope DCFunction

2 mV to 33 V 6.59% to 0.33% Using Multi ProductCalibrator 5502A withScope Option by DirectMethod

68 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Power Factor 50 Hz 0.1 to UPF 0.05PF Using Multi ProductCalibrator 5502A withPower Option by DirectMethod

69 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Voltage ratio 0.8 to 400 0.5% to 0.3% Using TransformerRatio Standard byComparison Method

70 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Voltage Ratio 400 to 2021 0.5% to 0.3% Using TransformerRatio Standard byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 12 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 14: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

71 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -RTD(Calibration oftemperature Indicator /Controller/ Recorder/Calibrator)

(-)200 °C to 800 °C 0.51°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

72 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)C-Type

0 °C to 2315 °C 1.4°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

73 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)E-Type

(-)270 °C to 1300 °C 0.37°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

74 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)J-Type

(-)200 °C to 1200 °C 0.35°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 13 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 15: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

75 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)K-Type

(-)250 °C to 1137 °C 0.47°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

76 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)L-Type

(-)200 °C to 600 °C 0.37°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

77 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)N-Type

(-)270 °C to 1300 °C 0.47°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

78 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)R-Type

(-)50 °C to 1767 °C 0.9°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 14 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 16: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

79 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)S-Type

(-)50 °C to 1767 °C 0.8°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

80 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)T-Type

(-)270 °C to 400 °C 0.35°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

81 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)U-Type

(-)200 °C to 600 °C 0.5°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

82 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -RTD(Calibration oftemperature Indicator /Controller/ Recorder/Calibrator)

(-) 200 °C to 800 °C 0.27°C Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 15 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 17: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

83 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)EType

(-)200 °C to 1000 °C 0.55°C Using Multi ProductCalibrator 5502A byDirect Method

84 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)JType

(-)210 °C to 1200 °C 0.37°C Using Multi ProductCalibrator 5502A byDirect Method

85 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)KType

(-)200 °C to 1372 °C 0.51°C Using Multi ProductCalibrator 5502A byDirect Method

86 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)NType

(-)200 °C to 1300 °C 0.61°C Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 16 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 18: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

87 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)RType

0 °C to 1700 °C 0.51°C Using Multi ProductCalibrator 5502A byDirect Method

88 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)SType

0 °C to 1700 °C 0.58°C Using Multi ProductCalibrator 5502A byDirect Method

89 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)TType

(-)200 °C to 400 °C 0.70°C Using Multi ProductCalibrator 5502A byDirect Method

90 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)UType

(-)200 °C to 600 °C 0.65°C Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 17 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 19: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

91 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)BType

600 °C to 1820 °C 0.55°C Using Multi ProductCalibrator 5502A byDirect Method

92 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)CType

200 °C to 2315 °C 0.95°C Using Multi ProductCalibrator 5502A byDirect Method

93 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Frequency 10 Hz to 1 MHz 0.084% to 0.014% Using 6 ½ DigitMultimeter 8846A byDirect Method

94 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Time 0.1 sec to 1 hr 0.013sec to 2.32sec Using Digital TimeInterval Meter By DirectMethod

95 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Time 1 hr to 24 hr 2.32sec to 5.18sec Using Digital TimeInterval Meter Bycomparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 18 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 20: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

96 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)

Bandwidth 50 MHz to 250 MHz 6.21% Using Multi ProductCalibrator 5502A withScope Option by DirectMethod

97 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)

Frequency 10 Hz to 300 kHz 0.086% to 0.0094% Using Multi ProductCalibrator 5502A byDirect Method

98 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)

Frequency 300 kHz to 2 MHz 0.0094% to 0.49% Using Multi ProductCalibrator 5502A byDirect Method

99 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)

Time Base 2 n Sec to 5 Sec 0.3% to 0.67% Using Multi ProductCalibrator 5502A withScope Option by DirectMethod

100 MECHANICAL-ACCELERATION ANDSPEED

Speed RPM - (Non-Contact)

10 RPM to 5000 RPM 2.1RPM Using Tachometer.

101 MECHANICAL-ACCELERATION ANDSPEED

Speed RPM - (Non-Contact)

5000 RPM to 30000RPM

19RPM Using Tachometer.

102 MECHANICAL-ACOUSTICS

Acoustic Pressure -Sound Level Meter @ 1kHz

114 db 1.1db Using Sound LevelCalibrator.

103 MECHANICAL-ACOUSTICS

Acoustic Pressure -Sound Level Meter @ 1kHz

94 db 0.96db Using Sound LevelCalibrator.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 19 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 21: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

104 MECHANICAL-DENSITY ANDVISCOSITY

Hydrometer - DensityHydrometer, TwaddleHydrometer, BaumeHydrometer, SpecificGravity Hydrometer,BrixHydrometer,Lactometer, Alcometer.

0.6 g/ml to 2.000 g/ml 0.00025g/ml Using WeighingBalance with resolution0.01 mg by Hydrostaticweighing method asper NISP SP 250-78Standard.

105 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Bevel ProtractorL.C.: 1' 0 ° to 360 ° 3.1arc min Using Angle GaugeBlock & Surface Plate.

106 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Bore Gauge(Transmission Only)L.C.: 1 µm

Up to 1 mm 3.1µm Using Dial CalibrationTester.

107 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Caliper(Vernier/Dial/Digital)L.C.:10 µm

1000 mm to 2000 mm 22.5µm Using Caliper Checker;Gauge Block; LengthBar & ExternalMicrometer.

108 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Caliper(Vernier/Dial/Digital)L.C.:10 µm

0 to 1000 mm 14µm Using Caliper Checker;Gauge Block; LengthBar & ExternalMicrometer.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 20 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 22: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

109 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Caliper(Vernier/Dial/Digital)L.C.:10 µm

0 to 600 mm 10µm Using Caliper Checker;Gauge Block; LengthBar & ExternalMicrometer.

110 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Coating ThicknessGauge(Digimatic/Analog)L.C.:0.1 µm

0 to 6 mm 3.55µm Using Thickness Foils.

111 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Comparator Stand 300 mm to 300 mm 3.7µm Using Gauge Block;Elec. Probe & SurfacePlate.

112 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Cube Mould 0 to 300 mm, 0 to 360° 16.8 µm, 3.6 min Using Profile Projector& Digital Verniercaliper, BevelProtractor byComparison Method

113 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Cylindrical MeasuringPin

0.5 mm to 25 mm 1.3µm Using Gauge Block Set& Electronic Probe.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 21 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 23: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

114 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Cylindrical SettingMasterDiameterVariation

>100 mm to 160 mm 3.3µm Using Gauge Block Set& Electronic Probe

115 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Cylindrical SettingMasterDiameterVariation

Up to 100 mm 2.8µm Using Gauge Block Set& Electronic Probe.

116 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Degree Protractor /Combination Set L.C.:1°

0 ° to 360 ° 20arc min Using Slip GaugeBlock, Sine Bar; AngleGauge Block & SurfacePlate.

117 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Depth Caliper(Vernier/Dial/Digital)L.C.: 10 µm

0 to 300 mm 8µm Using Gauge Block;Long Gauge Block &Surface Plate

118 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Depth MicrometerL.C. :1 µm

0 to 300 mm 4.0µm Using Gauge BlockSet; Long Gauge Block& Surface Plate.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 22 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 24: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

119 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Dial CalibrationTesterL.C.: 0.1 µm

0 to 25 mm 1.12µm Using Electronic Probe

120 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Dial ComparatorL.C.:0.5 µm

-50 µm to 50 µm 1.5µm Using Gauge Block Set& Comparator Stand.

121 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Dial ThicknessGaugeL.C.: 1 µm

0 to 10 mm 0.8µm Using Gauge BlockSet.

122 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Dial ThicknessGaugeL.C.: 10 µm

0 to 100 mm 8µm Using Gauge BlockSet.

123 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

ElectronicComparatorL.C.: 0.1µm

0 to 25 mm 0.9µm Using Gauge BlockSet.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 23 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 25: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

124 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Electronic HeightGaugeL.C.: 0.5 µm

0 to 600 mm 8µm Using Gauge Block;Long Gauge & SurfacePlate.

125 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

External Micrometer(Analog/Dial/Digital)L.C.: 1 µm

>25 mm to 300 mm 4.0µm Using Gauges BlockSets; Long GaugeBlock & Mic Check Set.

126 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

External Micrometer(Analog/Dial/Digital)L.C.: 1 µm

0 to 25 mm 0.8µm Using Mic.Check Set.

127 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

External Micrometer(Analog/Dial/Digital)L.C.: 10 µm

>300 mm to 600 mm 12.9µm Using Gauges BlockSets; Long GaugeBlock & Mic Check Set.

128 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

External Micrometer(Analog/Dial/Digital)L.C.: 10 µm

>600 mm to 1000 mm 14.0µm Using Gauges BlockSets; Long GaugeBlock & Mic Check Set.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 24 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 26: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

129 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

External Micrometer(Analog/Dial/Digital)L.C.: 10 µm

0 to 300 mm 6.9µm Using Gauges BlockSets; Long GaugeBlock & Mic Check Set.

130 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

External Micrometer(Digital)L.C.: 0.1µm

0 to 25 mm 0.4µm Using Gauge BlockSet.

131 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Feeler Gauge Up to 2 mm 1.5µm Using DigimaticMicrometer.

132 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

FlankinessGauge/ElongationGauge/Receiver Gauge

0 to 100 mm 16.8 µm Using Profile Projector& Digital VernierCaliper

133 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Height Gauge(Vernier/Dial/Digital)L.C.: 10 µm

0 to 1000 mm 20µm Using Caliper Checker;Gauge Block & SurfacePlate.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 25 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 27: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

134 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Height Gauge(Vernier/Dial/Digital)L.C.: 10 µm

0 to 600 mm 11µm Using Caliper Checker;Gauge Block & SurfacePlate.

135 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Inside Micrometer /Caliper TypeL.C. : 1µm

5 mm to 50 mm 3.0µm Using Gauge Block Set& Gauge BlockAccessories.

136 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Inside Micrometer /Caliper TypeL.C. : 10µm

5 mm to 300 mm 5.0µm Using Gauge Block Set& Gauge BlockAccessories.

137 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Inside Micrometer /Stick Micrometer (2points)Basic Travel ofMicrometer (Plain)L.C.:10 µm

50 mm to 63 mm 4.2µm Using Gauge BlockSet; Long Gauge BlockSet & Mic Check Set &Electronic Probe withComparator.

138 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Inside Micrometer /Stick Micrometer (2points)Overall LengthAccuracy withExtension Rod(Stick)L.C.: 10 µm

50 mm to 1050 mm 12.0µm Using Gauge BlockSet; Long Gauge BlockSet & Mic Check Set &Electronic Probe withComparator.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 26 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 28: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

139 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Internal CaliperGaugeL.C.: 10 µm

2.5 mm to 250 mm 7.1µm Using Gauge Block Set& Gauge BlockAccessories.

140 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Lever Dial GaugeL.C. :1 µm

Up to 0.14 mm 1.6µm Using Dial CalibrationTester.

141 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Lever Dial GaugeL.C. :10 µm

Up to 1 mm 5.9µm Using Dial CalibrationTester.

142 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Lever Dial GaugeL.C. :2 µm

Up to 2 mm 1.9µm Using Dial CalibrationTester.

143 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Limit Gauge/CDGauge/PCD Gauge

0 to 200 mm 16.8 µm Using Profile Projector,LMM & 2Height Gaugeby Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 27 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 29: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

144 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Measuring Scale 0 to 2000 mm 120 X Sq rt(L) µm, L inmtr.

Using Scale & TapeCalibration Unit

145 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Measuring Tape 0 to 100000 mm 120 X Sq rt(L) µm, L inmtr.

Using Scale & TapeCalibration Unit.

146 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Measuring Tape (PiTape)

0 to 15000 mm 120 X Sq rt(L) µm, L inmtr.

Using Scale & TapeCalibration Unit.

147 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Micrometer Setting Rod 300 mm to 1000 mm 10.0µm Using Gauge BlockSet; Long Gauge BlockSet & ElectronicComparator withProbe.

148 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Micrometer Setting Rod Up to 300 mm 4.5µm Using Gauge BlockSet; Long Gauge BlockSet & ElectronicComparator withProbe.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 28 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 30: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

149 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Parallel Thread PlugGauge

2 mm to 100 mm 4.1µm Using FCDM,Cylindrical SettingMaster & ThreadMeasuring Wires.

150 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Pistol CaliperL.C.: 50µm

Up to 100 mm 26.1µm Using Gauge BlockSet.

151 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Plain Plug Gauge 1 mm mm to 100 mm 2.0µm Using Gauge Block Set& Electronic Probe.

152 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Plain Plug Gauge 100 mm to 160 mm 5.0µm Using Gauge Block Set& Electronic Probe.

153 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Plain Ring Gauge 3 mm to 100 mm 2.0µm Using ULM & MasterRing.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 29 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 31: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

154 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Plunger DialGaugeL.C. : 1 µm

0 mm to 25 mm 1.7µm Using Dial CalibrationTester & Gauge BlockSet.

155 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Plunger DialGaugeL.C. : 1 µm

0 mm to 50 mm 2.2µm Using Dial CalibrationTester & Gauge BlockSet.

156 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Radius Gauge/RadiusTemplete

Up to 40 mm 15.2µm Using Profile Projector.

157 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Screw / Thread PitchGauge - Angle

55° & 60° 9.5Min of Arc Using Profile Projector

158 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Screw / Thread PitchGauge - Pitch

Up to 7 mm 12µm Using Profile Projector

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 30 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 32: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

159 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Sine Bar Up to 300 mm 9arc sec Using Gauge Block;Angle Gauge Block;Elec. Probe & SurfacePlate.

160 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Snap Gauge >100 mm to 200 mm 6.0µm Using Gauge BlockSet.

161 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Snap Gauge 1 mm to 100 mm 3.0µm Using Gauge Block Set

162 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Spirit / PrecisionLevelL.C.: 20 µm/m

Up to 300 mm 14.4µm/m Using Electronic Level.

163 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Surface Plate 1000 mm X 1000 mm 1.2 xv(L+W)/125 L&Win mm

Using Electronic Level,L.C.1µm/m.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 31 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 33: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

164 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Taper Scale 0.1 mm to 16 mm 13.6µm Using Profile Projector.

165 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Taper Thread PlugGauge

2 mm to 100 mm 4.1µm Using FCDM,Cylindrical SettingMaster & ThreadMeasuring Wires.

166 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Templates/Plain WorkPiece/Inspection JigandFixture/Moulds/LeverArm/ MasterConnecting Rod

0 to 600 mm, 0 to 360° 16.8 µm, 3.6 min Using Gauge Block set,2D Height Gauge,Profile Projector, LLM,Electronic Comparator,Digital Vernier Caliper,Digital Micrometer,Bevel Protractor, AngleGauge Set, Sine Bar,Gauge Block set andElectronic Comparatorby Comparison Method

167 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Test Sieves Up to 125 mm 13µm Using Profile Projector& Digital VernierCaliper

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 32 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 34: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

168 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Thickness Foils 0.01 mm to 12 mm 1.12µm Using Electronic Probe.

169 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Thread MeasuringWires

0.17 mm to 6.35 mm 1.3µm Using Electronic Probe.

170 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Thread Ring Gauge 3 mm to 100 mm 2.0µm Using ULM & MasterRing.

171 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Ultrasonic ThicknessGauge.L.C.: 0.01 mm

0 to 200 mm 72µm Using Steel MasterRod.

172 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Vickers/Knoop/Rockwell Diamond ConeIndenter

0 to 100 mm, 0 to 360° 16.8 µm, 3.6 min Using Profile Projector

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 33 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 35: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

173 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Weld Fillet Gauge Up to 100 mm 16.8 µm Using Profile Projectorby Comparison method

174 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Weld/Hi-Lo Gauge,Bridge Cam Gauge

0 to 60 mm, 0 to 60° 16.8 µm, 3.6 min Using Profile Projector,Gauge Block Set, Tapeand Scale calibrator byComparison method

175 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Width Gauge / HeightSetting Master

>100 mm to 160 mm 5.0µm Using Gauge Block Set& Electronic Probe.

176 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Width Gauge / HeightSetting Master

0 mm to 100 mm 2.0µm Using Gauge Block Set& Electronic Probe.

177 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Width Gauge/PaddleGauge

0 to 200 mm 16.8 µm Using Profile ProfileProjector, Gauge BlockSet, & Electronic Probe

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 34 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 36: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

178 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Wire Gauge 0.025 mm to 8 mm 12µm Using Profile Projector.

179 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Electronic LevelL.C.: 1µm/m

Up to 150 mm 7.2µm/m Using Electronic Level.

180 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Gauge Block - Carbide >25 mm to 50 mm 0.10µm Using Gauge Block Gr.’K” & Gauge BlockCalibrator.

181 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Gauge Block - Carbide >50 mm to 75 mm 0.13µm Using Gauge Block Gr.’K” & Gauge BlockCalibrator.

182 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Gauge Block - Carbide >75 mm to 100 mm 0.15µm Using Gauge Block Gr.’K” & Gauge BlockCalibrator.

183 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Gauge Block - Carbide 0.5 mm to 25 mm 0.09µm Using Gauge Block Gr.’K” & Gauge BlockCalibrator.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 35 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 37: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

184 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Gauge Block - Steel >25 mm to 50 mm 0.15µm Using Gauge Block Gr.’K” & Gauge BlockCalibrator.

185 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Gauge Block - Steel >50 mm to 75 mm 0.18µm Using Gauge Block Gr.’K” & Gauge BlockCalibrator.

186 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Gauge Block - Steel >75 mm to 100 mm 0.22µm Using Gauge Block Gr.’K” & Gauge BlockCalibrator.

187 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Gauge Block - Steel 0.5 mm to 25 mm 0.12µm Using Gauge Block Gr.’K” & Gauge BlockCalibrator.

188 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Length MeasuringMachine.L.C.:0.0001mm

0 to 100 mm 0.66µm Using Slip Gauge SetGrade 'k' byComparison Method.

189 MECHANICAL-DUROMETER

Rubber HardnessTester

0 to 100 Shore A 0.5Shore A Rubber hardnessCalibration (MicrometerHead with Fixture) ByDepth IndentationMethod as per ISO18898

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 36 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 38: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

190 MECHANICAL-DUROMETER

Rubber HardnessTester

0 to 100 Shore D 0.5Shore D Rubber hardnessCalibration (MicrometerHead with Fixture) ByDepth IndentationMethod as per ISO18898

191 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure -Dial/Digital PressureIndicating Device;PressureTransmitter/Switches.

0 bar to 1000 bar 0.12bar Using Digital PressureCalibrator & HydraulicComparator as perDKD- R - 6 -1

192 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure -Dial/Digital PressureIndicating Device;PressureTransmitter/Switches.

0 bar to 250 bar 0.05bar Using Digital PressureCalibrator & HydraulicComparator as perDKD- R - 6 -1

193 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure -Dial/Digital PressureIndicating Device;PressureTransmitter/Switches.

0 bar to 700 bar 0.12bar Using Digital PressureCalibrator & HydraulicComparator as perDKD- R - 6 -1

194 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure -Dial/Digital PressureIndicating Device;PressureTransmitter/Switches.

0 bar to 20 bar 5.6mbar Using Digital PressureCalibrator & HydraulicComparator as perDKD- R - 6 -1

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 37 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 39: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

195 MECHANICAL-PRESSUREINDICATING DEVICES

Negative Pressure -Dial / Digital VacuumGauge, PressureIndicating Devices.

0 mbar to -200 mbar 0.25mbar Using Digital PressureCalibrator with Vacuum/ Pneumatic HandPressure Pump byComparison MethodDKD- R - 6 -1

196 MECHANICAL-PRESSUREINDICATING DEVICES

Negative Pressure -Dial / Digital VacuumGauge, PressureIndicating Devices,PressureTransmitter/Switches.

0 bar to -0.95 bar 1.3mbar Using Digital PressureCalibrator with Vacuum/ Pneumatic HandPressure Pump byComparison MethodDKD- R - 6 -1

197 MECHANICAL-PRESSUREINDICATING DEVICES

Pneumatic Pressure -Dial / Digital PressureGauge, PressureIndicating Devices &PressureTransmitter/Switches

0 bar to 20 bar 5.6mbar Using Digital PressureCalibrator with Vacuum/ Pneumatic HandPressure Pump byComparison Method asper DKD- R - 6 -1

198 MECHANICAL-PRESSUREINDICATING DEVICES

Pneumatic Pressure -Dial / Digital PressurePressure IndicatingDevice & PressureTransmitter/Switches

0 bar to 2 bar 1mbar Using Digital PressureCalibrator with Vacuum/ Pneumatic HandPressure Pump byComparison Method asper DKD- R - 6 -1

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 38 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 40: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

199 MECHANICAL-PRESSUREINDICATING DEVICES

Pneumatic Pressure -Dial/Digital LowPressure IndicatingDevices & PressureTransmitter.

0 mbar to 200 mbar 0.25mbar Using DigitalManometer withVacuum / PneumaticHand Pressure Pumpby Comparison Methodas per DKD- R - 6 -1

200 MECHANICAL-PRESSUREINDICATING DEVICES

Pneumatic Pressure -Pressure IndicatingDevices(MaghnelicGauges, Low PressureGauges, Calibrators)

0 Pa to 500 Pa 0.35 Pa Using Digital PressureCalibrator byComparison MethodDKD- R - 6 -1

201 MECHANICAL-TORQUEMEASURINGDEVICES

Torque Screw Driver/Torque Wrench /Torque DialGaugesType I & II (AllClass)(Clock wise andanti-clock wise)

0.1 Nm to 10 Nm 1.3% Using ElectronicTorque WrenchCalibrator as per ISO6789:2003 / 2017

202 MECHANICAL-TORQUEMEASURINGDEVICES

Torque Screw Driver/Torque Wrench /Torque DialGaugesType I & II (AllClass)(Clock wise andanti-clock wise)

10 Nm to 200 Nm 1.3% Using ElectronicTorque WrenchCalibrator as per ISO6789:2003 / 2017

203 MECHANICAL-TORQUEMEASURINGDEVICES

Torque Screw Driver/Torque Wrench /Torque DialGaugesType I & II (AllClass)(Clock wise andanti-clock wise)

200 Nm to 1000 Nm 1.3% Using ElectronicTorque WrenchCalibrator as per ISO6789:2003 / 2017

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 39 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 41: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

204 MECHANICAL-VOLUME

Glass Pipettes &Burette

>10 ml to 100 ml 5µl Using Weighingbalance of Readability(0.01 / 0.1) mg &Distilled water byGravimetric methodbased on ISO 4787

205 MECHANICAL-VOLUME

Glass Pipettes &Burette

0.1 ml to 10 ml 2.4µl Using Weighingbalance of Readability(0.01 / 0.1) mg &Distilled water byGravimetric methodbased on ISO 4787

206 MECHANICAL-VOLUME

Glassware & Plasticwares (Such asMeasuring Cylinder /Volumetric Flask / Jar /Can etc.)

>10 ml to 100 ml 19.5µl Using Weighingbalance of Readability(0.01 / 0.1) mg &Distilled water byGravimetric methodbased on ISO 4787

207 MECHANICAL-VOLUME

Glassware & Plasticwares (Such asMeasuring Cylinder /Volumetric Flask / Jar /Can etc.)

>100 ml to 500 ml 21.3µl Using Weighingbalance of d:1 mg &Distilled water byGravimetric methodbased on ISO 4787

208 MECHANICAL-VOLUME

Glassware & Plasticwares (Such asMeasuring Cylinder /Volumetric Flask / Jar /Can etc.)

>2 L to 5 L 316.4µl Using Weighingbalance of d:10 mg &Distilled water byGravimetric methodbased on ISO 4787

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 40 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 42: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

209 MECHANICAL-VOLUME

Glassware & Plasticwares (Such asMeasuring Cylinder /Volumetric Flask / Jar /Can etc.)

>500 ml to 2 L 19.5µl to 249ul Using Weighingbalance of d:10 mg &Distilled water byGravimetric methodbased on ISO 4787

210 MECHANICAL-VOLUME

Glassware & Plasticwares (Such asMeasuring Cylinder /Volumetric Flask / Jar /Can etc.)

0.1 ml to 10 ml 2.9µl Using Weighingbalance of Readability(0.01 / 0.1) mg &Distilled water byGravimetric methodbased on ISO 4787

211 MECHANICAL-VOLUME

Micro pipettes (PistonOperated-Apparatus)

1 ml to 5 ml 4µl Using Weighingbalance of Readability(0.01 / 0.1) mg &Distilled water byGravimetric methodbased on ISO 8655-6

212 MECHANICAL-VOLUME

Micro pipettes(PistonOperated-Apparatus)

100 µl to 1000 µl 0.35µl Using Weighingbalance of Readability(0.01 / 0.1) mg &Distilled water byGravimetric methodbased on ISO 8655-6

213 MECHANICAL-VOLUME

Micro pipettes(PistonOperated-Apparatus)

10 µl to 100 µl 0.30µl Using Weighingbalance of Readability(0.01 / 0.1) mg &Distilled water byGravimetric methodbased on ISO 8655-6

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 41 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 43: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

214 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =0.01 g)

0 kg to 5 kg 0.1g Using Weights ofaccuracy class E2 & F1as per OIML R-76

215 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =0.01 mg)

0 g to 100 g 0.03mg Using Weights ofaccuracy class E2 asper OIML R-76

216 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =0.1 g)

0 kg to 30 kg 0.1g Using Weights ofaccuracy class E2 & F1as per OIML R-76

217 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =0.1 mg)

0 g to 200 g 0.10mg Using Weights ofaccuracy class E2 asper OIML R-76

218 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =1 mg)

0 kg to 1 kg 1.0mg Using Weights ofaccuracy class E2 asper OIML R-76

219 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F2 Class &Coarser

2 kg 8.6mg Using E2 class weightsand balance of d: 0.01g as per OIML R-111.Based on ABBAMethod

220 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F2 Class &Coarser

5 kg 9mg Using E2 class weightsand balance of d: 0.01g as per OIML R-111.Based on ABBAMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 42 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 44: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

221 MECHANICAL-WEIGHTS

Weights (ConventionalMass)M1 Class &Coarser

10 kg 80mg Using E2 & F1 classweights and balance ofd: 0.1 g as per OIML R-111. Based on ABBAMethod

222 MECHANICAL-WEIGHTS

Weights (ConventionalMass)M1 Class &Coarser

20 kg 100mg Using E2 & F1 classweights and balance ofd: 0.1 g as per OIML R-111. Based on ABBAMethod

223 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

1 g 0.02mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

224 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

1 kg 1.2mg Using E2 class weightsand balance of d: 0.001g as per OIML R-111.Based on ABBAMethod

225 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

10 g 0.02mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 43 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 45: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

226 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

10 mg 0.011mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod

227 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

100 g 0.08mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

228 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

100 mg 0.02mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod

229 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

2 g 0.02mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 44 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 46: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

230 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

2 mg 0.011mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod

231 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

20 g 0.05mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

232 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

20 mg 0.01mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod

233 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

200 g 0.15mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 45 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 47: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

234 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

5 g 0.02mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

235 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

5 mg 0.011mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod

236 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

50 g 0.05mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

237 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

50 mg 0.01mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod

238 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

500 g 1.0mg Using E2 class weightsand balance of d: 0.001g as per OIML R-111.Based on ABBAMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 46 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 48: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

239 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

500 mg 0.02mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

240 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F2 Class &Coarser

200 mg 0.02mg Using E2 classstandard weights andBalance of d: 0.01 mg /0.1 mg as per OIML R-111. Based on ABBAMethod.

241 MECHANICAL-WEIGHTS

Weights (ConventionalMass)F1 Class &Coarser

1 mg 0.011mg Using E2 classstandard weights andBalance of d: 0.01 mgas per OIML R-111.Based on ABBAMethod.

242 THERMAL-MISCELLANEOUS

Temperature Gauge 30°C to 250°C0.30°C Using PRT with

6.5DMM orMUltifunction Calibratorand Liquid bath bycomparison method

243 THERMAL-TEMPERATURE

Glass Thermometers -80 °C to 250 °C 0.24°C Using SPRT with 6.5DMM or Multifunctioncalibrator and Liquidbath by ComparisonMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 47 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 49: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

244 THERMAL-TEMPERATURE

Infrared/Non-contactType Thermometer

>300 °C to 650 °C 6.1°C Using RTD/N type/IRPyrometer Sensor withMultifunction calibratorAnd Black body sourceby Comparison Method

245 THERMAL-TEMPERATURE

Infrared/Non-contactType Thermometer

50 °C to 300 °C 5.2°C Using RTD/N type/IRPyrometer Sensor withMultifunction calibratorAnd Black body sourceby Comparison Method

246 THERMAL-TEMPERATURE

Temperature Gauge 0 °C to 600 °C 0.24°C Using SPRT with 6.5DMM or Multifunctioncalibrator and Liquidbath by ComparisonMethod

247 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block &Furnace.

>1000 °C to 1200 °C 2.44°C Using R- TypeThermocouple withMultifunction calibratorand Liquid bath /Dryblock by ComparisonMethod Single PositionCalibration

248 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block &Furnace.

>400 °C to 650 °C 0.16°C Using SPRT with 6.5DMM or Multifunctioncalibrator and Liquidbath /Dry block byComparison MethodSingle PositionCalibration

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 48 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 50: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

249 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block &Furnace.

>650 °C to 1000 °C 2.32°C Using R- TypeThermocouple withMultifunction calibratorand Liquid bath /Dryblock by ComparisonMethod Single PositionCalibration

250 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block/Liquid Bath / Incubators/ Oven / Furnace

>250 °C to 400 °C 0.21°C Using SPRT with 6.5DMM or Multifunctioncalibrator and Liquidbath /Dry block byComparison MethodSingle PositionCalibration

251 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block/Liquid Bath / Incubators/ Oven / Furnace

>50 °C to 250 °C 0.19°C Using SPRT with 6.5DMM or Multifunctioncalibrator and Liquidbath /Dry block byComparison MethodSingle PositionCalibration

252 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block/Liquid Bath / Incubators/ Oven / Furnace

-80 °C to 50 °C 0.22°C Using SPRT with 6.5DMM or Multifunctioncalibrator and Liquidbath /Dry block byComparison MethodSingle PositionCalibration

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 49 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 51: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

253 THERMAL-TEMPERATURE

Temperature sensor(RTD/ Thermocouple)With & withoutindicator, &Temperaturetransmitter

>1000 °C to 1200 °C 2.43°C Using R- TypeThermocouple withMultifunction calibrator& dry blockTemperature baths byComparison Method

254 THERMAL-TEMPERATURE

Temperature sensor(RTD/ Thermocouple)With & withoutindicator, &Temperaturetransmitter

>250 °C to 400 °C 0.21°C Using SPRT with6.5DMM orMultifunction calibratorwith Liquid/ Dry blockbath By ComparisonMethod

255 THERMAL-TEMPERATURE

Temperature sensor(RTD/ Thermocouple)With & withoutindicator, &Temperaturetransmitter

>400 °C to 650 °C 0.21°C Using SPRT with6.5DMM orMultifunction calibratorwith Liquid/ Dry blockbath By ComparisonMethod

256 THERMAL-TEMPERATURE

Temperature sensor(RTD/ Thermocouple)With & withoutindicator, &Temperaturetransmitter

>50 °C to 250 °C 0.19°C Using SPRT with6.5DMM orMultifunction calibratorwith Liquid/ Dry blockbath By ComparisonMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 50 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 52: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

257 THERMAL-TEMPERATURE

Temperature sensor(RTD/ Thermocouple)With & withoutindicator, &Temperaturetransmitter

>650 °C to 1000 °C 2.32°C Using R- TypeThermocouple withMultifunction calibrator& dry blockTemperature baths byComparison Method

258 THERMAL-TEMPERATURE

Temperature sensor(RTD/ Thermocouple)With & withoutindicator, &Temperaturetransmitter

-80 °C to 50 °C 0.22°C Using SPRT with6.5DMM orMultifunction calibratorwith Liquid/ Dry blockbath By ComparisonMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 51 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 53: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

Site Facility

1 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current 1 kHz 100 µA to 200 µA 0.26% to 0.39% Using 6 ½ DigitMultimeter 8846A byDirect Method

2 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current 50 Hz 10 A to 3200 A 0.4% Using Standard CurrentTransformer with 6 ½Digit Multimeter 8846Aby Direct Method

3 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current 50 Hz to 1kHz

0.2 mA to 100 mA 0.39% to 0.18% Using 6 ½ DigitMultimeter 8846A byDirect Method

4 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)

AC Current 50 Hz to 1kHz

100 mA to 10 A 0.18% to 0.29% Using 6 ½ DigitMultimeter 8846A byDirect Method

5 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 1 kHz to 5kHz

3 mA to 3 A 0.16% to 3.54% Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 52 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 54: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

6 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 50 Hz 20 A to 1000 A 0.79% Using MPF 5502 A withCurrent Coil by DirectMethod

7 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 50 Hz to 1kHz

10 A to 20 A 0.17% to 0.36% Using Multi ProductCalibrator 5502A byDirect Method

8 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 50 Hz to 1kHz

3 A to 10 A 0.16% to 0.17% Using Multi ProductCalibrator 5502A byDirect Method

9 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)

AC Current 50 Hz to 1KHz

30 uA to 3 mA 1.44% to 0.16% Using Multi ProductCalibrator 5502A byDirect Method

10 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 10 uA to 100 mA 0.36% to 0.064% Using 6 ½ DigitMultimeter 8846A byDirect Method

11 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 100 mA to 3 A 0.064% to 0.19% Using 6 ½ DigitMultimeter 8846A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 53 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 55: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

12 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)

DC Current 3 A to 10 A 0.19% Using 6 ½ DigitMultimeter 8846A byDirect Method

13 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 10 µA to 100 µA 0.26% to 0.044% Using Multi ProductCalibrator 5502A byDirect Method

14 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 100 uA to 3 mA 0.044% to 0.031% Using Multi ProductCalibrator 5502A byDirect Method

15 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 20 A to 1000 A 0.64% Using MPF 5502 AWith Current Coil byDirect Method

16 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 3 A to 20 A 0.11% to 0.38% Using Multi ProductCalibrator 5502A byDirect Method

17 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)

DC Current 3 mA to 3 A 0.031% to 0.11% Using Multi ProductCalibrator 5502A byDirect Method

18 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

CurrentTransformer(PhaseError)5A (Secondary)1% to 120% of ratedcurrent

1 A to 3200 A 3.34min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

19 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

CurrentTransformer(RatioError)5A (Secondary)1% to 120% of ratedcurrent

1 A to 3200 A 0.047% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 54 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 56: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

20 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC HIGH VOLTAGE 1 kV to 40 kV 3.2% Using HV Divider withKV Meter

21 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC High Voltage 50 Hz 1 kV to 5 kV 3.2% to 3.8% Using Udeyraj makeHV Divider withindicator by DirectMethod

22 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC High Voltage 50 Hz 5 kV to 100 kV 3.8% to 3.2% Using Udeyraj makeHV Divider withindicator by DirectMethod

23 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC Voltage 1 kHz 1 mV to 20 mV 4.72% to 0.33% Using 6 ½ DigitMultimeter 8846A byDirect Method

24 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

AC Voltage 50 Hz to 1kHz

20 mV to 1000 V 0.54% to 0.10% Using 6 ½ DigitMultimeter 8846A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 55 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 57: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

25 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Phase Error)1A(Secondary)From 1%to 120% of ratedcurrent

1 A to 3200 A 3.36min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

26 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Phase Error)1A/5ASecondary from 1% to120% of rated current

2000 A to 7500 A 3.4min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

27 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Phase Error)5A (Secondary)From 5% to120% of rated current

3000 A to 6000 A 3.36min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

28 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Phase Error)5A(Secondary)From 1%to 5% of rated current

3000 A to 6000 A 3.36min Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

29 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Ratio Error)1A(Secondary)From 1%to 120% of ratedcurrent

1 A to 3200 A 0.05% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 56 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 58: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

30 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Ratio Error)1A/5ASecondary from 1% to120% of Rated Current

2000 A to 7500 A 0.085% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

31 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Ratio Error)5A (Secondary)From 5% to120% of rated current

3000 A to 6000 A 0.1% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

32 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Current Transformer(Ratio Error)5A(Secondary)From 1%to 5% of rated current

3000 A to 6000 A 0.1% Using Standard Currenttransformer &Instrument transformertest set up ByComparison Method

33 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC High Voltage 1 kV to 5 kV 3.2% Using Udeyraj makeHV Divider withindicator by DirectMethod

34 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC High Voltage 5 kV to 40 kV 3.2% Using Udeyraj makeHV Divider withindicator by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 57 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 59: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

35 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Resistance 1 mOhm to 1 Ohm 0.42% to 0.36% Using 6 ½ DigitMultimeter 8846A byV/I Method

36 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Resistance 1 MOhm to 1000MOhm

0.013% to 2.31% Using 6 ½ DigitMultimeter 8846A byDirect Method

37 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Resistance 1 Ohm to 1 MOhm 0.36% to 0.013% Using 6 ½ DigitMultimeter 8846A byDirect Method andVoltage/ CurrentMethod

38 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Voltage 1 mV to 100 mV 0.42% to 0.085% Using 6 ½ DigitMultimeter 8846A byDirect Method

39 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

DC Voltage 100 mV to 1000 V 0.085% to 0.007% Using 6 ½ DigitMultimeter 8846A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 58 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 60: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

40 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Potential Transformer(Phase Error)

11000/110 4.03min Using StandardPotential transformer &Instrument transformertest set up ByComparison Method

41 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Potential Transformer(Phase Error)

11000/63.5 3.16min Using StandardPotential transformer &Instrument transformertest set up ByComparison Method

42 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Potential Transformer(Ratio Error)

11000/110 0.16% Using StandardPotential transformer &Instrument transformertest set up ByComparison Method

43 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Potential Transformer(Ratio Error)

11000/63.5 0.16% Using StandardPotential transformer &Instrument transformertest set up ByComparison Method

44 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)

Voltage Ratio 0.8 to 2021 ratio 0.16% Using 6 ½ DigitMultimeter 8846A andFluke 5502A by DirectMethod

45 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Power 50 [email protected] UPF 120/2240 V 0.1A to 20 A

1.2 W to 4800 W 3.47% to 0.23% Using Multi ProductCalibrator 5502A withPower Option by DirectMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 59 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 61: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

46 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Vertical deflectionHorizontal Deflection 1kHz

5m V to 50 V 2.64% to 0.40% Using Multi ProductCalibrator 5502A withScope Option by DirectMethod

47 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 1 kHz 2 mV to 10 mV 1.76% to 0.41% Using Multi ProductCalibrator 5502A byDirect Method

48 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 1 kHz to 10kHz

30 V to 1000 V 0.1% Using Multi ProductCalibrator 5502A byDirect Method

49 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 50 Hz to 1kHz

10 mV to 300 mV 0.41% to 0.044% Using Multi ProductCalibrator 5502A byDirect Method

50 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 50 Hz to 1kHz

300 mV to 300 V 0.044% to 0.1% Using Multi ProductCalibrator 5502A byDirect Method

51 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

AC Voltage 50 Hz to 1kHz

300 V to 1000 V 0.1% Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 60 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 62: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

52 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Capacitance @ 1 kHz 10 pF to 1 nF 2.98% to 1.2% Using DecadeCapacitance Box &Multi Product Calibrator5502A by DirectMethod

53 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Capacitance @1 kHz 9 nF to 1 mF 1.26% to 0.80% Using DecadeCapacitance Box &Multi Product Calibrator5502A by DirectMethod

54 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Capacitance @1kHz 1 nF to 9 nF 1.7% to 1.26% Using DecadeCapacitance Box &Multi Product Calibrator5502A by DirectMethod

55 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Power @ 1 V to600 V, 1A to 20 A

1 W to 12 KW 0.37% to 0.25% Using MultiproductCalibrator 5502A withPower Option by DirectMethod.

56 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Resistance 1 mohm to 1 mohm 0.6% to 0.6% Using Shunt By DirectMethod

57 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Resistance 1 Ohm to 100 Ohm 0.25% to 0.072% Using Multi ProductCalibrator 5502A byDirect Method.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 61 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 63: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

58 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Resistance 100 kOhm to 100MOhm

0.021% to 0.58% Using Multi ProductCalibrator 5502A byDirect Method

59 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Resistance 100 ohm to 100 kOhm 0.072% to 0.021% Using Multi ProductCalibrator 5502A byDirect Method

60 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Voltage 1 mV to 100 mV 0.36% to 0.013% Using Multi ProductCalibrator 5502A byDirect Method.

61 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Voltage 100 mV to 30 V 0.013% to 0.0064% Using Multi ProductCalibrator 5502A byDirect Method.

62 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

DC Voltage 30 V to 1000 V 0.0064% to 0.007% Using Multi ProductCalibrator 5502A byDirect Method.

63 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

High Resistance 100 kohm to 1 Tohm 1.2% to 2.4% Using DecadeResistance Box byDirect Method

64 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Inductance @1 kHz 1 mH to 10 H 1.53% to 2.03% Using DecadeInductance Box byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 62 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 64: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

65 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

ODD Harmonics(Current) upto 41st

100 mA to 6 A 0.75% Using FlukeMultiProduct Calibrator5502A

66 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

ODD Harmonics upto41st Voltage

100 mV to 60 V 0.75% Using FlukeMultiproduct Calibrator5502A

67 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Oscilloscope DCFunction

2 mV to 33 V 6.59% to 0.33% Using Multi ProductCalibrator 5502A withScope Option by DirectMethod

68 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Power Factor 50 Hz 0.1 to UPF 0.05PF Using Multi ProductCalibrator 5502A withPower Option by DirectMethod

69 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Voltage ratio 0.8 to 400 0.5% to 0.3% Using TransformerRatio Standard byComparison Method

70 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT (Source)

Voltage Ratio 400 to 2021 0.5% to 0.3% Using TransformerRatio Standard byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 63 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 65: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

71 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -RTD(Calibration oftemperature Indicator /Controller/ Recorder/Calibrator)

(-)200 °C to 800 °C 0.51°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

72 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)C-Type

0 °C to 2315 °C 1.4°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

73 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)E-Type

(-)270 °C to 1300 °C 0.37°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

74 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)J-Type

(-)200 °C to 1200 °C 0.35°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 64 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 66: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

75 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)K-Type

(-)250 °C to 1137 °C 0.47°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

76 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)L-Type

(-)200 °C to 600 °C 0.37°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

77 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)N-Type

(-)270 °C to 1300 °C 0.47°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

78 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)R-Type

(-)50 °C to 1767 °C 0.9°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 65 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 67: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

79 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)S-Type

(-)50 °C to 1767 °C 0.8°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

80 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)T-Type

(-)270 °C to 400 °C 0.35°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

81 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator /Controller/Recorder/Calibrator)U-Type

(-)200 °C to 600 °C 0.5°C Using MultifunctionCalibrator(BeamextoMC2) byDirect Method

82 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -RTD(Calibration oftemperature Indicator /Controller/ Recorder/Calibrator)

(-) 200 °C to 800 °C 0.27°C Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 66 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 68: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

83 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)EType

(-)200 °C to 1000 °C 0.55°C Using Multi ProductCalibrator 5502A byDirect Method

84 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)JType

(-)210 °C to 1200 °C 0.37°C Using Multi ProductCalibrator 5502A byDirect Method

85 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)KType

(-)200 °C to 1372 °C 0.51°C Using Multi ProductCalibrator 5502A byDirect Method

86 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)NType

(-)200 °C to 1300 °C 0.61°C Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 67 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 69: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

87 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)RType

0 °C to 1700 °C 0.51°C Using Multi ProductCalibrator 5502A byDirect Method

88 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)SType

0 °C to 1700 °C 0.58°C Using Multi ProductCalibrator 5502A byDirect Method

89 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)TType

(-)200 °C to 400 °C 0.70°C Using Multi ProductCalibrator 5502A byDirect Method

90 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)UType

(-)200 °C to 600 °C 0.65°C Using Multi ProductCalibrator 5502A byDirect Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 68 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 70: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

91 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)BType

600 °C to 1820 °C 0.55°C Using Multi ProductCalibrator 5502A byDirect Method

92 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)

TemperatureSimulation -Thermocouple(Calibration of temperatureIndicator / Controller/Recorder/ Calibrator)CType

200 °C to 2315 °C 0.95°C Using Multi ProductCalibrator 5502A byDirect Method

93 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Frequency 10 Hz to 1 MHz 0.084% to 0.014% Using 6 ½ DigitMultimeter 8846A byDirect Method

94 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Time 0.1 sec to 1 hr 0.013sec to 2.32sec Using Digital TimeInterval Meter By DirectMethod

95 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)

Time 1 hr to 24 hr 2.32sec to 5.18sec Using Digital TimeInterval Meter Bycomparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 69 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 71: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

96 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)

Bandwidth 50 MHz to 250 MHz 6.21% Using Multi ProductCalibrator 5502A withScope Option by DirectMethod

97 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)

Frequency 10 Hz to 300 kHz 0.086% to 0.0094% Using Multi ProductCalibrator 5502A byDirect Method

98 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)

Frequency 300 kHz to 2 MHz 0.0094% to 0.49% Using Multi ProductCalibrator 5502A byDirect Method

99 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)

Time Base 2 n Sec to 5 Sec 0.3% to 0.67% Using Multi ProductCalibrator 5502A withScope Option by DirectMethod

100 MECHANICAL-ACCELERATION ANDSPEED

Speed RPM - (Non-Contact)

10 RPM to 5000 RPM 2.1RPM Using Tachometer.

101 MECHANICAL-ACCELERATION ANDSPEED

Speed RPM - (Non-Contact)

5000 RPM to 30000RPM

19RPM Using Tachometer.

102 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Caliper(Vernier/Dial/Digital)L.C.:10 µm

1000 mm to 2000 mm 22.5µm Using Caliper Checker;Gauge Block; LengthBar & ExternalMicrometer.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 70 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 72: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

103 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Dial CalibrationTesterL.C.: 0.1 µm

0 to 25 mm 1.12µm Using Electronic Probe

104 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Electronic HeightGaugeL.C.: 0.5 µm

0 to 600 mm 8µm Using Gauge Block;Long Gauge & SurfacePlate.

105 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Height Gauge(Vernier/Dial/Digital)L.C.: 10 µm

0 to 1000 mm 20µm Using Caliper Checker;Gauge Block & SurfacePlate.

106 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Height Gauge(Vernier/Dial/Digital)L.C.: 10 µm

0 to 600 mm 11µm Using Caliper Checker;Gauge Block & SurfacePlate.

107 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

MetallurgicalMicroscope

100 x to 1000 x 0.3% Using Micro Scale ofL.C. 10 µm.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 71 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 73: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

108 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Straight Edge Upto 2000 mm 1.8 x v(L/125) µm L inmm

Using Electronic LevelL.C.1µm/m & GaugeBlock.

109 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)

Surface Plate 4000 mm X3000 mm 1.2 xv(L+W)/125 L&Win mm

Using Electronic Level,L.C.1 µm/m

110 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Length MeasuringMachine.L.C.:0.0001mm

0 to 100 mm 0.66µm Using Slip Gauge SetGrade 'k' byComparison Method.

111 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Profile Projector/VideoMeasuringMachineAngleL.C.:1sec

0 ° to 360 ° 6.0arc sec. Using AngularGraticule / AngleGauge Set.

112 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Profile Projector/VideoMeasuringMachineLinear : L.C.:0.1 µm

0 to 200 mm 4.8µm Using Glass Scale./Gauge Block.

113 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)

Profile Projector/VideoMeasuringMachineMagnification

Up to 100 X 1.2% Using Glass Scale &Digimatic Caliper.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 72 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 74: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

114 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure -Dial/Digital PressureIndicating Device;PressureTransmitter/Switches.

0 bar to 1000 bar 0.12bar Using Digital PressureCalibrator & HydraulicComparator as perDKD- R - 6 -1

115 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure -Dial/Digital PressureIndicating Device;PressureTransmitter/Switches.

0 bar to 250 bar 0.05bar Using Digital PressureCalibrator & HydraulicComparator as perDKD- R - 6 -1

116 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure -Dial/Digital PressureIndicating Device;PressureTransmitter/Switches.

0 bar to 700 bar 0.12bar Using Digital PressureCalibrator & HydraulicComparator as perDKD- R - 6 -1

117 MECHANICAL-PRESSUREINDICATING DEVICES

Hydraulic Pressure -Dial/Digital PressureIndicating Device;PressureTransmitter/Switches.

0 bar to 20 bar 5.6mbar Using Digital PressureCalibrator & HydraulicComparator as perDKD- R - 6 -1

118 MECHANICAL-PRESSUREINDICATING DEVICES

Negative Pressure -Dial / Digital VacuumGauge, PressureIndicating Devices.

0 mbar to -200 mbar 0.25mbar Using Digital PressureCalibrator with Vacuum/ Pneumatic HandPressure Pump byComparison MethodDKD- R - 6 -1

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 73 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 75: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

119 MECHANICAL-PRESSUREINDICATING DEVICES

Negative Pressure -Dial / Digital VacuumGauge, PressureIndicating Devices,PressureTransmitter/Switches.

0 bar to -0.95 bar 1.3mbar Using Digital PressureCalibrator with Vacuum/ Pneumatic HandPressure Pump byComparison MethodDKD- R - 6 -1

120 MECHANICAL-PRESSUREINDICATING DEVICES

Pneumatic Pressure -Dial / Digital PressureGauge, PressureIndicating Devices &PressureTransmitter/Switches

0 bar to 20 bar 5.6mbar Using Digital PressureCalibrator with Vacuum/ Pneumatic HandPressure Pump byComparison Method asper DKD- R - 6 -1

121 MECHANICAL-PRESSUREINDICATING DEVICES

Pneumatic Pressure -Dial / Digital PressurePressure IndicatingDevice & PressureTransmitter/Switches

0 bar to 2 bar 1mbar Using Digital PressureCalibrator with Vacuum/ Pneumatic HandPressure Pump byComparison Method asper DKD- R - 6 -1

122 MECHANICAL-PRESSUREINDICATING DEVICES

Pneumatic Pressure -Dial/Digital LowPressure IndicatingDevices & PressureTransmitter.

0 mbar to 200 mbar 0.25mbar Using DigitalManometer withVacuum / PneumaticHand Pressure Pumpby Comparison Methodas per DKD- R - 6 -1

123 MECHANICAL-PRESSUREINDICATING DEVICES

Pneumatic Pressure -Pressure IndicatingDevices(MaghnelicGauges, Low PressureGauges, Calibrators)

0 Pa to 500 Pa 0.35 Pa Using Digital PressureCalibrator byComparison MethodDKD- R - 6 -1

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 74 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 76: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

124 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =0.01 g)

0 kg to 5 kg 0.1g Using Weights ofaccuracy class E2 & F1as per OIML R-76

125 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =0.01 mg)

0 g to 100 g 0.03mg Using Weights ofaccuracy class E2 asper OIML R-76

126 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =0.1 g)

0 kg to 30 kg 0.1g Using Weights ofaccuracy class E2 & F1as per OIML R-76

127 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =0.1 mg)

0 g to 200 g 0.10mg Using Weights ofaccuracy class E2 asper OIML R-76

128 MECHANICAL-WEIGHING SCALEAND BALANCE

Weighing Balance(d =1 mg)

0 kg to 1 kg 1.0mg Using Weights ofaccuracy class E2 asper OIML R-76

129 THERMAL-MISCELLANEOUS

Temperature Gauge 30°C to 250°C0.30°C Using PRT with

6.5DMM orMUltifunction Calibratorand Liquid bath bycomparison method

130 THERMAL- SPECIFICHEAT & HUMIDITY

Humidity chamber/climatic chamber/Environmentalchamber

10 °C to 60 °C 0.31°C Using Temperature &humidity Data loggerswireless by Multi PointCalibration

131 THERMAL- SPECIFICHEAT & HUMIDITY

Humidity chamber/climatic chamber/Environmentalchamber @25°C

15% RH to 95% RH 2.44%RHRH Using Temperature &humidity Data loggerswireless by Multi PointCalibration

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 75 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 77: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

132 THERMAL-TEMPERATURE

Deep Freezer,Refrigerators,Incubator, Oven,Temperature Bath,Chamber, Furnace

-80 °C to 300 °C 2.37°C UsingRTD/Thermocouple(K/J/N/ other type)sensor with MultiChannel Data Loggermulti Point Calibration

133 THERMAL-TEMPERATURE

Glass Thermometers 30 °C to 250 °C 0.25°C Using PRT with6.5DMM orMultifunction calibratorAnd Liquid bath byComparison Method

134 THERMAL-TEMPERATURE

Infrared /Non-contactType Thermometer

>300 °C to 650 °C 6.1°C Using RTD/N type/IRPyrometer Sensor withMultifunction calibratorand Black body sourceBy Comparison Method

135 THERMAL-TEMPERATURE

Infrared /Non-contactType Thermometer

50 °C to 300 °C 5.1°C Using RTD/N type/IRPyrometer Sensor withMultifunction calibratorand Black body sourceBy Comparison Method

136 THERMAL-TEMPERATURE

Temperature Bath,Chamber & Furnace.

300 °C to 1200 °C 2.95°C UsingRTD/Thermocouple(K/J/N/ other type)sensor with MultiChannel Data Loggermulti Point Calibration

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 76 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 78: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

137 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block &Furnace.

>400 °C to 650 °C 1.70°C Using R- TypeThermocouple withMultifunction calibratorand Liquid bath /Dryblock by ComparisonMethod Single PositionCalibration

138 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block &Furnace.

>650 °C to 1000 °C 2.32°C Using R- TypeThermocouple withMultifunction calibratorand Liquid bath /Dryblock by ComparisonMethod Single PositionCalibration

139 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block/Furnace.

>1000 °C to 1200 °C 2.44°C°C Using R- TypeThermocouple withMultifunction calibratorand Liquid bath /Dryblock by ComparisonMethod Single PositionCalibration

140 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block/Liquid Bath / Incubators/ Oven / Furnace

>50 °C to 250 °C 0.30°C Using PRT with6.5DMM orMultifunction calibratorAnd Liquid bath /Dryblock by ComparisonMethod Single PositionCalibration

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 77 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 79: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

141 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block/Liquid Bath /Incubators.

-80 °C to 50 °C 0.25°C°C Using PRT with6.5DMM orMultifunction calibratorAnd Liquid bath /Dryblock by ComparisonMethod Single PositionCalibration

142 THERMAL-TEMPERATURE

Temperature Indicator/Controller with Sensorof Dry Well Block/Liquid Bath / Oven /Furnace

>250 °C to 400 °C 0.65°C°C Using PRT with6.5DMM orMultifunction calibratorAnd Liquid bath /Dryblock by ComparisonMethod Single PositionCalibration

143 THERMAL-TEMPERATURE

Temperature sensor(RTD/Thermocouple)With & without indicator &Temperaturetransmitter

>1000 °C to 1200 °C 2.44°C Using R- TypeThermocouple withMultifunction calibrator& dry blockTemperature baths andDry bath byComparison Method

144 THERMAL-TEMPERATURE

Temperature sensor(RTD/Thermocouple)With & without indicator &Temperaturetransmitter

>250 °C to 400 °C 0.65°C Using SPRT/PRT with6.5DMM orMultifunction calibratorand Liquid/Dry blockbath By ComparisonMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 78 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -

Page 80: CALIBRATION ISO/IEC 17025:2017 GROUP GODREJ & BOYCE … · 2019-07-24 · GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP has been assessed and accredited in accordance with the

S.No Discipline / Group Measurand orReferenceMaterial/Type ofinstrument ormaterial to becalibrated ormeasured / QuantityMeasured /Instrument

Measurement rangeand additionalparameters whereapplicable(Rangeand Frequency)

Calibration andMeasurementCapability(CMC)(±)

Calibration orMeasurementMethod orprocedure)

145 THERMAL-TEMPERATURE

Temperature sensor(RTD/Thermocouple)With & without indicator &Temperaturetransmitter

>400 °C to 650 °C 1.70°C Using R- TypeThermocouple withMultifunction calibrator& dry blockTemperature baths andDry bath byComparison Method

146 THERMAL-TEMPERATURE

Temperature sensor(RTD/Thermocouple)With & without indicator &Temperaturetransmitter

>50 °C to 250 °C 0.30°C Using SPRT/PRT with6.5DMM orMultifunction calibratorand Liquid/Dry blockbath By ComparisonMethod

147 THERMAL-TEMPERATURE

Temperature sensor(RTD/Thermocouple)With & without indicator &Temperaturetransmitter

>650 °C to 1000 °C 2.32°C Using R- TypeThermocouple withMultifunction calibrator& dry blockTemperature baths andDry bath byComparison Method

148 THERMAL-TEMPERATURE

Temperature sensor(RTD/Thermocouple)With & without indicator &Temperaturetransmitter

-35 °C to 50 °C 0.25°C Using SPRT/PRT with6.5DMM orMultifunction calibratorand Liquid/Dry blockbath By ComparisonMethod

This is annexure to 'Certificate of Accreditation' and does not require any signature.

Laboratory Name GODREJ & BOYCE MFG. CO. LTD., LAWKIM MOTORS GROUP, PLOT NO. 126,POR GIDC, POR, RAMANGAMDI, VADODARA, GUJARAT, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2205 Page No. : 79 / 79

Validity 17/01/2020 to 17/05/2021 Last Amended on -